Extech PH100 Operating Manual

User Guide
ExStikTM pH Waterproof Meters
Patent Pending
ExStikTM Description
Front Panel Controls
1. Battery compartment cap
2. LCD Display
3. MODE / HOLD button
4. CAL / RECALL button
5. ON/OFF button
6. Electrode collar
7. Electrode (refillable on PH110) (Electrode cap is not shown)
1
2
3 4 5
6
7
Display
1. Bargraph reading
2. Measurement reading
3. BAT (low battery) and HOLD (data hold) indicators
4. Temperature display
5. Bargraph scale designations
6. Units of measure
7. Calibration indicators
8. RENEW and CAL indicators
5
6
7 8
2
PH100-PH110 V5.1 05/07
1
2
3 4
Overview
pH Overview
pH is a unit of measure (ranging from 0 to 14pH) indicating the degree of acidity or alkalinity of a solution. pH tests are the most commonly performed measurements in water analysis and reports the negative log of the hydrogen ion activity of a solution which is an indicator of acidity or alkalinity. Solutions with a pH less than 7 are considered acidic, solutions with a pH higher than 7 are known as bases, and solutions with a pH of exactly 7 are neutral.
The pH scale is logarithmic so, for example, if sample A is 1 pH less than Sample B, this means that Sample A is 10 times more acidic then Sample B. A difference of 1 pH represents a ten-fold difference in acidity.
Getting Started
Remove the cap from the bottom of the ExStikTM to expose the electrode glass surface and reference junction
Before first use or after extended storage, soak the electrode (with its cap removed) in a pH 4 solution for about 10 minutes
White KCL crystals may be present in the cap. These crystals will dissolve in the soak or they can be simply rinsed with tap water
Always calibrate close to the expected measurement value
A sponge is located in the electrode protective cap. Keep this sponge soaked with a pH 4
solution to preserve Electrode life during storage
Replacing Electrodes
The ExStikTM is shipped with an electrode attached. Electrode life is limited and is dependent on (among other factors) frequency of use and care. If the electrode needs to be replaced, follow these steps for removing and connecting electrodes. Note that the PH110 has a refillable electrode and the PH100 does not.
1. To remove an electrode, unscrew and completely remove the electrode retaining collar.
2. Gently rock the electrode from side to side, pulling it away from the meter, until it disconnects.
3. To attach an electrode, carefully plug the electrode into the meter socket (note that the electrode connector is keyed, ensuring proper connection).
4. Secure the electrode in place by tightly turning the collar in place. (a rubber gasket seals the electrode with the meter).
Automatic Electrode Recognition
When the ExStikTM is turned on, it recognizes the type of electrode that is connected and displays the appropriate unit of measure. Attach electrode before turning the ExStik
TM
on.
Powering the ExStik™
If the batteries are weak, the ‘BAT’ indicator appears on the LCD. Press the ON/OFF key to turn the ExStik automatically after 10 minutes of inactivity to preserve battery life.
TM
on or off. The auto power off feature shuts the ExStikTM off
3
PH100-PH110 V5.1 05/07