Extech SDL200 Operating Manual

User's Guide
4-Channel Thermometer / Datalogger
RTD and Thermocouple Inputs
Model SDL200
Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter displays and stores temperature readings from up to four (4) Type K, J, T, E, R, or S thermocouple temperature probes and two (2) Pt100 RTD probes. Logged data readings are stored on an SD card for transfer to a PC. In addition, an RS232 port allows data streaming to a PC. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Meter Description
1. Thermocouple inputs
2. LCD Display
3. Power ON-OFF
4. HOLD and Backlight
5. MAX-MIN key
6. Up arrow ▲ / T1-T2 key (also offset key)
7. SET and Clock
8. Down arrow ▼ (also offset key)
9. ENTER and LOG key
10. PT1 and PT2 RTD input jacks
11. PC interface jack
12. Power Adaptor jack
13. SD card slot
Notes: Items 10 – 12 are located behind the snap-off compartment cover on meter’s right side.
Battery compartment, tilt stand, and tripod mount are located on the rear of the instrument
key
key
key
Display Description
1. Thermometer input jack T1 and PT1 (RTD) identifier
2. Temperature reading for T1 thermocouple or PT1 RTD input
3. Thermocouple input jack T2 and PT2 (RTD) identifier
4. Temperature reading for T2 thermocouple or PT2 RTD input
5. Thermocouple input jack T3 identifier
6. Thermometer input type
7. Temperature reading for T3 thermocouple input
8. Temperature reading for T4 thermocouple input
9. Unit of measure
10. Thermocouple input jack T4 identifier
2 SDL200-EU-EN-V1.2-7/11
Getting Started
Power ON-OFF
Power the meter by pressing and holding the power button
Press and hold the power button for at least 1.5 seconds to power OFF the meter.
This meter is powered by six (6) 1.5VDC ‘AA’ batteries or by optional AC adaptor. If the meter
will not switch ON please check that fresh batteries are installed in the rear battery compartment (refer to the battery replacement section later in this guide) or, in the case of the AC adaptor, check that the adaptor is connected correctly to the meter and to an AC source (refer to the AC Power Adaptor section later in this guide).
Connecting the RTD and Thermocouple Sensors
This meter offers two modes of operation: RTD temperature probe mode (platinum 100 ohm) and Thermocouple temperature probe mode (thermocouple Types K, J, T, E, R, or S).
Connect up to four (4) thermocouples to the sub-miniature jacks at the top of the meter labeled
T1 through T4.
Connect up to two (2) RTD probes to the 3.5mm ‘phono’ jacks labeled PT1 and PT2 located
under the snap-off compartment cover on the lower right side of the meter .
Select the desired mode of operation, and other setup features, in the Setup Mode. Refer to the
Setup Mode section below.
IMPORTANT NOTE: The supplied thermocouple probes are useable in the range of -20 to +250°C (-4 to +482°F) only; even though the meter can display temperature values outside of this range. For measuring outside of the range specified here, please obtain thermocouples rated accordingly.
for at least 1.5 seconds.
3 SDL200-EU-EN-V1.2-7/11
Setup Mode
Basic settings at a glance
To view the current configuration of the meter with regard to time, date, and datalogging sampling rate press the SET button momentarily. The meter will now display the configuration in quick succession. If the information is not gleaned on the first try, simply press the SET button again until all of the information is noted.
Accessing the Setup mode
1. Press and hold the SET button for at least 1.5 seconds to access the Setup menu.
2. Press the SET button momentarily to step through the available parameters. The parameter type is shown on the bottom of the LCD and the current selection for that type is shown above it.
3. When a parameter is displayed that is to be changed, use the arrow keys to change the setting. Press the ENTER button to confirm a change.
4. Press and hold the SET button for at least 1.5 seconds to exit the Setup mode. Note that the meter automatically switches out of the Setup mode if no key is pressed within 7 seconds.
5.
The available Setup parameters are listed below. Additional detailed information is provided below this list:
dAtE
SP-t
PoFF
bEEP
dEC
Sd F
t-CF
tYPE
Setting the Clock Time
1. Access the dAtE parameter.
2. Press the ENTER button.
3. Use the arrow buttons to change the YEAR value. Press ENTER to confirm setting.
4. Repeat step 1.5 for MONTH, DAY, HOUR, MINUTE and SECONDS.
5. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode (or wait 7 seconds for the meter to automatically switch to the normal operating mode).
6. The clock will keep accurate time even when the meter is switched off. However, if the batteries expire the clock will have to be reset after fresh batteries are installed.
Setting the Datalogger Sampling Time (Rate)
1. Access the SP-t parameter.
2. Use the arrow buttons to select the desired sampling rate. The available settings are: 10, 30, 60, 120, 300, 600, 1800, and 3600 seconds.
3.
Press the ENTER button to confirm the entry.
4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode (or simply wait 7 seconds for the meter to automatically switch to the normal operating mode).
Set the clock (Year/Month/Date; Hours/Minutes/Seconds)
Set the datalogger sampling rate
Automatic power-off management (Enable or disable the auto-power off function)
Set the beeper sound ON/OFF
Set the numerical format; USA (decimal: 20.00) or European (comma: 20,00)
Format the SD memory card
Select the temperature unit of measure (C or F)
Select the thermometer type (Pt, K, J, T, E, R, or S)
0, 1, 2, 5,
4 SDL200-EU-EN-V1.2-7/11
Loading...
+ 7 hidden pages