Eutech Instruments CON 100 C-CX User Manual

AAllpphhaa CCOONN 110000 CC//CCX
v
viittyy
CCoonndduucctti
i
X
CCoonnttrroolllleerr // TTrraannssmmiitttteer
INSTRUCTION MANUAL
MODE
68X103704 6/98
uS/mS
o
SETU
C
Technology Made Easy ...
READY
ENTER
RELAY
r
Preface
This manual serves to explain the use of the Alpha CON 100 Conductivity Controller/Transmitter. The units covered are Alpha C and CX Conductivity Controller/ Transmitter.
The instruction manual functions in two ways: first, as a step by step guide to help you operate and understand the operation of the unit and second, as a handy reference guide.
The information presented in this manual is subject to change as improvements are made, and does not represent a commitment of Eutech Instruments Pte Ltd.
This instruction manual is written to cover as many anticipated applications of the Alpha CON 100 C/CX Conductivity Controller/Transmitter as possible. If there are doubts in the use, please do not hesitate to contact the nearest Eutech Instruments’ Authorized Distributor.
Eutech Instruments cannot accept any responsibility for damage or malfunction of the unit due to improper use of instrument.
Copyright 1998 Eutech Instruments Pte Ltd. Version 1.1
Eutech Instruments (S) Pte Ltd Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139 949 Tel: (65) 778 6876; Fax: (65) 773 0836 E-MAIL: INFO@EUTECHINST.COM; WEB PAGE : HTTP://WWW.EUTECHINST.COM
1
TABLE OF CONTENTS
1. INTRODUCTION 1
2. GETTING ACQUAINTED 3
2.1 Front Panel 3
2.2 The Back Panel 4
2.3 Selecting Conductivi ty Measurement Range 4
2.4 Wiring 5
3. OPERATING THE CONTROLLER 6
The Main Display 6
4. SETTING UP THE CONTROLLER 7
4.1 Setting and Changing the Password 7
4.2 Setting the Controller Range (S oftware) 9
5. CALIBRATING THE CONTROLLER 10
5.1 The Lower Level Menus 10
5.2 Calibrating for Conductivity Measurement 10
5.3 Calibrating the Controller for Temperature Measurement 12
5.4 Setting the Alarm Feature 14
5.5 Setting Temperature Coef f i cient 18
6. SETTING TO FACTORY DEFAULT 19
7. USING THE CONTROLLER CURRENT LOOP FOR DATALOGGING (FOR TRANSMITTER MODEL ONLY) 20
8. ADDITIONAL INFORMATION 21
Appendix 1 23 Appendix 2 24 Technical Specifications 25
1. INTRODUCTION
The alpha CON 100 series Conductivit y Controller/Transmitter is an addition to the line of process controllers from Eutech Instruments. Incorporated with the ASIC (Application Specific Integrated Circuit) microprocessor technology, this panel­mounted on-line controller provides m any user-friendl y features desirable in conductivit y Cont rol l e rs.
This versatile conductivity control ler can be used for m easuring and controlling the Conductivity of a wide range of solutions in process streams.
Conductivity Ranges :
Range Resolution Cell Constant
0 - 99.9 uS
0 - 999 uS
0 - 999 uS 0 - 9.99 mS 0 - 99.9 mS
0 - 200 mS 0 - 200 mS
0.1 uS 1 uS 1 uS
0.01 mS
0.1 mS 1 mS 1 mS
0.1
0.1
1.0
1.0
1.0
1.0
10.0
1
Some of the features of this Controller are :
Automatic temperature compensation (with PT 100)
Manual temperature compensation setting without the
temperature probe
Adjustable temperature coefficient from 0. 0 to 10.0%
Push button calibration with cell correction function
High and Low alarm triggering relays
User-defined Password feature to prevent unauthorized
entry to change the calibration data, hysteresis function, high and low setpoints information
Alarm and operational message annunciators
Hysteresis function t o prevent chattering of relays around
the setpoint
High-end transmitter model with 4-20 mA current output
for datalogging purposes
Built-in memory backup to ensure that calibration and
other information will not be erased if power supply fails
Switchable mains voltages of 110 VAC or 220 VAC via
user selectable internal ju mper
2
2. Getting Acquainted
C
R
2.1 Front Panel
The front panel consists of a 3 digit LED display together with 5 LED annunciators. There are also 4 k eys as shown below.
MODE
888
ENTE
o
uS /mS
The keys available are the (UP/INCREMENT), (DOWN/DECREMENT), MODE and ENTER keys.
The annunciators are µS/mS, °C, SETUP, READY, and RELAY. The µS/mS annunciator lights up in Conductivity measurement mode. The READY annunciator lights up when the Conductivity stabilizes. RELAY lights up when any of the HI SET or LO SET relays is activated.
The MODE key allows you to select between Conductivity display, the temperat ure display or the SETUP menu display. While in one of the SETUP menus, it also functions as an ESCAPE key. For example, while sett ing the Hi SET point , you can press MODE key to return to the measurem ent mode.
You can confirm changes or enter into further levels of the lower menu by pressing the ENTER key. The (UP/INCREMENT) and (DOWN/DECREMENT) keys allow you to change information or t o select between different menus. Holding down the key increases the scrolling speed, i.e. changeover of 1st digit (ones) t o 2nd (tens) and then 3rd digit (hundreds).
SETUP
R E ADY REL AY
3
2.2 The Back Panel
The back panel consists of t wo connectors . The f irs t one is a 4­way screw terminal and the second is a 12-way screw term inal. Refer to the label on top of the unit for diagram. The connection for the 4-way screw termi nals are (from the left to right):
1. PT 100 connection
2. PT 100 connection
3. Conductivity input (Inner core)
4. Conductivity input (Outer shield) The connections for the 12-way screw term inals are (from left
to right),
5. High Set Relay deactivat ed position
6. High Set Relay center pole
7. High Set Relay activat ed position
8. Low Set Relay deactivated posit i o n
9. Low Set Relay center pole
10. Low Set Relay activated posit i o n
11. 4-20 mA - ve connection (for transmitter m odel s only)
12. 4-20 mA + ve connection (for transmitter model s only)
13. Protective earth
14. Protective earth
15. Neutral
16. Live
4
2.3 Selecting Conductivity Measurement Range
You can set the appropriate conductivity measurement range from the front panel. See S ection 4.2 for details on Setting the Controller Range (Software). Note that for “mA” the LED display shows “mA”.
2.4 Wiring
Connect the power supply to the GND (EARTH) - 13 or 14, NEUTRAL - 15 and LIVE - 16 screw terminals. Make sure that the power supply jumper setting matches the mains voltage (110 VAC or 220 VAC). See Appendix 2 for the jum per setting for the voltage selecti on.
Connect the Conductivity electrode to the 4-way sc rew terminal at the back panel and the PT 100 tem perature probe to the PT 100 connections.
Power on the controller and the display automatically s hows the Conductivity reading. The uS/m S annunciator lights up. Once the reading is stable, the READY annunciator lights up. However, if the PT 100 temperature probe is not connected, automatic temperature compensation does not function. You can set the temperature at a selected value. The tem perature is set to factory default at 25.0
NOTE : Eutech Instruments will not be responsible f or incorrect
application of the controller using improper voltage sources or wrong jumper settings.
o
C.
5
3. Operating the Controller
R
o
R
R
The Main Display
Press MODE key to switch to
SETUP
MODE
three main displays - the
ENTE
Conductivity display, the temperature display and “SEt” display.
MODE
Press MODE key once to get into
ENTE
the temperature measurement. The °C annunciat or lights up when you are measuring temperature. The display shows current measured temperature (with ATC)
MODE
or the temperature that was set in
ENTE
MTC mode. Press MODE key again and the display toggles to the “SEt” for SETUP menu.
7.4
uS/mS
24.0
C
SEt
Press ENTER key to go into the lower-level setup m enus while the display shows "SEt". These lower-level menus allow calibration of Conductivit y, Temperature and other parameters including set password, Hi or Lo Setpoints, Hi and Lo Hysteresis values and temperature coefficient.
6
4. Setting Up the Controller
R
R
R
R
R
4.1 Setting and Changing the Password
4.1.1 Setting New Password
To set a password, press MODE key until the “SEt” is displayed.
1. Press ENTER key and the
CAL
display shows "CAL” and “Con” alt ernately. Press ▲ key once and the display shows “Set” “id”.
2. Press ENTER to enter your desired password. Follow the
Con
steps below (e.g. using a password “123”).
3. Initially, the display shows "000" with the first digit blinking.
4. Use (UP/INCREMENT) key to
SEt
SETUP
enter “1” on the 1st digit of the password. Press ENTER once and the second digit will blink. Similarly, enter the second digit of the password “2” and press ENTER key again. Repeat with
SETUP
the third digit accordingl y.
5. Press ENTER key and the display shows “CAL” and ”COn” alternately. Use or key to get into the “SEt” and “id” menu,
123
press ENTER and set your desired password by following the above steps.
6. Once com pleted, press ENTER to conf irm and then press MODE key to return to the Conductivity display.
SETUP
SETUP
SETUP
id
MODE
ENTE
MODE
ENTE
MODE
ENTE
MODE
ENTE
MODE
ENTE
7
To calibrate the controller at any time, you may have to enter the password that you set, in order to access the c alibration mode. Once you have entered the password correctly, the display shows "CAL COn” indicating that you are in one of the lower-level SETUP menus.
If you enter the wrong password, the display revert s bac k t o t he Conductivity display. Alternatively, if you prefer no password protection, set the pass word to “000”, "CAL COn" immediately displays after you press ENTER key while you are in the "SEt" menu.
NOTE : The user set password is a prot ection c ode. Thus, it i s very important to keep this password strictly confidential to authorized personnel. You are advised to remember the password that you have set, in order to protect the controller settings and prevent any unauthorized tampering to the system!
IMPORTANT : In case the password set is forgotten, use the master password “555”.
4.1.2 Changing the Password
Enter the Set id menu with already set password or “555”. Change a new password as per steps mentioned for setting the password.
8
4.2 Setting the Controller Range (Software)
R
R
R
R
R
R
R
At the measurement mode, press both and keys simultaneously until the “CLr” display shows. Press or key to select the appropriate Conductivity measurement range. Press ENTER key to confirm.
Cell Constant either 0.1 or 1.0 blinks when the 999 uS range is selected
100
uS/mS
MODE
ENTE
Value scrolls from uS to mS, indicati ng
the measurement range or
100
uS/mS
MODE
ENTE
uS/mS
uS/mS
uS/mS
uS/mS
uS/mS
nnS
CLr
999
1.0
uS
MODE
ENTE
MODE
ENTE
MODE
ENTE
MODE
ENTE
MODE
ENTE
9
5. Calibrating the Controller
5.1 The Lower Level Menus
The "CAL COn" display is the fi rst menu seen upon entering the lower-level menus. Press or key to display the various menus as shown in the figure below. Pres s ENTER key to go to the lower level menus.
START
CAL COn
CAL oC
Hi SEt
Set id
Hys
Hi
Lo SEt
SEt tC
Hys
Lo
5.2 Calibrating for Conductivity Measurement
5.2.1 The CAL COn menu
1. Enter this menu by pressing the MODE key twice to the
“SEt” display from the measurement mode.
2. Press ENTER key. If the pas sword has been set earlier,
key in the password using the m et hod desc ribed in sec ti on
4.1. Press ENTER once to confirm. After you set the correct password, you can see the “CAL” and “COn” display blink alternat el y.
3. Press ENTER key to enter the Conductivity calibration
mode. Dip both the Conductivity electrode and temperature probe in the standard solution
4. NOTE: The Conductivity standard solution should
have a conductivity value between 40% to 100% of its full scale range selected
10
5. When “CAL COn” displays, press ENTER key to vi ew the
R
R
R
R
R
R
default cell constant at 100%. Press ENTER key again to show the measured conducti vity value. Use and keys to scroll to the standard’s Conductivi ty value
o
C.
at 25
6. Press ENTER key to calibrate the Conductivity value shown. Display will show the new cell constant being displayed. This
CAL
COn
MODE
ENTE
SETUP
MODE
ENTE
SETUP
value will be between 70% to 130%. Otherwise an error message “CAL“ and “Err“ alternatively blinks to indicate out of calibration range.
7. Press ENTER key to complete
CEL
MODE
ENTE
SETUP
the calibration, and you will return to “CAL” and “COn” menu.
8. Press MODE key and it brings you to the Measure mode.
100
MODE
ENTE
SETUP
NOTE : Gently stir the elect rode in a container filled with calibrating standard, and ENSURE that no bubble is trapped during the calibration process to avoid erroneous reading.
CEL
MODE
ENTE
SETUP
IMPORTANT : The correct Temperature Coefficient should be selected prior to Conductivity calibration. (Refer to Sec tion 5.5 for Setting Temperature Coeff i cient).
99
SETUP
MODE
ENTE
11
5.3 Calibrating the Controller for Temperature
R
R
R
R
A
R
R
Measurement
5.3.1 The CAL °°°°C menu
1. Enter this menu by pressing
MODE
SEt
SETUP
CAL
SETUP
o
C
SETUP
ON and ATC OFF.
tc
SETUP
On
SETUP
23.5
SETUP
MODE key twice to the “SEt” display
ENTE
if you are in measurement mode.
2. Press ENTER key. If the password has been set earlier, key i n the password using the method described in section 4.1. Press
MODE
ENTER once to confirm. After you
ENTE
set the correct password, you see the “CAL” and “COn” display blinks alternatively. Press key once, the display shows “CAL”
MODE
1. Press ENTER and the display
ENTE
alternatively shows “Atc” and “On” or “Atc” and “OFF”. Use and keys to choose between both ATC
ATC O N : If a PT 100 i s connected,
MODE
use and keys to adjust the
ENTE
temperature offset of t he PT 100 by ± 5 °C. Dip the probe into the sample liquid. Make sure that the display is set to ATC ON, press ENTER and the display now shows the actual
MODE
temperature reading (blinking).
ENTE
Use and keys to adjust the reading to its actual tem perature - as measured by an external thermometer. Once done, press
MODE
ENTER key and the display will flash
ENTE
“CAL” “Con” alternatively.
o
C” blinking.
12
ATC O F F : If a PT 100 i s not used,
R
A
R
R
then the ATC should be set to OFF. In step 3 above, choose by pressing and keys to select ATC OFF. Then press ENTER. The display will now show the default of 25
o
C or the last set val ue (blinking). Use and keys to set your desired value. Press ENTER to confirm and the display shows “CAL” “Con”.
tc
SETUP
OFF
SETUP
MODE
ENTE
MODE
ENTE
Note : For ATC OFF, you can adjust the set temperature values from 0.0 to 99.9
o
C. This value will be used for its temperature compensation e.g. Conductivity in the MTC mode.
25.0
SETUP
MODE
ENTE
13
5.4 Setting the Alarm Feature
R
R
R
This menu allows you to change the Hi and Lo Setpoint and Hysteresis values. S ee Section 8 for hysteresis applications.
IMPORTANT : When SETUP mode is entered, the 4-20 mA output (only for transmitter model) freezes and the relay de­activates (if it was in an alarm condition).
5.4.1 The Hi SET Menu
SEt
SETUP
Hi
SETUP
900
SETUP
4. Press ENTER key to access the Hi SET menu and the display shows the last Hi SET value or default (90% of Full Scale). Use and keys to change t he value of the Hi SET point.
5. Press ENTER to confirm the value of the Hi SET point.
6. Press MODE key to exit to the measurement mode. You can press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value is not stored into memory.
MODE
1. Enter this menu by pressing
ENTE
MODE key twice to the “SEt” display if you are in measurement mode.
2. Press ENTER key. If the password has been set earlier, key
MODE
in the password using the method
ENTE
described in section 4.1. Press
ENTER once to confirm.
3. After you set the correct
password, you see the “CAL” and “COn” display blinks alternately.
MODE
Press key twice, the display
ENTE
shows “Hi” “SEt” blinking alternately.
14
5.4.2 The Hi HYS Menu
R
R
R
1. Enter this menu by press ing the MODE key to the “SEt” display if you are in the measurement mode.
2. Press ENTER key. If the password has been set earlier, key in the password using the method described in s ection 4. 1. Press ENTER once to confirm. After you set the correct password, you see the “CAL” and “COn” display blinks alternatively. Press key thrice, the display shows “Hi” “HYS” blinking.
3. Press ENTER key to access the Hi HYS menu and the display shows the last Hi HYS value or default (2% of Full Scale). Use and keys to change the value of the Hi HYS point.
4. Press ENTER to confirm the value of the Hi HYS.
5. Press MODE key to exit to
measurement mode. You can press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value is not stored into mem ory.
HYS
SETUP
Hi
SETUP
02
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
NOTE : The maximum value of Hysteresis is 4% of its Full Scale (FS) selected. The HI HYS hysteresis is spread equally on either side of the Hi SET point.
15
5.4.3 The Lo SEt Menu
R
R
R
SEt
SETUP
Lo
SETUP
50
SETUP
MODE
ENTE
MODE
ENTE
MODE
ENTE
NOTE : The Lo SEt relay activates if the current Conductivity reading exceeds Lo SET point - ½ Lo HYS value. The Lo SET value cannot be higher than the Hi SET value.
1. Enter this menu by pressing MODE key twice to the “SEt” display if you are in meas urement mode.
2. Press ENTER key. If the password has been set earli er, k ey in the password usi ng the method described in section 4.1. Press
ENTER once to confirm.
3. After you set the correct
password, you see the “CAL” and “COn” display blinks alternately. Press
key four times, the display
shows “Lo” “SEt” blinking alternately.
4. Press ENTER key to access the Lo SET menu and the displ ay shows the last Lo SET value or default (5% of Full Sc ale). Use and keys to c hange t he value of the Lo SET point.
5. Press ENTER to confirm the value of the Lo SET point.
6. Press MODE key to exit to measurement mode. You can press MODE key (as an ESCAPE key) to revert to t he measurement value if ENTER is not pres s ed; the set value is not stored into memory.
16
5.4.4 The Lo HYS Menu
R
R
R
1. Enter this menu by pressing the MODE key to the “SEt” display if you are in the measurement mode.
HYS
SETUP
MODE
ENTE
2. Press ENTER key. If the password has been set earlier, key in the password using the method described in section 4.1.
3. Press ENTER once to
Lo
SETUP
MODE
ENTE
confirm. After you set the correct password, you see the “CAL” and “COn” display blinks alternatively. Press
key five times, the
display shows “Lo” “HYS” blinking alternately.
4. Press ENTER key to access the Lo HYS menu and the display shows the last Lo HYS value or default (2% of Full Scale). Use
and
keys to change the value of
02
NOTE : The maximum value of Hysteresis is 4% of its Full Scale (FS) selected. The Lo HYS hysteresis is spread equally on either side of the Lo SET point.
MODE
ENTE
SETUP
the Lo HYS point.
5. Press ENTER to confirm the value of the Lo HYS. Press MODE key to exit to the measurement mode. You can press MODE key (as an ESCAPE key) to revert to the measurement value if ENTER is not pressed; the set value is not stored into memory.
IMPORTANT : The Lo SET + ½ Lo HYS should be less than Hi SET - ½ Hi HYS i.e. Hi SET and Lo SET value (in consideration of the hysteres is band) can never overlap. Lo HYS and Hi HYS can be set independent of each other ; thi s allows non-symmetrical hysteresis option.
17
5.5 Setting Temperature Coefficient
R
R
R
5.5.1 Temperature Coefficient
This menu allows you to set the Temperature Coefficient corresponding to the solution whose Conductivity is being measured. In most control lers, this i s fixed at 2.10% per However, for alpha CON 100 C/CX models, it is adjust able from 0.0 to 10.0% per
o
C (please refer to Appendix 3 for
Temperature Coefficient determination and limits).
1. Enter this menu by pressing the MODE key to the “SEt”
SEt
SETUP
MODE
display if you are in the
ENTE
measurement mode.
2. Press ENTER key to access the SETUP menu. Press twice to enter t he “SEt tc” menu. I f
tc
SETUP
change and keep the Temperature
ENTE
Compensation as the previous
you do not wish to effect any
MODE
value, press MODE key. This will take you out of this mode and back to its measurement mode.
MODE
2.10
SETUP
3. Press ENTER and the di splay
ENTE
shows the current value of Temperature Coefficient (blinking).
4. Use
or key to change the
value of the Temperature Coeff icient. Pres s ENTER to confirm the value to be used. Press MODE key to exit to measurement mode.
o
key
C.
18
6. Setting to Factory Default
R
R
R
All the controller settings and user-defined password, except calibration data will be cleared when you enter into the “Usr” “Clr” menu.
Press both the and keys simultaneously while you are in any mode menu. The unit requests for the password if it was set previously.
CLr
SETUP
MODE
ENTE
Enter the password and press ENTER. The display will show “CLr”. Press ENTER to confirm. The display will blink briefly - the password is reset to its fact ory defaul t and toggles to its measurement mode.
Ent
SETUP
MODE
ENTE
The Master Password is “555” and default user set password is “000”. The Master Password can be used even if any other password was set previously.
SETUP
id
MODE
ENTE
Key the master pass word “555” after entering into the “Ënt” “id” display, you will see “Usr” and “Clr” display alternately. Press ENTER to confirm and the display returns to its measurement m ode.
19
7. Using the Controller Current Loop
for Datalogging (for Transmitter Model Only)
The 4-20 mA Current Loop
A 4-20 mA current loop can be connected if a remote data logging is required. The current will be proportional to the Conductivity displayed on the panel display. The 4-20 mA current loop can drive a load resist ance of no more than 200Ω.
If the calculated Conduc tivit y is less than 0 uS (due t o incorrect TC and temperature setting), the current will be set to approximately 0 mA. If Conductivit y exceeds the val ue that can be measured in the range, current will be set to approximately 21 mA. In this way, the rem ot e data l ogger can det ect when out of range conditions arise. Please note that the 4-20 mA loop is not galvanically isolated f rom the source.
20
8. Additional Information
A
A
The controller allows you to set High and Low alarms that switch on or off relays, and ac tivating or deactivating devices linked to the controller. In cases where the Conduc tivity values fluctuate close to the high or low setpoints, the relays will continuously switch on/off very quickly and may cause problems to the linked devi ces. The hysteres is band allows you to set an allowable range of fluctuat ions to prevent the relays from activating and deac tivating too quickly. See bel ow.
It is not possible t o set t he high set poi nt lower than the l ow set point and vice versa (taking the hysteresis band into consideration). For example, when the high set point is 8.50 mS and the high hysteresi s band is set at 0.40 mS , then the low setpoint cannot be set at hi gher than 8.50 - ½ Hi HYS.
NOTE : The shaded area indicates the hys teresis band.
D
Hi
Hys
Cond.
B
C
Hysteresis Band
E
ctual
Hi
Set
Lo
Hys
Hysteresis Band
F
H
Lo
Set
G
21
Explanation of the diagram in the previous page
F
A - Reading reaches (Hi SET - ½ HI HYS), the Hi-Set relay
remains inactivat ed.
B - Reading reaches above High Setpoint but below (Hi SET +
½ Hi HYS), Hi-Set relay remains inactivated.
C - Reading reaches between (Hi SET - ½ Hi HY S), the Hi-Set
relay remains inacti vated.
D - Reading reaches above (Hi SET + ½ Hi HYS), and the Hi-
Set relay is activat ed.
E - The Hi-Set relay is inac tivated only when the reading falls
below (Hi SET - ½ Hi HYS).
F - Reading reaches between (Lo SET - ½ Lo HYS), the Lo-
Set relay remains inactivated.
G - Reading reaches below (Lo SET - ½ Lo HYS), the Lo-Set
relay is activated.
H - The Lo-Set relay remains activated unt il the reading goes
above the (Lo SET + ½ Lo HYS)
22
Appendix 1
Jumper Positions - Internal to the Controller
JP1 Selects the input volt age between 110 VAC or
Fuse Note that there is a fuse internal to the
220 VAC.
Controller. Before opening the uni t, ENSURE that the power cable is physically separated from the mains s upply. Replace the fu se with the one recommended by the manufacturer.
Front
LED
Display
220VAC
2
Fuse
JP 1
Back
23
Appendix 2
(
)
Temperature Coefficient
The Temperature Coefficients (TC) of most solutions vary between 1.8 to 2.4 % per when the Conductivity exceeds about 100 uS. Thus, the def ault value of TC will be set to 2.10 % per
o
C. This is true of most salt solutions
o
C when the Controller is shipped from the factory. However, if the Temperature Coefficient is known to be dif ferent from the default, the user can set this value.
Temperature Coefficient can also be determined by taking 2 uncompensated readings at known temperatures. Uncompensated readings can be taken by setting TC t o 0.0 % or setting the temperature to 25
o
C by changing ATC ON to
ATC OFF. If G1 and G2 are the conductivity measured at two
temperatures T1 and T2, then TC is gi ven by the relation:
% TC = ----------------------------------------
G2 - G1
T2 - 25) - G2(T1-25
G1
* 100
24
Technical Specifications
Specifications Range Resolution Cell
Conductivity Range 0 - 99.9 uS
0 - 999 uS
0 - 999 uS 0 - 9.99 mS 0 - 99.9 mS
0 - 200 mS 0 - 200 mS
0.1 uS 1 uS 1 uS
0.01 mS
0.1 mS 1 mS 1 mS
Relative Accuracy +/- 1% full scale Cell Constant 0.1, 1.0 or 10.0 Measurement Range 7 separate ranges (software selectable,
non auto-ranging) Calibration Push-button (cell constant correction function) Temperature Coefficient 0.0 to 10.0 % from 5 to 50 oC Temperature
Automatic / Manual (0 to 100.0 oC) Compensation Temperature 0 to 100 oC
Resolution Relative Accuracy Sensor
0.1
+/- 0.5
PT 100
o
C
Output 4-20 mA, screw terminals (non-isolated);
Max. load 200 Display LED, 3 digits Inputs screw terminals Recommended Input
Less than 5 meters Cable Length Relays
No. Of Relays Maximum Voltage Maximum Current High Hys. Band Low Hys. Band
2 - High set & Low set SPDT
240 VAC
3A 4% of Full Scale Reading 4% of Full Scale Reading
Power Requirements 110 VAC or 220 VAC (Jumper selectable) Environmental Requirements Operating
Storage Humidity Limits
10 to 95% RH (non condensing)
0 to 50
-10 to 60
Storage Temp. Range 0 to 50 oC Dimensions 1/8 DIN size; 96 (L) x 48 (H) mm (Front panel)
Constant
0.1
0.1
1.0
1.0
1.0
1.0
10.0
o
C
o
C
o
C
25
NOTES
26
Loading...