Endress+Hauser Memocheck Sim CYP03D Operating Instructions Manual

Page 1
0
BA00481C/07/EN/18.18 71416644 2018-09-30
Products Solutions Services
Operating Instructions
Memocheck Sim CYP03D
Testing tool for analysis measuring points
Page 2
Declaration of Conformity Memocheck Sim CYP03D
Declaration of Conformity
2 Endress+Hauser
Page 3
Memocheck Sim CYP03D Table of contents

Table of contents

1 About this document ........... 4
1.1 Warnings ........................... 4
1.2 Symbols used ........................ 4
2 Basic safety instructions ....... 5
2.1 Requirements for personnel ........... 5
2.2 Designated use ...................... 5
2.3 Occupational safety .................. 5
2.4 Operational safety ................... 6
2.5 Product safety ....................... 6
3 Device description .............. 9
3.1 Measuring system .................... 9
3.2 Simulation values ................... 10
4 Incoming acceptance and
product identification ......... 11
4.1 Incoming acceptance ................ 11
4.2 Product identification ................ 11
4.3 Certificates and approvals ............ 12
5 Electrical connection .......... 13
5.1 Connection in hazardous areas ........ 13
5.2 Connecting the simulator ............ 13
11 Repair .......................... 44
11.1 Spare parts ........................ 44
11.2 Return ............................ 44
11.3 Disposal ........................... 44
12 Accessories .................... 45
12.1 Memosens data cable ................ 45
12.2 Storage case ....................... 45
13 Technical data ................. 46
13.1 Environment ....................... 46
13.2 Mechanical construction ............. 47
Index ................................. 49
6 Operation options ............. 14
6.1 Overview .......................... 14
6.2 Structure and function of the operating
menu ............................. 16
7 Commissioning ................ 18
7.1 Switching on the measuring device .... 18
7.2 Setting the operating language ........ 18
7.3 Quick Setup ........................ 18
8 Operation ...................... 20
8.1 Configuring the measuring device ..... 20
8.2 Extended functions .................. 23
9 Diagnosis ...................... 43
10 Maintenance .................. 43
10.1 Cleaning ........................... 43
10.2 Battery replacement ................. 43
10.3 Calibration and qualification .......... 43
Endress+Hauser 3
Page 4
About this document Memocheck Sim CYP03D

1 About this document

1.1 Warnings

Structure of information Meaning
DANGER
L
Causes (/consequences)
If necessary, Consequences of non­compliance (if applicable)
Corrective action
WARNING
L
Causes (/consequences)
If necessary, Consequences of non­compliance (if applicable)
Corrective action
CAUTION
L
Causes (/consequences)
If necessary, Consequences of non­compliance (if applicable)
Corrective action
NOTICE
Cause/situation
If necessary, Consequences of non­compliance (if applicable)
Action/note
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.

1.2 Symbols used

Symbol Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
4 Endress+Hauser
Page 5
Memocheck Sim CYP03D Basic safety instructions

2 Basic safety instructions

2.1 Requirements for personnel

• Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified
activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
• Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried out only directly at the manufacturer's site or by the service organization.

2.2 Designated use

Memocheck Sim CYP03D is a test tool for analysis measuring points. It enables the simulation of user-definable measured values and errors of all sensors incorporating Memosenstechnology.
The main areas of application are:
• Chemicals and process engineering
• Food, pharmaceutical industry and biotechnology
• Water and wastewater treatment
• Hazardous areas
Use of the device for any purpose other than that described, poses a threat to the safety of people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.

2.3 Occupational safety

As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
Endress+Hauser 5
Page 6
Basic safety instructions Memocheck Sim CYP03D

2.4 Operational safety

Before commissioning the entire measuring point:
1. Verify that all connections are correct.
2. Ensure that electrical cables and hose connections are undamaged.
3. Do not operate damaged products, and protect them against unintentional operation.
4. Label damaged products as defective.
During operation:
If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.

2.5 Product safety

2.5.1 State of the art

The product is designed to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. The relevant regulations and European standards have been observed.

2.5.2 Safety instructions for electrical equipment in hazardous areas

The Memosens inductive sensor cable connection system consists of:
• Memocheck Sim (Memosens technology) CYP03D
• Measuring cable CYK10 Memosens is approved for measuring applications in explosive atmospheres in accordance
with:
• IECEx design approval, IECEx BVS 12.0007
• ATEX design approval BVS 12 ATEX E 008 with amendments The EC Declaration of Conformity is an integral part of these Operating Instructions.
The Memocheck Sim CYP03D is powered by three alkaline batteries with a total nominal voltage of 4.5 V.
Use only the following battery types!
Type
Manufacturer Energizer
Type EN91
Designation LR6 (IEC)
U
battery, nominal
Chemical composition Zinc/manganese dioxide (Zn / MnO2)
Do not open the Memocheck Sim CYP03D in hazardous areas.
6 Endress+Hauser
1.5 V
Page 7
Memocheck Sim CYP03D Basic safety instructions
Before commissioning the Memocheck Sim CYP03D, make sure that the battery
compartment is closed securely with a screw. The maximum permissible cable length is 100 m (330 ft).
Pay attention to the requirements for electrical installations in explosive atmospheres
(EN/IEC 60079-14). The conductive protective coating on the device is part of the Ex-related safety concept.
Make sure that there is no damage >4 cm2 to the protective coating.
Hazardous area versions of digital sensors and simulators incorporating Memosens technology are indicated by a red-orange ring on the plug-in head.
IECEx
Connecting the approved digital Memocheck Sim CYP03D sensor simulator to the IECEx­certified, intrinsically safe sensor output circuit of the Liquiline M CM42 (IECEx TUR
11.0007X) transmitter or alternatively to an IECEx-certified, intrinsically safe Memosens sensor output: Use only the following IECEx-certified measuring cables: – CYK10-G*** (IECEx BVS 11.0052X) – or a Memosens measuring cable that has IECEx certification and is identical in terms of
design, appliance technology and function.
The electrical connection must be made according to the wiring diagram.
The transmitter's Memosens input must support the following maximum values. In
particular, the effective inner inductance and the capacitance of the approved, intrinsically safe sensor output may not exceed these values:
1. Entity Parameters
U0 = 5.1 V U0 = 5.04 V
I0 = 130 mA I0 = 80 mA
P0 = 166 mW (linear output curve) P0 = 112 mW (trapezoid output curve)
Ci = 15 μF Ci = 14.1 μF
Li = 95 μΗ Li = 237.2 μΗ
1) Ex-relevant electrical connection parameters
1)
2. Entity Parameters
1)
ATEX
Connecting the approved digital Memocheck Sim CYP03D sensor simulator to the ATEX­certified, intrinsically safe sensor output circuit of the Liquiline M CM42 transmitter or alternatively to an ATEX-certified, intrinsically safe Memosens sensor output: Use only the following ATEX-certified measuring cables: – CYK10-G*** (BVS 04 ATEX E 121 X incl. amendments) – or a Memosens measuring cable that has ATEX certification and is identical in terms of
design, appliance technology and function.
The electrical connection must be made according to the wiring diagram.
Endress+Hauser 7
Page 8
Basic safety instructions Memocheck Sim CYP03D
The transmitter's Memosens input must support the following maximum values. In
particular, the effective inner inductance and the capacitance of the approved, intrinsically safe sensor output may not exceed these values:
1. Entity Parameters
U0 = 5.1 V U0 = 5.04 V
I0 = 130 mA I0 = 80 mA
P0 = 166 mW (linear output curve) P0 = 112 mW (trapezoid output curve)
Ci = 15 μF Ci = 14.1 μF
Li = 95 μΗ Li = 237.2 μΗ
1) Ex-relevant electrical connection parameters
1)
2. Entity Parameters
1)
Temperature classes
Simulator Ambient temperature range T
Memocheck Sim CYP03D-**+** -20 to +50 ˚C (-4 to 122 ˚F) T4
a
Temperature class
If the ambient temperatures shown above are not exceeded no invalid temperatures for the particular temperature class will occur at the simulator.
CSA
Pay attention to the documentation and control drawings for the transmitter.

2.5.3 IT security

We only provide a warranty if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional protection for the device and device data transfer must be implemented by the operators themselves.
8 Endress+Hauser
Page 9
Memocheck Sim CYP03D Device description
3

3 Device description

3.1 Measuring system

Sensor simulation in a complete measuring system consists of:
• Memocheck Sim CYP03D
• Transmitter with Memosens technology, e.g. Liquiline M CM42 or Liquiline CM44x
• CYK10 Memosens data cable
A0025995
 1 Measuring system with Memocheck Sim CYP03D
1 Liquiline CM44x transmitter 2 CYK10 Memosens data cable 3 Memocheck Sim CYP03D
Endress+Hauser 9
Page 10
Device description Memocheck Sim CYP03D

3.2 Simulation values

With the Memocheck Sim CYP03D you can simulate the following data:
• Simulation values
– Main values – Raw values – Temperature
• Parameter
– pH glass (pH glass) – pH glass, SIL sensor (pH glass SIL) – pH ISFET (pH Isfet) – ORP (ORP) – pH + ORP combined sensor (pH + ORP) – Conductivity conductive (Cond c) – conductivity, conductive, 4-pin (Cond c 4-pol) – conductivity, inductiveCond i) – oxygen, amperometric (Oxygen (amp.)) – Oxygen, optical, Memosens (Oxygen (opt.Memo.)) – Oxygen, optical, fixed cable (Oxy. (opt.fixed)) – Chlorine (Chlorine) – TurbidityTurbidity) – Nitrate (Nitrate) – SAC (SAC)
• The main simulation values can be selected as required within the context of the sensor
specification values
• Repeated ramp with any increment
• Error, e. g. glass breakage, alarm and warning
• Calibration values
You can freely configure all the values so that they match your process. The data listed above are displayed on the transmitters.
10 Endress+Hauser
Page 11
Memocheck Sim CYP03D Incoming acceptance and product identification

4 Incoming acceptance and product identification

4.1 Incoming acceptance

1. Verify that the packaging is undamaged.  Notify the supplier of any damage to the packaging.
Keep the damaged packaging until the issue has been resolved.
2. Verify that the contents are undamaged.  Notify the supplier of any damage to the delivery contents.
Keep the damaged goods until the issue has been resolved.
3. Check that the delivery is complete and nothing is missing.  Compare the shipping documents with your order.
4. Pack the product for storage and transportation in such a way that it is protected against impact and moisture.
 The original packaging offers the best protection.
Make sure to comply with the permitted ambient conditions.
If you have any questions, please contact your supplier or your local Sales Center.

4.2 Product identification

4.2.1 Nameplate

The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Input and output values
• Safety information and warnings
Compare the information on the nameplate with the order.

4.2.2 Product page

www.endress.com/cyp03d

4.2.3 Interpreting the order code

The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Go to www.endress.com.
2. Call up the site search (magnifying glass).
Endress+Hauser 11
Page 12
Incoming acceptance and product identification Memocheck Sim CYP03D
3. Enter a valid serial number.
4. Search.  The product structure is displayed in a popup window.
5. Click on the product image in the popup window.  A new window (Device Viewer) opens. All of the information relating to your
device is displayed in this window as well as the product documentation.

4.2.4 Manufacturer address

Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 D-70839 Gerlingen

4.2.5 Scope of delivery Memocheck Sim CYP03D

• Memocheck Sim CYP03D
• Operating Instructions
• 1 quality certificate as ordered
• Cable as per order (optional)
• Case to store CYP03D and cable (optional)
• Certificate of calibration (optional)

4.3 Certificates and approvals

4.3.1  mark
The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EU directives. The manufacturer confirms successful testing of the product by affixing to it the mark.

4.3.2 Ex approvals Memocheck Sim CYP03D

• ATEX II 2G Ex ia IIC T4 Gb
• IECEx Ex ia IIC T4 Gb
• CSA IS NI Cl. I, Div. 1&2, Group A-D
12 Endress+Hauser
Page 13
Memocheck Sim CYP03D Electrical connection
II 2G
II 1G
II 3G
Ex ia
Zone 1
Zone 0
Zone 2
CYK10
CYP03D
not permitted
only with ATEX­Approval
GN/YE
YE
GN
BN
WH
GND
+
Com A
Com B
GY

5 Electrical connection

5.1 Connection in hazardous areas

A0026198-EN
 2 Operation in hazardous areas

5.2 Connecting the simulator

The electrical connection of the sensor's to the transmitter is established using measuring cable CYK10.
 3 Measuring cable CYK10
Endress+Hauser 13
A0024019
Page 14
Operation options Memocheck Sim CYP03D
2
3
4
5
6

6 Operation options

6.1 Overview

6.1.1 Display

1
Menu path and/or device designation
2
Battery status
3
Simulated parameter
4
Assignment of soft keys, e.g. menu
5
Assignment of soft keys, e.g.  and 
6
Sim. main value
A0026099-EN
 4 Display (e.g. simulation mode)

6.1.2 Key functions

ENTER key
• Switches on the device (press for at least 3 seconds)
• Calls up menu when in simulation mode
• Saves (confirms) data entered
• Selects a menu option
• Switches off the device (press for at least 3 seconds)
 or  MINUS key or PLUS key
In setup mode, the MINUS and PLUS keys have the following functions:
• Configuration of parameters and numerical values
• Navigation through menu In simulation mode, the MINUS and PLUS keys have the following
functions: "Run through" of sim. main values, with each values being changed by the amount of the delta value
and Escape function
Press MINUS and PLUS key simultaneously
• Pressing for a short time: takes you up one level in the menu.
• Pressing for a longer time in the main menu: takes you directly to the simulation mode.
14 Endress+Hauser
Page 15
Memocheck Sim CYP03D Operation options
and Reset
Press MINUS and ENTER key simultaneously for longer Saved setups remain intact.
and Factory settings
Press PLUS and ENTER key simultaneously for longer
• This resets the device to the factory settings.
• All the setups saved are deleted.
Endress+Hauser 15
Page 16
Operation options Memocheck Sim CYP03D
pH glass
pH Isfet
ORP Cond c
Oxygen (amp.)
Chlorine
pH glass SIL
Cond i Oxy. (opt.fixed) Turbidity Nitrate SAC
Language Display
Setup
Illumination Contrast
Simulation
Parameter
Memosens
Fixed cable
Sim. main value
Test values Save setup Load setup Advanced
Date
Time
Switch off auto.
Advanc. function
Calibration value Error simulation Ramp (continuous)
Diagnostics
Diagnostic list Order code Serial number Software Hardware Reset Factory settings
Switch off
pH + ORP
Cond c 4-pol
Oxygen (opt.Memo.)
Ramp (gradual)

6.2 Structure and function of the operating menu

6.2.1 Menu structure

16 Endress+Hauser
Page 17
Memocheck Sim CYP03D Operation options
-
-
-

6.2.2 Operating concept

Example of changing values within a menu: defining the start value of a ramp
Maximum and minimum values are displayed in the editor. You can only configure values within these limits.
1. Use the arrow keys to select the digit of the value that you want to change.
2. Press  to change the value.  The digit flashes
3. Press  or  to increase or decrease the value.
4. Confirm entry with .  The "Escape" function ( and  pressed simultaneously) is disabled here in order
to prevent incorrect entries.
5. Select the check mark (arrow key) and press .  The edited value is accepted.
Endress+Hauser 17
Page 18
Commissioning Memocheck Sim CYP03D

7 Commissioning

7.1 Switching on the measuring device

Switch on the device
Press and hold  for at least 3 seconds.
 This loads the last setup that was saved.
Switch off the device
Press and hold  for at least 3 seconds.

7.2 Setting the operating language

Configure language
Available languages (factory settings in bold)
• Deutsch
English
• Français
• Español
• Italiano
1. Press key for Menu .
2. Language Select (using ).
3. Select language, e.g.English.
4. Confirm selection, answer the prompt that follows with .  From now on, you will be guided through the menu in the language of your choice.
Pressing and takes you back to the main menu.

7.3 Quick Setup

Selecting a parameter and configuring test values
1. Under Setup/Simulation select the desired parameter, e. g.ORP (available parameters →  10).
 Confirm entry with .
2. Under Setup/Simulation/Test values , enter the test values (→  21).
3. Press  and  simultaneously.  You are in the simulation mode.
18 Endress+Hauser
Page 19
Memocheck Sim CYP03D Commissioning
You can now simulate the selected parameter using the selected settings.
Once you connect the Memocheck Sim CYP03D to a Memosens transmitter, the Memosens icon appears in the status line of the simulator. It indicates that the simulator is communicating with the transmitter. The simulation symbol shown on the transmitter display indicates that the transmitter is in simulation mode (→ Operating Instructions of the transmitter).
Endress+Hauser 19
Page 20
Operation Memocheck Sim CYP03D

8 Operation

8.1 Configuring the measuring device

8.1.1 Display behavior Possible settings

• Illumination
• Contrast
Path: Menu/Display
Change the illumination or contrast setting for the display using  or .
 Confirm entry with  .

8.1.2 General settings

Path: Menu/Setup/Advanced
Function Configuration options
(factory settings in bold)
Date
Year 09 to 99
10
Month 01 to 12
03
Day 01 to 31
28
Time
Hour 00 to 23
06
Minute 00 to 59
30
Second 00 to 59
21
Switch off auto. 5 to 100 min
20 min
Info
Set current date.
Set current time.
Configure automatic switch-off.
If the simulator is connected to a
transmitter, this function is deactivated. In this case, the simulator does not switch off automatically.
20 Endress+Hauser
Page 21
Memocheck Sim CYP03D Operation

8.1.3 Selecting parameters

In the setup, you configure which parameters are to be simulated.
1. Path: Menu/Setup/Simulation/Parameter.  The parameters currently configured are displayed.
2. Select new parameter: Parameter/Memosens or Fixed cable. Select the desired parameter.
 If the parameter is changed, the following prompt appears:
ESC = cancel  = direct parameter change, the last settings for the current parameter are lost  = save setup, followed by a prompt to specify where to save the setup. Select a free location so that setups already saved are not overwritten.
3. Select desired option.
Under Menu/Setup/Save setup , you can save up to 10 setups under the relevant parameter name. If you want to use a saved setup, select it under the "Setup/Load setup" menu.

8.1.4 Configuring sim. main value and test values

A setup comprises the selected parameter, a simulation main value and test values. You can change the simulation main value in the simulation mode using a user-definable
delta value. The delta value is the increment by which you change the simulation value by pressing  and .
Test values are all other measured values (including raw measured values) of a test setup. Test values are set to a fixed value, which cannot be modified in the simulation mode.
1. Menu/Setup/Simulation/Parameter/Memosens or .../Fixed cable: Select the parameter, e. g. pH glass.
2. Menu/Setup/Simulation/Sim. main value: Select the measured value to be simulated, e. g. pH value.
 A prompt asking if you wish to change the delta value () or not ().
3. Select .  The current delta value is now displayed, e. g. 00.10 pH.
4. Change the current value, e. g. to 00.50 pH.
5. Accept value (use  to select the  beside the value and then ).
Endress+Hauser 21
Page 22
Operation Memocheck Sim CYP03D
6. Menu/Setup/Simulation/Test values: set other measured values, which are shown on the display ( Temperatureonly) or on the transmitter as a fixed value.
7. Change the value to the desired display, e. g. 25.00 °C.  The test value remains at the value set here and cannot be modified in the
simulation mode.
8. Pressing  and  for longer takes you directly to the simulation mode.
The Sim. main value is displayed as the first value in the simulation mode. You can press  or  to change the set delta value. The temperature is displayed as the second value. Pressing  and  does not affect this value. The other test values, which are not visible on the Memocheck display, can only be read off the transmitter or output to transmitter outputs.
Parameter Chlorine
To ensure that the simulator and transmitter display the identical chlorine concentration, the pH value which is used for calculation purposes must be the same for the transmitter and the simulator.
Parameter Oxygen (amp.) or Oxygen (opt.Memo.)
To ensure that the simulator and transmitter display the identical oxygen concentration, the following values used for calculation purposes must be the same for the transmitter and the simulator: Salinity and Process pressure/Altitude.
22 Endress+Hauser
Page 23
Memocheck Sim CYP03D Operation
Sim main values Memosens (factory settings in bold)
pH glass pH glass SIL pH Isfet pH + ORP
pH value
• Temperature
• Raw value
• Raw value temp.
• SCS resistance
ORP Cond c Cond c 4-pol Oxygen (amp.)
ORP potential
• ORP [%]
• Temperature
• Raw value
• Raw value temp.
Oxygen (opt.Memo.) Chlorine
• Partial pressure
• Saturation
Conc. liquid
• Conc. gas.
• Temperature
• Salinity
• Process pressure
• Altitude
• Raw value temp.
pH value
• Temperature
• Raw value
• Raw value temp.
• SCS resistance
Conductivity
• Temperature
• Resistance
• Raw value temp.
• Phase
Chlor. concentr.
• Current
• Temperature
• pH value
• Raw value current
• Raw value temp.
pH value
• Temperature
• Raw value
• Raw value temp.
Conductivity
• Temperature
• Resistance
• Raw value temp.
pH value
• ORP potential
• rH value
• Temperature
Conc. liquid
• Current
• Saturation
• Partial pressure
• Conc. gas.
• Temperature
• Salinity
• Process pressure
• Altitude
• Raw value current
• Raw value temp.
Sim main values Fixed cable (factory settings in bold)
Cond i Oxy. (opt.fixed) Turbidity Nitrate SAC
Conductivity
• Temperature
• Resistance
• Raw value temp.
• Partial pressure
• Saturation
Conc. liquid
• Temperature
• Salinity
• Raw value temp.
• Slope
• Temperature
TU value [FNU]
• TU value [g/l]
• Raw value temp.
Content NO3
• Content NO3-N
• Temperature
• Raw value temp.
• TOC
• CSB
• Temperature
SAC value
• Raw value temp.

8.2 Extended functions

In the Advanc. function menu, you can enter the following values. These values always refer to the last parameters selected under Simulation .
• Calibration value
• Error simulation
• Ramp (continuous)
• Ramp (gradual)
Endress+Hauser 23
Page 24
Operation Memocheck Sim CYP03D
t
Value
Max.
Min.
Value
Max.
Min.
t

8.2.1 Calibration value

The calibration value of the simulated sensor is the value to which the sensor adjustment refers.
If you change a calibration value, there will be a brief interruption to communication between the Memocheck Sim CYP03D and the transmitter to allow the transmitter to accept the calibration settings.
An unfavorable configuration may result in measured values outside the specified measuring range. This can lead to fault states in the transmitter. More information on calibration can be found in the Operating Instructions for your transmitter.

8.2.2 Error simulation

You receive a list of possible errors which you can combine with one another. The error categories in accordance with NAMUR (F, M, C, S) are treated as follows:
• The errors declared as "F" in the transmitter are indicated by a flashing display.
• Errors in other categories are displayed in the transmitter's diagnostic list.

8.2.3 Ramps Ramp (continuous): You define the start and stop value as well as the duration of a ramp

(Starting value, Stop value, Duration). Ramp (gradual): Instead of a total time, you define the number and the duration of the
increments (Number of steps, Time per step).
Ramp (continuous)
The ramp rises continuously without any
Ramp (gradual)
The ramp is gradual.
jumps.
A0017397-EN
• The ramp repeats itself automatically until you stop it.
• The ramp is a simple linear function y = mx + n. The linear element "n" is normally equal to the zero point. The factor "m" is the slope of the line.
• You will obtain a negative slope for the ramp if you set the starting value higher than the stop value.
24 Endress+Hauser
A0017398-EN
Page 25
Memocheck Sim CYP03D Operation
x
y
n
y1
y2
x1
x2
m=
y2-y1 x2-x1
x
y
n
y2
y1
x1
x2
m=
y2-y1 x2-x1
 5 Linear function
n Zero point m Slope y1 Start Value y2 Stop value x Time
x2-x1 = duration
A0017356
 6 Negative slope
n Zero point m Slope y1 Start Value y2 Stop value x Time
x2-x1 = duration
A0017359

8.2.4 Extended functions: pH glass and pH glass SIL

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Temp. offset -10.0 to 10.0 ˚C
pH comp. isoth. 0.0 to 12.0 pH
mV comp. isoth. -300.0 to 300.0 mV
Slope 0.01 to 65.0 mV/pH
Zero point 0.0 to 12.0 pH
(factory settings in bold)
0.0 ˚C
7.0 pH
0.0 mV
59.16 mV/pH
7.0 pH
Endress+Hauser 25
Page 26
Operation Memocheck Sim CYP03D
Path: Menu/Advanc. function
Function Configuration options
Error simulation Glass SCS failure
Ramp Ramp
Sim. main value Starting
pH value -2.0 to 16.0
Temperature -40.0 to
Raw value -750.0 to
Raw value temp. -40.0 to
SCS resistance 0.001 MΩ to
(factory settings in bold)
Temp. sens. defect Glass SCS warning
value
pH
-2.0 pH
150.0 ˚C
-40.0 °C
750.0 mV
-750.0 mV
150.0 ˚C
-40.0 °C
1.0 TΩ
0.001 MΩ
Stop value Duration Number of
-2.0 to 16.0 pH
16.0 pH
-40.0 to
150.0 ˚C
150.0 °C
-750.0 to
750.0 mV
750.0 mV
-40.0 to
150.0 ˚C
150.0 °C
0.001 MΩ to
1.0 TΩ
1.0 TΩ
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s

8.2.5 Extended functions: pH Isfet

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Temp. offset -10.0 to 10.0 ˚C
Slope 0.01 to 65.0 mV/pH
pH comp. isoth. 0.0 to 12.0 pH
mV comp. isoth. -300.0 to 300.0 mV
Asymmetry -300.0 to 300.0 mV
26 Endress+Hauser
(factory settings in bold)
0.0 ˚C
59.16 mV/pH
7.0 pH
0.0 mV
0.0 mV
Page 27
Memocheck Sim CYP03D Operation
Path: Menu/Advanc. function
Function Configuration options
Error simulation Leak. curr. alarm
Ramp Ramp
Sim. main value Starting
pH value -2.0 to 16.0
Temperature -40.0 to
Raw value -750.0 to
Raw value temp. -40.0 to
(factory settings in bold)
Leak. curr. warn Temp. sens. defect Sensor supply
value
pH
-2.0 pH
150.0 ˚C
-40.0 °C
750.0 mV
-750.0 mV
150.0 ˚C
-40.0 °C
Stop value Duration Number of
-2.0 to 16.0 pH
16.0 pH
-40.0 to
150.0 ˚C
150.0 °C
-750.0 to
750.0 mV
750.0 mV
-40.0 to
150.0 ˚C
150.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
Endress+Hauser 27
Page 28
Operation Memocheck Sim CYP03D

8.2.6 Extended functions: pH + ORP

Path: Menu/Advanc. function
Function Configuration options
Calibration value
pH value -200.0 mV
pH comp. isoth. 0.0 to 12.0 pH
mV comp. isoth. -300.0 to 300.0 mV
Slope 0.01 to 65.0 mV/pH
Zero point 0.0 to 12.0 pH
ORP potential
Cal. point 1 [mV] -2.0 to 2.0 V
Cal. point 2 [mV] -2.0 to 2.0 V
Cal. point 1 [%] 0.0 to 100.0 %
Cal. point 2 [%] 0.0 to 100.0 %
ORP % slope -30.0 to 30.0 mV/%
ORP % zero point -1.0 to 1.0 V
rH value 0.0 to 100.0 %
rH offset -300.0 to 300.0 rH
Temperature 0.0 to 100.0 %
Temp. offset -10.0 to 10.0 ˚C
(factory settings in bold)
7.0 pH
0.0 mV
56.12 mV/pH
7.0 pH
-200.0 mV
200.0 mV
10.0 %
30.0 %
20.0 mV/%
-400.0 mV
10.0 %
0.0 rH
30.0 %
0.0 ˚C
28 Endress+Hauser
Page 29
Memocheck Sim CYP03D Operation
Path: Menu/Advanc. function
Function Configuration options
Error simulation Glass SCS failure
Ramp Ramp
Sim. main value Starting
pH value
pH value -2.0 to 16.0
Raw v. C1-C2 (pH) -750.0 to
SCS resistance 0.001 MΩ to
Refer. SCS resist. 0.01 to
ORP potential
ORP potential -2.0 to 2.0 V
ORP [%] 0.0 to
Raw val. -C2 (ORP) -2.0 to 2.0 V
Raw value C1 -3.0 to 3.0 V
rH value
rH value -40.0 to
(factory settings in bold)
Ref. SCS failure Temp. sens. defect Glass SCS warning Ref. SCS warning Counter spillover Meas. value inval.
value
pH
-2.00 pH
750.0 mV
-750.0 mV
1.000 TΩ
0.001 MΩ
60.0 kΩ
0.01 kΩ
-2.0 V
100.0 %
0.0 %
-2.0 V
-3.0 V
50.0 rH
-40.0 rH
Stop value Duration Number of
-2.0 to 16.0 pH
16.0 pH
-750.0 to
750.0 mV
750.0 mV
0.001 MΩ to
1.000 TΩ
1.000 TΩ
-2.0 to 16.0 pH
16.0 pH
-2.0 to 2.0 V
2.0 V
0.0 to
100.0 %
100.0 %
-2.0 to 2.0 V
2.0 V
-3.0 to 3.0 V
3.0 V
-40.0 to
50.0 rH
50.0 rH
Once you have selected an error using
, the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
Endress+Hauser 29
Page 30
Operation Memocheck Sim CYP03D
Path: Menu/Advanc. function
Function Configuration options
Temperature
Temperature -40.0 to
Raw value temp. -40.0 to
(factory settings in bold)
150.0 ˚C
-40.0 °C
150.0 ˚C
-40.0 °C
-40.0 to
150.0 ˚C
150.0 °C
-40.0 to
150.0 ˚C
150.0 °C
10 to 6000 s
60 s
10 to 6000 s
60 s
1 to 200
10
1 to 200
10
0.5 to 600 s
1 s
0.5 to 600 s
1 s

8.2.7 Extended functions: ORP

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Cal. point 1 [mV] -2.0 to 2.0 V
Cal. point 2 [mV] -2.0 to 2.0 V
Cal. point 1 [%] 0.0 to 100.0 %
Cal. point 2 [%] 0.0 to 100.0 %
ORP mV offset -1.0 to 1.0 V
Temp. offset -10.0 to 10.0 ˚C
ORP % slope -30.0 to 30.0 mV/%
ORP % zero point -1.0 to 1.0 V
(factory settings in bold)
-200.0 mV
200.0 mV
10.0 %
30.0 %
0.0 mV
0.0 °C
16.47 mV/%
-833.3 mV
30 Endress+Hauser
Page 31
Memocheck Sim CYP03D Operation
Path: Menu/Advanc. function
Function Configuration options
Error simulation Temp. sens. defect
Ramp Ramp
Sim. main value Starting
ORP potential -2.0 to 2.0 V
ORP [%] 0.0 to
Temperature -40.0 to
Raw value -2.0 to 2.0 V
Raw value temp. -40.0 to
(factory settings in bold)
Sensor supply
value
-2.0 V
100.0 %
0.0 %
150.0 ˚C
-40.0 °C
-2.0 V
150.0 ˚C
-40.0 °C
Stop value Duration Number of
-2.0 to 2.0 V
2.0 V
0.0 to
100.0 %
100.0 %
-40.0 to
150.0 ˚C
150.0 °C
-2.0 to 2.0 V
2.0 V
-40.0 to
150.0 ˚C
150.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s

8.2.8 Extended functions: Cond c, Cond c 4-pol and Cond i

Path: Menu/Advanc. function
Function Configuration options
Calibration value Conductive Toroidal
Cell constant 0.001 E-03 to
Reference value 0.001 mS/cm
Reference temp. 0.0 to 60.0 ˚C
Temp. offset -10.0 to 10.0 ˚C
Temp. gradient -3.0 to 3.0
Endress+Hauser 31
(factory settings in bold)
0.001 E-03 to
-1
10.0 cm
10 E-03 cm
to 1.0 S/cm
0.005 mS/cm
25.58 °C
0.0 ˚C
1.0
10.0 cm
-1
2.9 cm
0.001 mS/cm to 1.0 S/cm
100.0 mS/cm
0.0 to 60.0 ˚C
25.0 °C
-1
-1
Page 32
Operation Memocheck Sim CYP03D
Path: Menu/Advanc. function
Function Configuration options
Error simulation Cond c
Ramp Ramp
Sim. main value Starting
Conductivity 0.001 µS/cm
Temperature -40.0 to
Resistance 0.001 mΩ to
Raw value temp. -40.0 to
(factory settings in bold)
• Polarizat. warn.
• No Cond. display
• Temp. sens. defect
• Sensor supply
Cond c 4-pol
• Polarizat. warn.
• Cond. val. invalid
• Temp. sens. defect
• Broken connector
• Meas. value inval.
• Resist. maximum
Cond i
• Cond. sens. defect
• Cond. val. invalid
• Temp. sens. defect
• Temp. value inval.
• Ind. curr. too high
• Ind. curr. too low
value
to 2000 S/cm
0.001 µS/cm
150.0 ˚C
-40.0 °C
1.0 GΩ
0.001 mΩ
150.0 ˚C
-40.0 °C
Stop value Duration Number of
0.001 µS/cm to 2000 S/cm
2000 S/cm
-40.0 to
150.0 ˚C
150.0 °C
0.001 mΩ to
1.0 GΩ
1.0 GΩ
-40.0 to
150.0 ˚C
150.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
32 Endress+Hauser
Page 33
Memocheck Sim CYP03D Operation

8.2.9 Extended functions: Oxygen (amp.)

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Slope 0.1 pA/hPa to 5.0 nA/hPa
Zero point -3.2 nA to 3.2 nA
Temp. offset -10.0 to 10.0 ˚C
Temp. gradient -3.0 to 3.0
Temp. coeff. 1 20.00 E-03 to 40.00 E-03
Temp. coeff. 2 100.0 E-06 to 500.0 E-06
Temp. coeff. 3 500.0 E-09 to 5.000 E-06
Temp. coeff. 4 1.000 E-09 to 70.00 E-09
(factory settings in bold)
313.5 pA/hPa
0.0 pA
0.0 ˚C
1.0
30.79 E-03
447.6 E-06
4.224 E-06
66.75 E-09
Endress+Hauser 33
Page 34
Operation Memocheck Sim CYP03D
Path: Menu/Advanc. function
Function Configuration options
Error simulation Leak. curr. alarm
Ramp Ramp
Sim. main value Starting
Conc. liquid -0.02 to
Current 0.0 pA to
Saturation -0.02 to
Partial pressure 0.0 to 440.0
Conc. gas. 0.0 to
Temperature -40.0 to 60.0
Salinity 0.0 to 40.0
Process pressure 500 to 9999
Altitude -300 to 4000
Raw value current 0.0 pA to
Raw value temp. -40.0 to 60.0
(factory settings in bold)
Leak. curr. warn Temp. sens. defect Sensor supply
value
120.0 mg/l
-0.02 mg/l
640.0 nA
0.0 nA
1000 % sat
-0.02 % sat
hPa
0.0 hPa
100.0 %
0.0 %
˚C
-40.0 °C
g/kg
0.0 g/kg
hPa
500 hPa
m
-300 m
640.0 nA
0.0 pA
˚C
-40.0 °C
Stop value Duration Number of
-0.02 to
120.0 mg/l
120.0 mg/l
0.0 pA to
640.0 nA
640.0 nA
-0.02 to 1000 % sat
1000 % sat
0.0 to 440.0 hPa
440.0 hPa
0.0 to
100.0 %
100.0 %
-40.0 to 60.0 ˚C
60.0 °C
0.0 to 40.0 g/kg
40.0 g/kg
500 to 9999 hPa
9999 hPa
-300 to 4000 m
4000 m
0.0 pA to
640.0 nA
640.0 nA
-40.0 to 60.0 ˚C
60.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
34 Endress+Hauser
Page 35
Memocheck Sim CYP03D Operation

8.2.10 Extended functions: Oxygen (opt.Memo.)

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Temp. offset -10.0 to 10.0 ˚C
Temp. gradient -3.0 to 3.0
Cal. Quality 0.0 to 100.0 %
(factory settings in bold)
0.0 ˚C
1.0
100.0 %
Endress+Hauser 35
Page 36
Operation Memocheck Sim CYP03D
Path: Menu/Advanc. function
Function Configuration options
Error simulation Measurement stop
Ramp Ramp
Sim. main value Starting
Partial pressure 0.0 to 440.0
Saturation -0.02 to
Conc. liquid -0.02 to
Conc. gas. 0.0 to
Temperature -40.0 to 60.0
Salinity 0.0 to 40.0
Process pressure 500 to 9999
Altitude -300 to 4000
Raw value temp. -40.0 to 60.0
(factory settings in bold)
No ref. cal. P1 No ref. cal. P2 No amplitude Tau too low Tau too high Wave form Temp. sens. defect Temp. out of spec. Electronics temp. Electronic-error
value
hPa
0.0 hPa
1000 % sat
-0.02 % sat
120.0 mg/l
-0.02 mg/l
100.0 %
0.0 %
˚C
-40.0 °C
g/kg
0.0 g/kg
hPa
500 hPa
m
-300 m
˚C
-40.0 °C
Stop value Duration Number of
0.0 to 440.0 hPa
440.0 hPa
-0.02 to 1000 % sat
1000 % sat
-0.02 to
120.0 mg/l
120.0 mg/l
0.0 to
100.0 %
100.0 %
-40.0 to 60.0 ˚C
60.0 °C
0.0 to 40.0 g/kg
40.0 g/kg
500 to 9999 hPa
9999 hPa
-300 to 4000 m
4000 m
-40.0 to 60.0 ˚C
60.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
36 Endress+Hauser
Page 37
Memocheck Sim CYP03D Operation

8.2.11 Extended functions: Oxy. (opt.fixed)

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Slope 0 ... 200 %
Tau -5.0 to 105.0 µs
Temp. offset -10.0 to 10.0 ˚C
Temp. gradient -3.0 to 3.0
(factory settings in bold)
100%
20.0 µs
0.0 ˚C
1.0
Endress+Hauser 37
Page 38
Operation Memocheck Sim CYP03D
Path: Menu/Advanc. function
Function Configuration options
Error simulation Tau too low
Ramp Ramp
Sim. main value Starting
Partial pressure 0.0 to 440.0
Saturation -0.02 to
Conc. liquid -0.02 to
Temperature -40.0 to 60.0
Salinity 0.0 to 40.0
Raw value temp. -40.0 to 60.0
Slope 0.0 to
(factory settings in bold)
Tau too high No signal dropout No amplitude Temp. too low Temp. too high LED voltage No LED current Dynamic error
value
hPa
0.0 hPa
1000 % sat
-0.02 % sat
120.0 mg/l
-0.02 mg/l
˚C
-40.0 °C
g/kg
0.0 g/kg
˚C
-40.0 °C
200.0 %
0.0 %
Stop value Duration Number of
0.0 to 440.0 hPa
440.0 hPa
-0.02 to 1000 % sat
1000 % sat
-0.02 to
120.0 mg/l
120.0 mg/l
-40.0 to 60.0 ˚C
60.0 °C
0.0 to 40.0 g/kg
40.0 g/kg
-40.0 to 60.0 ˚C
60.0 °C
0.0 to
200.0 %
200.0 %
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
38 Endress+Hauser
Page 39
Memocheck Sim CYP03D Operation

8.2.12 Extended functions: Chlorine

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Slope -500.0 to -0.01 nA/g/l
Zero point -100.0 nA to 100.0 nA
Temp. offset -10.0 to 10.0 ˚C
Temp. gradient -3.0 to 3.0
Error simulation Temp. sens. defect
Ramp Ramp
Sim. main value Starting
Chlor. concentr. 0.0 to 200.0
Current -1.5 µA to 0.0
Temperature -40.0 to 60.0
pH value -2.0 to 16.0
Raw value current -1.5 µA to 0.0
Raw value temp. -40.0 to 60.0
(factory settings in bold)
-25.0 nA/g/l
0.0 pA
0.0 ˚C
1.0
Sensor supply
value
mg/l
0.0 mg/l
pA
-1.5 µA
˚C
-40.0 °C
pH
-2.0 pH
pA
-1.5 µA
˚C
-40.0 °C
Stop value Duration Number of
-0.02 to
120.0 mg/l
120.0 mg/l
-1.5 µA to 0.0 pA
0.0 pA
-40.0 to 60.0 ˚C
60.0 °C
-2.0 to 16.0 pH
16.0 pH
-1.5 µA to 0.0 pA
0.0 pA
-40.0 to 60.0 ˚C
60.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
Endress+Hauser 39
Page 40
Operation Memocheck Sim CYP03D

8.2.13 Extended functions: Turbidity

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Temp. offset -10.0 to 10.0 ˚C
Error simulation LED error
Ramp Ramp
Sim. main value Starting
Temperature -40.0 to
TU value [FNU] 0.0 to
TU value [g/l] 0.001 to 1000
Raw value temp. -40.0 to
(factory settings in bold)
0.0 ˚C
Turbidit. too high Electronics test Sensor polluted Meas. val. uncert. No cal. data Temperature err. Meas. value inval. Measurement stop
value
150.0 ˚C
-40.0 °C
10000.0 FNU
0.0 FNU
g/l
0.001 g/l
150.0 ˚C
-40.0 °C
Stop value Duration Number of
-40.0 to
150.0 ˚C
150.0 °C
0.0 to
10000.0 FNU
10000.0 FNU
0.001 to 1000 g/l
1000 g/l
-40.0 to
150.0 ˚C
150.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
40 Endress+Hauser
Page 41
Memocheck Sim CYP03D Operation

8.2.14 Extended functions: Nitrate

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Temp. offset -10.0 to 10.0 ˚C
Error simulation Flash lamp defect
Ramp Ramp
Sim. main value Starting
Content NO3 0.001 mg/l to
Content NO3-N 0.0 µg/l to
Temperature -40.0 to
Raw value temp. -40.0 to
(factory settings in bold)
0.0 ˚C
Turbidit. too high Electronics test Filter change Meas. val. uncert. No cal. data Temperature err. Measurement stop Meas. value inval.
value
550 mg/l
0.001 mg/l
500.0 mg/l
0.0 µg/l
150.0 ˚C
-40.0 °C
150.0 ˚C
-40.0 °C
Stop value Duration Number of
0.001 mg/l to 550 mg/l
550 mg/l
0.0 µg/l to
500.0 mg/l
500.0 mg/l
-40.0 to
150.0 ˚C
150.0 °C
-40.0 to
150.0 ˚C
150.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
Endress+Hauser 41
Page 42
Operation Memocheck Sim CYP03D

8.2.15 Extended functions: SAC

Path: Menu/Advanc. function
Function Configuration options
Calibration value
Temp. offset -10.0 to 10.0 ˚C
Error simulation Flash lamp defect
Ramp Ramp
Sim. main value Starting
TOC 0.000 mg/l to
CSB 0.000 mg/l to
Temperature -40.0 to
SAC value 0.0 to
Raw value temp. -40.0 to
(factory settings in bold)
0.0 ˚C
Turbidit. too high Electronics test Filter change Meas. val. uncert. No cal. data Temperature err. Measurement stop
value
100.0 g/l
0.000 mg/l
100.0 g/l
0.000 mg/l
150.0 ˚C
-40.0 °C
100.0 E03
-1
m
-1
0.0 m
150.0 ˚C
-40.0 °C
Stop value Duration Number of
0.000 mg/l to
100.0 g/l
100.0 g/l
0.000 mg/l to
100.0 g/l
100.0 g/l
-40.0 to
150.0 ˚C
150.0 °C
0.0 to
100.0 E03
-1
m
100.0 E03
-1
m
-40.0 to
150.0 ˚C
150.0 °C
Once you have selected an error using ,
the error is output at the transmitter. To delete the error again, remove the check mark again using .
(continuous)
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
10 to 6000 s
60 s
Ramp (gradual)
steps
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
1 to 200
10
Time per step
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
0.5 to 600 s
1 s
42 Endress+Hauser
Page 43
Memocheck Sim CYP03D Diagnosis

9 Diagnosis

Path: Menu/Diagnostics
Function Info
Diagnostic list The diagnostic messages provide information to
Order code CYP03D-xxx
Serial number xxxxxxxxxxx
Software x.xx.xx
Hardware x.xx.xx
Reset Device is restarted. Your stored settings will be
Factory settings All device settings are reset to the factory settings.
Endress+Hauser service personnel.
preserved.
All the setups saved are deleted.

10 Maintenance

10.1 Cleaning

Clean the device using only commercially available, mild, household cleaning agents.

10.2 Battery replacement

The battery compartment is accessed from behind. Permitted battery types: →  6.
Only ever open the battery compartment in the non-hazardous area!

10.3 Calibration and qualification

The Memocheck Sim CYP03D can, with the quality or calibration certificate, also be used as a qualification tool for your measuring point.
The quality and calibration certificates can be renewed: For such services, you will need to return Memocheck Sim CYP03D to Endress+Hauser. In the case of requalification, the device is tested fully and a new quality certificate is issued. In the case of recalibration, in addition to requalification the device is also incorporated into a calibration procedure. A quality certificate and a certificate of calibration are issued. The recommended testing interval is 1 year.
Endress+Hauser 43
Page 44
Repair Memocheck Sim CYP03D

11 Repair

11.1 Spare parts

Battery compartment cover
Order No. 71138380

11.2 Return

The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
To ensure the swift, safe and professional return of the device:
Refer to the website www.endress.com/support/return-material for information on the
procedure and conditions for returning devices.

11.3 Disposal

The device contains electronic components. and must therefore be disposed of in accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
Always dispose of batteries in accordance with local regulations on battery disposal.
44 Endress+Hauser
Page 45
Memocheck Sim CYP03D Accessories

12 Accessories

The following are the most important accessories available at the time this documentation was issued.
For accessories not listed here, please contact your Service or Sales Center.

12.1 Memosens data cable

Order No. Memosens data cable CYK10 (optional)
71128718 CYK10-A032 + adapter, cable ends; Non-Ex
71128721 CYK10-G032 + adapter; only for CYP03D, Ex
To connect Memocheck Sim CYP03D to transmitters with M12 sockets and Pg couplings, you require the Memosens data cable CYK10 supplied. The cable is always supplied with an adapter piece so that it fits both M12 sockets and Pg couplings. If you wish to simulate fixed cable sensors (turbidity, nitrate, toroidal conductivity, oxygen optical) with Memocheck Sim CYP03D, you need this cable. When using sensors with an inductive Memosens plug-in head (pH/ORP; oxygen, conductive conductivity, chlorine), the appropriate cable is already included in the measuring point.

12.2 Storage case

Order No.
71183327 Ex
In the hazardous area, the Memocheck simulator case should only be opened to remove or put back the Memocheck simulator. When opened, the case should never be exposed to process­related intensive electrostatic charges.
Endress+Hauser 45
Case for Memocheck Sim CYP03D
Page 46
Technical data Memocheck Sim CYP03D

13 Technical data

13.1 Environment

13.1.1 Ambient temperature range

–20 to 50 °C (–4 to 120 °F)

13.1.2 Storage temperature

-20 to 55 °C (-4 to 130 °F)

13.1.3 Degree of protection

IP55

13.1.4 Electromagnetic compatibility

Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry
46 Endress+Hauser
Page 47
Memocheck Sim CYP03D Technical data
40.0 (1.57)
25.0 (0.98)
58.0 (2.28)
94.0 (3.7)
36.0 (1.42)
200.0 (7.87)

13.2 Mechanical construction

13.2.1 Dimensions

 7 Memocheck Sim CYP03D
All dimensions in mm (inch)
A0026005

13.2.2 Weight (incl. batteries)

0.3 kg (0.7 lbs)

13.2.3 Materials

Housing:
Endress+Hauser 47
ABS (UL 94 HB)
Page 48
Technical data Memocheck Sim CYP03D

13.2.4 Batteries

For the Memocheck Sim CYP03D, use only the following battery types, as only these are covered by the Ex approval:
• Energizer, EN91 (AA, 1.5 V, LR6 as per IEC), x 3
• Battery storage temperature: -20 to 35 ˚C (-4 to 95 ˚F)
48 Endress+Hauser
Page 49
Memocheck Sim CYP03D Index

Index

Symbols
 mark ....................... 12
A
Accessories
Memosens data cable ............ 45
Storage case ..................45
Adapting the device ................ 20
Ambient temperature range ........... 46
B
Batteries .......................48
Battery replacement ................43
C
Calibration and qualification ........... 43
Calibration value .................. 24
Certificates and approvals .............12
Cleaning .......................43
Commissioning ...................18
Connection ..................... 13
D
Declaration of Conformity ............. 2
Degree of protection ................46
Designated use ....................5
Device description .................. 9
Diagnosis ...................... 43
Dimensions ..................... 47
Display ........................14
Display behavior .................. 20
Disposal ....................... 44
E
Electrical connection ................13
Electromagnetic compatibility .......... 46
Error simulation .................. 24
Ex approvals .................... 12
Extended functions ................ 23
Chlorine .................... 39
Conductivity ..................31
Nitrate .....................41
ORP ...................... 30
Oxy. (opt.fixed) ................37
Oxygen (amp.) ................ 33
Oxygen (opt.Memo.) .............35
pH + ORP ................... 28
pH glass .................... 25
pH glass SIL ..................25
pH Isfet .................... 26
SAC .......................42
Turbidity ....................40
I
Incoming acceptance ............... 11
K
Keys ......................... 14
M
Maintenance .................... 43
Manufacturer address ...............12
Materials ...................... 47
Measuring system .................. 9
Mechanical construction ............. 47
Menu structure ...................16
N
Nameplate ..................... 11
O
Occupational safety ................. 5
Operating concept ................. 17
Operating elements ................ 14
Operating language ................ 18
Operating menu .................. 16
Operation ...................... 20
Operation options ................. 14
Operational safety ..................6
Order code ......................11
P
Personnel .......................5
Product identification ............... 11
Product page .................... 11
Product safety .................... 6
Q
Quick Setup ..................... 18
Endress+Hauser 49
Page 50
Index Memocheck Sim CYP03D
R
Ramp (continuous) .................24
Ramp (gradual) ...................24
Repair ........................ 44
Requirements for personnel ............ 5
Return ........................ 44
S
Safety instructions
Hazardous areas ................ 6
Operational safety ...............6
Requirements for personnel ......... 5
Scope of delivery .................. 12
Settings
Extended functions ........... 23, 25
General .................... 20
Operating language ............. 18
Quick Setup .................. 18
Selecting parameters ............ 21
Sim. main value and test values ...... 21
Simulation values ................. 10
Spare parts ..................... 44
State of the art .................... 6
Storage temperature ................46
Switching on .................... 18
Symbols ........................ 4
T
Technical data ................... 46
U
Use ...........................5
W
Warnings ....................... 4
Weight ........................47
50 Endress+Hauser