Congratulations on your purchase of the Extech DM220 Pocket Autoranging
Multimeter. This meter measures AC/DC Voltage, AC/DC Current with
200mA/500Vresettable fuse, Resistance, Capacitance, Frequency, Duty Cycle,
Diode Test, and Continuity. It features a built-in non-contact AC voltage detector
plus flashlight. Proper use and care of this meter will provide many years of reliable
service.
Safety
This symbol adjacent to another symbol, terminal or
operating device indicates that the operator must refer to an
explanation in the Operating Instructions to avoid personal
injury or damage to the meter.
WARNING
This meter has been designed for safe use, but must be operated with caution.
The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified
maximum:
V DC or V AC 600V DC/AC
µA AC/DC 200mA 500Vfast acting resettable fuse
Resistance, Diode Test,
Continuity Test
2. USE EXTREME CAUTION when working with voltages greater than 25VAC
or 35VDC. These voltages are considered a shock hazard.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds
600V above earth ground.
4. NEVER connect the meter leads across a voltage source while the function
switch is in the current, resistance, or diode mode. Doing so can damage the
meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the
power when making resistance, continuity or diode tests.
6. ALWAYS turn off the power and disconnect the test leads from the circuit
before opening the cover to replace the batteries.
7. NEVER operate the meter unless the battery cover is in place and fastened
securely.
This WARNING symbol indicates a potentially hazardous
situation, which if not avoided, could result in death or
serious injury.
This symbol indicates that a device is protected throughout
by double insulation or reinforced insulation.
Function Maximum Input
Input Protection Limits
600V DC/AC
Description
1. Non-contact ACV detector probe tip
2. Non-contact ACV indicator light
3. 3-3/4 digit (4000 count) LCD
4. Mode switch
5. Flashlight button
6. Function switch
7. Battery cover
8. Test Leads
9. Hz/%Duty Cycle button
10. Flashlight
1
10
2
3
4
5
9
8
6
7
DM220 V3.0 01/06 2
Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very
dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in
use.
2. If “OL” appears in the display during a measurement, the value exceeds the
range you have selected. Change to a higher range.
NOTE: On some low AC and DC voltage ranges, with the test leads not connected
to a device, the display may show a random, changing reading. This is normal
and is caused by the high-input sensitivity. The reading will stabilize and give a
proper measurement when connected to a circuit.
AC/DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure AC/DC voltages if a motor on the circuit is being
switched ON or OFF. Large voltage surges may occur that can damage the
meter.
1. Set the function switch to the V/Hz/% position.
2. Press the MODE button to indicate “AC” or “DC” on the display.
3. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
4. Read the voltage in the display.
AC/DC CURRENT MEASUREMENTS
1. For current measurements up to 4000µA AC/DC, set the function switch to the
µA position
2. For current measurements up to 200mA AC/DC, set the function switch to the
mA position.
3. Press the MODE button to indicate “AC” or “DC” on the display.
4. Remove power from the circuit under test, then open up the circuit at the point
where you wish to measure current.
5. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
6. Apply power to the circuit.
7. Read the current in the display.
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any resistance measurements.
Remove the batteries and unplug the line cords.
1. Set the function switch to the Ω CAP position.
2. Press the MODE button to indicate Ω on the display.
3. Touch the test probe tips across the circuit or part under test. It is best to
disconnect one side of the part under test so the rest of the circuit will not
interfere with the resistance reading.
4. Read the resistance in the display.
DM220 V3.0 01/06 3
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or
wires that have voltage on them.
1. Set the function switch to the Ω CAP position.
2. Press the MODE button to indicate on the display.
3. Touch the test probe tips to the circuit or wire you wish to check.
4. If the resistance is less than approximately 100Ω, the audible signal will sound.
If the circuit is open, the display will indicate “OL”.
DIODE TEST
1. Set the function switch to the Ω CAP position.
2. Press the MODE button to indicate on the display.
3. Touch the test probes to the diode under test. Forward voltage will typically
indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”. Shorted devices
will indicate near 0V and an open device will indicate “OL” in both polarities.
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any capacitance measurements.
Remove the batteries and unplug the line cords.
1. Set the rotary function switch to the Ω CAP position.
2. Press the MODE button to indicate nF on the display.
3. Touch the test leads to the capacitor to be tested.
4. Read the capacitance value in the display
FREQUENCY MEASUREMENTS
1. Set the rotary function switch to the V/Hz/% position.
2. Press the Hz /% button to indicate Hz on the display.
3. Touch the test probe tips to the circuit under test.
4. Read the frequency on the display.
% DUTY CYCLE
1. Set the rotary function switch to the V/Hz/% position.
2. Press the Hz/% button to select % on the display.
3. Touch the test probe tips to the circuit under test.
4. Read the % duty cycle on the display.
NON-CONTACT AC VOLTAGE DETECTOR
WARNING: Risk of electrocution. Before use, always test the Voltage Detector on a
known live circuit to verify proper operation.
1. The Voltage Detector works in any function switch position, including OFF.
2. Place the detector probe tip next to the hot conductor or insert into the hot side
of the electrical outlet.
3. If AC voltage is present, the detector light will illuminate.
NOTE: The conductors in electrical cord sets are often twisted. For best results,
rub the probe tip along a length of the cord to assure placing the tip in
close proximity to the live conductor.
NOTE: The detector is designed with high sensitivity. Static electricity or other
sources of energy may randomly trip the sensor. This is normal
operation.
DM220 V3.0 01/06 4
FLASHLIGHT
Press and hold the
the flashlight off.
button to turn the flashlight on. Release the button to turn
AUTO POWER OFF
The auto off feature will turn the meter off after 15 minutes.
LOW BATTERY INDICATION
The icon will appear in the lower left corner of the display when the battery
voltage becomes low. Replace the battery when this appears.
Maintenance
WARNING: To avoid electric shock, disconnect the test leads from any source
of voltage before removing the battery cover.
WARNING: To avoid electric shock, do not operate your meter until the battery
cover is in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following
guidelines are followed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature
extremes can shorten the life of the electronic parts and distort or melt plastic
parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage
the electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth.
DO NOT use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE.
Remove old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the
batteries should be removed to prevent damage to the unit.
BATTERY INSTALLATION
WARNING: To avoid electric shock, disconnect the test leads from any source of
voltage before removing the battery cover.
1. Turn power off.
2. Open the rear battery cover by removing the Phillips head screw located on the
bottom of the meter.
3. Insert the batteries into battery holder, observing the correct polarity.
4. Put the battery cover back in place. Secure with the screw.
WARNING: To avoid electric shock, do not operate the meter until the battery
cover is in place and fastened securely.
NOTE: If your meter does not work properly, check the batteries to make sure that
they are still good and that they are properly inserted.
RESETTABLE FUSE
1. The fast acting, 0.2A/500V resettable fuse will open if the current limits of the
meter are exceeded.
2. The fuse will automatically reset itself when the suspect current is removed
from the meter’s input.
DM220 V3.0 01/06 5
Loading...
+ 10 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.