Fluke 2700G-G70M/C, 2700G-G20M/C, 2700G-BG2M/C, 2700G-BG700K/C, 2700G-BG200K/C User manual

...
Page 1
2700G Series
Reference Pressure Gauge
Users Manual
September 2012 © 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption. Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country. Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point). THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
11/99
uke assumes no risk
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 3

Table of Contents

Title Page
Introduction .................................................................................................................... 1
How to Contact Fluke Calibration ................................................................................... 1
Standard Equipment....................................................................................................... 2
Safety Information .......................................................................................................... 2
Special Conditions for Safe Use ................................................................................ 3
Symbols ..................................................................................................................... 4
Display and Buttons ....................................................................................................... 5
Operation ....................................................................................................................... 6
How to Setup the Product .......................................................................................... 6
Engineering Units ................................................................................................. 6
Set Auto Off .......................................................................................................... 7
Show Battery Voltage ........................................................................................... 7
Display Actual Temperature ................................................................................. 7
Set Damping ......................................................................................................... 7
i
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 4
2700G Series
Users Manual
Set Sample Rate .................................................................................................. 7
Set Tare ............................................................................................................... 7
Function Lock ....................................................................................................... 8
Supervisory Mode ..................................................................................................... 8
Available Pressure Ranges ....................................................................................... 8
How to Set a Custom Engineering Unit or Scale....................................................... 9
Battery Life .................................................................................................................... 9
Maintenance .................................................................................................................. 9
How to Clean the Product ......................................................................................... 9
How to Change the Batteries .................................................................................... 10
Accessories ................................................................................................................... 11
USB Interface Cable ................................................................................................. 11
Power Module ........................................................................................................... 11
Serial Interface Instructions ........................................................................................... 13
Initiating Communication ........................................................................................... 13
List of Commands ..................................................................................................... 13
Parameter Units ............................................................................................................. 14
Error Codes ................................................................................................................... 15
Units Conversion ........................................................................................................... 16
Specifications ................................................................................................................ 18
Accuracy ................................................................................................................... 18
Media Compatibility ................................................................................................... 18
Environmental ........................................................................................................... 18
Mechanical Specifications ......................................................................................... 19
Available Pressure Ranges ....................................................................................... 20
ii
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 5

List of Tables

Table Title Page
1. Symbols ................................................................................................................................. 4
2. Display and Buttons .............................................................................................................. 5
3. Units Conversion ................................................................................................................... 16
iii
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 6

List of Figures

Figure Title Page
1. The Product ........................................................................................................................... 5
2. Change the Batteries ............................................................................................................. 10
3. USB Power Module and Universal Connectors ..................................................................... 12
v
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 7

Introduction

The 2700G Series Reference Pressure Gauges (the Product) are high-accuracy digital pressure test gauges. Accurate to 0.02 % FS, the Product can be used as a calibration reference, or in applications where high­accuracy pressure measurement is required.
The Product features user-configurable functions that include:
Sampling rate
Tare
Damping
Auto off
Min Max
When the Product is configured, you can lock its settings and use password protection to prevent configuration changes. See the “Supervisory Mode” section.
1
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 8
2700G Series
Users Manual

Standard Equipment

The Product ships with:
Protective Cover (installed)
Three AA Alkaline Batteries (installed)
Safety Information (printed)
Report of Calibration
Manuals CD-ROM with translated Users Manuals
USB Cable
USB Power Adapter
NPT to ¼ BSP Male Adapter
NPT to M20 x 1.5 Male Adapter

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
XWWarning
To prevent injury, only assemble and operate high-pressure systems if you know the correct safety procedures. High-pressure liquids and gases are hazardous and the energy from them can be released without warning.
To prevent possible electrical shock, fire, or personal injury:
Read all safety Information before you
use the Product.
Use the Product only as specified, or the
protection supplied by the Product can be compromised.
Do not use the Product around explosive
gas, vapor, or in damp or wet environments.
Do not use and disable the Product if it
is damaged.
Remove the batteries if the Product is
not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 9
Reference Pressure Gauge
Safety Information
Replace the batteries when the low
battery indicator shows to prevent incorrect measurements.
The battery door must be closed and
locked before you operate the Product.
WCaution
To prevent possible damage to Product or to equipment under test:
The display reads “OL” when the
pressure source is above the Product range limit. The pressure source must immediately be removed.
Do not apply more than the maximum
torque specified. Maximum torque specified is 20 Nm = 15 ft-lb.

Special Conditions for Safe Use

Misuse
If the Product is exposed to overpressure or sudden physical shock (such as being dropped) examine it for damage that can cause a safety issue. If necessary, send the Product for examination to Fluke. Refer to the How to Contact Fluke Calibration section.
WWarning
To prevent possible fire, or personal injury:
Do not use the Product with flammable
substances.
The Product is intended for installation
only in locations with protection against the entry of solid unwanted objects or water that can compromise safety.
3
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 10
2700G Series
Users Manual

Symbols

Symbols used on the Product and in this manual are in Table 1.
Table 1. Symbols
Symbol Meaning Symbol Meaning
W
Risk of danger. Important information. See manual.
X Hazardous voltage. Risk of electrical shock. )
Conforms to relevant Australian standards. ~
4
P Conforms to European Union directives.
Conforms to relevant North American Safety Standards.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 11
Reference Pressure Gauge
Display and Buttons

Display and Buttons

The Display and Buttons are shown in Figure 1. The Buttons are in Table 2.
2
1
Figure 1. The Product
3
4
gsn001.eps
Table 2. Display and Buttons
Item Function
Push to turn on the Product. Push again to turn off the Product.
Zeros the display. In Configure Mode, push the button to move forward through the menus.
MIN MAX records minimum and maximum pressure values and saves them to memory. Push to show maximum (MAX) indication. Push again to show minimum (MIN) indication. After 2 seconds, the gauge goes back to live operation.
To clear the MIN MAX memory values, push and hold for 2 seconds until CLr is shown.
In Configure Mode, push  () to move backward through the menus.
5
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 12
2700G Series
Users Manual
Table 2. Display and Buttons (cont.)
Item Function
Push to go to setup and configuration menus.
Push to make a selection. When the Product is not in Configuration mode, push to turn on the backlight. Push again to turn off the backlight.
NPT Connector
Pressure Display
Engineering Units
Bargraph

Operation

The subsequent sections tell you how to operate the Product. Push to turn on the Product.
The analog bar graph at the bottom of the display shows the applied-pressure level relative to the full range of the gauge.
Note
If you record a Tare value, the pressure shown is not the actual pressure applied.

How to Setup the Product

Before you use the product, it is necessary to configure it for your application. Push to go to the Setup menu.
Each time is pushed, the display goes to the subsequent function. Push or to change the parameter value. When a parameter is set, push to exit the configuration menu or to move to the next parameter.

Engineering Units

The Product’s default engineering unit shows psi. To change this, push and to move through the standard engineering units plus one custom unit/scale. When the necessary unit shows, push or . Pressure now shows in the chosen engineering units. See the Specifications section for a list of available
6
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 13
Reference Pressure Gauge
Operation
engineering units. See the Supervisory Mode section for instructions to set up custom units.

Set Auto Off

Auto Off can be set in 1 minute increments from 1 to 30 minutes or you can turn off the function for continuous Product operation. The Product is configured for 30 minutes. Push and to set the necessary interval. The “off” position is at the low end of the selections, less than 1 minute.

Show Battery Voltage

Actual battery voltage and a percent-of-life bargraph show the battery charge. No adjustments are made in this parameter.

Display Actual Temperature

The Product is temperature compensated This parameter shows the temperature measured by the internal sensor. Push or to show degrees F or C.

Set Damping

Selections are “on” and “off” . Damping smooths readings from pulsating pressure sources.

Set Sample Rate

This function finds how many times pressure is sampled and the display is updated. Selections are 0.5, 1, 3, and 10 samples/second. Note that 10/second gives the fastest response time.

Set Tare

Use this function to set a constant offset value that is then subtracted from the measured pressure. For example, if a tare is set at 30 psi, and the measured pressure is 37 psi, 7 psi is shown.
A pressure of 27 psi is shown as -3 psi.
Push and to set the tare value. The value is in relation to the engineering units and resolution selected for display. Tare value can be set to the maximum range of the gauge.
For safety, the bar graph always shows the actual pressure in relation to the full range of the gauge regardless of the tare position. This is done to make sure that with a “0” reading pressure is being applied to the gauge.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 14
2700G Series
Users Manual

Function Lock

When set, access to each of the settable parameters above can be turned “off” to prevent unauthorized configuration changes. This is done with password protection in Supervisory Mode. Push to access Supervisory Mode or to go back to normal operation.

Supervisory Mode

If necessary, each user-configurable parameter can be edited when you receive the Product. Some parameters are locked and must be unlocked to configure them. Use Supervisory mode to do this.
When you are in the Configure menu, and FUnC LOCK is shown, it means that there are locked parameters.
To disable function lock:
1. Push  . 0 PWRD is then shown.
2. The password “101” is required to unlock
Supervisory mode. Push to put in the password entry. Hold or down to move faster through the selections by a factor of 10. When you stop the counter, push and again to move forward or backward by a factor of 1. The password is factory set and cannot be changed.
3. Push .
From this point each parameter can be locked or unlocked. Push and to select UnLOC or LOC for
each parameter. To move to the subsequent parameter, push .
You can access, lock, or unlock these functions:
Zero function (enable/disable)
Set pressure units (enable/disable)
Auto shutdown adjustment (enable/disable)
Damping settings (enable/disable)
Sample rate setting (enable/disable)
Tare setting (enable/disable)
Custom engineering units (set scale factor)
When a function is locked, it cannot be accessed or changed from its current condition until you go to Supervisory Mode and unlock it.

Available Pressure Ranges

Available pressure ranges are shown in the Specifications section.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 15
Reference Pressure Gauge
Battery Life

How to Set a Custom Engineering Unit or Scale

The last menu selection in Supervisory Mode is SET FACTR. You can set a multiplier factor from 0.001 to 100 to make a custom scale. The set factor is multiplied by the psi measured and the result is shown.
Example: 40 psi is the equivalent of 1000 lbs of product in a tank. It is necessary to show the product weight with a 100 psi gauge. If you set a factor of 25, 40 psi pressure shows as 1000 (40 x 25). The engineering unit shown is Cust (custom).

Battery Life

Battery life is approximately 75 hours of operation with the backlight off. When the battery voltage is low, the low­battery icon () shows on the top left of the display. To replace the batteries, see the How to Change the Batteries section.

Maintenance

How to Clean the Product

Clean the Product with a soft cloth dampened with water or water and weak soap.
WCaution
To prevent possible damage to the Product, do not use solvents or abrasive cleansers.
WCaution
For safe operation and maintenance of the product:
Repair the Product before use if the
battery leaks.
Remove batteries to prevent battery
leakage and damage to the Product if it is not used for an extended period.
Be sure that the battery polarity is
correct to prevent battery leakage.
Have an approved technician repair the
Product.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 16
2700G Series
Users Manual

How to Change the Batteries

XWWarning
To prevent possible electrical shock, fire, or personal injury, have an approved technician repair the Product.
To change the batteries, see Figure 2:
1. Pull off the Product cover.
2. Use a Phillips screwdriver to loosen the captive screw on the battery door.
3. Remove the battery door.
4. Replace the three AA batteries.
5. Install the battery door again.
6. Tighten the captive screw on the battery door.
7. Return the Product to its Holster.
gsn002.eps
Figure 2. Change the Batteries
10
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 17
Reference Pressure Gauge
Accessories

Accessories

USB Interface Cable

The Product includes an USB interface cable. The input jack is on the back of the Product. You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC. For specifications on the interface, see the Specifications section.

Power Module

The Product includes a USB power module. See Figure 3.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 18
2700G Series
Users Manual
gsn003.eps
Figure 3. USB Power Module and Universal Connectors
12
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 19
Reference Pressure Gauge
Serial Interface Instructions

Serial Interface Instructions

The subsequent sections tell you how to set up the Product for serial communication.

Initiating Communication

Terminal communication can be setup with terminal communication Software on a PC. The terminal must be set as follows:
Bits per second: 9600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: None
Local echo on

List of Commands

Use the subsequent list of commands to communicate with the Gauge:
*CLS Clears the error queue
FAULT? Returns an error code from the
error queue
*IDN? Identification query. Returns the
manufacturer, model number, and firmware revision level of the Calibrator.
TARE Tares the offset pressure of the
reading on the calibrator
TARE? Returns the current tare value
PRES_UNIT? Returns the pressure unit for the
upper display
PRES_UNIT Sets the pressure unit for the
display
ZERO_MEAS Zeros pressure of the calibrator
ZERO_MEAS? Returns the current zero offset
value
MINMAX_RST Resets the minimum and
maximum recorded values
13
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 20
2700G Series
Users Manual
MIN? Returns the minimum recorded
value
MAX? Returns the maximum recorded
value
TEMP? Returns temperature in the chosen
units
CAL_STORE Stores calibration data.
CUST_MULT? Sets the multiplier for the custom
unit type
STREAM_OFF Turns off streaming data
STREAM_ON Turns on streaming data
VAL? Returns the measured pressure
value in selected units
TEMP_UNIT Used to set temperature unit
TEMP_UNIT? Returns temperature unit

Parameter Units

The subsequent list of units is used with the Product:
psi Pressure in pounds per
square inch
bar Pressure in bars
mBar Pressure in millibars
kg/cm2 Pressure in kilograms per
centimeter squared
inH2O4C Pressure in inches of water at 4 °C
inH2O20C Pressure in inches of water at
20 °C
inH2O60F Pressure in inches of water at
60 °F
mH2O4C Pressure in meters of water at 4 °C
mH2O20C Pressure in meters of water at
20 °C
cmH2O4C Pressure in centimeters of water at
4 °C
cmH2O0C Pressure in centimeters of water at
20 °C
ftH2O4C Pressure in feet of water at 4 °C
ftH2O20C Pressure in feet of water at 20 °C
14
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 21
Reference Pressure Gauge
Error Codes
ftH2O60F Pressure in feet of water at 60 °F
iNHg0C Pressure in inches of mercury at
0 °C
mmHg0C Pressure in millimeters of mercury
at 0 °C
kpal Pressure in kilopascals
Far Temperature in Fahrenheit
Cel Temperature in Celsius
mSW Pressure in meters of SeaWater
ftSW Pressure in feet of SeaWater
MPA Pressure in MegaPascal
torr Pressure in Torr (mmHG0C)

Error Codes

The gauge uses the subsequent error codes:
101 A non-numeric entry was received
where it should be a numeric entry
102 Too many significant digits entered
103 Invalid units or parameter value
received
105 Entry is above the upper limit of
the allowable range
106 Entry is below the lower limit of the
allowable range
108 A required command parameter
was missing
109 An invalid pressure unit was
received
117 An unknown command was
received
120 The serial input buffer overflowed
121 Too many entries in the command
line
122 Pressure module not connected
15
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 22
2700G Series
Users Manual

Units Conversion

See Table 3 for units and their conversion factors.
Table 3. Units Conversion
Unit Name Conversion Factor (from kPa) Conversion Factor (to kPa)
psi 0.14503773773 6.894759
bar 0.01 100
MPa 0.001 1000
kgf/cm2 0.010197162130 98.06652
inH2O @ 4 °C 4.014742 0.249082
inH2O @ 20 °C 4.021845 0.248642
inH2O @ 60 °F 4.0185886 0.248844
ftH2O @ 4 °C 0.33456183 2.988984
ftH2O @ 20 °C 0.33515375 2.983705
ftH2O @ 60 °F 0.33488238 2.986123
16
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 23
Reference Pressure Gauge
Units Conversion
Table 3. Units Conversion (cont.)
Unit Name Conversion Factor (from kPa) Conversion Factor (to kPa)
cmH2O @ 4 °C 10.19744 0.09806383
cmH2O @ 20 °C 10.21549 0.09789056
mH2O @ 4 °C 0.1019744 9.806383
mH2O @ 20 °C 0.1021549 9.789056
kPa 1 1
mbar 10 0.1
inHg @ 0 °C 0.2952998 3.386387
mmHg @ 0 °C 7.500618 0.133322
Torr 7.500618 0.133322
ftSW @ 0 °C 0.325408 3.073062
mSW 0.09918444 10.08222
17
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 24
2700G Series
Users Manual

Specifications

Accuracy

Positive Pressure ............................................................................ ±0.02 % FS
Vacuum .......................................................................................... ±0.05 % FS
Temperature Compensation ........................................................... 18 °C to 28 °C (65 °F to 82 °F) to rated accuracy
Note: For temperatures from 0 °C to 18 °C and 28 °C to 50 °C, add .003 % FS/°C

Media Compatibility

15, 30 psi ........................................................................................ any clean dry non-corrosive gas
100, 300, 500, 1000 psi .................................................................. any liquids or gases compatible with 316 stainless steel
Above 1000 psi ............................................................................... any non-flammable, non-toxic, non-explosive,
non-oxidizing liquid or gas compatible with 316 stainless steel.

Environmental

Operating Temperature .................................................................. 0 °C to 50 °C (32 °F to 122 °F)
Storage .......................................................................................... -20 °C to +70 °C (-4 °F to +158 °F)
Humidity ......................................................................................... 10 % to 90 % RH Non-condensing
Altitude ............................................................................................ 2000 meters (6561.68 ft.)
Pollution Degree ............................................................................ 2
Agency Approvals ........................................................................... P, , )
18
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 25
Reference Pressure Gauge
Specifications

Mechanical Specifications

Dimensions .................................................................................... (11.4 x 12.7) cm, depth = 3.7 cm
(4.5 x 5) in, depth = 1.5 in
(Without Protective Cover)
Pressure
Connection ............................................................................. ¼ in NPT Male
Housing .................................................................................. Cast ZNAL
Display ............................................................................................ 5-1/2 Digits, 16.53 mm (0.65 in) high
20-Segment bar graph, 0 to 100 %
Power
Battery ................................................................................... three size AA alkaline batteries
Battery Life ............................................................................ 75 hours typical without backlight
19
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 26
2700G Series
Users Manual

Available Pressure Ranges

Model
Number
Pressure Range (psi)
Pressure Range (MPa)
Vacuum Range (psi)
Vacuum Range (kPa)
Burst Pressure (psi)
Burst Pressure (MPa)
Proof Pressure (psi)
Proof Pressure (MPa)
2030-
BG100K
15 30 100 300 500 1000 3000 5000 10000
0.1 0.2 0.7 2 3.5 7 20 35 70
-15 -15 -12 -12 -12 -12 0 0 0
-100 -100 -80 -80 -80 -80 0 0 0
45 90 1000 2000 2000 10000 10000 10000 15000
0.3 0.6 7 14 14 70 70 70 100
30 60 200 600 1000 2000 6000 8000 13000
0.2 0.4 1.4 4 7 14 40 55 90
2030-
BG200K
2030-
BG700K
2030-
BG2M
2030-
BG3.5M
2030-
BG7M
2030-
G20M
2030-
G35M
2030-
G70M
20
Find Quality Products Online at: sales@GlobalTestSupply.com