Fieldpiece SC77 User Manual

5 (1)
Maintenance
Clean the exterior with clean dry
cloth. Do not use liquid.
Battery replacement: When the
multimeter displays " " the battery must be replaced. Disconnect and unplug leads, turn meter off, and remove the battery cover. Replace the battery with a NEDA type 1604 9V bat­tery.
Symbols used:
Caution, risk of electric shock
Caution, refer to manual.
Ground
Double insulation
For your safety...
General: Disconnect the test leads
All voltage tests: For model SC77
all voltage ranges will withstand up to 600VDC or 600VAC rms. Do not apply more than 600VDC or 600VAC rms.
AC tests: Disconnect the meter
from the circuit before turning any inductor off, including motors, trans­formers, and solenoids. High voltage transients can damage the meter beyond repair. Do not use during elec­trical storms.
Limited warranty
This meter is warranted against defects in material or workmanship for one years from date of purchase. Fieldpiece will replace or repair the defective unit, at its option, subject to verification of the defect.
This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alter­ation, or unreasonable use of the instrument.
Any implied warranties arising from the sale of a Fieldpiece product, including but not limited to implied war­ranties of merchantability and fitness for a particular purpose, are limited to the above. Fieldpiece shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim of such damage, expenses, or economic loss.
State laws vary. The above limita­tions or exclusions may not apply to you.
Obtaining service
Return any defective SC77 to Fieldpiece for service. For warranty service, please provide a proof of pur­chase. Contact Fieldpiece for out of warranty repair charges. Contact infor­mation can be found at our website www.fieldpiece.com.
OPERATOR’S MANUAL
400
DISCONNECT TESTLEADS
340
SC77
µ
Auto
4000m
400m
HOLD
NCV
m A K M Ω
m V V
APO HOLD n m m F AC DC
Works with Fieldpiece accessory heads!
AUTO RANGING T-RMS DIGITAL CLAMP METER
Model SC77
WARNINGS
DISCONNECT AND UNPLUG TEST
LEADS
before opening case.
TEST NCV FUNCTION ON KNOWN
LIVE WIRE before using.
DO NOT APPLY VOLTAGE greater than 30VAC or 60VDC to the thermocou­ple or the jacks when the rotary dial is on °F.
REMOVE
THE THERMOCOUPLE
when taking voltage measurements.
DISCONNECT
THE TEST LEADS when taking temperature measure­ments.
DO NOT APPLY VOLTAGE TO THE JACKS when the rotary dial is on microamps. Even low voltages can cause a current overload and potentially dam­age the circuit.
!
!
Works with Fieldpiece accessory heads
Connect your Fieldpiece accessory head to SC series meter through included deluxe silicone leads. Use 400mVDC or 4000mVDC range for most heads.
SC77 connected to
ARH4 through leads.
Available Fieldpiece accessory heads
There is a Fieldpiece accessory head available for just about any job. There are heads to measure tempera­ture, superheat, subcooling, RH%, wet bulb, dew point, vacuum (microns of mercury), manometer (inches of water column), amps AC&DC, high voltage,
CO, CO2, air velocity, and many more.
Non-contact voltage
With the NCV tab on the tip of the clamp close to an AC voltage, press the NCV button. The NCV LED will light and the beeper will beep. The NCV function is sensitive enough to detect 24VAC on thermostats.
Hi voltage indicator
In any VAC/VDC range, when you touch a voltage greater than 30V, the beeper will beep and the red Hi-V LED will blink. BE CAREFUL!
Current measurements
The fixed jaw is longer than the moveable jaw to make it easier to select just one wire from a bundle. With the jaws closed, separate one wire using the long fixed jaw. Slide it to the corner where the two jaws meet. Then open the jaws to let it in. You can select the wire without having to hold the jaws open.
Microamps
For measuring the flame diode cur­rent in a heater control.
Capacitance
For motor-start and motor-run capacitors. Disconnect the capacitor from power first. Short the terminals to discharge the capacitors. Disconnect any resistors that might be between the terminals of the capacitor. The SC77 displays a “dsc” symbol when capacitor needs discharging.
Temperature
Plug any K-type thermocouple directly into the meter to measure tem­perature. Temperature measurement will maintain good accuracy in fast changing environments. One thermo­couple is included. No adapter is required.
Field Temp. calibration
For accuracies of ±1° calibrate the SC77 to a known temperature. A glass of stabilized ice water is very close to 32°F (0°C) and is usually very conven­ient.
1. Select the °F range on the SC77.
2. Remove back case. Hold the bat-
tery in place with a rubber band.
3. Stabilize a large cup of ice water by
stirring.
4. Immerse the probe and let it stabi-
lize. Continue stirring.
5. Adjust VR3 (lower right corner of
PCB) to get close to 32°F (0°C) then adjust VR4 (below VR3) to get within 0.1°F/C of 32°F (0°C).
6. To calibrate in °C, close the jumper
that is to the left of VR3.
Display ºC or ºF
Remove back of meter, locate jumper on lower right corner of PCB (just below VR3) and close jumper to read temperature in °C.
Using &storing test leads
Because the wire insulation is sili­cone they will stay flexible in cold weather and will not melt if bumped by a soldering iron.
Use the single test probe holder on clamp to make voltage testing easy.
For convenient lead storage, wrap the leads as shown.
!