Fluke 18-222 User Manual

Page 1
Nuclear Associates 18-222 and 18-223
Tissue-Equivalent Mammography Phantom
Instruction Manual
Page 2
(Blank Page)
Page 3
Contents
Section 1: Introduction .................................................................................................................... 1-1
1.1 Background .................................................................................................................................................................1-1
1.2 The Development of the Standard of Reference .................................................................................1-1
1.3 The Realistically Shaped, Tissue-Equivalent Series of Breast Phantom ...................................1-1
1.4 Clinical Usefulness...................................................................................................................................................1-3
Section 2: Operation......................................................................................................................... 2-1
2.1 How to Use the Mammographic Phantom.............................................................................................2-1
2.2 References....................................................................................................................................................................2-8
Page 4
(Blank Page)
Page 5
Section 1
Introduction
1.1 Background
The American Cancer Society and American College of Radiology guidelines for the screening of asymptomatic women have made over 50 million women candidates for mammography. In view of the staggering numbers involved, it is critically important that simple but reliable methods be developed to assess system performance, to assure consistent system performance, and to assure consistent production of diagnostically useful images (1, 17).
1.2 The Development of the Standard of Reference
Phantoms for use in mammography should simulate a real breast as closely as possible (2). A list of desirable features for such a phantom can be found in Section 2, page 2-2. Note that the phantom should be able to test for both image quality and dose if system performance is to be evaluated. The phantoms must also be easy to use and yield images that may be unambiguously interpreted.
In developing the tissue-equivalent/realistically shaped phantom:
Image Contrast may be measured quantitatively with standard densitometers though the use of the embedded step wedge.
Dose may be calculated by "TLD" or by ion chamber placed on top of the phantom and converted to average glandular dose through conversion tables (3.6 and 3.7) in NCRP Report #985 (2). A suggested dose chart is shown in Figure 2-1.
Resolution - Simulated tumors and microcalcifications of known size and location are embedded in the phantom for qualitative evaluation. The smallest microcalcifications and tumors are small enough that they will not normally be detected.
1.3 The Realistically Shaped, Tissue-Equivalent Series of Breast Phantoms
Shape
Standard dental modeling techniques were used to obtain molds of the compressed right breast of a volunteer female subject. This breast is 4.5 cm thick and 18 cm in width.
Materials
Tissue-equivalent resin molding techniques were used. The system of resins used have been developed over the past six years to permit mimicking of any body tissue at different diagnostic x-ray levels. The elemental composition of the simulating tissue as compared to Hammerstein's analysis (11) of human tissue is shown in Table 2-1. Also shown in Table 2-2 are comparisons of linear attenuation coefficients for actual and simulated tissue.
1-1
Page 6
The basic phantom (Model 18-222) matches the composition of an average firm breast consisting of 50% adipose tissue and 50% glandular tissue and is realistically shaped. The phantom is suitable for evaluating the mammographic process in the laboratory as well as for monitoring system performance in the clinic. The phantom may be used for screen-film mammography or xeromammography. Each molded breast is surrounded with a .5 cm adipose­equivalent tissue. Thus, the glandular portion of the standard phantom is 3.5 cm in thickness.
The materials used in this phantom have been formulated for optimum response in the film-screen mammographic range of x-ray exposure (24 to 34 kVp), but will generally provide similar results at higher (xeromammographic) exposure ranges.
The resin materials mimic the photon attenuation coefficients of a range of breast tissues. The average elemental composition of the human breast being mimicked is based on the individual elemental compositions of adipose and glandular tissues as reported by Hammerstein (11). See Tables 2-1 and 2-2 for comparative data.
The attenuation coefficients are calculated using the "mixture rule" and the photon mass attenuation and energy absorption coefficients table of J.H. Hubbell (16).
Optional Size Phantoms
The Model 18-222 Phantom is 4.5 cm in compressed thickness. Other sizes available are 4 cm, 5 cm, and 6 cm thickness.
Optional Tissue Densities
Densities ranging from 20% glandular/80% adipose to 70% glandular/30% adipose are available on request.
Details
The Standard Phantom (Model 18-222) has embedded details (Figure 2-2) consisting of:
Seven masses that are 75% glandular and hemispherical in shape.
A wax insert with embedded nylon fibers.
The Model 18-222 has an optical density reference zone. This allows OD measurements, which are position
dependent, to be taken from the same area each time. This helps factor out OD variances.
The Model 18-222 has two edges of beam targets. This enables precise localization of the x-ray beam's edge - for
example, is the machine penetrating the chest wall, or is it not close enough to the chest wall such that something may be missed in a clinical setting?
One line pair test target with line pair tests between 5 and 20 line pair/mm.
The Physicist Research Model
Three tissue equivalent phantoms with removable outer fat layer and with embedded details similar to Model 18-222
- 4 cm - 50/50 (dense)
- 5 cm – 30/70 (normal) (or 4.5 cm - 50/50)
- 6 cm - 20/80 (fatty)
Three tissue equivalent slab combinations of plates ranging from .5 cm thickness to 2 cm. This permits test imaging in .5 cm increments from a thickness of .5 cm to 7 cm.
- 30% glandular/70% adipose
- 50% glandular/50% adipose
- 70% glandular/30% adipose
(Model 18-223) includes:
1-2
Page 7
Each set has one removable detail plate (50% glandular) containing:
- Step Wedge
- Simulated tumors (100% glandular)
- Microcalcifications (CaCO
- Tabular alumina specs (AL
3
)
2O3
)
- Fibril (8.7 micron) plus cladding
- Line pair test target (20 LP/mm)
1.4 Clinical Usefulness
The phantom approaches the desirable features see Section 2, page 2-2. The phantom is realistically shaped and has the tissue equivalency of an average, firm breast. Breast detail components closely mimic the radiographic properties and shapes of normal and pathological breast structures. The shape and configuration of the phantom makes it easy to use by both technologists and physicists. Since the phantom is both realistically shaped and tissue equivalent, it can be reliably used to test for radiation dose as well as image quality. A recently completed field study confirms this assumption (6). Finally, the phantom provides valuable image quality information. The subjective assessment of detail visibility is easy to use for routine clinical assessment while densitometric analysis provides necessary accuracy for laboratory work. Hence the phantoms may be used to assess the mammographic process as well as assuring consistent image performance.
Table 2-3 provides a comparison of composite attenuation for various mammographic phantoms currently commercially available. Also shown are similar calculations for breast tissue using Hammerstein's methodology (11).
1-3
Page 8
(Blank page)
Page 9
Section 2
Operation
2.1 How To Use the Mammographic Phantom
What To Do First
Select the technique you would use on a normal 4.5 cm compressed breast of average glandular composition.
Take one photo-timed image at the technique normally used for the average breast patient.
With standard densitometer, read central background density in the center of the phantom image. This
background density should be 1.0 to 1.2 optical density.
If first film does not give OD of 1.0, then adjust technique to obtain a background OD of 1.0.
Record technique and retain image. This now becomes your image control film.
Quantitative Procedures (at least once a week)
Count the number of microcalcification groups visible and record the number.
Count the number of simulated tumors and record the number.
With optical densitometer, read fat and gland steps of the step wedge. Record the values, and the difference
(i.e., contrast). The fat/gland (steps 1 vs. step 5) should be .28 or greater.
With a magnification lens, identify the number of line pair/mm, which are discernible.
Record values on the record sheet (Figure 2-3).
Long Term Comparisons
Once a quarter, take one of the weekly test films and compare visually to the initial film. You should see identical images. If not, then corrective actions should be initiated.
Records To Keep
Daily record of processor function (temperature and OD of step 10 or 11). This requirement is well understood and not discussed further herein.
Weekly record of step wedge contrast and detail visibility.
Retained films of weekly phantom checks.
Keep the QA record sheets (see Figure 2-3) in a file. These records of system performance are valuable to you
as a management tool and as proof of good "QA" should your system performance ever be challenged.
Care And Handling
These phantoms are manufactured from high quality materials but, like anatomy they represent, they can be broken. Please
handle with care.
If you will treat these phantoms as you would any fragile piece of technical equipment, they will serve you well for many years.
When not in use, the phantom should be stored in a safe location. Store at normal room temperature. If subjected to temperatures above 110° for any extended period of time, return the phantom to Cardinal Health, Radiation Management Service for re-certification.
2-1
Page 10
Cleaning may be accomplished by using mild
p
cause the surface of the to become tacky. We use "Armor-All" vinyl protectant and experience no difficulty with tacky surfaces.
Avoid contact with corrosive substances and with radiographic contrast media. Wash thoroughly if such contact occurs.
soap and water solution. It has been reported that some detergents
Desirable Features of A Breast Phantom
1. Structural characteristics of the phantom:
a. Phantom should be realistically shaped.
b. Phantom should be tissue equivalent.
c. Phantom should have a realistic background
d. Phantom components should mimic features of breast disease (calcifications, tumors.).
2. Phantom should be easy to use.
3. Phantom should test relevant parameters including absorbed dose and image quality.
4. Phantom images should be easy to interpret and provide an accurate, unambiguous measure of image quality.
Suggested Dose Curve for
Tissue-Equivalent Phantom with
CGR-500 Unit
(HVL = .344 @ 30 kVp)
Dg
(RADS)
Average Glandular Dose (2)
kV
Figure 2-1. Suggested Dose Chart
(1) ½ RAD is considered the maximum acceptable dose for 1 view mammogram of the average patient per the National
Council on Radiation Protection and Measurements (NCRP-80).
(2) NCRP-85; Pages 40 – 56.
(3) Measurements were taken at exposure settings that produced background photographic density of 1.0 using Ortho-M
Film, Min-R screen, Grid, and General Purpose film processor.
2-2
Page 11
Table 2-1. Actual vs. Simulated Tissue (Weight Fractions)
Tissue C O H N Ca P Al G1
Specific
Gravity
30% Glandular
Actual Tissue 48.850 46.400 10.700 2.150 .200 .930 Simulated 75.510 9.960 11.730 1.230 .350 1.180 .930
50% Glandular
Actual Tissue 40.150 46.400 10.700 2.450 .300 .982 Simulated 75.070 10.160 11.670 1.230 .670 1.170 .982
70% Glandular
Actual Tissue 31.150 54.920 2.750 .380 1.004 Simulated 74.650 10.350 11.620 1.230 .970 1.170 1.004
100% Glandular
Actual Tissue 18.400 67.700 10.200 3.200 .5000 1.040 Simulated 70.210 12.510 10.930 1.150 .610 3.460 1.100 1.040
100% Adipose
Actual Tissue 61.900 25.100 11.200 1.700 .100 .930
Simulated 75.950 9.820 11.760 1.230 1.170 .924
2-3
Page 12
Table 2-2. Actual vs. Simulated Linear Attenuation Coefficients (u)
30% Glandular 100% Glandular
keV Actual Simulated keV Actual Simulated
10.0 3.400820 3.262850 10.0 4.919490 4.685870
15.0 1.112980 1.098010 15.0 1.560170 1.535910
20.0 0.574784 0.574784 20.0 0.768012 0.768012
30.0 0.302201 0.304501 30.0 0.368387 0.371520
40.0 0.232988 0.234800 40.0 0.268827 0.270928
50.0 0.205005 0.206445 50.0 0.229969 0.231386
60.0 0.189907 0.191118 60.0 0.209919 0.210931
80.0 0.172653 0.173673 80.0 0.188308 0.189033
100.0 0.161710 0.162638 100.0 0.175442 0.176050
50% Glandular 100% Adipose
keV Actual Simulated keV Actual Simulated
10.0 3.812000 3.622000 10.0 2.975010 2.837940
15.0 1.234110 1.213450 15.0 0.995056 0.960624
20.0 0.627163 0.627163 20.0 0.530186 0.511345
30.0 0.320188 0.323307 30.0 0.294058 0.280331
40.0 0.242758 0.245163 40.0 0.233268 0.220533
50.0 0.211829 0.213702 50.0 0.208080 0.195861
60.0 0.195389 0.196938 60.0 0.194093 0.182268
80.0 0.176952 0.178232 80.0 0.177546 0.166402
100.0 0.165485 0.166639 100.0 0.166690 0.156112
70% Glandular
keV Actual Simulated
10.0 4.231200 3.984430
15.0 1.357150 1.329930
20.0 0.679992 0.679992
30.0 0.337905 0.342249
40.0 0.252103 0.255583
50.0 0.218186 0.220987
60.0 0.200393 0.202773
80.0 0.180781 0.182797
100.0 0.168807 0.170643
NOTE
Our simulated materials are formulated to maximize simulation properties at 20 keV for the mammographic energy range and 70 keV for the diagnostic energy range.
2-4
Page 13
Table 2-3. Attenuation Comparison (µx) for Uses in the Well Known Relationship Various
µ
Phantom Densities and Sizes
I I
O
= e
-
x
Tissue:
Thickness:
MFGR:
Fat Layer:
keV
Acrylic Acrylic Acrylic BR-12 50/50 30/70 50/50 30/70 50/50 20/80
4.4 cm 5.0 cm 4.55 cm 4.5 cm 4.0 cm 4.5 cm 4.5 cm 5.0 cm 5.0 cm 6.0 cm ACR MFR #1 MFR #2 NAD NAD NAD NAD NAD NAD NAD
No No No No Yes Yes Yes Yes Yes Yes
10 15.542 17.769 16.995 15.942 13.703 14.257 15.514 15.889 17.325 18.294
15 5.127 5.891 5.608 5.298 4.601 4.803 5.207 5.352 5.814 6.175 20 (28 kVp) 2.705 3.103 2.936 2.728 2.392 2.523 2.706 2.811 3.021 3.251 30 1.477 1.691 1.582 1.402 1.251 1.346 1.412 1.498 1.573 1.758 40 1.162 1.328 1.236 1.062 .956 1.042 1.078 1.159 1.201 1.371 50 1.033 1.179 1.095 .925 .836 .918 .944 1.021 1.051 1.211
This chart compares the composite attenuation for various phantom size/density compositions.
The linear attenuation coefficient (µ) for each type of material (wax/lucite/gland/fat/etc.) applied to the thickness of the material in each phantom design permits calculation of total attenuation for each phantom design.
Actual Breast Tissue per Hammerstein
Tissue
Thickness:
MFGR:
Fat Layer:
keV
10 15.689 14.411 16.317 18.223 16.578 18.996 15 5.066 4.697 5.314 5.931 5.446 6.271 20 (28 kVp) 2.571 2.411 2.725 3.038 2.829 3.279 30 1.307 1.254 1.414 1.574 1.502 1.762 40 .989 .961 1.082 1.204 1.165 1.375 50 .862 .843 .949 1.055 1.028 1.217
70/30 50/50 50/50 50/50 30/70 20/80
4.0 cm 4.0 cm 4.5 cm 5.0 cm 5.0 cm 6.0 cm Actual Actual Actual Actual Actual Actual
Yes Yes Yes Yes Yes Yes
2-5
Page 14
Figure 2-2. Embedded Details (Model 18-222 Specifications)
Line Pair Target Nylon Fibers 20 lp/mm
Ca Co
Specs
3
Diameter size (mm)
19. 1.25
Grain Size (mm) 20. 0.83
2. .130 21. 0.71
3. .165 22. 0.53
4. .196 23. 0.30
5. .230
Hemispheric Masses
6. .275 75% Glandular/ 25%
7. .400 adipose, thickness (mm)
8. .230 24. 4.76
9. .196 25. 3.16
10. .165 26. 2.38
11. .230 27. 1.98
12. .196 28. 1.59
13. .165 29. 1.19
Step Wedge
1 cm thick
30. 0.90
Optical Density
14. 100% Gland 31. Reference zone
15. 70% Gland
16. 50% Gland
Edge of Beam
17. 30% Gland 32. Localization target
18. 100% Adipose
2-6
Page 15
Location: Week/Month
Quality Assurance Record
Mammography
Baseline Day/WK-1 2 3 4 5 Remarks
Film Type
kVp (with photomtimer)
Processor Temperature °F
Processor Speed (Sensitometry – Step 10)
Processor Contrast (Step 9 – 11)
Phantom Contrast (Step 1 – 5)
Phantom Central Background Density
Phantom Calcifications
Phantom Low Contrast Masses
Phantom Line Pair Visible
Dose (Mean glandular dose for a
4.5 cm 50% Glandular Breast)
For daily readings, use 1 sheet/week
For weekly readings, use 1 sheet/month
Mammo Unit: QC Phantom No.
Processor Type:
Processor Cycle: 90 sec/2.5 min/3.0 min
Record film type and record serial number of film box in use. Record kVp used for phantom test measurement. Use kVp normally used for an average density 4.5 cm breast. Record processor temp at 9:00 A.M. each day. Keep a box of film set aside ­sensitize and process. Read step 10 with optical densitometer – record value. Again, read steps 9 – 11 on the sensitized film. Subtract step 9 value from step 11. Record contrast. On the phantom test image, read stepwedge step 1 and step 5. Subtract values. Record contrast. On phantom test image, read background density in the middle of phantom with optical densitometer. Record value. On phantom test image, count the number of micro calcification groupings visible. Record value. On the phantom test image, count the number of low contrast masses visible. Records value. On phantom test image, view line pair test target with microscope. Record the number of line pairs/mm visible. Calculate the exposure monthly with ion chamber and convert to mean glandular dose or, contact CIRS for QC kit.
Figure 2-3. Quality Assurance Record
2-7
Page 16
2.2 References
1. American Cancer Society, CA32: 226-230, 1982.
2. National Council on Radiation Protection and Measurements (NCRP) Report #985. March 86.
3. Stanton L. Villafana, Day, Lightfoot. Dosage Evaluation in Mammography. RADIOLOGY 50:577-584, 1984.
4. Johns HE, Cunningham Jr. THE PHYSICS OF RADIOLOGY. 4th Edition.
5. McCrohon JL, Thompson WE, Butler PF, Goldstein HA, Phillips PR, Jane RG. Mammographic Phantom
Evaluation Project. HHS Publication. 83: 8213. 1983.
6. Fatouros PF, Skubic S. The Development and Use of Realistically-shaped, Tissue Equivalent Phantom for
Assessing the Mammographic Process, RADIOLOGY 157P:32. 1985.
7. Fatouros PF. Resolution and Noise in Xeromammography. MEDICAL PHYSICS 9:819-829
8. Fatouros PF, Goodman H, Rao G, Beachley M, Janis S, Bourland P. Absorbed Dose and Image Quality in
Xeromammography. Proceedings SPIE. Vol 419. 1983.
9. White DR, Tucker A. Test Object for Assessing Image Quality in Mammography. BJR 53:331-335. 1980.
10. Hessler C, Depeusinge C, Greceescu M, et al. Objective Assessment of Mammography Systems, part I: Method.
Radiology 156:215-219. 1985
11. Hammerstein R, Miller D, White D, et al. Absorbed Dose in Mammography. RADIOLOGY 156:215-219. 1985
12. Skubic S. An Investigation of the Relationship Between Image Quality and Breast Dose in Xeromammography.
Dissertation, Medical College of Virginia. May 1986.
13. Skubic S, Fatouros PF. The Dependence of Absorbed Breast Dose on X-Ray Modality, X-Ray Technique, and
Breast Thickness. RADIOLOGY 161:263-270. 1986.
14. Fewell, Shuping. Handbook of Mammographic Spectra. FDA Publication.
15. Skubic SE, Fatouros PP, Goodman H. The Effect of Breast Composition on Absorbed Dose and Image Contrast
in Publication.
16. Hubbell JH. INTERNATIONAL JOURNAL RADIATION ISOT Vol. 33:1269-1290. 1982.
17. American College of Radiology Guidelines on Mammographic Screening.
2-8
Page 17
(Blank Page)
Page 18
Radiation Management Services
All rights reserved. Manual No. 18-222-1 Rev. 2 08 apr 04
Cardinal Health 6045 Cochran Road Cleveland, Ohio 44139
120 Andrews Road Hicksville, New York 11801
www.cardinal.com/rms
For additional information, please contact Radiation Management Services business of Cardinal Health at 440.248.9300, or
516.870.0100.
© 2004 Cardinal Health, Inc. or one of its subsidiaries
Loading...