Eutech instruments ALPHA COND 560 User Manual

Alpha COND 560
Controller
mS / µS / ºC / ºF
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO­BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO­STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, T RIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U.S. patent 6,168,707. PerpHecT ROSS electrodes are protected by U.S. patent 6,168,707. ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.ionplus electrodes and Optimum Results solutions ar e protected by U.S. patent 5,830,338.ROSS Ultra electrodes are protected by U.S. patent 6,793,787.ORP standard is protected by U.S. patent 6,350,367. No Cal electrodes are protected by U.S. patent 7,276,142.© 2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.This publication supersedes all previous publications on this subject.
Preface
This manual serves to explain the use of the Alpha COND 560 Controller. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover as many anticipated applications of the Alpha COND 560 Controller. If you have doubts in the use of the instrument, please do not hesitate to contact your nearest Alpha Authorised Distributor.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Thermo Scientific.
Thermo Scientific cannot accept any responsibilit y for damage or malfunction of the unit due to improper use of the instrument.
Copyright © 2009 All rights reserved
TABLE OF CONTENTS 1 INTRODUCTION 1
1.1 Before You Begin................................................................................... 1
1.2 Intended Use ......................................................................................... 1
1.3 Safety Instructions .................................................................................2
1.4 Taking Out of Service / Correct Disposal of the Unit .............................. 2
2 GETTING STARTED 3
2.1 Description of Instrument.......................................................................3
2.2 Measurement System............................................................................ 4
2.3 Connecting Peripherals..................................................................... 5
2.4 Installation ...........................................................................................8
2.5 Di s p l ay & K e yp a d ............................................................................... 10
3 OPERATION 13
3.1 Measurement mode............................................................................... 13
3.2 Menu Overview...................................................................................... 14
4 CALIBRATION MODE 15
4.1 About Calibration ................................................................................... 15
4.2 Calibration.............................................................................................. 16
5 SETUP MODE 18
5.1 Enter Setup mode.................................................................................. 18
5.2 Temperature Coefficient Settings...........................................................19
5.3 Temperature Settings ............................................................................ 20
5.4 Relay 1/Relay 2 Settings........................................................................22
5.5 Controller Settings ................................................................................. 24
5.6 Measuring Range Settings..................................................................... 26
5.7 Configuration Settings............................................................................27
5.8 Viewing Electrode Properties................................................................. 30
6 TECHNICAL SPECIFICATIONS 31 7 LIST OF ACCESSORIES 34
7.1 Thermo Scientific Order Codes.............................................................. 34
7.2 Eutech Instruments Order Code ............................................................ 34
8 TROUBLESHOOTING 35 9 GENERAL INFORMATION 36
9.1 Warranty................................................................................................ 36
9.2 Return of Goods.....................................................................................36
9.3 Guidelines for Returning Unit for Repair................................................ 36
9.4 Maintenance and Cleaning .................................................................... 37
10 APPENDICES 38
10.1 Appendix 2 Conductivity of various aqueous solutions.........................38
10.2 Appendix 2 Graphical Representation of the Function of Hysteresis....39
10.3 Appendix 3 General Instructions Concerning Controller Setting...........40
10.4 Appendix 4 - Abbreviations Used in LCD...............................................41

1 INTRODUCTION

1.1 Before You Begin

Thank you for purchasing the Alpha COND 560 Controller. The construction of the Alpha COND 560 Controller employs leading edge
technology and complies with safety regulations c urrently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment. Therefore, the operating instructions must be read and understood by the persons involved before working with the Conductivity Controller.
The instruction manual must always be stored close at ha nd, in a place accessible to all people working with the Conductivity Controller.
If you have questions, which are not or insufficiently answered in this instruction manual, please contact your authorized supplier. They will be glad to assist you.

1.2 Intended Use

Alpha COND 560 Controllers are intended solely for Conductivity and temperatur e measurement, as described in this instruction manual.
Any other use, or use not mentioned here, that is incompatible with the technical specifications is deemed inappropriate. The oper ator is solely responsible for any damage arising from such use.
Other prerequisites for appropriate use include: – Comply with the instructions, notes and requirements set out in this instruction
manual.
Comply with all local safety regulations concerning safety at work. Comply with all information a nd warnings in the documentation de aling with the
products used together with the Conductivity Controller (housing, sensors, etc.).
– Comply with the local environmental and operational conditions.
- 1 -

1.3 Safety Instructions

The Alpha COND 560 Controller should be installed and operated only by personnel familiar with the instrument and who are qualified for such work.
A defective Conductivity Controller must neither be installed nor put into service.
The Alpha COND 560 must only be operated under the specified operating conditions (see section 6).
The Alpha COND 560 must not be repaired by the customer.
No modifications to the Alpha COND 560 are allowed. The manufacturer/supplier accepts no responsibility for damage caused by unauthorized modifications. The risk is borne entirely by the user.

1.4 Taking Out of Service / Correct Disposal of the Unit

Taking out of Service
First disconnect the unit from the power supply and then undo all electrical connections.
Remove the unit from the wall.
Correct Disposal of the Instrument
When the Alpha COND 560 is permanently taken out of service, obey the local environmental regulations for correct disposal or send the instrument to your local distributor, they will take care of proper disposal.
- 2 - - 3 -

2 GETTING STARTED

2.1 Description of Instrument

The Eutech Instruments’ Alpha CON 560 Controller is used for measuring conductivity and temperature values. The controller can be used for applications such as water treatment and controlling, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
The controller has many user-friendly and safety features which include:
Push-button keypad for calibration and setup
Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails
Menu-driven program that simplifies setup
Automatic temperature compensation (ATC)
Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
Large dual display LCD for e asy reading with clear multiple annunciators, operational mode indicators and error indicators
Two switching contacts as set-point relays
Limit (on/off) control mode
Separately adjustable high and low set-point hysteresis (dead bands) to
prevent chattering of relays around the set points
0 to 2000 seconds time delay adjustment on all relays – minimizes false alarms
Backlight LCD display

2.2 Measurement System

A typical measurement system consists of:
A conductivity process controller
A conductivity sensor with integrated or separate Pt100/Pt1000
temperature sensor
An appropriate measurement cable
An immersion, flow or process assembly
A final control element such as pump or valve
Alpha COND 560
Alpha CO N 5 60 Controller
Alpha COND 560
lpha pH560
ESC
CAL
Conductivity Controller
Conductivity Controller
ENT
pH Monitor
Pump or Valve
Relay Output
Housing and Sensors
Measur ement Ca b le
Power Adaptor (+9 V DC)

2.3 Connecting Peripherals

2.3.1 Connection Terminals Remove Back Cover:

Remove the screws from the four corners at the back of the conductivity controller. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below.
Connectors:
J11 – 24V DC power connector
J8 – 9V DC power connector (1.2mm diameter)
J9 – Relay connections
J10 - Conductivity electrode & Temperature probe connections (wiring
has to be done in the detachable connector
J11 Connections
1. +ve
2. Gnd
J10 Connections
1. CON Low Current
2. CON Low Voltage
3. CON High Voltage
4. CON High Current
5. Pt 100 Compensate
6. Pt 100 Sense
7. Pt 100 GND
8. (No connection)
J9 Connections
Relay 1 & 2
Figure 1: Outer Side of Back PCBA
1)+ve
2)Gnd
24V DC
9V DC Power
J8
J9
REL 1
REL 2
J11
Screw
J10

2.3.2 Switching Between PT100 & PT1000 Temperature Sensors

The controller supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below.
Remove Back Cover:
Remove screws from the four corners at the back of the conductivity controller. Remove the back cover.
Remove Back PCBA:
Remove the screw located center of the back PCBA (Figure 1). Detach the back PCBA from the controller. Turn over the back PCBA. Locate J7 jumper o n the inner side of the back PCBA as shown in figure 2 below.
Figure 2: Inner Side of Back PCBA
Set Jumper J7:
Set the J7 jumper to required sensor (Pt100 or Pt1000) type

2.3.3 Connecting Conductivity Electrode

Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity controller.
2-Cell Conductivity Probe:
1. Short Pin 1 & 2 of J10 connector using a jumper wire.
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (black) to Pin 3 of J10 conn ector.
4. Short Pin 3 & 4 of J10 connector using a jumper wire.
4-Cell Conductivity Probe:
5. Connect CON low input wire (black) to Pin 1 of J10 con nector
6. Connect CON low input wire (white) to Pin 2 of J10 connector.
7. Connect CON high input wire (orange) to Pin 3 of J10 connector.
8. Connect CON high input wire (blue) to Pin 4 of J10 con nector.

2.3.4 Connecting Temperature Probe

For Automatic Temperature Compensated (ATC) conductivity readings, an in-built temperature sensor is usually integrated with industrial conductivity probes. Alternatively, a separate 100Ω Pt RTD temperature probe (2-wire or 3-wire) can be connected to the controller.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J10 connector.
2. Connect PT100 sense wire to Pin 6 of J10 connector.
3. Connect PT100 GND wire to Pin 7 of J10 co nnector.
2-Wire Probe:
4. Short Pin 5 & 6 of J10 connector using a jumper wire.
5. Connect PT100 sense wire to Pin 6 of J10 connector.
6. Connect PT100 GND wire to Pin 7 of J10 co nnector.

2.4 Installation

2.4.1 Mechanical Dimensions

Alpha COND 560
CAL

2.4.2 Wall Mount

P
Alpha pH 600
ESC
H/ORP RF Transmitter
Conductivity Monitor
ENT
Pierce through holes at both sides
11
Cover the catch slots at both sides
2
with overlays
f
f
A
r

2.4.3 Panel Mount

Prepare panel cut-out o
1
1
92.0 mm X 92.0 mm
Remove back cover o
2
pH Monitor and slide it through panel cut-out
Panel
Gasket
Panel
Insert threaded rods through catch
4
until pH Meter is held against panel
ttach catch to both sides
3
of pH Mete

2.5 Display & Keypad

2.5.1 Display Overview

The Liquid Crystal Display (LCD) of Alpha COND 560 conductivity controller has two alpha-numerical displays (Upper and a Lower).
Upper display: Measured conductivity value is d ispl aye d when the c ontr oller is
in normal operation (measurement) mode.
Lower display: Measured temperatur e value is displayed when the controller
is in normal operation (measurement) mode. In conductivity calibration mode, measured conductivity values are displayed here.
The two displays indicate function names, options & settings in Setup m ode. Refer ‘Appendix 4 - Abbreviations Used in LCD’ for more details.
The LCD also consists of various mode indicators, status annunciators and unit of measurement indicators.
Upper Display
L
SETUP MEAS
READY HOLD
ERR
K=
8.8.
-
-
CA
8.8
.8
Low
.8.8
er Display
R.
mS
µS
°C ATC
%
°F
Mode Indicators MEAS SETUP
CAL Status Annunciators
READY HOLD
ATC
ERR
K
Measurement mode (Refer Section 3.1 ) Setup mode (Refer Section 4.1) Calibration mode (Refer Section 4)
Appears when the reading is stable Appears in Setup mode and Calibration mode to indicate that the relay
function is frozen Appears when Automatic Temperature Compensation (ATC) is enabled.
Not visible when Manual Temperature Compensation (MTC) is enabled. Flashes if the temperature probe is faulty in its ATC mode. (Refer Section 5.3)
Appears when an error occurs Electrode annunciator. Blinks during calibration error (Refer Section 4.2)
Buffer annunciator. Appears in calibration mode (Refer Section 4.2) Cell Constant (Refer Section 4.2)
Units of Measurement Indicators mS µS R. %
ºC ºF
Conductivity in Milli Siemens Conductivity in Micro Siemens Line Resistance in Ohms (Refer Section 5.7) Temperature Coefficient (Refer Section 5.2)
Temperature in Celsius (Refer Section 5.3) Temperature in Fahrenheit (Refer Section 5.3)

2.5.2 Key Functions

Alpha COND 560
lpha CON560
ESC
Key Description
CAL
Enter Calibration mode
Enter Setup mode Access sub screens (parameters) within a group of settings in Setup
ENT
mode Confirm (save) setup parameters and numerical values Start/Confirm calibration in Calibration mode. Select a group of settings in Setup mode. Select parameters and increment/decrement numerical values in Setup
and Calibration modes. (When pressed continuously, speed of value increment/decrement
increases) Returns to Measurement mode when both keys are pressed
simultaneously.
CAL
Conductivity Monito
ENT
r

3 OPERATION

3.1 Measurement mode

When the conductivity controller is powered on, the display shows all the LCD segments briefly, and then automatically enters into the measurement mode.
MEAS
µS
READY
°C
25.0
The mode indicator ‘MEAS’ at the top of the display indicates that the conductivity controller is in measurement mode. The upper alpha-num erical display shows the measured conductivity value, while the lower display shows the temperature value.
The controller has six selectable measuring ranges. The unit of measurement and resolution of the conductivity reading depend on the selec ted measuring range. If the controller reads a conductivity reading b eyond the selected measuring range, the display shows ‘Or’ (over range) error indicator.
NOTE:
From measurement mode you can access:
For more details, refer section 4 for Calibration mode and section 5 for Setup mode.)
To guarantee accurate readings, the measuring system (the conductivity controller and the sensor) must be calibrated regularly.
Calibration mode (by pressing CAL key)
Setup mode (by pressing ENT key)
ATC
Y
A

3.2 Menu Overview

READ
MEAS
ENT
µS
CAL ENT
°C
TC
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT

4 CALIBRATION MODE

4.1 About Calibration

Calibration should be carried out each time a new electrode is attached to the controller or when you suspect that the controller/electrode is out of calibration. T he controller allows you to perform temperature calibration and conductivity calibration.
About Temperature Calibration
It is important to ensure that temperature calibration is carried out prior to conductivity calibration since temperature readings affect the accuracy of conductivity measurements. It is recommended that temperature calibration should be carried out only if the temperature value displayed on t he controller is different from that of a calibrated thermometer. A temperature offset calibration of ± 10 °C/± 18 °F from the default reading is allowed. Once a temperature calibration is performed, conductivity calibration should be carried out to ensure the accuracy of conductivity measurements.
About Measuring Ranges & Conductivity Calibration
The controller has six selectable conductivity measuring ranges. You can only calibrate one-point in the measuring range th at you have selected for your proces s application. You have to recalibrate the controller every time you change the measuring range since the calibration data of the controlle r will be reset every time a new measuring range is selected.
About Conductivity Calibration
During conductivity calibration, the controller allows a calibration window of ± 40% from the measured default reading of the calibration standard. The minimum allowable calibration point is 10% of the full scale reading of the measur ing range selected.
For best results, select a standard conductivity solution close to the sa mple value you are measuring.

4.2 Calibration

The controller requires one-point calibration. Conductivity calibration is always carried out in the specific measuring range selected for the controller. M ake sure, the controller is set to desired measuring range before you enter cal ibration mode (Refer Section 5.6 for selecting measuring ranges).
Select a suitable standard conductivity solution so that its conductivity value is within 10% to 100% of the full scale of the selected measuring range of the controller. (For better accuracy, select the calibration standard to be nearly 50% of the full scale of the measuring range selected.) Prepare calibration standard solution in two beakers – one for rinsing and the other for calibration. Prepare separate de-ionized water for electrode rinsing.
All new calibration values will automatically override existing calibration data. If you wish to abort the calibration, press and keys simultaneously and the controller reverts to measurement mode and the controller will revert to previous calibration data.
MEAS
READY
11 2
3
OR
2
HOLD
K=
HOLD
K=
1
µS
CAL
HOLD
°C
ATC
HOLD
From measurement mode press CAL key to enter calibration mode. The display briefly shows the cell type (“2 CELL” or “4 CELL”) to which
the controller is configured (Refer section 5.7 for cell type settings)
2
CAL
CAL
ENT
READY HOLD
2
READY HOLD
3
CAL
µS
READY HOLD
3
CAL
READY
µS
ENT
Next, the last configured cell constant (k) is displayed in the upper display. Use or keys to adjust the cell constant to mach the cell constant of the conductivity probe. Adjustable range is 0.10 to 10.0. Press ENT key to confirm the new cell constant value. Rinse the electrode in de-ionized water and then rinse with the calibration standard. Immerse the probe in calibration standard solution intended for calibration and swirl gently to create a homogenous sample. Make sure there are no any air-bubbles trapped in the sensor.
The upper display shows the currently measured conductivity value of the solution based on last calibration (if any). The lower display shows
MEAS
4
CAL
ENT
µS
°C ATC
uncalibrated (default) reading of the solution. The buffer annunciator appears in LCD. The ‘READY’ annunciator appears when the reading is stable. Once the reading stabilizes press the or key to adjust the upper display value to that of the standard solution. Press the ENT key to accept the value.
NOTES:
Calibration is completed. The display shows the calibration factor
4
(FCt). This serves as an indication of the effectiveness of the sensor which will degrade with time and usage. An effective sensor should have a calibration factor of 0.60 to 1.40 Press ENT key to return to measurement mode.
The acceptable calibration window is ±40% of the displayed (default) value. For example, if the display is 1000 µS, the values to which it can be adjusted is 600 to 1400 µS.
If the conductivity reading of the solution is not within 10% to 100% of the full scale of the selected measuring range, calibration error occurs; LCD displays “ERR” and flash electrode annunciator. If the conductivity reading of the solution is beyond the selected measuring range, calibration error occurs; LCD displays “Or” (over range), “ERR” and flashes electrode annunciator. If calibration error occurs, exit form calibration mode by pressing and keys simultaneously.
NOTES:
If you use a 2-cell conductivity sensor, make sure that the pin 1 and 2 are
shorted and pin 3 and 4 are shorted on the J10 connector. (Refer section 2 for more details)
When calibrating with manual temperature compensation, the controller
automatically changes from the preset process temperature to the calibratio n temperature. After leaving the Calibration mode, the controller switches back to the process temperature (for setting the calibration temperature and the process temperature, see section 5.3)
WARNING: If you change the measuring range of the controll er after calibration is done, the controller will reset back to its factory default calibration.

5 SETUP MODE

5.1 Enter Setup mode

The setup mode allows you to customize the settings of the conductivity controller to suite your requirements. While in measurement mode, press the ENT key to access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup (rAng – Range settings). Press ▲ or ▼ key to access other pages of the setup mode.
READY
MEAS
µS
°C ATC
ENT
SETUP
HOLD
ENT
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
NOTES:
To exit setup mode at any time press ▲ and keys simultaneously (escape). The controller returns to the
measurement mode
If you do not press any key for 2.5 minutes, the controller automatically exists from the setup mode.

5.2 Temperature Coefficient Settings

Setting the temperature coefficient to match your process/calibration li quids allows controller to achieve conductivity readings with higher accuracy.
SETUP
HOLD
1
SETUP
ENT
HOLD
SETUP
HOLD
2
SETUP
ENT
HOLD
2
ENT
3
SETUP
% %
ENT
HOLD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode
4
ENT
(rAng). Press or key to select temperature coefficient settings screen (tC).
Press ENT key to access temperature coefficient settings (tC). Select type of temperature coefficient: The upper display shows the
2
last configured temperature coefficient type (Lin or PUr). Press or key to select the desired type based on type of your process liquid.
PUr : Pure water temperature coefficient: Select this if your
process liquid is pure water or ultra pure water.
Lin : Linear temperature coefficient: Select this for all other
process liquids
Press ENT key to confirm your selection. If PUr was selected: The controller reverts back to tC screen If Lin was selected: Select temperature coefficient for process liquid: The lower display
3
shows ‘P.tC’. The upper display shows last configured temperature coefficient for process liquid (if any). The default value is 2.1%. Press
or key to select the desired value. Allowable range: 0 to 10%. Press ENT key to confirm your value. Select temperature coefficient for calibration liquid: The lower
4
display shows ‘C.tC’. The upper display shows last configured temperature coefficient for calibration liquid (if any). The default value is
2.1%. Press or key to select the desired value. Allowable range: 0 to 10%. Press ENT key to confirm your value. The controller reverts to tC screen.
Press or key to access other setup screens or press and ▼ key simultaneously (escape) to return to measurement mode.
A

5.3 Temperature Settings

Automatic Temperature Compensation (ATC):
The conductivity value of a sample is affected by temperature. Use ATC feature of the controller to compensate for conductivity changes when the temperat ure of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the controller if an integrated conductivity/Temperature probe is not available.
Manual Temperature Compensation (MTC):
Set the controller to MTC (disable ATC) when the temperature of sample or process liquid is constant and a temperature probe is not available. If you disable AT C, the controller allows you to set your process temperature (P.ºC) and calibration temperature (C.ºC). This allows calibration at a different temperature other than the process temperature. Example: Setting calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your process temperature is different from 25°C.
SETUP
HOLD
1
SETUP
ENT
HOLD
2
SETUP
ENT
HOLD
°C
3
SETUP
ENT
HOLD
°C
4
SETUP
HOLD
°C
TC
5
ENT
°C ATC
SETUP
HOLD
2
SETUP
HOLD
ENT
°F
3
SETUP
ENT
HOLD
°C
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode
6 7
SETUP
ENT
HOLD
ENT
(rAng). Press ▲ or ▼ key to select Temperature settings screen (SET ºCF).
Press ENT key to access temperature settings (SET ºCF). Selecting unit of measurement for temperature: The upper display
2
shows ‘Unit’ and the lower display shows the last configured unit of measurement for temperature.
Press or key to select the desired units for temperature (ºC or ºF). Press ENT key to confirm your selection.
Enable/disable ATC: The lower display shows ‘AtC’ and the upper
3
display shows the last configured ATC selection (‘On’ or ‘OFF’).
Press or key to enable (ATC On) or disable (ATC OFF)
automatic temperature compensation. Press ENT key to confirm your selection.
If ATC enabled (ATC On): Selecting temperature offset: The upper display shows the last
4
configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner.
Place a thermometer, which is known to be accurate, in your sample or process liquid. Make sure your temperature probe is placed in the same liquid. Compare the stabilized temperature read ing displayed on the controller with the thermometer. If there is a difference between the two readings (offset), you can adjust the reading of the controller. Press or key to adjust the lower display to the correct temperature value.
As the lower display value changes, the controller adjust the upper
5
5
display reading automatically to suit the new offset value. Up to ± 10 ºC/ ± 18 ºF offset is allowed.
Press ENT key to confirm the value. The controller reverts to SET ºCF screen.
If ATC disabled (ATC OFF): Setting process temperature: The lower display shows ‘P.ºC’ and the
6
upper display shows the last configured process temperature. Press or key to adjust the upper display to desired process temperature.
Allowable range: –10.0 to 125.0°C / 14.0 to 257.0°F.
Press ENT key to confirm the process temperature. Setting calibration temperature: The lower display shows ‘C.ºC’ and
7
the upper display shows the last configured calibration temperature. Press or key to adjust the upper display to desired calibration temperature. Allowable range: –10.0 to 125.0°C / 14.0 to 257.0°F.
Press ENT key to confirm the calibration temperature. The controller reverts to SET ºCF screen.
Press or key to access other setup screens or press and ▼ key simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously (escape). The controller returns to the first screen of temperature settings SET ºCF.

5.4 Relay 1/Relay 2 Settings

NOTE:
The SP1 settings determine the operating parameters for Relay 1; while SP2 settings define the operating parameters for Relay 2. Since these two settings have identical setup parameters, only SP1 is described here.
Note that when you change measuring range of the controller, the values you have set for SP1 & SP2 will not be affected. However, the unit of measurement & resolution displayed in LCD for SP1 & SP2 will change according to the newly selected measuring range.
If controller type is set to ‘OFF’ (Refer section 5.5 for controller settings), the ‘Relay 1/Relay 2 Setting’ is not available.
SETUP
HOLD
1
SETUP
ENT
HOLD
2
SETUP
µS
ENT
HOLD
SETUP
HOLD
3
SETUP
µS
ENT
HOLD
3
µS
ENT
4
SETUP
µS
ENT
HOLD
SETUP
HOLD
5
ENT
6
ENT
From measurement mode press ENT key to enter setup mode as
1
1
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press or key to select relay 1 (Set point 1) settings screen (SP1).
Press ENT key to access relay 1 settings (SP1). Setting set point value: The upper display shows the last configured
2
set point value, otherwise the default value 100 µS. Press or key to enter the value for set point 1, at which control function of the conductivity controller should activate. Press ENT key to confirm your set point value.
Select relay function: The upper display shows the last configured
3
relay function (LO = Low or HI=High). Press or key to select the desired function.
Press ENT key to confirm your selection.
NOTE:
Setting hysteresis value: The lower display shows ‘HYS’. The upper
4
display shows the last configured value, otherwise the default value 20 µS. Press or key to enter the hysteresis value for set point 1. Allowable range: 0 to 10% of full-scale of the selected measuring range.
Press ENT key to confirm your hysteresis value. NOTE:
Setting on-delay time: The lower display shows ‘On.d’. The upper
5
5
display shows last configured on-delay value. Press or key to enter the on-delay time for set point 1. Allowable range: 0 to 2000 seconds. When the conductivity reading is reached the set point 1, the controller delays the relay activation (switch on) by the on-delay time you set here.
Press ENT key to confirm your on-delay time. Setting off-delay time: The lower display shows ‘OF.d’. The upper
6
display shows last configured off-delay value. Press or key to enter the off-delay time for set point 1. Allowable range: 0 to 2000 seconds. When the conductivity reading is reached the set point 1 plus hysteresis value, the controller delays the relay deactivation (switch off) by the off-delay time you set here.
Press ENT key to confirm your off-delay time. The controller reverts to SP1 screen.
Press or key to access other setup screens or press and ▼ key simultaneously (escape) to return to measurement mode.
This parameter lets you choose the relay function. Select ‘LO’ to activate the relay when the conductivity reading undershoots the set point (SP1) or select ‘HI’ to activate the relay when the conductivity reading overshoots the set point (SP1). SP1 and SP2 can be selected as ‘LO/LO’, ‘LO/HI’, ‘HI/LO’, or ‘HI/HI’.
Hysteresis prevents rapid relay contact switching if the conductivity reading is fluctuating near the set point. Refer ‘Appendix 2 Graphical Representation of the Function of Hysteresis’.

5.5 Controller Settings

The controller settings allow you to set the relay type and relay status of the controller.
SETUP
HOLD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup
1
SETUP
ENT
HOLD
SETUP
HOLD
2
ENT
2
SETUP
ENT
HOLD
SETUP
HOLD
mode (rAng). Press or key to select controller settings screen (Cntr).
Press ENT key to access controller settings (Cntr). Select the controller type: The lower display shows ‘tyP’. The upper
2
display shows the last configured controller type. (‘OFF’ or ‘L.Ct’). Press or key to select the desired controller type.
OFF : Controller Off
Selecting ‘OFF’, the conductivity controller stops functioning as a controller by preventing relay functions. Once the controller type is set to ‘OFF’, the setup mode does not show you the Relay 1/Relay 2 settings.
3
ENT
3
ENT
L.Ct: Limit value control (Controller On)
Selecting ‘L.Ct’, the conductivity controller is set to function as a controller. When the controller type is set to ‘L.Ct’, the Relay 1/Relay 2 settings are available in the setup mode for you to configure relay operating parameters.
Press ENT key to confirm your selection.
If Controller type was set to Off (OFF): The controller reverts to Cntr screen. If Controller type was set to On (L.Ct): Select the relay status under Non-Alarm condition: The lower
3
display shows ‘rEL’. The upper display shows the last configured relay status. (‘dEEn’ or ‘En’). Press or key to select the desired relay status.
dEEn: En:
Press ENT key to confirm your selection. The controller reverts to Cntr screen.
De-energized Energized
Press or key to access other setup screens or press and ▼ key simultaneously (escape) to return to measurement mode.

5.6 Measuring Range Settings

The conductivity controller has 6 different measuring ranges. You need to select a range which is the most suitable for conductivity reading of your applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm).
The measuring range you select here will affect the way the control ler behaves in the measurement mode and calibration mode. In measurement mode, if the conductivity reading of the controller is above the selected measuring range, the LCD shows “Or” (over range) error. In calibration mode, the controller adjusts calibration tolerance based on full scale of the selected measuring range. The calibration is carried out only for the selected measuring range.
WARNING: If you change the measuring range of the controll er after calibration is done, the controller will reset back to its factory default calibration.
From measurement mode press ENT
11
key to enter setup mode as described in
section 5.1. The LCD shows the first screen of setup mode (rAng). Press ENT key to access measurement range setting (rAng). The upper display shows the currently
2
selected measuring range. The factory default is 2000 µS. The lower display shows the currently configured cell constant (k). (Refer section 4.2 for setting cell constant) Press or key to select the desired measuring range. Press ENT key to confirm your selection.
The controller reverts to rAng screen.
Measuring range Resolution
0.000 to 2.000 µS/cm 0.001 µS/cm
0.00 to 20.00 µS/cm 0.01 µS/cm
0.0 to 200.0 µS/cm 0.1 µS/cm 0 to 2000 µS/cm 1 µS/cm
0.00 to 20.00 mS/cm 0.01 mS/cm
0.0 to 200.0 mS/cm 0.1 mS/cm
Press or key to access other setup screens or press and ▼ key simultaneously (escape) to return to measurement mode.
SETUP
HOLD
1
SETUP
ENT
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
2
µS
ENT
2
µS
ENT
2
µS
ENT
2
µS
ENT
2
mS
ENT
2
mS
ENT

5.7 Configuration Settings

Configuration settings let you to enable/disable backlight of the LCD, select between 2-cell & 4-cell and reset the controller to factory defaults.
SETUP
HOLD
1
From measurement mode press ENT key to enter setup mode as
1
1
described in section 5.1. The LCD shows the first screen of setup mode
SETUP
ENT
HOLD
SETUP
HOLD
2
2
SETUP
ENT
HOLD
SETUP
HOLD
ENT
SETUP
ENT
HOLD
SETUP
HOLD
SETUP
HOLD
3
3
5
5
5
SETUP
ENT
HOLD
This screen appears only
when 2-cell is selected
and measuring range is set to
0.0 to 200.0 mS/cm
ENT
READY
ENT
ENT
MEAS
(rAng). Press or key to select configuration settings screen (Cnfg).
Press ENT key to access configuration settings (Cnfg). Enable/disable LCD backlight: The upper display shows ‘LitE’. The
2
lower display shows last configured backlight selection (‘On’ or ‘OFF’). Press or key to enable (LitE On) or disable (LitE OFF) backlight
4
R.
ENT
µS
°C
ATC
of the LCD.
Press ENT key to confirm your selection. Select cell type of the sensor: The lower display shows ‘CEL’. The
3
upper display shows the last configured cell type (‘2’ or ‘4’). Press or key to select the cell type of your conductivity probe. Press ENT key to confirm your selection.
NOTE: If you use a 2-cell conductivity sensor, make sure that the pin 1
and 2 are shorted and pin 3 and 4 are shorted on the J10 connector. (Refer section 2 for more details)
Setting line resistance: The lower display shows ‘L.Ad’ and the upper
4
display shows the last configured line resistance; otherwise the default is 0.00 R.
NOTE: Line resistance of sensor cable constitutes to an error in measurement of high conductivity values. The controller will compensate this error if you specify line resistance value in this setting. The setting is available only if the cell type is set to 2-cell (step 3 above) and the measuring range is selected to ‘0.0 to 200.0 mS/cm’.
Press or key to set the line resistance of the sensor cable connected to the controller.
Press ENT key to confirm your selection. Reset to factory defaults: The display shows ‘nO dEF’.
5
5
Press or key to select:
nO dEF :
FCt dEF:
CAL dEF:
NOTE: Refer section 6 for factory/calibration default settings Press ENT key to confirm your selection.
- If ‘nO dEF’ was selected, the controller reverts to Cnfg screen.
- If ‘FCt dEF’ or ‘CAL dEF’ was selected, the controller performs the selected reset and returns to measurement mode.
Not default reset of the controller. User configured values remain active, when confirmed by pressing ENT key.
Reset all settings of the controller to factory defaults, when confirmed by pressing ENT key.
Reset all calibration settings of the controller to factory defaults, when confirmed by pressing ENT key.
Press or key to access other setup screens or press and key simultaneously (escape) to return to
measurement mode.
Press ▲ or ▼ key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously (escape). The controller returns to the first screen configuration settings (Cnfg).

5.8 Viewing Electrode Properties

Each time you calibrate the conductivity sensor, the controller re-calculates the calibration factor of the sensor and shown in the LCD at the end of calibration. The setup mode allows you to view the last saved calibration factor and cell constant at any time.
SETUP
HOLD
1
ENT
SETUP
HOLD
2
ENT
SET
HOLUPD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press or key to select viewing electrode properties screen (CdAt).
Press ENT key to view electrode properties (CdAt). Cell constant: The lower display shows ’CEL’ and the upper display
2
shows the last configured cell constant.
Press ENT key. Calibration factor: The lower display shows ‘FCt’ and the upper
3
display shows the calibration factor based on the last conductivity calibration.
NOTE: The upper display shows ‘- - - ‘ if the controller has not been calibrated yet or the controller has been reset to factory defaults.
Press ENT key to quit from electrode properties screen. The controller reverts to CdAt screen.
Press or key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
ENT

6 TECHNICAL SPECIFICATIONS

General Specification Conductivity
Measuring Range: Resolution:
0.000 - 2.000 µS/cm 0.001 µS/cm
0.00 - 20.00 µS/cm 0.01 µS/cm
0.0 - 200.0 µS/cm 0.1 µS/cm 0- 2000 µS/cm 1 µS/cm
0.00 - 20.00 mS/cm 0.01 mS/cm
0.0 - 200.0 mS/cm 0.1 mS/cm
Relative accuracy ± 1% of full scale reading System accuracy The effective measuring range and the
Temperature
Measuring range -10.0 to +125.0 oC Resolution 0.1
Relative accuracy ±0.5 oC / ±1.0 oF Sensor 2 wire / 3 wire - Pt100 or Pt1000 (Jumper
Compensation Automatic or Manual (reference at 25oC)
Calibration Conductivity
Number of calibration points 1 point Calibration window ±40% of default reading Calibration standard solution window
Temperature
Offset Adjustment ± 10 ˚C/ ± 18 ˚F
Controller
Output 2 SPST Relays Relay 1/Relay 2 function (for Set Point 1 (SP1) & Set Point(SP2))
Hysteresis window 0% to 10 % of selected measuring range On-delay & Off-delay time 0 to 2000 seconds
Display
LCD 7 segments display with symbols for status Back light On/Off selectable
total system accuracy depend on the sensor used with the controller. Please refer respective sensor documentation.
+14.0 to +257.0
o
C / 0.1 oF
Selectable)
10% to 100% of the selected measuring range
LO (Relay is active when measured value < Set point) HI (Relay is active when measured value > Set point)
information
o
F
Electrical Data and Connections
Conductivity Input Screw Terminal (3.5mm pitch) Connection terminal 8-pin, 2-pin & 5-pin terminal blocks, DC
EMC Specifications
Emission EN 61 36 Susceptibility EN 61326
Other
Power Input Universal +9 V DC Adaptor and 24V DC Dimensions (W x H x D) 96mm x 96mm x 66 mm
Weight (Estimated) 210g Ambient Temp. operating range 0 to 40 oC Maximum Relative Humidity 80% up to 31 oC decreasing linearly to
Calibration Default Settings (CAL dEF) Conductivity
Calibration factor Reset to 1.000
Factory Default Settings (FCt dEF) Conductivity
Calibration factor Reset to 1.000 Cell constant 1.00 Cell type 2-cell Measuring range 0 to 2000 µS/cm Line resistance 0.000
Temperature
Unit of measurement Temperature compensation ATC On
Temperature coefficient 2.1% (Both process & calibration temp. Temperature coefficient type Linear
MTC (ATC Off) 25 oC (Both process & calibration
Controller Output
Controller type Limit control Relay status De-energized
Relay 1 (SP1)
Set point value 100 µS Relay function Low
power connector (1.2mm diameter)
Supply
o
50% at 40
o
C
C
coefficients)
temperature)
Display
Hysteresis 20 µS On-delay time 0 seconds Off-delay time 0 seconds
Relay 2 (SP2)
Set point value 1900 µS Relay function High Hysteresis 20 µS On-delay time 0 seconds Off-delay time 0 seconds
Backlight On

7 LIST OF ACCESSORIES

7.1 Thermo Scientific Order Codes

Conductivity Controller Replacement and Accessories
Item Description Order Code
Alpha CON 560 Controller TSCONCP0560 Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 0.1 ECCONSEN89X Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 1.0 ECCONSEN88X Conductivity 2 Cell type probe, 0.1 - 200μS; Cell constant, K=0.1 with
integrated Pt 100, Material SS316 and 25ft cable (open-ended) Conductivity 2 Cell type probe, up to 200 mS; Cell constant, K=1.0 with
integrated Pt 100, Material SS316 and 25ft cable (open-ended)

7.2 Eutech Instruments Order Code

Conductivity Controller Replacement and Accessories
Item Description Order Code
Alpha COND 560 Controller 19505-40 Eutech Instruments Four-Electrode Conductivity Cell 3/4" NPT(M) 19503-00 100 Ohm Pt RTD Industrial Conductivity Cells Conductivity Cell; K=0.1, SS, In-Line; 100 Ohm Pt RTD ATC 100 Ohm Pt RTD Industrial Conductivity Cells Conductivity Cell; K=1.0, SS, In-Line; 100 Ohm Pt RTD ATC 100 Ohm Pt RTD Industrial Conductivity Cells Conductivity Cell; K=10.0 CPVC, In-Line; 100 Ohm Pt RTD ATC
NOTE: Please contact your authorised distributor or dealer for the prices of extension measuring cables and other accessories such as tee joints, electrode assembly, and calibration solutions.
ECCS1001S
ECCS1010S
19500-44
19500-45
19500-46

8 TROUBLESHOOTING

Problem
Power on, but no display
Unstable conductivity reading
Oscillating temperature readings
Slow response a) Dirty / Oily probe a) Clean probe Blinking ATC a) No temperature probe
Blinking electrode annunciator
Or (Conductivity) a) Conductivity reading is
Or/Ur (Temperature) a) Temperature probe is not
Cause
a) Loose connections b) Incorrect output voltage
of the power adaptor
c) Cables not in correct
polarity (+ and – position)
a) Air bubbles in the probe b) Dirty probe c) Probe not deep enough
in sample
a) Electrical noise
interference
connection during ATC
mode a) Error in calibration b) Conductivity value of the
calibration standard is not
within acceptable range
above the selected
measuring range
connected when ATC
enabled b) Temperature value
exceeds above/ below
the specified range
Solution
a) Ensure cables make good
contact
b) Use an power adaptor with
specified output voltage
c) Re-wire loop cables with
correct polarity
a) Tap probe to remove air
bubbles
b) Clean probe and
recalibrate
c) Ensure that the sample
entirely covers the sensor of the probe
a) Ensure shield wire is
properly connected to pin 7
a) Ensure temperature
sensing cable makes good contact
a) Ensure calibration
standard solution is not contaminated. Ensure electrode is clean
b) Ensure calibration
standard is within 10 to 100% of full-scale of the selected measuring range
a) Ensure correct conductivity
measurement range is selected
a) Ensure electrode makes
good contact with controller
b) Ensure temperature
calibration is done properly. Ensure process temperature is within the specified range

9 GENERAL INFORMATION

9.1 Warranty

This Controller is supplied with a one-year warranty against significant dev iations in material and workmanship from date of purchase and a six-month warranty for probe. Each instrument will have a warranty card with a specific ser ial nu mber. T he warranty card must be endorsed by the Authorized Distributor at the point of sale.
If repair or adjustment is necessary and has not been the result of ab use or misuse within the designated period, please return – freight pre-paid – and c orrection will be made without charge. Thermo Scientific/ Eutech will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of the products

9.2 Return of Goods

Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Goods Authorization” (RGA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipm ent and insured against possible damage or loss. Thermo Scientific will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be ma de on all unauthorized returns.
NOTE: Thermo Scientific reserves the right to make improvements in design, construction, and appearance of products without notice.

9.3 Guidelines for Returning Unit for Repair

Use the original packaging material if possible when shipping the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for additional
protection. Include a brief description of any faults suspected for the c onvenie nce of Customer Service Dept., if possible.

9.4 Maintenance and Cleaning

Maintenance
The Alpha COND 560 Controller contains no user repairable components. Please contact your authorized distributor if there is any problem with the unit.
Cleaning
To remove dust, dirt and spots, the external surfaces of the Conductivity Controller may be wiped with a damp, lint-free cloth. A mild household cleaner can also be used if necessary.

10 APPENDICES

10.1 Appendix 2 Conductivity of various aqueous solutions

At Temperature 25 oC / 77 oF
Solution Conductivity Resistivity
Ultra-pure Water 0.055 µS/cm 18.18 MΩ-cm Power Plant Boiler Water 0.05 - 1 µS/cm 1 - 18 MΩ-cm Distilled Water 0.5 µS/cm 2 MΩ-cm De-ionized Water 0.1 - 10 µS/cm 0.1 - 10 MΩ-cm De-mineralized Water 1 - 80 µS/cm 0.01 - 1 MΩ-cm Mountain Water 10 µS/cm 0.1 MΩ-cm Drinking Water 0.5 - 1 mS/cm 1 - 2 kΩ-cm Waste-water 0.9 - 9 mS/cm 0.1 - 1 kΩ-cm Potable Water Maximum 1.5 mS/cm 0.7 kΩ-cm Brackish Water 1 - 80 mS/cm 0.01 - 1 kΩ-cm Industrial Process Water 7 - 140 mS/cm Rarely stated Ocean Water 53 mS/cm Rarely stated

10.2 Appendix 2 Graphical Representation of the Function of Hysteresis

SP1 = 100 & Set to LO SP2 = 1900 & Set to HI
µS
µS
RELAY ON
µS
µS
Default = 20
HYSTERESIS BAND
100
SP1
120
RELAY OFF
18
The relay of the conductivity controller activates when the set-point is reached (forward direction). In the reverse direction, it does not de-activate when the value reaches the set-point. Instead, it continues to be active till the value reaches the amount set by the Hysteresis band.
Example: You have set your high set point at 1900 µS and your hysteresis value at 20 µS. If your measured value overshoots 1900 µS, the controll er’s relay activates. The actions of the external device will caus e the solutio n’s conductivit y to drop. The relay will deactivate, when the conductivity value drops below 1880 µS.
Default = 20
HYSTERESIS BAND
80 1900
SP2
Fo
rward Direction
R
everse Direction
µS

10.3 Appendix 3 General Instructions Concerning Controller Setting

Control characteristic of controller used as limit value switch
MIN function MAX function
-Xw
100% Yh
50%
0%
SP 2
Xp = 0
+Xw

10.4 Appendix 4 - Abbreviations Used in LCD

Abbreviation Description

AtC Automatic Temperature Compensation CAL Calibration mode C.ºC Calibration temperature CdAt Electrode properties CEL Cell CnFg Configuration Cntr Controller C.tC Temperature coefficient for calibration liquid dEF Default values dEEn De-energized En Energized FCt Factory (defaults) HI High limit HOLD Relay functions are hold HyS Hysteresis K Cell constant L.Ad Line ar resistance L.Ct Limit control Lin Linear LitE Backlight LO Low limit MEAS Measurement mode OF.d Off delay On.d On delay Or Reading is over range P.ºC Process temperature P.tC Temperature coefficient for process liquid PUr Pure water rAng Range

Set Setting SETUP Setup mode SP1 1st Set point SP2 2nd Set point tC Temperature Coefficient tyP Type Ur Reading is under range

Water Analysis Instruments
North America
166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836
www.thermo.com/process
© 2009 Thermo Fisher Scientific Inc.
All rights reserved. Thermo Fisher Scientific Inc.
68X216878 Rev 5
Loading...