Endress+Hauser M TW15 Specifications

TI00265T/02/EN/14.13 71226885
Products Solutions Services
Technical Information
Thermowell Omnigrad M TW15
Heavy duty - General purpose Weld-in or flanged process connection
TW15 is designed for use with resistance and thermocouple thermometers in heavy duty applications. Modular configuration according to DIN 43772 (form 4/4F) enable a use in all industrial processes with severe thermal and mechanical stresses.

Your benefits

• TW15 is a barstock thermowell.
• The process connection can either be weld-in or flanged.
• Extension, immersion length and total length can be chosen according to process requirements.
• A wide choice of dimensions, materials and process connections is available.
• Special versions can be manufactured according to customer requirements.

Equipment architecture

TW15

Function and system design

A0017835
Equipment architecture of the Omnigrad M TW15
The Omnigrad M TW15 thermowells are designed according to DIN 43772 and can therefore guarantee a good level of resistance to the most typical and common industrial processes. The thermowell is made from a metal bar with a diameter of 18, 24 or 26 mm (0.71, 0.94 or 1.02 in). The end of the thermowell is conical with a tip diameter of 9 or 12.5 mm (0.35 or 0.49 in). As process connection the TW15 offers weld-in or different flanged versions.
2 Endress+Hauser
TW15
0
5
10
15
20
25
30
35
40
45
100 200 300 400
v (m/s)
A
B
0
4 8 12 16
0
15
30
50
65
80
100
115
130
145
v (ft/s)
U1 (mm)
U1 (in)
50
165 160
0
5
10
15
20
25
30
35
40
45
100 200 300 400
v (m/s)
0
4 8 12 16
0
15
30
50
65
80
100
115
130
145
v (ft/s)
U1 (mm)
U1 (in)

Performance characteristics

Operating conditions Process pressure (static)

Process connection Standard Max. process pressure
Weld-in version - 400 bar (5800 psi)
Flange EN1092-1 or ISO 7005-1 20, 40 or 50 bar depending on the

Permitted flow velocity depending on the immersion length

The highest flow velocity tolerated by the thermometer diminishes with increasing immersion length exposed to the stream of the fluid. In addition it is dependent on the geometry of the thermometer, the process medium, the process temperature and the process pressure. The following figures exemplify the maximum permitted flow velocities in water and superheated steam at a process pressure of 5 MPa (50 bar = 725 PSI).
flange pressure rating PNxx
ANSI B16.5 Depending on the flange pressure
rating 150 or 300 psi
JIS B 2220 Depending on the flange pressure
rating 20K, 25K or 40K
Thermowell with D = 18 mm (0.71 in), U = 65 mm (2.56 in) -----------­Thermowell with D = 24 mm (0.94 in), U = 125 mm (4.9 in) - - - - - -
Endress+Hauser 3
A Medium water at T = 50 °C (122 °F) U1 Immersion length thermowell, material 1.4571
B Medium superheated steam at T = 400 °C (752 °F) v Flow velocity
(316Ti)
A0011123