Extech CMM-17 Data Sheet

Specifications subject to change without notice. Copyright © 2007-2008 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com
9/24/08 - R2
Precision DC, mA, Voltage, Frequency Calibrator
Combined with True RMS MultiMeter
Process Calibrator MultiMeter
Simultaneous source/simulate and measure on
large dual display
Precision 0.03% accurate voltage and
current calibration source
Constant current output for loads up to 1200Ω
Bipolar current (±25mA) and voltage (±1.5V, ±15V) outputs
Simultaneously source 15V excitation voltage and
measure mA output of 2-wire transmitters
% of scale readout for 4 to 20mA or 0 to 20mA
Wide range Frequency (square wave) source with
adjustable amplitude, pulse width, and duty cycle
Programmable 16 continuous or timed step outputs, or 999
step ramp
Optional communications software allows you to remotely
operate the CMM-17 from your PC and document readings
DMM Features:
Measure DCmA, ACmA, DCV, ACV, AC+DC, Temperature,
Frequency, Duty Cycle, Pulse Width, Resistance, Continuity, and Diode
1μV or 1μA resolution and 0.03% basic DCV accuracy
(400mV-40V)
1ms Peak Hold for glitch capture, plus Data Hold, Auto Hold,
relative and dynamic recording
True RMS measurement for non-linear voltage and current
loads
Common Features:
Built-in bi-directional RS-232 interface
Built-in NiMH rechargeable battery
Large 50,000 count backlit LCD displays primary, secondary,
source/measure ranges and functions selected
Separate input/output banana jacks with fused protection
Auto power off with disable function
Complete kit with input/output test leads, Type K temperature
probe, 8 NiMH rechargeable batteries, 100V-250V AC adap­tor/recharger, and carrying case
Ordering Information:
CMM-17 ..............Process Calibrator MultiMeter
CMM-17-NIST ....CMM-17 with NIST Certificate
CMM-17-S ..........PC Software and RS-232 cable
Specifications
Source (Output) Ranges Max Res. Accuracy (% output + digits) DC Voltage ±1.5 100μV
±15V 1mV ±(0.03% + 3)
DC Current ±25mA 1μA ±(0.03% + 5)
(Max load 1200Ω)
Square Wave 28 Frequencies 0.01Hz ±(0.005% + 1)
(0.5,1,2,5,10,15,20,25,30,40,50,60,75,80,100,120,150,200,240,300,400,
480,600,800,1200,1600,2400,4800 Hz) Duty Cycle 0.39% to 99.60% 0.390625% ±(0.01% + 0.02%) Pulse Width 1/frequency Range/256 ±(0.01% + 0.3ms) Amplitude 5V, ±5V, 12V, ±12V 0.1V ±(2% + 0.2V)
Measure (Input) Ranges Max Res. Accuracy (% rdg + digits) DC Voltage 50mV 1μV ±(0.05% + 50d)
500mV, 5V, 50V 10μV ±(0.03% + 5d)
250V 10mV
AC Voltage (45-20kHz) 50,500mV, 5,50,250V 1μV ±(1.5% + 20d) @50/60Hz DC Current 50mA, 500mA 1μA ±(0.3% + 5d) AC Current (45Hz-2kHz) 50mA, 500mA 1μA ±(0.6% + 20d) AC + DC Voltage 50,500mV, 5,50,250V 1μV ±(0.8% + 25d) @50/60H AC + DC Current 50mA, 500mA 1μA ±(1.2% + 10d) 1ms Peak Hold (Glitch capture) 50,500mV, 5,50,250V 1μV
±(2% + 400d)
50mA, 500mA 1μA
Resist ance 500,5k,50k,500k,5MV 0.01V ±(0.15% + 5d)
50MV 1kV ±(1% + 8d)
Temperature (Type K) -40°F to 2502°F 1/0.1°F ±(0.3% + 6°F)
-40°C to 1372°C 1/0.1°C ±(0.3% + 3°C) Frequ ency 100Hz,1,10,100,200kHz 0.001Hz ±(0.02% + 3d) Duty Cycle 0.1-99.9% ±(0.3% per kHz +0.3%)FS Pulse Width 0.1-1999.9mS ±(0.2% + 3d) Diode Check / Continuity 0.45mA @ <4.8V / Beeper < 10Ω on 400Ω range Dimensions/Weight: 2.1 x 3.54 x 7.56"(54 x 90 x 192mm) / 3.75lbs 1.71kg)
V
V
V
V
V
V
V
mV
mA
mA
mA
mA
mA
mA
OFF
PROCESS CALIBRATOR MULTIMETER + RS232
MODE
SHIFT MAX/MIN °C/°F AUTO OUTPUT
Ω
250V
500mA
MAX
FUSED
DC 15V
25mA
MAX
FUSED
V•mA
V•mA•Ω
True RMS
CAT II 250V
CMM-17
Optional software records output and input and generates calibration report
Loading...