This manual serves to explain the use of the Alpha CON
100 Conductivity Controller/Transmitter. The units
covered are Alpha C and CX Conductivity Controller/
Transmitter.
The instruction manual functions in two ways: first, as a
step by step guide to help you operate and understand
the operation of the unit and second, as a handy
reference guide.
The information presented in this manual is subject to
change as improvements are made, and does not
represent a commitment of Eutech Instruments Pte Ltd.
This instruction manual is written to cover as many
anticipated applications of the Alpha CON 100 C/CX
Conductivity Controller/Transmitter as possible. If there
are doubts in the use, please do not hesitate to contact
the nearest Eutech Instruments’ Authorized Distributor.
Eutech Instruments cannot accept any responsibility for
damage or malfunction of the unit due to improper use
of instrument.
Copyright 1998 Eutech Instruments Pte Ltd.
Version 1.1
The alpha CON 100 series Conductivit y Controller/Transmitter
is an addition to the line of process controllers from Eutech
Instruments. Incorporated with the ASIC (Application Specific
Integrated Circuit) microprocessor technology, this panelmounted on-line controller provides m any user-friendl y features
desirable in conductivit y Cont rol l e rs.
This versatile conductivity control ler can be used for m easuring
and controlling the Conductivity of a wide range of solutions in
process streams.
Conductivity Ranges :
RangeResolutionCell Constant
0 - 99.9 uS
0 - 999 uS
0 - 999 uS
0 - 9.99 mS
0 - 99.9 mS
0 - 200 mS
0 - 200 mS
0.1 uS
1 uS
1 uS
0.01 mS
0.1 mS
1 mS
1 mS
0.1
0.1
1.0
1.0
1.0
1.0
10.0
1
Some of the features of this Controller are :
• Automatic temperature compensation (with PT 100)
•Manual temperature compensation setting without the
temperature probe
•Adjustable temperature coefficient from 0. 0 to 10.0%
•Push button calibration with cell correction function
•High and Low alarm triggering relays
•User-defined Password feature to prevent unauthorized
entry to change the calibration data, hysteresis function,
high and low setpoints information
•Alarm and operational message annunciators
•Hysteresis function t o prevent chattering of relays around
the setpoint
•High-end transmitter model with 4-20 mA current output
for datalogging purposes
•Built-in memory backup to ensure that calibration and
other information will not be erased if power supply fails
•Switchable mains voltages of 110 VAC or 220 VAC via
user selectable internal ju mper
2
2. Getting Acquainted
C
R
2.1Front Panel
The front panel consists of a 3 digit LED display together with 5
LED annunciators. There are also 4 k eys as shown below.
MODE
888
ENTE
o
uS /mS
The keys available are the ▲ (UP/INCREMENT), ▼
(DOWN/DECREMENT), MODE and ENTER keys.
The annunciators are µS/mS, °C, SETUP, READY, and
RELAY. The µS/mS annunciator lights up in Conductivity
measurement mode. The READY annunciator lights up when
the Conductivity stabilizes. RELAY lights up when any of the HI
SET or LO SET relays is activated.
The MODE key allows you to select between Conductivity
display, the temperat ure display or the SETUP menu display.
While in one of the SETUP menus, it also functions as an
ESCAPE key. For example, while sett ing the Hi SET point , you
can press MODE key to return to the measurem ent mode.
You can confirm changes or enter into further levels of the
lower menu by pressing the ENTER key. The ▲
(UP/INCREMENT) and ▼ (DOWN/DECREMENT) keys allow
you to change information or t o select between different menus.
Holding down the key increases the scrolling speed, i.e.
changeover of 1st digit (ones) t o 2nd (tens) and then 3rd digit
(hundreds).
SETUP
R E ADYREL AY
3
2.2The Back Panel
The back panel consists of t wo connectors . The f irs t one is a 4way screw terminal and the second is a 12-way screw term inal.
Refer to the label on top of the unit for diagram.
The connection for the 4-way screw termi nals are (from the left
to right):
1. PT 100 connection
2. PT 100 connection
3. Conductivity input (Inner core)
4. Conductivity input (Outer shield)
The connections for the 12-way screw term inals are (from left
to right),
5. High Set Relay deactivat ed position
6. High Set Relay center pole
7. High Set Relay activat ed position
8. Low Set Relay deactivated posit i o n
9. Low Set Relay center pole
10. Low Set Relay activated posit i o n
11. 4-20 mA - ve connection (for transmitter m odel s only)
12. 4-20 mA + ve connection (for transmitter model s only)
13. Protective earth
14. Protective earth
15. Neutral
16. Live
4
2.3Selecting Conductivity Measurement Range
You can set the appropriate conductivity measurement range
from the front panel. See S ection 4.2 for details on Setting the
Controller Range (Software). Note that for “mA” the LED
display shows “mA”.
2.4Wiring
Connect the power supply to the GND (EARTH) - 13 or 14,
NEUTRAL - 15 and LIVE - 16 screw terminals. Make sure that
the power supply jumper setting matches the mains voltage
(110 VAC or 220 VAC). See Appendix 2 for the jum per setting
for the voltage selecti on.
Connect the Conductivity electrode to the 4-way sc rew terminal
at the back panel and the PT 100 tem perature probe to the PT
100 connections.
Power on the controller and the display automatically s hows the
Conductivity reading. The uS/m S annunciator lights up. Once
the reading is stable, the READY annunciator lights up.
However, if the PT 100 temperature probe is not connected,
automatic temperature compensation does not function. You
can set the temperature at a selected value. The tem perature is
set to factory default at 25.0
NOTE : Eutech Instruments will not be responsible f or incorrect
application of the controller using improper voltage
sources or wrong jumper settings.
o
C.
5
3. Operating the Controller
R
o
R
R
The Main Display
Press MODE key to switch to
SETUP
MODE
three main displays - the
ENTE
Conductivity display, the
temperature display and “SEt”
display.
MODE
Press MODE key once to get into
ENTE
the temperature measurement.
The °C annunciat or lights up when
you are measuring temperature.
The display shows current
measured temperature (with ATC)
MODE
or the temperature that was set in
ENTE
MTC mode. Press MODE key
again and the display toggles to
the “SEt” for SETUP menu.
7.4
uS/mS
24.0
C
SEt
Press ENTER key to go into the lower-level setup m enus while
the display shows "SEt". These lower-level menus allow
calibration of Conductivit y, Temperature and other parameters
including set password, Hi or Lo Setpoints, Hi and Lo
Hysteresis values and temperature coefficient.
6
Loading...
+ 20 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.