This Technical Manual is intended for clinical engineering / technical personnel. It defines the technical
specifications, setup, servicing and troubleshooting information, for the AIRVO 2 humidifier. It applies to all
lot numbers from 140910 and above.
OTHER REFERENCES
• Refer to the AIRVO 2 User Manual for detailed instructions for use.
• Watch the AIRVO 2 DVD to learn how to set up and use the AIRVO 2. Also available on YouTube.
• Download the AIRVO 2 Simulator App to learn how to use the AIRVO 2.
You can change settings, simulate faults and test your skills. Available
from the Apple, Google Play and Windows App stores.
• Visit the Fisher & Paykel education & resources website (https://www.fphcare.co.nz/education/) to find
self-paced online courses and local training events.
• If the unit is ever used by multiple patients, the unit must be cleaned and disinfected between patients
according to instructions in the Disinfection Kit Manual (900PT600).
• For further assistance, please contact your Fisher & Paykel Healthcare representative.
•
TABLE OF CONTENTS
1. General information ................................................................................................................................. 4
5. Spare Parts .................................................................................................................................................. 16
The AIRVO 2 is a humidifier with integrated flow generator that delivers warmed and humidified respiratory
gases to spontaneously breathing patients through a variety of patient interfaces.
PACKAGE CONTENTS
AIRVO 2 humidifier
(PT101xx)
Oxygen inlet extension kit
(900PT422)
AIRVO 2 User ManualAIRVO 2 SwingtagAIRVO 2 DVD
Disinfection Kit
(900PT600)
WARNING
UNDER NO CIRCUMSTANCES SHOULD THE
AIRVO 2 BE OPENED OR ANY OF THE SIX
FASTENING SCREWS ON THE UNDERNEATH
SIDE OF THE DEVICE BE LOOSENED.
OPENING THE UNIT WILL AFFECT THE
OXYGEN SEALS INSTALLED INSIDE, WHICH
WILL COMPROMISE THE SAFETY OF THE
DEVICE.
Air filter (x2)
(900PT913)
Power cord
(900PT410xx)
4
HOSPITAL
STAND
POLE
MOUNTING
TRAY
ON/OFF (STANDBY)
MUTE
UP
DOWN
MODE
AIRVO 2 AND ACCESSORIES
DISPLAY
OXYGEN
INLET PORT
SERIAL PORT
Heated
breathing
tube
Patient
interface
HEATED BREATHING TUBE
CONNECTION PORT
MEASUREMENT POINT OF
DISPLAYED DEW POINT
TEMPERATURE
CHAMBER PORTS
Water chamber
POWER CORD
and
CONNECTOR
AIR FILTER
FILTER COVER
AIRVO2
(PT101xx)
HEATER
PLATE
FINGER
GUARD
AUTO-FILL WATER
CHAMBER (MR290)
(with adapter fitted)
5
2. SETTING UP AIRVO 2 FOR FIRST USE
1. REMOVE THE AIRVO 2 FROM ITS PACKAGING
Place the AIRVO 2 on the 900PT405 pole mounting tray, on the
900PT421 hospital stand.
2. CONNECT THE POWER CORD
Plug the power cord connector into the socket on the back of the
AIRVO 2.
3. ATTACH THE OXYGEN INLET EXTENSION KIT
Refer to the instruction sheet included with the kit itself.
4. ATTACH WATER CHAMBER AND HEATED BREATHING TUBE
The water chamber and heated breathing tube must be connected to
carry out the following setup and testing procedures.
If you have not been supplied with a reusable HC360 water chamber, you
can use an MR290 chamber instead.
6
5. SWITCH ON UNIT
Switch on the unit by pressing the On/O button.
6. WARM-UP
The unit will begin to warm up.
“Warm-up” symbol
7. READY FOR USE
The “Ready for use” symbol means that the system is ready for the
patient to use.
“Ready for use” symbol
7
ADVANCED SETTINGS
pl
ar
ru
tr
he
el
ro
When you see the “Warm-up” or “Ready for use” symbols, hold a
combination of three buttons (Up, Down and Mute) for 5 seconds, to
view and change advanced settings.
This button combination is for use by clinical engineering / technical
personnel only.
AIRVO 2 / myAIRVO 2 MODE
You can change the unit from “AIRVO” (hospital) mode to “myAIRVO”
(home / long-term care) mode, eg. for patients going home.
Contact Fisher & Paykel Healthcare for a myAIRVO 2 User Manual.
To change the mode:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up button to select myAIRVO 2.
Press the Mode button to confirm the change and/or move on to the
next screen. Note that the unit will reset itself if it is switched between
AIRVO 2 and myAIRVO 2 modes.
LANGUAGE
You can set the AIRVO 2 / myAIRVO 2 to one of 22 language settings:
EnglishNederlandsSvenskaPolski
DeutschPortuguês
EspañolDansk
FrançaisSuomiΕλληνικά
ItalianoNorsk
简体中文
繁体中文
日本語
[simp.]
[trad.]
Русский
עברית
Română
To change the language:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up and Down buttons to select the desired
language.
العربية
Türkçe
Press the Mode button to confirm the change and/or move on to the
next screen.
8
95
ENVIRONMENT SETTINGS (FOR DEFAULT MODE)
A clinician may change the “Environment Settings”, to customise
individual AIRVOs for dierent environments (eg. intensive care, general
care areas, emergency departments). The “Environment Settings”
chosen will put limits on the “Patient Settings” that the operator can
choose when in normal use.
This screen defines the “Environment Settings” for the AIRVO 2 when in
Default Mode (ie. non-“Junior Mode”).
Minimum dew-point
temperature (°C)
Minimum flow (L/min)The lowest flow that the operator will be able to
Maximum flow (L/min)The highest flow that the operator will be able
Maximum oxygen fraction (%) The highest oxygen fraction that the operator
95
Minimum oxygen fraction (%) The lowest oxygen fraction that the operator
Note that, for Oxygen display, this is a measurement only, not a control setting. The
operator changes the measured oxygen fraction by altering the AIRVO 2 target
flow setting and the flow of oxygen connected to the unit (e.g. from a flowmeter)
- there is no closed-loop control.
The lowest target dew-point temperature that
the operator will be able to select.
Possible Settings: 31, 34, 37 °C
If this is set to 31, the operator can select a target
dew-point temperature between 31 and 37. ie. 31,
34 or 37 (°C).
If the patient is tracheostomised, a clinician may
wish to set this value to 37, so that the operator
can only select a target dew-point temperature
between 37 and 37, ie. only 37 (°C).
Note: The maximum dew-point temperature
setting is always 37 °C in Default Mode.
select.
Possible Settings: 10 to 60 in increments of 5 L/min,
always less than or equal to Maximum Flow setting.
Example: If this is set to 10, the operator will be
able to select flows down to 10 L/min.
If this is set to 25, the operator will be able to
select flows down to 25 L/min.
to select.
Possible Settings: 10 to 60 in increments of 5 L/min,
always greater than or equal to Minimum Flow setting.
Example: If this is set to 60, the operator can
select flows up to 60 L/min.
If this is set to 35, the operator can select flows up
to 35 L/min.
may set the unit to.
Possible settings: 30 - 100% in increments of 5% O2.
The unit will alarm if the measured oxygen
fraction exceeds this value.
Note: Even if this ‘Maximum oxygen fraction‘
setting is set to 100%, any time the measured
oxygen fraction exceeds 95%, the oxygen reading
will pulse red and the device will beep.
may set the unit to.
Possible settings: 21 or 25% O2.
When set to 25% the unit will alarm if the
measured oxygen fraction is below this value. This
allows detection of oxygen being disconnected.
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the first setting.
95
Use the Up and Down buttons to change the setting,
then press the Mode button to progress to the next
setting.
Press the Mode button to confirm the change and/or move on to the
next screen.
9
95
100
ENVIRONMENT SETTINGS (FOR JUNIOR MODE)
This screen defines the “Environment Settings” for the AIRVO 2 when in
Junior Mode.
Junior Mode Enable/Disable
Enabled
Disabled
Dew-point
temperature (°C)
Minimum flow (L/min)The lowest flow that the operator will be able to
Maximum flow (L/min)The highest flow that the operator will be able
Maximum oxygen fraction (%)
95
Minimum oxygen fraction (%) The lowest oxygen fraction that the operator
Note that, for Oxygen display, this is a measurement only, not a control setting. The
operator changes the measured oxygen fraction by altering the AIRVO 2 target
flow setting and the flow of oxygen connected to the unit (e.g. from a flowmeter)
- there is no closed-loop control.
When this option is enabled (default), the
operator can enter Junior Mode from the Home
Screen, by holding the Mode button for 5
seconds.
When this option is disabled, entering Junior
mode is not possible.
Consider disabling this option if the unit will never
be used on pediatric patients.
The only dew-point setting in Junior Mode is
34 °C.
select.
Possible Settings: 2 to 25 in increments of 1 L/min,
always less than or equal to Maximum Flow setting.
If this is set to 10, the operator will be able to
select flows down to 10 L/min.
to select.
Possible Settings: 2 to 25 in increments of 1 L/min,
always greater than or equal to Minimum Flow setting.
If this is set to 15, the operator can select flows up
to 15 L/min.
The highest oxygen fraction that the operator
may set the unit to.
Possible settings: 30 - 100% in increments of 5% O2.
The unit will alarm if the measured oxygen
fraction exceeds this value.
Note: Even if this ‘Maximum oxygen fraction‘
setting is set to 100%, any time the measured
oxygen fraction exceeds 95%, the oxygen reading
will pulse red and the device will beep.
may set the unit to.
Possible settings: 21 or 25% O2.
When set to 25% the unit will alarm if the
measured oxygen fraction is below this value. This
allows detection of oxygen being disconnected.
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the first setting.
Use the Up and Down buttons to change the setting,
then press the Mode button to progress to the next
setting.
Press the Mode button to confirm the change and/or move on to the
next screen..
10
25
FLOW INCREMENT SETTINGS
This screen defines the “Flow Increment Settings” for the AIRVO 2 when
in either Default Mode or Junior Mode.
You can define the flow rate above which the increments are 5 L/min
and below which the increments are 1 L/min.
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up and Down buttons to change the setting.
25
Press the Mode button to confirm the change and/or move on to the
next screen.
Flows > 25 L/min will increment in steps of 5 L/min
Flows < 25 L/min will increment in steps of 1 L/min
DISINFECTION STOP-GATE SETTINGS
This screen defines the “Disinfection Stop-Gate Settings” for the
AIRVO 2 when in either Default Mode or Junior Mode.
Confirmation required
At startup, disinfection number and time
information is shown, confirmation from the
operator is required before progressing to
warmup
Confirmation not required (default)
At startup, disinfection number and time
information is shown before progressing to
warmup
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up and Down buttons to change the setting.
Press the Mode button to confirm the change and/or move on to the
next screen.
11
OXYGEN INPUT SETTINGS
This screen defines the “Oxygen Input Settings” for the AIRVO 2 when
in either Default Mode or Junior Mode.
The 100% setting is for hospital oxygen supplies, liquid oxygen or
standard bottled oxygen and is the default setting for AIRVO 2.
The 95% setting is for use with oxygen concentrators.
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up and Down buttons to change the setting.
Press the Mode button to confirm the change and/or move on to the
next screen.
TRANSPORT MODE SETTINGS
This screen defines the “Transport Mode Settings” for the AIRVO 2
when in either Default Mode or Junior Mode.
Transport Mode enabled
Transport mode may be activated by the operator.
(see “Transport mode” in the AIRVO 2 User
Manual)
Transport Mode disabled (default)
Transport mode may not be activated by the
operator..
To change the environment settings:
Hold the Up and Down buttons for 3 seconds to
“unlock” the setting.
Use the Up and Down buttons to change the setting.
Press the Mode button to return to the “Warm-up”/”Ready for use”
screen. You can now conduct the Performance/Acceptance checks..
12
3. ACCEPTANCE/PERFORMANCE CHECKS
This section contains performance checks which can be carried out on the AIRVO 2, however there is
no manufacturer requirement to carry out these checks on a routine basis. These checks test the basic
functions of the unit, the operation of the flow sensor and the audible alarm signal.
SENSOR CHECKS
To ensure quality and patient safety, Fisher & Paykel Healthcare undertakes stringent testing to each
and every unit manufactured. The AIRVO 2’s sensors, measuring temperature, flow and oxygen, have
been carefully designed to exacting criteria, and are calibrated and tested in our controlled work
environment to strict limits. Furthermore, the AIRVO 2 itself carries out regular self-checks during
normal use, comparing sensor readings against expected values.
Do not perform additional tests of internal sensor accuracy, as these often produce erroneous
results due to limitations of the external test environment and equipment used, particularly given the
temperature, humidity content and/or flow of the gases being delivered by the AIRVO 2.
ACCEPTANCE/PERFORMANCE CHECKS
The acceptance/performance checks should be conducted under the following ambient conditions:
Temperature: 22 ± 2°C, Humidity: 50 ± 5% RH.
The following equipment is required:
AIRVO 2 humidifierMR290 chamber
(or alternatively a
HC360 chamber)
Heated breathing tube
(from 900PT500 or
900PT501 kit)
Nasal interface
(OPT842, OPT844 or
OPT846)
A. HEATERPLATE TEST
1. Add 150 mL of room temperature (not hot) water to the water chamber and fit the chamber onto the
heater plate of the device. Fit the chamber tightly on to the chamber ports.
2. Connect the heated breathing tube to the Heated Breathing Tube Connection port. Connect the nasal
cannula interface to the heated breathing tube.
3. Turn on the device, by pressing the power button for 2 seconds. Warm-up bars will be displayed as the
unit warms up. Ensure the flow is set to 30 L/min.
4. Check that the “Ready for use” symbol (a “tick” or “check”) is displayed within 30 minutes.
B. “CHECK FOR LEAKS” TEST
After the “Ready for use” symbol is displayed, the “Check for leaks” sensor test can be tested as follows:
1. Remove the chamber completely from the unit.
2. Check that the display shows the “Check for leaks” error (in the appropriate language) and that the
audible alarm sounds, within 2 minutes.
3. Reconnect the chamber and check that this flashing display disappears, the audible alarm stops and the
display reverts back to the Warm-up/Ready-for-use screen.
13
C. “CHECK FOR BLOCKAGES” TEST
After completing the “Check for leaks” test, the “Check for blockages” test can be tested as follows:
1. Disconnect the cannula from the Heated Breathing Tube.
2. Completely block the end of the Heated Breathing Tube with your hand.
3. Check that the display shows the “Check for blockages” error (in the appropriate language) and that the
audible alarm sounds, within 30 seconds.
4. Unblock the end of tube and check that this flashing display disappears, the audible alarm stops and the
display reverts back to the previous display.
5. Reconnect the cannula to the Heated Breathing Tube.
D. “CHECK TUBE” TEST
After completing the above flow tests, the Tube Missing alarm can be tested as follows:
1. Remove the Heated Breathing Tube (pull the blue sleeve up first).
2. Check that within 10 seconds the display flashes the Breathing Tube symbol and the alarm sounds.
3. Refit the Heated Breathing Tube, check the alarm stops and that the display reverts back to the previous
display.
Note: If any of the tests above fail, please contact your Fisher & Paykel Healthcare representative.
14
4. SERVICING
AIRVO 2 and myAIRVO 2 humidifiers do NOT require routine servicing or calibration.
The only checks that can be carried out are the Acceptance/Performance Checks in the previous section,
and the Electrical Safety Test detailed below.
WARNING
UNDER NO CIRCUMSTANCES SHOULD THE
AIRVO 2 BE OPENED OR ANY OF THE SIX
FASTENING SCREWS ON THE UNDERNEATH
SIDE OF THE DEVICE BE LOOSENED. OPENING
THE UNIT WILL AFFECT THE OXYGEN
SEALS INSTALLED INSIDE, WHICH WILL
COMPROMISE THE SAFETY OF THE DEVICE.
ELECTRICAL SAFETY TESTS
To test for electrical safety perform the following electrical safety tests and any others required by local
regulations.
InspectionCheck the power cord for damage - cuts, stretching, wear, adequate cable restraint, bent
Insulation ResistanceUse a 500 VDC insulation tester to measure the resistance between the mains plug phase pin
pins. Replace with F&P approved cord if necessary.
and the heaterplate* - it should be > 10 Mohm. Repeat test from the mains plug neutral pin to
the heaterplate*.
* Note: The exposed surface of the heaterplate is anodised (high resistance). Contact MUST
be made to the bottom lip of the heaterplate at the front of the device to make proper
connection - depress the finger guard and slip the tester probe beneath the heaterplate to
ensure contact to unanodised aluminium.
STORAGE AND DISPOSAL
Refer to AIRVO 2 User Manual.
15
5. SPARE PARTS
5.1 POWER CORDS
900PT410AZ (Aus/NZ)
900PT410EW (European)
900PT410UK (UK)
900PT410US (US/Canada)
900PT410KR (Korea)
900PT410JP (Japan)
16
5.2 FOR AIRVO
Non-Return Valve (2-pack)
Serial Port Cover (10-pack)
900PT911
900PT912
AIRVO 2 Filter Holder
900PT913
Air Filter (2-pack)
900PT422
Oxygen Inlet Kit
900PT407
AIRVO Outlet O-Ring (10-pack)
5.3 FOR DISINFECTION
Disinfection Kit
Disinfection Filter
Cleaning Sponge Stick
Disinfection Storage Cover
900PT408
900PT600
900PT601
(2-pack)
900PT602
(20-pack)
900PT603
(20-pack)
17
5.4 HARDWARE
900PT405
Pole Mounting Tray
900PT421
Mobile Pole Stand
900PT423
Hook for 900PT421
900PT426
Plastic Basket
900PT427
Oxygen Bottle Holder
900PT409
3M Dual Lock Pad (4 pairs)
18
APPENDIX A: IEC60601-1-2 EMC TABLES
14of17
Appendix A: IEC60601-1-2 EMC tables
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B Applicable for countries with 100-115V and 220-240V mains voltage.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes
15of17
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should ensure that it is used in
such an environment.
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
±2 kV for power
supply lines
±2 kV
Electrical fast
transient/burst
±1 kV for input/output See note 2 below
IEC61000-4-4
lines
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV differential mode ±1 kV
Surge
±2 kV common mode ±2 kV
IEC 61000-4-5
Mains power quality should be that of a typical
commercial or hospital environment.
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 0.5 cycle for 0.5 cycle
40 % UT 40 % UT
(60 % dip in UT) (60 % dip in UT)
for 5 cycles for 5 cycles
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
70 % UT 70 % UT
IEC 61000-4-11 (30 % dip in UT) (30 % dip in UT)
for 25 cycles for 25 cycles
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 5 sec for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the device requires continued operation during
power interruptions, it is recommended the device
be powered from an uninterruptible power supply
or a battery.
3 A/m 3 A/m
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE 1: UT is the a.c. mains voltage prior to application of the test level.
NOTE 2: This testing is not necessary for the safe operation of the device.
19
16of17
Guidance and manufacturer's declaration - electromagnetic immunity
17of17
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in such
an environment.
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √
P
P 80 MHz to 800 MHz
d = 1.2 √
P 800 MHz to 2.5 GHz
d = 2.3 √
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF
compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,ª should be less than the compliance
level in each frequency range
Interference may occur in the vicinity of equipment marked with
the following symbol:
b
.
Recommended separation distances between portable and mobile RF communications equipment
and the device
The device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the device as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output power 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
of transmitter
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
d = 1.2 √
Pd =1.2 √P d = 2.3 √P
20
m
APPENDIX B: USER INTERFACE FLOW CHARTS
AIRVO 2
Junior Mode
New Target
Settings
Wait for
time out 60s
Oxygen
is left on
Disconnect
Oxygen
Disinfection:
Hold
5 s
Hold
3s
Last
#4
17 days
14 hours
ago
Last
Disinfection:
#5
2 days
4 hours
ago
Environment Settings
Hold
Press
Hold
AIRVO 2
V6.72
5 s
AIRVO 2
Transport
Mode
Hold
5 s
5 s
[Machine Data]
English
Hold
3s
Press
English
Press
Hold
3s
3760
3110
90
60
37
10
21
Hold
Hold
3s
After 3
scrolls
3s
Press
Press
Hold
3s
9090
6025
37
3131102
21
90
21
90
21
onitor patient O2
Patient Settings
Press
Press
90
25
34
2
21
34
60
30
10
Hold
3s
Press
34
34
21
Hold
3s
Press
30
Hold
95%
100%
Hold
3s
Press
3s
Press
100%
Hold
3s
19:23
timer
reaches
00:00
Transport
Mode
Press
Non-Transport
Mode
Press
Hold
5 s
Wait for
19:55
Hold
5 s
time out 60s
21
DISINFECTION
1000 hours use
- Replace lter
Press
Press
Any key
PressPress
Now
Later
Press
Now
Later
Filter timer
remains at 1000
hrs +
Time
Filter timer is
reset to 0 hrs
Wait for
time out 60s
22
APPENDIX C: DEFAULT VALUES
The following values are default non-alarm settings set by the factory:
ParameterModelModeValueUnit
Default set dewpoint
temperature
Max set dewpoint
temperature
Min set dewpoint
temperature
Default set flowAIRVO 2Default30L/min
Max set flowAIRVO 2 or myAIRVO 2 Default60L/min
Min set flowAIRVO 2 or myAIRVO 2 Default10L/min
Default upper oxygen limit
Max upper oxygen limit
Min upper oxygen limitAIRVO 2 or myAIRVO 2 Default or Junior30%
Default lower oxygen limitAIRVO 2 or myAIRVO 2 Default or Junior21%
Max lower oxygen limitAIRVO 2 or myAIRVO 2 Default or Junior25%
Min lower oxygen limitAIRVO 2 or myAIRVO 2 Default or Junior21%
LanguageAIRVO 2 or myAIRVO 2 Default or JuniorEnglish
Flow Increment CrossoverAIRVO 2 or myAIRVO 2 Default or Junior25L/min
Oxygen Input
Disinfection Stop-GateAIRVO 2Default or Junior
Transport ModeAIRVO 2Default or JuniorDisabled
Day/Night modemyAIRVO 2Default or JuniorDay
AIRVO 2 or myAIRVO 2 Default37°C
AIRVO 2 or myAIRVO 2 Junior34°C
AIRVO 2 or myAIRVO 2 Default37°C
AIRVO 2 or myAIRVO 2 Junior34°C
AIRVO 2 or myAIRVO 2 Default 31°C
AIRVO 2 or myAIRVO 2 Junior34°C
myAIRVO 2Default25L/min
AIRVO 2 or myAIRVO 2 Junior15L/min
AIRVO 2 or myAIRVO 2 Junior25L/min
AIRVO 2 or myAIRVO 2 Junior2L/min
AIRVO 2Default or Junior95%
myAIRVO 2Default or Junior90%
AIRVO 2Default or Junior100%
myAIRVO 2Default or Junior90%
AIRVO 2Default or Junior100%
myAIRVO 2Default or Junior95%
Confirmation not required
23
APPENDIX D: TROUBLESHOOTING GUIDE
This Troubleshooting Guide is intended for technical users, including clinical/biomedical engineers and
technical personnel, of the AIRVO™ 2 humidifier. It applies to all AIRVO 2 humidifiers from lot numbers
140910 and above.
If this troubleshooting guide does not resolve your issue, please contact your local Fisher & Paykel
Healthcare representative.
D.1 AIRVO does not turn on
A. Press and hold the ON/OFF button for at least 2 seconds.
B. Is the AIRVO 2 plugged into mains power?
C. Is the power cord securely inserted into the back of the AIRVO 2?
D. Is the power cord damaged?
• If yes, replace the damaged cord. See Section 5.1 for a 900PT410xx replacement power cord.
• If no, try using another power cord.
E. Connect the AIRVO 2 into another power outlet.
F. Connect a different electrical device into the same power outlet. Turn on the device to confirm that the
power outlet is working.
G. The AIRVO 2 may be ‘on’ with a broken display.
Turn the AIRVO 2 on without the heated breathing tube and check that the audible alarm activates.
D.2 Power out (black screen)
The auditory alarm will sound for at least 120 seconds.
The most likely cause is a dislodged or disconnected power cord.
A. Please follow the instructions in Section D.1.
Note: Press “audio pause” button to permanently silence the alarm ().
The device will not automatically restart.
D.3 “Check tube”
Fig. 1
or “E38“
A. Is the heated breathing tube attached correctly?
• Even if it appears to be, unplug and reconnect the heated breathing
tube.
B. Is the heated breathing tube visibly damaged?
• Check the electrical pins and the tube itself.
C. Try using a new heated breathing tube.
Figure 1
24
D.4 “Check for blockages”
Fig. 2
or “E121”
D.4.1 WATER CHAMBER AND NON-RETURN VALVE
A. Have the silicone flaps of the non-return valve, found inside the left-hand
chamber port, been displaced
Fig. 3
?
• If yes, return them to the correct position using a non-sharp tool, such
as a pair of non-sharp tweezers
Fig. 5
.
Note: If the Non-return valve is damaged or missing, replace with part
900PT911. Upon replacement, ensure the spine is sitting vertically
Fig. 5
.
If placed horizontally, this may cause the bottom flap to open due to
gravity
Fig. 4
. This may cause both “Check for leaks” and “Check for blockages”
warnings.
B. Is the MR290 water chamber overfilled above the black line?
• If yes, replace with a new water chamber. Contact your local Fisher &
Paykel Healthcare representative about the faulty chamber.
D.4.2 HEATED BREATHING TUBE
A. Is the heated breathing tube visibly blocked or kinked
Fig. 6
?
D.4.3 PATIENT INTERFACE AND AIRVO MODE
A. Is the patient interface visibly blocked or kinked?
B. Should the unit be in Junior mode
Fig. 7
?
• If the AIRVO is in Default mode and the 900PT531 Junior tube is used
with the OPT316 and OPT318 cannula interfaces it may generate a
“Check for blockages” alarm.
See Appendix E for identification of the Default and Junior Tube and
Chamber Kits, according to their labels.
C. Are you using an unsuitable cannula?
• The OPT312 and OPT314 cannot be used with the AIRVO 2.
See the User Manual for information regarding patient interfaces.
D.4.4 AIR FILTER
A. Is the air filter significantly discolored/dirty?
• Replace with part 900PT913.
Note: A prompt
Fig. 8
for filter change will occur at the start of the Disinfection
Cycle once the AIRVO 2 has counted 1,000 hours of use. Choose ‘Now’ or
‘Later’
Fig. 9
by using the “up” or “down” buttons and press the “mode” button
() to confirm. Selecting ‘Now‘ will zero the counter. Selecting “Later“
will activate the prompt at the start of the next Disinfection Cycle.
B. Is there a foreign object blocking the air filter or filter holder?
D.4.5 CONDENSATION
Please see Section D.12.
D.4.5 ALTITUDE
A. The myAIRVO 2 is designed to operate at an altitude below 2,000 meters.
Figure 2
Figure 3
Figure 6
Figure 4
Figure 5
25
Figure 7
Figure 8
Figure 9
D.5 “Check for leaks”
Fig. 10
or “E122”
The most likely cause is a missing water chamber or the existing chamber has
not been pushed into place correctly.
D.5.1 WATER CHAMBER
A. Is the water chamber fitted correctly? Even if it appears to be:
• Remove the water chamber.
• Push the chamber on firmly, until the finger guard “clicks” into place
Fig. 11
.
Warning: The heater-plate and base of the water chamber may be hot.
D.5.2 HEATED BREATHING TUBE
A. Is the heated breathing tube attached to the device correctly?
Even if it appears to be:
• Disconnect the heated breathing tube.
• Check that the black O-ring is in place
Fig.12
.
If the O-ring is damaged or missing, replace with part 900PT408.
• Reconnect the heated breathing tube.
B. Confirm that the heated breathing tube is not visibly damaged.
D.5.3 PATIENT INTERFACE
A. Is the patient interface correctly fitted to the heated breathing tube?
• Even if it appears to be, disconnect and reconnect the patient interface.
It should make a “click” sound when it is connected properly.
B. Should the unit be in Default (adult) mode?
• If the AIRVO is in Junior mode and the 900PT501 Default tube is used
with the OPT842/44/46/70 or RT013 interfaces, it may generate a
“Check for leaks” alarm.
See Appendix E for identification of the Default and Junior Tube and
Chamber Kits, according to their labels.
D.5.4 AIR FILTER & FILTER COVER
A. Is the air filter and filter cover (at the back of the device) correctly fitted, as
per the User Manual?
Figure 10
Figure 11
Figure 12
26
D.6 “O2 too low”
Fig. 13
The measured oxygen level has fallen below the allowed limit.
A. Adjust the level of oxygen from the oxygen source as necessary, i.e. increase
the oxygen flow rate through the oxygen flow meter.
B. Is the oxygen source (wall or cylinder flow meter) turned on?
C. Is the oxygen source empty or faulty?
D. Is the “AIRVO 2 oxygen inlet kit”
Fig. 14
installed correctly, as per the
instructions included with part 900PT422 and confirmed that there are no
kinks in the “AIRVO 2 oxygen inlet kit” oxygen tubing?
E. Is the oxygen source tubing correctly and securely fitted to the AIRVO 2?
F. Allow the device to sufficiently warm up; rapid changes in temperature can
affect the sensor.
G. Is the minimum oxygen limit set to 25%?
• A prompt will appear with an option to change this lower limit to 21 %.
Select “Yes” or “No” by using the “Up“ and “Down” buttons. Press the
“mode” button () to confirm selection
Fig. 15
.
See Section 2 - Advanced Settings to change this lower oxygen limit.
Figure 13
D.7 “O2 too high”
Fig. 16
The measured oxygen level has risen above the allowed limit.
A. Adjust the level of oxygen from the oxygen source as necessary, i.e. decrease
the oxygen flow rate through the oxygen flow meter.
B. See Section 2 - Advanced Settings to change this lower oxygen limit.
D.8 “Cannot reach target flow”
Fig. 17
A. Press the “mode” button () to continue normal operation at a lower
(maximum achievable) flow rate.
B. Is the target flow setting too high for the patient interface?
• Check the swing tag/User Manual for the appropriate flow range for each
patient interface.
Note: If the AIRVO 2 cannot reach the target flow setting, it will automatically
select a maximum achievable flow rate and prompt the user to press the
“mode” button () to confirm.
C. Follow steps in Section D.4 — “Check for blockages”.
D. Is the altitude above 2,000 m?
The AIRVO 2 is designed to operate at an altitude below 2,000 meters.
Figure 14
Figure 15
27
Figure 16
Figure 17
D.9 “Cannot reach target temperature”
Fig. 18
The most likely cause is operating the AIRVO 2 at a high flow rate in a cold room.
Consider decreasing the target flow setting.
A. Press “mode” button () to continue.
Note: The humidity level may be compromised.
B. Is the ambient room temperature below 18 °C (64 °F)?
• If yes, proactive management of condensation may be required.
See Section D.12 on prevention and management of condensation.
Figure 18
D.10 “Check water”
Fig. 19
A. Is the water bag empty?
If yes, refill or replace the water bag and press the “mode” button () to
reset the alarm.
B. Is the water chamber empty?
• If yes, replace the water chamber as it may be damaged.
Warning: The heater-plate and base of the water chamber may be hot.
C. Is there a kink in the fluid line, preventing water from flowing into the
chamber?
D. Open the vent cap near the water bag spike. This allows the pressure to
equalize, letting the water flow into the water chamber.
D.11 “Check operating conditions”
Fig. 20
This alarm may be caused by a sudden change in ambient room temperature,
e.g. storing the unit in a cold place, then using it in a warm place.
A. Is the ambient room temperature less than 10 °C (50 °F) or greater than
30 °C (86 °F)?
B. Leave the unit running for 30 minutes.
Switch the unit off, then restart.
Figure 19
28
Figure 20
D.13 Exxx
Fig. 23
A. Follow the instructions in Appendix F if a fault with an error code is
displayed on the AIRVO screen.
D.12 Condensation
D.12.1 PREVENTION OF EXCESSIVE CONDENSATION
A. Is the AIRVO 2 being used in ambient conditions between 18 – 28 °C(64 - 82 °F)?
B. Is the AIRVO 2 placed below head height
C. Is there a local source of cooling acting on the heated breathing tube?
D.12.2 CONDENSATION MANAGEMENT
A. Implement a system to check the heated breathing tube for condensate
B. If condensation is present, drain it back into the water chamber
C. If condensate persists, consider turning the target temperature down.
D.12 Condensation
D.12.1 PREVENTION OF EXCESSIVE CONDENSATION
A. Is the AIRVO 2 being used in ambient conditions between 18 – 28 °C
(64 - 82 °F)?
• If the room is less than 18 °C (64 °F), condensation is more likely to
occur.
B. Is there a local source of cooling acting on the heated breathing tube?
• A fan to cool the patient,
• An air-conditioning unit, vent or an open window?
• Are you able to remove or minimize these sources of cooling, e.g.
redirect the fan, cooling the patient, away from the heated breathing
tube?
D.12.2 CONDENSATION MANAGEMENT
A. Implement a system to check the heated breathing tube for condensate
regularly.
B. Is the AIRVO 2 placed below head height
Fig. 21
?
• This will allow condensate to drain towards the water chamber, away
from the patient.
C. If condensation is present, drain it back into the water chamber
Fig. 22
:
• Disconnect the patient interface from the heated breathing tube.
• Drain the tube by lifting the patient end of the tube, allowing the
condensate to run into the water chamber.
• At higher target flow rates, it may be necessary to first reduce the target
flow rate to 30 L/min or below, to ensure the condensate drains into the
water chamber.
D. If condensate persists, consider turning the target temperature down.
• A lower target temperature will decrease the humidity output of the
AIRVO 2, decreasing the level of condensation.
Note: The temperature and humidity level delivered to the patient will also
be reduced.
Figure 21
Figure 22
29
Figure 23
APPENDIX E: TUBE AND CHAMBER KIT LABELS
AIRVO
TM
2
900PT531
Tube and Chamber Kit (Junior)
PEDIATRICINFANT
Single
Use
Fisher & Paykel Healthcare Ltd. 15 Maurice Paykel Place, East Tamaki,
Auckland 2013, New Zealand
Cordwallis Park, Clivemont Road, Maidenhead, Berkshire, SL6 7BU, UK
Rx Only
0123
Fisher & Paykel Healthcare Ltd. Unit 16,
Made in New Zealand
Label for default (adult)
Tube and Chamber Kit
Label for Junior
Tube and Chamber kit
APPENDIX F: ERROR FLOWCHART
The following pages provide troubleshooting advice for fault / error / “E” codes that may appear during use of the AIRVO 2.