ETS-Lindgren HI-1710A User Manual

HI-1710A
Microwave Measurement
System
User Manual
ii
ETS-Lindgren Inc. reserves the right to make changes to any product described herein in
Revision
Description
Date
A
Initial Release
May, 2001
B
Added Alarm Out pin definitions; rebranding
June, 2008
C
Updated Error Analysis table in Specifications
June, 2009 D
Updated for v3.0 firmware release
May, 2012
E
Updated Menu Option Settings table in Filter (Response) command
August, 2014
order to improve function, design, or for any other reason. Nothing contained herein shall constitute ETS-Lindgren Inc. assuming any liability whatsoever arising out of the application or use of any product or circuit described herein. ETS-Lindgren Inc. does not convey any license under its patent rights or the rights of others.
© Copyright 2001–2014 by ETS-Lindgren Inc. All Rights Reserved. No part of this document may be copied by any means without written permission from ETS-Lindgren Inc.
Trademarks used in this document: The ETS-Lindgren logo is a trademark of ETS-Lindgren Inc.; Avery is a registered trademark of Avery Dennison Corporation.
Revision Record
MANUAL,HI-1710A | Part #H-600094, Rev. E
www.ets-lindgren.com
iii
This page intentionally left blank.
iv
Table of Contents
Notes, Cautions, and Warnings .............................................. vii
General Safety Considerations ............................................. viii
1.0 Introduction .......................................................................... 9
HI-1710A Features ..................................................................................... 9
Standard Configuration ............................................................................. 10
Optional External Zero Receptacle ........................................................... 11
ETS-Lindgren Product Information Bulletin ............................................... 11
2.0 Maintenance ....................................................................... 13
Software/Firmware Updates (Version 3.XX Only) ...................................... 13
Replacing the Fuse ................................................................................... 14
Replacement and Optional Parts .............................................................. 15
Annual Calibration .................................................................................... 15
Service Procedures .................................................................................. 15
3.0 Specifications ..................................................................... 17
Electrical Specifications ............................................................................ 17
Physical Specifications ............................................................................. 18
RS-232 Specifications ............................................................................... 18
Calibration ................................................................................................ 19
Power Cable ............................................................................................. 19
Error Analysis ........................................................................................... 20
4.0 Operation ............................................................................ 21
About Zeroing ........................................................................................... 21
Menu Navigation ....................................................................................... 22
Back Panel Controls and Connectors........................................................ 23
ZERO ............................................................................................... 23
ANALOG OUT .................................................................................. 23
ALARM OUT ..................................................................................... 24
RS-232 (Digital Input/Output) ............................................................ 24
Starting the HI-1710A ............................................................................... 25
Using the HI-1710A .................................................................................. 25
Connecting and Using the Zero Receptacle .............................................. 26
www.ets-lindgren.com
v
5.0 Menu Options, Settings, Command Set .......................... 27
Menu Options ........................................................................................... 27
Menu Option Settings and Command Set ................................................. 28
3 OR 4 DIGITS ................................................................................. 28
ABOUT (Version 3.XX Only) ............................................................. 29
ALARM SETPOINT ........................................................................... 30
AUDIO .............................................................................................. 31
AUTO-RANGE .................................................................................. 31
BAUD RATE ................................................................ ..................... 32
CONTRAST ...................................................................................... 33
ECHO ............................................................................................... 33
FILTER (RESPONSE) ...................................................................... 34
PARAMETER SCREEN .................................................................... 35
PEAK HOLD ................................................................ ..................... 36
PRINT MODE ................................................................................... 37
RANGE ............................................................................................. 39
SELF TEST ...................................................................................... 40
STATUS ........................................................................................... 41
STIM ................................................................................................. 44
RESET.............................................................................................. 44
TEST ALARM ................................................................................... 45
VERSION (Firmware) – Version 3.XX Only ....................................... 45
ZERO ............................................................................................... 45
Factory Default Settings ............................................................................ 46
6.0 Theory of Operation ........................................................... 47
Appendix A: Warranty ............................................................. 49
Appendix B: Error Messages .................................................. 51
Self-Test Error Messages ......................................................................... 51
Startup Error Messages ............................................................................ 52
REF C – OUTSIDE LIMIT ................................................................. 52
+5SUP C – OUTSIDE LIMIT ............................................................. 52
+8SUP C – OUTSIDE LIMIT ............................................................. 52
-8SUP C – OUTSIDE LIMIT .............................................................. 52
BIAS C – OUTSIDE LIMIT ................................................................ 53
STIM C – OUTSIDE LIMIT ................................................................ 53
vi
Appendix C: Summary of RS-232 Command Set ................. 55
Appendix D: EC Declaration of Conformity .......................... 57
www.ets-lindgren.com
vii

Notes, Cautions, and Warnings

Note: Denotes helpful information intended to
provide tips for better use of the product.
Caution: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures.
Warning: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures.
See the ETS-Lindgren Product Information Bulletin for safety, regulatory, and other product marking information.
viii

General Safety Considerations

Safety Symbol
Definition
Warning: This is a Safety Class I product
(provided with a protective earthing ground incorporated in the power cord). The mains plus shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited. DO NOT defeat the earth­grounding protection by using an extension cable, power cable, or autotransformer without a protective ground conductor.
Caution: This instrument is shipped with a three-wire power cable, in accordance with international safety standards. When connected to an appropriate power line outlet, this cable grounds the instrument cabinet.
Warning: For indoor use only, do not expose to rain.
www.ets-lindgren.com
Introduction
9
The ETS-Lindgren
Holaday HI-1710A Microwave Measurement System combines the
proven and reliable diode sensing technology of the ETS-Lindgren Holaday Microwave Survey Meters with the digital processing techniques developed for automatic oven scanning.
HI-1710A with
HI-2623 Microwave Oven Probe

1.0 Introduction

HI-1710A Features

HI-1710A Front Panel
Provides time response characteristics, or filters:
raw (Version 3.XX only), fast, and slow. These allow standard
hand scan measurements, as well as the potential for faster scanning
similar to that of automated scanning systems. The unit is shielded with the necessary isolating and attenuating
devices to provide immunity to stray leakage of microwave energy. The circuitry is powered by a regulated supply with normal industrial
transient protection devices.
10
Introduction
www.ets-lindgren.com
The diode detection array of eight hot carrier diodes is housed in the
large end of the plastic probe assembly. This antenna array has the
unique feature of being able to sum microwave electric fields of any
plane perpendicular to the axis of the probe. The antenna lobe is also
very broad, making it easy to use when measuring leakage around an
oven door.
The cone spacer is designed to provide 5-cm spacing (as required by
the CDRH performance standard) from the tip of the probe cone to the
center of the array. The shape of the cone provides minimum
perturbation of the electric field impinging on the diode array.
The probe is attached permanently to the preamplifier assembly by a
9-ft shielded cable. Each detection probe and amplifier assembly is
calibrated as a unit and can be used interchangeably with any
HI-1710A readout. The digital indicator reads to two or three decimal
places from 0 to 9.999 mW/cm2. Temporary over exposure to
microwave fields up to 2000 mW/cm2 will not cause probe burnout or
damage the instrument or the calibration.
Either auto-range or four ranges with full-scale indications of 1, 2, 5,
and 10 mW/cm2 can be selected from the front panel or with a remote
serial port command. Selecting the range determines the full-scale
setting of the bar graph display located below the digital readout for a
continuous analog indication of the measured leakage level. The audio alarm circuit can be set to any level up to 9.99 mW/cm2.
When the level of the microwave leakage exceeds the pre-selected
value the audio alarm will sound if the audio is enabled, alerting the
operator that the chosen level has been exceeded. For more
information about the alarm, see page 30.

Standard Configuration

HI-1710A Microwave Measurement System  HI-2623 Microwave Oven Probe  Power cable
www.ets-lindgren.com
Introduction
11

Optional External Zero Receptacle

The external zero receptacle will help shield the probe from stray RF fields while being zeroed. Additionally, in high-volume testing situations, the zero receptacle can reduce wear on the ZERO button located on the HI-1710A front panel.
The zero receptacle ships with a 10-foot cable.
For ordering information, see Replacement and Optional Parts on page 15. For information on using the zero receptacle, see Connecting and Using the Zero Receptacle on page 26.

ETS-Lindgren Product Information Bulletin

See the ETS-Lindgren Product Information Bulletin included with your shipment for the following:
Warranty information  Safety, regulatory, and other product marking information  Steps to receive your shipment  Steps to return a component for service  ETS-Lindgren calibration service  ETS-Lindgren contact information
12
Introduction
www.ets-lindgren.com
This page intentionally left blank.
www.ets-lindgren.com
Maintenance
13
Before performing any maintenance, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
Maintenance of the HI-1710A is limited to external components such as cables or connectors.
Clean the exterior of the cabinet using a damp cloth and mild cleaner. Always unplug the unit before cleaning.
To prevent electrical shock, do not remove cover.
Warranty may be void if the housing is opened.
If you have any questions concerning maintenance, contact ETS-Lindgren Customer Service.
WARRANTY

2.0 Maintenance

Software/Firmware Updates (Version 3.XX Only)

Periodically you should check for updates to the software/firmware for your HI-1710A Microwave Measurement System.
Go to www.ets-lindgren.com.  On the Resources menu, click Software/Firmware. In the Software/Firmware column, look for HI-1710A files. Save the
zip file to the desired location on your computer. In the same section where the zip file is located, click the link to
download the installation instructions. Follow those instructions to
install the new software/firmware.
14
Maintenance
www.ets-lindgren.com
Disconnect the HI-1710A from power before replacing a fuse.

Replacing the Fuse

If the HI-1710A fails to operate, check for a blown fuse inside the power entry module. A blown fuse must be replaced with the same value and type of fuse, or an unsafe condition may result. Use only 250 Volt, 1.0 Amp, Type T (5 mm x 20 mm) fuses.
To replace a fuse:
1. Two fuses are located in the fuse drawer in the power input module.
Use a screwdriver to open the drawer.
2. The fuse towards the outside of the drawer is the spare. Remove the
spare fuse from the module.
3. Replace the blown fuse with the spare fuse.
4. Slide the fuse drawer back into the module. Make sure that the drawer
snaps securely into its locked position.
www.ets-lindgren.com
Maintenance
15
ETS-Lindgren may substitute a similar part or new part number with the same functionality for another part/part number. Contact ETS-Lindgren for questions about part numbers and ordering parts.
Part Description
Part Number
Microwave Oven Probe
HI-2623
The HI-2623 Microwave Oven Probe may be used interchangeably with the HI-1710A. A calibrated assembly may be used to replace the existing probe, eliminating downtime. Many service problems are related to probe or cable damage and can be corrected without sending the HI-1710A for service.
Spacer Cone KitIncludes 10 cones and 20 shields.
H-540013
To maintain stated accuracy, use only ETS-Lindgren replacement cones.
Zero Receptacle
H-490930

Replacement and Optional Parts

Following are the part numbers for ordering replacement or optional parts for the HI-1710A.

Annual Calibration

See the Product Information Bulletin included with your shipment for information on ETS-Lindgren calibration services.

Service Procedures

The inner case of the HI-1710A is sealed to maintain calibration. Normally, user repair operations are limited to verification of proper input voltage and replacement of the probe/preamplifier assembly.
For the steps to return a system or system component to ETS-Lindgren for service, see the Product Information Bulletin included with your shipment.
16
Maintenance
www.ets-lindgren.com
This page intentionally left blank.
www.ets-lindgren.com
Specifications
17
Ranges:
Four ranges:
0–1, 0–2, 0–5, and 0–10 mW/cm2
Response Characteristics:
0 to 90% of final value for step input:
F0, Raw (Version 3.XX only)
unfiltered
F1, Slowless than three
seconds
F2, Fastless than one second
Maximum Power Density:
2.0 Watts/cm2 (2000 mW/cm2)
Power Input:
100–240 VAC
0.8–0.4 A max 50/60 Hz
Fuses:
250 Volt, 1.0 Amp, Type T (5 mm x 20 mm)
Control Inputs:
ZeroDry switch closure to zero
the HI-1710A RS-232—Standard 9-pin, 2-way
communications input
Control Outputs:
Alarm OutputForm C contact;
24 V 1 Amp Analog Output—4 VDC max;
1 mA max RS-232— Standard 9-pin, 2-way
communications output

3.0 Specifications

Electrical Specifications

18
Specifications
www.ets-lindgren.com
Environmental:
Operating Temperature:
15°C–40°C Humidity: 5% to 95% relative
humidity, non-condensing
Dimensions:
Height: 127 mm (5.0 in) Width: 311 mm (12.25 in) Depth: 254 mm (10.0 in)
Probe:
Probe Length: 305 mm (12 in) Cable Length: 2.74 m (9 ft)
Weight:
3.6 kg (8 lb)
Connector:
Standard 9-pin, 2-way communications input/output
Communications Protocol:
8 data bits, 1 stop bit, no parity, no flow control, 9600 baud (default)
Pin Designations:
Pin 2: Data received by HI-1710A from computer
Pin 3: Data transmitted from
HI-1710A to computer
Pin 5: Ground

Physical Specifications

RS-232 Specifications

www.ets-lindgren.com
Specifications
19
Calibration:
Calibrated at 2450 MHz for use in the ISM band

Calibration

Calibration is performed at a standard temperature of 75°F in an
anechoic chamber, utilizing a slot radiator.
A CW field is generated using a crystal-controlled solid state source.  The power monitoring system is the HI-2795 Calibration Comparison
System. The calibration field level is established using a reference survey meter
with CDRH calibration correction factors. All ranges are calibrated at mid-scale.

Power Cable

The HI-1710A Microwave Measurement System is shipped with a three-wire power cable. When the cable is connected to an appropriate AC power source, it connects the chassis to earth ground. The type of power cable shipped with each HI-1710A depends on the country of destination.
20
Specifications
www.ets-lindgren.com
1.0 mW/cm2
Absolute Calibration
+0.30 dB
Precision
±0.13
Linearity and AM response
±0.15
Frequency Response*
±0.04
Near field response
±0.29
Polarization*
±0.21
Pattern
+0 / -0.11
Temperature
+0.46 / -0.15
Supply Voltage
±0.01
RFI*
±0.01
Overload
±0.01
Drift*
±0.04
-1.068 / +1.568 dB
* Values are combined in an RMS manner.

Error Analysis

www.ets-lindgren.com
Operation
21
Before connecting any components, follow the safety information in the ETS-Lindgren Product Information Bulletin included with your shipment.
Do not use without spacer cone in place.
RF must not be present in the vicinity of the probe when the zero switch is activated, or it will result in an inaccurate zero.

4.0 Operation

About Zeroing

Zeroing determines the offset value that will be subtracted from all subsequent measurements.
There are three ways to zero a probe connected to the HI-1710A Microwave Measurement System:
Press the ZERO button on the front panel.  Send the serial port command Z.  Attach an external device to the ZERO connector on the back panel.
The zero will drift as the electronic circuitry and diodes warm up to normal operating temperature. There will also be small variations as the ambient temperature changes during the day. For these reasons, the zero point must be reset periodically.
In addition, the bias and offset readings are checked against preset internal limits each time the zero switch is activated. If the limits are exceeded, an error message is displayed. Zeroing in a sufficiently high RF field or zeroing with the probe removed will also cause an error. These tests measure the performance characteristics of the probe and serve as failure indicators.
22
Operation
www.ets-lindgren.com
See Menu Options, Settings, and Command Set on page 27.
(Version 3.XX only) Press the ZERO button once on the front of the HI-1710A to go back one level in the menu. Do not to press the
button more than once or you will zero the probe.

Menu Navigation

HI-1710A Front Panel
1. To display the menu, press any arrow button on the front of the
HI-1710A.
2. Press the down arrow button to move the highlighted selection down
one item at a time, and press the up arrow button to move the
highlighted selection up one item at a time.
3. With the item you want to change highlighted, press the left/right arrow
button to move to the settings for that item.
4. Press the up or down arrow buttons to scroll through the available
settings for the highlighted item.
5. Highlight the desired setting and then release the buttons. After
approximately three to four seconds the display will return to the main
screen with the new/highlighted settings in effect.
www.ets-lindgren.com
Operation
23
The resistance of the external load connected to the analog output receptacle should not be less than 5000 ohms.

Back Panel Controls and Connectors

HI-1710A Back Panel

ZERO

Can be used to attach an external device to the ZERO connector. See About Zeroing on page 21 for more information on the zero function.

ANALOG OUT

This analog output BNC connector is a DC voltage varying from 0 to 4 volts proportional to the full scale range setting. For example, with the instrument on range 2 (0–2 mW/cm2), a leakage value of 2 mW/cm2 will have an analog signal of 4 VDC and a 1 mW/cm2 value will have a signal of 2 VDC. On range 4 (0–10 mW/cm2), the 4 VDC analog output signal will indicate 10 mW/cm2.
24
Operation
www.ets-lindgren.com
Pin
Definition
1
NC 2 NC 3 Normal
4
Common
5
Alarm

ALARM OUT

RS-232 (DIGITAL INPUT/OUTPUT)

This two-way, standard 9-pin RS-232 serial data port is used to connect external devices. These external devices can monitor and/or control the HI-1710A.
All functions normally accessed by the touch switch inputs can be controlled externally and all data normally displayed can be transmitted through the serial port. This is accomplished using a simple command structure of ASCII character codes and a controller such as a personal computer or a video display terminal (VDT). See Appendix C on page 55 for a summary of commands.
The default mode of the HI-1710A generates a formatted test summary. When on, the default conditions are 8 data bits, 1 stop bit, no parity, and no flow control. The HI-1710A is set up at the factory for 9600 baud. This makes it possible to connect a monitoring device such as a VDT or printer.
The baud rate of the serial port can be changed either through the HI-1710A menu options or by an external command through the RS-232 serial port.
www.ets-lindgren.com
Operation
25
Check the probe spacer cone for damage, discoloration, or contamination. Accuracy may be affected by cone wear or damage or by contamination by metallic particles which may become imbedded in the cone. If necessary, replace the cone. See Replacement and Optional Parts on page 15 for part numbers.

Starting the HI-1710A

1. With the power off, attach the HI-2623 Microwave Oven Probe to the
probe connector on the front of the HI-1710A. Insert the connector so
that the locating screw is facing upward.
2. Connect the HI-1710A to AC power.
3. Turn on the power switch on the front of the HI-1710A.
The screen will display the firmware version and additional
ETS-Lindgren information. During this time a series of self-tests are
performed; when the self-tests are complete, the probe is ready to be
zeroed.
4. Place the probe in a zero field environment, and then press ZERO on
the front of the HI-1710A.
When zeroing is complete, the HI-1710A is ready for operation.

Using the HI-1710A

1. Allow the HI-1710A to warm up for at least 10 minutes.
2. Select the operating settings for the test needs. For normal tests or a
beginning point, use the default conditions. To select settings, see
Menu Options, Settings, and Command Set on page 27.
3. Zero the probe before beginning the test. With the oven operating with
a suitable load, move the probe across the oven surfaces, keeping the
probe axis (handle) perpendicular to the potential leak area.
Hold the probe by the red handle to avoid potential interference. Keep
other parts of the body and the probe cord away from the immediate
vicinity of the probe head.
26
Operation
www.ets-lindgren.com
The zero receptacle should be located in an area that is relatively free of significant RF fields. The fields near the zero receptacle should be checked periodically to confirm that no significant stray fields are present.
The analog bar graph display may be used to determine areas of highest leakage. In the normal mode (not Peak Hold) the bar graph display will track the displayed digital value and respond to the filtered RF signal. When the instrument is operated in the Peak Hold mode, the bar graph display is not filtered and responds very rapidly to changes in RF intensity. The Peak Hold mode may be used while scanning to capture the area of highest leakage, while the bar graph will continuously display instantaneous variations in leakage.
Changing the filter selection to FAST will decrease the response time of the system. This will more effectively show the effect of rapid variations in RF levels due to mode stirrers or other modulating devices. In both the SLOW and FAST filter positions, the response time meets the CDRH requirement for RF measuring instrument response times while the SLOW response characteristic provides maximum smoothing of RF variations.
After the test is completed and the RF reading recorded, zero the HI-1710A again before the next test. The zero point may gradually drift over long periods. Re-zero before beginning a test.

Connecting and Using the Zero Receptacle

1. Attach the cable to the ZERO BNC connector located on the back of
the HI-1710A.
2. Attach the other end of the cable to the connector on the side of the
zero receptacle.
The zero receptacle is most often used with Peak Hold mode. When scanning an oven in Peak Hold, the highest reading is displayed on the HI-1710A. When the probe is inserted into the zero receptacle, the RF reading is sent to the RS-232 serial port for automatic data collection and the probe is zeroed.
For manual data recording, the RF reading must be recorded before the probe is inserted into the zero receptacle.
www.ets-lindgren.com
Menu Options, Settings, Command Set
27
Use the arrow buttons on the front of the HI-1710A to navigate the menu. See Menu Navigation on page 22 for a description of the arrow buttons.
Menu Option & Serial Port Command
Serial Port Command Only
3 OR 4 DIGITSsee page 28
AUTO-RANGEsee page 31 **
ABOUTsee page 29
** Auto-Range is also a setting
ALARM SETPOINT—see page 30
in the RANGE menu option;
AUDIO—see page 31
see page 39.
BAUD RATEsee page 32
ECHOsee page 33
CONTRASTsee page 33
STATUSsee page 41
FILTER (RESPONSE)see page 34
VERSION (firmware)see page 45
PARAMETER SCREENsee page 35
ZEROsee page 45
PEAK HOLDsee page 36
PRINT MODEsee page 37
RANGEsee page 39
RESETsee page 44
SELF TESTsee page 40
STIMsee page 44
TEST ALARMsee page 45

5.0 Menu Options, Settings, Command Set

Menu Options

Following are the menu options and command set for the HI-1710A Microwave Measurement System; each is described in the following pages.
28
Menu Options, Settings, Command Set
www.ets-lindgren.com
All serial port commands must be completed with a carriage return (CR).
Menu Option
Setting
Serial Port Command
3
D3 4 D4 — D?

Menu Option Settings and Command Set

Following is a description of each menu option and the available settings. The majority of the menu options have an equivalent serial port command that can be used to set or retrieve a status; in a few cases, there is no menu option, and only the serial port command can be used.

3 OR 4 DIGITS

Specifies whether 3 or 4 digits will display; includes digit to the left of
the decimal point and either 2 or 3 digits to the right of the decimal
point.
All numeric commands return a total of 3 or 4 digits.
In 3-digit mode, the fourth digit is rounded and not displayed.
In 4-digit mode, some least significant bit noise will be seen, resulting
in the least significant digit changing.
www.ets-lindgren.com
Menu Options, Settings, Command Set
29
Menu Option
Setting
Serial Port Command
OFF — ON — —
*IDN?

ABOUT (VERSION 3.XX ONLY)

Displays/returns the installed firmware version and other ETS-Lindgren
and system information. When ON is selected, the information will
display for approximately 10 seconds.
30
Menu Options, Settings, Command Set
www.ets-lindgren.com
The alarm will sound only if the audio is turned on. To turn on the alarm, see Audio on page 31.
Menu Option
Setting
Serial Port Command
XX.XX
T#### — T?
Where #### is a value from 0 to 9999 in mW/cm2. For example, 5000 is 5.0 mW/cm2, 5 is 0.005 mW/cm2, and so on.

ALARM SETPOINT

Sets a level after which an alarm will sound to indicate that the
measured leakage has exceeded the level.
Each press of the up/down arrow button will raise/lower the setting
Any time a change in direction is made up or down, the change
by 0.01 for the first ten presses (9.99, 9.98, 9.97, and so on), by 0.10 for the next ten presses (9.89, 9.79, 9.69, and so on), and by 1.00 for the remaining presses (8.89, 7.89, 6.89, and so on).
increment begins at 0.01.
The alarm will sound only when the displayed RF value exceeds the
alarm set point. For example, if the alarm is set at 4.99 mW/cm2, the
alarm will only sound when the displayed RF level is 5.00 or greater.
Once the RF level decreases below the alarm set point, the alarm will
turn off.
www.ets-lindgren.com
Menu Options, Settings, Command Set
31
Although the alarm set point may be specified, the alarm will not sound unless the audio is turned on. For set point information, see page 30.
Menu Option
Setting
Serial Port Command
OFF
Q0
ON
Q1 — Q?
Menu Option
Setting
Serial Port Command
Function
A0
OFF — A1
ON — A?
Retrieve status
The auto-range setting only affects the bar graph display and the analog output.
When in Peak Hold with the range set to AUTO, the range will change as the field strength increases; it will remain in the range that shows the highest peak value until the ZERO button is pressed. This is so the peak value can be displayed on the bar graph. When the ZERO button is pressed with no field present, the range will revert back to the lowest (1 mW/cm2).

AUDIO

AUTO-RANGE

Turns on the audible alarm indicator, which will sound to indicate that
the measured leakage has exceeded a preset level (alarm set point).
When set to auto-range, the best range for each measuring condition
will automatically be selected.
32
Menu Options, Settings, Command Set
www.ets-lindgren.com
Menu Option
Setting
Serial Port Command
57600
B57
19200
B19
9600
B96
4800
B48
2400
B24
1200
B12
600
B60
300
B30 — B?

BAUD RATE

Sets the communication rate between the HI-1710A and an
external device.
After the baud rate is changed on the HI-1710A, it also must be
changed on the external device. When baud rates are incorrectly set, a video display terminal (VDT) may display nonsense or non-printing characters.
The baud rate is changed immediately when a B command is sent.
The baud rate setting is saved in memory when the power is turned off.
www.ets-lindgren.com
Menu Options, Settings, Command Set
33
Menu Option
Setting
Serial Port
Command
0–8
C0–8 — C?
Menu Option
Setting
Serial Port
Command
OFF
E0
ON
E1 — E?

CONTRAST

ECHO

Increase the value to darken the display or decrease it to lighten the
display.
When echo is on, keyboard entries on the VDT will also be displayed
on the VDT screen.
The echo mode can be used when marginal transmission conditions
exist or when the HI-1710A is connected to a full duplex terminal that requires the characters be re-transmitted.
34
Menu Options, Settings, Command Set
www.ets-lindgren.com
Menu Option
Setting
Serial Port Command
RAW (Version 3.XX only)
F0
SLOW
F1
FAST
F2 — F?

FILTER (RESPONSE)

Digital filtering, or response times, changes the response to
rapidly-changing RF or leakage levels; see Theory of Operation on
page 47 for more information.
RAW (Version 3.XX only)Unfiltered.
SLOWEquivalent to slow response mode on analog meters. SLOW
is the default setting, and is used for most hand-scanning of microwave ovens. It responds more slowly to rapidly-changing RF or leakage levels and tends to average the response to peak levels caused by oven mode stirring devices. This is designed to meet the CDRH requirement that an instrument respond to a step CW microwave input to 90% of its steady state reading within a maximum of three seconds, while still providing maximum smoothing of RF peaks.
FASTEquivalent to fast response mode on analog meters. It is used
to observe rapid variations in RF leakage and to evaluate the effect of stirrer modulation on leakage.
www.ets-lindgren.com
Menu Options, Settings, Command Set
35
Change the setting to ON, and exit the menu to display the parameter screen. Press the ZERO button to escape the parameter screen.
Menu Option
Setting
Serial Port Command
OFF — ON

PARAMETER SCREEN

Displays system board voltages and probe voltages and offsets.
Sample Parameter Screen
RFProbe RF level currently detected.
BIASProbe bias reading; value should be between 0.30 and
0.40 volts.
REFPower supply reference voltage; value should be between 1.22
and 1.26 volts.
+5SUPPower supply +5 voltage; value should be between 4.75 and
5.25 volts.
+8SUPPower supply +8 voltage; value should be between 7.60 and
8.40 volts.
-8SUPPower supply -8 voltage; value should be between -7.92 and
-8.08 volts.
OFSTProbe offset reading; value should be between 0.02 and
0.40 mW (0.005 and 0.100 volts).
36
Menu Options, Settings, Command Set
www.ets-lindgren.com
If an error occurs in any of the parameter readings, the parameter screen displays and the erroneous reading flashes. Turn the HI-1710A off, correct the problem, and then turn it on again.
Menu Option
Setting
Serial Port Command
ON
P1
OFF
P0 — P?

PEAK HOLD

STIMApplies and displays a DC voltage to the input of the RF signal
circuitry; value should be between 2.44 and 4.88. Readings outside this range indicate a fault in the HI-2623 probe/preamplifier.
TESTVoltage at the analog output jack; value should be between
3.80 and 4.20 volts.
Captures and holds the highest leakage reading displayed.
While in Peak Hold the bar graph on the display shows the current
instantaneous RF reading, without filtering. A bar graph segment remains lit to indicate the relative location of the peak reading. The bar graph display responds very rapidly and will show stirrer/turntable modulation as well as minima and maxima of leakage around the microwave oven periphery.
Because of the fast, unfiltered, response of the bar graph display, the
digital peak reading may lag behind the bar graph indication, depending on the filter factor chosen. When the instantaneous RF indication on the bar graph display is greater than the reading on the digital display, the bar graph segment proportional to the digital reading is unlit. In this way the Peak Hold mode of the meter can be used to observe the effects of the various digital filters on fast changing RF inputs.
In the normal mode (not Peak Hold), the bar graph display follows the
filtered display reading.
www.ets-lindgren.com
Menu Options, Settings, Command Set
37
See page 42 for the status message formats for modes 0, 1, and 2, and page 43 for mode 3.
Menu Option
Setting
Serial Port Command
0
M0 1 M1 2 M2 3 M3 — M?

PRINT MODE

There are four modes that specify the content and length of the
information provided during testing.
Mode 0Print mode off. Select this mode when no automatic output
from the HI-1710A is needed; for example, when all output is specifically requested by the controller. This mode might be chosen in a fully-automatic test system where the external device requests the measurement, data gathering, and the zeroing of the meter.
Status messages are transmitted in the same form as in mode 2, but no message is transmitted by the HI-1710A when the zero command is received.
Mode 1Prints continuous reading. In this mode the HI-1710A
continually sends the RF reading through the RS-232 serial port. Data is transmitted at a high rate and can only be used at baud rates of 4800 and above; data is sent at a rate of 45 Hz, the same rate at which the probe is sampled. The controlling device must be able to receive a serially-transmitted 4-digit number 45 times a second.
This mode might be used in the laboratory to investigate the effects of different filter constants and scan rates. The output from the HI-1710A is a three- or four-digit ASCII string in the format X.XX or X.XXX.
38
Menu Options, Settings, Command Set
www.ets-lindgren.com
Mode 2Prints summary ticket. This is the default mode. The
HI-1710A sends a pre-formatted summary report through the RS-232 serial port each time the probe is zeroed. This is most useful when using Peak Hold mode. With this configuration, a serial input printer set up to print adhesive-backed labels can be connected to the RS-232 serial port. After scanning an oven, the meter is zeroed. On zeroing, a ticket is printed with the maximum leakage observed as well as several system test parameters.
The output is formatted for a standard 3 1/2-in x 15/16-in label (for example, Avery® label 4013). Line feed characters are transmitted following the formatted data to advance one additional label space with each formatted output. By printing the report on every other label, the labels can be removed from the backing more conveniently.
Mode 3Does not print summary ticket when zeroed. Changes the
output format of the status messages S0–S8. The transmitted summary messages are shortened for easier adaptation to a computer interface program.
The zero command does not transmit a formatted report as in mode 2. In mode 3 the transmitted status messages end with only a carriage return terminator (CR). In modes 0, 1, and 2, the messages end with a carriage return and a line feed (CR/LF).
www.ets-lindgren.com
Menu Options, Settings, Command Set
39
For more information on auto-range, see page 31.
Menu Option
Setting
Serial Port Command
AUTO
A1
0-1mW
R1
0-2mW
R2
0-5mW
R3
0-10mw
R4
R?

RANGE

The range determines the full-scale setting of the bar graph display
located on the front panel digital readout for a continuous analog
indication of the measured leakage level. The four ranges have
full-scale indications of 1, 2, 5, and 10 mW/cm2; the digital display will
always indicate any value between 0 and 9.999 mW/cm2. When set to
auto-range, the best range for each measuring condition will
automatically be selected.
With no RF input, the instrument will shift to the 0–1 mW/cm2 scale.
When the displayed value exceeds 95% of full scale, the next higher
scale is automatically selected.
While the digital reading will always indicate any value up to
9.999 mW/cm2, the bar graph display will change to reflect the correct percentage of the newly-selected range. When the displayed value drops below 30% of full scale, the next lower range is selected.
If the unfiltered RF signal from the probe exceeds 20 mW/cm2 or the
filtered RF signal exceeds 10 mW/cm2, the HI-1710A will indicate an over range condition by sounding a continuous beep. The over range condition will remain until the filtered RF signal drops below 10 mW/cm2.
40
Menu Options, Settings, Command Set
www.ets-lindgren.com
Menu Option
Setting
Serial Port
Command

SELF TEST

ST
SELF TEST
Performs a series of self-tests; when the self-tests are complete, the
probe is ready to be zeroed.
After successful completion of the self-test, the HI-1710A sends
SELF TEST PASSED message, plus three blank lines.
If the self-test fails, the parameter screen displays. Press the
ZERO button to initiate another self-test.
If an error occurs, turn off the HI-1710A, correct the error, and then
turn on the HI-1710A.
www.ets-lindgren.com
Menu Options, Settings, Command Set
41
Menu
Option
Serial Port
Command
Function
S0
Send System Status
S1
Send Alarm Set Point
S2
Send Range Number
S3
Send Filter Number
S4
Send Bias Voltage
S5
Send Offset — S6
Send STIM Value
S7
Send Power Supply Voltages
S8
Send RF Reading*
S9
(Version 3.XX only)
Send Non-Filtered RF Reading*
*(Version 3.XX only) In Peak Hold mode S8 will return peak
RF value. For the current RF reading, use S9.
The format of information returned depends on the current print mode. See page 42 for the status message formats for print modes 0, 1, and 2, and page 43 for print mode 3.
For more information on print modes, see page 37.

STATUS

Checks status for the following:
42
Menu Options, Settings, Command Set
www.ets-lindgren.com
S0
DIGITS=V/AUTO RNG WWW/PEAK HOLD XXX/YY RNG/PRINT MODE Z
Where:
V is 3 or 4 WWW is ON or OFF XXX is ON or OFF YY is IN or OUT Z is 0,1, or 2
S1
ALARM --- X.XX or X.XXX
S2
SCALE --- X
S3
FILTER --- XXXX Where XXXX is RAW (Version 3.XX only), FAST,
or SLOW
S4
BIAS --- X.XX or X.XXX
S5
OFFSET --- X.XX or X.XXX
S6
STIM --- X.XX or X.XXX
S7
REFERENCE --- X.XXX 5 VOLTS --- X.XXX 8 VOLTS --- X.XXX
-8 VOLTS --- X.XXX
S8
RDNG OK --- X.XX or X.XXX The RF value depends on the filter setting.
S9
(Version 3.XX only) RDNG OK --- X.XX or X.XXX The RF value is raw/unfiltered.
Print Modes 0, 1, 2
www.ets-lindgren.com
Menu Options, Settings, Command Set
43
S0:
XXXXX
First digit3 or 4 for number of digits
displayed and transmitted
Second digit0 or 1 for auto-range off
or on
Third digit0 or 1 for Peak Hold off or
on
Fourth digit0 or 1 for in range or
over range
 Fifth digit0, 1, 2, or 3 for operating
mode
S1:
X.XX
Alarm trigger point (X.XXX in four-digit mode)
S2: X Range number 1–4
S3: X 1 = F1 (Slow)
S4:
X.XXX
Bias voltage (X.XXX in four-digit mode)
S5:
X.XXX
Offset; milliwatts (X.XXX in four-digit mode)
S6:
X.XX
STIM test value (X.XXX in four-digit mode)
S7:
X.XXX
Reference voltage level
X.XXX
5 volt power supply level
X.XXX
8 volt power supply level
X.XXX
-8 volt power supply level
S8:
X.XX
RF reading (X.XXX in four-digit mode)
S9:
X.XX
(Version 3.XX only) Raw/unfiltered RF value (X.XXX in
four-digit mode)
Print Mode 3
44
Menu Options, Settings, Command Set
www.ets-lindgren.com
Menu Option
Setting
Serial Port Command
NO — YES
S6
Menu Option
Setting
Serial Port Command
NO — YES
I

STIM

Performs a stimulus routine to verify that the probe is operational.
When activated, the HI-1710A will apply a DC voltage to the input of
the RF signal.
The STIM value should be between 2.44 and 4.88. Readings outside
this range indicate a fault in the HI-2623 probe/preamplifier.

RESET

This will reset the HI-1710A to factory settings and overwrite any
settings you changed. See Factory Default Settings on page 46 for
more information.
In addition to resetting the system to factory defaults, RESET performs
the self-test sequence.
www.ets-lindgren.com
Menu Options, Settings, Command Set
45
Menu Option
Setting
Serial Port Command
ON — OFF
Menu Option
Setting
Serial Port Command
VER?
Menu Option
Setting
Serial Port Command
Z
When the zero command is received and the RS-232 output is in mode 2 (default mode), a formatted output message is sent through the serial port and the probe is zeroed. The formatted message transmits the RF reading only when the front panel Peak Hold mode is ON.

TEST ALARM

Sets or resets the alarm relay located on the back panel of the
HI-1710A.

VERSION (FIRMWARE) VERSION 3.XX ONLY

Returns the version of installed firmware.
Format: X.XX

ZERO

Zeros the probe; performs the same function as pressing the ZERO button on the front panel.
46
Menu Options, Settings, Command Set
www.ets-lindgren.com
All settings are stored in memory. The last settings used will be restored each time the HI-1710A is turned off.
RESET returns the HI-1710 to factory default settings. For more information on RESET, see page 44.
3 or 4 Digits:
3 (two places to the right of the decimal)
Alarm Set Point:
5.0
Audio Alarm:
OFF
Baud Rate:
9600
Display Contrast:
7
Filter (Response):
Slow
Peak Hold:
OFF
Print Mode:
2
Range:
0–1 mW/cm2

Factory Default Settings

When you receive the HI-1710A, it is configured with the following settings; these default settings can be used for normal production line testing.
www.ets-lindgren.com
Theory of Operation
47

6.0 Theory of Operation

The HI-1710A Microwave Measurement System uses a unique diode/dipole antenna sensor that is coupled with digital filtering techniques developed for Automatic Microwave Oven Scanner Systems. A radial array of eight antennas is located perpendicular to the axis of the probe handle. The electric field is detected by hot carrier diodes operating in the square law region. A negative voltage is applied to the cathode connection of the diodes to bias them in an optimum operating region. The bias is adjusted individually for each probe/preamplifier assembly. The diode signals are summed and input to two stages of amplification. The amplifying stages perform analog filtering to minimize external interference at 60 Hz.
The analog output of the preamplifier is connected to one input of a seven channel analog-to-digital converter. The bias input is connected to another channel. Scaled values of the various power supply voltages are connected to the remaining inputs. The digital equivalent of the RF signal is input to the microprocessor. A software-implemented filter is used to condition the RF signal for most effective RF measurement. The filtering in the HI-1710A is accomplished by an integrating software filter. Sets of fixed filter parameters can be selected from the front panel of the HI-1710A.
Selecting filter 1 (F1 / SLOW) results in a digital filter characteristic with a rise time 90% of less than three seconds. This response is equivalent to the SLOW response of the common analog microwave survey meter.
Filter 2 (F2 / FAST) results in a digital filter characteristic with a rise time 90% of less than one second. The response of this filter is much faster and is equivalent to the FAST response of an analog survey meter.
After filtering, the final leakage value is displayed as a 3- or 4-digit value. The microprocessor also performs the peak hold and alarm functions. In the Peak Hold mode, the bar graph display is driven without any digital filtering to display the real time variations in the RF level. In the Peak Hold mode, the highest reading sent to the digital display by the processor is held and continuously displayed until either a higher reading is transmitted or until the probe is zeroed. The peak value displayed is the value obtained after all filtering is accomplished.
48
Theory of Operation
www.ets-lindgren.com
When not in the Peak Hold mode, the bar graph display responds to the filtered signal as displayed.
The system performs a series of self-tests at startup and each time the probe is zeroed. The bias voltage is measured to check for defective diodes and for a proper bias supply, and then the offset voltage of the array is measured. The offset voltage is the signal at the preamp output with the probe in a zero RF field. Limits on the offset voltage check for proper probe/preamplifier operation. A zero chamber option is available to minimize the RF present at the probe head during zeroing; see Zero Receptacle Option on page 11. A stimulus voltage is applied to the input of the first operational amplifier stage in the preamp. This signal is used to test the integrity of the RF signal path. The level of each of the four power supply outputs is tested on startup. If an error is detected, the parameter screen will be displayed with the erroneous value flashing. Upon zeroing, the self-test is repeated until all errors are corrected.
www.ets-lindgren.com
Warranty
49
See the Product Information Bulletin included with your shipment for the complete ETS-Lindgren warranty for your HI-1710A Microwave Measurement System.
Product Warranted
Duration of Warranty Period
HI-1710A Microwave Measurement System
1 Year

Appendix A: Warranty

DURATION OF WARRANTIES FOR HI-1710A
All product warranties, except the warranty of title, and all remedies for warranty failures are limited to one year.
50
Warranty
www.ets-lindgren.com
This page intentionally left blank.
www.ets-lindgren.com
Error Messages
51
BIAS
OUTSIDE LIMIT
REF
OUTSIDE LIMIT
+5SUP
OUTSIDE LIMIT
+8SUP
OUTSIDE LIMIT
-8SUP
OUTSIDE LIMIT
STIM
OUTSIDE LIMIT

Appendix B: Error Messages

In any mode, an invalid command will produce the error message
ENTRY ERROR -- PLEASE RETRY.

Self-Test Error Messages

The following error messages will be transmitted if during the self-test an out of specification condition is detected.
52
Error Messages
www.ets-lindgren.com

Startup Error Messages

As part of the self-test performed on startup, a number of error messages can be displayed indicating operating problems. In addition, the offset and bias values of the probe are measured each time the probe is zeroed. If an erroneous reading is detected, the HI-1710A Microwave Measurement System will display the parameter screen and the erroneous reading will flash. A message will also be sent to the RS-232 port. Following is the action to be taken for each error.

REF C OUTSIDE LIMIT

1.24-volt supply low  1.24-volt supply high: The 1.24 volt reference to the A/D converter is
beyond specification. Return for service if error message persists.

+5SUP C OUTSIDE LIMIT

5-volt supply low  5-volt supply high: The 5-volt logic supply is beyond specification.
Voltage is not user-adjustable; return for repair.

+8SUP C OUTSIDE LIMIT

8-volt supply low  8-volt supply high: The 8-volt supply to the analog circuitry is beyond
specification. Voltage is not user-adjustable; return for repair.

-8SUP C OUTSIDE LIMIT

-8-volt supply low  -8-volt supply high: The negative 8-volt supply to the analog circuitry is
beyond specification. Voltage is not user-adjustable; return for repair.
www.ets-lindgren.com
Error Messages
53

BIAS C OUTSIDE LIMIT

Bias voltage low  Bias voltage high: The probe bias voltage is beyond specification.
Check the negative 8-volt supply level. If the supply voltage is within
specification, the probe/preamplifier assembly should be replaced. If
the bias voltage is still beyond specification, return for service.

STIM C OUTSIDE LIMIT

STIM test low  STIM test high: A fault exists in the RF signal processing path. Replace
the probe/preamplifier assembly; if the error persists, return for service.
54
Error Messages
www.ets-lindgren.com
This page intentionally left blank.
www.ets-lindgren.com
Summary of RS-232 Command Set
55
For a list of factory settings, see page 46.
Function
Command
Range
Default
3 or 4 digit display
D#
3–4
3
Alarm trigger point
T####
0–9999
5000
Audio mode (1=ON)
Q#
0–1
0
Auto-range (1=ON)
A#
0–1
0
Baud rate
B##
12–96
96
Echo (1=ON)
E#
0–1
0
Filter (Response) number (0 is available only in
Version 3.XX)
F#
0–2 1
Re-initialize system (Reset)
I — —
Peak Hold (1=ON)
P#
0–1
0
Print mode
M#
0–3
2
Range
R#
1–4
1
Self-test
ST — —
Zero meter
Z — —

Appendix C: Summary of RS-232 Command Set

56
Summary of RS-232 Command Set
www.ets-lindgren.com
This page intentionally left blank.
www.ets-lindgren.com
EC Declaration of Conformity
57

Appendix D: EC Declaration of Conformity

Loading...