Escort 3136A User manual

O
S
E
E
E
S
S
O
O
T
T
T
50000 COUNT
DUAL DISPLAY MULTIMETER
Oppeerraattiioonnaall
O
Maannuuaall
M
Section 1. Introduction
1-1 Introducing the 50,000 Count Dual Display Multimeter 1 1-2 Features 2 1-3 Options and Accessories 3 1-4 How to use this manual 4 1-5 Safety 5
Section 2. Getting Started
2-1 Introduction 7 2-2 Getting Started 7 2-3 Turning the Meter ON 11 2-4 Selecting Current Input Terminals and 11
Measurement Range 2-5 Using the Pushbuttons 11 2-6 Basic Measurement Examples 12 2-7 Rack Mounting 18

Table of Contents

Section 3. Front Panel Operation
3-1 Introduction 19 3-2 Front Panel Operations 19 3-3 Primary and Secondary Displays 20 3-4 Input Terminals 22 3-5 Initialization of Measurement Conditions 23 3-6 Selecting A Measurement Function 23 3-7 Selecting Measurement Range 26 3-8 Selecting Secondary Display 27 3-9 Entering Setup Mode 29 3-10 Selecting Local Operation Mode 31 3-11 Operating Arithmetic Functions 31 3-12 Selecting Trigger Mode 38
Section 4. Measurement Applications Examples
4-1 Introduction 41 4-2 Applications for Using Dual Display 41
4-2-1 Dual Display Operation Examples 42 4-3 Measuring Resistance 45 4-4 Measuring True RMS AC+DC 46
I
Section 5. Calibrating the Meter
5-1 Introduction 47 5-2 Environment Condition 47 5-3 Warm up 47 5-4 Recommended Test Equipment 48
Section 6. RS-232 Remote Operation
6-1 Introduction 49 6-2 RS-232 Interface Overview 49 6-3 RS-232 Interface Parameter Set up 49 6-4 Using Commands 51
6-4-1 Types of Commands 51 6-4-2 Command Syntax 52
6-5 Instructions of Command Sets 53
6-5-1 Key Commands 53 6-5-2 Set Commands 54 6-5-3 Query Commands 59
6-6 Remote Program Examples Using RS-232 interface 65
Section 7. GPIB Remote Operation
7-1 Introduction 71 7-2 Description of the GPIB 71 7-3 GPIB Interface Parameters Setup 71 7-4 Commands Summary 72
7-4-1 Overview of Command Type and Formats 72 7-4-2 Response Message Data Types 73 7-4-3 Status Reporting 73
7-5 Instructions of Command Sets 77
7-5-1 IEEE 488.2 Common Commands 77 7-5-2 SCPI Commands 79 7-5-3 SCPI Commands Summary 87
7-6 Remote Program Examples Using GPIB interface 89
Appendices
A. Specifications 93 B. Maintenance 101
II
Section 1
Introduction

1-1 Introducing the 50,000 Count Dual Display Multimeter

NOTE
1. This operation manual contains information and warning that must
be followed to ensure user operation safety and to retain the meter
safety condition.
Precaution!
TO ENSURE PERSONAL SAFETY AND TO AVOID DAMAGING THE
METER AND THE EQUIPMENT CONNECTED, READ “GETTING
STARTED” IN SECTION 2-2 BEFORE USING THE METER.
The meter is 50,000 count
bench-top, field service, and system applications with a high
performance/price ratio.
With the RS-232 computer interface (standard), the meter is fully
programmable for use on the RS-232 interface.
With the IEEE-488 computer interface (optional) installed, the meter is fully
programmable for use on IEEE-488.1 interface bus (1978). The meter is also
designed in compliance with supplemental standard IEEE-488.2 (1987).
Dual Display Multi-meter. The meter is designed for
1

1-2 Features

The main features provided by the meter are:
50,000 Count Dual Display
Vacuum-fluorescent Display (VFD)
Low Cost and High Performances
DCV, ACV, DCA, ACA, Frequency, Diode Continuity
DCV Measurement to 1000V, ACV to 750V (Up 1200Vdc, 1000Vac are
measurable)
AC/DC Current Measurements to 10A (Up to 20A is measurable in less
than 20 seconds).
True RMS (AC, AC+DC), 30Hz to 100kHz Measurement Bandwidth.
AC Current Measurement Bandwidth from 30Hz to 20kHz.
Frequency Measurements Up to 500KHz, 0.01 Hz Resolution.
Resistance Measurement Up to 50 M , 10m Resolutions.
dBm measurement with variable reference impedance from 2 to 8000Ω.
Auto or Lock Ranging Relative Calculation.
Auto or Lock Ranging Dynamic Recording (MIN/MAX) with elapsed
time.
Compare (Hi/Lo/Pass) function for quick in-tolerance test.
Percentage function transfers the measuring value to proportional
percentage (%) display.
Fast Electronic and Closed-case calibration.
Data Hold to freeze displayed value.
Refresh Hold for difficult measuring place.
External trigger a one-time measurement to get the result as your needs.
RS 232 Interface.
GPIB Interface (Option).
2

1-3 Options and Accessories

At the moment, one option is available for the meter, which option can be
installed at the factory and a field installable retrofit kit option is also
available:
IEEE-488 interface (Optional) provides full programmability. There are
two types of programming commands: IEEE 488.2 Common Commands
and Standard Commands for Programmable Instruments (SCPI). The
SCPI commands used in this device is conformance with the SCPI
Standard Version 1993.0.
Standard accessories come with the meter are:
Power cord
Protective holsters (Front and Rear)
Operation Manual
Test leads (Tip-type probe)
Available optional accessories are listed as below:
Test leads (Lantern-type probe)
Test leads (Tip-type probe)
Insulation piercing clip
RS-232 PC Link software and cable
IEEE-488 Interface
Rack-mount kit (used for single meter)
3

1-4 How to use this manual

This manual is designed to help the user to get a quick start. Though it is not
necessary to read the entire manual to operate the unit effectively, we recommend
the manual to be read thoroughly in order to use the meter to its full advantages.
First scan the Tables of contents to be familiar with the outline of the manual.
Then read “Getting Started” in Section 2-2. Refer to the appropriate section of the
manual as needed. The contents of each section are summarized below.
Section 1. Introduction
Introducing the general information of features, options, accessories, and
operation manual for the 50,000 count Dual Display Multi-meter.
Section 2. Getting Started
Introducing how to prepare the meter for operation and to start taking basic
front panel operations and measurements quickly.
Section 3. Operating the Meter from the Front Panel
Providing a complete description of each operation, which can be performed by
using the pushbuttons on the front panel. All related information for operations
and functions are grouped together.
Section 4. Measurement Application Examples
Describing how to use the meter in more advanced and sophisticated operations
and applications.
Section 5. Calibrating the Meter
Describing the basic information to calibrate the meter if necessary.
Section 6. RS-232 Remote Operation
Describing how to connect the meter to a terminal or a host computer and operate
the meter via RS-232 interface.
Section 7. GPIB Remote Operation (Option)
Describing how to connect the meter to a terminal or a host computer and operate
the meter via GPIB interface.
Appendices
Appendix A: Specifications Appendix B: Maintenance
4

1-5. SAFETY

This meter has been designed and tested according to EN61010-1 (IEC1010-1), Safety
Requirements for Electronic Measuring Apparatus. This manual contains information and
warns which must be followed to ensure safe operation and retain the meter in safe
condition. Use of this instrument in a manner not specified herein may impair the
protection provided by the equipment. Some common international electrical symbols used
in this manual are shown below Table:
Table 1-1. International Electrical Symbols
AC - Alternating Current
DC - Direct Current
AC and DC - Alternating and Direct Current
Ground
See Explanation In The Manual
Before using the meter, read the following safety information carefully. In this manual,
"WARNING", is reserved for conditions and actions that pose hazard(s) to the user; "CAUTION", is reserved for conditions and actions that may damage your meter.
WARNING
TO AVOID ELECTRICAL SHOCK OR OTHER INJURY:
Be sure the meter is in good operating condition and avoid working alone.
Follow all safety procedures for equipment being tested.
Inspect the test leads for damaged insulation or exposed metal. Check test lead
continuity. Damaged leads should be replaced.
This equipment operates from a power source that does not apply more than 250V
rms between the supply conductors or each supply conductor and ground. A
protective ground connection by way of the grounding conductor in the power cord is
essential for safe operation.
This equipment is grounded through the grounding conductor of the power cord. To
avoid electrical shock, plug the power cord into a properly wired receptacle before
connecting to the equipment input or output terminals.
5
To avoid explosion, do not operate this product in an explosive atmosphere.
To avoid personal injury, do not remove the cover or panel. Refer servicing to
qualified personnel.
Select the proper function for your measurement.
To avoid electrical shock, use caution when working above 60V dc or 30V ac RMS.
Disconnect the live test lead before disconnecting the common test lead.
Disconnect the power and discharge high-voltage capacitors before testing in and
diode.
When making a current measurement, turn the circuit power off before connecting the
meter in the circuit.
To avoid fire hazard, always use a specified fuse.
Use clamp-on probes when measuring circuits exceeding 10 amps.
When servicing the meter, use only the replacement parts specified.
Do not allow meter to be used if it is damaged or if its safety is impaired.
The meter is safety-certified in compliance with EN61010-1 and EN61010-2-31
(IEC1010-1 & IEC1010-2-31) Installation Category ΙΙI 600V and CAT II 1000V
Pollution Degree 2. In order to maintain its insulation properties, please be sure to use
with the standard or compatible test probes.
CE requirement: Under the influence of R.F field according to standard, the supplied
test leads will pick up induced noise. To have better shielding effect, a short-twisted
lead should be used.
6
Getting Started
d

2-1 Introduction

Section 2 describes the front panel operational keys, displays, input
terminals and rear panel of the meter, adjusting handle, explains general
operating features.

2-2 Getting Started

Unpacking and Inspecting the Meter
Carefully remove the meter from its shipping container and inspect it for
possible damage or missing items. If the meter is damaged or something
is missing, contact the place of purchase immediately. Save the container
and packing material in case user has to return the meter.
Section 2
Front Panel
The front panel (shown in Figure 2-1) has three main elements: the input
terminals on the left, the primary/secondary displays, and the
pushbuttons. The pushbuttons are used to select major functions, ranging
operations, and function modifiers. These elements are described in
detail in Section 3.
mA Fuse
Receptacle
Front Panel
Protective Holster
Primary
Display
Secondary
Display
Power
Switch
POWER
Input
Terminals
Function
Buttons
Ranging
Buttons
Arithmetic
Buttons
n
2
Display
Selection
Shift
Button
Figure 2-1. Front Panel
7
Rear Panel
The rear panel (shown in Figure 2-2) contains a line fuse, the power-line
cord connector, an RS-232 interface connector, and a cutout for
IEEE-488 interface (optional) connector.
Line Power
Figure 2-2 illustrates the location of the Line Voltage Selector with
Fuse Holder housing. If user has already done so, plug the line cord
into the connector on the rear of the meter. The meter will operate at
any line voltage between 90Vac and 264Vac when “line voltage
selector” is set properly, and its frequency range is at 50/60Hz. For
operation safety, DO NOT APPLIES a line voltage that exceeds the
range specified to line cord connector on the rear panel of the meter.
Power-Line Cord Connector
IEEE-488 I/F Connector *
RS-232 Connector
Line Voltage Selector with
Fuse Holder Housing
Earth Ground Terminal Screw
16VA 50-60Hz
~LINE VOLTAGE
WARNING
FUSE
Line Voltage Fuse
Selection Table
*Available with IEEE-488 Interface Option only.
Otherwise covered with Plastic Decal.
GPIB
RS-232C
Real Panel
Protective Holster
Figure 2-2. Rear Panel
CAUTION!
BEFORE TURNING THE METER ON, MAKE SURE THE LINE VOLTAGE SELECTOR IS SET TO THE CORRECT POSITION FOR APPLIED LINE VOLTAGE TO THE POWER-LINE CORD CONNECTOR.
8
The “line voltage selector” is settable for 100Vac, 120Vac, 220Vac,
and 240Vac line voltages.
The correct fuse ratings: 250mA fuse for 100Vac or 120Vac is selected,
and 125mA fuse for 220Vac or 240Vac is selected.
Case, Panels and Holsters
To avoid electric shock or injury, do not operate the meter without panels
or case in place.
The meter is provided with special designed anti-slippery protective
holsters on the front and rear panel sides (shown in Figure 2-1 and 2-2).
The holsters provide a protection to both front and rear panels of the meter
as well as its corners. User may stack up one meter on the top of the other
without concerning the slide off of the units (shown in Picture 2-1).
The holsters can be easily removed when install the rack-mounted ears to
the meter in order to mount the meter into a 19-inch standard rack. Refer
to Section 2-7 for Rack Mounting procedures.
Picture 2-1. Stack up the Meters with Holsters
9
Grounding the Meter
The meter is grounded through power cord. To avoid electric shock or
injury, grounding wire in the power line cord must be connected.
Operating in Explosive Atmospheres
The meter does not provide explosion protection for explosive gasses or
arcing components. Do not operate the meter in such circumstances.
Adjusting Handle
For bench-top use, the handle can be adjusted to provide three viewing
angles. For viewing positions, pull the ends out to a hard stop (about 1/4
inch on each side) and rotate it to one of four stop positions (shown in
Figure 2-3).
WARNING
Be sure to put the meter on a table before removing the handle.
To remove the handle, adjust it to the vertical stop position and pull the
ends all the way out.
1. Viewing Position 1
2. Carrying Position
3. Remove Handle Position
4. Viewing Position 2
5. Viewing Position 3
Figure 2-3. Adjusting Handle
10

2-3 Turning the Meter ON

To turn the meter on, press the
lower right of the front panel. When the meter is turned on, the primary and
secondary displays light for about 2 seconds while an internal self-test
running by its digital circuitry. If the
power-up sequence is in progress, all segments and annunciators of the entire
display remain on until another button is pressed. Then the power-up
sequence continues.
After the meter completing its power-up sequence, it resumes the power-up
measurement configuration stored in non-volatile memory. The power-up
default configuration status set at factory is shown in Table 3-2.
Power
button to “IN” position located on the
button is pressed while the
Hold

2-4 Selecting Current Input Terminals and Measurement Range

If current (dc or ac) is being measured in the Auto-ranging mode, with a
signal input on the 500mA terminal, the meter will select the range
500µA~500mA automatically.
If a signal input is applied to the 10A input terminal, the meter will select the
5A or 10A range automatically.

2-5 Using the Pushbuttons

The meter functions and operations can be selected by pressing the
pushbuttons on the front panel select.
A summary of pushbuttons is shown in Figure 2-4.
Pushbuttons can be used in three ways. User can:
Press a single button to select a function or operation.
EXAMPLE:
(Press) to select AC voltage measurement for the primary display.
Press a combination of buttons, one after the other.
EXAMPLE:
then followed by to select dBm calculation.
11
4
POWER
2 3 8 10 7 9 6 5 1
1. Power Switch 6. 2
nd
Display Selection / Escape key
2. Measurement Function Selection 7. Local / Setup key
3. Range Selection 8. Dynamic Recording / Comparator
4. Calibration Mode Button 9. Arithmetic Function Selection
5. Shift / Enter key 10. Hold / Trigger key
Figure 2-4. Front Panel Pushbuttons
Press multiple buttons simultaneously.
EXAMPLE:
and simultaneously to select True RMS AC volts and DC
volts (calculated) on the primary display.
More detail operations are described in Section 3.

2-6 Basic Measurement Examples

This section describes the basic measurement procedures via operations in
front panel. These procedures as follows provide the user who wants to get a
quick start, but does not want to read the entire manual thoroughly. But it is
still recommended to read this manual thoroughly in order to fully utilize all
advantages in the meter.
12
Voltage, Resistance or Frequency Measurements
Press the desired function button and connect the test leads as shown in
Figure 2-5 to measure voltage, resistance, or frequency. The meter will
select the appropriate range in the auto-range mode, and an annunciator
on the display will indicate measurement units.
NOTE
Excessive error may occur when making measurements with 10 to 100 µV resolutions after measuring high voltage up to 1000 volts dc. It requires two minutes before making low-level measurements.
Current Measurements
To measure current, connect the test leads to mA input terminal or 10A
input terminal for measured current above 500mA as shown in Figure 2-6.
Be sure to turn off the power in the circuit to be measured before taking
connection.
Break the circuit on the groundside to minimize the common mode
voltage) to be measured, and place the meter in series at that point.
Turn on power to the circuit, and then read the display. The meter will
select the appropriate range automatically, and an annunciator on the
display will indicate the units of the measurement value shown.
Turn off power to the circuit and disconnect the meter from the tested
circuit.
NOTE
After making a high current measurement using the 10A input, thermal voltages are generated that may create errors when making high-resolution low-level dc measurements of volts, amps, or ohms. It requires ten minutes to allow the thermals to settle out before making low-level measurements in order to obtain the best accuracy.
13
Figure 2-5. Voltage, Resistance or Frequency Measurements
POWER
Figure 2-6. Current or Frequency Measurements
VFD Bright ness Control
Press
and hold, then press or step by step to select the VFD
brightness to darker level (4 steps in this function and factory setting is set at
POWER
highest light level),
14
Overload Alert
The meter has provided an overload alert for voltage and current measurements.
The beeper sounds tones periodically once the measuring value is exceeded the
value shown as below Table 2-1:
Table 2-1. Beeper Responses for overload Alert Points
Measuring Function Start Alerting Value
DC V >1000V AC V >750V DC + AC V >750V DC A >10A AC A >10A DC + AC A >10A
If the input values over above points, the beeper still sounds tones whatever the
beeper has been set to OFF state, or not.
15
Diode Continuity Tests
Press
to select diode continuity function, then connect the test leads
across the diode under test as shown in Figure 2-7 (Reversing the polarity
will reverse-bias the diode).
The diode test measures the forward voltage of a semiconductor junction
at approximately 0.5mA. The beeper generates a single beep tone when
the input voltage drops below +0.7V (approximately 1.4k) and generates
a continuous beep tone when the input voltage drops below +50mV
(approximately 100Ω).
Readings are displayed in the 2.3V range. “OL” is displayed for voltage
above 2.3V. If the diode continuity test is performed, readings are
displayed in 0.1mV resolution on the 2.3000V range.
WARNING!
To avoid damaging to the meter, do not apply a voltage
exceeding 500 V peak between the input terminals.
Figure 2-7. Diode Continuity Test
POWER
16
Resistance/Continuity Tests
In Ohm test, press button momentarily to set continuity function ON. The sign will be lit and lock the range to 500 . Then
connect the test leads and across the tested circuit as shown in Figure
2-8. While testing continuity, the beeper will sound if the resistance is
less than 10
Ω. For other ranges, the beeper will sound if the resistance
falls below the typical values indicated in Table 2-2.
Table 2-2. Beeper Responses in Continuity Test
Measuring range Beeper On
500.00 <10
5.0000 k <100
50.000 k <1 k
500.00 k <10 k
5.0000 M <100 k
50.000 M <1 M
WARNING!
To avoid damaging to the meter, do not apply a voltage
exceeding 500 V peak between the input terminals.
Figure 2-8. /Continuity Test
POWER
17

2-7 Rack Mounting

User can mount the meter into a standard 19-inch rack using RK-01 (for
single meter) Rack Mount Kit. The front and rear protective holsters can be
removed when mount the meter into a rack.
To install RK-01 rack mount kit, refer to following procedures and Figure 2-9
or the instructions provided with it:
1. Adjusting the handle of the meter to its upward vertical stop position
(refer to Figure 2-3) and pull the ends all the way out.
2. Removing two protective holsters out of the front panel and rear panel of
the meter.
3. Installing the rack mount ears onto the left and right hand side of the
meter frame by using four screws provided with RK-01.
4. Paste two blind plates on the handle hole.
5. Mount the meter with RK-01 into the standard 19” rack.
Figure 2-9. Installing the Rack Mount Kit
18
Front Panel Operation

3-1 Introduction

This section provides a complete description of each operation that can be
performed by using the pushbuttons on the front panel.
All related information for operations and functions are grouped together.

3-2 Front Panel Operations

The following operations can be performed from the front panel:
Select a measuring function (Vdc, Vac, Adc, Aac, resistance/continuity,
frequency, and diode continuity test) for the primary display.
Base on primary display press to select the related function for
Section 3
secondary display.
Take a measurement and display a reading.
Select the manual or auto-range mode (AUTO)
Manually select a measuring range for the primary display.
Select function modifier that cause the meter to display relative readings
(REL), minimum or maximum values (MIN MAX) or decibels (dBm and
dB), or to enter the Data Hold mode or Refresh Hold (HOLD) to hold a
reading on the primary display.
Set the dB reference impedance (REFΩ).
Take a measurement and compare (COMP) it against a tolerance range
(Hi, Lo, or Pass).
Take a measurement and percentage (%) display.
Select the brightness for VFD display.
Use the ”editor” to select from option list, to enter a HI-LO range for the
compare mode and percentage mode.
Configure the computer interface (RS-232 or IEEE-488).
Send measurement directly to a printer or terminal through the RS-232
interface (RS-232 print only mode)
These and other front panel operations are described in the remainder of
Section 3.
19

3-3 Primary and Secondary Displays

The meter is 50,000 count, Vacuum-Fluorescent dual display (VFD). This
display shows measuring readings, annunciator, and messages. The
annunciator indicates measuring units and the meter’s operating
configuration.
The dual display allows you to see two properties (e.g. Vac and frequency) of
the input signal you are measuring. The display contains two major parts,
primary display and secondary display (See figure 3-1).
The primary display contains of larger digits and annunciators and is located
on the left side of the dual display. Readings using the relative (REL),
minimum maximum (MIN MAX), data/refresh hold (HOLD), or decibels
(dBm) modifier can be shown on the primary display.
The secondary display contains of a set of smaller digits on the right side of
the dual display. To press
to turn the secondary display on and select
the related function with the primary display. Press cycling to turn the
secondary display off or Press
then followed by to turn off the
secondary display directly and all arithmetic functions will be disabled.
20
2 3 4 5 6 9 10
14
1
7
8
16
11 12 13
15
Primary Display Secondary Display
Primary Display indications: Secondary Display indications:
1. Auto range (AUTO)
2. Data/Refresh Hold mode (HOLD)
3. Dynamic recording (MIN MAX)
4. Relative mode (REL)
5. dBm indicator (dBm)
6. Diode/Continuity mode
7. Primary Numeric data display
st
8. 1
Measurement unit
9. 2nd Numeric data display
nd
10. 2
11. Auto range (AUTO)
12. Remote state (REMOTE)
13. Compare function (COMP)
14. Shift mode (S)
15. Calibration mode (CAL)
16. Trigger mode
Measurement unit
Figure 3-1 Dual Display Illustrations
21

3-4 Input Terminals

The input terminals, shown in Figure 3-2 are located on the left side of the
front panel. The meter is protected against overloads up to the limits shown in
Table 3-1. Exceeding these limits poses a hazard to both the meter and
operator.
Input-High Terminal for V olts, Ohms, Hz, Diode/Continuity Measurements
500µA~500mA Range DC/AC Current Input
Terminal for DC/AC Current Measurement
Common T erminal (COM)
5A/10A Range Current Input Terminal for DC/AC Current Measurement
500µA/5mA/50mA/500mA Fuse & Holder
(1A/250V Fuse Fast)
Figure 3-2 Input Terminals
Table 3-1 Input Protection Limits
Function Input Terminal Maximum Allowable Input
Vdc to COM
Vac, Hz to COM
mA, Hz mA to COM 500mA dc or ac rms 10A, Hz 10A to COM
All functions Any terminal to earth 1000V dc or peak ac
(1)
In Vdc 1000V range, 1200Vdc is readable with audio warning
(2)
In Vac 750V range, 1000Vdc is readable with audio warning
(3)
10A dc or ac rms continuous, and >10A dc or ac rms for 20 seconds maximum
to COM 500V dc or ac rms to COM 500V dc or ac rms
1200V 750V
normal mode, or 1x10
10A
(1)
dc
(2)
ac rms, 1100V peak, 2x107 V-Hz
6
(3)
dc or ac rms
V-Hz common mode
22

3.5 Initialization of Measurement Conditions

Power up default configuration Status:
When turning the meter on, it assumes its power-up configuration. The
power-up configuration set at the factory is shown in Table 3-2.
As configuration data for IEEE-488 address, RS-232 baud rate, data bit,
stop bit, parity, echo and so on are stored in the non-volatile memory, they
are not changed when power is cycled off and on until the configurations
are changed by the user.
Table 3-2 Default configuration Status
Parameters Default Settings Function DCV Range Auto Range Remote/Local Local Data / Refresh Hold OFF Trigger Type Internal Compare mode HI: 10000(10000E+0)
OFF
Percentage (%) Relative mode OFF Dynamic Recording OFF Secondary Display mode OFF CAL mode OFF

3-6 Selecting A Measurement Function

LO: 00000(00000E+0)
Press a function button shown in Figure 3-3, to select a measuring function.
To select ac + dc total true RMS readings, press and or and
, simultaneously.
Press
toggling to turn the secondary display on and select the related
function.
The summary of ranges and scale values are shown in Table 3-3
23
4
POWER
1. Power Switch 6. 2nd Display Selection / Escape key
2 3 8 10 7 9 6 5 1
2. Measurement Function Selection 7. Local / Setup key
3. Range Selection 8. Dynamic Recording / Comparator
4. Calibration Mode Button 9. Arithmetic Function Selection
5. Shift / Enter key 10. Hold / Trigger key
Figure 3-3 Front Panel Pushbuttons
24
Table 3-3 Range Scale Value
Function Range Scale Auto Ranging
, +
, , + , , +
500mV, 5V, 50V, 500V, 1000V 500mV, 5V, 50V, 500V, 750V 500µA, 5mA, 50mA, 500mA 5A, 10A
(3)
(1)
(2)
Hz 500Hz, 5kHz, 50kHz, 500kHz
(4)
.
(1)
In Vdc 1000V range, 1200Vdc is readable with audio warning
(2)
In Vac 750V range, 1000Vdc is readable with audio warning
(3)
10A dc or ac rms continuous, and 20A dc or ac rms for 20 seconds maximum
(with audio warning)
(4)
In order to eliminate the noise interference, which might be induced to the test
leads, it is recommended to use a shielded test cable for measuring resistance above 500K.
500, 5k, 50k, 500k, 5M, 50M
2.3V Fixed range 500, 5k, 50k, 500k, 5M, 50M (Continuity Mode)
More operations of selecting a measurement function are described below:
to select DC voltage measurement
to select AC voltage measurement
to select DC current measurement
to select AC current measurement
to select frequency measurement
to select Diode Continuity measurements
to select resistance or resistance Continuity measurements mode by
toggling
then to select dBm calculation
( and ) simultaneously to select DC+AC RMS volts calculation
( and ) simultaneously to select DC+AC RMS amps calculation
25
3-7 Selecting Measurement Range ( , and )
Ranging operations are performed by using the , and
buttons (see Figure 3-3).
Measuring ranges can be selected automatically by the meter in
“Auto-ranging” or manually operated by the user.
The range setting is synchronous for dual display for current or voltage
measurement. In auto ranging mode, the range setting for both the primary
and secondary display are corresponding to the higher range of two displays.
In manual ranging mode, the range setting for secondary display is following
to the range setting of primary display.
Auto-Ranging
Press
to toggle in and out of manual ranging. When meter is in
auto-range mode, the AUTO annunciator is lit.
In auto-range, the meter selects the next higher range automatically
when a reading is greater than full scale. If no higher range is available,
OL’ (overload) will be displayed on primary or secondary display.
Likewise, the meter will automatically selects a lower range when a
reading is less than approximately 9.5% of the full scale.
That will be another way to set Auto-ranging for Current measurement.
You can press
then to toggle auto-ranging path for mA and
A terminals.
Manual Range
Press
to toggle in and out of manual ranging. The range user is in
when user enters the manual range mode will become the selected range.
In manual range, the meter remains in the selected range regardless of
input.
Press
to back auto-ranging. The range setting is performed both
on readings shown in the primary display and secondary display.
Selecting A Measuring Range
To manually select a range,
to toggle in (and out) of the manual ranging mode, or
or to select higher range or lower range directly.
26
Loading...
+ 78 hidden pages