MultiPro™ Digital MultiMeter Series
With PC Interface
Models:
MP510
MP520
MP530
1
WARRANTY
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects
in parts and workmanship for three years from date of shipment (a six month limited
warranty applies on sensors and cables). If it should become necessary to return the
instrument for service during or beyond the warranty period, contact the Customer Service
Department at (781) 890-7440 ext. 210 for authorization. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is
responsible for shipping charges, freight, insurance and proper packaging to prevent
damage in transit. This warranty does not apply to defects resulting from action of the
user such as misuse, improper wiring, operation outside of specification, improper
maintenance or repair, or unauthorized modification. Extech specifically disclaims any
implied warranties or merchantability or fitness for a specific purpose and will not be liable
for any direct, indirect, incidental or consequential damages. Extech's total liability is
limited to repair or replacement of the product. The warranty set forth above is inclusive
and no other warranty, whether written or oral, is expressed or implied.
REPAIR AND CALIBRATION SERVICES
Extech offers complete repair and calibration services for all of the
products we sell. For periodic calibration, NIST certification or repair of
any Extech product, call customer service for details on services
available. Extech recommends that calibration be performed on an annual
basis to insure calibration integrity.
All rights reserved including the right of reproduction in whole or in part in any form.
MP Series V2.2 9/03
P/N: 7M1C-0241-A000
2
Introduction
Congratulations on your purchase of Extech model MP510, MP520, or
MP530 digital multimeter. Properly used, this meter will provide many
years of reliable service.
MultiPro™ Meters measure AC/DC Voltage, AC/DC Current, Resistance,
Frequency, Capacitance, Diode, and Continuity. The MultiPro™ also
offers RS-232 PC Interface data acquisition capabilities as an option.
Models MP520 and MP530 are True RMS AC sensing. In addition, the
MP530 offers Temperature, Relative, Peak capture and MIN/MAX
functions.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates
the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal
use, hazardous voltages may be present
Double insulation
Safety Precautions
1. Improper use of this meter can cause damage, shock, injury or death.
Read and understand this users manual before operating the meter.
2. Make sure any covers or battery doors are properly closed and
secured.
3. Always remove the test leads before replacing the battery or fuses.
4. Inspect the condition of the test leads and the meter itself for any
damage before operating the meter. Repair or replace any damage
before use.
5. Do not exceed the maximum rated input limits.
6. Use great care when making measurements if the voltages are
greater than 25VAC rms or 35VDC. These voltages are considered a
shock hazard.
7. Always discharge capacitors and remove power from the device
under test before performing Capacitance, Diode, Resistance or
Continuity tests.
8. Remove the battery from the meter if the meter is to be stored for
long periods.
9. To avoid electric shock, do not measure AC current on any circuit
whose voltage exceeds 500V AC.
10. Voltage checks on electrical outlets can be difficult and misleading
because of the uncertainty of connection to the electrical contacts.
Other means should be used to ensure that the terminals are not
"live".
11. The product is intended only for indoor use
12. If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
13. Pollution degree: 2
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for
3
connection to circuits in which measures are taken to limit the transient
overvoltages to an appropriate low level. Note – Examples include
protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming
equipment to be supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed
installations.
Note – Examples include switches in the fixed installation and some
equipment for industrial use with permanent connection to the fixed
installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of
the installation.
Note – Examples include electricity meters and primary over-current
protection equipment
50.00mV, 500.0mV, 5.000V, 50.00V, 500.0V, 1000V0.5% + 3d
40Hz to 500Hz
50.00mV, 500.0mV 0.8% + 3d
5.000V, 50.00V, 500.0V 1.0% + 4d
1000V 1.2% + 4d
Up to 20kHz
50.00mV, 500.0mV 0.5dB*
5.000V, 50.00V, 500.0V 3dB*
1000V Unspecified
*Specified from 30% to 100% of range
CMRR: >60dB @ DC to 60Hz, Rs=1kΩ
Input Impedance: 10MΩ, 16pF nominal (44pF nominal for 50mV &
500mV ranges)
DC CURRENT
RANGE Accuracy Burden Voltage
500.0µA, 5000µA 0.15mV/µA
50.00mA, 500.0mA 3.3mV/mA
5.000A, 10.00A*
0.2% + 4d
0.03V/A
*10A continuous, 20A for 30 seconds max with a 5 minute cool down
interval