Instruction Manual
αlpha CON 200
1/8 DIN Conductivity Controller with
Temperature display and Transmitter
Technol o
adeEa
PREFACE
This manual serves to explain the use of the αlpha CON 200 controller/transmitter. This manual functions in two ways: first, as a step­by-step guide to help you operate the meter; second, it serves as a handy reference guide.
This manual is written to cover as many anticipated applications of the αlpha CON 200 controller/transmitter as possible. If there are doubts in the use of the αlpha CON 200 controller/transmitter, do not hesitate to contact the nearest Eutech Instruments Authorized Distributor.
Eutech Instruments cannot accept any responsibility for damage or malfunction to the controller/transmitter caused by improper use of the instrument. Remember to fill in the guarantee card and mail it mail to your Authorized Distributor or Eutech Instruments Pte Ltd.
The information presented in this manual is subjected to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd.
Copyright© Rev. 1 09/2003
Eutech Instruments Pte Ltd. All rights reserved.
TABLE OF CONTENTS
1 INTRODUCTION 1 2 SAFETY INFORMATION 2 3 OVERVIEW 3
3.1 FRONT PANEL 3
3.2 BACK PANEL 4
3.3 WIRING 5
3.4 PANEL-MOUNTING THE CONTROLLER 6
4 MEASUREMENT MODE 7 5 PASSWORD 8 6 CONDUCTIVITY CALIBRATION 10 7 TEMPERATURE CALIBRATION 12 8 SETUP MODE 13
8.1 GENERAL INFORMATION 13
8.2 SETUP MODE OVERVIEW 14
8.3 SET POINT 1 P1.0 15
8.4 SET POINT 2 P2.0 17
8.5 MEASUREMENT RANGE SELECTION P3.0 19
8.6 CONFIGURE TEMPERATURE SETTINGS P4.0 20
8.7 VIEWING CONDUCTIVITY CALIBRATION DATA P5.0 21
8.8 VIEWING CONDUCTIVITY/ TEMPERATURE ELECTRODE DATA P6.0 22
8.9 CONTROLLER RESET P7.0 23
9 RELAYS 24 10 TRANSMITTER FUNCTION 24 11 SPECIFICATIONS 25 12 ACCESSORIES 26 13 GENERAL INFORMATION 29
1 INTRODUCTION
Thank you for purchasing a CON 200 DIN Conductivity Controller. This controller is part of a series of quality process controllers available from Eutech Instruments. These sturdy, economical conductivity controllers are designed with the features and reliability of a much more expensive instrument.
Your controller includes:
Removable terminal blocks for easy connections;
Two mounting brackets for easy panel mounting;
Some features of this controller are:
Two set point, two SPDT relay operation
Scrolling, 14-segment LED guides user easily through setup functions
Reliable power supply from 85 to 260 V AC, 50/60 Hz or DC withstands
voltage fluctuations
Push-button operation from the front panel
Single-point calibration for each individual range
Adjustable hysteresis band prevents rapid contact switching around set-point
Selectable automatic or manual temperature compensation
Two-level password protection
Removable terminal strips for quick and easy connections
Built-in memory backup retains setup even if power fails, and lets you
configure unit before installation
Isolated 4-20 mA output for remote monitoring or hard copy recording
1
2 SAFETY INFORMATION
The Eutech Controller/Transmitter shall be installed and operated only in the manner specified in the Instruction manual. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument.
Before powering up the unit, make sure that power source it is connected to, is as specified in the top label. Failure to do so may result in a permanent damage to the unit.
The unit has live and exposed parts inside. If it has to be opened, make sure that the power to the unit is off and disconnected.
The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the manual.
The degree of protection against electric shock will be achieved only by observance of the corresponding installation rules.
2
3 OVERVIEW
3.1 Front Panel
The front panel consists of a 4-digit LED display, 8 LED annunciators and 4 keys.
Annunciators
1. REL 1 Displayed when Relay 1 is activated
2. REL 2 Displayed when Relay 2 is activated
3. MEAS Displayed in measurement mode
4. CAL Displayed in calibration mode
5. SETUP Displayed in setup mode
6. µS Displayed when measurement range in micro-Siemens / cm
7. oC Unit of the displayed parameter (temperature)
8. mS Displayed when measurement range in milli-Siemens / cm
Keys
9. MODE
10. ENTER
11. ▲(increment)
11. (decrement)
12. LED Display 14 segment LED display
Toggle between Conductivity and temperature display Confirm changes or to enter into further levels of sub-menu Use during calibration and setup modes to increment values. Press /keys together to escape to Measurement mode Use during calibration and setup modes to decrease values. Press
/keys together to escape to Measurement mode
8.8.8.8.
3
3.2 Back Panel
The back panel consists of three different connectors that can be used with removable terminal blocks (included):
153514
13
1820 1921
- 4 - 20mA
+ 4 - 20mA
12 11 10 9
RELAY1
Caution
NC
RELAY2
8177166
IN
NC
GND
- CON
+ CON
PT100
NC NC N L
SENSE
24
1
1. VAC live wire
2. VAC neutral wire
3. unused
4. unused
5. Relay 2 deactivated position (normally closed)
6. Relay 2 center pole
7. Relay 2 activated position (normally open)
8. Relay 1 deactivated position (normally closed)
9. Relay 1 center pole
10. Relay 1 activated position (normally open)
11. 4-20 mA connection, negative
12. 4-20 mA connection, positive
Pt 100 connection: sense (jumper to terminal 14 if using 2-wire RTD)
13.
14. Pt 100 connection: input
15. Pt 100 connection: ground
16. Conductivity positive terminal
17. Conductivity negative terminal
18. unused
19. unused
20. VAC protective ground wire
21. VAC protective ground wire
4
3.3 Wiring
Caution: Ensure electrical mains are disconnected before proceeding.
1. Connect the power supply to the three-pin terminal block
VAC live wire = 1
VAC neutral wire = 2
VAC protective ground wire = 20 OR 21
alpha CON 200 controller accepts voltages from 85 to 260 VAC, 50/60 Hz or DC.
2. Connect the Pt 100 leads to terminals 13 to 15 of the seven-pin terminal block. Either wire can be connected to either terminal. Terminals 13 and 14 must be shunted unless using a 3-wire RTD.
NOTE: CON 200 is factory set for Automatic temperature compensation. MTC can be selected in Program P4.0.
3. Power on the controller. The display automatically shows the Conductivity reading, the µS and ‘MEAS’ annunciators lights.
NOTE: In the event Pt 100 is not connected or the connection is broken in the ATC mode, the display flashes ‘OR’ to alert you.
5
3.4 Panel-mounting the controller
a
d
The supplied mounting hardware allows surface mounting to all panels and protective enclosures. Mounting cut-out size is 91 x 45 mm.
To attach the mounting to the controller:
1. Align the catch to the side of the controller, and insert threaded rods through
catch.
2. Screw the threaded-rod through the catch in a clock-wise direction. Tighten
until the catch holds the controller firmly against the back of the panel or protective housing. Repeat on the other side.
h c t a C
h c t a C
2
0
.
TOP VIEW
e d a e r h T
0 0 1 . x o r p p a
d o r
d e d a e r h T
1
5
4
.
x
p
o
p
a
r
p l l a
W
SIDE VIEW
6
4 MEASUREMENT MODE
Press MODE to toggle between:
Conductivity measurement mode and
Temperature measurement mode
Conductivity Measurement Mode
The controller displays the selected Conductivity range number, R X (X ranges from 1 to 8), for 2 seconds before displaying the Conductivity measurement.
Conductivity Range Range No., R Resolution Cell Constant
0.00 – 20.00 µS/cm 1 0.01 µS/cm 0.1
0.0 – 200.0 µS/cm 2 0.1 µS/cm 0.1
0.0 – 200.0 µS/cm 3 0.1 µS/cm 1 0 – 2000 µS/cm 4 1 µS/cm 1
0.00 – 10.00 mS/cm 5 0.01 mS/cm 1
0.00 – 20.00 mS/cm 6 0.01 mS/cm 1
0.0 – 200.0 mS/cm 7 0.1 mS/cm 1
0.0 – 200.0 mS/cm 8 0.1 mS/cm 10
Temperature Measurement mode
Press MODE key once to view the temperature measurement. The display shows ATC (Automatic Temperature Compensation) or MTC (Manual Temperature Compensation), then the current measured temperature (for ATC) or the set temperature (for MTC). The temperature.
NOTE: After pressing the MODE key to display Temperature, if there is no further key-press, the Controller will automatically revert to Conductivity Measurement mode after about 30 seconds.
See Setup program P4.0 for further instructions.