Fluke 174x Instructions Manual

Page 1
174x
IP65 Voltage Adapter
Instructions
The IP65 Voltage Adapter (the Product or Adapter) enables safe and reliable operation of the 174x Logger family in harsh environments. The adapter seals and protects the terminals of the voltage measurement inputs and power supply from water and dust.
How to Contact Fluke
To contact Fluke, use one of these telephone numbers:
USA: 1-800-760-4523
Canada: 1-800-36-FLUKE (1-800-363-5853)
Brazil: +55-11-3530-8901
Europe: +31 402-675-200
Japan: +81-3-6714-3114
Singapore: +65-6799-5566
China: +86-400-810-3435
Anywhere in the world: +1-425-446-5500 Go to www.fluke.com
to register your Product and find more
information. To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
PN 4947664 November 2017
©2017 Fluke Corporation. All rights reserved. Specifications are subject to change without notification. All product names are trademarks of their respective companies.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
ООО «Флюк СИАЙЭС» 125167, г. Москва,
Ленинградский проспект дом 37, корпус 9, подъезд 4, 1 этаж
Page 2
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user.
XW Warning
To prevent possible electrical shock, fire, or personal injury:
• Read all safety information before you use the Product.
• Do not alter the Product and use only as specified, or the protection supplied by the Product can be compromised.
• Comply with local and national safety codes. Use personal protective equipment (approved rubber gloves, face protection, and flame-resistant clothes) to prevent shock and arc blast injury where hazardous live conductors are exposed.
• Use Product-approved measurement category (CAT), voltage, and amperage rated accessories (probes, test leads, and adapters) for all measurements.
• Do not use test leads if they are damaged. Examine the test leads for damaged insulation and measure a known voltage.
• Do not use the Product if it is altered or damaged.
• Do not work alone.
• Do not make connections on hazardous live conductors in damp or wet environments.
• Do not use the Product around gas or vapor.
• Protect the Product against direct solar radiation.
• De-energize the circuit or wear personal protective equipment (PPE) when you remove test leads installed outdoors because the insulation properties can be compromised by dust and moisture.
• Do not apply or remove test leads on energized circuits under wet conditions.
• Do not exceed the Measurement Category (CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
• Keep fingers behind the finger guards on the probes.
• Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.
• Do not apply more than the rated voltage, between the terminals or between each terminal and earth ground.
• Measure a known voltage first to make sure that the Product operates correctly.
• Do not touch the metal parts of one test lead when the other is still connected to hazardous voltage.
• Have an approved technician repair the Product.
• Remove the input signals before you clean the Product.
Page 3
Table 1 is a list of symbols used on the Product or in these instructions.
Table 1. Symbols
Symbol Description
W
X
P
)
Consult user documentation.
WARNING. RISK OF DANGER.
WARNING. HAZARDOUS VOLTAGE. Risk of electric shock.
Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building’s low-voltage MAINS installation.
Measurement Category IV is applicable to test and measuring circuits connected at the source of the building’s low-voltage MAINS installation.
Conforms to European Union directives.
Certified by CSA Group to North American safety standards.
~
Conforms to relevant Australian EMC standards.
This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.
Page 4
Before You Start
Below is a list of the items included with purchase. Carefully unpack and inspect each item.
1x IP65 Voltage Adapter
3x Screw M3 X 12 mm
2x Coupler F/F 4 mm
2x Alligator Clips (blue)
1x Wire Clip Set To prepare the Adapter for installation:
1. Attach the cable marker clips to the test leads for the three phases and neutral N at the plug end.
2. Use the markers A, B, C, N or 1, 2, 3, N as appropriate for the region. See Figure 1.
Figure 1. Cable Marker Clips
Installation
1. Insert the Adapter into the input terminals of the Logger.
Note Apply force to the Adapter body only. Do not push on the test leads.
2. Attach the Adapter with the three screws (M3 x 12 mm). See Figure 2.
3. Use the mains adapter MA-C8 and the included couplers to connect the Logger to a mains outlet. See Table 2.
Page 5
2
3
4
1
1
748
P
OW
E
R
Q
U
A
L
IT
Y LO
G
G
ER
M
A
X
3
0
V
6
0
0
C
V
A
T
6
I
0
N
0
P
C
C
U
V
U
TS
A
RRE
T
NT
1
0
0
0
C
V
AT
NC
5
0
5
V
0/
A
6
/
1
0
L3
00-6
H
z
0
0
V
B
/
L2
Wi
Fi
H
O
TS P
O
A
T
W
iFi
3
SE
S
C
TO
P
E
THER
S
T
A
N
R
E
T
T
T
I
M
E
V O
/
L
L
T
1
M
A
A
G
X
E
1 I N
0
P
0
0
UT
V
S
R
M
S
Figure 2. Adapter Attachment
Table 2. Mains Connection
Item Description
   
Safety Plugs Couplers MA-C8 Power Cord
Removal
For frequent use in IP65 environments, keep the Adapter installed on the Logger.
To remove the Adapter from the Logger:
1. Unfasten the screws.
Note
Page 6
2. Pull on both the left and the right sides simultaneously to avoid damage to the Adapter.
As an option, insert a screwdriver 3/16 inch (5 mm) into the slots on the left and right side and twist to lift the Adapter.
Note
Do not pull on the test leads.
Maintenance
Clean the Adapter with a soft cloth, mild soap, and water. To avoid damage, do not clean with abrasives or solvents.
Safety Specifications
General
Test lead length............ 2.5m (8.2 ft)
Test lead gauge............ 1 mm² (AWG 18)
Temperature
Operating...................... -25 °C to +50 °C (-13 °F to +122 °F)
Storage......................... -25 °C to +60 °C (-13 °F to +140 °F)
Operating Humidity ........IEC60721-3-3: 3K6:
-25 °C to +30 °C (-13 °F to +86 °F): ≤100 % 40 °C (104 °F): 55 % 50 °C (122 °F): 35 %
Altitude
Operating...................... 2000 m (up to 4000 m derate to 1000 V
CAT II/600 V CAT III/300 V CAT IV)
Storage......................... 12 000 m
Safety
Overvoltage.................. IEC 61010-1: Overvoltage Category IV
500 V, Pollution Degree 2
Measurement ...............IEC 61010-031: CAT IV 600 V / CAT III
1000 V
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.