Fluke 320 Series, 330 Series Technical Data

Fluke 320 and 330 Series Clamp Meters
Narrow body and jaw fit into small spaces
Backlight available on most models
Accurate low current measurements
Inrush current feature on most models
Auto shut-off
New Inrush Feature
A new innovation from Fluke! Fluke Clamp Meters with the inrush feature measure motor starting current like a circuit breaker or overload unit sees
. One button lets you troubleshoot current surges
it that can cause nuisance tripping. Fluke’s inrush feature gives you a synchronized reading of the current draw profile from motor start up
or more information ab
F
feature go to www
Inrush, A T
e
.fluke.com/330 to view
gy Brief
chnolo
out Fluke
.
.
’s new inrush
Choose the model that fits your work.
Looking for a clamp meter that’s right for you? Match the functions and features listed below to the Fluke Clamp Meter best suited for your job.
321 322
AC Amps AC Amps A Ohms Ohms AC Volts AC Volts A Auto-Off Auto-Off Hold Hold Compact Compact
DC Volts
Measure ac current 400.0 A
Measure ac voltage
Measure continuity 30 Ω≤30
Measure dc voltage
Higher resolution for loads below 40 amperes
Measure resistance 400.0 400.0 Measure inrush Work in poorly lit areas Measure non-linear loads Measure dc current Variable speed drives and stand-by power
400.0 V 400.0 V 600 V 600 V
40.00 A
400.0 A
400.0 V 600 V
0 - 40.00A
40.0 - 400.0A
333 334 335 336 337
AC Amps AC Amps AC Amps AC Amps AC Amps Ohms Ohms Ohms Ohms Ohms AC Volts AC Volts AC Volts AC Volts AC Volts Auto-Off Auto-Off Auto-Off Auto-Off Auto-Off Hold Hold Hold Hold Hold Full Size Full Size Full Size Full Size Full Size DC Volts DC Volts DC Volts DC Volts DC Volts
In-rush In-rush In-rush In-rush Backlight Backlight Backlight Backlight
True-rms True-rms True-rms
DC Amps DC Amps
Frequency Large Jaw MIN/MAX
400.0 A 600.0 A 600.0 A 600.0 A 999.9 A
600.0 V 600.0 V 600.0 V 600.0 V 600.0 V
30 Ω≤30 Ω≤30 Ω≤30 Ω≤30
600.0 V 600.0 V 600.0 V 600.0 V 600.0 V
600.0 6000 6000 6000 6000 100 mS 100 mS 100 mS 100 mS Backlight Backlight Backlight Backlight
True-rms True-rms True-rms
600.0 A 999.9 A
400.0 Hz
600V
CAT III
Designed and  Conforms to
IEC 1010-2-031
Specifications (@23 ºC + 5 ºC RH 0 - 90%)
Fluke 321 Fluke 322 Fluke 333 Fluke 334 Fluke 335 Fluke 336 Fluke 337
Range 0 - 400.0 A 0 - 40.00 A 0 - 400.0 A 0 - 600.0 A 0 - 600.0 A 0 - 600.0 A 0 - 999.9 A
Accuracy 1.8% ± 5 counts 1.8% ± 5 counts 2% ± 5 counts 2% ± 5 counts 2% ± 5 counts 2% ± 5 counts 2% ± 5 counts
(50 - 60 Hz) (50 - 60 Hz) (50 - 60 Hz) (50 - 60 Hz) (50 - 60 Hz) (10 - 100 Hz) (10 - 100 Hz)
3.0% ± 5 counts 3.0% ± 5 counts 6% ± 5 counts 6% ± 5 counts (60 Hz - 400 Hz) (60 Hz - 400 Hz) (100 - 400 Hz) (100 - 400 Hz)
Crest Factor 2.4 @ 500 A 3 @ 500 A 3 @ 500 A (50/60 Hz) 2.0 @ 600 A 2.5 @ 600 A 2.5 @ 600 A
AC Response Averaging Averaging Averaging Averaging True-rms True-rms True-rms
Inrush Integration Time 100 mS 100 mS 100 mS 100 mS
Range 0 - 600.0 A 0 - 999.9 A Accuracy 2% + 3 counts 2% + 3 counts Range 0 - 400.0 V 0 - 400.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V
0 - 600 V 0 - 600 V
Accuracy 1.2% + 5 counts 1.2% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts
(50 - 400 Hz) (50 - 400 Hz) (50/60 Hz) (50/60 Hz) (50/60 Hz) (20 - 100 Hz) (20 - 100 Hz)
AC Response Averaging Averaging Averaging Averaging True-rms True-rms True-rms Range 0 - 400.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V 0 - 600.0 V
Accuracy 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts 1% + 5 counts Range 0 - 400.0 0 - 400.0 0 - 600.0 0 - 600.0 0 - 600.0 0 - 600.0 0 - 600.0
Accuracy 1% ± 5 counts 1% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts 1.5% ± 5 counts
Continuity 30 Ω≤30 Ω≤30 Ω≤30 Ω≤30 Ω≤30 Ω≤30
Range 5.0 - 400.0 Hz Accuracy 0.5% ± 5 counts Trigger Level 10 - 100 Hz 5 A
MIN/MAX NoNoNoNoNoNoYes
Backlight No No No No Yes Yes Yes
Display Hold Yes Yes Yes Yes Yes Yes Yes
Height 7.5” 7.5” 9.375” 9.375” 9.375” 9.875” 9.875”
Size Width 2.5” 2.5” 3.125” 3.125” 3.125” 3.125” 3.125”
Depth 1.4” 1.4” 1.625” 1.625” 1.625” 1.625” 1.625” Jaw Opening 1” (25.4 mm) 1” (25.4 mm) 1.2” (30.5 mm) 1.2” (30.5 mm) 1.2” (30.5 mm) 1.2” (30.5 mm) 1.2” (30.5 mm)
Max. Wire Size 500 MCM 500 MCM 750 MCM 750 MCM 750 MCM 750 MCM or 750 MCM or
Weight 8 oz. 8 oz. 11 oz. 11 oz. 11 oz. 11 oz. 11 oz.
40.0 - 400.0 A
0 - 600 V
add 2% for C.F.>2 add 2% for C.F.>2 1.42 @ 1000 A
6% + 5 counts 6% + 5 counts (100 - 400 Hz) (100 - 400 Hz)
601 - 6000 601 - 6000 601 - 6000 601 - 6000
two 500 MCM two 500 MCM
add 2% for C.F.>2
5 - 10 Hz,
100 - 400 Hz 10 A
Models 321 and 322
Operating temperature
-14 °F to 122 °F (-10 °C to 50 °C)
Altitude
2,000 meters
Battery life
(2) AAA carbon zinc approxi­mately 100 hours continuous, automatic low battery indicator
Safety rating
IEC 1010-2-031, 600 V CAT III
Temperature de-rating
< 18 °C, > 28 °C add 0.1 x specified accuracy / °C
Storage temperature
-22 °F to 140 °F (-30 °C to 60 °C)
Warranty
Two years
Models 333, 334, 335, 336 and 337
Operating temperature
-14 °F to 122 °F (-10 °C to 50 °C)
Altitude
2,500 meters
Battery life
(2) AA alkaline approximately 150 hours continuous, auto­matic low battery indicator
Safety rating
IEC 1010-2-031, 600 V CAT III
Temperature de-rating
Add 0.1 x specified accuracy for each ºC above 28 ºC or below 18 ºC
Storage temperature
-40 ºF to 140 ºF (-40 ºC to 60 ºC)
Warranty
Three years