Fluke 525B Specifications

Page 1
525B Temperature/
Pressure Calibrator
Superior accuracy and functionality in
an economical benchtop package
Technical Data
The Fluke Calibration 525B Temperature/ Pressure Calibrator gives you a workhorse combina­tion of high accuracy and broad functionality for temperature and pressure instrument calibration. Compact and economical, the 525B has an interface for auto­mated calibration, providing wide workload coverage in instrument shops and calibration labs, as well as in ATE applications.
The 525B is the most accurate Fluke Calibration temperature calibrator, sourc­ing and measuring a complete range of RTDs, thermocouples, and the YSI 400 thermistor. It also measures pressure cover­ing common ranges from 1 inch (6900 Pa) of water up to 10,000 PSI (69 MPa) using the Fluke 700 Series and 525A-P Series Pressure Modules. Plus, the dc voltage and current specifications of the 525B
enable you to calibrate other process calibrators and a wide variety of other instruments with accuracy that rivals any calibrator in its price range.
Key features:
Simulates and measures all
ANSI thermocouples, as well as L and U types, and provides cold junction compensation to enable calibration of a wide variety of thermocouple instru­mentation.
Direct input for storage of
ITS-90 RTD constants.
RTD source uncertainties to
0.03 °C.
Direct measurement of all
Fluke 700 Series and 525A-P Series Pressure Modules cover­ing the most common pressure ranges from 1˝ H to 10,000 PSI.
Converts easily to any pres-
sure unit directly from the front panel or through remote communications.
Sources dc voltage and current
for multifunction workload coverage, enabling calibration of data loggers, strip chart recorders, multimeters, and other industrial instruments.
Supports automation using Fluke
Calibration’s MET/CAL Calibration Management Software or custom automated programs with standard RS-232-C and IEEE-488 interface.
Eight user programmable
setpoints allow quick recall of values for Zero, Span and linearity checks during calibration.
Programmable setpoint dwell
times for automated calibration and troubleshooting.
0
2
®
Plus
Selectable internal or external
CJC (Cold junction compen­sation) provides you with the ability to make exacting thermocouple measurements using remote junctions.
Easy five-way binding posts for
low-loss convenient hookup. Banana plug, screw terminal, spade lug, wraparound terminal, wire pass-through terminal.
NIST-traceable calibration with
data is included.
At home in the cal lab
The 525B is a compact bench instrument well suited for traditional calibration labora­tories with a large temperature workload. Its 0 to 100 V output and 0 to 100 mA current capa­bilities—plus its ability to serve as a highly-accurate pressure standard—make it a versatile performer.
Filling the gap in the instrument shop
In process plant environments, instrument shop technicians responsible for maintaining the equipment pool need to calibrate and repair a wide range of instru­ments, transmitters, field calibra­tors, P/l converters, transducers, and temperature probes. With its ability to simulate almost any temperature sensor, precise pres­sure monitoring capability, and dc voltage and current accuracy rivaling any calibrator in its price range, the 525B fills the gap between lower accuracy hand­held field calibrators and more expensive high-end multifunction calibrators.
Page 2
A good fit for ATE test applications
With precise calibration-grade instrument control enabled via an IEEE bus, the 525B is a hardworking addition to auto­mated test systems in quality and manufacturing test applications requiring capabilities such as voltage and current sourcing, sensor simulation, and tempera­ture monitoring and validation.
Powerful functionality that is easy to use
From start to finish, the 525B is designed for ease of opera­tion. The intuitive front panel design features large keypads and display that help reduce training time and make the 525B comfortable to use even for long periods of time. Plus, you can store frequently used constants for a variety of probes in memory for faster setup on the job.
MET/CAL® Plus automates 525B calibration and docu­mentation management
You can use the 525B with MET/CAL Plus calibration software (V.6.11 or later) to meet the strin­gent documentation and reporting requirements imposed by quality standards such as ISO 9000. MET/CAL Plus is a powerful software environment for creating, editing, testing, and documenting calibration procedures, and for performing automated calibrations.
At the core of MET/CAL Plus is MET/TRACK database management system for test and measurement assets created specifically to manage equipment. It enables you to track the information you need to maintain quality calibrations. And it supports the traceability and record-keeping requirements of modern quality and accreditation standards.
MET/TRACK provides you with a variety of standard report formats prepared with Crystal Reports Professional, which is included with MET/CAL Plus. These report formats cover a wide range of information to help you meet many of your documentation
®
software, a
requirements, including those for ISO 9000 and similar quality standards. They report on forward and reverse traceability, allowing you to easily document traceabil­ity from any asset to the equip­ment calibrated. It even includes measurement uncertainty report­ing to meet the requirements of ISO/IEC 17025.
Pressure modules
An optional set of external pres­sure modules provides pressure measurement capability. The 525B can accept either the Fluke 700 Series or the 525A-P pres­sure modules. Both module types plug directly into the calibrator’s front panel Lemo connector; the 525B firmware auto-detects the type and value of the attached module.
World-class temperature and pressure calibration solutions
In addition to the 525B Tempera­ture/Pressure Calibrator, Fluke Calibration offers a broad range of solutions in temperature and humidity calibration as well as pressure and flow calibration.
Fluke Calibration makes every­thing you need for calibrating temperature sensors including SPRTs, PRTs, thermistors, and thermocouples. The constant temperature baths and dry-block calibrators offer unmatched stabil­ity, and the primary temperature standards are used in national metrology institutes around the world. For precision thermometry, Fluke Calibration thermometer readouts and probes are excep­tionally accurate and easy to use, and the 1620A “DewK” tempera­ture and humidity data logger handles your environmental monitoring needs.
Fluke Calibration also offers a range of high-end products focus­ing on pressure and flow calibra­tors and standards. More than 30 national measurement institutes rely on Fluke Calibration pressure standards to establish their pres­sure references. Fluke Calibration pressure standards, transfer stan­dards, pressure controller/calibra­tors and monitors are recognized
for their advanced features and the realistic specifications that are the hallmark of a true metrology company.
Calibration service, repair, and support
Fluke Calibration provides exten­sive calibration support and service to maximize the value of your calibration investment. Our worldwide network of Calibration Centers has traceability to national standards. Service and support are just a telephone call or fax away. We offer fast, quality repair and calibration services includ­ing a module exchange program, comprehensive training, and full support in setting up your lab.
General specifications
Warm-up time Twice the time since last warmed up, to a
Settling time Less than 5 seconds for all functions and
Standard interface Standard, RS-232; IEEE-488 (GPIB)
Temperature performance
Electromagnetic compatibility
Temperature coefficient
Relative humidity Operating: < 80 % to 30 °C, < 70 % to 40 °C,
Altitude Operating: 3,050 m (10,000 ft) maximum
Safety EN 61010 Second, ANSI/ISA-S82.01-1994;
Analog low isolation
EMC Designed to comply with IEC 61326-1/1997
Line power Line voltage (selectable): 100 V/120 V or
Power consumption
Dimensions Height: 13.3 cm (5.25 in) plus 1.5 cm (0.6 in)
Weight (without options)
maximum of 30 minutes
ranges except as noted
Operating: 0 °C to 50 °C Calibration (tcal): 15 °C to 35 °C Storage: -20 °C to 70 °C
Meets CE requirements, including susceptibility
Temperature coefficient for temperatures outside tcal +5 °C is 10 % of the 90 day speci­fication (or 1 year, as applicable) per °C
< 40 % to 50 °C Storage: < 95 %, non-condensing
Non-operating: 12,200 m (40,000 ft) maxi­mum
CAN/CSA-C22.2 No. 1010.1-92, NRTL 20 V
(EMC)
220 V/240 V Line frequency: 47 Hz to 63 Hz Line voltage variation: ± 10 % about line voltage setting
15 VA maximum
four feet on bottom Width: 0.75 in standard rack width Depth: 47.3 cm (18.6 in) overall
4 kg (9 lb)
2 Fluke Calibration 525B Temperature/Pressure Calibrator
Page 3
General Specifications
Warm up time................................................................Twice the time since last warmed up, to a maximum of 30 minutes.
Settling time..................................................................Less than 5 seconds for all functions and ranges except as noted.
Standard interface.......................................................RS-232, IEEE-488 (GPIB)
Temperature performance
Operating ....................................................................0 °C to 50 °C
Calibration (tcal) ........................................................18 °C to 28 °C
Storage ........................................................................-20 °C to 70 °C
Electromagnetic compatibility..................................CE: Conforms to EN61326
Temperature coefficient.............................................Temperature coefficient for temperatures outside tcal ± 5 °C is
10 % of the 90 day specification (or 1 year if applicable) per °C
Relative humidity
Operating ....................................................................< 80 % to 30 °C, < 70 % to 40 °C, < 40 % to 50 °C
Storage ........................................................................< 95 % noncondensing
Altitude
Operating ....................................................................3,050 m (10,000 ft) maximum
Nonoperating..............................................................12,200 m (40,000 ft) maximum
Safety...............................................................................EN 61010 Second, ANSI/ISA-S82.01-1994,
CAN/CSA-C22.2 No. 1010.1-92, NRTL
Analog low isolation....................................................20 V
Line power
Line Voltage (selectable) ..........................................00 V/120 V or 220 V/240 V
Line Frequency ..........................................................47 to 63 Hz
Line Voltage Variation ..............................................± 10 % about line voltage setting
Power consumption.....................................................15 VA maximum
Dimensions
Height..........................................................................13.3 cm (5.25 in) plus 1.5 cm (0.6 in) four feet on bottom
Width.......................................................................... standard rack width
Depth...........................................................................47.3 cm (18.6 in) overall
Weight (without options)..............................................4 kg (9 lb)
Electrical Specifications
DC Voltage Specifications, Output
V6
V6
V6
DC Current Specifications, Output
Absolute Uncertainty,
Ranges
[1]
±
(ppm of output + µV)
90 days 1 year
0 to 100.000 mV 25 3303 5 + 2 1 µV 10 mA
0 to 1.00000 V 25 10 30 10 4 + 10 10 µV10 mA 0 to 10.0000 V 25 10030100 4 + 100 100 µV 10 mA 0 to 100.000 V 25 1 mV 30 1 mV 5 + 1 mV 1 mV 1 mA
-10 to 75.000 mV 25 3303 5 + 21µV10
Notes: [1] All outputs are positive only. [2] Remote sensing is not provided. Output resistance is < 1.
Ranges
0 to 100.000 mV 1 µ
0 to 1.00000 V 10 µ 0 to 10.0000 V 100 µ 0 to 100.000 V 10 ppm+1 mV 20 mV
0 to 100.000 mA 40 1501 1 μA 12 V 100 mH
Note: [1] All outputs are positive only.
Ranges
[1]
tcal ± 5 °C ± (ppm of output + µA)
Ranges
0 to 100.000 mA 2000 nA 20 μA
5 °C
tcal ±
TC Output
Bandwidth 0.1 to 10 Hz
(ppm of output + µV p-p)
±
Absolute Uncertainty
90 days 1 year
Bandwidth
0.1 to 10 Hz p-p
Stability
24 hours, ±
(ppm of output + µV)
±
1 °C
Noise
Resolution
Noise
Bandwidth 10 Hz to 10 kHz
Maximum
Compliance
Voltage
10 Hz to 10 kHz rms
Resistance Specifications, Output
[1]
Ranges
5 to 400.00 Ω 0.012 0.015 0.001 Ω 1 to 3 mA
5 to 4.0000 kΩ 0.25 0.3 0.01 Ω 100 μA to 1 mA
Notes: [1] Continuously variable from 0 to 4 kΩ.
Absolute Uncertainty
tcal ± 5°C, ±
Ω
Resolution Allowable Current
90 days 1 year
Resolution
V rms)
µV
0 µV
00 µV
Bandwidth
Maximum
[2]
Burden
Maximum
Inductive Load
[2]
[2] For currents lower than shown, the floor adder increases by Floor
For example, a 500 μA stimulus measuring 100 Ω has a floor uncertainty of 0.015 Ω x 1 mA/500 μA = 0.03 Ω.
(new)
= Floor
(old)
x I
min/Iactual
.
Resistance Specifications, Input
[1]
Ranges
5 to 400.00 Ω 20 0.0035 20 0.004 0.001 Ω 1mA
5 to 4.00000 kΩ 20 0.0035 20 0.004 0.01 Ω 0.1 mA
Note: [1] 4-wire mode.
Absolute Uncertainty
tcal ± 5°C, ± (ppm of output + Ω)
90 days 1 year
Resolution Stimulus Current
525B Temperature/Pressure Calibrator Fluke Calibration 3
Page 4
Thermocouple Specification, Output and Input
TC Type
Range (°C)
Minimum Maximum 90 days 1 year
600 °C
B
C
E
J
K
L
N
R
S
800 °C 1000 °C 1550 °C
0 °C 150 °C 650 °C
1000 °C 1800 °C
-270 °C
-100 °C
-25 °C 350 °C 650 °C
-270 °C
-100 °C
-25 °C 120 °C
1000 °C
-270 °C
-100 °C
-25 °C 120 °C
1000 °C
-200 °C
-100 °C 800 °C
-270 °C
-100 °C
-25 °C 120 °C 410 °C
-50 °C 250 °C 400 °C
1000 °C
0 °C
250 °C
1000 °C 1400 °C
-270 °C
T
-150 °C
0 °C
120 °C
U
-200 °C
0 °C
mV -10 to 75.000 mV
Note: [1] Does not include thermocouple wire error.
800 °C
1000 °C
1550 °C 1820 °C
150 °C
650 °C 1000 °C 1800 °C 2316 °C
-100 °C
-25 °C
350 °C
650 °C 1820 °C
-100 °C
-30 °C
150 °C
760 °C 1200 °C
-100 °C
-25 °C
120 °C 1000 °C 1372 °C
-100 °C
800 °C
900 °C
-100 °C
-25 °C
120 °C
410 °C 1300 °C
250 °C
400 °C 1000 °C 1760 °C
250 °C 1000 °C 1400 °C 1750 °C
-150 °C
0 °C 120 °C 400 °C
0 °C 600 °C
Absolute Uncertainty
tcal ± 5 °C, ± (°C)
Output/Input
0.42 °C
0.39 °C
0.40 °C
0.44 °C
0.25 °C
0.21 °C
0.23 °C
0.38 °C
0.63 °C
0.38 °C
0.16 °C
0.14 °C
0.14 °C
0.16 °C
0.20 °C
0.18 °C
0.14 °C
0.14 °C
0.18 °C
0.25 °C
0.19 °C
0.14 °C
0.19 °C
0.30 °C
0.37 °C
0.26 °C
0.17 °C
0.33 °C
0.20 °C
0.16 °C
0.14 °C
0.21 °C
0.58 °C
0.34 °C
0.31 °C
0.30 °C
0.56 °C
0.36 °C
0.30 °C
0.35 °C
0.51 °C
0.18 °C
0.13 °C
0.12 °C
0.56 °C
0.27 °C
[1]
0.46 °C
0.39 °C
0.40 °C
0.45 °C
0.30 °C
0.26 °C
0.31 °C
0.50 °C
0.84 °C
0.50 °C
0.18 °C
0.15 °C
0.16 °C
0.21 °C
0.27 °C
0.20 °C
0.16 °C
0.17 °C
0.23 °C
0.33 °C
0.22 °C
0.16 °C
0.26 °C
0.40 °C
0.37 °C
0.26 °C
0.17 °C
0.40 °C
0.24 °C
0.19 °C
0.18 °C
0.27 °C
0.58 °C
0.35 °C
0.33 °C
0.40 °C
0.56 °C
0.36 °C
0.37 °C
0.46 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
0.56 °C
0.27 °C
4 Fluke Calibration 525B Temperature/Pressure Calibrator
Page 5
RTD and Thermistor Specification, Output
RTD Type
C0
Pt 385, 100
Pt 3926, 100
Pt 3916, 100
Pt 385, 200
Pt 385, 500
Pt 385, 1000
Ni 120, 120
Cu 427, 10
YSI 400 15 °C50 °C0.005 °C0.007 °C
Notes: [1] 2-wire output. [2] Based on MINCO Application Aid No. 18.
Range (°C)
Absolute Uncertainty
tcal ±
5 °C, ± (°C)
[1]
Minimum Maximum 90 days 1 year
-200 °C
-80 °C 0 °C
100 °C 300 °C 400 °C 630 °C
-200 °C
-80 °C 0 °C
100 °C 300 °C 400 °C
-200 °C
-190 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C
-200 °C
-80 °C 0 °C
100 °C
260 °C 300 °C 400 °C 600 °C
-200 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C
-200 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C
-80 °C 0 °C
[2]
100 °C
-100 °C260 °C0.30 °
-80 °C 0 °C
100 °C 300 °C 400 °C 630 °C 800 °C
-80 °C 0 °C
100 °C 300 °C 400 °C 630 °C
-190 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
0 °C
100 °C 260 °C
0.03 °C
0.04 °C
0.04 °C
0.03 °C
0.04 °C
0.04 °C
0.04 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.04 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.04 °C
0.04 °C
0.31 °C
0.32 °C
0.33 °C
0.33 °C
0.36 °C
0.36 °C
0.42 °C
0.42 °C
0.13 °C
0.13 °C
0.13 °C
0.14 °C
0.14 °C
0.15 °C
0.16 °C
0.16 °C
0.06 °C
0.06 °C
0.07 °C
0.07 °C
0.07 °C
0.07 °C
0.08 °C
0.08 °C
0.02 °C
0.02 °C
0.01 °C
0.04 °C
0.05 °C
0.05 °C
0.04 °C
0.04 °C
0.05 °C
0.05 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
0.03 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
0.05 °C
0.38 °C
0.38 °C
0.39 °C
0.39 °C
0.43 °C
0.43 °C
0.50 °C
0.50 °C
0.15 °C
0.15 °C
0.16 °C
0.17 °C
0.17 °C
0.18 °C
0.19 °C
0.19 °C
0.07 °C
0.08 °C
0.08 °C
0.08 °C
0.09 °C
0.09 °C
0.09 °C
0.09 °C
0.02 °C
0.02 °C
0.02 °C .38 °C
525B Temperature/Pressure Calibrator Fluke Calibration 5
Page 6
RTD and Thermistor Specification, Input
RTD Type
C0
Range (°C)
Minimum Maximum 90 days 1 year
-200 °C
-80 °C 0 °C
Pt 385, 100
Pt 3926, 100
Pt 3916, 100
100 °C 300 °C 400 °C 630 °C
-200 °C
-80 °C 0 °C
100 °C 300 °C 400 °C
-200 °C
-190 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C
-200 °C
-80 °C 0 °C
Pt 385, 200
Pt 385, 500
Pt 385, 1000
PtNi 385, 120
(Ni120)
Cu 427, 10
[2]
100 °C 260 °C 300 °C 400 °C 600 °C
-200 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C
-200 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C
-80 °C 0 °C
100 °C
-100 °C260 °C0.067 °C0.069 °C
YSI 400 15 °C50 °C0.005 °C0.007 °C
SPRT -200 °C660 °C0.05 °
Notes: [1] 4-wire mode. Uncertainties listed do not include probe uncertainties. [2] Based on MINCO Application Aid No. 18.
-80 °C 0 °C
100 °C 300 °C 400 °C 630 °C 800 °C
-80 °C 0 °C
100 °C 300 °C 400 °C 630 °C
-190 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
-80 °C 0 °C
100 °C 260 °C 300 °C 400 °C 600 °C 630 °C
0 °C
100 °C 260 °C
Absolute Uncertainty
5 °C, ± (°C)
tcal ±
0.011 °C
0.018 °C
0.018 °C
0.027 °C
0.031 °C
0.042 °C
0.050 °C
0.011 °C
0.014 °C
0.018 °C
0.026 °C
0.031 °C
0.041 °C
0.006 °C
0.011 °C
0.014 °C
0.018 °C
0.025 °C
0.026 °C
0.031 °C
0.040 °C
0.042 °C
0.008 °C
0.012 °C
0.015 °C
0.020 °C
0.050 °C
0.053 °C
0.070 °C
0.071 °C
0.007 °C
0.019 °C
0.023 °C
0.030 °C
0.032 °C
0.037 °C
0.047 °C
0.048 °C
0.011 °C
0.014 °C
0.019 °C
0.025 °C
0.027 °C
0.030 °C
0.041 °C
0.042 °C
0.009 °C
0.010 °C
0.011 °C
[1]
0.012 °C
0.020 °C
0.020 °C
0.030 °C
0.035 °C
0.047 °C
0.057 °C
0.011 °C
0.015 °C
0.019 °C
0.029 °C
0.034 °C
0.046 °C
0.006 °C
0.012 °C
0.015 °C
0.019 °C
0.028 °C
0.029 °C
0.034 °C
0.045 °C
0.047 °C
0.009 °C
0.013 °C
0.017 °C
0.022 °C
0.053 °C
0.057 °C
0.075 °C
0.076 °C
0.008 °C
0.020 °C
0.025 °C
0.033 °C
0.035 °C
0.041 °C
0.052 °C
0.053 °C
0.012 °C
0.015 °C
0.020 °C
0.028 °C
0.030 °C
0.034 °C
0.045 °C
0.047 °C
0.010 °C
0.011 °C
0.012 °C
.06 °C
6 Fluke Calibration 525B Temperature/Pressure Calibrator
Page 7
700 Series Pressure Modules Specifications
Model
Differential Fluke 700P00 1 in. H2O/0.001 0.25 kPa/0.0002 0.300 % Dry Dry 316 SS
Fluke 700P01 10 in. H2O/0.01 2.5 kPa/0.002 0.200 % Dry Dry 316 SS Fluke 700P02 1 psi/0.0001 6900 Pa/0.7 0.150 % Dry Dry 316 SS Fluke 700P22 1 psi/0.0001 6900 Pa/0.7 0.100 % 316 SS Dry 316 SS Fluke 700P03 5 psi/0.0001 34 kPa/0.001 0.050 % Dry Dry 316 SS Fluke 700P23 5 psi/0.0001 34 kPa/0.001 0.025 % 316 SS Dry 316 SS Fluke 700P04 15 psi/0.001 103 kPa/0.01 0.025 % Dry Dry 316 SS Fluke 700P24 15 psi/0.001 103 kPa/0.01 0.025 % 316 SS Dry 316 SS
Gage Fluke 700P05 30 psi/0.001 207 kPa/0.01 0.025 % 316 SS N/A 316 SS
Fluke 700P06 100 psi/0.01 690 kPa/0.07 0.025 % 316 SS N /A 316 SS Fluke 700P27 300 psi/0.01 2070 kPa/0.1 0.025 % 316 SS N /A 316 SS Fluke 700P07 500 psi/0.01 3400 kPa/0.1 0.025 % 316 SS N /A 316 SS Fluke 700P08 1000 psi/0.1 6900 kPa/0.7 0.025 % 316 SS N /A 316 SS Fluke 700P09 1500 psi/0.1 10 M Pa/0.001 0.025 % 316 SS N /A 316 SS
Absolute Fluke 700PA3 5 psi/0.0001 34 kPa/0.001 0.050 % 316 SS N/A 316 SS
Fluke 700PA4 150 psi/0.001 103 kPa/0.001 0.050 % 316 SS N /A 316 SS Fluke 700PA5 30 psi/0.001 207 kPa/0.01 0.050 % 316 SS N/A 316 SS Fluke 700PA6 100 psi/0.01 690 kPa/0.001 0.050 % 316 SS N/A 316 SS
Vacuum Fluke 700PV3 -5 psi/0.0001 -34 kPa/0.001 0.040 % 316 SS Dry 316 SS
Fluke 700PV4 -15 psi/0.001 -103 kPa/0.01 0.040 % 316 SS Dry 316 SS
Dual Fluke 700PD2 ± 1 psi/0.0001 ± 6900 Pa/0.7 0.150 % 316 SS Dry 316 SS
Fluke 700PD3 ± 5 psi/0.0001 ± 34 kPa/0.001 0.040 % 316 SS Dry 316 SS Fluke 700PD4 ± 15 psi/0.001 ± 103 kPa/0.01 0.025 % 316 SS Dry 316 SS Fluke 700PD5 -15/30 psi/0.001 -100/207
Fluke 700PD6 -15/100 psi/0.01 -100/690
Fluke 700PD7 -15/200 psi/0.01 -100/1380
High Fluke 700P29 3000 psi/0.1 20.7 M Pa/0.001 0.050 % C276 N /A C276
Fluke 700P30 5000 psi/0.1 34 M Pa/0.001 0.050 % C276 N/A C276
1
Total uncertainty, one year for temperature range O °C to +5O °C. Total uncertainty, 1.0% of full span for temperature range -10 °C to O °C. For POO module only, compensated temperature range is 15 °C to 35 °C. 2 “Dry” indicates dry air or non-corrosive gas as compatible media. “316SS” indicates media compatible with Type 316 Stainless Steel. “C276” indicates media compatible with Hastelloy C276. Use of pressure zero is required prior to measurement or source. Max overpressure specification includes common mode pressure. Modules are CE rated. Metric adapter(s}: 1/4” NPr female to male BSP/ISO 1/4­19, tapered thread, included with all modules except P29, P3O, and P31, all modules include a NIST traceable certificate and test data.
Fluke 700P31 10000 psi/1 69 M Pa/0.007 0.050 % C276 N/A C276
Range/
Resolution
Range (approx)/
Resolution
kPa/0.01
kPa/0.07
kPa/0.1
Reference uncertainty (23 ± 3 °C) High
1
2
side media Low2 side media Fitting material
0.025 % 316 SS N/A 316 SS
0.025 % 316 SS N/A 316 SS
0.040 % 316 SS N/A 316 SS
525A-P Series Precision Pressure Transducers
Type Model Range/Resolution Range/Resolution Reference Uncertainty (23 ± 3 °C)
Differential 525A-P02 1 psi/0.00001 6900 Pa/0.01 0.008 % FS
Gage 525A-P03 5 psi/0.00001 34 kPa/0.001 0.008 % FS
Gage 525A-P04 15 psi/0.001 103 kPa/0.001 0.008 % FS
Gage 525A-P05 30 psi/0.0001 207 kPa/0.001 0.008 % FS
Gage 525A-P06 100 psi/0.001 690 kPa/0.001 0.008 % FS
Gage 525A-P07 500 psi/0.001 3400 kPa/0.01 0.008 % FS
Gage 525A-P08 1000 psi/0.01 6900 kPa/0.01 0.008 % FS
Gage 525A-P29 3000 psi/0.01 20.7 M Pa/0.0001 0.008 % FS
Absolute 525A-PA4 15 psi/0.0001 103 kPa/0.001 0.008 % FS
Absolute 525A-PA5 30 psi/0.0001 207 kPa/0.001 0.008 % FS
Absolute 525A-PA6 100 psi/0.001 690 kPa/0.001 0.008 % FS
Absolute 525A-PA7 500 psi/0.001 3400 kPa/0.01 0.008 % FS
Absolute 525A-PA8 1000 psi/0.01 6900 kPa/0.01 0.008 % FS
Vacuum 525A-PV4 -15 TO 0 psi/0.0001 -34 kPa/0.001 0.008 % FS
525B Temperature/Pressure Calibrator Fluke Calibration 7
Page 8
Pressure Measurement
The Calibrator can accept either the Fluke 700 or 525A-P series pressure modules. Pressure modules plug directly into the front panel Lemo connector with the Calibrator firmware autodetecting the type and value of the module you are attaching.
Range ............................................Determined by pressure module
Accuracy/Resolution.................Determined by pressure module
Units
PSI ..............................................pounds per square inch
inH2O4°C ..................................inches of water at 4 degrees Celsius
inH2O20°C................................inches of water at 20 degrees Celsius
cmH2O4°C.................................centimeters of water at 4 degrees Celsius
cmH2O20°C ..............................centimeters of water at 20 degrees Celsius
BAR ............................................bars
mBAR.........................................millibars
KPAL ..........................................kilopascals
inHG 0°C ...................................inches of mercury at 0 degrees Celsius
mmHG 0°C ................................millimeters of mercury at 0 degrees Celsius
Kg/cm2......................................kilograms per square centimeter
Model Description
525B Temperature/Pressure Calibrator
Software
MET/CAL Plus Automated Calibration Software (IEEE-488 and RS-232)
Accessories
5520A-525A/ Leads
Y525 19” Rack Mount Kit Fluke 700PD3 Dual Pressure Module ± 5 psi / ± 34 kPa Fluke 700P31 High Pressure Module 10000 psi/69 M Pa Fluke 700P22 Differential Pressure Module 1 psi / 6900 Pa Fluke 700P24 Differential Pressure Module 15 psi / 103 kPa Fluke 700PD2 Dual Pressure Module ± 1 psi / ± 6900 Pa Fluke 700P08 Gage Pressure Module 1000 psi / 6900 kPa Fluke 700PA4 Absolute Pressure Module 15 psi / 103 kPa Fluke 700P30 High Pressure Module 5000 psi / 34 M Pa Fluke 700PD7 Dual Pressure Module -15/200 psi / -100/1380 kPa Fluke 700P05 Gage Pressure Module 30 psi / 207 kPa Fluke 700P06 Gage Pressure Module 100 psi / 690 kPa Fluke 700P29 High Pressure Module 3000 psi / 20.7 M Pa Fluke 700PD5 Dual Pressure Module -15 / 30 psi / -100 / 207 kPa Fluke 700PV3 Vacuum Pressure Module -5 psi / -34 kPa Fluke 700P23 Differential Pressure Module 5 psi / 34 kPa Fluke 700PA5 Absolute Pressure Module 30 psi / 207 kPa
Model Description
Accessories continued
Thermocouple and test lead set 1 meter TC wire with male mini-plugs, All ANSI, L and U types; TC shorting plug; 1, 80AK-8001 TC adapter; 1, bead thermocouple type K; 1 meter shielded test lead with double banana plugs (M); 2, 30 A test leads, banana type connectors
Fluke 700P03 Differential Pressure Module 5 psi / 34 kPa Fluke 700PD6 Dual Pressure Module -15/100 psi / -100/690 kPa Fluke 700P04 Differential Pressure Module 15 psi / 103 kPa Fluke 700PA6 Absolute Pressure Module 100 psi / 6900 kPa Fluke 700P01 Differential Pressure Module 10.00 in. H2O / 2.5 kPa Fluke 700PD4 Dual Pressure Module ± 15 psi / ± 103 kPa 525A-P02 Differential 1 psi/6900 Pa 525A-P03 Gage 5 psi/34 kPa 525A-P04 Gage 15 psi/103 kPa 525A-P05 Gage 30 psi/207 kPa 525A-P06 Gage 100 psi/690 kPa 525A-P07 Gage 500 psi/3400 kPa 525A-P08 Gage 1000 psi/6900 kPa 525A-P29 Gage 3000 psi/20.7 M Pa 525A-PA4 Absolute 15 psi/103 kPa 525A-PA5 Absolute 30 psi/207 kPa 525A-PA6 Absolute 100 psi/690 kPa 525A-PA7 Absolute 500 psi/3400 kPa 525A-PA8 Absolute 1000 psi/6900 kPa 525A-PV4 Vacuum -15 to 0 psi/-34 kPa
Support
MET/SUPPORT Gold
MET/SUPPORT GLDNW
Software Support Program, one-year membership
MET/SUPPORT Gold Program, up to four workstations, one-year membership
8 Fluke Calibration 525B Temperature/Pressure Calibrator
Fluke Calibration. Precision, performance, confidence.
Fluke Corporation PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands
©2001-2012 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 6/2012 1611141D_EN Pub_ID: 10884-eng Rev 02
For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1(425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
Loading...