Fieldpiece LT16A User Manual

DIGITAL MULTIMETER Model: LT16A
OPERATOR’S MANUAL
General description
The LT16A is an excellent true RMS digital multi­meter for testing motors and electrical equipment. The LT16A meter measures current, resistance, voltage, capacitance, frequency, continuity, and phase rotation. Test leads store within the meter and can connect the LT16A to Fieldpiece accesso­ry heads. The body is made out of durable ABS plastic and comes with an alligator clip test lead for measuring phase rotation. It also has a backlight.
User maintenance
Battery Replacement: When the multimeter dis­plays the " " the battery must be replaced to maintain proper operation. To prevent electrical shock, turn off the meter and disconnect leads before removing the back cover.
Fuse Replacement: When only certain ranges quit working, check the fuse. Open the case and replace according to the Overload Protection chart.
All other repairs must be performed by a Fieldpiece service center
For your safety...
General: Inspect the test leads for damage to the insulation or exposed metal. Replace if suspect. Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating materi­al. When disconnecting from a circuit, disconnect the "RED" lead first, then the common lead. Work with oth­ers. Use one hand for testing. Turn off power to the cir­cuit under test before cutting, unsoldering, or breaking the circuit. Keep your fingers on the plastic and behind the ridge of the probes.
All Voltage Tests: All voltage ranges will with­stand up to 600VAC or 600VDC. Do not apply more than 600VDC or 600VAC.
AC Tests: Disconnect the meter from the circuit before turning any inductor off, including motors, transformers, and solenoids. Hi voltage transients can damage the meter beyond repair. Do not use during electrical storms.
Safety: Designed to meet IEC 61010-1 (EN61010-1), CATIII 600V, Class II, pollution deg.2, indoor use. CE UL61010-1. C-Tick certified.
Symbols used:
Caution, refer to manual.
Ground
Double insulation
Overload protection
Service
Call Fieldpiece Instruments for one-price-fix-all warranty service pricing. Send check or money order for the amount quoted. Send the meter freight prepaid to Fieldpiece Instruments. Send proof of date and location of purchase for in-warranty serv­ice. The meter will be repaired or replaced, at the option of Fieldpiece, and returned via least cost transportation.
www.fieldpiece.com
CAT.III
MAX
600V
15SEC FOR
200mV
RANGE
MAX
500V 3 PHASE
600V FOR 3O SEC
FUSED
2A
MAX
V
MF
D
V
2A
MF
D
A
A
TRUE RMS PHASE ROTATION AUTO-OFF CAPACITANCE
LT1 6A
L1
L2
L3
MAX
MIN
General specifications
Auto-off: off after 25 minutes to extend battery. Indicators: Continuity beeper (<100Ω). Low bat. Overrange: "OL" or “-OL” is displayed. Operating environment: 32 to 122°F (0 to 50°C)
<70%RH
Storage environment (with batteries removed): -4 to
140°F (-20 to 60°C) <80%RH
Altitude: 6561.7 feet (2000m). Battery life: 150 hours typical. Battery type: 9V NEDA 1604 type Stated accuracies: 74°F±8°F (23°C±4°C), 75%RH Temperature coefficient: 0.1 x (specified accuracy)
1°F/°C (32 to 66°F (0-19°C), 82 to 122°F (28-50°C)).
MIN/MAX
Record the minimum or maximum reading during a measurement. To exit the MIN/MAX function, hold the but­ton for more than 2 seconds.
Backlight
Always be very careful when testing in the dark. You should always have a separate light on the area you are testing. The backlight will shine for 4.5 minutes when the backlight button is pressed.
True RMS
Digital multimeters use one of two types of AC sensing. The most common is average sensing, normalized to a true RMS value of a sine wave. The other is true RMS sensing. The actual true RMS value is sensed for a wave form with­in the limits of the crest factor. True RMS is needed to cal­culate power. Either sensing method will give the same results on a clean sine wave but they may differ on a non­sinusoidal waveform.
DC voltage
Ranges: 200mV, 2000mV, 20V, 200V, 600V Resolution: 0.1mV
Accuracy: 0.5%±1 Input impedence: 10MΩ
AC voltage True RMS (50Hz-500Hz)
Ranges: 200mV, 2000mV, 20V, 200V, 600V Conversion: True RMS
Crest factor: Less than or equal to 3 Resolution: 0.1mV Accuracy: 1.5%±5(200mV to 20V ranges),
2%±5(200V, 600V ranges)
Input impedence: 10MΩ
DC Current (through meter)
Ranges: 200μA, 20mA, 200mA, 2A Resolution: 0.1μA
Accuracy: 1.0%±1 (200μA to 200mA ranges), 2.5%±1
(on 2A range)
Voltage burden: 800mV
AC Current True RMS (through meter)
Ranges: 200μA, 20mA, 200mA, 2A Crest factor: Less than or equal to 3
Resolution: 0.1μA Accuracy: 1.5%±5 (200μA to 200mA ranges), 3.0%±5
(on 2A range)
Voltage burden: 800mV
Frequency (autoranging)
Range: 10Hz to 40KHz Resolution: 1Hz
Sensitivity: 3.5V rms min Accuracy: 0.1%± 3
Resistance (ohms)
Ranges: 200Ω, 2kΩ, 200kΩ, 20MΩ, 2000MΩ Resolution: 0.1Ω Accuracy: 1.0%±4 (200Ω to 200kΩ ranges),
2.0%±4 (20MΩ range), (5.0%-10)+10 (2000MΩ range)
Open circuit voltage: 0.3VDC typical, (3.0VDC
on 200Ω and 2000MΩ range)
Diode test
Accuracy: 1.5%±3 Test current: Approx. 1.0mA Open circuit voltage: 3.0VDC typical
Capacitance
(MFD)
Ranges: 200μF, 2kμF, 20kμF Resolution: 0.1μF Accuracy: 4%±10 Test frequency: 21Hz Test voltage: <3.0V
Phase rotation
Range: 80VAC to 500VAC (45Hz to 450Hz)
This function enables you to connect 3-phase power to the correct leads of a motor to insure the motor turns in the intended direction. The terminals the motor are marked L1, L2, and L3. The wires supplying power are not marked. Connect meter jacks marked L1, L2, and L3 to the power wires in any order. Make sure all three leads are connected (L1, L2, and L3 will be “on”). “OK” indicates “for-
ward”. Connect L1 on the meter to L1 on the motor, L2 on the meter to L2 on the motor, etc. The motor will turn in the direction designed. If you get “ER” (reverse), swap any two leads. It should then say “OK.”
Limited warranty
This meter is warranted against defects in material or workmanship for one year from date of purchase. Fieldpiece will replace or repair the defective unit, at its option, subject to verification of the defect.
This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alteration, or unreasonable use of the instrument.
ANY IMPLIED WARRANTIES ARISING OUT OF THE SALE OF A FIELDPIECE INSTRUMENT’S PRODUCT, INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE ABOVE. FIELDPIECE SHALL NOT BE LIABLE FOR LOSS OF USE OF THE INSTRU­MENT OR OTHER INCIDENTAL OR CONSEQUEN­TIAL DAMAGES, EXPENSES, OR ECONOMIC LOSS, OR FOR ANY CLAIM FOR SUCH DAMAGE, EXPENS­ES, OR ECONOMIC LOSS.
State laws vary. The above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
>200mV range 600VAC/DC rms
Re si sta n ce 500VAC/DC rms Diode Test 500VAC/DC rms Continuity 500V AC/DC rms
Capacitance
200mV range
VAC/DC Phase Rot.
AAC/DC
600VAC/DC rms
for 15 sec
600VAC/DC rms
for 30 sec
2A/600V fuse
(6.35X25.4mm)
model RFM70
0.25A/500V fuse (6.3X32mm)
model RFM66
!
v14
V
MF
D
V
A
A
VW
COM
How To Use Your Multimeter
For DC voltage and currents, set the meter to the DC parameter instead of AC as shown to the left. For all ranges and functions choose range just above value
you expect. If display reads "OL” or “-OL" (overload), select a higher range. If display shows less than three numbers, select a lower range for better resolution.
Voltage Amps Microamps Resistance & Continuity Beeper
V
MF
D
V
A
A
COM
2A
V
MF
D
V
A
A
COM
2A
V
MF
D
V
A
A
COM
V
W
Optional accessory heads
Fieldpiece accessory heads convert the desired parameter into a millivolt signal. For example, the ARH4 converts one °F into one milli-volt DC. The multimeter must then be set to read millivolts DC. One exception is the ACH4 current clamp which must be set to read AC millivolts. Any digital multi­meter with these scales can be used in conjunction with Fieldpiece Accessory Heads.
Use Fieldpiece deluxe silicone test leads (model# ADLS2) with removable probe tips to use the accessory heads with the LT16A.
AirVelocity &Temperature Head
English
READLOBATT
ON
LCDX 100
Average
(16sec)
Metric
Real time
AAV3
AUTO-
OFF
English Metric
Ft/min M/s
KM/hrMPH ºF ºC
OFF
V
MF
D
V
A
A
COM
MFD
AUTO­OFF
T/C Cal
STABLE
Pressure
Temp
SHor SC
ON
LOBATT
Air Conditioner Superheat andSubcooling
ENGLISH
METRIC
R22
R410A
Set ATM
ASX14
SH
SC
Frequency AC Current Clamp Capacitance (MFD)
V
MF
D
V
A
A
COM
VW
V
MF
D
V
A
A
L1
L2
L3
Phase Rotation Diode Test Accessory Heads
V
V
COM
V
Ω
A
D
MF
Ω
V
V
V
COM
ACH4
!
400AAC MAX
A
A
D
MF
A
0 0
4 V 0
0 3
1AAC / 1mVAC
Clamp
AC Current
AT.III
C
P
M
A L
C
V
A
D
MF
V
300V
CLAMP
CAT.III
A
400A
AC Current Clamp
1AAC / 1mVAC 400AAC MAX
ACH4
AUTO OFF
!
ARH4
COM
V
Ω
A