1. Deionized or distilled water for solution and standard preparation.
2. Eutech Calcium Standard, 0.1M CaCl2, Code. no. EC-SCS-CA1-BT. To prepare this
solution from your own laboratory stock, half fill a one liter volumetric flask with distilled
water and add 14.7 grams of reagent-grade calcium chloride (CaCl2.2H2O). Swirl the flask
gently to dissolve the solid. Fill to the mark with distilled water, cap, and upend several
times to mix the solution.
3. Eutech Calcium Standard, 1,000 ppm Ca
solution from you own laboratory stock, half fill a one liter volumetric flask with distilled
water and add 3.67 grams of reagent-grade calcium chloride (CaCl
gently to dissolve the solid. Fill to the mark with distilled water, cap, and upend several
times to mix the solution.
4. Eutech Calcium Standard, 100 ppm Ca+2 as CaCO3, Code. no. EC-SCS-CA3-BT. To
prepare this solution from your own laboratory stock, half fill a one liter volumetric flask
with distilled water and add 0.15 grams of reagent-grade calcium chloride (CaCl2.2H2O).
Swirl the flask gently to dissolve the solid. Fill to the mark with distilled water, cap, and
upend several times to mix the solution.
5. Eutech Ionic Strength Adjuster (ISA), 4 M KCl, Code. no. EC-ISA-CA1-BT. To prepare
this solution from your own laboratory stock, half fill a 1,000 ml volumetric flask with
distilled water and add 298 grams of reagent-grade potassium chloride (KCl). Swirl the
+2
, Code. no. EC-SCS-CA2-BT. To prepare this
.2H2O). Swirl the flask
2
3
Calcium Electrode Instruction Manual
flask gently to dissolve the solid. Fill to the mark with distilled water, cap, and upend
several times to mix the solution.
6. EDTA titrant, 1M stock solution, for the titration of calcium. To prepare this titrant, add
37.2 grams of reagent grade Na2EDTA.2H2O, ethylenediaminetetraacetic acid dihydrate,
disodium salt, to a 100 ml volumetric flask, add about 75 ml of distilled water, and swirl
the flask gently to dissolve the solid. Fill to the mark with distilled water, cap, and upend
several times to mix the solution.
GENERAL PREPARATION
Electrode Preparation
Remove the rubber caps covering the electrode tips and the rubber insert covering the filling hole of
the reference electrode. Fill the combination electrode or the reference electrode with the filling
solution shipped with the electrode to a level just below the fill hole. No preparation is required
with a sealed reference electrode. Gently shake the electrode downward in the same manner as a
clinical thermometer to remove any air bubbles which might be trapped behind the calcium
membrane. Prior to first usage, or after long-term storage, immerse the calcium membrane in
calcium standard for thirty minutes. The electrode is now ready for use.
Connect the electrodes to the proper terminals as recommended by the meter manufacturer.
Electrode Slope Check (with pH/mV meter)
(Check electrodes each day)
1. To a 150 ml beaker, add 100 ml of distilled water. Place the beaker on a magnetic stirrer
and begin stirring at a constant rate. After assuring that the meter is in the millivolt mode,
lower the electrode tips into the solution. If drifting or instability is observed, see the
TROUBLESHOOTING section.
2. Using a pipet, add 1 ml of 0.1M, 1,000 ppm, or 100 ppm (as calcium carbonate) standard
and 2 ml of ISA to the beaker. When the reading is stable, record the mV reading.
3. Using a pipet, add 10 ml of the same calcium standard used above to the beaker. When the
reading has stabilized, record the mV reading.
4. Determine the difference between the two readings. The electrode is operating correctly if
the mV potential has changed by 27±2 mV, assuming the solution temperature is between
20o and 25oC. See the TROUBLESHOOTING section if the potential change is not within this
range.
Slope is defined as the change in potential observed when the concentration changes by a factor of
10.
4
Instruction Manual Calcium Electrode
Electrode Slope Check (with ion meter)
(Check electrodes each day)
1. Prepare standard calcium solutions whose concentrations vary by tenfold. Use either the
0.1M Ca+2, 1,000 ppm Ca+2, or the 100 ppm Ca+2 (as calcium carbonate) standard stock
solutions. Use the serial dilution method for this preparation.
2. To a 150 ml beaker, add 100 ml of the lower value standard and 2 ml of ISA. Place the
beaker on the magnetic stirrer and begin stirring at a constant rate. Lower the electrode tips
into the solution. Assure that the meter is in the concentration mode.
3. Adjust the meter to the concentration of the standard and fix the value in the memory
according to the meter manufacturer's instructions.
4. Rinse the electrodes with distilled water and blot dry.
5. To another 150 ml beaker, add 100 ml of the higher value standard and 2 ml of ISA. Place
the beaker on the magnetic stirrer and begin stirring at a constant rate. Lower the electrode
tips into the solution.
6. Adjust the meter to the concentration of the standard and fix the value in the memory.
7. Read the electrode slope according to the meter manufacturer's instructions. Correct
electrode operation is indicated by a slope of 90-100%. See the TROUBLESHOOTING
sections if the slope is not within this range.
MEASUREMENT
Measuring Hints
All samples and standards should be at the same temperature for precise measurement. A difference
o
C in temperature will result in a 4% measurement error.
of 1
The sensing membrane is normally subject to water uptake and might appear milky. This has no
effect on performance.
Constant, but not violent, stirring is necessary for accurate measurement. Magnetic stirrers can
generate sufficient heat to change the solution temperature. To counteract this effect, place a piece
of insulating material, such as a styrofoam sheet, between the stirrer and beaker.
Always rinse the electrode tips with distilled water and blot dry. Use a clean, dry tissue to prevent
cross-contamination.
For samples with high ionic strength, prepare standards with compositions similar to that of the
sample. Always check to see that the membrane is free from air bubbles after immersion into
standard or sample.
A slow responding electrode may be caused by interferences to the electrode. To restore proper
performance, soak the electrode in distilled water for about 5 minutes to clean the membrane, rinse,
and soak in diluted standard solution for about 5 minutes.
5
Calcium Electrode Instruction Manual
Dilute concentrated samples (over 0.1M) before measurement. Recalibrate every few hours for
routine measurement.
Sample Requirements
All samples and standards must be aqueous and not contain organics which can dissolve in the
membrane or extract out the liquid ion exchanger.
The temperature of the standard solutions and of the sample solutions should be the same and below
40oC. About a 2% error will be introduced for a 1oC difference in temperature.
The pH range for the calcium ion electrode is 3-10. Neutralize samples outside this range with
NaOH or HCl to bring them in range.
Interferences should be absent. If they are present, use the procedures found in the Interferences
section to remove them.
Units of Measurement
Calcium concentrations are measured in units of ppm as calcium, ppm as CaCO3, moles per liter, or
any other convenient concentration unit. Table 1 indicates some of the concentration units.
TABLE 1: Concentration Unit Conversion Factors
ppm Ca+2 ppm CaCO3 moles/liter
4.01 10.0 1.0X10-4
10.00 24.9 2.5X10-4
40.10 100.1 1.0X10-3
400.80 1,000.9 1.0X10
-2
MEASUREMENT PROCEDURE
Direct Measurement
Direct measurement is a simple procedure for measuring a large number of samples. A single meter
reading is all that is required for each sample. The temperature of both sample solution and of
standard solutions should be the same.
Direct Measurement of Calcium (using a pH/mV meter)
1. By serial dilution prepare three standard solutions from the 0.1M, 1,000 ppm, or 100 ppm
-2
standard. The resultant concentrations should be 10
, 10-3, and 10-4M or 100, 10 and 1
ppm standards. Add 2 ml of ISA to each 100 ml of standard.
2. Place the most dilute solution (1.0X10-4M or 1 ppm) in a 150 ml beaker on the magnetic
stirrer and begin stirring at a constant rate. After assuring that the meter is in the mV mode,
lower the electrode tips into the solution. After the reading has stabilized, record the mV
reading.
6
Instruction Manual Calcium Electrode
3. Place the mid-range solution (1.0X10-3M or 10 ppm) in a 150 ml beaker on the magnetic
stirrer and begin stirring. After rinsing the electrodes with distilled water, blot dry, and
immerse the electrode tips in the solution. When the reading has stabilized, record the mV
value.
4. Place the most concentrated solution (1.0X10-2M or 100 ppm) in a 150 ml beaker on the
magnetic stirrer and begin stirring. After rinsing the electrodes in distilled water, blot dry
and immerse the electrode tips in the solution. When the reading has stabilized, record the
mV reading.
5. Using the semi-logarithmic graph paper, plot the mV reading (linear axis) against the
concentration (log axis). A typical calibration curve can be found in Figure 1.
A calibration curve is constructed on semi-logarithmic paper when using a
pH/mV meter in the millivolt mode. The measured electrode potential in mV
(linear axis) is plotted against the standard concentration (log axis). In the linear
region of the curve, only three standards are necessary to determine a calibration
curve. In the non-linear region, additional points must be measured. The direct
measurement procedures given are for the linear portion of the curve. The nonlinear portion of the curve requires the use of low level procedures.
6. To a clean, dry 150 ml beaker, add 100 ml of sample and 2 ml of ISA. Place the beaker on
the magnetic stirrer and begin stirring at a constant rate. Rinse the electrode tips with
distilled water, blot dry, and lower the electrode tips in the solution. When the reading has
stabilized, record the mV reading. Using the calibration curve, determine the sample
concentration.
7
Calcium Electrode Instruction Manual
7. The calibration should be checked every two hours. Assuming no change in ambient
temperature, place the electrode tips in the mid-range standard. After the reading has
stabilized, compare it to the original reading recorded in Step 3 above. A reading differing
by more than 0.5 mV or a change in the ambient temperature will necessitate the repetition
of Steps 2-5 above. A new calibration curve should be prepared daily.
Direct Measurement of Calcium (using an ion meter)
1. By serial dilution of the 0.1M, 1,000 ppm, or 100 ppm calcium standard, prepare two
calcium standards whose concentration is near the expected sample concentration.
Measure out 100 ml of each standard into individual 150 ml beakers and add 2 ml of ISA
to each.
2. Place the more dilute solution on the magnetic stirrer and begin stirring at a constant rate.
Assure that the meter is in the concentration mode. Lower the electrode tips into the
solution.
3. Adjust the meter to the concentration of the calcium standard and fix the value in the
memory according to the meter manufacturer's instructions after stabilization of the
reading. Rinse the electrode tips with distilled water and blot dry.
4. Place the more concentrated solution on the magnetic stirrer and begin stirring at a
constant rate. Lower the electrode tips into the solution.
5. Adjust the meter to the concentration of the calcium standard and fix the value in the
memory according to the meter manufacturer's instructions after stabilization of the
reading.
6. For low level measurements, place the rinsed, dried electrodes into a solution containing
100 ml of distilled water and 2 ml of ISA. After stabilization, fix the blank value in the
meter according to the meter manufacturer's instructions.
7. Place 100 ml of the sample and 2 ml of ISA in a 150 ml beaker. Place the beaker on the
magnetic stirrer and begin stirring.
8. Immerse the electrode tips in the solution and wait for the reading to stabilize. Read the
concentration directly from the meter display.
9. The calibration should be checked every two hours. Assuming no change in ambient
temperature, place the electrode tips in the first calcium standard. After the reading has
stabilized, compare it to the original reading in Step 3 above. A reading differing by more
than 0.5 mV or a change in ambient temperature will necessitate the repetition of Steps 2-6
above. The meter should be re-calibrated daily.
Low Level Calcium Determination (using a pH/mV meter)
-2
This procedure is recommended for solutions with ionic strengths less than 1.0X10
M. If the
solution is high in ionic strength, but low in calcium, use the same procedure, but prepare a
calibration solution with a composition similar to the sample.
8
Instruction Manual Calcium Electrode
1. Dilute 10 ml of the 0.1M standard to 1,000 ml to prepare a 1.0x10-3M standard solution
for measurements in moles per liter. Dilute 10 ml of the 1,000 ppm or 1 ml of the 100 ppm
standard to 1,000 ml to prepare a 10 ppm standard solution for measurements in ppm.
2. Soak the calcium electrode for at least 1 hour in 1.0x10-3M or 100 ppm calcium standard
solution.
3. To a 150 ml beaker, add 100 ml of distilled water. Place the beaker on the magnetic
stirrer and begin stirring at a constant rate.
4. Place the electrode tips in the solution. Assure that the meter is in the mV mode.
5. Add increments of the 1.0X10-3M or 10 ppm standard as given in Table 2 below.
6. After the reading has stabilized, record the mV reading after each addition.
TABLE 2: Step-wise Calibration for Low Level Calcium Measurements
Added Concentration
Step Pipet Volume (ml) M ppm
1 A 0.1 1.0X10-6 1.0X10-2
2 A 0.1 2.0X10-6 2.0X10-2
3 A 0.2 4.0X10-6 4.0X10-2
4 A 0.2 6.0X10-6 6.0X10
-2
5 A 0.4 9.9X10-6 1.0X10-1
6 B 2 2.9X10-5 2.9X10
7 B 2 4.8X10-5 4.8X10
-1
-1
Pipet A = 1 ml graduated pipet
Pipet B = 2 ml pipet
Solutions: additions of 1.0x10-3M or 10 ppm standard to 100 ml of distilled water.
7. On semi-logarithmic graph paper, plot mV reading (linear axis) against the concentration
(log axis) as in Figure 1.
8. Rinse the electrodes in distilled water and blot dry.
9. Measure out 100 ml of the sample into a 150 ml beaker and place the beaker on the
magnetic stirrer and begin stirring. Lower the electrode tips into the solution. After the
reading has stabilized, record the mV reading and determine the concentration from the
low level calibration curve.
10. Prepare a new low level calibration curve daily. Check the calibration curve every 1-2
hours by repeating Steps 2-7 above.
Low Level Calcium Determination (using an ion meter)
Follow the procedure given for normal calcium determinations using an ion meter and the blank
correction procedure.
Titration
9
Calcium Electrode Instruction Manual
The progressive and quantitative addition of a reagent to a measured sample until neither active
species (reagent or sample) is in excess. Ion selective electrodes are excellent endpoint detectors
since they are not influenced by solution color or turbidity. Though titration is more time
consuming than direct measurement, it is about 10 times more accurate.
Titration of Calcium
The method outlined in this section makes use of the calcium ion electrode as a highly sensitive
endpoint detector for calcium-containing samples. The titrant used is EDTA.
EDTA complexes calcium as well as other cations. The sample pH can be adjusted to pH 10 by
adding ammonia to eliminate unwanted ion complexes. Masking agents can be added in some
cases.
1. Soak the calcium ion electrode tip in 10-3M or 100 ppm calcium standard solution for a
minimum of one hour prior to use.
2. Prepare the stock EDTA titrant as given in the section Required Solutions. Dilute the
EDTA to 10 to 20 times as concentrated as the suspected sample concentration. The
sample should contain at least 1.0X10-3M calcium for a good detection of the endpoint.
3. Fill a 50 ml buret with the EDTA solution. Pipet 100 ml of the sample into a 150 ml
beaker, place the beaker on the magnetic stirrer and begin stirring at a constant rate. Adjust
the sample to pH 10 by adding ammonia.
4. Position the buret tip in the beaker, slightly above the liquid level in the beaker and slightly
off center. Position the electrode tips in the solution about half way between the center of
the beaker and the beaker wall.
5. Begin adding the EDTA in 0.5 ml to 1.0 ml increments and about 0.1 ml to 0.2 ml
increments as the potential begins to change more rapidly. Record the mV potential after
each addition. Continue the additions several milliliters past the endpoint.
6. Plot the milliliters of EDTA added against the mV potential on standard coordinate graph
paper. (See Figure 2). The point of greatest potential change is the endpoint.
7. The calcium ion concentration from the unknown is calculated as follows.
10
Instruction Manual Calcium Electrode
VtMt
M
V
+2
= ———
Ca
Ca
+2
where
M
+2
= concentration of calcium ion in the unknown (moles/liter)
Ca
Vt = volume of EDTA added at endpoint
Mt = EDTA concentration (moles/liter)
V
+2
= volume of unknown sample
Ca
ELECTRODE CHARACTERISTICS
Reproducibility
Electrode measurements reproducible to ±4% can be obtained if the electrode is calibrated every
hour. Factors such as temperature fluctuations, drift, and noise limit reproducibility. Reproducibility
is independent of concentration within the electrode's operating range.
Interferences
Table 3 lists some common cations that, if present in high enough levels, will cause electrode
interferences and measurement errors or electrode drift when using the calcium electrodes.
Electrode drift and slow response could indicate the presence of high interferences from the ions
listed. Soak the electrode in distilled water for five minutes, then for five minutes in calcium
standard solution to restore proper response.
TABLE 3: Concentration of Possible Interferences Causing a 10%
Error at Various Levels of Calcium.
Interferences
(moles/liter) 10-2M 10-3M 10-4M
+2
Mg
1.0X10+1 1.0X10
Zn+2 1.0X10+1 1.0X10
Ba+2 7.0X10
K
Sulfate, bicarbonate, and carbonate are the most common species that complex calcium ions. The
level of calcium ions, the level of the complexing ion, the pH of the solution, and the total ionic
strength of the solution determine the extent of the complexation. Complexation reduces the free
calcium ion concentration and, since the electrode responds only to free calcium ions, a false
reading results.
To avoid formation of CaSO4, the sulfate concentrations must be less than 5X10-4M (50 ppm). To
avoid formation of CaCO3 or formation of the CaHCO
+
complex, the pH of the solution should be
3
less than 7, and the total carbonate/bicarbonate concentration should be less than 3X10-3M (280
ppm carbonate).
Temperature Influences
Samples and standards should be at the same temperature, since electrode potentials are influenced
by changes in temperature. A 1oC difference in temperature results in a 4% error at the 1.0X10-3M
level.
Provided that temperature equilibrium has occurred, the calcium electrodes can be used at
temperatures from 0o-40oC. Room temperature measurements are recommended, since
measurements at temperatures quite different from room temperature may require equilibrium times
up to one hour. Table 4 indicates the variation of theoretical slope with temperature.
TABLE 4: Temperature vs. Value for the Electrode Slope
Plotting the mV potential against the calcium concentration on semi-logarithmic paper results in a
straight line with a slope of about 27 mV per decade (Refer to Figure 1).
The time needed to reach 99% of the stable electrode potential reading, the electrode response time,
varies from one minute or less for calcium concentration above 1.0x10-4M to several minutes near
the detection limit. (Refer to Figure 3.)
Limits of Detection
The upper limit of detection in pure calcium chloride solutions is 1M. In the presence of other ions,
the upper limit of detection is above 1.0x10-1M, but the possibility of a liquid junction potential
developing at the reference electrode and the "salt extraction effect" are two limiting factors. Some
salts may infuse into the electrode membrane at high salt concentrations causing deviation from
theoretical response. Calibrate the electrode at four or five intermediate points, or dilute the sample,
-1
to measure samples between 1.0x10
M and 1M.
The lower limit of detection is influenced by the slight water solubility of the ion exchanger used in
the sensing portion of the electrode. Refer to Figure 1 for a comparison of the theoretical response
to actual response at low levels of calcium chloride.
pH Effects
The operating range of the calcium electrode is from pH 3 to pH 10. Use at other pH values can
adversely affect the membrane. Hydrogen ion interferes with measurements of very low levels of
calcium. Hydroxide ion will complex calcium ions.
Electrode Life
The calcium electrode will last six months in normal laboratory use. On-line measurement might
shorten operational lifetime to several months. In time, the response time will increase and the
calibration slope will decrease to the point calibration is difficult and electrode replacement is
required.
13
Calcium Electrode Instruction Manual
Electrode Storage
The calcium electrodes may be stored for short periods of time in 1.0X10-2M calcium standard. For
longer storage (longer than two weeks), rinse and dry the calcium membrane and cover the tip with
any protective cap shipped with the electrodes. The reference portion of the combination electrode
(or the outer chamber of the reference electrode) should be drained of filling solution, if refillable,
and the rubber insert placed over the filling hole.
ELECTRODE THEORY
Electrode Operation
Eutech Calcium Electrode consists of an electrode body containing an ion exchanger in a sensing
module. This sensing module contains a liquid internal filling solution in contact with a gelled
organophilic membrane containing a calcium selective ion exchanger.
An electrode potential develops across the membrane when the membrane is in contact with a
calcium solution. Measurement of this potential against a constant reference potential with a digital
pH/mV meter or with a specific ion meter depends on the level of free calcium ion in solution. The
level of calcium ions, corresponding to the measured potential, is described by the Nernst equation:
E = Eo + S log X
where: E = measured electrode potential
Eo= reference potential (a constant)
S = electrode slope (∼27 mV/decade)
X = level of calcium ions in solution
The activity, X, represents the effective concentration of the ions in solution. Total calcium
concentration, Ct, includes free calcium ions, Cf, plus bound or complexed calcium ions, Cb. Since
the calcium electrodes only respond to free ion, the free ion concentration is:
Cf = Ct - Cb
The activity is related to the free ion concentration, Cf, by the activity coefficient, γ , by:
X = γ Cf
Activity coefficients vary, depending on total ionic strength, I, defined as:
I = ½ Σ CxZ
2
x
where: Cx = concentration of ion X
Zx = charge on ion X
Σ = sum of all of the types of ions in the solution
In the case of high and constant ionic strength relative to the sensed ion concentration, the activity
coefficient, γ , is constant and the activity, X, is directly proportional to the concentration.
To adjust the background ionic strength to a high and constant value, ionic strength adjuster (ISA)
is added to samples and standards. The recommended ISA for calcium is potassium chloride, KCl.
14
Instruction Manual Calcium Electrode
Solutions other than this may be used as ISA's as long as ions that they contain do not interfere with
the electrode's response to calcium ions.
The reference electrode must also be considered. When two solutions of different composition are
brought into contact with one another, liquid junction potentials arise. Millivolt potentials occur
from the inter-diffusion of ions into the two solutions. Electrode charge will be carried unequally
across the solution boundary resulting in a potential difference between the two solutions, since
ions diffuse at different rates. When making measurements, it is important to remember that this
potential be the same when the reference is in the standardizing solution as well as in the sample
solution or the change in liquid junction potential will appear as an error in the measured electrode
potential.
The composition of the liquid junction filling solution in the reference electrode is most important.
The speed with which the positive and negative ions in the filling solution diffuse into the sample
should be equitransferent. No junction potential can result if the rate at which positive and negative
charge carried into the sample is equal.
TROUBLESHOOTING GUIDE
The goal of troubleshooting is the isolation of a problem through checking each of the system
components in turn: the meter, the glass-ware, the electrodes, the standards & reagents, the sample,
and the technique.
Meter
The meter may be checked by following the check-out procedure in the instrument instruction
manual.
Glass-ware
Clean glass-ware will drain clean. That is, when rinsed with distilled or deionized water, the water
does not bead on the inside walls of the glass-ware.
Electrodes
The electrodes may be checked by using the procedure found in the sections entitled Electrode
Slope Check
.
1. Be sure to use distilled or deionized water when following the procedures given in
Electrode Slope Check.
2. If the electrode fails to respond as expected, see the section Measuring Hints. Repeat the
slope check.
3. If the electrodes still fail to respond as expected, substitute another calcium electrode that
is known to be in good working order for the questionable electrode. If the problem
persists and you are using an electrode pair, try the same routine with a working reference
electrode.
4. If the problem persists, the standards and/or reagents may be of poor quality, interferences
in the sample may be present or the technique may be faulty. (See Standards and
Reagents, Sample,
and Technique sections below.)
15
Calcium Electrode Instruction Manual
5. If another electrode is not available for test purposes, or if the electrode in use is suspect,
review the instruction manual and be sure to:
- Clean and rinse the electrodes thoroughly.
- Prepare the electrodes properly.
- Use the proper filling solution.
- Adjust the pH of the solution according to the method being used for the analysis.
- Measure correctly and accurately.
- Review TROUBLESHOOTING HINTS.
Standards and Reagents
Whenever problems arise with the measuring procedure that has been used successfully in the past,
be sure to check the standard and reagent solutions. If in doubt about the credibility of any of the
solutions, prepare them again. Errors may result from contamination of the ISA, incorrect dilution
of standards, poor quality distilled/deionized water, or a simple mathematical miscalculation.
Sample
Look for possible interferences, complexing agents, or substances which could affect the response
or physically damage the sensing electrode (or the reference electrode) if the electrodes work
perfectly in the standard, but not in the sample.
Try to determine the composition of the samples prior to testing to eliminate a problem before it
starts. (See Measuring Hints, Sample Requirements, and Interferences.)
Technique
Be sure that the electrodes' limit of detection has not been exceeded. Be sure that the analysis
method is clearly understood and is compatible with the sample. Refer to the instruction manual
again. Reread GENERAL PREPARATION and ELECTRODE CHARACTERISTICS.
If trouble still persists, call Eutech Instruments Pte Ltd. at (65) 6778-6876 and ask for the Customer
Services Department.
16
Instruction Manual Calcium Electrode
TROUBLESHOOTING HINTS
Symptom Possible Causes Next Step
Out of Range defective meter check meter with shorting strap
Reading (see meter instruction manual)
defective electrode check electrode operation
electrodes not unplug electrodes and reset
plugged in properly
reference electrode be sure reference electrode is filled
not filled
air bubbles on remove bubbles by
membrane re-dipping electrode
electrodes not put electrodes in solution
in solution
Noisy or Unstable defective meter check meter with
Readings (readings shorting strap
continuously or
randomly changing air bubble on remove bubble by
membrane re-dipping electrode
meter or stirrer ground meter or stirrer
not grounded
outer filling solution fill electrode to level just
too low below fill hole
defective electrode replace electrode
electrode exposed to soak electrode in calcium standard
interferences
"Incorrect incorrect scaling plot millivolts on the linear
Answer" of semi-log paper axis. On the log axis, be sure
(but calibration concentration numbers within
curve is good) each decade are increasing
increasing concentration
incorrect sign be sure to note sign of
millivolt number correctly
Drift (reading samples and standards allow solutions to come to room
slowly changing at different temperature before measurement
in one direction) temperatures
electrode exposed soak electrode in
to interferences calcium standard
incorrect reference use recommended
filling solution filling solution
incorrect pH adjust to pH 3-10
with NaOH or HCl
Low Slope or standards contaminated prepare fresh standards
No Slope or incorrectly made
defective electrode check electrode operation
air bubble on membrane remove bubble by
re-dipping probe
electrode exposed to soak electrode in
interferences calcium standard
standard used as ISA use ISA
ISA not used use ISA
18
Instruction Manual Calcium Electrode
SPECIFICATIONS
Concentration Range: 1M to 5x10-6M
pH Range: 3 to 10
Temperature Range: 0o to 40 oC
Resistance: 100 Mohms
Reproducibility: ±4%
Samples: Aqueous solutions only; no organic solvents
Size: 110 mm length; 12 mm diameter; 1 m cable length
Storage: Store in dilute calcium standard
ORDERING INFORMATION
CODE NO. DESCRIPTION
EC-CAL-03 Calcium Electrode, epoxy body combination
EC-SCS-CA1-BT Calcium Standard, 0.1M CaCl2
EC-SCS-CA2-BT Calcium Standard, 1,000 ppm Ca
+2
EC-SCS-CA3-BT Calcium Standard, 100 ppm Ca+2 as CaCO3
EC-ISA-CA1-BT Calcium ISA (Ionic Strength Adjuster), 4 M KCl
19
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.