Installation Proline Promass
26 Endress+Hauser
Promass P, S
DN A B C
[mm] [in] [mm] [in] [mm] [in] [mm] [in]
8 ³⁄₈ 298 11,73 33 1,3 28 1,1
15 ½ 402 15,83 33 1,3 28 1,1
25 1 542 21,34 33 1,3 38 1,5
40 1 ½ 658 25,91 36,5 1,44 56 2,2
50 2 772 30,39 44,1 1,74 75 2,95
Promass I
DN A B C
[mm] [in] [mm] [in] [mm] [in] [mm] [in]
8 8 373 14,69 20 0,79 40 1,57
15 15 409 16,1 20 0,79 40 1,57
15 FB 15 FB 539 21,22 30 1,18 44,5 1,75
25 25 539 21,22 30 1,18 44,5 1,75
25 FB 25 FB 668 26,3 28 1,1 60 2,36
40 40 668 26,3 28 1,1 60 2,36
40 FB 40 FB 780 30,71 35 1,38 80 3,15
50 50 780 30,71 35 1,38 80 3,15
50 FB 50 FB 1 152 45,35 57 2,24 90 3,54
80 80 1 152 45,35 57 2,24 90 3,54
Complete drainability guaranteed (Promass I, P)
When the sensor is installed in a horizontal line, eccentric clamps can be used to ensure
complete drainability. When the system is pitched in a specific direction and at a specific
slope, gravity can be used to achieve complete drainability. The sensor must be mounted in
the correct position to ensure full drainability in the horizontal position. Markings on the
sensor show the correct mounting position to optimize drainability.
Promass I