Extech Instruments Corporation, 285 Bear Hill Road, Waltham, MA 02451
3 Standard Accessories
Ensure that the package contains all of the accessories listed below. If items are damaged, please contact the vendor from
which the product was purchased.
z Two (2) sets of Industrial Test Leads
z One set of standard Test Leads
z One (1) set of Alligator clips
z User's Guide
z Two (2) Fuses 50mA/250V and 63mA/250V
Page 3 PRC70-EN-V2.1 12/11
4 Safety Information
For the correct and safe use of the instrument, be sure to follow the cautionary notes stated in this manual whenever
handling the instrument. Extech Instruments shall not be held liable for any damage resulting from use of the instrument in a
manner other than prescribed in the cautionary notes.
Warning identifies conditions and actions that pose hazards to the user; a Caution identifies conditions and actions that
A
may damage the meter or the equipment under test.
Refer to Table 2 for the international electric symbols adopted by the calibrator.
Table 2 - International Electrical Symbols
EARTH GROUND
Warning
To avoid possible electric shock or personal injury:
zDo not apply more than the rated voltage, as marked on the calibrator, between terminals or between any terminal and
earth ground;
z Before use, verify the meter’s operation by measuring a known voltage;
z Follow all equipment safety procedures;
z Do not connect the probe of the test leads with any live power when the other end has been inserted into the current
jack;
zDo not use the meter if it is damaged. Before using the meter, inspect the case. Look for cracks or missing plastic. Pay
particular attention to the insulation surrounding the connectors;
z Select the proper function and range for the measurement;
z Make sure the battery door is closed and latched before operating the meter;
z Remove test leads from the meter before opening the battery door;
z Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Replace damaged test leads
before using the meter;
z When using the probes, keep fingers behind the finger guards on the probes;
z Connect the common test lead before connecting the live test lead. When disconnecting test leads, disconnect the live
test lead first;
z Do not use the meter if it operates abnormally. Protection may be impaired. When in doubt, have the meter inspected;
z Do not operate this instrument in areas where inflammable or explosive gases or vapor exists. It is extremely hazardous
to use the instrument under such environments;
z Do not operate the meter near explosive gas, vapor, or dust;
z Use only four (4) AAA 1.5V batteries, properly installed in the meter case, to power the meter;
z Disconnect the test leads before shifting to different source or measurement functions;
z When servicing the meter, use only specified replacement parts.
z To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as
the low battery indicator (
) appears.
WARNING
INFORMATION
Caution
To avoid possible damage to meter or to equipment under test:
z Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity.
z Use the proper jacks, functions, and ranges for the measurement or source operation.
Page 4 PRC70-EN-V2.1 12/11
5 Calibrator Description
Refer to Figure 1
1. LCD display
2. Digit up/down adjust buttons
3. Control keypad
4. Measurement terminals
5. Source terminals
2
4
Figure 1 – Meter Description
5.1 Measurement/ Source Terminals
Figure 2 shows the calibrator’s measurement /source terminals. Table 3 explains their use.
Figure 2 - Measurement/ Source Terminals
Table 3 - Measurement/ Source Terminals
1
3
5
Terminal Function
1
2
3
4
5
6
7
8
Common (COM) return(-)terminal for all
measurement functions
Measurement Signals(+):DCV, OHM, FREQ,
TC, RTD, SWITCH, and CONT (continuity)
Measurement Signal(+):mADC
3W Terminal:measurement terminal of the 3W
OHM, RTD
LOOP Terminal:+24VDC Loop Power Terminal
Source Signal:(+)DCmA
Source Signals:(-)OHM、RTD
Common (COM) return(-)terminal for source
functions
Source Signal(+)DCV, OHM, TC, RTD, XMT,
FREQ, CYC, and SWITCH
Source Signal:(-)mADC
Page 5 PRC70-EN-V2.1 12/11
5.2 Keys
Figure 3 lists the calibrator’s function keys and Table 4 explains their use.
Figure 3 – Calibrator Function Keys
Table 4 – Function Key Descriptions
No. Name Function
1~5 Source value set key Increment source set point
6~10 Source value set key Decrement source set point
11
12
13 Power key Switch the power ON or OFF
14 Backlight Switch the backlight ON or OFF
15
16
17
18
19
20
21
22
23
24
25
26
Measurement ON key
Source ON key
Measurement FUNC Key
Source FUNC Key
TEMP Key
ZERO Key
Measurement RANGE Key
Source RANGE Key
AVG Key
25/100% key
HOLD Key
START Key
LOOP Key
Key
Switch measurement functions ON or OFF
Switch source function ON or OFF
Select measurement function
Select source function
Switch room temperature monitoring function ON or OFF
In TC source or measurement mode, switch the RJ compensation ON/OFF
Set the source value to zero-point; In pulse source function, set the pulse number.
Select measurement range; Measurement mA and percentage shifting.
Select source range; Source mA and percentage shifting.
Measuring average value
In mA source function, select 25% or 100% manual step output mode.
In pulse number, frequency or switch source, set the frequency value.
Freeze measured value
Source auto-pulse number, switch ON mA auto-stepping or sweeping function.
Convert sourced TC temperature and mV, and sourced RTD temperature and Ohms.
24v Loop circuit power: Convert the measured TC temperature and the mV, the
measured RTD temperature and the Ohms.
In DCmA source mode, select auto-wave; In frequency/pulse source mode, set
amplitude
Page 6 PRC70-EN-V2.1 12/11
5.3 Display Screen
a. Battery level indicator
b. Measure Mode icon
c. Measurement function ON
d. Measurement function OFF
e. Average value for measurement
f. Display HOLD for measured value
g. Switch measurement
h. Room temperature measurement
i. Beeper for measurement continuity
j. Measured value
k. Unit of measure
l. Dividing line between measurement and source mode displays
m. Type of RTD measurement / source
n. Dividing line between measurement and source mode displays
o. Type of Thermocouple (TC) measurement / source
p. Reference Junction Compensation ON
q. 24V Loop Power Supply ON
r. Source Mode icon
s. Source function ON
t. Source function OFF
u. Set -point for source
v. RUN icon for auto-step and auto-sweep
w. Unit of measure for setpoint of source
x. Not used
y. Not used
m
Figure 4 - LCD display
cd e
a
b
I
j
x
l
n
r
s
t
u
y
g
f
v
h
k
o
q
p
w
Page 7 PRC70-EN-V2.1 12/11
6 Source/Measurement Preparations
Precautions for Safe Use of the Instrument
zWhen using the instrument for the first time, be sure to read the instructions given in Section 4 “Precautions for Safe
Use of the Instrument.”
z Do not open the instrument’s case. Contact the vendor for service.
z In case of failure: Should the instrument begin to emit smoke, give off an unusual odor, or show any other anomaly,
immediately switch OFF the POWER. If using an AC Charger, disconnect the plug from the wall outlet. Also switch OFF
power to the object under test that is connected to the input terminals. Contact the vendor.
zCharger: Use an AC Charger dedicated to the instrument. Avoid placing any load on the Charger; prevent any heat-
emitting object from coming into contact with the adapter.
General Handling Precautions
zBefore transporting the instrument, switch OFF power to the object under test, and then power off the meter. When
using an AC Charger, disconnect the power cord from the wall outlet. Finally, detach all lead cables from the instrument.
Use a dedicated carry case when transporting the instrument.
z Do not bring any electrified object close to the input terminals, since this may damage the internal circuitry.
z Do not apply any volatile chemical to the instrument’s case or operation panel. Do not leave the instrument in contact
with any product made of rubber or vinyl for a prolonged period. Be careful not to let a soldering iron or any other heat-
emitting object come into contact with the operation panel (panel is made of thermoplastic resin).
zBefore cleaning the instrument’s case or operation panel, disconnect the power cord plug from the wall outlet if using an
AC Charger. Use a soft, damp cloth to gently wipe the outer surfaces of the instrument. Water that is allowed into the
instrument can result in malfunction.
z If using an AC Charger, disconnect the power cord plug from the wall outlet when the meter is not in use.
z For handling precautions regarding the batteries, see “Installing or Replacing the Batteries”.
z Never use the instrument with the cover of the battery holder opened.
Environmental Requirements
Use the instrument in locations that meet the following environmental requirements:
zAmbient temperature and humidity:
o
Ambient temperature range: 32 to 122
F (0 to 50oC); ambient humidity range: 30 to 80% RH; non-condensing.
zPlace the instrument on flat and level areas
Do not use the instrument in locations that are:
z Exposed to direct sunlight or close to any heat source.
z Exposed to frequent mechanical vibration.
z Close to an electrical noise source, such as high-voltage equipment or motive power sources.
z Close to any source of intensive electric or electromagnetic fields.
z Exposed to large amounts of grease fumes, hot steam, dust or corrosive gases.
z Exposed to unstable areas such as where a risk of explosion due to the presence of flammable gases exists.
Notes:
zUse the instrument under the following environmental conditions when precise measurements are required:
o
Ambient temperature range: 74
zWhen using the instrument within a temperature range of 32 to 64
F (23oC) ±5o; ambient humidity range: 20 to 80% RH (non-condensing)
o
F (0 to 18oC) or 82 to 122oF (28 to 50oC), add a value
based on the temperature coefficient shown in the “Specifications” section to the given accuracy rating.
zWhen using the instrument at an ambient humidity of 30% or lower, prevent electrostatic charges from being produced
by using an antistatic mat or other alternative means.
Page 8 PRC70-EN-V2.1 12/11
zCondensation may occur if the instrument is relocated from places with low temperature and humidity to places with
high temperature and humidity, or if the instrument experiences a sudden temperature change. In these cases, leave
the instrument under the given ambient temperature for at least one hour to ensure that the instrument is free from
condensation before using the instrument.
Indication of Battery Level
The battery replacement indicator shows the battery level in five steps according to the measured voltage of the batteries.
Full battery:
The battery level is below 50% full:
The battery level is below 25% full:
Low battery:
The battery icon flashes in sequence when charging.
Note that the battery replacement indicator is driven by directly measuring the battery voltage when the calibrator is in actual
operation. Consequently, the indicator may read differently depending on the battery load condition (e.g., the load condition
of the source output or on/off state of the measurement function).
Replace the batteries and fuses as described in the section “Replacing or installing batteries and fuses” toward the end of
this User Guide.
Connecting the Charger
Warning
zMake sure the voltage of the AC power source matches the rated supply voltage of the Charger before connecting the
Charger to the AC power source.
z Do not use a Charger other than the dedicated, supplied Charger.
z Do not charge non Ni-Cd, Ni-MH batteries.
Step 1: Ensure that the calibrator is OFF.
Step 2: Insert the plug of the optional Charger into the Charger connection jack.
Notes:
z Switch OFF the calibrator before connecting or disconnecting the Charger from AC power.
z Unplug the Charger from the Charger connection jack of the calibrator when discharging.
z Do not charge the calibrator without batteries installed.
Page 9 PRC70-EN-V2.1 12/11
Switching Power ON
Press the power key once to switch ON the calibrator. Press and hold the power key for 2 seconds to switch OFF the
calibrator.
Switching MEASURE Mode ON and OFF
The measurement function defaults to the OFF state each time the calibrator is switched ON.
z Use the MEASURE (ON) key to switch the MEASURE mode ON/OFF.
z If the MEASURE function is not needed and therefore switched OFF, power to the measurement circuit is also switched
OFF within the calibrator. Thus, battery power is saved.
zTurning off the MEASURE function causes the on-screen measured value to switch off and the “OFF” indicator to
appear.
zTo resume measurement when the MEASURE function is off, press the MEASURE (ON) key again.
Automatic Power OFF
When the calibrator is running on battery power, and no key is pressed for approximately ten minutes, the calibrator switches
OFF automatically. The automatic power-off time can be programmed or set to the factory default state, see the
“Programming the Global Settings” section.
Switching the LCD Backlight ON and OFF
Press the backlight key to switch ON the backlight; press again to switch it OFF. Battery life is shortened by backlight
overuse.
Notes: The backlight automatically switches OFF after 30 seconds. The time can be programmed or reset to the factory
default condition, see the “Programming the Global Settings” section.
Page 10 PRC70-EN-V2.1 12/11
7 Source
The calibrator can source a DC voltage, DC current, resistance, thermocouple, RTD, frequency, pulse signal or contact
output.
Warning
To avoid electrical shock, do not apply more than the rated voltage, as marked on the calibrator, between terminals or
between any terminal and earth ground. Always use the calibrator in locations with a voltage to ground value below 30V.
Caution
zDo not apply voltage to the output terminals for ranges other than 4-20mA simulating transmitter outputs. Otherwise, the
internal circuitry may be damaged.
zThe instrument has been calibrated without taking into account the voltage drop due to the resistance component of the
lead cables. Care must be taken therefore when drawing a load current since the voltage drop due to the resistance
component (approximately 0.1 Ω) of the lead cables induces a slight error.
7.1 Connecting Cables to Terminals
For DC voltage, thermocouple, frequency, pulse or contact output (see Figure 5)
Step 1: Connect the black lead cable for source to the COM output terminal and the red lead cable to the “VhzTcmA-” output
terminal.
Step 2: Connect the other ends of the cables to the input of equipment under test ensuring the polarities are correct.
Figure 5 - Sourcing DC voltage, TC, frequency, pulse and contact output
For DC current (see Figure 6)
Step 1: Connect the black lead cable for source to the “VhzTcmA-” output terminal and the red lead cable to the ‘’mA+’’
output terminal.
Step 2: Connect the other ends of the cables to the input of equipment under test ensuring correct polarity.
Figure 6 - Sourcing DC Current
Page 11 PRC70-EN-V2.1 12/11
For resistance and RTD signal (see Figure 7)
Step 1: Connect the two black lead cables for source to the ‘’ΩRTD’’ terminal and the red lead cable to the “VhzTcmA-”
terminal.
Step 2: Connect the other ends of the cables to the input of equipment under test ensuring correct polarity.
Figure 7 - Sourcing Resistances and RTD
7.2 Sourcing DC Voltage
Step 1: Using the Function selector switch (FUNC) select the DC voltage source function, select the desired range from
100mV, 1V, and 10V by pressing the〔RANGE〕key. The default value, the unit of the selected source function, and the
range is displayed in the lower area of the LCD.
Step 2: Set the output value digit by digit using the 〔S〕/〔T〕keys.
Each pair of 〔S〕/〔T〕keys corresponds to each digit of the LCD reading. Each press of the 〔S〕/〔T〕key increases
or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to overflow or underflow, allowing
the output value to be set without interruption. Holding down the 〔S〕/〔T〕key changes the digit in question quickly. The
value will not change if it is increased or decreased to the Maximum or Minimum value. Pressing the (ZERO) key initializes
the output set point to the default value (0).
Step 3: Pressing the (ON) key causes the indicator on the LCD to change from ‘’OFF‘’ to ‘’ON ‘’. The calibrator sources the
preset DC voltage between the output terminals.
Step 4: To switch OFF the output, press the (ON) key again. The’’ OFF ‘’ icon appears on the LCD and no signals are
sourced between the terminals.
7.3 Sourcing DC Current
Step 1: Using the Function selector switch (FUNC) select the desired source function 20mA. The default value and unit of
the selected source function is displayed in the lower area of the LCD.
Step 2: Set the output value digit by digit using〔S〕/〔T〕keys.
Each pair of〔S〕/〔T〕keys corresponds to each digit of the LCD reading. Each press of the〔S〕/〔T〕key increases
or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to overflow or underflow, allowing
the output value to be set without interruption. Holding down the〔S〕/〔T〕key continuously changes the digit in question
quickly. The value won’t change if it is increased or decreased to the Maximum or Minimum value. Pressing the (ZERO) key
initializes the output set point to the default value (0).
Step 3: Pressing the (ON) key causes the indicator on the LCD to change from ‘’OFF‘’ to ‘’ON ‘’. The calibrator sources the
preset DC voltage between the output terminals.
Step 4: To switch OFF the output, press the (ON) key again. The OFF icon appears on the LCD.
Page 12 PRC70-EN-V2.1 12/11
7.3.1 Manually Set 25%, 100% 4–20 mA Function
The source can be set within the 4–20 mA current range.
Step 1: In DC current mode, press the〔25%100%〕key to display “25%SET” on the lower part of the screen, and press
again to display “100%SET”. The default source values will be shown.
Step 2: Using each pair of 〔S〕/〔T〕output setting keys, set the signal in a step-by-step manner. In the 25% set point
condition, the signal can be set in 4 mA increments or decrements in the sequential order 4-8-12-16-20 by each press of the
key. In the 100% set point condition, the signal can set in 16 mA increments or decrements by each press of the key.
Pressing the (ZERO) key initializes the signal set point to the default value (4.00).
Step 3: Pressing the (ON) key causes the indicator on the LCD to change from ‘’OFF‘’ to ‘’ON ‘’. The calibrator sources the
preset 4–20 mA current signal between the output terminals.
Step 4: To switch OFF the output, press the (ON) key again. The’’ OFF ‘’appears on the LCD and no signals are sourced
between the terminals.
6.3.2 Auto-stepping and auto-sweeping 4-20mA function
The user can set a 4–20 mA range in which to auto-step or auto-sweep. 80 seconds are required to finish a 4-20mA cycle in
auto-sweeping mode and 20 seconds are required for auto-stepping mode.
Step 1: In the DC current mode, press the 〔
the screen; press again to display the auto-sweeping mode icon “
〕key to display the auto-stepping mode icon “” on the lower part of
”. The default source value will display.
Step 2: Pressing the (ON) key causes the indicator on the LCD to change from ‘’OFF‘’ to ‘’ON ‘’. The calibrator sources the
default 4–20 mA current signal between the output terminals.
Step 3: Pressing the (START) key starts the auto-stepping or the auto-sweeping mode. The “RUN” icon is displayed in the
lower part of the LCD.
Step 4: Pressing the (START) key again stops the auto-stepping or the auto-sweeping mode. The “RUN” icon switches
OFF.
Step 5: Pressing the (ON) key cancels the sourcing and the ‘’OFF’’ icon is displayed on the screen. No signals are sourced
between the terminals.
Notes:
z Press the (START) key again to continue the auto-stepping or the auto-sweeping mode after stopping them; The “RUN”
icon is displayed on the lower part of the screen.
z The ability to use the (START) key to start mA auto-stepping and auto-sweeping mode is only available when the
source function is in the ON state.
z Starting the mA auto-sweeping mode switches OFF the measurement mode. The mA auto-stepping mode and the
measurement function cannot run simultaneously.
Page 13 PRC70-EN-V2.1 12/11
7.3.3 mA% display
In the mA source mode, press the (RANGE)key to convert the preset source value to mA% mode (shown on the lower
part of the LCD) as described in the following equation.
100 (current measured value mA-4mA)
mA % = %
16 mA
Press the (RANGE) key to return to the current preset value (shown on the lower part of the LCD).
Notes:
Increment/decrement setting functionality is not available in mA% mode. Press the (RANGE) key again to return to the
source set mode.
7.3.4 4-20 mA simulate transmitter source
Connect the calibrator and loop power as shown in Figure 8. Operate as described in the DC current Source section.
Figure 8 - 4-20 mA simulation transmitter source
7.4 Sourcing Resistance
zThe calibrator sources a resistance signal by receiving the resistance-measuring current (I) supplied from the device
being calibrated (such as a resistance meter) and then delivering the voltage (V) proportional to the preset resistance
(R) between the output terminals, and thus producing the equivalent resistance R =V/I. Consequently, the calibrator
sources the signal correctly only for devices that use this method of measurement.
zThe allowable range of the resistance measuring current that the calibrator receives from a resistance measuring device
under calibration is rated as 0.1 to 3 mA. To ensure accuracy, the resistance measuring current from the device under
calibration is strictly confined within the range. See the “Specifications” section.
zAny resistance signal being sourced does not include the resistance component of the lead cables. The whole
resistance, when measured at the ends of the lead cables for sourcing, is found by adding the resistance of the lead
cables (approximately 0.1 ohm) to the sourced resistance signal. For the sourcing of precise resistance signals, use a
three-wire or a four-wire connection.
zIf capacitance between the terminals of a device under calibration is greater than 0.1µF, the calibrator may fail to source
correct resistance signals.
Step 1: Using the function selector switch (FUNC), select the Ohm mode. Using the (RANGE) key, select the range. The
selected function, the default range source value, and the unit are shown in the lower part of the LCD.
Step 2: Set the output value, digit by digit, using each pair of〔S〕/〔T〕keys.
Each pair of〔S〕/〔T〕keys corresponds to each digit of the LCD reading. Each press of the〔S〕/〔T〕key increases
or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to overflow or underflow, allowing
the output value to be set without interruption. Holding down the〔S〕/〔T〕key changes the digit in question quickly. The
Page 14 PRC70-EN-V2.1 12/11
value won’t change if it is increased or decreased to the Maximum or Minimum value. Pressing the (ZERO) key initializes the
output set point to the default value (0).
Figure 9 - Connection method based on three-wire and four-wire
Step 3: Pressing the (ON) key causes the SOURCE indicator on the LCD to change from ‘’OFF‘’ to ‘’ON ‘’. The calibrator
sources the preset resistance value between the output terminals.
Step 4: To switch OFF the output, press the (ON) key again. The ‘’OFF‘’ icon appears on the LCD and no signals are
sourced between the terminals. The connection method based on three-wire and four-wire is shown in Figure 9: