ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
U.S. patent 6,793,787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT,
PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink,
ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered
trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTOBAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTOSTIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT,
DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority,
Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy,
KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement,
MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement,
Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture
Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB,
SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator
21, Stat Face, The Enhanced Lab, ThermaSense, Triode, T RIUMpH, Unbreakable
pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
This manual serves to explain the use of the Alpha COND 560 Controller. The
manual functions in two ways, firstly as a step by step guide to help the user
operate the instrument. Secondly, it serves as a handy reference guide. This
instruction manual is written to cover as many anticipated applications of the Alpha
COND 560 Controller. If you have doubts in the use of the instrument, please do not
hesitate to contact your nearest Alpha Authorised Distributor.
The information presented in this manual is subject to change without notice as
improvements are made, and does not represent a commitment on part of Thermo
Scientific.
Thermo Scientific cannot accept any responsibilit y for damage or malfunction of the
unit due to improper use of the instrument.
9.2 Return of Goods.....................................................................................36
9.3 Guidelines for Returning Unit for Repair................................................ 36
9.4 Maintenance and Cleaning .................................................................... 37
10 APPENDICES 38
10.1 Appendix 2 Conductivity of various aqueous solutions.........................38
10.2 Appendix 2 Graphical Representation of the Function of Hysteresis....39
10.3 Appendix 3 General Instructions Concerning Controller Setting...........40
10.4 Appendix 4 - Abbreviations Used in LCD...............................................41
1 INTRODUCTION
1.1 Before You Begin
Thank you for purchasing the Alpha COND 560 Controller.
The construction of the Alpha COND 560 Controller employs leading edge
technology and complies with safety regulations c urrently in force. Notwithstanding
this, improper use could lead to hazards for the user or a third-party, and/or adverse
effects on the plant or other equipment. Therefore, the operating instructions must
be read and understood by the persons involved before working with the
Conductivity Controller.
The instruction manual must always be stored close at ha nd, in a place accessible
to all people working with the Conductivity Controller.
If you have questions, which are not or insufficiently answered in this instruction
manual, please contact your authorized supplier. They will be glad to assist you.
1.2 Intended Use
Alpha COND 560 Controllers are intended solely for Conductivity and temperatur e
measurement, as described in this instruction manual.
Any other use, or use not mentioned here, that is incompatible with the technical
specifications is deemed inappropriate. The oper ator is solely responsible for any
damage arising from such use.
Other prerequisites for appropriate use include:
– Comply with the instructions, notes and requirements set out in this instruction
manual.
– Comply with all local safety regulations concerning safety at work.
– Comply with all information a nd warnings in the documentation de aling with the
products used together with the Conductivity Controller (housing, sensors, etc.).
– Comply with the local environmental and operational conditions.
- 1 -
1.3 Safety Instructions
• The Alpha COND 560 Controller should be installed and
operated only by personnel familiar with the instrument and
who are qualified for such work.
• A defective Conductivity Controller must neither be installed
nor put into service.
• The Alpha COND 560 must only be operated under the
specified operating conditions (see section 6).
• The Alpha COND 560 must not be repaired by the
customer.
• No modifications to the Alpha COND 560 are allowed. The
manufacturer/supplier accepts no responsibility for damage
caused by unauthorized modifications. The risk is borne
entirely by the user.
1.4 Taking Out of Service / Correct Disposal of the Unit
Taking out of Service
•First disconnect the unit from the power supply and then undo all electrical
connections.
•Remove the unit from the wall.
Correct Disposal of the Instrument
When the Alpha COND 560 is permanently taken out of service, obey the local
environmental regulations for correct disposal or send the instrument to your local
distributor, they will take care of proper disposal.
- 2 - - 3 -
2 GETTING STARTED
2.1 Description of Instrument
The Eutech Instruments’ Alpha CON 560 Controller is used for measuring
conductivity and temperature values. The controller can be used for applications
such as water treatment and controlling, galvanic-decontamination, chemical
processing, food processing, clean or wastewater control and neutralization
processes.
The controller has many user-friendly and safety features which include:
• Push-button keypad for calibration and setup
• Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails
• Menu-driven program that simplifies setup
• Automatic temperature compensation (ATC)
• Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
•Large dual display LCD for e asy reading with clear multiple annunciators,
operational mode indicators and error indicators
• Two switching contacts as set-point relays
• Limit (on/off) control mode
• Separately adjustable high and low set-point hysteresis (dead bands) to
prevent chattering of relays around the set points
•0 to 2000 seconds time delay adjustment on all relays – minimizes false
alarms
•Backlight LCD display
2.2 Measurement System
A typical measurement system consists of:
• A conductivity process controller
• A conductivity sensor with integrated or separate Pt100/Pt1000
temperature sensor
• An appropriate measurement cable
• An immersion, flow or process assembly
• A final control element such as pump or valve
Alpha COND 560
Alpha CON 560 Controller
Alpha COND 560
lpha pH560
ESC
CAL
Conductivity Controller
Conductivity Controller
ENT
pH Monitor
Pump or Valve
Relay Output
Housing and
Sensors
Measur ement Ca b le
Power Adaptor
(+9 V DC)
2.3 Connecting Peripherals
2.3.1 Connection Terminals
Remove Back Cover:
Remove the screws from the four corners at the back of the conductivity
controller. Remove the back cover. The connectors are exposed on the
back PCBA as shown in the Figure 1 below.
Connectors:
• J11 – 24V DC power connector
• J8 – 9V DC power connector (1.2mm diameter)
• J9 – Relay connections
• J10 - Conductivity electrode & Temperature probe connections (wiring
has to be done in the detachable connector
J11 Connections
1. +ve
2. Gnd
J10 Connections
1. CON Low Current
2. CON Low Voltage
3. CON High Voltage
4. CON High Current
5. Pt 100 Compensate
6. Pt 100 Sense
7. Pt 100 GND
8. (No connection)
J9 Connections
Relay 1 & 2
Figure 1: Outer Side of Back PCBA
1)+ve
2)Gnd
24V DC
9V DC Power
J8
J9
REL 1
REL 2
J11
Screw
J10
2.3.2 Switching Between PT100 & PT1000 Temperature Sensors
The controller supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature
sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature
sensor, you have to change the jumper setting (J7) as described below.
Remove Back Cover:
Remove screws from the four corners at the back of the conductivity controller.
Remove the back cover.
Remove Back PCBA:
Remove the screw located center of the back PCBA (Figure 1). Detach the back
PCBA from the controller. Turn over the back PCBA. Locate J7 jumper o n the inner
side of the back PCBA as shown in figure 2 below.
Figure 2: Inner Side of Back PCBA
Set Jumper J7:
Set the J7 jumper to required sensor (Pt100 or Pt1000) type
2.3.3 Connecting Conductivity Electrode
Any industrial conductivity probe (2-cell or 4-cell) can be connected to the
conductivity controller.
2-Cell Conductivity Probe:
1. Short Pin 1 & 2 of J10 connector using a jumper wire.
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (black) to Pin 3 of J10 conn ector.
4. Short Pin 3 & 4 of J10 connector using a jumper wire.
4-Cell Conductivity Probe:
5. Connect CON low input wire (black) to Pin 1 of J10 con nector
6. Connect CON low input wire (white) to Pin 2 of J10 connector.
7. Connect CON high input wire (orange) to Pin 3 of J10 connector.
8. Connect CON high input wire (blue) to Pin 4 of J10 con nector.
2.3.4 Connecting Temperature Probe
For Automatic Temperature Compensated (ATC) conductivity readings, an in-built
temperature sensor is usually integrated with industrial conductivity probes.
Alternatively, a separate 100Ω Pt RTD temperature probe (2-wire or 3-wire) can be
connected to the controller.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J10 connector.
2. Connect PT100 sense wire to Pin 6 of J10 connector.
3. Connect PT100 GND wire to Pin 7 of J10 co nnector.
2-Wire Probe:
4. Short Pin 5 & 6 of J10 connector using a jumper wire.
5. Connect PT100 sense wire to Pin 6 of J10 connector.
6. Connect PT100 GND wire to Pin 7 of J10 co nnector.
2.4 Installation
2.4.1 Mechanical Dimensions
Alpha COND 560
CAL
2.4.2 Wall Mount
P
Alpha pH 600
ESC
H/ORP RF Transmitter
Conductivity Monitor
ENT
Pierce through holes at both sides
11
Cover the catch slots at both sides
2
with overlays
f
f
A
r
2.4.3 Panel Mount
Prepare panel cut-out o
1
1
92.0 mm X 92.0 mm
Remove back cover o
2
pH Monitor and slide it
through panel cut-out
Panel
Gasket
Panel
Insert threaded rods through catch
4
until pH Meter is held against panel
ttach catch to both sides
3
of pH Mete
2.5 Display & Keypad
2.5.1 Display Overview
The Liquid Crystal Display (LCD) of Alpha COND 560 conductivity controller has
two alpha-numerical displays (Upper and a Lower).
•Upper display: Measured conductivity value is d ispl aye d when the c ontr oller is
in normal operation (measurement) mode.
•Lower display: Measured temperatur e value is displayed when the controller
is in normal operation (measurement) mode. In conductivity calibration mode,
measured conductivity values are displayed here.
The two displays indicate function names, options & settings in Setup m ode. Refer
‘Appendix 4 - Abbreviations Used in LCD’ for more details.
The LCD also consists of various mode indicators, status annunciators and unit of
measurement indicators.
Conductivity in Milli Siemens
Conductivity in Micro Siemens
Line Resistance in Ohms (Refer Section 5.7)
Temperature Coefficient (Refer Section 5.2)
Temperature in Celsius (Refer Section 5.3) Temperature in Fahrenheit (Refer Section 5.3)
2.5.2 Key Functions
Alpha COND 560
lpha CON560
ESC
Key Description
CAL
Enter Calibration mode
Enter Setup mode
Access sub screens (parameters) within a group of settings in Setup
ENT
mode
Confirm (save) setup parameters and numerical values
Start/Confirm calibration in Calibration mode.
Select a group of settings in Setup mode. Select parameters and increment/decrement numerical values in Setup
and Calibration modes. (When pressed continuously, speed of value increment/decrement
increases) Returns to Measurement mode when both keys are pressed
simultaneously.
CAL
Conductivity Monito
ENT
r
3 OPERATION
3.1 Measurement mode
When the conductivity controller is powered on, the display shows all the LCD
segments briefly, and then automatically enters into the measurement mode.
MEAS
µS
READY
°C
25.0
The mode indicator ‘MEAS’ at the top of the display indicates that the conductivity
controller is in measurement mode. The upper alpha-num erical display shows the
measured conductivity value, while the lower display shows the temperature value.
The controller has six selectable measuring ranges. The unit of measurement and
resolution of the conductivity reading depend on the selec ted measuring range. If
the controller reads a conductivity reading b eyond the selected measuring range,
the display shows ‘Or’ (over range) error indicator.
NOTE:
From measurement mode you can access:
For more details, refer section 4 for Calibration mode and section 5 for Setup
mode.)
To guarantee accurate readings, the measuring system (the
conductivity controller and the sensor) must be calibrated regularly.
• Calibration mode (by pressing CAL key)
• Setup mode (by pressing ENT key)
ATC
Y
A
3.2 Menu Overview
READ
MEAS
ENT
µS
CAL ENT
°C
TC
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
4 CALIBRATION MODE
4.1 About Calibration
Calibration should be carried out each time a new electrode is attached to the
controller or when you suspect that the controller/electrode is out of calibration. T he
controller allows you to perform temperature calibration and conductivity calibration.
About Temperature Calibration
It is important to ensure that temperature calibration is carried out prior to
conductivity calibration since temperature readings affect the accuracy of
conductivity measurements. It is recommended that temperature calibration should
be carried out only if the temperature value displayed on t he controller is different
from that of a calibrated thermometer. A temperature offset calibration of ± 10 °C/±
18 °F from the default reading is allowed. Once a temperature calibration is
performed, conductivity calibration should be carried out to ensure the accuracy of
conductivity measurements.
About Measuring Ranges & Conductivity Calibration
The controller has six selectable conductivity measuring ranges. You can only
calibrate one-point in the measuring range th at you have selected for your proces s
application. You have to recalibrate the controller every time you change the
measuring range since the calibration data of the controlle r will be reset every time
a new measuring range is selected.
About Conductivity Calibration
During conductivity calibration, the controller allows a calibration window of ± 40%
from the measured default reading of the calibration standard. The minimum
allowable calibration point is 10% of the full scale reading of the measur ing range
selected.
For best results, select a standard conductivity solution close to the sa mple value
you are measuring.
4.2 Calibration
The controller requires one-point calibration. Conductivity calibration is always
carried out in the specific measuring range selected for the controller. M ake sure,
the controller is set to desired measuring range before you enter cal ibration mode
(Refer Section 5.6 for selecting measuring ranges).
Select a suitable standard conductivity solution so that its conductivity value is
within 10% to 100% of the full scale of the selected measuring range of the
controller. (For better accuracy, select the calibration standard to be nearly 50% of
the full scale of the measuring range selected.) Prepare calibration standard
solution in two beakers – one for rinsing and the other for calibration. Prepare
separate de-ionized water for electrode rinsing.
All new calibration values will automatically override existing calibration data. If you
wish to abort the calibration, press ▲ and ▼ keys simultaneously and the controller
reverts to measurement mode and the controller will revert to previous calibration
data.
MEAS
READY
11
2
3
OR
2
HOLD
K=
HOLD
K=
1
µS
CAL
HOLD
°C
ATC
HOLD
From measurement mode press CAL key to enter calibration mode.
The display briefly shows the cell type (“2 CELL” or “4 CELL”) to which
the controller is configured (Refer section 5.7 for cell type settings)
2
CAL
CAL
ENT
READY
HOLD
2
READY
HOLD
3
CAL
µS
READY
HOLD
3
CAL
READY
µS
ENT
Next, the last configured cell constant (k) is displayed in the upper
display. Use ▲ or ▼ keys to adjust the cell constant to mach the cell
constant of the conductivity probe. Adjustable range is 0.10 to 10.0.
Press ENT key to confirm the new cell constant value.
Rinse the electrode in de-ionized water and then rinse with the
calibration standard. Immerse the probe in calibration standard
solution intended for calibration and swirl gently to create a
homogenous sample. Make sure there are no any air-bubbles trapped
in the sensor.
The upper display shows the currently measured conductivity value of
the solution based on last calibration (if any). The lower display shows
MEAS
4
CAL
ENT
µS
°C
ATC
uncalibrated (default) reading of the solution. The buffer annunciator
appears in LCD. The ‘READY’ annunciator appears when the reading
is stable.
Once the reading stabilizes press the ▲ or ▼ key to adjust the upper
display value to that of the standard solution. Press the ENT key to
accept the value.
NOTES:
Calibration is completed. The display shows the calibration factor
4
(FCt). This serves as an indication of the effectiveness of the sensor
which will degrade with time and usage. An effective sensor should
have a calibration factor of 0.60 to 1.40
Press ENT key to return to measurement mode.
The acceptable calibration window is ±40% of the
displayed (default) value. For example, if the display is
1000 µS, the values to which it can be adjusted is 600 to
1400 µS.
If the conductivity reading of the solution is not within 10%
to 100% of the full scale of the selected measuring range,
calibration error occurs; LCD displays “ERR” and flash
electrode annunciator.
If the conductivity reading of the solution is beyond the
selected measuring range, calibration error occurs; LCD
displays “Or” (over range), “ERR” and flashes electrode
annunciator.
If calibration error occurs, exit form calibration mode by
pressing ▲ and ▼ keys simultaneously.
NOTES:
• If you use a 2-cell conductivity sensor, make sure that the pin 1 and 2 are
shorted and pin 3 and 4 are shorted on the J10 connector. (Refer section 2 for
more details)
• When calibrating with manual temperature compensation, the controller
automatically changes from the preset process temperature to the calibratio n
temperature. After leaving the Calibration mode, the controller switches back to
the process temperature (for setting the calibration temperature and the
process temperature, see section 5.3)
WARNING: If you change the measuring range of the controll er after calibration is
done, the controller will reset back to its factory default calibration.
5 SETUP MODE
5.1 Enter Setup mode
The setup mode allows you to customize the settings of the conductivity controller
to suite your requirements. While in measurement mode, press the ENT key to
access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup
(rAng – Range settings). Press ▲ or ▼ key to access other pages of the setup
mode.
READY
MEAS
µS
°C
ATC
ENT
SETUP
HOLD
ENT
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
NOTES:
• To exit setup mode at any time press ▲ and ▼ keys
simultaneously (escape). The controller returns to the
measurement mode
• If you do not press any key for 2.5 minutes, the controller
automatically exists from the setup mode.
5.2 Temperature Coefficient Settings
Setting the temperature coefficient to match your process/calibration li quids allows
controller to achieve conductivity readings with higher accuracy.
SETUP
HOLD
1
SETUP
ENT
HOLD
SETUP
HOLD
2
SETUP
ENT
HOLD
2
ENT
3
SETUP
%%
ENT
HOLD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode
4
ENT
(rAng). Press ▲ or ▼ key to select temperature coefficient settings
screen (tC).
Press ENT key to access temperature coefficient settings (tC).
Select type of temperature coefficient: The upper display shows the
2
last configured temperature coefficient type (Lin or PUr). Press ▲ or
▼ key to select the desired type based on type of your process liquid.
PUr : Pure water temperature coefficient: Select this if your
process liquid is pure water or ultra pure water.
Lin : Linear temperature coefficient: Select this for all other
process liquids
Press ENT key to confirm your selection.
If PUr was selected: The controller reverts back to tC screen
If Lin was selected:
Select temperature coefficient for process liquid: The lower display
3
shows ‘P.tC’. The upper display shows last configured temperature
coefficient for process liquid (if any). The default value is 2.1%. Press
▲ or ▼ key to select the desired value. Allowable range: 0 to 10%.
Press ENT key to confirm your value.
Select temperature coefficient for calibration liquid: The lower
4
display shows ‘C.tC’. The upper display shows last configured
temperature coefficient for calibration liquid (if any). The default value is
2.1%. Press ▲ or ▼ key to select the desired value. Allowable range:
0 to 10%. Press ENT key to confirm your value. The controller reverts
to tC screen.
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
A
5.3 Temperature Settings
Automatic Temperature Compensation (ATC):
The conductivity value of a sample is affected by temperature. Use ATC feature of
the controller to compensate for conductivity changes when the temperat ure of the
sample or process liquid fluctuates. Connect a separate temperature probe
(Pt100/pt1000) to the controller if an integrated conductivity/Temperature probe is
not available.
Manual Temperature Compensation (MTC):
Set the controller to MTC (disable ATC) when the temperature of sample or process
liquid is constant and a temperature probe is not available. If you disable AT C, the
controller allows you to set your process temperature (P.ºC) and calibration
temperature (C.ºC). This allows calibration at a different temperature other than the
process temperature. Example: Setting calibration temperature of 25°C lets you
calibrate using standard solutions at 25°C, even if your process temperature is
different from 25°C.
SETUP
HOLD
1
SETUP
ENT
HOLD
2
SETUP
ENT
HOLD
°C
3
SETUP
ENT
HOLD
°C
4
SETUP
HOLD
°C
TC
5
ENT
°C
ATC
SETUP
HOLD
2
SETUP
HOLD
ENT
°F
3
SETUP
ENT
HOLD
°C
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode
67
SETUP
ENT
HOLD
ENT
(rAng). Press ▲ or ▼ key to select Temperature settings screen (SET
ºCF).
Press ENT key to access temperature settings (SET ºCF).
Selecting unit of measurement for temperature: The upper display
2
shows ‘Unit’ and the lower display shows the last configured unit of
measurement for temperature.
Press ▲ or ▼ key to select the desired units for temperature (ºC or
ºF). Press ENT key to confirm your selection.
Enable/disable ATC: The lower display shows ‘AtC’ and the upper
3
display shows the last configured ATC selection (‘On’ or ‘OFF’).
Press ▲ or ▼ key to enable (ATC On) or disable (ATC OFF)
automatic temperature compensation. Press ENT key to confirm your
selection.
If ATC enabled (ATC On):
Selecting temperature offset: The upper display shows the last
4
configured temperature offset value (if any), otherwise the default is
zero. The lower display shows currently measured temperature reading
(including last configured offset value). LCD shows ‘ATC’ annunciator
in lower-right corner.
Place a thermometer, which is known to be accurate, in your sample or
process liquid. Make sure your temperature probe is placed in the same
liquid. Compare the stabilized temperature read ing displayed on the
controller with the thermometer. If there is a difference between the two
readings (offset), you can adjust the reading of the controller. Press ▲ or ▼ key to adjust the lower display to the correct temperature value.
As the lower display value changes, the controller adjust the upper
5
5
display reading automatically to suit the new offset value. Up to ± 10 ºC/
± 18 ºF offset is allowed.
Press ENT key to confirm the value. The controller reverts to SET ºCF
screen.
If ATC disabled (ATC OFF):
Setting process temperature: The lower display shows ‘P.ºC’ and the
6
upper display shows the last configured process temperature. Press ▲
or ▼ key to adjust the upper display to desired process temperature.
Allowable range: –10.0 to 125.0°C / 14.0 to 257.0°F.
Press ENT key to confirm the process temperature.
Setting calibration temperature: The lower display shows ‘C.ºC’ and
7
the upper display shows the last configured calibration temperature.
Press ▲ or ▼ key to adjust the upper display to desired calibration
temperature. Allowable range: –10.0 to 125.0°C / 14.0 to 257.0°F.
Press ENT key to confirm the calibration temperature. The controller
reverts to SET ºCF screen.
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press ▲ and ▼ keys simultaneously (escape).
The controller returns to the first screen of temperature settings SET ºCF.
5.4 Relay 1/Relay 2 Settings
NOTE:
The SP1 settings determine the operating parameters for Relay 1; while SP2
settings define the operating parameters for Relay 2. Since these two settings have
identical setup parameters, only SP1 is described here.
Note that when you change measuring range of the controller, the values you have
set for SP1 & SP2 will not be affected. However, the unit of measurement &
resolution displayed in LCD for SP1 & SP2 will change according to the newly
selected measuring range.
If controller type is set to ‘OFF’ (Refer section 5.5 for controller
settings), the ‘Relay 1/Relay 2 Setting’ is not available.
SETUP
HOLD
1
SETUP
ENT
HOLD
2
SETUP
µS
ENT
HOLD
SETUP
HOLD
3
SETUP
µS
ENT
HOLD
3
µS
ENT
4
SETUP
µS
ENT
HOLD
SETUP
HOLD
5
ENT
6
ENT
From measurement mode press ENT key to enter setup mode as
1
1
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press ▲ or ▼ key to select relay 1 (Set point 1) settings
screen (SP1).
Press ENT key to access relay 1 settings (SP1).
Setting set point value: The upper display shows the last configured
2
set point value, otherwise the default value 100 µS. Press ▲ or ▼ key
to enter the value for set point 1, at which control function of the
conductivity controller should activate. Press ENT key to confirm your
set point value.
Select relay function: The upper display shows the last configured
3
relay function (LO = Low or HI=High). Press ▲ or ▼ key to select the
desired function.
Press ENT key to confirm your selection.
NOTE:
Setting hysteresis value: The lower display shows ‘HYS’. The upper
4
display shows the last configured value, otherwise the default value 20
µS. Press ▲ or ▼ key to enter the hysteresis value for set point 1.
Allowable range: 0 to 10% of full-scale of the selected measuring range.
Press ENT key to confirm your hysteresis value.
NOTE:
Setting on-delay time: The lower display shows ‘On.d’. The upper
5
5
display shows last configured on-delay value. Press ▲ or ▼ key to
enter the on-delay time for set point 1. Allowable range: 0 to 2000
seconds. When the conductivity reading is reached the set point 1, the
controller delays the relay activation (switch on) by the on-delay time
you set here.
Press ENT key to confirm your on-delay time.
Setting off-delay time: The lower display shows ‘OF.d’. The upper
6
display shows last configured off-delay value. Press ▲ or ▼ key to
enter the off-delay time for set point 1. Allowable range: 0 to 2000
seconds. When the conductivity reading is reached the set point 1 plus
hysteresis value, the controller delays the relay deactivation (switch off)
by the off-delay time you set here.
Press ENT key to confirm your off-delay time. The controller reverts to
SP1 screen.
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
This parameter lets you choose the relay function. Select
‘LO’ to activate the relay when the conductivity reading
undershoots the set point (SP1) or select ‘HI’ to activate the
relay when the conductivity reading overshoots the set point
(SP1). SP1 and SP2 can be selected as ‘LO/LO’, ‘LO/HI’,
‘HI/LO’, or ‘HI/HI’.
Hysteresis prevents rapid relay contact switching if the
conductivity reading is fluctuating near the set point. Refer
‘Appendix 2
Graphical Representation of the Function of Hysteresis’.
5.5 Controller Settings
The controller settings allow you to set the relay type and relay status of the
controller.
SETUP
HOLD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup
1
SETUP
ENT
HOLD
SETUP
HOLD
2
ENT
2
SETUP
ENT
HOLD
SETUP
HOLD
mode (rAng). Press ▲ or ▼ key to select controller settings screen
(Cntr).
Press ENT key to access controller settings (Cntr).
Select the controller type: The lower display shows ‘tyP’. The upper
2
display shows the last configured controller type. (‘OFF’ or ‘L.Ct’).
Press ▲ or ▼ key to select the desired controller type.
OFF : Controller Off
Selecting ‘OFF’, the conductivity controller stops
functioning as a controller by preventing relay functions.
Once the controller type is set to ‘OFF’, the setup mode
does not show you the Relay 1/Relay 2 settings.
3
ENT
3
ENT
L.Ct: Limit value control (Controller On)
Selecting ‘L.Ct’, the conductivity controller is set to function
as a controller. When the controller type is set to ‘L.Ct’, the
Relay 1/Relay 2 settings are available in the setup mode
for you to configure relay operating parameters.
Press ENT key to confirm your selection.
If Controller type was set to Off (OFF):
The controller reverts to Cntr screen.
If Controller type was set to On (L.Ct):
Select the relay status under Non-Alarm condition: The lower
3
display shows ‘rEL’. The upper display shows the last configured relay
status. (‘dEEn’ or ‘En’). Press ▲ or ▼ key to select the desired relay
status.
dEEn:
En:
Press ENT key to confirm your selection. The controller reverts to
Cntr screen.
De-energized
Energized
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
5.6 Measuring Range Settings
The conductivity controller has 6 different measuring ranges. You need to select a
range which is the most suitable for conductivity reading of your
applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm).
The measuring range you select here will affect the way the control ler behaves in
the measurement mode and calibration mode. In measurement mode, if the
conductivity reading of the controller is above the selected measuring range, the
LCD shows “Or” (over range) error. In calibration mode, the controller adjusts
calibration tolerance based on full scale of the selected measuring range. The
calibration is carried out only for the selected measuring range.
WARNING: If you change the measuring range of the controll er after calibration is
done, the controller will reset back to its factory default calibration.
From measurement mode press ENT
11
key to enter setup mode as described in
section 5.1. The LCD shows the first
screen of setup mode (rAng). Press ENT key to access measurement range
setting (rAng).
The upper display shows the currently
2
selected measuring range. The factory
default is 2000 µS. The lower display
shows the currently configured cell
constant (k). (Refer section 4.2 for
setting cell constant)
Press ▲ or ▼ key to select the desired
measuring range. Press ENT key to
confirm your selection.
The controller reverts to rAng screen.
Measuring range Resolution
0.000 to 2.000 µS/cm 0.001 µS/cm
0.00 to 20.00 µS/cm 0.01 µS/cm
0.0 to 200.0 µS/cm 0.1 µS/cm
0 to 2000 µS/cm 1 µS/cm
0.00 to 20.00 mS/cm 0.01 mS/cm
0.0 to 200.0 mS/cm 0.1 mS/cm
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
SETUP
HOLD
1
SETUP
ENT
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
2
µS
ENT
2
µS
ENT
2
µS
ENT
2
µS
ENT
2
mS
ENT
2
mS
ENT
5.7 Configuration Settings
Configuration settings let you to enable/disable backlight of the LCD, select
between 2-cell & 4-cell and reset the controller to factory defaults.
SETUP
HOLD
1
From measurement mode press ENT key to enter setup mode as
1
1
described in section 5.1. The LCD shows the first screen of setup mode
SETUP
ENT
HOLD
SETUP
HOLD
2
2
SETUP
ENT
HOLD
SETUP
HOLD
ENT
SETUP
ENT
HOLD
SETUP
HOLD
SETUP
HOLD
3
3
5
5
5
SETUP
ENT
HOLD
This screen appears only
when 2-cell is selected
and measuring range is set to
0.0 to 200.0 mS/cm
ENT
READY
ENT
ENT
MEAS
(rAng). Press ▲ or ▼ key to select configuration settings screen
(Cnfg).
Press ENT key to access configuration settings (Cnfg).
Enable/disable LCD backlight: The upper display shows ‘LitE’. The
2
lower display shows last configured backlight selection (‘On’ or ‘OFF’).
Press ▲ or ▼ key to enable (LitE On) or disable (LitE OFF) backlight
4
R.
ENT
µS
°C
ATC
of the LCD.
Press ENT key to confirm your selection.
Select cell type of the sensor: The lower display shows ‘CEL’. The
3
upper display shows the last configured cell type (‘2’ or ‘4’).
Press ▲ or ▼ key to select the cell type of your conductivity probe.
Press ENT key to confirm your selection.
NOTE: If you use a 2-cell conductivity sensor, make sure that the pin 1
and 2 are shorted and pin 3 and 4 are shorted on the J10 connector.
(Refer section 2 for more details)
Setting line resistance: The lower display shows ‘L.Ad’ and the upper
4
display shows the last configured line resistance; otherwise the default
is 0.00 R.
NOTE: Line resistance of sensor cable constitutes to an error in
measurement of high conductivity values. The controller will
compensate this error if you specify line resistance value in this setting.
The setting is available only if the cell type is set to 2-cell (step 3 above)
and the measuring range is selected to ‘0.0 to 200.0 mS/cm’.
Press ▲ or ▼ key to set the line resistance of the sensor cable
connected to the controller.
Press ENT key to confirm your selection.
Reset to factory defaults: The display shows ‘nO dEF’.
5
5
Press ▲ or ▼ key to select:
nO dEF :
FCt dEF:
CAL dEF:
NOTE: Refer section 6 for factory/calibration default settings
Press ENT key to confirm your selection.
- If ‘nO dEF’ was selected, the controller reverts to Cnfg screen.
- If ‘FCt dEF’ or ‘CAL dEF’ was selected, the controller performs the
selected reset and returns to measurement mode.
Not default reset of the controller. User configured
values remain active, when confirmed by pressing ENT
key.
Reset all settings of the controller to factory defaults,
when confirmed by pressing ENT key.
Reset all calibration settings of the controller to factory
defaults, when confirmed by pressing ENT key.
Press ▲ or ▼ key to access other setup screens or press ▲
and ▼ key simultaneously (escape) to return to
measurement mode.
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press ▲ and ▼ keys simultaneously
(escape). The controller returns to the first screen configuration settings (Cnfg).
5.8 Viewing Electrode Properties
Each time you calibrate the conductivity sensor, the controller re-calculates the
calibration factor of the sensor and shown in the LCD at the end of calibration. The
setup mode allows you to view the last saved calibration factor and cell constant at
any time.
SETUP
HOLD
1
ENT
SETUP
HOLD
2
ENT
SET
HOLUPD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press ▲ or ▼ key to select viewing electrode properties
screen (CdAt).
Press ENT key to view electrode properties (CdAt).
Cell constant: The lower display shows ’CEL’ and the upper display
2
shows the last configured cell constant.
Press ENT key.
Calibration factor: The lower display shows ‘FCt’ and the upper
3
display shows the calibration factor based on the last conductivity
calibration.
NOTE: The upper display shows ‘- - - ‘ if the controller has not been
calibrated yet or the controller has been reset to factory defaults.
Press ENT key to quit from electrode properties screen. The controller
reverts to CdAt screen.
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key
simultaneously (escape) to return to measurement mode.
ENT
6 TECHNICAL SPECIFICATIONS
General Specification
Conductivity
Measuring Range: Resolution:
0.000 - 2.000 µS/cm 0.001 µS/cm
0.00 - 20.00 µS/cm 0.01 µS/cm
0.0 - 200.0 µS/cm 0.1 µS/cm
0- 2000 µS/cm 1 µS/cm
0.00 - 20.00 mS/cm 0.01 mS/cm
0.0 - 200.0 mS/cm 0.1 mS/cm
Relative accuracy ± 1% of full scale reading
System accuracy The effective measuring range and the
Temperature
Measuring range -10.0 to +125.0 oC
Resolution 0.1
Relative accuracy ±0.5 oC / ±1.0 oF
Sensor 2 wire / 3 wire - Pt100 or Pt1000 (Jumper
Compensation Automatic or Manual (reference at 25oC)
Calibration
Conductivity
Number of calibration points 1 point
Calibration window ±40% of default reading
Calibration standard solution
window
Temperature
Offset Adjustment ± 10 ˚C/ ± 18 ˚F
Controller
Output 2 SPST Relays
Relay 1/Relay 2 function (for
Set Point 1 (SP1) & Set
Point(SP2))
Hysteresis window 0% to 10 % of selected measuring range
On-delay & Off-delay time 0 to 2000 seconds
Display
LCD 7 segments display with symbols for status
Back light On/Off selectable
total system accuracy depend on the
sensor used with the controller. Please
refer respective sensor documentation.
+14.0 to +257.0
o
C / 0.1 oF
Selectable)
10% to 100% of the selected measuring
range
LO (Relay is active when measured value
< Set point)
HI (Relay is active when measured value >
Set point)
NOTE: Please contact your authorised distributor or dealer for the prices of
extension measuring cables and other accessories such as tee joints, electrode
assembly, and calibration solutions.
ECCS1001S
ECCS1010S
19500-44
19500-45
19500-46
8 TROUBLESHOOTING
Problem
Power on, but no
display
Unstable conductivity
reading
Oscillating
temperature readings
Slow response a) Dirty / Oily probe a) Clean probe
Blinking ATC a) No temperature probe
Blinking electrode
annunciator
Or (Conductivity) a) Conductivity reading is
Or/Ur (Temperature) a)Temperature probe is not
Cause
a) Loose connections
b) Incorrect output voltage
of the power adaptor
c) Cables not in correct
polarity (+ and – position)
a) Air bubbles in the probe
b) Dirty probe
c) Probe not deep enough
in sample
a) Electrical noise
interference
connection during ATC
mode
a) Error in calibration
b) Conductivity value of the
calibration standard is not
within acceptable range
above the selected
measuring range
connected when ATC
enabled b) Temperature value
exceeds above/ below
the specified range
Solution
a) Ensure cables make good
contact
b) Use an power adaptor with
specified output voltage
c) Re-wire loop cables with
correct polarity
a) Tap probe to remove air
bubbles
b) Clean probe and
recalibrate
c) Ensure that the sample
entirely covers the sensor
of the probe
a) Ensure shield wire is
properly connected to pin 7
a) Ensure temperature
sensing cable makes good
contact
a) Ensure calibration
standard solution is not
contaminated. Ensure
electrode is clean
b) Ensure calibration
standard is within 10 to
100% of full-scale of the
selected measuring range
a) Ensure correct conductivity
measurement range is
selected
a)Ensure electrode makes
good contact with controller
b) Ensure temperature
calibration is done properly. Ensure process temperature is within the specified range
9 GENERAL INFORMATION
9.1 Warranty
This Controller is supplied with a one-year warranty against significant dev iations in
material and workmanship from date of purchase and a six-month warranty for
probe. Each instrument will have a warranty card with a specific ser ial nu mber. T he
warranty card must be endorsed by the Authorized Distributor at the point of sale.
If repair or adjustment is necessary and has not been the result of ab use or misuse
within the designated period, please return – freight pre-paid – and c orrection will
be made without charge. Thermo Scientific/ Eutech will determine if the product
problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
• Improper or inadequate maintenance by customer
• Unauthorized modification or misuse
• Operation outside of the environment specifications of the products
9.2 Return of Goods
Authorization must be obtained from our Customer Service Department or
authorized distributor before returning items for any reason. A “Return Goods
Authorization” (RGA) form is available through our authorized distributor. Please
include data regarding the reason the items are to be returned. For your protection,
items must be carefully packed to prevent damage in shipm ent and insured against
possible damage or loss. Thermo Scientific will not be responsible for damage
resulting from careless or insufficient packing. A restocking charge will be ma de on
all unauthorized returns.
NOTE: Thermo Scientific reserves the right to make improvements in design,
construction, and appearance of products without notice.
9.3 Guidelines for Returning Unit for Repair
Use the original packaging material if possible when shipping the unit for repair.
Otherwise wrap it with bubble pack and use a corrugated box for additional
protection. Include a brief description of any faults suspected for the c onvenie nce of
Customer Service Dept., if possible.
9.4 Maintenance and Cleaning
Maintenance
The Alpha COND 560 Controller contains no user repairable components. Please
contact your authorized distributor if there is any problem with the unit.
Cleaning
To remove dust, dirt and spots, the external surfaces of the Conductivity Controller
may be wiped with a damp, lint-free cloth. A mild household cleaner can also be
used if necessary.
10 APPENDICES
10.1 Appendix 2
Conductivity of various aqueous solutions
At Temperature 25 oC / 77 oF
Solution Conductivity Resistivity
Ultra-pure Water 0.055 µS/cm 18.18 MΩ-cm
Power Plant Boiler Water 0.05 - 1 µS/cm 1 - 18 MΩ-cm
Distilled Water 0.5 µS/cm 2 MΩ-cm
De-ionized Water 0.1 - 10 µS/cm 0.1 - 10 MΩ-cm
De-mineralized Water 1 - 80 µS/cm 0.01 - 1 MΩ-cm
Mountain Water 10 µS/cm 0.1 MΩ-cm
Drinking Water 0.5 - 1 mS/cm 1 - 2 kΩ-cm
Waste-water 0.9 - 9 mS/cm 0.1 - 1 kΩ-cm
Potable Water Maximum 1.5 mS/cm 0.7 kΩ-cm
Brackish Water 1 - 80 mS/cm 0.01 - 1 kΩ-cm
Industrial Process Water 7 - 140 mS/cm Rarely stated
Ocean Water 53 mS/cm Rarely stated
10.2 Appendix 2
Graphical Representation of the Function of Hysteresis
SP1 = 100 & Set to LO
SP2 = 1900 & Set to HI
µS
µS
RELAY ON
µS
µS
Default = 20
HYSTERESIS BAND
100
SP1
120
RELAY OFF
18
The relay of the conductivity controller activates when the set-point is reached
(forward direction). In the reverse direction, it does not de-activate when the value
reaches the set-point. Instead, it continues to be active till the value reaches the
amount set by the Hysteresis band.
Example: You have set your high set point at 1900 µS and your hysteresis value at
20 µS. If your measured value overshoots 1900 µS, the controll er’s relay activates.
The actions of the external device will caus e the solutio n’s conductivit y to drop. The
relay will deactivate, when the conductivity value drops below 1880 µS.
Default = 20
HYSTERESIS BAND
801900
SP2
Fo
rward Direction
R
everse Direction
µS
10.3 Appendix 3
General Instructions Concerning Controller Setting
Control characteristic of controller used as limit value switch
MIN functionMAX function
-Xw
100%Yh
50%
0%
SP 2
Xp = 0
+Xw
10.4 Appendix 4 - Abbreviations Used in LCD
Abbreviation Description
AtC Automatic Temperature Compensation
CAL Calibration mode
C.ºC Calibration temperature
CdAt Electrode properties
CEL Cell
CnFg Configuration
Cntr Controller
C.tC Temperature coefficient for calibration liquid
dEF Default values
dEEn De-energized
En Energized
FCt Factory (defaults)
HI High limit
HOLD Relay functions are hold
HyS Hysteresis
K Cell constant
L.Ad Line ar resistance
L.Ct Limit control
Lin Linear
LitE Backlight
LO Low limit
MEAS Measurement mode
OF.d Off delay
On.d On delay
Or Reading is over range
P.ºC Process temperature
P.tC Temperature coefficient for process liquid
PUr Pure water
rAng Range
Set Setting
SETUP Setup mode
SP1 1st Set point
SP2 2nd Set point
tC Temperature Coefficient
tyP Type
UrReading is under range
Water Analysis Instruments
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
Dom. Fax: 1-978-232-6015
Int’l Fax: 978-232-6031