This manual serves to explain the use of the CON 500 transmitter. It functions in
two ways, firstly as a step by step guide to help you to operate the transmitter.
Secondly, it serves as a handy reference guide. It is written to cover as many
anticipated applications of the transmitter as possible. If there are doubts in the use
of the transmitter, please do not hesitate to contact the nearest Authorized
Distributor.
Eutech Instruments will not accept any responsibility for damage or malfunction to
the transmitter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as
improvements are made, and does not represent a commitment on the part of
Eutech Instruments.
12 RETURN OF ITEMS........................................................................27
Instruction Manual CON 500
A
1 INTRODUCTION
Thank you for selecting a CON 500 LCD Transmitter. This isolated output 4-20mA
transmitter is a sturdy microprocessor-based instrument that measures conductivity
and temperature and transmits its output via the 2-wire power supply loop.
This transmitter has many user-friendly features – all of which are completely
accessible through the water-resistant membrane keypad. Your transmitter includes
an instruction manual and a warranty card.
Please read this manual thoroughly before operating your transmitter.
lpha CON500
MEAS
READY
µ
S
1413
°C
.
TC
2-wire Conductivity Transmitter
- 1 -
Instruction Manual CON 500
2 PREPARATION
Remove screws from the four corners at the back of the Transmitter, and remove
back cover.
Connectors should be exposed as follows:
NC
NC
NC
Figure 1 – Connection Guide
All wiring is done on 2 detachable connectors: –
1. 9-pin connector (located on SL1 position) for Conductivity electrode and
temperature sensor; &
2. 3-pin connector (located on SL2 position) for power supply.
Using a suitable screwdriver, loosen screws from top of connector.
When inserting the wires, always hold connector with top screws facing up.
- 2 -
Instruction Manual CON 500
2.1 Pow er Supply Requirements (SL2 Position)
The CON 500 transmitter requires a 12 to 24V DC power supply. Other
Transmitters and/or a chart recorder may be connected in the loop.
1. Insert positive loop wire from power supply to pin 1, tighten screw.
2. Insert negative loop wire to pin 2, tighten screw. This wire may be linked to
a chart recorder or to negative terminal of power supply.
Alpha
CON 500
1
2
Power
Supply
Chart
Recorder
2.2 Connecting the Electrode and Temperature Sensor (SL1
Position)
2.2.1 To connect the Conductivity electrode:
The CON 500 transmitter accepts any industrial electrodes for conductivity
measurements and they are usually offered with bare wires.
1. Insert the electrode’s conductivity wire Low into pin 3 of SL1 connector.
Tighten the screw.
2. Insert the electrode’s conductivity wire High into pin 4 of SL1 connector.
Tighten the screw.
Pt 100 Input wire
Pt 100 Sense wire
Pt 100 Return wire
Conductivity wire High
Conductivity wire Low
Shield wire
To Electrode
- 3 -
Instruction Manual CON 500
2.2.2 To connect ATC temperature sensor:
For Automatic Temperature Compensated (ATC) Conductivity readings, an in-built
temperature sensor is usually integrated in the industrial conductivity electrodes.
Otherwise, a separate 100Ω Pt RTD temperature probe can be used.
3 wire temperature sensor
1. Insert the Pt100 input wire to pin 5 of SL1 connector. Tighten the screw.
2. Insert the Pt100 sense wire to pin 6 of SL1 connector. Tighten the screw.
3. Insert the Pt100 return wire and the shield wire to pin 7 of SL1 connector.
Tighten the screw.
2 wire temperature sensor
1. Take a small length of wire and short pins 5 and 6 of SL1 connector. Insert
Pt 100 sense wire into pin 6 of SL1 connector and tighten the screws.
2. Insert Pt 100 return wire and the shield wire to pin 7 of SL1 connector.
Tighten the screw.
IMPORTANT: It is necessary to connect the SHIELD wire to the pin 7 as stat ed
above to eliminate possible oscillation of the conductivity and/or temperature
reading due to electrical noise or interference.
Ensure that no bare wires are exposed and all screws are tighten ed fo r pro per
contact.
- 4 -
Instruction Manual CON 500
3 INSTALLATION
3.1 Mechanical Dimensions
lpha CON500
2-wire Conductivity Transmitter
3.2 Wall Mount
- 5 -
Instruction Manual CON 500
3.3 Panel Mount
1. Prepare panel cut-out of
92.0 mm by 92.0 mm
2. Remove back cover of
transmitter and slide it
through panel cut-out
Panel (side)
4. Thread rods through
lugs until transmitter
is held in place against
panel
3. Attach lugs to either
side of transmitter
- 6 -
Instruction Manual CON 500
r
r
4 DISPLAY AND KEYPAD FUNCTIONS
4.1 Display
The LCD has a primary and secondary display.
• The primary display shows the measured conductivity value.
• The secondary display shows the measured temperature. In Calibration
mode, measured conductivity values are displayed here.
Primary Display
(Upper Display)
123
4
16
15
14
13
1. SETUP - Setup mode
indicator
2. MEAS - Measurement
mode indicator
3. CAL - Calibration mode
indicator
4. R. – Line Resistance in
Ohms
SETUPMEASCAL
ON
K=
-.8.8.8
ERR
-1.8.8.8
11
12
5. % - Temperature
6. mS - Milli Siemens
7. µS - Micro Siemens
8. °C - Temperature
Coefficient
indicato
indicator
indicator
indicator
10
Secondary Display
(Lower Display)
9. ATC - Automatic Temp.
10. K = - Cell Constant
11. ERR – Wrong key
12. - Electrode indicator
Compensation
indicator.
press error
indicato
R.
%
mS
µ
S
°C
ATC
13. - Calibration
standard indicator
14. ON /OFF– READY set
up enable/disable
indicator.
15. HOLD – Hold indicator
16. READY – Ready
5
6
7
8
9
indicator
- 7 -
Instruction Manual CON 500
A
4.2 Keypad
lpha CON500
MEAS
READY
1413
2-wire Conductivity Transmitter
µ
S
°C
.
TC
The four-button keypad allows easy and quick operations of the Transmitter.
Key Function
CAL Brings you directly into the Calibration mode.
If you were in Conductivity Measurement mode, press CAL to enter
Conductivity Calibration mode.
▲ / ▼To scroll through the various submenus
To increment/decrement values or toggle between options (in the
SETUP/CAL modes)
When pressed together, serves as escape function to return to MEAS mode
from any point (CAL or SETUP modes).
ENT To confirm your calibration values in Calibration mode.
To confirm selections in SETUP mode.
While in MEAS mode, pressing ENT takes you directly to SETUP submenu
NOTE: Error indicator ‘ERR’ will be displayed if an invalid key is pressed during a
particular operation.
- 8 -
Instruction Manual CON 500
5 CALIBRATION
5.1 Impo rtant Information on Transmitter Calibration
Calibration should be carried when you are using your transmitter with a new
electrode for the first time or when you suspect that the transmitter/electrode is out
of calibration. Your transmitter allows you to perform temperature calibration and
conductivity calibration.
As temperature readings affect the accuracy of conductivity measurements, it is
recommended that temperature calibration should be carried out only if the
temperature value displayed on the transmitter is different from that of a calibrated
thermometer. A temperature offset calibration of ± 5 °C from the default reading is
allowed. Once a temperature calibration is performed, conductivity calibration
should be carried out to ensure the accuracy of conductivity measurements are
maintained.
Your transmitter has up to six selectable conductivity measuring ranges. The
selection for the sixth range, range 6 (0.00 to 199.9 mS/cm), only appears in the
Setup mode if the cell constant is selected to be K =1.0 or K =10 whereas range
1(0.00 to 19.99uS/cm), will only appear when cell constant is selected to be K = 0.1
or K = 1.0. Other wise, the selection will be limited to five ranges only.
You can only calibrate one point in the measuring range that you have selected for
your process application. You have to recalibrate your transmitter every time you
change the measuring range since the calibration data of your transmitter will be
erased every time a new measuring range is selected.
During conductivity calibration, the transmitter allows a calibration window of ± 40%
from the measured default reading of the calibration standard. The minimum
allowable calibration point is 10% of the full scale reading of the range selected.
For best results, select a standard value close to the sample value you are
measuring. Alternatively use a calibration solution value that is approximately 2/3
the full-scale value of the measurement range you plan to use. The following table 1
lists the corresponding conductivity ranges. You should calibrate, in the range that
you have selected, using a solution that falls between the values in the
“recommended calibration solution range” column.
- 9 -
Instruction Manual CON 500
Table 1
Conductivity
Measuring Range
0.00 Æ 19.99 µS 6.00 to 17.00 µS
0.0 Æ 199.9 µS 60.0 to 170.0 µS
0 Æ 1999 µS 600 to 1700 µS
0.00 Æ 10.00 mS 6.00 to 7.00 mS
0.00 Æ 19.99 mS 6.00 to 17.00 mS
0.0 Æ 199.9 mS 60.0 to 170.0 mS
To view current calibration point, see Section 6.8 - Viewing Calibration .
Do not reuse calibration solutions after calibration. Contaminants in the solution can
affect the calibration, and eventually the accuracy of the measurements. Use fresh
calibration solution each time you calibrate your meter.
NOTE: Your transmitter is factory set to a temperature coefficient of 2.1% per °C.
For most applications this will provide good results. The factory default value
for normalization temperature is 25 °C. If you need to set a different
temperature coefficient or change the normalization temperature value, see
Section 6.5 - Temperature Coefficient and Normalization Temperature
Setting for further information.
NOTE: In the CAL mode, the 4/20 mA output current will be automatically held.
‘HOLD’ will be displayed on LCD. After calibration is performed and when the
transmitter is back to the measurement mode, the ‘HOLD’ is automatically released.
Please refer to section 6.3 for more details.
Recommended Calibration Solution
Range
- 10 -
Instruction Manual CON 500
5.2 Temperature Calibration
You need to perform temperature calibration if your transmitter temperature reading
is inaccurate or if the transmitter’s cell constant setting is changed. It is because the
transmitter’s temperature offset calibration will be erased once a new cell constant
is selected.
It is important to ensure that temperature calibration is carried out first prior to
conductivity calibration since temperature readings affect the accuracy of
conductivity measurements. Your transmitter allows a one point temperature
calibration using reference of a thermometer known to be accurate. Please refer to
Section 6.2 - Temperature Compensation Setting for temperature calibration
procedure.
- 11 -
Instruction Manual CON 500
5.3 Conductivity Calibration
A 1-point Calibration is required for this transmitter. If
the calibration process is aborted, your transmitter
will revert to the previous calibration data.
1. Organize your calibration standard solution
in two beakers – one for rinsing and the
other for calibration. Prepare separate deionized water for electrode rinsing.
2. Rinse the electrode in de-ionized water and
then rinse with the calibration standard.
3. Dip the electrode to the calibration standard
intended for calibration and swirl gently to
create a homogenous sample. Wait for the
reading to stabilize.
4. Press the CAL key to enter calibration
mode. The upper display will show the
measured reading based on the last
calibration setting whereas the lower display
will show the measured default reading of
the standard.
5. Use the ▲ or ▼ keys to adjust the upper
display to the calibration standard value.
Note: You can offset the conductivity reading within
± 40% from the default reading of the standard. The
minimum allowable calibration point is 10% of the full
scale reading of the range selected.
6. Press the ENTER key to confirm the calibration value and to return to the
measurement mode.
Note: To exit this program without confirming the calibration, press the ▲ and ▼
keys together (Escape).
Note: When there is a calibration error, the buffer icon and ERR annunciator will
appear together with a blinking electrode icon. Press both ▲ and ▼ keys together
TWICE to return to the measurement mode.
HOLD
HOLD
MEAS
1578
25.5
CAL
1578
1578
CAL
1413
1578
ATC
µ
S
°C
µ
S
µ
S
- 12 -
Instruction Manual CON 500
6 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter’s preferences and
defaults. This transmitter features different sub groups that organize all setup
parameters. The sub-groups are:
6.1 Range and Zooming Selection Se tting
Your CON 500 transmitter provides six selections of conductivity measurement
ranges to suit your process application needs. Zooming selection is available where
both the 4 mA and 20 mA output current can be assigned to specific conductivity
values for a more refined output.
From the measurement mode,
1. Press the ENT key to enter the setup mode.
Transmitter displays ‘SEt rng’ indicating the
range setup mode.
2. Press the ENT key again. The display will show
the last set conductivity measurement range.
3. Use the ▲ or ▼ keys to scroll through the six
ranges and press the ENT key to select your
desired range. The transmitter will now switch to
the zooming selection and display the low zoom
value setup.
NOTE: Range 6 selection page will only appear in the
setup mode when the cell constant is selected as K =
1.0 or K = 10. Range 1 will only appear when cell
constant is selected as K =0.1 or K =1.0
SETUP
HOLD
SETUP
HOLD
19.99
µ
S
1
Six Range selection
- 13 -
Instruction Manual CON 500
4. For zooming setting, use the ▲ or ▼ keys to set
the conductivity low zoom value for the 4 mA
current and press the ENT key to confirm. The
transmitter now displays the high zoom value
SETUP
HOLD
0.00
µ
L
setup.
5. Use the ▲ or ▼ keys again to set the
conductivity high zoom value for the 20 mA
current and press the ENT key to confirm and
return to the range setup mode.
SETUP
HOLD
19.99
Hi
6. Press both ▲ and ▼ keys together to return to
the measurement mode.
S
µ
S
- 14 -
Instruction Manual CON 500
6.2 Temperature Compensation Setting
Conductivity readings are affected by temperature. Under varying temperature
conditions, use ATC to compensate for the conductivity values. If temperature of
sample is constant, and a temperature sensor/probe is not available, Manual
Temperature Compensation can be utilized.
6.2.1 Automatic Temperature Compensation
For automatic temperature compensation (ATC) selection, connect the ATC sensor
to the transmitter, as described in Section 2.2.2.
1. Press the ENT key to enter the setup mode. Use the ▲ or ▼ keys to scroll
through the sub-menus till LCD displays ‘
SEt oC’.
2. Press the ENT key to enter the temperature compensation mode. Display
shows ‘
AtC On’ or ‘AtC Off’. Use S or T keys to select ‘AtC On’ and
then press the ENT key.
3. Dip ATC sensor into a solution of known temperature (i.e. a temperature
bath). Allow time for temperature sensor to acclimatize with surrounding
bath temperature.
4. Use S or T keys to adjust the displayed reading to the correct
temperature value according to the temperature bath. Maximum
adjustments allowed is ± 5
o
C (± 9 oF) from the default reading.
5. Once you have adjusted to the correct temperature value, press the ENT
key to confirm the setting.
6. Press S and T keys together to return to the measurement mode.
The ATC indicator will light up on the LCD.
NOTE: If you are using a temperature sensor, the sensor must be submersed in
the liquid you are measuring.
- 15 -
Instruction Manual CON 500
6.2.2 Manual Temperature Compensation
For manual temperature compensation you can set the process and calibration
temperatures. This allows calibration at a temperature other than the process
temperature. Example: setting a calibration temperature of 25°C lets you calibrate
using standard solutions at 25°C, even if your process temperature is different from
25°C.
From the measurement mode,
1. Press the ENT key and use ▲ or ▼ keys to select ‘
SEt oC’.
2. Press the ENT key to enter the temperature compensation mode. Display
shows ‘
AtC On’ or ‘AtC Off’. Use S or T keys to select ‘AtC Off’ and
then press the ENT key.
3. The primary display will show temperature (default is 25.0), and the
secondary display shows ‘P °C’’. This is the page for setting the
temperature of your process or sample by using an accurate thermometer
as reference.
4. Use ▲ or ▼ keys to set the transmitter to the temperature of your process
or sample. Press the ENT key.
5. The primary display shows temperature (default is 25.0), and secondary
display shows ‘C °C’. This is the page for setting the temperature of your
calibration solutions.
6. Use ▲ or ▼ keys to set the transmitter to the temperature of your
calibration solutions. Press the ENT key.
7. Press S and T keys together, to return to the measurement mode.
The transmitter will now compensate conductivity readings for the manually set
temperature (values taken from P °C’).
- 16 -
Instruction Manual CON 500
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
.
°C
.
ATC
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
.
.
Temperature Compensation Setting Chart – alpha CON 500 Transmitter
- 17 -
Instruction Manual CON 500
6.3 HOLD Current Setting
When Transmitter is in CAL or SETUP modes, it
SETUP
HOLD
automatically goes into a ‘HOLD’ mode.
To indicate Transmitter is in ‘HOLD’ mode, output current
can be set to 22 mA output by activating the ‘HLD On’.
1. Press the ENT key and use ▲ or ▼ keys to scroll
till LCD displays ‘SEt’ in the upper display; and
‘HLd’ in the lower display. Press the ENT key
SETUP
HOLD
again.
2. Upper display now shows ‘HLd’. Lower display will
show either ‘OFF’ or ‘On’. Use ▲ or ▼ keys to
toggle between ‘On’ or ‘OFF’. Press ENT to accept
selection.
3. Press S and T keys together, to return to the
measurement mode.
SETUP
HOLD
NOTE: If ‘HLd’ is set to ‘OFF’, then current output will be
equivalent to last measured value.
6.4 Out-of-Range Current Setting
The transmitter can be set to give a fix current output of 3.8 mA when the
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
conductivity readings exceed the zoom setting range. This Out
of Range indicator feature is applicable in the uS ranges only.
1. Press the ENT key and use ▲ or ▼ keys to scroll till
LCD displays ‘SEt’ in the upper display; and ‘org’ in
the lower display.
2. Press the ENT key again. Upper display now shows
‘org’. Lower display will show either ‘OFF’ or ‘On’. Use
▲ or ▼ keys to toggle between ‘On’ or ‘OFF’. Press
the ENT key to confirm the selection.
3. Press S and T keys together, to return to the
measurement mode.
NOTE: If ‘org’ is set to ‘OFF’, current output below the low
zoom setting will be at 4 mA, while current output above the
high zoom setting will be at 20 mA.
- 18 -
Instruction Manual CON 500
6.5 Temperature Coefficient and Normalization Temperature
Setting
Since different process liquid may require different temperature coefficient factor for
its temperature compensation calculation, CON 500 transmitter allows 0 to 10%
temperature coefficient factor adjustment to cater for your different application
needs. You also have the selection of 25.0 °C or 20.0°C for the conductivity
measurement normalization temperature.
6.5.1 Setting the temperature coefficient
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
the ▲ or ▼ keys to scroll till LCD displays ‘SEt’
in the upper display; and ‘tPr’ in the lower
display.
2. Press the ENT key to view the temperature
coefficient setup page, ‘tC.O’.
3. Press the ENT key to enter the setup page and
use the ▲ or ▼ keys to set the temperature
coefficient value.
4. Press the ENT key to confirm your setting. The
transmitter will now switch to the normalization
setup page ‘tn.r’
6.5.2 Setting the normalization temperature
From Step 4 above,
1. Press the ENT key to enter the normalization
setup page.
2. Use the ▲ or ▼ keys to select the desired
normalization temperature (25.0 °C or 20.0°C).
SETUP
HOLD
.
°C
3. Press the ENT key to set the normalization
temperature.
4. Press S and T keys together, to return to
the measurement mode.
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
%
.
%
2.1
.
- 19 -
Instruction Manual CON 500
6.6 Cell Constant Setting
You can select up to three types of cell constant in the setup mode: - K=1.0, K=0.1
or K=10.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
SETUP
HOLD
the ▲ or ▼ keys to scroll till LCD displays ‘SEt’ in
the upper display; and ‘CEL’ in the lower display.
CEL
2. Press the ENT key to enter the cell constant setup
page.
3. Use the ▲ or ▼ keys to select the desired cell
constant and press the ENT key to confirm the
selection.
4. Press S and T keys together, to return to the
SETUP
K =
0.1
CEL
measurement mode.
6.7 Line resistance Setting
Line resistance of an electrode cable constitutes to error in measurement of high
conductivity values as in range 6. This error can be compensated by inputting the
electrode cable line resistance value in the setup mode.
From Range 6 measurement mode,
1. Press the ENT key to enter the setup mode. Use
the ▲ or ▼ keys to scroll till LCD displays ‘SEt’ in
the upper display; and ‘Lnr’ in the lower display.
2. Press the ENT key to enter the line resistance
setup page.
3. Use the ▲ or ▼ keys to set the line resistance
value (0 to 50 ohms) and press the ENT key to
confirm the setting.
4. Press S and T keys together, to return to the
measurement mode.
NOTE: The line resistance setup mode can be accessed
only from range 6 measurement mode.
SETUP
HOLD
SETUP
HOLD
0.00
R.
- 20 -
Instruction Manual CON 500
6.8 Viewing Calibration Point
This mode lets you view the current calibration point and its range.
From the measurement mode of a selected measuring
range,
1. Press the ENT key to enter the setup mode. Use
the ▲ or ▼ keys to scroll till LCD displays ‘CAL’
in the upper display.
2. Press the ENT key to enter the calibration point
viewing page. The display will show the current
calibration point and its range.
3. Press S and T keys together, to return to the
measurement mode.
NOTE: This setup mode will only show the calibration point for the selected
measuring range. If there is no calibration done in the selected range, the display
will show ‘---‘.
SETUP
HOLD
SETUP
12.88
CAL
CAL
mS
5
6.9 Viewing the Electrode Data
The CON 500 transmitter lets you check the electrode’s parameters for diagnostic
purposes. This option shows you the current effective cell constant and its range.
The cell constant is adjusted according to your calibration.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
the ▲ or ▼ keys to scroll till LCD displays ‘ELE’
in the upper display.
2. Press the ENT key to enter the electrode data
viewing page. The display will show the current
effective cell constant and its range.
3. Press S and T keys together, to return to the
measurement mode.
SETUP
HOLD
SETUP
K=
1.000
ELE
6
- 21 -
Instruction Manual CON 500
6.10 Reset Function
The reset function lets you choose to either reset the
SETUP
HOLD
transmitter’s conductivity calibration only or conductivity
calibration plus all setup functions that you might have
changed back to factory default settings. However, the
temperature compensation setting, as in Section 6.2, will
remain unchanged.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
SETUP
HOLD
the ▲ or ▼ keys to scroll till LCD displays ‘rSt’ in
the upper display.
2. Press the ENT key to enter the reset page. Use
the ▲ or ▼ keys to toggle between ‘CAL’ and
‘FCT’ for reset type selection.
•CAL – reset conductivity calibration only
SETUP
HOLD
• FCT – reset conductivity calibration and all
setup functions back to factory default
3. Press the ENT key to confirm your selection.
The lower display will blink momentarily before returning to the reset
function setup mode.
4. Press S and T keys together, to return to the measurement mode.
CAL
FC
- 22 -
Instruction Manual CON 500
7 FACTORY DEFAULT SETTINGS
SETUP
PAGE
FUNCTION
Selection of effective conductivity range
SET rng
Setting low zooming selection
Setting high zooming selection
SET °C
SET HLd
Selection of Automatic or Manual
Temperature Compensation
Selection of HOLD current to be of 22 mA or
the last current output value
Selecting Out of Range current output to be
SET org
constant 3.8 mA or 4mA for under range
and 20 mA for over range
SET tPr
Setting Temperature Coefficient
Setting Normalization Temperature
SET CEL Selecting the various cell constant K = 1.0
SET Lnr Setting the line resistance value 0.00
CAL Viewing current calibration point --- ; r3
ELE Viewing electrode’s current cell constant K = 1.000; r3
rSt
Selection of calibration reset or factory
default reset
DEFAULT
SETTING
1999 uS, r3
0 uS, Lo
1999 uS, Hi
AtC On
HLd OFF
org OFF
2.1%
25.0 °C
rSt CAL; rSt FCt
- 23 -
Instruction Manual CON 500
8 TROUBLE SHOOTING GUIDE
Problem Cause Solution
Power on but
no display
a). Loose connections
b). Cables not in correct
polarity (+ and – position).
a). Ensure cables are making
good contact.
b). Re-wire loop cables with
correct polarity.
Unstable
conductivity
readings
a). Air bubbles in probe.
b). Dirty probe.
c). Probe not deep enough in
sample.
a). Tap probe to remove
bubbles.
b). Clean probe and recalibrate.
c). Ensure that sample entirely
covers the probe sensors.
Oscillating
temperature
readings
a). Electrical noise
interference
a) Ensure shield wire is properly
connected to pin 7
Conductivity 2 Cell type probe, 0.1 - 200µS; Cell
constant, K=0.1 with integrated Pt 100, Material SS316
and 25ft cable (open-ended)
Conductivity 2 Cell type probe, up to 200 mS; Cell
constant, K=1.0 with integrated Pt 100, Material SS316
and 25ft cable (open-ended)
NOTE: Please contact your authorised distributor or dealer for the prices of
extension measuring cables and other accessories like tee joints, electrode
assembly, and calibration solutions.
ECCONSEN89X
ECCONSEN88X
EC-CS10-0-1S
EC-CS10-1-0S
- 26 -
Instruction Manual CON 500
11 WARRANTY
This transmitter is supplied with a one-year warranty against significant deviations
in material and workmanship from date of purchase and a six-month warranty for
probe. Each instrument will have a warranty card with a specific serial number. The
warranty card must be endorsed by the Authorized Distributor at the point of sale.
If repair or adjustment is necessary and has not been the result of abuse or misuse
within the designated period, please return – freight pre-paid – and correction will
be made without charge. Eutech Instruments will determine if the product problem
is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
• Improper or inadequate maintenance by customer
• Unauthorized modification or misuse
• Operation outside of the environment specifications of the products
12 RETURN OF ITEMS
Authorization must be obtained from our Customer Service Department or
authorized distributor before returning items for any reason. A “Return Goods
Authorization” (RGA) form is available through our authorized distributor. Please
include data regarding the reason the items are to be returned. For your protection,
items must be carefully packed to prevent damage in shipment and insured against
possible damage or loss. Eutech Instruments will not be responsible for damage
resulting from careless or insufficient packing. A restocking charge will be made on
all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd reserves the right to make improvements in
design, construction, and appearance of products without notice.
- 27 -
For more information on our products, contact your nearest distributor or visit our
website listed below:
Eutech Instruments Pte Ltd.
Blk 55, Ayer Rajah Crescent,
#04-16/24 Singapore 139949
Tel: (65) 6778 6876
Fax: (65) 6773 0836
E-mail: marketing@eutechinst.com
Web-site: www.eutechinst.com
Distributed by:
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.