Extreme Networks Anritsu User Manual

Anritsu User’s Manual
Using the Anritsu Spectrum Analyzer with
Summit® WMScanner
TTaabbllee ooff CCoonntteennttss
1. Overview ......................................................................................................................... 2
2. Recording Real-Time Measurement Data With the Anritsu MS2711 HHSA ................ 2
2.1 Configure the Anritsu MS2711 HHSA ..................................................................... 2
2.2 Configure Summit WMScanner................................................................................ 3
3. WVC Anritsu MS2711 HHSA Logfiles ......................................................................... 6
3.1 Overview ................................................................................................................... 6
3.2 WVC Anritsu MS2711 HHSA Logfile Format ........................................................ 6
3.2.1 WVC Anritsu MS2711 HHSA Logfile Header Section .................................... 6
Extreme Networks, Inc. 3585 Monroe Street Santa Clara, California 95051 (888) 257-3000 (408) 579-2800 http://www.extremenetworks.com
Published: December 2009 Part Number: 100361-00 Rev 01
Anritsu User’s Manual - Page
1 of 13.
1. Overview
This manual explains how to use the Anritsu MS2711 in conjunction with Summit® WMScanner. The Anritsu MS2711 Hand-Held Spectrum Analyzer (HHSA) is a battery-powered, field-operable measurement tool with a real-time serial interface. The Anritsu HHSA is capable of measuring radio signals from 100 kHz to 3 GHz with a measurement range of –90 dBm to +20 dBm. Additionally, the Anritsu HHSA has a power monitor mode whereby an RF detector may be attached to measure specific power levels. Both spectrum analyzer mode and power monitor mode are fully supported within Summit WMScanner, enabling you to record measurement data in real-time for either mode.
The serial interface on the Anritsu HHSA makes it easy for Summit WMScanner to handle all the interaction with the HHSA. All configuration of the HHSA is handled from within Summit WMScanner as discussed below. Summit WMScanner can acquire data from the Anritsu HHSA in real-time, using the specified options. Depending on the sweep configuration, acquiring a single measurement point using the spectrum analyzer mode can take from 3 seconds (for a 0 Hz sweep) to over 30 seconds (for a full 3 GHz sweep). When using the Anritsu HHSA in power monitor mode, measurements can be made much more rapidly, with typical operation permitting over 30 samples a second.
2. Recording Real-Time Measurement Data With the Anritsu MS2711 HHSA
2.1 Configure the Anritsu MS2711 HHSA
Follow the steps outlined below before beginning any measurement run with Summit WMScanner using the Anritsu HHSA:
1. Connect all necessary cables, antennas, and any other accessories to the Anritsu HHSA as desired. For information about such procedures, see the documentation accompanying the Anritsu HHSA. Additionally, connect a null-modem serial cable from the Anritsu HHSA to the PC with Summit WMScanner installed.
2. Toggle the Anritsu HHSA power switch to the ON position.
3. When the system boots, verify that the Anritsu HHSA firmware version is 1.35 or higher. Summit WMScanner does not support Anritsu HHSA models with firmware versions prior to 1.35.
4. After the system has booted, press the Enter key as prompted on the Anritsu HHSA display.
5. To ensure that proper communications will occur with the Anritsu HHSA, place the unit into “Spectrum Analyzer” mode. To do this, press the MODE button under the LCD display, and use the UP/DOWN arrow keys to select “Spectrum Analyzer” mode. NOTE: Even if power monitor measurements are being made, make sure the Anritsu HHSA is running in Spectrum Analyzer mode.
Anritsu User’s Manual - Page
2 of 13.
2.2 Configure Summit WMScanner
From the Summit WMScanner menu, select “Measurement Survey”, “Record
New Data…”, choose the “Anritsu Site Master / MS2711 / MS2711A / S33XB / S33XC” item from the “Select Measurement Mode” box located at the bottom of the dialog, and then click “Next >>”. This will open the Configure Serial Port dialog box shown in Figure 1.
Figure 1: Configure Serial Port dialog box
Referring to Figure 1, select the appropriate serial port number where the Anritsu
HHSA is attached (e.g., COM1, COM2, etc.) The baud rate should be 9600. The other settings are grayed out because the HHSA does not support any other modes different from those shown in the dialog box. After these have been accomplished, click OK and the Anritsu MS2711 HHSA options dialog box shown in Figure 2 should be presented.
Anritsu User’s Manual - Page
3 of 13.
Figure 2: Anritsu MS2711 HHSA Options Dialog Box
Referring to Figure 2, follow the steps given below to configure the desired
options for collecting data:
1. Select the measurement mode: either Spectrum Analyzer, Channel Number, or Power Monitor.
If spectrum analyzer mode was selected, select the desired sweep options. For more information on how to choose these settings, consult the Anritsu MS2711 HHSA User’s Guide published by the device manufacturer.
Channel Number mode is another way of selecting settings for operating the Anritsu in Spectrum Analyzer mode. Simply select the channel standard and the channel number you wish to monitor.
If power monitor mode was selected, make sure an RF detector is
Anritsu User’s Manual - Page
4 of 13.