ETS-Lindgren HI-4433 User Manual

Model HI-4433
Broadband Isotropic
Field Probe
User Manual
ii |
| iii
ETS-Lindgren L.P. reserves the right to make changes to any product described
herein in order to improve function, design, or for any other reason. Nothing
contained herein shall constitute ETS-Lindgren L.P. assuming any liability
whatsoever arising out of the application or use of any product or circuit
described herein. ETS-Lindgren L.P. does not convey any license under its
patent rights or the rights of others.
© Copyright 1993–2016 by ETS-Lindgren L.P. All Rights Reserved. No part
of this document may be copied by any means without written permission
from ETS-Lindgren L.P.
Trademarks used in this document: The ETS-Lindgren logo is a trademark of
ETS-Lindgren L.P.
Revision Record | HI-4433 Series Probe, MANUAL | Part #600056, Rev. L
Revision Description Date
Initial Release October, 1993
A Revised September, 1994
B Revised June, 1995
C Added Probes September, 1995
D Revised August, 1996
E Changed Charger October, 1997
F Added CE Label
October, 1997
G Revised
January 1998
H Changed Charger Specs
August, 1999
I
Changed Area Code February, 2000
J
Revised August, 2005
iv |
Revision Description Date
K
Updated branding: Revised to meet Style Guide specifications; reformatted to half-size; PIB included with release
March, 2009
L
Updated format; updated manual to include dynamic range for each model.
August, 2016
| v
Table of Contents
Notes, Cautions, and Warnings .............................................. vii
Safety Information ................................................................... vii
Introduction ................................................................................ 9
ETS-Lindgren Product Information Bulletin ................................................. 10
Maintenance ............................................................................. 11
Maintenance Recommendations ................................................................. 11
Annual Calibration ....................................................................................... 11
Replacement and Optional Parts ................................................................. 12
Upgrade Policies .......................................................................................... 13
Service Procedures ..................................................................................... 13
Specifications ........................................................................... 15
General Specifications ................................................................................. 15
H-Field Probes ............................................................................................. 16
HI-4433-HCH ............................................................................................... 16
HI-4433-LFH ................................................................................................ 16
HI-4433-CH .................................................................................................. 17
E-Field Probe Specifications ........................................................................ 17
HI-4433-STE ................................................................................................ 17
HI-4433-GRE ............................................................................................... 18
HI-4433-MSE ............................................................................................... 18
Practical Application and Use ................................................ 19
System Theory ............................................................................................. 19
Automatic Zero ............................................................................................ 19
Probe Power Supply .................................................................................... 20
Application Considerations .......................................................................... 20
HI-4416 Hand Held Readout ............................................................... 21
HI-4413P Fiber Optic Interface ............................................................ 22
HI-4413USB ......................................................................................... 22
Operation .................................................................................. 23
Xmit/Rcv ...................................................................................................... 23
ARM/OFF ..................................................................................................... 24
vi |
Charger ........................................................................................................ 24
Assembling the Probe .................................................................................. 26
Appendix A: Warranty ............................................................. 29
Appendix B: EC Declaration of Conformity .......................... 31
Appendix C: Series H-491198-36 Battery Charger for NiCd
Batteries .................................................................................... 33
Safety Precautions ....................................................................................... 33
Introduction .................................................................................................. 33
Battery Life ................................................................................................... 34
Charging a Battery ....................................................................................... 34
Specifications ............................................................................................... 36
Maintenance Recommendations ................................................................. 37
Appendix D: HI-4433 Operating Protocols ............................ 39
| vii
Notes, Cautions, and Warnings
Note: Denotes helpful information intended to provide tips for better
use of the product.
Caution: Denotes a hazard. Failure to follow instructions
could result in minor personal injury and/or property
damage. Included text gives proper procedures.
Warning: Denotes a hazard. Failure to follow instructions
could result in SEVERE personal injury and/or property
damage. Included text gives proper procedures.
Note: See the ETS-Lindgren Product Information Bulletin for safety,
regulatory, and other product marking information.
Safety Information
High Voltage: Indicates presence of hazardous voltage.
Unsafe practice could result in severe personal injury or
death.
viii |
Introduction | 9
Introduction
The HI-4433 series comprises a family of battery-operated broadband radio
frequency (RF) isotropic field probes designed for use in making RF exposure
measurements in the vicinity of broadcast facilities and industrial RF sources.
These probes use optical isolation to minimize field perturbation during
measurements. Each probe in this series has characteristics that optimize its
performance for a particular application. Model number differentiates individual
probes for example HI-4433-XXX, where XXX defines the specific probe.
Each HI-4433 series probe assembly consists of a spherical casing containing
the sensor mounted on one end of a shaft. The other end of the shaft is attached
to an extrusion that houses the electronics. The sensor and electronics housing
operate, and are calibrated, as a unit. Since the appearance and layout of the
electronics housing are the same for all probes in this series, both the probe
assembly and the housing are coded with identical serial numbers to prevent
mismatch.
HI-4433 series probes measure the field strength in each of three axes. They
perform a vector addition calculation on the readings and send the result to the
readout via a fiber optic cable. Data from each axis can be viewed individually, or
can be combined. The measurement capability and frequency response of
HI-4433 series probes depend upon which model is selected. In addition, both
magnetic and electric field probes are available.
10 | Introduction
ETS-Lindgren Product Information Bulletin
See the ETS-Lindgren Product Information Bulletin included with your shipment
for the following:
Warranty information
Safety, regulatory, and other product marking information
Steps to receive your shipment
Steps to return a component for service
ETS-Lindgren calibration service
ETS-Lindgren contact information
Maintenance | 11
Maintenance
Before performing any maintenance, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
WARRANTY
Maintenance of the HI-4433 Broadband
Isotropic Field Probe is limited to external
components such as cables or connectors.
To prevent electrical shock, do not remove
cover.
Warranty may be void if the housing is
opened.
If you have any questions concerning
maintenance, contact ETS-Lindgren
Customer Service.
Maintenance Recommendations
Maintenance of the HI-4433 probe is limited to external components such as
cables or connectors.
Annual Calibration
The HI-4433 probe assembly (sensor and electronics housing) requires an
annual calibration check to verify that they are performing within specifications. See the Product Information Bulletin included with your shipment for information
on ETS-Lindgren calibration services.
12 | Maintenance
Replacement and Optional Parts
Following are the part numbers for ordering replacement or optional parts for the
HI-4433 Broadband Isotropic Field Probe.
Replacement Parts
Part Description Part Number
Battery Pack, 3.6 VDC,
Rechargeable
H-491038
Standard Fast Charger (115/230
Volt)
H-491198-36
Cable, Fiber Optic, Glass, 2 Meter H-491106-xx (where “xx” = length”)
Probe Cover (2 required) H-470286A
Handle Assembly H-491001
HI-4433 User Manual H-600056
Shipping Carton H-390066
Optional Parts
Part Description Part Number
Tripod, Dielectric, HI-4433 H-491009
Fiber Optic/RS232 Interface HI-4413P
Fiber Optic to USB Converter HI-4413USB
ProbeView EMF Analysis Software ProbeView II
Digital Remote Readout HI-4416
Carrying Case H-491083
Maintenance | 13
Upgrade Policies
Periodically, probes are upgraded to enhance functionality. Contact ETS-
Lindgren Customer Service for the upgrade status of your probe.
Service Procedures
For the steps to return a system or system component to ETS-Lindgren for service, see the Product Information Bulletin included with your shipment.
Loading...