Eutech Instruments ALPHA CON 500 User Manual

Instruction Manual
A
gyM
sy...
CON 500 LCD Transmitter
Conductivity Transmitter with Display
lpha CON500
MEAS
READY
µ
S
1413
.
TC
2-wire Conductivity Transmitter
Tec hn ol o
adeEa
68X216867
Rev 0 11/04
Preface
This manual serves to explain the use of the CON 500 transmitter. It functions in two ways, firstly as a step by step guide to help you to operate the transmitter. Secondly, it serves as a handy reference guide. It is written to cover as many anticipated applications of the transmitter as possible. If there are doubts in the use of the transmitter, please do not hesitate to contact the nearest Authorized Distributor.
Eutech Instruments will not accept any responsibility for damage or malfunction to the transmitter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments.
Copyright © 2004
All rights reserved
Eutech Instruments
Rev 0 11/04
TABLE OF CONTENTS
1
INTRODUCTION...................................................................................1
2 PREPARATION ....................................................................................2
2.1 POWER SUPPLY REQUIREMENTS (SL2 POSITION)........................................... 3
2.2 CONNECTING THE ELECTRODE AND TEMPERATURE SENSOR (SL1 POSITION).... 3
2.2.1 To connect the Conductivity electrode:.............................................3
2.2.2 To connect ATC temperature sensor:...............................................4
3 INSTALLATION ....................................................................................5
3.1 MECHANICAL DIMENSIONS ............................................................................ 5
3.2 WALL MOUNT..............................................................................................5
3.3 PANEL MOUNT ............................................................................................ 6
4 DISPLAY AND KEYPAD FUNCTIONS................................................7
4.1 DISPLAY ..................................................................................................... 7
4.2 KEYPAD ...................................................................................................... 8
5 CALIBRATION......................................................................................9
5.1 IMPORTANT INFORMATION ON TRANSMITTER CALIBRATION ............................... 9
5.2 TEMPERATURE CALIBRATION ...................................................................... 11
5.3 CONDUCTIVITY CALIBRATION ...................................................................... 12
6 ADVANCED SETUP FUNCTIONS.....................................................13
6.1 RANGE AND ZOOMING SELECTION SETTING.................................................. 13
6.2 TEMPERATURE COMPENSATION SETTING .....................................................15
6.2.1 Automatic Temperature Compensation ..........................................15
6.2.2 Manual Temperature Compensation...............................................16
6.3 HOLD CURRENT SETTING.......................................................................... 18
6.4 OUT-OF-RANGE CURRENT SETTING ............................................................ 18
6.5 TEMPERATURE COEFFICIENT AND NORMALIZATION TEMPERATURE SETTING ....19
6.5.1 Setting the temperature coefficient .................................................19
6.5.2 Setting the normalization temperature............................................19
6.6 CELL CONSTANT SETTING .......................................................................... 20
6.7 LINE RESISTANCE SETTING ......................................................................... 20
6.8 VIEWING CALIBRATION POINT......................................................................21
6.9 VIEWING THE ELECTRODE DATA.................................................................. 21
6.10 RESET FUNCTION ...................................................................................... 22
7 FACTORY DEFAULT SETTINGS ......................................................23
8 TROUBLE SHOOTING GUIDE...........................................................24
9 SPECIFICATIONS ..............................................................................25
10 ACCESSORIES...............................................................................26
11 WARRANTY....................................................................................27
12 RETURN OF ITEMS........................................................................27
Instruction Manual CON 500
A
1 INTRODUCTION
Thank you for selecting a CON 500 LCD Transmitter. This isolated output 4-20mA transmitter is a sturdy microprocessor-based instrument that measures conductivity and temperature and transmits its output via the 2-wire power supply loop.
This transmitter has many user-friendly features – all of which are completely accessible through the water-resistant membrane keypad. Your transmitter includes an instruction manual and a warranty card.
Please read this manual thoroughly before operating your transmitter.
lpha CON500
MEAS
READY
µ
S
1413
.
TC
2-wire Conductivity Transmitter
- 1 -
Instruction Manual CON 500
2 PREPARATION
Remove screws from the four corners at the back of the Transmitter, and remove back cover.
Connectors should be exposed as follows:
NC NC
NC
Figure 1 – Connection Guide
All wiring is done on 2 detachable connectors: –
1. 9-pin connector (located on SL1 position) for Conductivity electrode and temperature sensor; &
2. 3-pin connector (located on SL2 position) for power supply.
Using a suitable screwdriver, loosen screws from top of connector.
When inserting the wires, always hold connector with top screws facing up.
- 2 -
Instruction Manual CON 500
2.1 Pow er Supply Requirements (SL2 Position)
The CON 500 transmitter requires a 12 to 24V DC power supply. Other Transmitters and/or a chart recorder may be connected in the loop.
1. Insert positive loop wire from power supply to pin 1, tighten screw.
2. Insert negative loop wire to pin 2, tighten screw. This wire may be linked to a chart recorder or to negative terminal of power supply.
Alpha CON 500
1
2
Power Supply
Chart Recorder
2.2 Connecting the Electrode and Temperature Sensor (SL1 Position)
2.2.1 To connect the Conductivity electrode:
The CON 500 transmitter accepts any industrial electrodes for conductivity
measurements and they are usually offered with bare wires.
1. Insert the electrode’s conductivity wire Low into pin 3 of SL1 connector. Tighten the screw.
2. Insert the electrode’s conductivity wire High into pin 4 of SL1 connector. Tighten the screw.
Pt 100 Input wire
Pt 100 Sense wire
Pt 100 Return wire
Conductivity wire High
Conductivity wire Low
Shield wire
To Electrode
- 3 -
Instruction Manual CON 500
2.2.2 To connect ATC temperature sensor:
For Automatic Temperature Compensated (ATC) Conductivity readings, an in-built temperature sensor is usually integrated in the industrial conductivity electrodes. Otherwise, a separate 100 Pt RTD temperature probe can be used.
3 wire temperature sensor
1. Insert the Pt100 input wire to pin 5 of SL1 connector. Tighten the screw.
2. Insert the Pt100 sense wire to pin 6 of SL1 connector. Tighten the screw.
3. Insert the Pt100 return wire and the shield wire to pin 7 of SL1 connector. Tighten the screw.
2 wire temperature sensor
1. Take a small length of wire and short pins 5 and 6 of SL1 connector. Insert Pt 100 sense wire into pin 6 of SL1 connector and tighten the screws.
2. Insert Pt 100 return wire and the shield wire to pin 7 of SL1 connector. Tighten the screw.
IMPORTANT: It is necessary to connect the SHIELD wire to the pin 7 as stat ed above to eliminate possible oscillation of the conductivity and/or temperature reading due to electrical noise or interference.
Ensure that no bare wires are exposed and all screws are tighten ed fo r pro per contact.
- 4 -
Instruction Manual CON 500
3 INSTALLATION
3.1 Mechanical Dimensions
lpha CON500
2-wire Conductivity Transmitter
3.2 Wall Mount
- 5 -
Instruction Manual CON 500
3.3 Panel Mount
1. Prepare panel cut-out of
92.0 mm by 92.0 mm
2. Remove back cover of transmitter and slide it through panel cut-out
Panel (side)
4. Thread rods through lugs until transmitter is held in place against
panel
3. Attach lugs to either side of transmitter
- 6 -
Instruction Manual CON 500
r
r
4 DISPLAY AND KEYPAD FUNCTIONS
4.1 Display
The LCD has a primary and secondary display.
The primary display shows the measured conductivity value.
The secondary display shows the measured temperature. In Calibration
mode, measured conductivity values are displayed here.
Primary Display
(Upper Display)
123
4
16 15
14
13
1. SETUP - Setup mode
indicator
2. MEAS - Measurement
mode indicator
3. CAL - Calibration mode
indicator
4. R. – Line Resistance in
Ohms
SETUP MEAS CAL
ON
K=
-.8.8.8
ERR
-1.8.8.8
11
12
5. % - Temperature
6. mS - Milli Siemens
7. µS - Micro Siemens
8. °C - Temperature
Coefficient indicato
indicator
indicator
indicator
10
Secondary Display
(Lower Display)
9. ATC - Automatic Temp.
10. K = - Cell Constant
11. ERR – Wrong key
12. - Electrode indicator
Compensation
indicator.
press error indicato
R.
%
mS
µ
S
°C
ATC
13. - Calibration standard indicator
14. ON /OFF– READY set
up enable/disable indicator.
15. HOLD – Hold indicator
16. READY – Ready
5 6 7
8
9
indicator
- 7 -
Instruction Manual CON 500
A
4.2 Keypad
lpha CON500
MEAS
READY
1413
2-wire Conductivity Transmitter
µ
S
°C
.
TC
The four-button keypad allows easy and quick operations of the Transmitter.
Key Function
CAL Brings you directly into the Calibration mode.
If you were in Conductivity Measurement mode, press CAL to enter Conductivity Calibration mode.
/ To scroll through the various submenus
To increment/decrement values or toggle between options (in the SETUP/CAL modes)
When pressed together, serves as escape function to return to MEAS mode from any point (CAL or SETUP modes).
ENT To confirm your calibration values in Calibration mode.
To confirm selections in SETUP mode.
While in MEAS mode, pressing ENT takes you directly to SETUP submenu
NOTE: Error indicator ‘ERR’ will be displayed if an invalid key is pressed during a particular operation.
- 8 -
Instruction Manual CON 500
5 CALIBRATION
5.1 Impo rtant Information on Transmitter Calibration
Calibration should be carried when you are using your transmitter with a new electrode for the first time or when you suspect that the transmitter/electrode is out of calibration. Your transmitter allows you to perform temperature calibration and conductivity calibration.
As temperature readings affect the accuracy of conductivity measurements, it is recommended that temperature calibration should be carried out only if the temperature value displayed on the transmitter is different from that of a calibrated thermometer. A temperature offset calibration of ± 5 °C from the default reading is allowed. Once a temperature calibration is performed, conductivity calibration should be carried out to ensure the accuracy of conductivity measurements are maintained.
Your transmitter has up to six selectable conductivity measuring ranges. The selection for the sixth range, range 6 (0.00 to 199.9 mS/cm), only appears in the Setup mode if the cell constant is selected to be K =1.0 or K =10 whereas range 1(0.00 to 19.99uS/cm), will only appear when cell constant is selected to be K = 0.1 or K = 1.0. Other wise, the selection will be limited to five ranges only.
You can only calibrate one point in the measuring range that you have selected for your process application. You have to recalibrate your transmitter every time you change the measuring range since the calibration data of your transmitter will be erased every time a new measuring range is selected.
During conductivity calibration, the transmitter allows a calibration window of ± 40% from the measured default reading of the calibration standard. The minimum allowable calibration point is 10% of the full scale reading of the range selected.
For best results, select a standard value close to the sample value you are measuring. Alternatively use a calibration solution value that is approximately 2/3 the full-scale value of the measurement range you plan to use. The following table 1 lists the corresponding conductivity ranges. You should calibrate, in the range that you have selected, using a solution that falls between the values in the “recommended calibration solution range” column.
- 9 -
Instruction Manual CON 500
Table 1
Conductivity
Measuring Range
0.00 Æ 19.99 µS 6.00 to 17.00 µS
0.0 Æ 199.9 µS 60.0 to 170.0 µS
0 Æ 1999 µS 600 to 1700 µS
0.00 Æ 10.00 mS 6.00 to 7.00 mS
0.00 Æ 19.99 mS 6.00 to 17.00 mS
0.0 Æ 199.9 mS 60.0 to 170.0 mS
To view current calibration point, see Section 6.8 - Viewing Calibration .
Do not reuse calibration solutions after calibration. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. Use fresh calibration solution each time you calibrate your meter.
NOTE: Your transmitter is factory set to a temperature coefficient of 2.1% per °C.
For most applications this will provide good results. The factory default value for normalization temperature is 25 °C. If you need to set a different temperature coefficient or change the normalization temperature value, see Section 6.5 - Temperature Coefficient and Normalization Temperature Setting for further information.
NOTE: In the CAL mode, the 4/20 mA output current will be automatically held.
‘HOLD’ will be displayed on LCD. After calibration is performed and when the transmitter is back to the measurement mode, the ‘HOLD’ is automatically released. Please refer to section 6.3 for more details.
Recommended Calibration Solution
Range
- 10 -
Instruction Manual CON 500
5.2 Temperature Calibration
You need to perform temperature calibration if your transmitter temperature reading is inaccurate or if the transmitter’s cell constant setting is changed. It is because the transmitter’s temperature offset calibration will be erased once a new cell constant is selected.
It is important to ensure that temperature calibration is carried out first prior to conductivity calibration since temperature readings affect the accuracy of conductivity measurements. Your transmitter allows a one point temperature calibration using reference of a thermometer known to be accurate. Please refer to Section 6.2 - Temperature Compensation Setting for temperature calibration procedure.
- 11 -
Instruction Manual CON 500
5.3 Conductivity Calibration
A 1-point Calibration is required for this transmitter. If the calibration process is aborted, your transmitter will revert to the previous calibration data.
1. Organize your calibration standard solution in two beakers – one for rinsing and the other for calibration. Prepare separate de­ionized water for electrode rinsing.
2. Rinse the electrode in de-ionized water and then rinse with the calibration standard.
3. Dip the electrode to the calibration standard intended for calibration and swirl gently to create a homogenous sample. Wait for the reading to stabilize.
4. Press the CAL key to enter calibration mode. The upper display will show the measured reading based on the last calibration setting whereas the lower display will show the measured default reading of the standard.
5. Use the or keys to adjust the upper display to the calibration standard value.
Note: You can offset the conductivity reading within
± 40% from the default reading of the standard. The minimum allowable calibration point is 10% of the full scale reading of the range selected.
6. Press the ENTER key to confirm the calibration value and to return to the measurement mode.
Note: To exit this program without confirming the calibration, press the ▲ and
keys together (Escape).
Note: When there is a calibration error, the buffer icon and ERR annunciator will
appear together with a blinking electrode icon. Press both and keys together TWICE to return to the measurement mode.
HOLD
HOLD
MEAS
1578
25.5
CAL
1578
1578
CAL
1413
1578
ATC
µ
S
°C
µ
S
µ
S
- 12 -
Instruction Manual CON 500
6 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter’s preferences and defaults. This transmitter features different sub groups that organize all setup parameters. The sub-groups are:
6.1 Range and Zooming Selection Se tting
Your CON 500 transmitter provides six selections of conductivity measurement ranges to suit your process application needs. Zooming selection is available where both the 4 mA and 20 mA output current can be assigned to specific conductivity values for a more refined output.
From the measurement mode,
1. Press the ENT key to enter the setup mode.
Transmitter displays ‘SEt rng’ indicating the
range setup mode.
2. Press the ENT key again. The display will show the last set conductivity measurement range.
3. Use the ▲ or ▼ keys to scroll through the six ranges and press the ENT key to select your desired range. The transmitter will now switch to the zooming selection and display the low zoom value setup.
NOTE: Range 6 selection page will only appear in the setup mode when the cell constant is selected as K =
1.0 or K = 10. Range 1 will only appear when cell
constant is selected as K =0.1 or K =1.0
SETUP
HOLD
SETUP
HOLD
19.99
µ
S
1
Six Range selection
- 13 -
Instruction Manual CON 500
4. For zooming setting, use the or keys to set the conductivity low zoom value for the 4 mA current and press the ENT key to confirm. The transmitter now displays the high zoom value
SETUP
HOLD
0.00
µ
L
setup.
5. Use the ▲ or ▼ keys again to set the conductivity high zoom value for the 20 mA current and press the ENT key to confirm and return to the range setup mode.
SETUP
HOLD
19.99
Hi
6. Press both ▲ and ▼ keys together to return to the measurement mode.
S
µ
S
- 14 -
Instruction Manual CON 500
6.2 Temperature Compensation Setting
Conductivity readings are affected by temperature. Under varying temperature conditions, use ATC to compensate for the conductivity values. If temperature of sample is constant, and a temperature sensor/probe is not available, Manual Temperature Compensation can be utilized.
6.2.1 Automatic Temperature Compensation
For automatic temperature compensation (ATC) selection, connect the ATC sensor to the transmitter, as described in Section 2.2.2.
1. Press the ENT key to enter the setup mode. Use the or keys to scroll through the sub-menus till LCD displays ‘
SEt oC’.
2. Press the ENT key to enter the temperature compensation mode. Display shows ‘
AtC On’ or ‘AtC Off’. Use S or T keys to select ‘AtC On’ and
then press the ENT key.
3. Dip ATC sensor into a solution of known temperature (i.e. a temperature bath). Allow time for temperature sensor to acclimatize with surrounding bath temperature.
4. Use S or T keys to adjust the displayed reading to the correct temperature value according to the temperature bath. Maximum adjustments allowed is ± 5
o
C (± 9 oF) from the default reading.
5. Once you have adjusted to the correct temperature value, press the ENT key to confirm the setting.
6. Press S and T keys together to return to the measurement mode.
The ATC indicator will light up on the LCD.
NOTE: If you are using a temperature sensor, the sensor must be submersed in the liquid you are measuring.
- 15 -
Instruction Manual CON 500
6.2.2 Manual Temperature Compensation
For manual temperature compensation you can set the process and calibration temperatures. This allows calibration at a temperature other than the process temperature. Example: setting a calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your process temperature is different from 25°C.
From the measurement mode,
1. Press the ENT key and use or keys to select ‘
SEt oC’.
2. Press the ENT key to enter the temperature compensation mode. Display shows ‘
AtC On’ or ‘AtC Off’. Use S or T keys to select ‘AtC Off’ and
then press the ENT key.
3. The primary display will show temperature (default is 25.0), and the secondary display shows ‘P °C’’. This is the page for setting the temperature of your process or sample by using an accurate thermometer as reference.
4. Use ▲ or ▼ keys to set the transmitter to the temperature of your process or sample. Press the ENT key.
5. The primary display shows temperature (default is 25.0), and secondary display shows ‘C °C’. This is the page for setting the temperature of your calibration solutions.
6. Use ▲ or ▼ keys to set the transmitter to the temperature of your calibration solutions. Press the ENT key.
7. Press S and T keys together, to return to the measurement mode.
The transmitter will now compensate conductivity readings for the manually set temperature (values taken from P °C’).
- 16 -
Instruction Manual CON 500
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
.
°C
.
ATC
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
.
.
Temperature Compensation Setting Chart – alpha CON 500 Transmitter
- 17 -
Instruction Manual CON 500
6.3 HOLD Current Setting
When Transmitter is in CAL or SETUP modes, it
SETUP
HOLD
automatically goes into a ‘HOLD’ mode.
To indicate Transmitter is in ‘HOLD’ mode, output current can be set to 22 mA output by activating the ‘HLD On’.
1. Press the ENT key and use or keys to scroll till LCD displays ‘SEt’ in the upper display; and ‘HLd’ in the lower display. Press the ENT key
SETUP
HOLD
again.
2. Upper display now shows ‘HLd’. Lower display will show either ‘OFF’ or ‘On’. Use or keys to toggle between ‘On’ or ‘OFF’. Press ENT to accept selection.
3. Press S and T keys together, to return to the measurement mode.
SETUP
HOLD
NOTE: If ‘HLd’ is set to ‘OFF’, then current output will be equivalent to last measured value.
6.4 Out-of-Range Current Setting
The transmitter can be set to give a fix current output of 3.8 mA when the
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
conductivity readings exceed the zoom setting range. This Out of Range indicator feature is applicable in the uS ranges only.
1. Press the ENT key and use or keys to scroll till LCD displays ‘SEt’ in the upper display; and ‘org’ in the lower display.
2. Press the ENT key again. Upper display now shows ‘org’. Lower display will show either ‘OFF’ or ‘On’. Use or keys to toggle between ‘On’ or ‘OFF’. Press the ENT key to confirm the selection.
3. Press S and T keys together, to return to the measurement mode.
NOTE: If ‘org’ is set to ‘OFF’, current output below the low zoom setting will be at 4 mA, while current output above the high zoom setting will be at 20 mA.
- 18 -
Instruction Manual CON 500
6.5 Temperature Coefficient and Normalization Temperature Setting
Since different process liquid may require different temperature coefficient factor for its temperature compensation calculation, CON 500 transmitter allows 0 to 10% temperature coefficient factor adjustment to cater for your different application needs. You also have the selection of 25.0 °C or 20.0°C for the conductivity measurement normalization temperature.
6.5.1 Setting the temperature coefficient
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use the or keys to scroll till LCD displays ‘SEt’ in the upper display; and ‘tPr’ in the lower display.
2. Press the ENT key to view the temperature coefficient setup page, ‘tC.O’.
3. Press the ENT key to enter the setup page and use the or keys to set the temperature coefficient value.
4. Press the ENT key to confirm your setting. The transmitter will now switch to the normalization setup page ‘tn.r’
6.5.2 Setting the normalization temperature
From Step 4 above,
1. Press the ENT key to enter the normalization setup page.
2. Use the ▲ or ▼ keys to select the desired normalization temperature (25.0 °C or 20.0°C).
SETUP
HOLD
.
°C
3. Press the ENT key to set the normalization temperature.
4. Press S and T keys together, to return to the measurement mode.
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
%
.
%
2.1
.
- 19 -
Instruction Manual CON 500
6.6 Cell Constant Setting
You can select up to three types of cell constant in the setup mode: - K=1.0, K=0.1 or K=10.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
SETUP
HOLD
the or keys to scroll till LCD displays ‘SEt’ in the upper display; and ‘CEL’ in the lower display.
CEL
2. Press the ENT key to enter the cell constant setup page.
3. Use the ▲ or ▼ keys to select the desired cell constant and press the ENT key to confirm the selection.
4. Press S and T keys together, to return to the
SETUP
K =
0.1
CEL
measurement mode.
6.7 Line resistance Setting
Line resistance of an electrode cable constitutes to error in measurement of high conductivity values as in range 6. This error can be compensated by inputting the electrode cable line resistance value in the setup mode.
From Range 6 measurement mode,
1. Press the ENT key to enter the setup mode. Use the or keys to scroll till LCD displays ‘SEt’ in the upper display; and ‘Lnr’ in the lower display.
2. Press the ENT key to enter the line resistance setup page.
3. Use the ▲ or ▼ keys to set the line resistance value (0 to 50 ohms) and press the ENT key to confirm the setting.
4. Press S and T keys together, to return to the measurement mode.
NOTE: The line resistance setup mode can be accessed only from range 6 measurement mode.
SETUP
HOLD
SETUP
HOLD
0.00
R.
- 20 -
Instruction Manual CON 500
6.8 Viewing Calibration Point
This mode lets you view the current calibration point and its range.
From the measurement mode of a selected measuring range,
1. Press the ENT key to enter the setup mode. Use the or keys to scroll till LCD displays ‘CAL’ in the upper display.
2. Press the ENT key to enter the calibration point viewing page. The display will show the current calibration point and its range.
3. Press S and T keys together, to return to the measurement mode.
NOTE: This setup mode will only show the calibration point for the selected measuring range. If there is no calibration done in the selected range, the display will show ‘---‘.
SETUP
HOLD
SETUP
12.88
CAL
CAL
mS
5
6.9 Viewing the Electrode Data
The CON 500 transmitter lets you check the electrode’s parameters for diagnostic purposes. This option shows you the current effective cell constant and its range. The cell constant is adjusted according to your calibration.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use the or keys to scroll till LCD displays ‘ELE’ in the upper display.
2. Press the ENT key to enter the electrode data viewing page. The display will show the current effective cell constant and its range.
3. Press S and T keys together, to return to the measurement mode.
SETUP
HOLD
SETUP
K=
1.000
ELE
6
- 21 -
Instruction Manual CON 500
6.10 Reset Function
The reset function lets you choose to either reset the
SETUP
HOLD
transmitter’s conductivity calibration only or conductivity calibration plus all setup functions that you might have changed back to factory default settings. However, the temperature compensation setting, as in Section 6.2, will remain unchanged.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
SETUP
HOLD
the or keys to scroll till LCD displays ‘rSt’ in the upper display.
2. Press the ENT key to enter the reset page. Use the or keys to toggle between ‘CAL’ and ‘FCT’ for reset type selection.
CAL – reset conductivity calibration only
SETUP
HOLD
FCT – reset conductivity calibration and all
setup functions back to factory default
3. Press the ENT key to confirm your selection. The lower display will blink momentarily before returning to the reset function setup mode.
4. Press S and T keys together, to return to the measurement mode.
CAL
FC
- 22 -
Instruction Manual CON 500
7 FACTORY DEFAULT SETTINGS
SETUP
PAGE
FUNCTION
Selection of effective conductivity range
SET rng
Setting low zooming selection Setting high zooming selection
SET °C
SET HLd
Selection of Automatic or Manual Temperature Compensation
Selection of HOLD current to be of 22 mA or the last current output value
Selecting Out of Range current output to be
SET org
constant 3.8 mA or 4mA for under range and 20 mA for over range
SET tPr
Setting Temperature Coefficient Setting Normalization Temperature
SET CEL Selecting the various cell constant K = 1.0
SET Lnr Setting the line resistance value 0.00
CAL Viewing current calibration point --- ; r3
ELE Viewing electrode’s current cell constant K = 1.000; r3
rSt
Selection of calibration reset or factory default reset
DEFAULT
SETTING
1999 uS, r3
0 uS, Lo
1999 uS, Hi
AtC On
HLd OFF
org OFF
2.1%
25.0 °C
rSt CAL; rSt FCt
- 23 -
Instruction Manual CON 500
8 TROUBLE SHOOTING GUIDE
Problem Cause Solution
Power on but no display
a). Loose connections
b). Cables not in correct polarity (+ and – position).
a). Ensure cables are making good contact.
b). Re-wire loop cables with correct polarity.
Unstable conductivity readings
a). Air bubbles in probe.
b). Dirty probe.
c). Probe not deep enough in sample.
a). Tap probe to remove bubbles.
b). Clean probe and recalibrate.
c). Ensure that sample entirely covers the probe sensors.
Oscillating temperature readings
a). Electrical noise interference
a) Ensure shield wire is properly connected to pin 7
Slow response a). Dirty / Oily probe. a). Clean probe.
Blinking ATC
Blinking electrode indicator
Or (Conductivity)
a) No temperature sensor connection during ATC mode
a) Error in calibration
a) Conductivity measurement value is above
a) Ensure temperature sensing cable are making good contact
a) Ensure calibration standard solution is not contaminated. Ensure clean probe is used.
a) Ensure correct conductivity measurement range is selected.
the specified range
Or / Ur (Temperature)
b) Temperature value exceeds above/ below the specified range.
b) Ensure temperature calibration is done properly. Ensure process temperature is within the specified range.
- 24 -
Instruction Manual CON 500
9 SPECIFICATIONS
Cell Constant
(K)
0.1, 1.0
0.1, 1.0, 10.0
0.1, 1.0, 10.0
0.1, 1.0, 10.0
0.1, 1.0, 10.0
1.0, 10.0
Conductivity Range
(r1 to r6)
Range 1 = 0 – 19.99 uS/cm
Range 2 = 0 – 199.9 uS/cm
Range 3 = 0 – 1999 uS/cm
Range 4 = 0 – 10.00 mS/cm
Range 5 = 0 – 19.99 mS/cm
Range 6 = 0 – 199.9 mS/cm
Resolution
0.01 uS
0.1uS
1 uS
0.01 mS
0.01 mS
0.1 mS
Temperature Range 0-100 oC
Temperature Resolution 0.1 oC
Temperature Relative Accuracy
Temperature Compensation
Temperature Coefficient Factor Setting 0 to 10%
Normalization Temperature Setting 25.0 oC or 20.0 oC
Number of calibration points 1
Output 4.0 to 20.0 mA
Over- / Under range current output
(Only applicable in the uS range)
HOLD current output ‘ON’ Select - 22 mA
Display Custom Dual LCD
Averaging/Stability (READY indicator) Yes
Operating Temperature 0 to 100 oC
Power Requirements 12 to 24 VDC
Load Resistance Max. 600
Dimension / Weight 96 x 96 x 66 mm / 210 g
o
± 0.5
C
Auto (Pt 100) (from 0 to 100 °C)
Manual (from 0 to 100 °C)
‘ON’ Select - 3.8 mA
‘OFF’ Select - 4 mA (under-range) & 20 mA (over-range)
‘OFF’ Select - Last measured current value
- 25 -
Instruction Manual CON 500
10 ACCESSORIES
Replacement Unit
Product Description Eutech Instruments
Order Code
CON 500 LCD Transmitter EC-CONCTP0500
Assembly Accessories
Product Description Eutech Instruments
Order Code
Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 0.1
Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 1.0
Conductivity 2 Cell type probe, 0.1 - 200µS; Cell constant, K=0.1 with integrated Pt 100, Material SS316 and 25ft cable (open-ended)
Conductivity 2 Cell type probe, up to 200 mS; Cell constant, K=1.0 with integrated Pt 100, Material SS316 and 25ft cable (open-ended)
NOTE: Please contact your authorised distributor or dealer for the prices of extension measuring cables and other accessories like tee joints, electrode assembly, and calibration solutions.
ECCONSEN89X
ECCONSEN88X
EC-CS10-0-1S
EC-CS10-1-0S
- 26 -
Instruction Manual CON 500
11 WARRANTY
This transmitter is supplied with a one-year warranty against significant deviations in material and workmanship from date of purchase and a six-month warranty for probe. Each instrument will have a warranty card with a specific serial number. The warranty card must be endorsed by the Authorized Distributor at the point of sale.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Eutech Instruments will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of the products
12 RETURN OF ITEMS
Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Goods Authorization” (RGA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd reserves the right to make improvements in design, construction, and appearance of products without notice.
- 27 -
For more information on our products, contact your nearest distributor or visit our website listed below:
Eutech Instruments Pte Ltd.
Blk 55, Ayer Rajah Crescent,
#04-16/24 Singapore 139949
Tel: (65) 6778 6876
Fax: (65) 6773 0836
E-mail: marketing@eutechinst.com
Web-site: www.eutechinst.com
Distributed by:
Loading...