Fluke 3125 Operating Manual

Page 1
Model 3125
Detachable Hot Plate
Users’ Manual
Rev. 8C1501
Page 2
© Copyright,1998 All rights reserved Hart Scientific, Inc.
799 E. Utah Valley Drive American Fork, Utah 84003-9775 Telephone (801) 763-1600 • Fax (801) 763-1010 Internet: http://www.hartscientific.com
Page 3
WARNING
The plate gets extremely hot. DO NOT TOUCH.
DO NOT TOUCH the metal parts on the plate.
WARNING
To ensure the safety of operating personnel, and to avoid damage to this equipment:
DO NOT connect this unit to any other instrument or outlet other than the 2200 controller.
WARNING
HIGH VOLTAGE
is used in the operation of this equipment.
SEVERE INJURY OR DEATH
may result if personnel fail to observe safety precautions.
Before working inside the equipment, turn power off and disconnect power cord.
WARNING
HIGH TEMPERATURE MAY BE PRESENT
in this equipment.
FIRES AND SEVERE BURNS
may result if personnel fail to observe safety precautions.
a
Page 4
WARNING
To ensure the safety of personnel, and to avoid damage to equipment:
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the user’s manual.
Continuous use of this equipment at high temperatures for extended periods of time requires caution.
Completely unattended high temperature operation is not recommended for safety reasons.
Components and heater lifetimes can be shortened by continuous high temperature operation.
Follow all safety guidelines listed in the user’s manual.
WARNING
The 2200 Controller and the 3125 Detachable Hot Plate are a matched set.
DO NOT mix and match sets.
WARNING
CALIBRATION EQUIPMENT SHOULD ONLY BE USED BY TRAINED PERSONNEL.
b
Page 5
Table of Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . 1
2 Specifications and Environmental Conditions. . . . 3
2.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Environmental Conditions . . . . . . . . . . . . . . . . . 3
2.3 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Safety Guidelines . . . . . . . . . . . . . . . . . . . 5
4 Troubleshooting . . . . . . . . . . . . . . . . . . . . 7
i
Page 6
1 Introduction
1 Introduction
Hart Scientific’s Model 3125 Detachable Hot Plate is controlled by Hart Scien
­tific’s 2200 controller and uses a precision platinum RTD as a sensor with a heater to control the temperature. The 3125 and 2200 are a matched set. Both of them have the same serial number.
The controller display shows the temperature and also the set-point tempera
­ture. The temperature may be set to any temperature within the range of 35°C to 400°C in 0.01°C or °F increments by using the buttons on the control panel. The controller’s multiple fault protection devices insure user and instrument safety and protection.
The Detachable Hot Plate (see Figure 1) consists of a controlled plate made of aluminum. A heater is attached to the bottom with an RTD temperature sensor used to control it. The housing consists of a top cover and a base, which serve as a clamp and strain relief point for the wiring. The Reference Well is avail
­able for use with a 3/16-inch diameter probe, which may be used to check the plate temperature. This well is used to calibrate the system.
1150/2300W
10 A
115/230V
115V
1
50/60 Hz
HEATER
2
3
PROBE
4
RS-232
115/230VAC 50/60 Hz
2200 Controller Back Panel
3125 Hot Plate Assembly
Figure 1 Controller and Hotplate Assembly
3125 Manual Rev. 8C1501 1
Page 7
1 Introduction
2 Manual Rev. 8C1501 Hart Scientific
Page 8
2 Specifications and Environmental Conditions
2 Specifications and Environmental
Conditions
2.1 Specifications
Operating Range 35°C to 400°C (95°F to 752°F) Accuracy ±0.5°C to 200°C
±1.0°C to 400°C Traceable to NIST
Resolution 0.01°C or °F Readout °C or °F switchable Stability ±0.2°C to 200°C
±0.3°C to 400°C
Controller Hart Model 2200 Microprocessor based with
RS232 serial port
Sensor RTD Heater 325 Watt, solid state controlled Plate (accessible) 3.8 inch diameter Weight 4.5 lb. (2.1 kg)
2.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. The instrument oper ates safely under the following conditions:
temperature range: 5-40°C (41-104°F)
ambient relative humidity: 15-50%
pressure: 75kPa-106kPa
mains voltage within ±10% of nominal
vibrations in the calibration environment should be minimized
altitude does not effect the performance or safety of the unit
-
-
2.3 Warranty
Hart Scientific, Inc. (Hart) warrants this product to be free from defects in ma terial and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in
3125 Manual Rev. 8C1501 3
-
Page 9
2 Specifications and Environmental Conditions
Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc tions on appropriate Hart products. It is not warranted to be error free. Hart’s obligation under this warranty is limited to repair or replacement of a product, which is returned to Hart within the warranty period and is determined, upon examination by Hart, to be defective. If Hart determines that the defect or mal-function has been caused by misuse, alteration, abuse or abnormal condi tions or operation or handling, Hart will repair the product and bill the pur chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing Hart for authorization. Hart assumes NO risk for in-transit damage. For service or assistance, please contact the manufacturer.
Hart Scientific, Inc. 799 East Utah Valley Drive American Fork, UT 84003-9775 Phone: (801) 763-1600 Fax: (801) 763-1010 E-mail: support@hartscientific.com
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, IN­CIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
-
-
-
4 Manual Rev. 8C1501 Hart Scientific
Page 10
3 Safety Guidelines
Operate the instrument in room temperatures between 5-50°C
(41-122°F). Allow sufficient air circulation by leaving at least 15 cm (6 inches) of space between the instrument and nearby objects. Overhead clearance 45 cm (18 inches) to allow for safe and easy placement and re moval of sensors for calibration.
The hot plate is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with care. The instrument should not be operated in excessively dusty or dirty environments or near liquids thatmay present an electrical shock hazard. Do not operate near flammable materials.
DO NOT use fluids to clean off the surface.
The instrument generates extreme temperatures. Precautions must be
takentoprevent personalinjuryordamage toobjects.Sensorsmay beex tremely hot when removed from the instrument. Cautiously handle sensors to prevent personal injury. Carefully place sensors on a heat re sistant surface or rack until they are at room temperature. Allow the sur face to cool before transporting the instrument.
3 Safety Guidelines
-
-
-
-
Never introduce any foreign material onto the surface plate. Fluids, etc.
can leak into the instrument causing damage.
Theunit is fused throughthe2200 Controller.Refertothe2200Controller
manual for fuse information.
3125 Manual Rev. 8C1501 5
Page 11
3 Safety Guidelines
6 Manual Rev. 8C1501 Hart Scientific
Page 12
4 Troubleshooting
4 Troubleshooting
If problems arise while operating the 3125, this section provides some sugges tions that may help you solve the problem. Opening the unit without contacting Hart Scientific Customer Service may void the warranty.
Unit will not heat
Check the power cord to see if it is connected to the 2200 controller.
Out of calibration
Ensure that the 2200 and 3125 are a matched set. Check the serial num
bertothe hotplateon boththehotplate andthe2200 controller.These two serial numbers must match.
CheckthatthecalibrationparametersR0,Alpha,andDeltamatchtheRe
port of Calibration.
Fuses blow
Make sure the correct fuse rating and type are being used.
Onlyoperatethehotplateatthevoltagespecifiedontheserialnumberla
bel.
Unit is unstable
Check that the proportional band on the controller matches the Report of
Calibration.
Make sure ambient conditions are stable
-
-
-
-
3125 Manual Rev. 8C1501 7