Extech PCR70 Data sheet

Ordering Information:
PRC70 ............Multifunction Calibrator
PRC70-NIST ..PRC70 with NIST Certificate
Specifications subject to change without notice. Copyright © 2010 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com
7/7/11 - R1
Features:
mA output of 2-wire transmitters
Source 0 to 22mA with 1μA resolution
Measure 0 to 50mA (1μA resolution)
Select Type R, S, J, K, T, E, N, or B thermocouple and
RTD (Pt100, Pt1000) for precision calibrations in °F/°C
Output 0 to 100mV, 0 to 1V, or 0 to 10.0000V
Measure 0 to 50.00V
Single/continuous step function for Voltage and Current
Self-contained 24V loop power supply drives
current loads up to 1000Ω
Complete with four AAA batteries and test leads
Precision Simultaneous Source and Measure
For thermocouples (Types R, S, J, K, T, E, N and B), RTD, mA, mV, V, Hz,
and Ω devices
Multifunction Process Calibrator
Specifications Range Accuracy Source
Current 0.000 to 22.00mA ±(0.02% of reading ± 0.02%Range) Voltage -1.0000V to 11.0000V ±(0.02% of reading ± 0.01%Range) Resistance 0.00 to 40.000KΩ ±(0.02% of reading ± 0.025%Range) Type J -328 to 2192°F (-200 to 1200°C) ±1.2°F or 0.7°C Type K -328 to 2501°F (-200 to 1372°C) ±1.6°F or 0.9°C RTD (Pt100) -328 to 1472°F (-200 to 800°C) ±1.4°F or 0.8°C Frequency 1.00Hz to 110kHz ±2 digits Pulse 1.0 to 100000 cycles ±2 digits Switch (open/close) 1.00Hz to 110kHz ±2 digits Loop Power 24V ±10% Max. Load 1000 @ 20mA
Measure
Current -5.000 to 55.000mA ±(0.02% of reading ± 0.01%Range) Voltage -5.000mV to 55.00V ±(0.02% of reading ± 0.01%Range) Resistance 0.00 to 5.5000kΩ ±(0.05% of reading ± 0.02%Range) Type J -76 to 2192°F (-60 to 1200°C) ±1.8°F or 1.0°C Type K -148 to 2501°F (-100 to 1372°C) ±2.1°F or 1.2°C RTD (Pt100) -328 to 1472°F (-200 to 800°C) ±1.4°F or 0.8°C Frequency 3Hz to 50.000kHz ±2 digits Continuity <50Ω
Dimensions/Weight 8 x 3.7 x 1.7" (205 x 95 x 42mm)/1.1lbs (500g)