Eutech Instruments ALPHA PH 800 User Manual

EUTECH INSTRUMENTS
ααααlpha-pH800
pH and ORP Controller/Transmitter
Operating Instructions
68X216813 03/99
Thank you for purchasing the αlpha-pH800 series pH/ORP controller/transmitter.
This manual serves to explain the us e of th e αlpha-pH800 series pH/ORP controller/transm itter. The manual functions in two ways, fir stly as a ste p by step guide to help t he user op erate the instr um ent. Second l y, it serv es as a handy referenc e guide. T his instr uctio n m anual is writ ten to c over as many anticipated applications of the αlpha-pH800 pH/ORP controller/transmitter. If you have doubts in the use of the instrument, please do not hesitate to contact the nearest Eutech Instruments’ Authorised Distributor.
The information presented in this manual is subject to change without notice as improvements are made, an d does not r eprese nt a com m itment on part of Eutech Instruments Pte Ltd.
Eutech Instruments cannot accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.
Copyright 1999 Eutech Instruments Pte Ltd. Version 1.0. All rights reserved.
Eutech Instruments Pte Ltd. Blk 55, Ayer Rajah Crescent #04-14/24, Singapore 139 949. Tel: (65) 778 6876; Fax: (65) 773 0836; e-mail:
marketing@eutechinst.com; Home page: http://www.eutechinst.com
TABLE OF CONTENTS
1 INTRODUCTION ................................................................................................ 1
1.1 Description of Unit.................................................................................... 1
1.2 Applications................................................................................................ 2
2 ASSEMBLY AND INSTALLATIONS............................................................. 3
2.1 Measurement and Control System........................................................ 3
2.2 Unit Dimensions.......................................................................................3
3 ELECTRICAL CONNECTIONS....................................................................... 4
3.1 Connection Diagram................................................................................ 4
3.2 Back Panel................................................................................................ 6
4 OVERVIEW....................................................................................................... 8
4.1 Keypad and Display................................................................................. 8
Function Groups................................................................................................... 9
4.2 Control Concept ....................................................................................... 9
5 MEASUREMENT............................................................................................ 10
5.1 Display in Measurem ent m ode ............................................................ 10
6 CALIBRATION............................................................................................... 11
6.1 pH Calibration......................................................................................... 11
6.2 ORP – mV Calibration........................................................................... 13
7 ADVANCED SET UP MODE......................................................................... 14
7.1 Electrode Offset (OFS) sub-function................................................... 14
7.2 Setting temperature (Set
7.3 Control Relay A/Control Relay B (SP1/SP2) sub-function............... 18
7.4 Configuration (ConF) sub-function ...................................................... 22
7.5 Calibration (CAL) sub-function............................................................. 23
8 TECHNICAL SPECIFICATIONS................................................................... 24
9 ACCESSORIES ............................................................................................... 25
10 GENERAL INFORMATION....................................................................... 25
10.1 Warranty.................................................................................................. 25
10.2 Packaging ............................................................................................... 25
10.3 Return of Goods..................................................................................... 26
10.4 Guidelines for Returning Unit for Repair ............................................ 26
11 APPENDICES.............................................................................................. 27
11.1 Appendix 1.............................................................................................. 27
11.2 Appendix 2.............................................................................................. 28
11.3 Appendix 3.............................................................................................. 29
o
C) sub-function ......................................... 15
1 INTRODUCTION
1.1 Description of Unit
Thank you for purchasing Eutech’s ¼ DIN alpha-800 series pH/ORP process controllers. This unit is used for measuring either pH or ORP parameter one at a time, and the oper atio nal m ode is switc hab le fr om the menu. You can use this unit to measure pH or ORP with limit control. This controller has many user-friendly and safety features which include:
Menu-driven program that simplifies set-up
Built-in non-volatile memory backup to ensure that calibr ation and
other information are not erased if power supply fails
Push-button two-point calibration and electrode offset adjustment from the keypad
Automatic temperature compensation (ATC)
Manual temperature comp ensation setting without the ATC probe,
with independent setting for calibration and process temperature
0 to 2000 second time delay adjustment on all relays – minimise false alarms
Separately adjustable high and low set point hysteresis (dead bands) prevent chattering of relays around the set points
Asymmetrical/symmetrical input for pH/ORP operation
Large dual display LCD for easy reading with clear multiple
annunciators, alarm status, operational and error messages
Two switching contacts as set-point triggering relays
Hold function freezes output current (4/20mA) and control relays
LED indicators signal control activities to monitor controller status
from a distance
Protection against electromagnetic interference – Input and
output are free from external inference
1
1.2 Applications
Use this controller in panel mounted enc losures for applica tions such as water treatment and monitoring, galvanic-decontamination, chemical processing food processing, clean or waste water control and
neutralisation process
.
2
2 ASSEMBLY AND INSTALLATIONS
(
)
(
r
)
y
p
2.1 Measurement and Control System
A typical measurement system consists of:
a pH/ORP process controller
a pH/ORP combination electrode with integrated or separate
temperature sensor Pt 100/1000,
an immersion, flow or pr ocess assembly with or without a potential matching pin (PMP)
a final control element such as pump or valve
a recorder
Flow Assemb
l
Measurement Cable
H Contr oller
Power Mains
Chart Record
er
2.2 Unit Dimensions
max. 175
1mm
max. 45
Flat Gasket
To be Inserted ByCustome
Note: The Taped Corners Have to Be On Top
92 + 0.5
92 + 0.5
MountingCut-Out
92
56
32
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54 (front).
96
3
3 ELECTRICAL CONNECTIONS
w
A
3.1 Connection Diagram
S/
V
S/S/
pH
AC:
Po
er Mains
1
18 19 20 21 22
Pt 100
Relay A
2
36
PE
NL
5
4
* ) indicated contact positionsare for currentlessc onditions
PE/S
Potential M atching Pin (PMP)
RelayB
78 9
SignalInput pH/ORP
10
11
BNC
12 13
Signal OutputHold Input
pH
m
+
-
16
151715
14
PE/S
4
PENL
Temp.
S
0/PA
20
21
123
22
19 18
S
pH
BNC
Symmetrical high-impedance
connections
Measuring Cable
Potential
Matching Pin (PM)
N
LS
PE
22
123
Potential
Matching Pin (PM)
0/PA
21
pH Combination
Electrode
Tem p .
20
19 18
pH
BNC
Asymmetrical connections
Measuring Cable
pH Combination
Electrode
5
3.2 Back Panel
The back panel consists of two conn ectors . The f irs t connector is the 17-wa y PCB edge connector and the other is the 5-way connector.
Connection for the 17-way screw terminals (from left to right):
1.AC mains live wire 10.No connection
2.AC mains neutral wire 11.No connection
3.AC mains protective earth wire 12.No connection
4.Low set relay resting position (NC) 13.Hold function switch terminal 1
5.Low set relay common 14.Hold function switch terminal 2
6.Low set relay working position (NO) 15.No connection
7.High set relay resting position (NC) 16.4 - 20 mA for -ve connection
8.High set relay common 17.4 - 20 mA for +ve connection
9.High set relay working position (NO)
Connections for the 5-way screw terminals:
18. Pt100/Pt1000 lead 1 terminal
19. Pt100/Pt1000 sense lead terminal
20. Pt100/Pt1000 lead 2 terminal
Note: If using a two-wire RTD, short terminal 19 to terminal 18. Pt 100/Pt 1000 is s electable via an internal jum per. Fac tory def ault is Pt1 00. See Appendix 1 for directions on switching the RTD type.
21. pH/ORP (potential matching pin)
21. pH/ORP (shield)
6
/
Pt100
/
/
Pt1000
0V PAL
S
FUSE 250VAC
(F)
100mA
LNPE
RELAY A
RELAY B
NC NCNC
HOLD
NC
+
-
7
4 OVERVIEW
4.1 Keypad and Display
4.1.1 Keypad
Perform rapid 2-point calibration and view electrode status
Allows entry to Set up mode
Selec t individual functions with in the function group of Set up
mode
Store input data in the Set up mode
Start calibration in the calibration mode
Select various function groups in the Set up mode.
Set parameters and n umerical values in sub functions of Set
up mode
If pressed continuously, the setting speed increases
Return to th e Meas urem ent m ode when bo th k eys are pr ess ed together
4.1.2 Display
The LCD display features two numerical displa ys t hat sh ow status messages and measured values for easy, quick reference. The dis play provides shor t­text information for setting parameters and configuration.
SETUP READY HOLD
MEAS
CAL
7.00
25.0 ATC
CON
pH
o
C
•••• MEAS: Measurement mode
•••• SETUP: Set-up mode of function groups
•••• CAL: Calibration mode of pH/ORP
•••• READY: Comes on after a successful
calibration
•••• HOLD: Relay position and c urrent output are frozen
•••• ATC: Comes on in the ATC mode. Disappears in the Manual temperature Compensation mode.
“ATC” flashes if the temperature probe is
faulty in its ATC mode
Display for RELAY A/B. Green LED indicates measured value within lim it while RED LED in dicates measured value outside limit.
Alarm display if limit value overshoot or the ATC probe fails.
8
Function Groups
The main function an d sub-function gr oups are orga nised in a matrix format for configuration and selection of parameters. The main function groups are:
1) Offset adjustment (OFS)
2) Temperature Measurement / compensation settings (Set
3) Control relay 1 configuration (SP1)
4) Control relay 2 configuration (SP2)
5) Configuration (ConF)
6) Calibration (CAL pH)
o
C)
4.2 Control Concept
The main function an d sub-function gr oups are orga nised in a matrix format as shown below. These functions can be access ed via the front keypad for configuration and selection of parameters.
9
5 MEASUREMENT
5.1 Display in Measurement mode
When the controller is initially powered on, it automatically enters into the Measurement mode after the large dual LCD displays all segments briefly.
The upper display s hows the measured pH or ORP valu e, while the lower display shows either the temperature value if the controller is set for pH measurement or “OrP” if it is set for ORP measurement.
Annunciators at the r ight s ide of the disp la y indicate whether the contr ol ler is set for pH or mV measurement..
Similarly annunc iat or s or ic o ns at th e t op or lef t side of the disp lay shows the current status of contro ller, e.g. “MEAS”, “S ETUP”, “CAL”, “READ Y”, etc. In addition, error messages also guide the user i n time of any fau lty conditions as indicated by “ERR” or icons flashing.
5.1.1 Check electrode performance
To read current electrode slope and offset values:
1) Press the CAL key followed by the ENTER key. The upper display shows electrode slop e. T he lower d ispla y read ing s hows t he p H r eadin g at 0 mV.
2) Press the
and ∇∇∇∇ keys to return to Measurement mode.
∆∆
10
6 CALIBRATION
Direct Calibration from the Measurement m ode is possible via the CAL key. The Calibration procedure c an also be accessed from the Advanced Setup mode.
6.1 pH Calibration
This unit features f ive preset buffer values (1.00, 4.00, 6.86 or 7.00, 9.00,
9.18 and 10.00) for qu ick, two-point auto calibratio n. When calibrating this instrument, standard pH buffer solutions must be used.
1) Enter Calibration mode. While in the Measurement mode, push the CAL key. The upper and lower display reads “CAL” & “pH”. Push the ENTER or CAL key again. The controller displays the electrode status based on the previous calibration. Electrode s lope and offset information is displayed. NOTE: If the display reads “CAL OrP”, see section
7.4 for procedures on how to switc h from ORP to p H readings.
2) Press the ENTER key again to begin calibration. The “CA L” indicator a ppears at th e top of display and the buffer ind icator app ear s in the bottom left corner of the dis p lay. The upper dis p lay is the on­line pH value as measured by the electrode. The lower display indicates the first compulsory
calibration value, pH 7.00 (or 6.86).
3) Immerse the clean e lectr ode in th e buffer solution. In ATC, the temperature probe must also be
SETUP
HOLD
READY HOLD
HOLD
SETUP
HOLD
READY HOLD
CAL
PH
CAL
57.2 mV
6.88
CAL
6.96
7.00
CAL
4.02
4.00
CAL
59.0 mV
7.00
pH
pH
pH
pH
pH
pH
immersed in the buffer solution. Likewise for
11
symmetrical mode, immerse the potential matching pin in the buffer.
4) Press the ENTER key to start the ca libration at pH 7 (or pH 6. 86). The electrode indicator and CAL indica tor both f lash. T he controller automatic all y adjusts the reading to match the buffer value.
5) The lower dis play will now show its nex t lower ‘pH’ buf fer. Use ∆∆∆ and ∇∇∇∇ keys to select the sec ond buffer f rom one of the 5 remaining preset va lues. Remove the electrode f rom the first buf fer, wash and then imm erse into the second buffer, Once the upper d isplay stab ilises pre ss the ENTER key, the electrode and CAL indicat or will flash. The controll er adjusts the reading to the buffer.
6) After calibrating to a second buffer value, the controller automatically displays electrode slope and offset. The zero point and slope are re­determined after each calibration.
7) Press the ENTER key. If the calibrat ion mode was access ed using the CAL key, the controller will return to the Measurement mode. If the calibration mode was accessed from the Advanced Set-up mode, the controller will return to the sub-function menu. Note: If there is a calibration error, the controller displays “ERR”. If this happens, push both and (escape) to restart the calibration beginning from step 1. Note: When calibrating with manual temperature compensation, the controller automatic ally changes from the preset process temperature to the calibration temperature. After leaving the Calibration mode, the controller switches back to proces s temperature (for settin g th e c al ibra tio n temperature and the process temperature, see section 7.2.3).
12
6.2 ORP – mV Calibration
This mode allows one-point calibration.
1) Enter Calibration mode. While in Measurement mode, push the CAL key. Press the ENTER key. The upper and lower display reads “CAL OrP”. Note: If the upper and lo wer display read “ CAL PH”, see section 7.4 for procedur es on ho w to s witch f rom
SETUP
HOLD
CAL
OrP
pH to ORP mV readings.
2) Place sensor in the ORP solution.
3) Press the ENTER key to begin calibration. The “CAL” indicator appears on the display. The upper display shows the current m V output of the electrode without any offset adjustment.
4) Press the ∆∆∆ and ∇∇∇∇ keys to adjust the m V value to that of the standard solution.
5) To confirm, press the ENTER key. The controller calculates the correction factor and returns to the Measurement mode.
HOLD
CAL
226
mV
13
7 ADVANCED SET UP MODE
7.1 Electrode Offset (OFS) sub-function
This option is available only in the pH mode. An allowance of ±120mV is allowed to correct for electrode deviations. This feature is useful for prolonged on-line applications.
The controller will add or subtract the value from the measured pH and display the correct value. Should the offset value be too high, consider performing maintenance on the electrode followed by calibration. If the calibration is unsuccessful, please replace the electrode.
SETUP
HOLD
SETUP
HOLD
Ofs
Ofs
OfsOfs
-0.05
-0.05
-0.05-0.05
7.00
5) Press the ENTER key. The SETUP annunciator appears at the top of the display. The lower displa y shows the on-l ine meas ured pH value. T he upper display shows the current offset value.
6) Press the ∆∆∆ or ∇∇∇∇ keys until the pH val ue coincides with the correc t pH value noted in Step 2. As the ∆∆∆ or ∇∇∇∇ keys is pressed to adjust th e pH valu e, the ‘offset’ value in the upper display also changes.
7) Press the ENTER key to enter the offset value. Note: The offset value is reset during full calibration . See section 6 for full calibration instructions.
1) Tak e a sample from the system and note t he pH reading displayed on the control ler .
2) Measure the pH of the sa mple using a calibrated pH tester, hand-held meter, or bench m eter. Record the correct pH value.
3) Enter Advanced set-up mode. Press ENTER.
4) Press the ∆∆∆ or ∇∇∇∇ keys to scroll until the upper display shows OFS.
Note: The offset parameter is blanked out during ORP operation.
14
7.2 Setting temperature (Set oC) sub-function
7.2.1 Selecting automatic or manual temperature compensation
1) Enter Advanced Set-up mode by pressing the ENTER key.
2) Press the ∆∆∆ or ∇∇∇∇ keys to scroll until the display shows “Set
3) Press the ENTER key. The lower display shows “Atc”; the upper display shows “on” or “oFF” depending on whether or not ATC is selected.
4) Press the ∆∆∆ or ∇∇∇∇ keys to toggle between ATC on and off.
5) Proceed with additional Advanced Set-up procedures (press ENTER again) or return to Measurement mode by pressing the ∆∆∆ and ∇∇∇∇ keys (escape) simultaneously.
o
C”.
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
7.2.2 Temperature calibration (ATC mode only)
1) Select “ATC on” as described above in Section 7.2.1.
2) Press the ENTER key. The upper display
SETUP
HOLD
SEt
SEt
SEtSEt
oooo
on
on
onon
Atc
oFF
oFF
oFFoFF
Atc
on
on
onon
indicates the current temperature offset. The current measured temperature is shown in the lower displa y.
3) Compar e the current m easured tem perature on the controller display to a thermometer known to be accurate. Note down the correct temperature value.
15
SETUP
HOLD
0.0
0.0
0.00.0
25.0
ATC
o
C
4) Press the ∆∆∆ or ∇∇∇∇ keys to scroll the lower display to match the correct value. The upper display will now show the offset value. You can offset
o
temperature up to ± 5
5) Press the ENTER key to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or return to the Measurement mode by pressing the ∆∆∆ and ∇∇∇∇ keys (escape) simultaneously.
C.
7.2.3 Setting manual temperature compensation
Note: This option is not available when the controller is set for ATC
SETUP
oFF
oFF
HOLD
oFFoFF
Atc
SETUP
25.0
25.0
HOLD
25.025.0
P.C
SETUP
25.0
25.0
HOLD
25.025.0
C.C
shows the current proces s temperature and the lo wer display shows “P. to indicate process temperature.
3) Press the ∆∆∆ or ∇∇∇∇ keys to adjust the process temperature value, bet ween –9.9 and 125
o
C.
operation. For manual temperature compensation, two different temperatures: process and calibration, can be input independently. This allows calibration at a temperature other than the process temperature. Example: setting a calibration temperature of 25 calibration using standard buffer solutions at
o
25
C, even if the process temperature is
different.
1) Select “ATC oFF” as described above in section 7.2.1
2) Press the ENTER key. The upp er disp la y
o
C allows
o
C”
4) Af ter settin g the pr ocess tem perature valu e, pr ess the ENTER ke y. The upper display shows the current calibration temperature and the lower display shows “C.
o
C”.
16
5) Press the ∆∆∆ or ∇∇∇∇ keys to adjust the calibration temperature value, between –9.9 and 125
o
C.
6) Press the ENTER key to confirm.
7) Continue with additional Advanced Set-up procedures, or return to Measurement mode by pressing the ∆∆∆ and ∇∇∇∇ keys (escape) simultaneously.
17
7.3 Control Relay A/Control Relay B (SP1/SP2) sub­function
The SP1 option sets th e operating parameters for Relay A; the SP2 option sets the operating parameters for relay B. Since these groups have the same set-up parameters, they are described together.
7.3.1 Entering the Set point 1 (Set point 2) sub-function
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
2) Press the ENTER key. The upper display shows the curr ent set point value and the lower display shows SP1 (SP2).
3) Press the ∆∆∆ or ∇∇∇∇ keys to enter the value for Set point 1 (Set point 2).
4) Press the ENTER key to confirm.
5) Proc eed to 7.3.3 step 3, or r eturn t o Meas urem ent m ode b y pressing th e ∆∆∆∆ and ∇∇∇∇ keys simultaneously (escape).
SP1
SP1
SP1SP1
4.00
SP1
Lo
SP1
pH
1) Enter Advanced Set-up mode. Push the ENTER key.
2) Press the ∆∆∆ or ∇∇∇∇ keys to scroll until the upper display shows SP1 (S P2).
7.3.2 Selecting the set point values
This lets you choos e the value that will cause your controller to activate the respective relays.
1) Follow directions in 7.3.1 to enter Control Relay mode. If you are in this mode, skip to step 2.
7.3.3 Choosing High or Low set points
Select a “lo” to activate relay wh en the pH/ORP value undersho ots the set point; select a “Hi” to activat e rela y when th e va lue ov ershoots the s et po int. Using both SP1 and S P2, the following configurati on for the set points can be selected: lo/lo, lo/hi, hi/lo, or hi/hi.
18
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press the ENTER key until the upper disp lay shows Lo or Hi and the lower display shows SP1 (SP2).
3) Press the ∆∆∆ or ∇∇∇∇ keys to select l ow (lo) or high (hi) set point for SP1 (SP2).
4) Press the ENTER key to confirm.
5) Proceed to 7.3.4 step 3, or retur n Measurement mode b y pressing the
∆∆∆∆ and ∇∇∇∇ keys simultaneously (escape).
7.3.4 Selecting a hysteresis (dead band) value
Hysteresis prevents rapid co ntact switching if your value is fluctuati ng near the set point. It does this by overshooting the set po int value to a specified hysteresis valu e. You can set the h ysteresis va lue from 0.1 to 1.0 pH, 10 to 100mV. Example: You have set your high set point at pH 9.0 and your hysteresis limit value at pH 0.5. If your m easured value overshoots the high s et point pH 9.0, the controller’s relay activates, which in turn activates external devices such as a pump or valve. The actions of the external device will cause the solution’s pH to drop. W hen the pH drops to 8.5 pH, t he rela y and hence the pump will switch off.
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆∆ or ∇∇∇∇
SETUP
HOLD
0.50
pH
keys until the upper display shows the h ysteresis (dead band) value and the lower display shows “HYS”.
3) Press the ∆∆∆ or ∇∇∇∇ keys to enter your hysteresis value for Set point 1 (S et point 2). Your controller will acti vate at the valu e you selec t.
4) Press the ENTER key to confirm your selection.
19
5) Proceed to 7.3.5 step 3, or return to Measurement mode by press ing the ∆∆∆ and ∇∇∇∇ keys simultaneously (escape). NOTE: Please refer to Appendix 3 for a graphical representation of the Hysteresis.
7.3.5 Setting an on-delay time lag
You can set as time delay for each relay, which stops the relay from switching on the moment the set p oint is exceeded. This c ontroller lets you set a 0 to 2000 second time delay before the relay activates.
SETUP
HOLD
0
On.d
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆∆ or ∇∇∇∇ keys until the upper displ ay shows “on delay” tim e
and the lower display shows “On.d”.
3) Press the ∆∆∆ and ∇∇∇∇ keys to enter o n-dela y tim e for Set p oin t 1 (Set point
2). The controller wil l delay activat ion for the n umber of seconds ( 0 to 2000) you select.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.3.6 step 3, or return to Measurement mode by press ing the ∆∆∆ and ∇∇∇∇ keys simultaneously (escape).
7.3.6 Setting an off-delay time lag
You can set a time delay for each relay, which delays the relay from switching off the moment the val ue re ac hes th e set point and hysteres is . T he controller allows a timer delay from 0 to 2000 seconds time delay before
deactivating.
SETUP
HOLD
0
OF.d
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆∆ or ∇∇∇∇ keys until the upper displa y shows “of f delay” tim e
and the lower display shows “OF.d”.
20
3) Press the ∆∆∆ or ∇∇∇∇ keys to enter of f-delay time for Set point 1 (Set point
2). The controller wil l delay activat ion for the n umber of seconds ( 0 to 2000) you select.
4) Press the ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to Measurement mode by pressing the ∆∆∆ and ∇∇∇∇ keys simultaneously (escape).
21
7.4 Configuration (ConF) sub-function
The options available in this sub-function allows the controller to be configured as a pH or ORP controller.
7.4.1 Entering the Configuration sub-function
1) Enter Advanced Set-up mode. Press the ENTER key.
2) Press the ∆∆∆ or ∇∇∇∇ keys to scroll until the upper display shows “ConF”.
7.4.2 Selecting pH or ORP mV measurement
SETUP
ConF
HOLD
SETUP
HOLD
SETUP
HOLD
PH
SY
PH
ASY
SETUP
HOLD
SETUP
HOLD
OrP
SY
OrP
ASY
mV
mV
Besides selecting the contro ll er to f unc tion as a pH or ORP ( mV) controller, it also allows the selection of the type of input: symmetric al or asymmetrical. Symmetrical m ode is used in conjunction with a potential matching pi n – in an electrically noisy environment. Please refer to section 3, page 6 for a graphical representation.
1) Follow directions in 7.4.1 to enter Configuration mode.
2) Press the ENT ER ke y. Scroll with the ∆∆∆ or ∇∇∇∇ keys until th e upper displa y shows the mode of op eration (pH or ORP mV and the lower displa y shows symmetrical (SY) or asymmetrical (ASY) input type.
3) Proceed to 7.4.3, or return to Measurement mode by pressing the ∆∆∆ and ∇∇∇∇ keys simultaneously (escape).
22
7.4.3 Reverting to factory default settings
Use this parameter to r es et all s ett ings t o f ac tory default. Changing f r om “ no” to “YES” and pressing the ENTER key resets all settings to factory default.
WARNING: If “Yes” is selected, all settings input will be overwritten!
1) Follow directions in 7.4.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the ∆∆∆ or ∇∇∇∇ keys until the upper display shows “no” or” YES”, and the lower display shows “deF” (default).
3) Press the ∆∆∆ or ∇∇∇∇ keys to select no or yes. Selecting “YES” a nd pres s ing the ENTER key will overwrite all user-settings and automatically return to Measurement mode.
4) Press the ENTER key to confirm your select ion and r eturn to Ad vanced Set-up mode, or return to Measur ement m ode by press ing the ∆∆∆ and ∇∇∇∇ keys simultaneously (escape).
SETUP
HOLD
no
dEF
7.5 Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to the procedure in the Calibration m ode. The only difference is that the control ler will revert back to Set-up mode (instead of Measurement mode) after calibration is com pleted.
7.5.1 Entering Calibration mode from Advanced Set-up mode
1) Enter Advanced Set-up mode. Press the ENTER key.
2) Press the ∆∆∆ or ∇∇∇∇ keys to scroll until the upper display shows “CAL”.
3) See section 6 for complete calibration procedures.
23
8 TECHNICAL SPECIFICATIONS
pH Range -2.00 to 16.00 pH Resolution 0.01 pH Relative Accuracy mV Range - 1000 to +1000 mV
Resolution 1 mV Relative Accuracy
Temperature - 9.9 to + 125.0 oC
Resolution 0.1 oC Relative Accuracy
Sensor Pt 100 /Pt 1000 (jumper selectable) Temperature Compensation Auto / manual (reference at 25.0 oC)
Set-point and Controller Functions
Function (switchable) Set-point controller Controller charac teristics Limit controller Pickup / Dropout delay 0 to 2000 sec. Switching pH hysteresis 0.1 to 1 pH Switching ORP hysteresis 10 to 100 mV Contact outputs, controller 2 potential-free change-over contacts Switching voltage max. 250 VAC Switching current max. 3A Switching power max. 600 VA
Electrical Data and Connections
Power Requirements 110 / 220 VAC (jumper selectable) Frequency 48 to 62 Hz Signal Output 4 to 20 mA, galvanically isolated Load pH / ORP input BNC Connection terminal Terminal blocks 5-pole / 17-pole, removable Mains fuse / fine wire fuse slow-blow 250 V / 100 mA
EMC Specifications
Emissions According to EN 50081-1 Susceptibility According to EN 50082-1
Environmental Conditions
Ambient temp. operating range 0 to +50 oC Relative humidity 10 to 95%, non-condensing
Mechanical Specifications
Dimensions (control panel housing - L x H x W) 175 x 96 x 96 mm Weights (control panel housing) max. 0.7 kg Material ABS with polycarbonate (front housing) Insulation (Front / Housing) IP 54 / IP 40
± 0.01 pH
± 1 mV
± 0.5 oC
max. 500
24
9 ACCESSORIES
Product Description Code no.
Ryton-body pH combination elec tr ode withou t AT C, without potential matching pin Ryton-body pH combination elec tr ode withou t AT C, without potential matching pin, high temperature and pressure rating Ryton-body pH combination elec tr ode withou t AT C, with potential matching pin Ryton-body pH com bination electr ode with ATC (Pt
100), with potential matc hing pin Ryton-body ORP Gold com bination electrode with potential matching pin Ryton-body ORP Platnium combination electrode with potential matching pin
Note: Above pH/ORP electrodes can withstand a pressure up to 6 bar
(except for ECARHTTSO05B – 9 bar). These electrodes have integral 5 meter measuring cable. Please ask your authorised distributor or dealer for the prices, extension cables, tee joints, electrode assembly and buffer solutions.
EC-ARTSO05B
EC-ARHTTSO05B
EC-ARGTSO05B
EC-100GTSO05B EC-HTAUTSO05B EC-HTPTTSO05B
10 GENERAL INFORMATION
10.1 Warranty
Eutech Instruments warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return by freight pre-paid. Eutech Instruments’ Customer Service Dept. will determine if problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
10.2 Packaging
The instrument is packaged in a corrugated box with a warranty card, instruction manual and the following accessories:
17-way and 5-way (right-angled) terminal block [1 unit each]
side threaded rod with catch [2 units]
receptacle cable lug [1 unit]
rubber gasket [1 unit]
25
10.3 Return of Goods
Authorisation must be obtained from Eutech Instruments’ Customer Service Dept. to issue a RGA (Return of Goods Authorisation) number before returning items for any reason. When applying for authorisation, please include data requiring t he reason of return. Items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments will not be responsible for any damage resulting from careless or insufficient packing.
Warning: Shipping damage as a result of inadequate packaging is
the user/distributor’s responsibility, whoever applicable. Please follow the guidelines below before shipment.
10.4 Guidelines for Returning Unit for Repair
Use the original packaging material, if possible when shipping the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Dept., if possible.
26
11 APPENDICES
11.1 Appendix 1
Jumper Positions - Internal to the controller
JP 1 Selects the input voltage 220 VAC. JP 2 Selects the input voltage 110 VAC. JP 3 Solder bridge selects between Pt100 and Pt1000.
Fuse Note that there is a fuse (slow-blow 100mA) internal
to the controller. Before opening the unit, ENSURE that the power cable is physically separated from the power supply. Replace fuse with the recommended type only.
27
11.2 Appendix 2
The following table shows the various pH values at different temperature of the solution during calibration.
Temperature
o
C)
(
0 1.67 4.01 6.98 7.12 9.47 10.32
5 1.67 4.01 6.95 7.09 9.38 10.25 10 1.67 4.00 6.92 7.06 9.32 10.18 15 1.67 4.00 6.90 7.04 9.27 10.12 20 1.68 4.00 6.88 7.02 9.22 10.06 25 1.68 4.01 6.86 7.00 9.18 10.01 30 1.69 4.01 6.85 6.99 9.14 9.97 35 1.69 4.02 6.84 6.98 9.10 9.93 40 1.70 4.03 6.84 6.97 9.07 9.89 45 1.70 4.04 6.83 6.97 9.04 9.86 50 1.71 4.06 6.83 6.97 9.01 9.83 55 1.72 4.08 6.83 6.97 8.99 9.81 60 1.73 4.10 6.84 6.98 8.96 9.79 70 1.74 4.12 6.85 6.99 8.92 9.76 80 1.77 4.16 6.86 7.00 8.89 9.74 90 1.80 4.20 6.88 7.02 8.85 9.73
pH 1.68
(oxalate)
pH 4.01
(phthalate)
pH 6.86
(tetraoxalate)
pH 7.00 (neutral
phosphate)
pH 9.18
(tetra borate )
pH 10.01
(carbonate)
28
11.3 Appendix 3
Simple Explanation on the Function of Hysteresis
SP1 Set to LO SP2 Set to HI
4.0 4.5 9.5 10.0
SP1
FORWARD DIRECTION REVERSE DIRECTION
The controller relay activates when the set-point is reached. In the reverse direction, it does not de-activate when the value reaches the set-point. Instead, it continues to be active till the value reaches the amount set by the Hysteresis band.
RELAY ON
RELAY OFF
7.0
SP2
HYSTERESIS BAND (DEFAULT = 0.5 pH)
29
CALIBRATION RECORD
Calibration Date Electrode Slope Electrode Offset
Loading...