5.2.6 Peak Mode
P
This mode is accessed through the Advanced Menu, see above. In Peak mode, the
meter captures and displays the positive and negative ACA and ACV peak values. The
Peak display values changes only when higher/lower values are registered.
1. Press OK to show the Peak Max and Peak Min readings on the display.
2. Press OK to switch this mode OFF.
5.2.7 Relative mode
This mode is accessed through the Advanced Menu, see above.
1. Press OK on this icon to capture a reference reading that subsequent
measurements can be compared with.
2. A blue dot is shown next to the icon when this mode is selected.
3. The reference value will be displayed next to the Relative icon.
4. The primary display will show the difference between the measured value and
the stored reference.
5. Press OK at this icon to switch the Relative mode OFF.
5.2.8 Multimeter Settings Menu
1. Press OK to open the main menu
2. Navigate to the Multimeter Settings icon using the arrow buttons
3. Open the Settings menu by pressing the OK button.
4. See the Settings menu example below. The items in the example may differ
from the meter that you have depending on the firmware version. Customize
the settings as described below:
Diode (Smart or Classic modes); see
Section 7.11, Classic Diode, and Section
7.12, Smart Diode. Use the OK button to
choose the desired setting.
Auto power off (APO): Press OK to
open the sub-menu. Use the arrow
buttons to choose the time period after
which the meter enters sleep mode.
Press OK to confirm the selection (a blue
dot will appear next to the selection).
Auto hold: Use the OK button to
toggle ON (blue dot) and OFF. For more
information, see Section 7.4, Data Hold
and Auto Hold.
Temperature units. Use the OK button to toggle
o
C and oF.