The Fluke Corporation is pleased to hereby grant GIDEP
(Government-Industry Data Exchange Program) permission to reproduce the
manuals listed in the attachment. Permission is granted on the condition
that you legibly reproduce on the copies the legend "Reproduced With
Permission".
It is our understanding from your request dated January 23, 2001,
that the purpose for copying these manuals is to include them in the GIDEP
database, and that the material won't be copied for any other purpose.
Thank you for your cooperation, and for your interest in Fluke.
Notwithstanding any provision of any agreement the following warranty is exclusive:
The JOHN FLUKE MFG. CO., INC., warrants each instrument it manufactures to be free from defects in material and
workmanship under normal use and service for the period of 1-year from date of purchase. Thiswarranty extends only
to the original purchaser. This warranty shall not apply to fuses, disposable batteries (rechargeable type batteriesare
warranted for go-days). or any product or parts which have been subject to misuse, neglect, accident, or abnormal
conditions of operations.
In the event of failure of a product covered by this warranty, John Fluke Mfg. Co., Inc., will repair and calibrate an
instrument returned to an authorized Service Facility within 1 year of the original purchase; provided the warrantor’s
examination discloses to its satisfaction that the product was defective. The warrantor may, at its option, replace the
product in lieu of repair. With regard to any instrument returned within 1 year of the original purchase, said repairs or
replacement will be made without charge. If the failure has been caused by misuse, neglect, accident, or abnormal
conditions of operations, repairs will be billed at a nominal cost. In such case, an estimate will be submitted before
work is started, if requested.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS,
OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. JOHN FLUKE MFG. CO., INC., SHALL
NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN
CONTRACT, TORT, OR OTHERWISE.
II any failure occurs, the following steps should be taken:
1. Notify the JOHN FLUKE MFG. CO., INC., or nearest Service facility, giving full details of the difficulty, and
include the model number, type number, and serial number. On receipt of this information, service data, or
shipping instructions will be forwarded to you.
2. On receipt of the shipping instructions, forward the instrument, transportation prepaid. Repairs will be
made at the Service Facility and the instrument returned, transportation prepaid.
SHIPPING TO MANUFACTURER FOR REPAIR OR ADJUSTMENT
All shipments of JOHN FLUKE MFG. CO., INC.. instruments should be madevia United Parcel Serviceor”BestWay”’
prepaid. The instrument should be shipped in the original packing carton; or if it IS not available, use any suitable
container that is rigid and of adequate size. If a substitute container is used, the instrument should bewrapped in paper
and surrounded with at least four inches of excelsior or similar shock-absorbing material.
CLAIM FOR DAMAGE IN SHIPMENT TO ORIGINAL PURCHASER
The instrument should be thoroughly inspected immediately upon original delivery to purchaser. All material in the
container should bechecked againsttheenclosed packing list. The manufacturerwill not be responsible for shortages
against the packing sheet unless notified immediately. If the instrument is damaged in anyway, aclaim should be filed
with the carrier immediately. (To obtain a quotation to reparr shipment damage, contact the nearest Fluke Technical
Center.) Final claim and negotiations with the carrier must be completed by the customer.
The JOHN FLUKE MFG. CO., INC, will be happy to answer all applications or use questions, which will enhance your
use of this instrument. Please address your requests or correspondence to: JOHN FLUKE MFG. CO., INC., P.O. BOX
C9090, EVERETT, WASHINGTON 98206, ATTN: Sales Dept. For European Customers: Fluke (Holland) B.V.. P.O. Box
5053, 5004 EB, Tilburg, The Netherlands.
*For European customers, Air Freight prepard.
John Fluke Mfg. Co., Inc., P.O. Box C9090, Everett, Washington 98206
Rev. 6181
Change/Errata Informal
Issue No:
This change/errata contains information necessary to ensure the accuracy of the following manual.
Enter the corrections in the manual if either one of the following conditions exist:
1. The instrument’s PCB revision letter is lower than that which is indicated at the beginnir
of the change/errata.
2. No revision letter is indicated at the beginning of the change/errata.
4
2J79
MANUAL
Title:
Print Date:
Rev and Date:
MODEL 2190A DIGITAL THERMOMETER
JUNE 1978
---
C/E PAGE EFFECTIVITY
Page No. Prlnt Date
1
2 1 O/78
3 10178
4 2179
5 2179
6 2179
1 O/78
.E RRATA #l
2190A
On page 4-2, replace Figure 4-l
CHANGE #l-l 1690
and
Table 4-1 with the following:
Line Voltage Switch Settings
120
220 right
240 right
left
left
right
left
Rev.-C, Main PCB Assembly
On page 5-10, make the following changes:
DELETE: S2-i/Switch/473454/89536/473454/ 1
CHANGE: Sl-l/Switch 473355 89536 473355
CHANGE #L-l 1732
/483891/ /48389’i
I
Rev. -C, Main PCB Assembly, make the following changes:
On page 8-3, Figure 8-1, Al Main PCB Assembly make the indicated changes:
1 O/78
CHANGE #3-l 1789
On page 8-5, Figure 8-1, Al Main PCB Assembly, make the following changes:
TO: pf----3
’ bxq
2190A
ERRATA#2
On page
On page 2-2, Figure 2-1, (Controls and Indicators) make the following changes:
On page 3-1, paragraph 3-S, line 9, change “microcomputed” to “microcomputer”.
IO/78
l-2,
Table l-3 ( Specifications) make the following changes:
FROM: Common Mode Rejection Z 160 dB at .50,60 and 400 Hz +O.l% with 1OOR unbalance at input.
TO:
FROM: Normal Mode Rejection > 90 dB at 50,60 and 400 Hz *0.10/o
TO:
FROM: Power: 12V dc or 100, 200 or 240V ac *IO’% selectable, 50 to 400
TO:
Common Mode Rejection Z 160 dB at dc, 50,60 and 400 Hz +O.l% with 1 OOs2 unbalance at input.
Normal Mode Rejection > 90 dB at dc, 50,60 and 400 Hz +0.1%x
HZ;
8W typical.
Power: 12V dc or 100, 120,200 or 240V ac *lo%, selectable, 50 to 400 Hz; 8W typical.
ADD an arrow from 4 to scale switch:
0
2
.
On page 3-3, paragraph 3-22, change line 4 to read:
“These voltages appear at FET switches Q12 and Q13 respectively.”
On page 3-3, Figure 3-2, change as shown below:
2190A
AUTO ZERO COMMAND (Q5, QlO, 012, a161
INTEGRATE COMMAND (Q15)
READ COMMAND, DE+ Qll)
SETTLING TIME COMMAND n2 (Q8)
COMPARE OUTPUT, CM
INTEGRATOR OUTPUT U8-7)
BUFFER OUTPUT
TPl 01
(TP7l
MEASUREMENT CYCLE = 300 ms
A
/ AUTO
-, ZERO f
’ lmt0 ’ ,,,,TE- ’
i
200
OV i
-15v :
ov
-15v I
ov 1
-15v :
ms 1 GRATE !
I
,
I
I
,
I
I
I
-1101 .O msl I
I I
I I
-hII+ ,
i i 1 .O
u
I
I
I
I
‘+
ms
u
I
I I
READ
\
I
i
I
I
I
I
I
I
I
r I
I
u IJ
DE+ is generated for positive input voltages. Negative Input voltages enable the DE- mode which, in turn, operates FET
switch, Q13.
Waveform is shown for positive input voltage. It is inverted for negative inputs. Shown for full-scale in.
3 The Q numbers in parenthesis represent FET switches operated by the commands. See Figure 3-3.
D
Figure 3-2. Measurement Cycle Waveforms
On page 3-4, Figure 3-3, Simplified Schematic Analog Section, make the following changes:
ADD: Arrow to emitter; defines NPN.
1 O/78
3
2190A
On page 4-2, paragraph 4-19, line 6, change to read as follows:
“
. . . ambient temperature of 25 +2”C (77.0 ?3.6”F).”
On page 4-3, Table 4-2, change minimum use specifications for Mercury Thermometer as follows:
FROM: “0.05”C resolution.”
TO: “0.05”F resolution.”
On page 4-3, paragraph 4-20, step 8, change to read as follows:
“Select that portion of Table 4-3, , . .”
On page 4-4, paragraph 4-24, step 4, change to read as follows:
“4. Select a switch setting of 9 (Cal), see Table 2-2.”
On page 600-2, paragraph 600-12, delete the last sentence.
On page 602-3, paragraph 602-24, line 4, change “PRI” to “PTI”.
On page 606-3, Table 606-2, step 3, change “+4V” to “+.5V”.
On page 7A-1/7A-2, make the following changes:
Change line 7, of Introduction to read as follows:
“
. . .
this manual with an X.”
Table on 7A-1/7A-2, is missing title, at top of table, place the following title there:
“Table 7A-1. Manual
Status
and Backdating Information”
On Table 7A-1, change assembly name of the A3 Assembly to: “Thermocouple Input PCB Assembly”.
On page 3-4, Figure 3-3, change integrator capacitor:
FROM: Cl3
TO: C2.
ERRATA #3
On page 5-9, Table 5-2, Al Main PCB Assembly (cant), make the following changes:
FROM: R14/Res, var, lOOI,+-lo%, ~W/369520/89536/369520/1/1
TO: R14, Res, var 500k &lo%, ‘/W/474387/1 1236/360T503A/l
FROM: U3/1C, CMOS, Hex open drain buffers/473389/12040/MM74C906N/l/l
On page 5-3, Table 5-1, Final Assembly, make the following changes:
ADD: U3/IC, CMOS, Hex open drain buffers/473389/i 2040/MM74C906N/l /l
ERRATA +l
On page 602-3, Figure 602-l make the following changes:
ADD to the end of Step 5: the degree symbol (
“““) is transmitted to the printer as a (“d”),
ERRATA #5
On Page 600-2 replace paragraph 600-12 with the following:
2190A
600-I 2 The device is a preset 100 to I divider to provide the precision millivolt outputs
from a DC Calibrator required for calibration.
couple input PCB in place of the thermocouple probe during calibration.
to the Y2022 instruction manual for schematic and additional informalion.
On Page 4-3, Paragraph 4-20 make the following changes:
Place the following note between steps 3 and 4:
On Page 4-6, paragraph 4-26, Step 6 make the following changes:
From: Adjust RI4 for a thermometer display of 10.0,. .
To : Adjust RI4 for a thermometer display of I .O,.
CHANGE #+l, 12061
Rev. E, A3 Thermocouple Input PCB Assembly.
On Page 5-13/5-14, Table 5-4 make the following changes:
Add: C3, C4/Cap. Ta. 1 .O fiF +-20%,
Add: UljlC, Linear/473819/04713/MC79LI2ACP/I
From: R2/Res, mtl film. 187K ?I%, 1/8W/291641/91637/CMF551873F/1
On Page 5-3, Table 5-l make the following changes:
From: F 1 -I 20/Fuse, Fast acting/ 196790/7 1400/AGC I -S/l /5/ 1
To: Fl-I?O/Fuse, Slo-blo. 1/8A/Ih648ti/7I400JMDL1-8/1/5/1
From: Fl-240/Fuse, Fast acting/109348/71400/AGCl-16!1/5/1
To: Fl-240/Fuse, Slo-blo, l/16A/l63030/7I400/MDLI -16/1/S/l
ERRATA #6
On Page 4-3 make the indicated changes to the steps of Paragraph 4-20 as shown below:
I. No change
2. No change
3. No change
4. No change
5. No change
6. No change
7. Select the temperature scale (C” or F”) on the 2190A to agree with the temperature
scale of the mercury thermometer in the lag bath.
2179
2190A
ERRATA #I%
8. Read the temperature of the mercury thermometer and find the corresponding mV reading
to this temperature (refer to either the NBS Monograph 125 or the DIN 43710 thermocouple
standard). Interpolation may be used as necessary.
9. Algebraically subtract the mV reading found in Step 8 from the first value found under the
“Divider Output MV”
10. Obtain a divider output equal to the result found in Step 9 by adjusting the calibrator and or
the Divider.
11. Verify that the 2190A reads the value listed in Table 4-3 within the limits specified.
12. Repeat steps 8. 9. 10 and I I for the remaining voltages in the “Divider Output mV” column
Table 4-3.
13. No change
14. No change
15. No change
16. No change
17. No change
(CONT)
column of Table 4-3 for the thermocouple type in use.
6
2179
CEAIYGWEERATA
ISSOB 110: 6 5/84
IlWOMATIO!J
This change/errata contains information necessary to ensure the accuracy of
the following manual. Enter the corrections in the manual if either one of
the following conditions exist:
1.
The revision letter stamped on the indicated PCS is equal to or
higher than that given with each change.
2.
No revision letter is indicated at the beginning of the change/errata.
Add VR4 and its symbol directly beneath and parallel to R49.
On page 8-3:
Add VR4 and its symbol between U14-10 and +15V (anode to Ul4-10).
On page 8-4:
Delete the connection between 55-16 (ANALOG COM) and 55-17 (RJ COM).
IALL 2190A MAIN PCBS WITH REV. LETTERS E AND LOWER 1
IARE m COMPATIBLE WITH THE FLUKE 2300A SCANNER OR (
/THE Y2001 THERMOCOUPLE MULTIPOINT SELECTOR. IF YOU ;
IHAVE SUCH AN INSTRUMENT AND YOU WISH TO INTERFACE !
IWITH THE
2300A
OR THE Y2001 You MUST CONTACT YOUR f
INEAREST FLUKE SERVICE CENTER FOR MODIFICATION.
iHOWEVER IF YOUR 2190A MAIN PCB HAS A REV. LETTER OFI
I F
IWITH THE 2300A OR THE
OR HIGHER THEN YOUR INSTRUMENT IS COMPATIBLE
Y2001
AND NO MOD IS REQUIRED.;
I :
I
1
CHANGE #9 -
Rev.- G,
13540
002 output option
On page 602-g:
FROM: U210PTO-ISOLATOR
TO: U210PTO-ISOLATOR
FROM: U310PTO-ISOLATOR
TO: U310PTO-ISOLATOR
FROM: U410PTO-ISOLATOR
TO:
~41
0PT0-ISOLATOR
ERRATA 15
On page
8-13,
change the
FROM: GND, TO: GND ISOLATED .
On page 8-5, change S2 to
On page 4-5:
The celicius equivalent for 400 OF should be 204.4 'C,
2 190A Digital Thermometer
Controls and Indicators
Thermocouple PCB Access
2190A Simplified Block Diagram
Measurement Cycle Waveforms
Simplified Schematic - Analog Section
Line Voltage Selection Switches
Equipment Connections
Calibration Adjustment Connections
.......................................
..........................................
........................................
....................................
..........................................
..................................
.............................
...................................
...............................
PAGE
;:2
2-4
3-1
3-3
3-4
4-2
4-3
4-6
\’
2190A
279OA Digital
Thermometer
2190A
Section 1
Introduction & Specifications
l-l. lNTRODUCTlON
1-2. The Model 2190A Digital Thermometer is a
portable, stackable, five digit, thermocouple
thermometer capable of temperature measurements over
a range of -252” C to +247 1°C (-486“ F to i-4480” F) with
a resolution of 0.1”. The instrument features automatic
reference-junction compensation, eliminating the
requirement for an ice bath reference-junction; front
panel switch selection of Fahrenheit or Celsius readings;
and digital linearization (microcomputer controlled) of
the thermocouple inputs.
l-3. The instrument display has six high-intensity, seven
segment LEDs, and leading zero suppression. Five of the
LEDs are used to display numeric data, including a minus
sign for negative temperature readings. The remaining
LED displays the selected temperature scale character
(OF or “C). The symbol “O.C.“is displayed when an openthermocouple is detected at the thermocouple input
terminals.
l-4. Screw-type input connectors, on a removable
module, are provided for attaching the selected
thermocouple. The terminals are an integral part of an
isothermal block which serves as a portion of the
reference-junction compensation circuit. This circuit
removes the need for an ice bath reference junction (O°C
or 32°F) since it provides the electrical equivalent
automatically.
l-6. The measurement range of the 2190A is determined
by the type of thermocouple used as the input device. The
type of microcomputer installed in your instrument will
determine the type of thermocouples that may be used.
The specifications table identifies the type of
thermocouples that are presently available for use in the
2190A, and Section 2 gives the procedure for identifying
the microcomputer type. The version required must be
specified when ordering the instrument. Choice of the
thermocouple type is accomplished by selecting the
applicable switch position
module using the table printed beside the switch on the
pcb. Recalibration is not required if the thermocouple
type is changed.
The letter C is not an ISA/ANSI
thermocouple desgnation. C is
convenience on!y to represent Tungsten-So/o
Rhenium versus Tungsten-26% Rhenium
allo>!.
l-7. The instrument is in a PTI (Portable Test
Instrument) case. This lightweight, durable plastic case
makes the instrument truly portable when used with the
Y2003 or Y2009 external power sources and can be
stacked for one handle portability.
l-8. Options and accessories available for the 2190A are
listed in Tables l-l and l-2 respectively. Detailed
information concerning each item listed is given in
Section 6 of this manual, Option and Accessory
Information.
NOTE
on
the removable input
used for
l-5. Selection of the temperature scale for display is
made with a front panel pushbutton. The scale selected,
Celsius or Fahrenheit, is displayed as either the character
OC or o F as the last character in the temperature reading.
A scale change can be made at any time, and has no effect
on the calibration of the instrument.
Table l-l. 2190A Options
j::::,
l-1
Loading...
+ 94 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.