ETS-Lindgren HI-3550 User Manual

HIn3550
Magnetic Field Monitor
User's Manual
Copyright
Manual #600053 10/97 $12.50
1993 by Holaday Industries, Inc.
©
Revision Record
Manual #600053
HIn3550 Magnetic Field Monitor
Revision Description Date
--- Release 11/93 A Added CE Label 10/97
TABLE OF CONTENTS
1.0 DESCRIPTION ........................ 1
Introduction ........................... 1
2.0 SPECIFICATIONS ...................... 3
3.0 UNPACKING AND ACCEPTANCE .......... 5
Introduction ........................... 5
Unpacking and Acceptance ................ 5
4.0 MAINTENANCE ....................... 7
User Precautions ....................... 7
Replacing Batteries ...................... 8
Battery Recommendations ................. 10
Maintenance Recommendations ............. 10
Return Procedures ...................... 10
Periodic/Preventive Maintenance ............. 11
5.0 DESCRIPTION OF HInn3550 OPERATION .... 13
Introduction ........................... 13
Measurement Cycle ..................... 13
Measurement Ranges .................... 13
5.1 Using the Integrating (IB) Mode ......... 14
Changing the Measurement Mode ........ 14
5.2 Using the Instantaneous Measurement ()B)
Mode ........................... 15
Peak Hold ........................ 15
Peak Hold Display Range .............. 15
5.3 Changing the Target Magnetic Fields ...... 16
5.4 Using the Stand-by (STBY) Mode ........ 16
5.5 Alarm Function .................... 17
Selecting the Alarm Mode ............. 17
Activating/Deactivating the Alarm ........ 17
Setting the Instantaneous Alarm Value .... 17
Setting the Integrating Alarm Value ...... 19
Changing the Alarm Settings ........... 20
Ending Alarm Set Mode ............... 20
Alarm Indicators by Measurement Mode ... 20
Alarm ........................... 22
Stopping the Integrating Mode Alarm ..... 23
Clearing a Manually Interrupted Alarm ..... 23
Limited Warranty
Holaday Industries, Inc. warrants each model HIn3550 Magnetic Field Monitor to be free from defects in material and workmanship for a period of one year from the date of shipment to the purchaser. This warranty extends to the original purchaser only, and does not apply to the batteries or to any products or parts subject to misuse, neglect, accident, unauthorized service or abnormal conditions of operation.
In the event an instrument covered by this warranty fails, Holaday Industries, Inc. will, without charge, repair and recalibrate the instrument if returned to their factory within one year of the original purchase—provided that Holaday Industries' examination discloses, to its satisfaction, that the product is defective. Holaday Industries, Inc. may, at its option, replace the product in lieu of repair. If the defect was caused by misuse, neglect, accident, unauthorized service or abnormal conditions of operation, repairs will be billed at a nominal cost. In such cases, an estimate will be provided before work is started, if requested by the purchaser.
For warranty service, contact Holaday Industries, Inc. Provide the serial number of the instrument and complete details regarding the failure mode. You will then be given either service information or shipping instructions. Return the instrument to the factory, transportation prepaid. Repairs will be made at the factory and the instrument will be returned to you, transportation prepaid. Holaday Industries, Inc., assumes no responsibility for loss of, or damage to, products in transit.
Warning!
EXTREME CAUTION IS ADVISED WHEN WORKING IN ENVIRONMENTS WHERE HIGH-INTENSITY ELECTROMAGNETIC FIELDS MAY EXIST AND WHERE CONTACT WITH HIGH VOLTAGE OR HIGH CURRENT CIRCUITS OR APPARATUS IS POSSIBLE. ACCIDENTAL CONTACT WITH OBJECTS OR CIRCUITS OPERATING AT HIGH VOLTAGES OR HIGH CURRENTS CAN BE LETHAL! HOLADAY INDUSTRIES, INC. ASSUMES NO LIABILITY FOR ANY DAMAGES OR PERSONAL INJURY WHICH MAY RESULT FROM ACCIDENTS ARISING FROM THE USE OF THIS EQUIPMENT.
HIn3550 Manual Page — 1
1.0 DESCRIPTION
Introduction
The model HIn3550 is a compact, lightweight personal magnetic field monitor employing three-axis detection, enabling accurate measurement regardless of field orientation. This instrument provides both instantaneous and time integrated field measurements. A dual alarm function (one for each measurement mode) is also included.
Holaday Industries, Inc. introduced the model HIn3550 in response to demand from researchers and operators working in the strong DC (static) magnetic fields generated by medical diagnostic devices such as MRI's, nuclear fusion experimental devices, metallurgical processes or superconductive magnets. This monitor measures the instantaneous and time integrated exposures to such fields.
Thank you for selecting our model HIn3550 Magnetic Field Monitor. Before using the instrument, please read this manual thoroughly to obtain optimum test results and keep it handy for reference.
Page — 2 HIn3550 Manual
Figure 1
HInn3550 Magnetic
Field Monitor
HIn3550 Manual Page — 3
2.0 SPECIFICATIONS
Target magnetic field: DC (static) magnetic fields;
ELF(50/60 Hz) AC magnetic fields
Measuring sensor: InAs Hall Effect Sensor,
three-axis (isotropic) response
Measurement range: 0.1 mT to 0.3 T
DC Measurement accuracy:
0.1 mT to 0.5 mT: ± 0.05 mT (24 °C, 50%
RH)
0.5 mT to 0.3 T: ± 10%
Resolution:
0 - 40 mT: 0.01 mT 40 mT - 0.3 T: 0.5 mT
Measurement update period: 3 seconds
Measurement modes: Instantaneous (with peak
hold function) Integrating (mT-hours) Standby (low power consumption)
Alarm function range:
Instantaneous measurement (3 seconds): 0.1 mT to 9999.9 mT Integrating measurement (60 seconds): 1 mT-h to 999 T-h
Page — 4 HIn3550 Manual
Operating Temperature Range:
0 °C to 50 °C
Display: FE Type White liquid crystal
Battery: Size AAA alkaline batteries
(4 required)
Battery Life at 8 hours/day continuous use):
measurement of 40 mT or less: 25 days measurement of 40 mT or more: 15 days
Controls:
Side of case: Field Select Switch (AC,DC)
Mode (STBY-IB-)B)
Rear of Case: LOCK/UNLOCK,
SELECT, SET, CLEAR (SELECT + SET)
Outside dimensions: 13.2 X 7.35 X 2.0 cm
(5.2 X 2.9 X 0.8 in)
Weight: Approximately 150 g (5.3
oz)
Belt Pack: Part #51560038
HIn3550 Manual Page — 5
3.0 UNPACKING AND ACCEPTANCE
Introduction
This section contains information on the procedure for the unpacking and acceptance of the HIn3550.
Unpacking and Acceptance
Step 1. Upon delivery of your order, inspect the
shipping container(s) for evidence of damage. Record any damage on the delivery receipt before signing. In case of concealed damage or loss, retain the packing materials for inspection by the carrier.
Step 2. Remove the instrument from its shipping
container. Save the boxes and any protective packing materials for future use.
Step 3. Check all materials against the packing list to
verify that the equipment received matches that which was ordered. If you find any discrepancies, note them and call Holaday Customer Service for further instructions.
Be sure that you are satisfied with the contents of your order and the condition of your equipment before using the instrument.
Loading...