Eutech instruments Alpha CON 550 User Manual

Alpha CON 550
Conductivity Monitor mS/µS/ºC/ºF meter
Preface
guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover many anticipated applications of the Alpha-CON 550 monitor. If you have doubts in the use of the instrument, please do not hesitate to contact the nearest Alpha Authorised Distributor.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Thermo Scientific.
Thermo Scientific cannot accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.
Copyright ©2006 All rights reserved.
TABLE OF CONTENTS
1. Introduction ...........................................................................................................1
1.1 Before You Begin................................................................................................................... 1
1.2 Intended Use.......................................................................................................................... 1
1.3 Safety Instructions.................................................................................................................. 2
1.4 Taking Out of Service / Correct Disposal of the Unit ............................................................. 2
2. Getting Started ......................................................................................................3
2.1 Description of Instrument....................................................................................................... 3
2.2 Measurement System ............................................................................................................ 4
2.3 Connecting Peripherals.......................................................................................................... 5
2.3.1 Connection Terminals.................................................................................................... 5
2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors.......................................... 6
2.3.3 Connecting Conductivity Probe..................................................................................... 7
2.3.4 Connecting Temperature Probe.................................................................................... 7
2.4 Installation.............................................................................................................................. 8
2.4.1 Mechanical Dimensions................................................................................................. 8
2.4.2 Wall Mount..................................................................................................................... 8
2.4.3 Panel Mount................................................................................................................... 9
2.5 Display & Keypad................................................................................................................. 10
2.5.1 Display Overview......................................................................................................... 10
2.5.2 Key Functions.............................................................................................................. 11
3. Operation .............................................................................................................12
3.1 Measurement mode ............................................................................................................. 12
3.2 Menu Overview .................................................................................................................... 13
4. Calibration Mode.................................................................................................14
4.1 About Calibration.................................................................................................................. 14
4.2 Calibration............................................................................................................................ 15
5. Setup Mode..........................................................................................................17
5.1 Enter Setup mode ................................................................................................................ 17
5.2 Temperature Coefficient Settings......................................................................................... 18
5.3 Temperature Settings........................................................................................................... 19
5.4 Measuring Range Settings................................................................................................... 21
5.5 Configuration Settings.......................................................................................................... 22
5.6 Viewing Electrode Properties............................................................................................... 24
6. Technical Specifications.....................................................................................25
7. List of Accessories .............................................................................................27
8. Troubleshooting..................................................................................................28
9. General Information ............................................................................................29
9.1 Warranty............................................................................................................................... 29
9.2 Return of Goods................................................................................................................... 29
9.3 Guidelines for Returning Unit for Repair.............................................................................. 29
10. Appendices.......................................................................................................30
10.1 Appendix 2 Conductivity of various aqueous solutions....................................................... 30
10.2 Appendix 4 - Abbreviations Used in LCD............................................................................. 31
Instruction Manual Alpha-CON 550
1. Introduction
1.1 Before You Begin
We thank you for purchasing the Alpha CON 550 monitor. The construction of the CON 550 monitor employs leading edge technology
and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment. Therefore, the operating instructions must be read and understood by the persons involved before working with the monitor.
The instruction manual must always be stored close at hand, in a place accessible to all people working with the conductivity monitor.
If you have questions, which are not or insufficiently answered in this instruction manual, please contact your authorized supplier. They will be glad to assist you.
1.2 Intended Use
Alpha CON 550 monitors are intended solely for conductivity and
temperature measurement, as described in this instruction manual. Any other use, or use not mentioned here, that is incompatible with the
technical specifications is deemed inappropriate. The operator is solely responsible for any damage arising from such use.
Other prerequisites for appropriate use include: – Comply with the instructions, notes and requirements set out in this
instruction manual.
Comply with all local safety regulations concerning safety at work. Comply with all information and warnings in the documentation dealing
with the products used together with the conductivity monitor (housings, sensors, etc.).
– Comply with local environmental and operational conditions.
1
Instruction Manual Alpha-CON 550
1.3 Safety Instructions
- The Alpha CON 550 monitor should be installed and operated only by personnel familiar with the instrument and who are qualified for such work.
- A defective CON monitor must neither be installed nor put into service.
- The Alpha CON 550 must only be operated under the specified operating conditions (see section 6).
- The Alpha CON 550 must not be repaired by the customer.
- No modifications to the Alpha CON 550 are allowed. The manufacturer/supplier accepts no responsibility for damage caused by unauthorized modifications. The risk is borne entirely by the user.
1.4 Taking Out of Service / Correct Disposal of the Unit
Taking out of Service
First disconnect the unit from the power supply and then undo all electrical connections.
Remove the unit from the wall / panel.
Correct Disposal of the Instrument
When the Alpha CON 550 monitor is permanently taken out of service, obey the local environmental regulations for correct disposal or send the instrument to your local distributor, they will take care of proper disposal.
2
Instruction Manual Alpha-CON 550
2. Getting Started
2.1 Description of Instrument
The Alpha CON 550 monitor is used for measuring conductivity and temperature values. The monitor can be used for applications such as water treatment and controlling, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
The monitor has many user-friendly and safety features which include:
Push-button keypad for calibration and setup
Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails
Menu-driven program that simplifies setup
Automatic temperature compensation (ATC)
Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
Large dual display LCD for easy reading with clear multiple annunciators,
operational mode indicators and error indicators
3
Instruction Manual Alpha-CON 550
A
2.2 Measurement System
A typical measurement system consists of:
A conductivity process monitor
A conductivity sensor with integrated or separate Pt100/Pt1000
temperature sensor
An appropriate measurement cable
An immersion, flow or process assembly
lpha CON 550 Monito r
lpha pH560
ESC
CAL
pH Monitor
ENT
Power Adap tor
Housing and Sensors
Meas u r ement C able
(+9 V DC)
4
Instruction Manual Alpha-CON 550
2.3 Connecting Peripherals
2.3.1 Connection Terminals
Remove Back Cover:
Remove the screws from the four corners at the back of the conductivity monitor. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below.
Connectors:
J11 – 24V DC power connector
J8 – 9V DC power connector (1.2mm diameter)
J9 – Relay connections
J10 - Conductivity electrode & Temperature probe connections (wiring has
to be done in the detachable connector
J11 Connections
1. +ve
2. Gnd
J10 Connections
1. CON Low Current
2. CON Low Voltage
3. CON High Voltage
4. CON High Current
5. Pt 100 Compensate
6. Pt 100 Sense
7. Pt 100 GND
8. (No connection)
Figure 1: Outer Side of Back PCBA
24V DC
1)+ve
2)Gnd
J10
J11
J8
9V DC Power
Screw
5
Instruction Manual Alpha-CON 550
2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors
The conductivity monitor supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default
factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below.
Remove Back Cover: Remove screws from the four corners at the back of the conductivity monitor.
Remove the back cover.
Remove Back PCBA:
Remove the screw located center of the back PCBA ( Figure 1). Detach the back PCBA from the monitor. Turn over the back PCBA. Locate J7 jumper on the inner side of the back PCBA as shown in Figure 2 below.
Figure 2: Inner Side of Back PCBA
Set Jumper J7:
Set the J7 jumper to required sensor (Pt100 or Pt1000) type.
6
Instruction Manual Alpha-CON 550
2.3.3 Connecting Conductivity Probe
Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity monitor.
2-Cell Conductivity
Probe:
1. Short Pin 1 & 2 of J10 connector using a jumper wire.
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (black) to Pin 3 of J10 connector.
4. Short Pin 3 & 4 of J10 connector using a jumper wire.
4-Cell Conductivity Probe:
1. Connect CON low input wire (black) to Pin 1 of J10 connector
2. Connect CON low input wire (white) to Pin 2 of J10 connector.
3. Connect CON high input wire (orange) to Pin 3 of J10 connector.
4. Connect CON high input wire (blue) to Pin 4 of J10 connector.
2.3.4 Connecting Temperature Probe
For Automatic Temperature Compensated (ATC) conductivity readings, an in-built temperature sensor is usually integrated with industrial conductivity probes. Alternatively, a separate 100Ω Pt RTD temperature probe (2-wire or 3-wire) can be connected to the monitor.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J10 connector.
2. Connect PT100 sense wire to Pin 6 of J10 connector.
3. Connect PT100 GND wire to Pin 7 of J10 connector.
2-Wire Probe:
1. Short Pin 5 & 6 of J10 connector using a jumper wire.
2. Connect PT100 sense wire to Pin 6 of J10 connector
3. Connect PT100 GND wire to Pin 7 of J10 connector.
7
Instruction Manual Alpha-CON 550
2.4 Installation
2.4.1 Mechanical Dimensions
lpha CON550
ESC
CAL
2.4.2 Wall Mount
ENT
Conductivity Monitor
Cover the catch slots at
2
both sides with overlays
Pierce through holes at
11
both sides
8
Instruction Manual Alpha-CON 550
2.4.3 Panel Mount
Prepare panel cut-out of
1
1
92.0 mm X 92.0 mm
Remove back cover of
2
conductivity monitor and slide it through panel cut-out
Gasket
Panel
Panel
Insert threaded rods through catch
4
until conductivity monitor is held against panel
Attach catch to both sides
3
of conductivity monitor
9
Instruction Manual Alpha-CON 550
2.5 Display & Keypad
2.5.1 Display Overview
The Liquid Crystal Display (LCD) of αlpha CON 550 conductivity monitor has two alpha-numerical displays (Upper and a Lower).
Upper display: Measured conductivity value is displayed when the
monitor is in normal operation (measurement) mode.
Lower display: Measured temperature value is displayed when the
monitor is in normal operation (measurement) mode. In conductivity calibration mode, measured conductivity values are displayed here.
The two displays indicate function names, options & settings in Setup mode. Refer ‘Appendix 4 - Abbreviations Used in LCD’ for more details.
The LCD also consists of various mode indicators, status annunciators and unit of measurement indicators.
Upper Display
Mode Indicators MEAS SETUP CAL Status Annunciators READY HOLD
ATC
Measurement mode (Refer Section 3.1 ) Setup mode (Refer Section 5) Calibration mode (Refer Section 4)
Appears when the reading is stable Appears in Setup mode and Calibration mode to indicate that the relay function
is frozen Appears when Automatic Temperature Compensation (ATC) is enabled.
Not visible when Manual Temperature Compensation (MTC) is enabled. Flashes if the temperature probe is faulty in its ATC mode.
SETUP ME AS
READY HOLD
ERR
K=
8.8.8.8
-
-
CAL
.8.8.8
Lower Displ a y
R.
%
mS
µS
°F
°C ATC
(Refer Section 5.3)
ERR
K
Appears when an error occurs Electrode annunciator. Blinks during calibration error (Refer Section 4.2)
Buffer annunciator. Appears in calibration mode (Refer Section 4.2) Cell Constant (Refer Section 4.2)
10
Instruction Manual Alpha-CON 550
Units of Measurement Indicators mS µS R. % ºC ºF
Conductivity in Milli Siemens Conductivity in Micro Siemens Line Resistance in Ohms (Refer Section 5.5) Temperature Coefficient (Refer Section 5.2) Temperature in Celsius (Refer Section 5.3) Temperature in Fahrenheit (Refer Section 5.3)
2.5.2 Key Functions
lpha CON550
Key Description
CAL
Enter Calibration mode
Enter Setup mode Access sub screens (parameters) within a group of settings in Setup
ENT
mode Confirm (save) setup parameters and numerical values Start/Confirm calibration in Calibration mode. Select a group of settings in Setup mode. Select parameters and increment/decrement numerical values in Setup
and Calibration modes. (When pressed continuously, speed of value increment/decrement
increases)
ESC
CAL
Conductivity Monitor
ENT
Returns to Measurement mode when both keys are pressed simultaneously.
11
Instruction Manual Alpha-CON 550
3. Operation
3.1 Measurement mode
When the conductivity monitor is powered on, the display shows all the LCD segments briefly, and then automatically enters into the measurement mode.
MEAS
READY
µS
°C
25.0
The mode indicator ‘MEAS’ at the top of the display indicates that the conductivity monitor is in measurement mode. The upper alpha-numerical display shows the measured conductivity value, while the lower display shows the temperature value.
The monitor has five selectable measuring ranges. The unit of measurement and resolution of the conductivity reading depend on the selected measuring range. If the monitor reads a conductivity reading beyond the selected measuring range, the display shows ‘Or’ (over range) error indicator.
NOTE:
From measurement mode you can access:
Calibration mode (by pressing CAL key)
Setup mode (by pressing ENT key)
For more details, refer section 4 for Calibration mode and section 5 for Setup mode.)
To guarantee accurate readings, the measuring system (the conductivity monitor and the sensor) must be calibrated regularly.
ATC
12
Instruction Manual Alpha-CON 550
3.2 Menu Overview
READY
MEAS
ENT
µS
CAL ENT
°C
ATC
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
13
Instruction Manual Alpha-CON 550
4. Calibration Mode
4.1 About Calibration
Calibration should be carried out each time a new electrode is attached to the monitor or when you suspect that the monitor/electrode is out of calibration. The monitor allows you to perform temperature calibration and conductivity calibration.
About Temperature Calibration
It is important to ensure that temperature calibration is carried out prior to conductivity calibration since temperature readings affect the accuracy of conductivity measurements. It is recommended that temperature calibration should be carried out only if the temperature value displayed on the monitor is different from that of a calibrated thermometer. A temperature offset calibration of ± 10 °C/± 18 °F from the default reading is allowed. Once a temperature calibration is performed, conductivity calibration should be carried out to ensure the accuracy of conductivity measurements.
About Measuring Ranges & Conductivity Calibration
The monitor has five selectable conductivity measuring ranges. You can only calibrate one-point in the measuring range that you have selected for your process application. You have to recalibrate the monitor every time you change the measuring range since the calibration data of the monitor will be reset every time a new measuring range is selected.
About Conductivity Calibration
During conductivity calibration, the monitor allows a calibration window of ± 40% from the measured default reading of the calibration standard. The minimum allowable calibration point is 10% of the full scale reading of the measuring range selected.
For best results, select a standard conductivity solution close to the sample value you are measuring.
14
Instruction Manual Alpha-CON 550
A
4.2 Calibration
The monitor requires one-point calibration. Conductivity calibration is always carried out in the specific measuring range selected for the monitor. Make sure, the monitor is set to desired measuring range before you enter calibration mode (Refer Section 5.4 for selecting measuring ranges).
Select a suitable standard conductivity solution so that its conductivity value is within 10% to 100% of the full scale of the selected measuring range of the monitor. (For better accuracy, select the calibration standard to be nearly 50% of the full scale of the measuring range selected.) Prepare calibration standard solution in two beakers – one for rinsing and the other for calibration. Prepare separate de-ionized water for electrode rinsing.
All new calibration values will automatically override existing calibration data. If you wish to abort the calibration, press and keys simultaneously and the monitor reverts to measurement mode and the monitor will revert to previous calibration data.
READY
MEAS
1
µS
CAL
°C
ATC
HOLD
HOLD
From measurement mode press CAL key to enter calibration mode.
11
The display briefly shows the cell type (“2 CELL” or “4 CELL”) to which
2
the monitor is configured (Refer section 5.5 for cell type settings) Next, the last configured cell constant (k) is displayed in the upper
display. Use or constant of the conductivity probe. Adjustable range is 0.10 to 10.0. Press ENT key to confirm the new cell constant value.
Rinse the electrode in de-ionized water and then rinse with the calibration standard. Immerse the probe in calibration standard solution intended for calibration and swirl gently to create a homogenous sample. Make sure there are no any air-bubbles trapped in the sensor.
OR
2
HOLD
K=
HOLD
K=
CAL
CAL
2
ENT
READY HOLD
2
READY HOLD
CAL
CAL
3
µS
3
µS
ENT
READY HOLD
READY
CAL
MEAS
keys to adjust the cell constant to mach the cell
ENT
4
µS
°C
TC
The upper display shows the currently measured conductivity value of
3
the solution based on last calibration (if any). The lower display shows uncalibrated (default) reading of the solution. The buffer annunciator appears in LCD. The ‘READY’ annunciator appears when the reading is stable.
Once the reading stabilizes press the or key to adjust the upper display value to that of the standard solution. Press the ENT key to accept the value.
15
Instruction Manual Alpha-CON 550
NOTES:
The acceptable calibration window is ±40% of the displayed (default) value. For example, if the display is 1000 µS, the values to which it can be adjusted is 600 to 1400 µS.
If the conductivity reading of the solution is not within 10% to 100% of the full scale of the selected measuring range, calibration error occurs; LCD displays “ERR” and flash electrode annunciator.
If the conductivity reading of the solution is beyond the selected measuring range, calibration error occurs; LCD displays “Or” (over range), “ERR” and flashes electrode annunciator.
If calibration error occurs, exit form calibration mode by pressing and keys simultaneously.
Calibration is completed. The display shows the calibration factor (FCt).
4
This serves as an indication of the effectiveness of the sensor which will degrade with time and usage. An effective sensor should have a calibration factor of 0.60 to 1.40
Press ENT key to return to measurement mode.
NOTES:
If you use a 2-cell conductivity sensor, make sure that the pin 1 and 2 are shorted and pin 3 and 4 are shorted on the J10 connector. (Refer section 2.3.3 for more details)
When calibrating with manual temperature compensation, the monitor automatically changes from the preset process temperature to the calibration temperature. After leaving the Calibration mode, the monitor switches back to the process temperature (for setting the calibration temperature and the process temperature, see section 5.3)
WARNING: If you change the measuring range of the monitor after calibration is done, the monitor will reset back to its factory default calibration.
16
Instruction Manual Alpha-CON 550
5. Setup Mode
5.1 Enter Setup mode
The setup mode allows you to customize the settings of the conductivity monitor to suite your requirements. While in measurement mode, press the ENT key to access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup (rAng – Range settings). Press or key to access other pages of the setup mode.
READY
MEAS
µS
°C
ATC
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
NOTES:
To exit setup mode at any time press and keys simultaneously (escape). The monitor returns to the measurement mode
If you do not press any key for 2.5 minutes, the monitor automatically exists from the setup mode.
17
Instruction Manual Alpha-CON 550
5.2 Temperature Coefficient Settings
Setting the temperature coefficient to match your process/calibration liquids allows monitor to achieve conductivity readings with higher accuracy.
SETUP
HOLD
From measurement mode press ENT key to enter setup mode as
11
1
ENT
SETUP
HOLD
SETUP
HOLD
2
2
ENT
ENT
SETUP
HOLD
3
%
ENT
SETUP
HOLD
4
%
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press or key to select temperature coefficient settings screen (tC).
Press ENT key to access temperature coefficient settings (tC). Select type of temperature coefficient: The upper display shows the
2
last configured temperature coefficient type (Lin or PUr). Press ▲ or key to select the desired type based on type of your process liquid.
PUr :
Pure water temperature coefficient: Select this if your process
liquid is pure water or ultra pure water.
ENT
Lin :
Linear temperature coefficient: Select this for all other
process liquids
Press ENT key to confirm your selection.
If PUr was selected:
The monitor reverts back to tC screen
If Lin was selected:
Select temperature coefficient for process liquid: The lower
3
display shows ‘P.tC’. The upper display shows last configured temperature coefficient for process liquid (if any). The default value is 2.1%. Press or key to select the desired value. Allowable range: 0 to 10%.
Press ENT key to confirm your value. Select temperature coefficient for calibration liquid: The lower
4
display shows ‘C.tC’. The upper display shows last configured temperature coefficient for calibration liquid (if any). The default value is 2.1%. Press or key to select the desired value. Allowable range: 0 to 10%. Press ENT key to confirm your value. The monitor reverts to tC screen.
Press ▲ or ▼ key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
18
Instruction Manual Alpha-CON 550
A
5.3 Temperature Settings
Automatic Temperature Compensation (ATC):
The conductivity value of a sample is affected by temperature. Use ATC feature of the monitor to compensate for conductivity changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the monitor if an integrated conductivity/Temperature probe is not available.
Manual Temperature Compensation (MTC):
Set the monitor to MTC (disable ATC) when the temperature of sample or process liquid is constant and a temperature probe is not available. If you disable ATC, the monitor allows you to set your process temperature (P.ºC) and calibration temperature (C.ºC). This allows calibration at a different temperature other than the process temperature. Example: Setting calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your process temperature is different from 25°C.
SETUP
HOLD
1
ENT
SETUP
HOLD
SETUP
HOLD
2
°C
2
°F
ENT
ENT
SETUP
HOLD
SETUP
HOLD
3
°C
3
°C
ENT
ENT
SETUP
HOLD
SETUP
HOLD
4
°C
TC
6
ENT
SETUP
HOLD
SETUP
HOLD
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press or key to select Temperature settings screen (SET
ºCF). Press ENT key to access temperature settings (SET ºCF). Selecting unit of measurement for temperature: The upper display
2
shows ‘Unit’ and the lower display shows the last configured unit of measurement for temperature.
5
ENT
°C ATC
7
ENT
Press or key to select the desired units for temperature (ºC or ºF). Press ENT key to confirm your selection.
Enable/disable ATC: The lower display shows ‘AtC’ and the upper
3
display shows the last configured ATC selection (‘On’ or ‘OFF’). Press or key to enable (ATC On) or disable (ATC OFF) automatic
temperature compensation. Press ENT key to confirm your selection.
19
Instruction Manual Alpha-CON 550
If ATC enabled (ATC On):
Selecting temperature offset: The upper display shows the last
4
configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner.
Place a thermometer, which is known to be accurate, in your sample or process liquid. Make sure your temperature probe is placed in the same liquid. Compare the stabilized temperature reading displayed on the monitor with the thermometer. If there is a difference between the two readings (offset), you can adjust the reading of the monitor. Press or key to adjust the lower display to the correct temperature value.
As the lower display value changes, the monitor adjust the upper
55
display reading automatically to suit the new offset value. Up to ± 10 ºC/ ± 18 ºF offset is allowed.
Press ENT key to confirm the value. The monitor reverts to SET ºCF screen.
If ATC disabled (ATC OFF):
Setting process temperature: The lower display shows ‘P.ºC’ and
6
the upper display shows the last configured process temperature. Press or key to adjust the upper display to desired process temperature. Allowable range: –10.0 to 125.0°C / 14.0 to 257.0°F.
Press ENT key to confirm the process temperature. Setting calibration temperature: The lower display shows ‘C.ºC
7
and the upper display shows the last configured calibration temperature. Press or key to adjust the upper display to desired calibration temperature. Allowable range: –10.0 to 125.0°C /
14.0 to 257.0°F. Press ENT key to confirm the calibration temperature. The monitor
reverts to SET ºCF screen.
Press ▲ or ▼ key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously (escape). The monitor returns to the first screen of temperature settings SET
ºCF.
20
Instruction Manual Alpha-CON 550
5.4 Measuring Range Settings
The conductivity monitor has 5 different measuring ranges. You need to select a range which is the most suitable for conductivity reading of your applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm).
The measuring range you select here will affect the way the monitor behaves in the measurement mode and calibration mode. In measurement mode, if the conductivity reading of the monitor is above the selected measuring range, the LCD shows “Or” (over range) error. In calibration mode, the monitor adjusts calibration tolerance based on full scale of the selected measuring range. The calibration is carried out only for the selected measuring range.
WARNING: If you change the measuring range of the monitor after calibration is done, the monitor will reset back to its factory default calibration.
From measurement mode press
11
ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press ENT key to access measurement range setting (rAng).
The upper display shows the
2
currently selected measuring range. The factory default is 2000 µS. The lower display shows the currently configured cell constant (k). (Refer section 4.2 for setting cell constant)
Press or key to select the desired measuring range. Press ENT key to confirm your selection.
The monitor reverts to rAng screen.
Measuring range Resolution
0.00 to 20.00 µS/cm 0.01 µS/cm
0.0 to 200.0 µS/cm 0.1 µS/cm 0 to 2000 µS/cm 1 µS/cm
0.00 to 20.00 mS/cm 0.01 mS/cm
0.0 to 200.0 mS/cm 0.1 mS/cm
SETUP
HOLD
1
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
2
µS
ENT
2
µS
ENT
2
µS
ENT
2
mS
ENT
2
mS
ENT
Press ▲ or ▼ key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
21
Instruction Manual Alpha-CON 550
5.5 Configuration Settings
Configuration settings let you select between 2-cell & 4-cell and reset the monitor to factory defaults.
SETUP
HOLD
1
ENT
SETUP
HOLD
SETUP
HOLD
2
2
SETUP
ENT
HOLD
This screen appears only
when 2-cell is selected
and measuring range is set to
0.0 to 200.0 mS/cm
ENT
3
R.
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
READY
MEAS
ATC
4
ENT
4
ENT
4
ENT
µS
°C
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press or key to select configuration settings screen (Cnfg).
Press ENT key to access configuration settings (Cnfg). Select cell type of the sensor: The lower display shows ‘CEL’. The
2
upper display shows the last configured cell type (‘2’ or ‘4’). Press or key to select the cell type of your conductivity probe. Press ENT key to confirm your selection. NOTE: If you use a 2-cell conductivity sensor, make sure that the pin 1
and 2 are shorted and pin 3 and 4 are shorted on the J10 connector. (Refer section 2.3.3 for more details)
Setting line resistance: The lower display shows ‘L.Ad’ and the upper
3
display shows the last configured line resistance; otherwise the default is
0.00 R. NOTE: Line resistance of sensor cable constitutes to an error in
measurement of high conductivity values. The monitor will compensate this error if you specify line resistance value in this setting. The setting is available only if the cell type is set to 2-cell (step 3 above) and the
22
Instruction Manual Alpha-CON 550
measuring range is selected to ‘0.0 to 200.0 mS/cm’. Press or key to set the line resistance of the sensor cable
connected to the monitor.
Press ENT key to confirm your selection. Reset to factory defaults: The display shows ‘nO dEF’.
4
Press or key to select:
nO dEF : Not default reset of the monitor. User configured values
remain active, when confirmed by pressing ENT key.
FCt dEF: Reset all settings of the monitor to factory defaults, when
confirmed by pressing ENT key.
CAL dEF: Reset all calibration settings of the monitor to factory
defaults, when confirmed by pressing ENT key.
NOTE: Refer section 6 for factory/calibration default settings Press ENT key to confirm your selection.
- If ‘nO dEF’ was selected, the monitor reverts to Cnfg screen. Press ▲ or ▼ key to access other setup screens or press
and key simultaneously (escape) to return to measurement
mode.
- If ‘FCt dEF’ or ‘CAL dEF’ was selected, the monitor performs the
selected reset and returns to measurement mode.
Press ▲ or ▼ key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously (escape). The monitor returns to the first screen configuration settings (Cnfg).
23
Instruction Manual Alpha-CON 550
5.6 Viewing Electrode Properties
Each time you calibrate the conductivity sensor, the monitor re-calculates the calibration factor of the sensor and shown in the LCD at the end of calibration. The setup mode allows you to view the last saved calibration factor and cell constant at any time.
SETUP
HOLD
1
ENT
SETUP
HOLD
2
ENT
SETUP
HOLD
ENT
From measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press or key to select viewing electrode properties screen (CdAt).
Press ENT key to view electrode properties (CdAt). Cell constant: The lower display shows ’CEL’ and the upper display
2
shows the last configured cell constant.
Press ENT key. Calibration factor: The lower display shows ‘FCt’ and the upper display
3
shows the calibration factor based on the last conductivity calibration. NOTE: The upper display shows ‘- - - ‘ if the monitor has not been
calibrated yet or the monitor has been reset to factory defaults. Press ENT key to quit from electrode properties screen. The monitor
reverts to CdAt screen.
Press ▲ or ▼ key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
24
Instruction Manual Alpha-CON 550
6. Technical Specifications
General Specification
Conductivity
Measuring Range: Resolution:
0.00 - 20.00 µS/cm 0.01 µS/cm
0.0 - 200.0 µS/cm 0.1 µS/cm 0- 2000 µS/cm 1 µS/cm
0.00 - 20.00 mS/cm 0.01 mS/cm
0.0 - 200.0 mS/cm 0.1 mS/cm Relative accuracy ± 1% of full scale reading System accuracy The effective measuring range and the total system
accuracy depend on the sensor used with the monitor. Please refer respective sensor documentation.
Temperature
Measuring range -10.0 to +125.0 oC
+14.0 to +257.0 oF
Resolution 0.1 Relative accuracy ±0.5 oC Sensor 2 wire / 3 wire - Pt100 or Pt1000 (Jumper Selectable)
Compensation Automatic or Manual (reference at 25oC)
o
C
0.1 oF ±1.0 oF
Calibration
Conductivity
Number of calibration points 1 point Calibration window ±40% of default reading Calibration standard solution window 10% to 100% of the selected measuring range
Temperature
Offset Adjustment ± 10 ˚C/ ± 18 ˚F
Display
LCD 7 segments display with symbols for status information
Electrical Data and Connections
Conductivity Input Screw Terminal (3.5mm pitch) Connection terminal 8-pin & 2-pin terminal blocks, DC power connector
EMC Specifications
Emission EN 61 36 Susceptibility EN 61326
Other
Power Input Universal +9 V DC Adaptor and 24V DC Supply Dimensions (W x H x D) 96mm x 96mm x 66 mm Weight (Estimated) 210g Ambient Temp. operating range 0 to 40 oC Maximum Relative Humidity 80% up to 31 oC decreasing linearly to 50% at 40oC
(1.2mm diameter)
25
Instruction Manual Alpha-CON 550
Calibration Default Settings (CAL dEF)
Conductivity
Calibration factor Reset to 1.000
Factory Default Settings (FCt dEF)
Conductivity
Calibration factor Reset to 1.000 Cell constant 1.00 Cell type 2-cell Measuring range 0 to 2000 µS/cm Line resistance 0.000
Temperature
Unit of measurement
o
C Temperature compensation ATC On Temperature coefficient 2.1% (Both process & calibration temp. coefficients) Temperature coefficient type Linear MTC (ATC Off) 25 oC (Both process & calibration temperature)
26
Instruction Manual Alpha-CON 550
7. List of Accessories
Monitor Replacement and Accessories
Item Description Order Code
Alpha CON 550 Monitor EC-PHCP0550 Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 0.1 EC-CONSEN89X Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 1.0 EC-CONSEN88X Conductivity 2 Cell type probe, 0.1 - 200μS; Cell constant, K=0.1 with
integrated Pt 100, Material SS316 and 25ft cable (open-ended) Conductivity 2 Cell type probe, up to 200 mS; Cell constant, K=1.0 with
integrated Pt 100, Material SS316 and 25ft cable (open-ended)
NOTE: Please contact your authorised distributor or dealer for the prices of extension measuring cables and other accessories such as tee joints, electrode assembly, and calibration solutions.
EC-CS10-0-1S
EC-CS10-1-0S
27
Instruction Manual Alpha-CON 550
8. Troubleshooting
Problem
Power on, but no display a) Loose connections
Unstable conductivity reading
Oscillating temperature readings
Slow response a) Dirty / Oily probe a) Clean probe Blinking ATC
Blinking electrode annunciator
Or (Conductivity)
Or/Ur (Temperature)
Cause
b) Incorrect output voltage of
the power adaptor
c) Cables not in correct polarity
(+ and – position)
a) Air bubbles in the probe b) Dirty probe c) Probe not deep enough in
sample
a) Electrical nois e interference
a) No temperature probe
connection during ATC mode
a) Error in calibration b) Conductiv ity value of the
calibration standard is not within acceptable range
a) Conductiv ity reading is
above the selected measuring range
a) Temperature probe is not
connected when ATC enabled
b) Temperature value exceeds
above/ below the specified range
Solution
a) Ensure cables make good
contact
b) Use an power adaptor with
specified output voltage
c) Re-wire loop cables with correct
polarity
a) Tap probe to remove air bubbles b) Clean probe and recalibrate c) Ensure that the sample entirely
covers the sensor of the probe
a) Ensure shield wire is properly
connected to pin 7
a) Ensure temperature sensing
cable makes good contact
a) Ensure calibration standard
solution is not contaminated. Ensure electrode is clean
b) Ensure calibration standard is
within 10 to 100% of full-scale of the selected measuring range
a) Ensure correct conductivity
measurement range is selected
a) Ensure electrode makes good
contact with monitor
b) Ensure temperature calibration is
done properly. Ensure process temperature is within the specified range
28
Instruction Manual Alpha-CON 550
9. General Information
9.1 Warranty
Eutech Instruments warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. Eutech Instrument Customer Service Dept. will determine if the product problem is due to deviations or customer abuse. Out-of-warranty products will be repaired on an exchange basis at cost.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of the products
9.2 Return of Goods
Before returning goods for any reason whatsoever, the Customer Service Dept. has to be informed in advance. Items must be carefully packed to prevent damage during shipment, and insured against possible damage or loss. Eutech Instruments will not be responsible for any damage resulting from careless or insufficient packing.
Warning: Shipping damage as a result of inadequate packaging is the user's/distributor’s responsibility. Please follow the guidelines below before shipment.
9.3 Guidelines for Returning Unit for Repair
Use the original packaging material if possible when shipping the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for additional protection. Include a brief description of any faults suspected for the convenience of Customer Service Dept., if possible.
29
Instruction Manual Alpha-CON 550
10. Appendices
10.1 Appendix 2 Conductivity of various aqueous solutions
At Temperature 25 oC /77 oF
Solution Conductivity Resistivity
Ultra-pure Water 0.055 µS/cm 18.18 MΩ-cm Power Plant Boiler Water 0.05 - 1 µS/cm 1 - 18 MΩ-cm Distilled Water 0.5 µS/cm 2 MΩ-cm De-ionized Water 0.1 - 10 µS/cm 0.1 - 10 MΩ-cm De-mineralized Water 1 - 80 µS/cm 0.01 - 1 M-cm Mountain Water 10 µS/cm 0.1 MΩ-cm Drinking Water 0.5 - 1 mS/cm 1 - 2 kΩ-cm Waste-water 0.9 - 9 mS/cm 0.1 - 1 kΩ-cm Potable Water Maximum 1.5 mS/cm 0.7 k-cm Brackish Water 1 - 80 mS/cm 0.01 - 1 kΩ-cm Industrial Process Water 7 - 140 mS/cm Rarely stated Ocean Water 53 mS/cm Rarely stated
30
Instruction Manual Alpha-CON 550
10.2 Appendix 4 - Abbreviations Used in LCD
Abbreviation Description
AtC Automatic Temperature Compensation CAL Calibration mode C.ºC Calibr ation temperature CdAt Electrode properties CEL Cell CnFg Configuration C.tC Temperature coefficient for calibration liquid dEF Default values FCt Factory (defaults) HOLD Relay functions are hold K Cell constant L.Ad Linear resistance Lin Linear MEAS Measurement mode Or Reading is over range P.ºC Process temperature P.tC Temperature coefficient for process liquid PUr Pure water rAng Range Set Setting SETUP Setup mode tC Temperature Coefficient tyP Type Ur Reading is under range
31
NOTES
Water Analysis Instruments
North America
166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836
www.thermo.com/process
© 2009 Thermo Fisher Scientific Inc.
All rights reserved. Thermo Fisher Scientific Inc.
68X216870 Rev 4
Loading...