The American Cancer Society and American College of Radiology guidelines for the screening of asymptomatic
women have made over 50 million women candidates for mammography. In view of the staggering numbers
involved, it is critically important that simple but reliable methods be developed to assess system performance, to
assure consistent system performance, and to assure consistent production of diagnostically useful images (1, 17).
1.2 The Development of the Standard of Reference
Phantoms for use in mammography should simulate a real breast as closely as possible (2). A list of desirable features
for such a phantom can be found in Section 2, page 2-2. Note that the phantom should be able to test for both
image quality and dose if system performance is to be evaluated. The phantoms must also be easy to use and yield
images that may be unambiguously interpreted.
In developing the tissue-equivalent/realistically shaped phantom:
• Image Contrast may be measured quantitatively with standard densitometers though the use of the
embedded step wedge.
• Dose may be calculated by "TLD" or by ion chamber placed on top of the phantom and converted to average
glandular dose through conversion tables (3.6 and 3.7) in NCRP Report #985 (2). A suggested dose chart is
shown in Figure 2-1.
• Resolution - Simulated tumors and microcalcifications of known size and location are embedded in the
phantom for qualitative evaluation. The smallest microcalcifications and tumors are small enough that they
will not normally be detected.
1.3 The Realistically Shaped, Tissue-Equivalent Series of Breast
Phantoms
Shape
Standard dental modeling techniques were used to obtain molds of the compressed right breast of a volunteer
female subject. This breast is 4.5 cm thick and 18 cm in width.
Materials
Tissue-equivalent resin molding techniques were used. The system of resins used have been developed over the past
six years to permit mimicking of any body tissue at different diagnostic x-ray levels. The elemental composition of the
simulating tissue as compared to Hammerstein's analysis (11) of human tissue is shown in Table 2-1. Also shown in
Table 2-2 are comparisons of linear attenuation coefficients for actual and simulated tissue.
1-1
Page 6
The basic phantom (Model 18-222) matches the composition of an average firm breast consisting of 50% adipose
tissue and 50% glandular tissue and is realistically shaped. The phantom is suitable for evaluating the mammographic
process in the laboratory as well as for monitoring system performance in the clinic. The phantom may be used for
screen-film mammography or xeromammography. Each molded breast is surrounded with a .5 cm adiposeequivalent tissue. Thus, the glandular portion of the standard phantom is 3.5 cm in thickness.
The materials used in this phantom have been formulated for optimum response in the film-screen mammographic
range of x-ray exposure (24 to 34 kVp), but will generally provide similar results at higher (xeromammographic)
exposure ranges.
The resin materials mimic the photon attenuation coefficients of a range of breast tissues. The average elemental
composition of the human breast being mimicked is based on the individual elemental compositions of adipose and
glandular tissues as reported by Hammerstein (11). See Tables 2-1 and 2-2 for comparative data.
The attenuation coefficients are calculated using the "mixture rule" and the photon mass attenuation and energy
absorption coefficients table of J.H. Hubbell (16).
Optional Size Phantoms
The Model 18-222 Phantom is 4.5 cm in compressed thickness. Other sizes available are 4 cm, 5 cm, and 6 cm
thickness.
Optional Tissue Densities
Densities ranging from 20% glandular/80% adipose to 70% glandular/30% adipose are available on request.
Details
The Standard Phantom (Model 18-222) has embedded details (Figure 2-2) consisting of:
• Seven masses that are 75% glandular and hemispherical in shape.
• A wax insert with embedded nylon fibers.
• The Model 18-222 has an optical density reference zone. This allows OD measurements, which are position
dependent, to be taken from the same area each time. This helps factor out OD variances.
• The Model 18-222 has two edges of beam targets. This enables precise localization of the x-ray beam's edge - for
example, is the machine penetrating the chest wall, or is it not close enough to the chest wall such that
something may be missed in a clinical setting?
• One line pair test target with line pair tests between 5 and 20 line pair/mm.
The Physicist Research Model
• Three tissue equivalent phantoms with removable outer fat layer and with embedded details similar to Model
18-222
- 4 cm - 50/50 (dense)
- 5 cm – 30/70 (normal) (or 4.5 cm - 50/50)
- 6 cm - 20/80 (fatty)
• Three tissue equivalent slab combinations of plates ranging from .5 cm thickness to 2 cm. This permits test
imaging in .5 cm increments from a thickness of .5 cm to 7 cm.
- 30% glandular/70% adipose
- 50% glandular/50% adipose
- 70% glandular/30% adipose
(Model 18-223) includes:
1-2
Page 7
• Each set has one removable detail plate (50% glandular) containing:
- Step Wedge
- Simulated tumors (100% glandular)
- Microcalcifications (CaCO
- Tabular alumina specs (AL
3
)
2O3
)
- Fibril (8.7 micron) plus cladding
- Line pair test target (20 LP/mm)
1.4 Clinical Usefulness
The phantom approaches the desirable features see Section 2, page 2-2. The phantom is realistically shaped and has
the tissue equivalency of an average, firm breast. Breast detail components closely mimic the radiographic properties
and shapes of normal and pathological breast structures. The shape and configuration of the phantom makes it easy
to use by both technologists and physicists. Since the phantom is both realistically shaped and tissue equivalent, it
can be reliably used to test for radiation dose as well as image quality. A recently completed field study confirms this
assumption (6). Finally, the phantom provides valuable image quality information. The subjective assessment of detail
visibility is easy to use for routine clinical assessment while densitometric analysis provides necessary accuracy for
laboratory work. Hence the phantoms may be used to assess the mammographic process as well as assuring
consistent image performance.
Table 2-3 provides a comparison of composite attenuation for various mammographic phantoms currently
commercially available. Also shown are similar calculations for breast tissue using Hammerstein's methodology (11).
1-3
Page 8
(Blank page)
Page 9
Section 2
Operation
2.1 How To Use the Mammographic Phantom
What To Do First
• Select the technique you would use on a normal 4.5 cm compressed breast of average glandular composition.
• Take one photo-timed image at the technique normally used for the average breast patient.
• With standard densitometer, read central background density in the center of the phantom image. This
background density should be 1.0 to 1.2 optical density.
• If first film does not give OD of 1.0, then adjust technique to obtain a background OD of 1.0.
• Record technique and retain image. This now becomes your image control film.
Quantitative Procedures (at least once a week)
• Count the number of microcalcification groups visible and record the number.
• Count the number of simulated tumors and record the number.
• With optical densitometer, read fat and gland steps of the step wedge. Record the values, and the difference
(i.e., contrast). The fat/gland (steps 1 vs. step 5) should be .28 or greater.
• With a magnification lens, identify the number of line pair/mm, which are discernible.
• Record values on the record sheet (Figure 2-3).
Long Term Comparisons
• Once a quarter, take one of the weekly test films and compare visually to the initial film. You should see
identical images. If not, then corrective actions should be initiated.
Records To Keep
• Daily record of processor function (temperature and OD of step 10 or 11). This requirement is well understood
and not discussed further herein.
• Weekly record of step wedge contrast and detail visibility.
• Retained films of weekly phantom checks.
• Keep the QA record sheets (see Figure 2-3) in a file. These records of system performance are valuable to you
as a management tool and as proof of good "QA" should your system performance ever be challenged.
Care And Handling
These phantoms are manufactured from high quality materials but, like anatomy they represent, they can be broken.
Please
handle with care.
If you will treat these phantoms as you would any fragile piece of technical equipment, they will serve you well for
many years.
When not in use, the phantom should be stored in a safe location. Store at normal room temperature. If subjected to
temperatures above 110° for any extended period of time, return the phantom to Cardinal Health, Radiation
Management Service for re-certification.
2-1
Page 10
Cleaning may be accomplished by using mild
p
cause the surface of the to become tacky. We use "Armor-All" vinyl protectant and experience no difficulty with tacky
surfaces.
Avoid contact with corrosive substances and with radiographic contrast media. Wash thoroughly if such contact
occurs.
soap and water solution. It has been reported that some detergents
Desirable Features of A Breast Phantom
1. Structural characteristics of the phantom:
a. Phantom should be realistically shaped.
b. Phantom should be tissue equivalent.
c. Phantom should have a realistic background
d. Phantom components should mimic features of breast disease (calcifications, tumors.).
2. Phantom should be easy to use.
3. Phantom should test relevant parameters including absorbed dose and image quality.
4. Phantom images should be easy to interpret and provide an accurate, unambiguous measure of image
quality.
Suggested Dose Curve for
Tissue-Equivalent Phantom with
CGR-500 Unit
(HVL = .344 @ 30 kVp)
Dg
(RADS)
Average Glandular Dose (2)
kV
Figure 2-1. Suggested Dose Chart
(1) ½ RAD is considered the maximum acceptable dose for 1 view mammogram of the average patient per the National
Council on Radiation Protection and Measurements (NCRP-80).
(2) NCRP-85; Pages 40 – 56.
(3) Measurements were taken at exposure settings that produced background photographic density of 1.0 using Ortho-M
Film, Min-R screen, Grid, and General Purpose film processor.
2-2
Page 11
Table 2-1. Actual vs. Simulated Tissue (Weight Fractions)
Actual Tissue 61.900 25.100 11.200 1.700 .100 .930
Simulated 75.950 9.820 11.760 1.230 1.170 .924
2-3
Page 12
Table 2-2. Actual vs. Simulated Linear Attenuation Coefficients (u)
30% Glandular 100% Glandular
keV Actual Simulated keV Actual Simulated
10.0 3.400820 3.262850 10.0 4.919490 4.685870
15.0 1.112980 1.098010 15.0 1.560170 1.535910
20.0 0.574784 0.574784 20.0 0.768012 0.768012
30.0 0.302201 0.304501 30.0 0.368387 0.371520
40.0 0.232988 0.234800 40.0 0.268827 0.270928
50.0 0.205005 0.206445 50.0 0.229969 0.231386
60.0 0.189907 0.191118 60.0 0.209919 0.210931
80.0 0.172653 0.173673 80.0 0.188308 0.189033
100.0 0.161710 0.162638 100.0 0.175442 0.176050
50% Glandular 100% Adipose
keV Actual Simulated keV Actual Simulated
10.0 3.812000 3.622000 10.0 2.975010 2.837940
15.0 1.234110 1.213450 15.0 0.995056 0.960624
20.0 0.627163 0.627163 20.0 0.530186 0.511345
30.0 0.320188 0.323307 30.0 0.294058 0.280331
40.0 0.242758 0.245163 40.0 0.233268 0.220533
50.0 0.211829 0.213702 50.0 0.208080 0.195861
60.0 0.195389 0.196938 60.0 0.194093 0.182268
80.0 0.176952 0.178232 80.0 0.177546 0.166402
100.0 0.165485 0.166639 100.0 0.166690 0.156112
70% Glandular
keV Actual Simulated
10.0 4.231200 3.984430
15.0 1.357150 1.329930
20.0 0.679992 0.679992
30.0 0.337905 0.342249
40.0 0.252103 0.255583
50.0 0.218186 0.220987
60.0 0.200393 0.202773
80.0 0.180781 0.182797
100.0 0.168807 0.170643
NOTE
Our simulated materials are formulated to maximize
simulation properties at 20 keV for the mammographic
energy range and 70 keV for the diagnostic energy range.
2-4
Page 13
Table 2-3. Attenuation Comparison (µx) for Uses in the Well Known Relationship Various
This chart compares the composite attenuation for various phantom size/density compositions.
The linear attenuation coefficient (µ) for each type of material (wax/lucite/gland/fat/etc.) applied to the thickness of
the material in each phantom design permits calculation of total attenuation for each phantom design.
Record film type and record
serial number of film box in use.
Record kVp used for phantom
test measurement. Use kVp
normally used for an average
density 4.5 cm breast.
Record processor temp at 9:00
A.M. each day.
Keep a box of film set aside sensitize and process. Read step
10 with optical densitometer –
record value.
Again, read steps 9 – 11 on the
sensitized film. Subtract step 9
value from step 11. Record
contrast.
On the phantom test image,
read stepwedge step 1 and
step 5. Subtract values. Record
contrast.
On phantom test image, read
background density in the
middle of phantom with optical
densitometer. Record value.
On phantom test image, count
the number of micro
calcification groupings visible.
Record value.
On the phantom test image,
count the number of low
contrast masses visible.
Records value.
On phantom test image, view
line pair test target with
microscope. Record the
number of line pairs/mm
visible.
Calculate the exposure monthly
with ion chamber and convert
to mean glandular dose or,
contact CIRS for QC kit.
Figure 2-3. Quality Assurance Record
2-7
Page 16
2.2 References
1. American Cancer Society, CA32: 226-230, 1982.
2. National Council on Radiation Protection and Measurements (NCRP) Report #985. March 86.
3. Stanton L. Villafana, Day, Lightfoot. Dosage Evaluation in Mammography. RADIOLOGY 50:577-584, 1984.
4. Johns HE, Cunningham Jr. THE PHYSICS OF RADIOLOGY. 4th Edition.
5. McCrohon JL, Thompson WE, Butler PF, Goldstein HA, Phillips PR, Jane RG. Mammographic Phantom