This Technical Manual is intended for clinical engineering / technical personnel. It defines the technical
specifications, setup, servicing and troubleshooting information, for the AIRVO 2 humidifier. It applies to all
lot numbers from 140910 and above.
OTHER REFERENCES
• Refer to the AIRVO 2 User Manual for detailed instructions for use.
• Watch the AIRVO 2 DVD to learn how to set up and use the AIRVO 2. Also available on YouTube.
• Download the AIRVO 2 Simulator App to learn how to use the AIRVO 2.
You can change settings, simulate faults and test your skills. Available
from the Apple, Google Play and Windows App stores.
• Visit the Fisher & Paykel education & resources website (https://www.fphcare.co.nz/education/) to find
self-paced online courses and local training events.
• If the unit is ever used by multiple patients, the unit must be cleaned and disinfected between patients
according to instructions in the Disinfection Kit Manual (900PT600).
• For further assistance, please contact your Fisher & Paykel Healthcare representative.
•
TABLE OF CONTENTS
1. General information ................................................................................................................................. 4
5. Spare Parts .................................................................................................................................................. 16
The AIRVO 2 is a humidifier with integrated flow generator that delivers warmed and humidified respiratory
gases to spontaneously breathing patients through a variety of patient interfaces.
PACKAGE CONTENTS
AIRVO 2 humidifier
(PT101xx)
Oxygen inlet extension kit
(900PT422)
AIRVO 2 User ManualAIRVO 2 SwingtagAIRVO 2 DVD
Disinfection Kit
(900PT600)
WARNING
UNDER NO CIRCUMSTANCES SHOULD THE
AIRVO 2 BE OPENED OR ANY OF THE SIX
FASTENING SCREWS ON THE UNDERNEATH
SIDE OF THE DEVICE BE LOOSENED.
OPENING THE UNIT WILL AFFECT THE
OXYGEN SEALS INSTALLED INSIDE, WHICH
WILL COMPROMISE THE SAFETY OF THE
DEVICE.
Air filter (x2)
(900PT913)
Power cord
(900PT410xx)
4
HOSPITAL
STAND
POLE
MOUNTING
TRAY
ON/OFF (STANDBY)
MUTE
UP
DOWN
MODE
AIRVO 2 AND ACCESSORIES
DISPLAY
OXYGEN
INLET PORT
SERIAL PORT
Heated
breathing
tube
Patient
interface
HEATED BREATHING TUBE
CONNECTION PORT
MEASUREMENT POINT OF
DISPLAYED DEW POINT
TEMPERATURE
CHAMBER PORTS
Water chamber
POWER CORD
and
CONNECTOR
AIR FILTER
FILTER COVER
AIRVO2
(PT101xx)
HEATER
PLATE
FINGER
GUARD
AUTO-FILL WATER
CHAMBER (MR290)
(with adapter fitted)
5
2. SETTING UP AIRVO 2 FOR FIRST USE
1. REMOVE THE AIRVO 2 FROM ITS PACKAGING
Place the AIRVO 2 on the 900PT405 pole mounting tray, on the
900PT421 hospital stand.
2. CONNECT THE POWER CORD
Plug the power cord connector into the socket on the back of the
AIRVO 2.
3. ATTACH THE OXYGEN INLET EXTENSION KIT
Refer to the instruction sheet included with the kit itself.
4. ATTACH WATER CHAMBER AND HEATED BREATHING TUBE
The water chamber and heated breathing tube must be connected to
carry out the following setup and testing procedures.
If you have not been supplied with a reusable HC360 water chamber, you
can use an MR290 chamber instead.
6
5. SWITCH ON UNIT
Switch on the unit by pressing the On/O button.
6. WARM-UP
The unit will begin to warm up.
“Warm-up” symbol
7. READY FOR USE
The “Ready for use” symbol means that the system is ready for the
patient to use.
“Ready for use” symbol
7
ADVANCED SETTINGS
pl
ar
ru
tr
he
el
ro
When you see the “Warm-up” or “Ready for use” symbols, hold a
combination of three buttons (Up, Down and Mute) for 5 seconds, to
view and change advanced settings.
This button combination is for use by clinical engineering / technical
personnel only.
AIRVO 2 / myAIRVO 2 MODE
You can change the unit from “AIRVO” (hospital) mode to “myAIRVO”
(home / long-term care) mode, eg. for patients going home.
Contact Fisher & Paykel Healthcare for a myAIRVO 2 User Manual.
To change the mode:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up button to select myAIRVO 2.
Press the Mode button to confirm the change and/or move on to the
next screen. Note that the unit will reset itself if it is switched between
AIRVO 2 and myAIRVO 2 modes.
LANGUAGE
You can set the AIRVO 2 / myAIRVO 2 to one of 22 language settings:
EnglishNederlandsSvenskaPolski
DeutschPortuguês
EspañolDansk
FrançaisSuomiΕλληνικά
ItalianoNorsk
简体中文
繁体中文
日本語
[simp.]
[trad.]
Русский
עברית
Română
To change the language:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up and Down buttons to select the desired
language.
العربية
Türkçe
Press the Mode button to confirm the change and/or move on to the
next screen.
8
95
ENVIRONMENT SETTINGS (FOR DEFAULT MODE)
A clinician may change the “Environment Settings”, to customise
individual AIRVOs for dierent environments (eg. intensive care, general
care areas, emergency departments). The “Environment Settings”
chosen will put limits on the “Patient Settings” that the operator can
choose when in normal use.
This screen defines the “Environment Settings” for the AIRVO 2 when in
Default Mode (ie. non-“Junior Mode”).
Minimum dew-point
temperature (°C)
Minimum flow (L/min)The lowest flow that the operator will be able to
Maximum flow (L/min)The highest flow that the operator will be able
Maximum oxygen fraction (%) The highest oxygen fraction that the operator
95
Minimum oxygen fraction (%) The lowest oxygen fraction that the operator
Note that, for Oxygen display, this is a measurement only, not a control setting. The
operator changes the measured oxygen fraction by altering the AIRVO 2 target
flow setting and the flow of oxygen connected to the unit (e.g. from a flowmeter)
- there is no closed-loop control.
The lowest target dew-point temperature that
the operator will be able to select.
Possible Settings: 31, 34, 37 °C
If this is set to 31, the operator can select a target
dew-point temperature between 31 and 37. ie. 31,
34 or 37 (°C).
If the patient is tracheostomised, a clinician may
wish to set this value to 37, so that the operator
can only select a target dew-point temperature
between 37 and 37, ie. only 37 (°C).
Note: The maximum dew-point temperature
setting is always 37 °C in Default Mode.
select.
Possible Settings: 10 to 60 in increments of 5 L/min,
always less than or equal to Maximum Flow setting.
Example: If this is set to 10, the operator will be
able to select flows down to 10 L/min.
If this is set to 25, the operator will be able to
select flows down to 25 L/min.
to select.
Possible Settings: 10 to 60 in increments of 5 L/min,
always greater than or equal to Minimum Flow setting.
Example: If this is set to 60, the operator can
select flows up to 60 L/min.
If this is set to 35, the operator can select flows up
to 35 L/min.
may set the unit to.
Possible settings: 30 - 100% in increments of 5% O2.
The unit will alarm if the measured oxygen
fraction exceeds this value.
Note: Even if this ‘Maximum oxygen fraction‘
setting is set to 100%, any time the measured
oxygen fraction exceeds 95%, the oxygen reading
will pulse red and the device will beep.
may set the unit to.
Possible settings: 21 or 25% O2.
When set to 25% the unit will alarm if the
measured oxygen fraction is below this value. This
allows detection of oxygen being disconnected.
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the first setting.
95
Use the Up and Down buttons to change the setting,
then press the Mode button to progress to the next
setting.
Press the Mode button to confirm the change and/or move on to the
next screen.
9
95
100
ENVIRONMENT SETTINGS (FOR JUNIOR MODE)
This screen defines the “Environment Settings” for the AIRVO 2 when in
Junior Mode.
Junior Mode Enable/Disable
Enabled
Disabled
Dew-point
temperature (°C)
Minimum flow (L/min)The lowest flow that the operator will be able to
Maximum flow (L/min)The highest flow that the operator will be able
Maximum oxygen fraction (%)
95
Minimum oxygen fraction (%) The lowest oxygen fraction that the operator
Note that, for Oxygen display, this is a measurement only, not a control setting. The
operator changes the measured oxygen fraction by altering the AIRVO 2 target
flow setting and the flow of oxygen connected to the unit (e.g. from a flowmeter)
- there is no closed-loop control.
When this option is enabled (default), the
operator can enter Junior Mode from the Home
Screen, by holding the Mode button for 5
seconds.
When this option is disabled, entering Junior
mode is not possible.
Consider disabling this option if the unit will never
be used on pediatric patients.
The only dew-point setting in Junior Mode is
34 °C.
select.
Possible Settings: 2 to 25 in increments of 1 L/min,
always less than or equal to Maximum Flow setting.
If this is set to 10, the operator will be able to
select flows down to 10 L/min.
to select.
Possible Settings: 2 to 25 in increments of 1 L/min,
always greater than or equal to Minimum Flow setting.
If this is set to 15, the operator can select flows up
to 15 L/min.
The highest oxygen fraction that the operator
may set the unit to.
Possible settings: 30 - 100% in increments of 5% O2.
The unit will alarm if the measured oxygen
fraction exceeds this value.
Note: Even if this ‘Maximum oxygen fraction‘
setting is set to 100%, any time the measured
oxygen fraction exceeds 95%, the oxygen reading
will pulse red and the device will beep.
may set the unit to.
Possible settings: 21 or 25% O2.
When set to 25% the unit will alarm if the
measured oxygen fraction is below this value. This
allows detection of oxygen being disconnected.
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the first setting.
Use the Up and Down buttons to change the setting,
then press the Mode button to progress to the next
setting.
Press the Mode button to confirm the change and/or move on to the
next screen..
10
Loading...
+ 23 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.