If necessary, Consequences of
non-compliance (if applicable)
Corrective action
‣
WARNING
L
Causes (/consequences)
If necessary, Consequences of
non-compliance (if applicable)
Corrective action
‣
CAUTION
L
Causes (/consequences)
If necessary, Consequences of
non-compliance (if applicable)
Corrective action
‣
NOTICE
Cause/situation
If necessary, Consequences of
non-compliance (if applicable)
Action/note
‣
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious
injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to
property.
1.2 Symbols
SymbolMeaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
1.3 Symbols on the device
SymbolMeaning
Reference to device documentation
Endress+Hauser5
About this documentLiquiport CSP44
1.4 Documentation
The following manuals which are available on the complement these Brief Operating
Instructions Operating Instructions:
• Brief Operating Instructions for Liquiport CSP44, BA00465C
• Operating Instructions for Memosens, BA01245C
• Software description for Memosens inputs
• Calibration of Memosens sensors
• Sensor-specific diagnostics and troubleshooting
• Guidelines for communication via fieldbus and web server
• Special Documentation: Sampler application manual SD01068C
• Documentation on other devices in the Liquiline platform:
• Liquiline CM44xR (DIN rail device)
• Liquiline System CA80 (analyzer)
• Liquiline System CAT8x0 (sample preparation)
• Liquistation CSFxx (sampler)
• Liquiport CSP44 (sampler)
6Endress+Hauser
Liquiport CSP44Basic safety instructions
2 Basic safety instructions
2.1 Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may
be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the
specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions
and must follow the instructions contained therein.
• Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried out only
directly at the manufacturer's site or by the service organization.
2.2 Designated use
Liquiport 2010 CSP44 is a portable sampler for liquid media in non-hazardous areas. The
samples are taken discontinuously using a peristaltic pump and are then distributed into
sampling containers.
The sampler is designed for use in the following applications:
• Communal and industrial wastewater treatment plants
• Laboratories and water management offices
• Monitoring of liquid media in industrial processes
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted. The
manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Workplace safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable international standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
Endress+Hauser7
Basic safety instructionsLiquiport CSP44
2.4 Operational safety
Before commissioning the entire measuring point:
1.Verify that all connections are correct.
2.Ensure that electrical cables and hose connections are undamaged.
3.Do not operate damaged products, and protect them against unintentional operation.
4.Label damaged products as defective.
During operation:
If faults cannot be rectified:
‣
products must be taken out of service and protected against unintentional operation.
8Endress+Hauser
Liquiport CSP44Basic safety instructions
2.5 Product safety
2.5.1 State of the art
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
international standards have been observed.
Devices connected to the sampler must comply with the applicable safety standards.
2.5.2 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
Endress+Hauser9
Product descriptionLiquiport CSP44
1
2
3
4
5
6
7
8
9
10
11
12
13
3 Product description
3.1 Device design
A complete sampling unit comprises:
• Controller with display, soft keys and navigator
• Vacuum or peristaltic pump for sampling
• PE or glass sample bottles for sample preservation
• Sampling chamber temperature regulator (optional) for safe sample storage
• Suction line with suction head
1Device cover
2Battery compartment cover
with switch
3Upper carrying handles
4Unit upper compartment
5Peristaltic pump with pump
Secure the sampler against unintentional start-up whilst you work on the opened hose
‣
pump.
10Endress+Hauser
Liquiport CSP44Product description
3.2 Terminal diagram
The unique terminal name is derived from:
Slot no. : Port no. : Terminal
Example, NO contact of a relay
Device with inputs for digital sensors, 4 current outputs and 4 relays
• Base module BASE2-E (contains 2 sensor inputs, 2 current outputs)
• 2AO module (2 current outputs)
• 4R module (4 relays)
Endress+Hauser11
Incoming acceptance and product identificationLiquiport CSP44
4 Incoming acceptance and product
identification
4.1 Incoming acceptance
1.Verify that the packaging is undamaged.
Notify the supplier of any damage to the packaging.
Keep the damaged packaging until the issue has been resolved.
2.Verify that the contents are undamaged.
Notify the supplier of any damage to the delivery contents.
Keep the damaged goods until the issue has been resolved.
3.Check that the delivery is complete and nothing is missing.
Compare the shipping documents with your order.
4.Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
Make sure to comply with the permitted ambient conditions.
If you have any questions, please contact your supplier or your local Sales Center.
4.2 Product identification
Nameplates can be found:
• On the inside of the door
• On the packaging (adhesive label, portrait format)
• On the inside of the device cover
4.2.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Firmware version
• Ambient and process conditions
• Input and output values
• Activation codes
• Safety information and warnings
• Certificate information
Compare the information on the nameplate with the order.
‣
4.3 Scope of delivery
The scope of delivery comprises:
• 1 Liquiport 2010 CSP44 with:
• The ordered bottle configuration
• Optional hardware
• 1 print version of Brief Operating Instructions in the language ordered
• Optional accessories
If you have any queries:
‣
Please contact your supplier or local sales center.
12Endress+Hauser
Liquiport CSP44Incoming acceptance and product identification
4.4 Certificates and approvals
4.4.1
Declaration of Conformity
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EU directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
MCERTS
The device has been assessed by Sira Certification Service and complies with "MCERTS
Performance Standards for Water Monitoring Equipment Part 1, Version 2.1 dated
November 2009"; certificate no.: Sira MC100176/02.
EAC
The product has been certified according to guidelines TP TC 004/2011 and TP TC
020/2011 which apply in the European Economic Area (EEA). The EAC conformity mark
is affixed to the product.
mark
Endress+Hauser13
InstallationLiquiport CSP44
480 (18.9)
675 (26.6)
320 (12.6)
355 (14.0)
5 Installation
5.1 Installation conditions
5.1.1 Dimensions
A0013473
1CSP44 standard version, dimensions in mm (in)
14Endress+Hauser
Liquiport CSP44Installation
5.1.2 Installation site
A0013474
2Installation site, example
The suction line must be routed with a downward slope to the sampling point. Avoid
siphon formation!
Note the following when erecting the device:
• Erect the device on a level surface.
• Securely connect the device at the fastening points to the surface underneath.
• Protect the device against additional heating (e.g. heater or direct sunlight in the case of
PS housing).
• Protect the device against mechanical vibrations.
• Protect the device against strong magnetic fields.
5.1.3 Connection for suctioning samples
• Maximum suction height: 8 m (26 ft)
• Maximum hose length: 30 m (98 ft)
• Diameter of hose connection: 10 mm (3/8")) internal diameter
• Intake speed:
> 0.5 m/s (> 1.6 ft/s) as per EN 25667, ISO 5667
> 0.6 m/s (> 1.9 ft/s) as per Ö 5893, US EPA
Note the following when erecting the device:
• Always route the suction line so that it slopes upwards from the sampling point to the
sampler.
• The sampler must be located above the sampling point.
• Avoid siphoning effects in the suction line.
Endress+Hauser15
InstallationLiquiport CSP44
Requirements for the sampling point:
• Do not connect the suction line to pressurized systems.
• Use the suction filter to impede coarse and abrasive solids and solids which can cause
clogging.
• Immerse the suction line in the direction of flow.
• Take the sample at a representative point (turbulent flow, not directly at the bottom of
the channel).
Useful sampling accessories
Suction filter:
Impedes coarser solids and solids which can cause clogging.
5.1.4 Connection for sample intake on version with sample pump
• Maximum suction height: 8 m (26 ft)
• Maximum hose length: 30 m (98 ft)
• Diameter of hose connection: 10 mm (3/8")) internal diameter
• Intake speed:
> 0.5 m/s (> 1.6 ft/s) as per EN 25667, ISO 5667
> 0.6 m/s (> 1.9 ft/s) as per Ö 5893, US EPA
Note the following when erecting the device:
• Always route the suction line so that it slopes upwards from the sampling point to the
sampler.
• The sampler must be located above the sampling point.
• Avoid siphoning effects in the suction line.
Requirements for the sampling point:
• Do not connect the suction line to pressurized systems.
• Use the suction filter to impede coarse and abrasive solids and solids which can cause
clogging.
• Immerse the suction line in the direction of flow.
• Take the sample at a representative point (turbulent flow, not directly at the bottom of
the channel).
Useful sampling accessories
Suction filter:
Impedes coarser solids and solids which can cause clogging.
5.2 Installation
5.3 Connecting the suction line
1.When installing the device, take the installation conditions into account.
2.Open the device cover at the front fastening clasp.
3.Route the suction line from the sampling point to the device.
4.Screw the suction line onto the device's hose connection.
5.4 Post-installation check
1.Verify that the suction line is securely connected to the device.
2.Visually check that the suction line is installed correctly from the sampling point to
the device.
3.Verify that the rotating arm is correctly engaged.
16Endress+Hauser
Liquiport CSP44Electrical connection
1
2
3
4
5
OFFON
6 Electrical connection
6.1 Connecting the sampler
WARNING
L
Device is live!
Incorrect connection may result in injury or death!
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions
‣
and must follow the instructions contained therein.
Prior to commencing connection work, ensure that no voltage is present on any cable.
‣
3Electrical connections of the controller
1Connection socket for charger
2Socket for M12 sensor connector (optional)
3Socket for M12 sensor connector (optional)
4Connection socket for signal cable (optional)
5Service interface
The polarity of the switch connections does not have to be taken into account.
6.1.1 Charging the battery
NOTICE
Defective batteries
The batteries can be destroyed if they are discharged completely.
To prevent complete discharge, move the switch to the "OFF" position.
‣
4Switch position
A0029150
A0035816
Endress+Hauser17
Electrical connectionLiquiport CSP44
Charge the battery before initial commissioning. It takes approx. 5 hours to fully charge
the battery. Please refer to the charger's operating manual for detailed information on the
charger.
Connect the device with the mains plug to the supply voltage.
‣
The battery starts charging as soon as the power unit is connected, regardless of
the switch position.
Only replace batteries with the following battery type: Panasonic LC-R127R2PG1.
Connecting the charger when the batteries are installed
The mains plug of the charger must be easily accessible so that the charger can be easily
disconnected from the power supply.
Connect the battery charger to the connection socket (item 1). If the battery is not fully
‣
charged, it is recharged by the charger.
Only use the chargers specified by the manufacturer. → 136
Connecting the charger when the batteries are removed
If you are charging batteries that have been removed, you require the adapter cable
(accessory no.: 71111882) to connect to the charger.
6.1.2 Removing the cover
WARNING
L
Device is live
Incorrect connection may result in injury or death
If a power unit or charger is connected, disconnect it from the power supply.
‣
1.Release both securing screws.
2.Remove the cover of the battery compartment.
3.Remove the old batteries and release the plug-in connections.
4.Connect the new batteries (pay attention to the battery polarity).
5.Insert the new batteries and secure the battery compartment cover.
A0035817
18Endress+Hauser
Liquiport CSP44Electrical connection
1
2
6.2 Connecting modules and sensors
6.2.1 Connecting the sensors
Sensor connection
A0028664
5Connection sockets for sensors
1Socket for M12 sensor connector (= channel 1 for version with one sensor)
2Socket for M12 sensor connector (= channel 2 for version with two sensors)
6.3 Terminal assignment for input/output signals
Input signals
• 2 analog signals 0/4 to 20 mA (optional)
• 2 binary signals > 100 ms pulse width or edge (optional)
Signals of digital sensors with Memosens protocol (optional)
Output signals
2 binary signals > 1 s pulse width or edge (optional)
2 current outputs 0/4 to 20 mA (optional)
Endress+Hauser19
Electrical connectionLiquiport CSP44
+
-
+
-
-
-
+
+
RR
ll
Ui
Ui
1
(A)
(K)
(J)
(H)
(M)
(G)
(F)
(E)
(D)
(C)
(L)
(B)
(A)
(B)
(C)
(D)
(E)
(F)
(G)
(H)
(J)
(K)
(L)
(M)
1
2
3
4
U1-
U1+
B 1+I
B 1-I
B 2+I
B 2-I
B 1+O
B 1-O
B 2+O
B 2-O
AI1
In
AI1
GND
WH
BN
BU
RD
GN
YE
GY
PK
BK
VT
GY/PK
RD/BU
+
-
Ui
-
+
Ui
1
(A)
(K)
(J)
(H)
(M)
(G)
(F)
(E)
(D)
(C)
(L)
(B)
(A)
(B)
(C)
(D)
(E)
(F)
(G)
(H)
(J)
(K)
(L)
(M)
1
2
3
4
-
+
-
+
U1-
U1+
B 1+I
B 1-I
B 1+O
B 1-O
AI1
OUT
AI1
IN
AI1
GND
AI2
OUT
AI2
IN
AI2
GND
WH
BN
BU
RD
GN
YE
GY
PK
BK
VT
GY/PK
RD/BU
6.4 Signal cable connection (optional)
A0014162
6Pin assignment and wiring diagram of signal cable (version K3)
1Auxiliary voltage U: 24 V max. 30 mA load capacity
2Binary inputs BI: > 20 ms, only extra-low voltage Ui £ 30 V DC
3Binary outputs BO: only extra-low voltage Ui £ 30 V DC, max. current when using ext. auxiliary voltage (max.
200 mA)
4Analog input AI: 0 to 20 mA, 4 to 20 mA
20Endress+Hauser
7Pin assignment and wiring diagram of signal cable (version K4)
1Auxiliary voltage U: 24 V max. 30 mA load capacity
2Binary input BI: > 20 ms, only extra-low voltage Ui £ 30 V DC
3Binary output BO: only extra-low voltage Ui £ 30 V DC, max. current when using ext. auxiliary voltage (max.
200 mA)
4Analog inputs AI: 0 to 20 mA, 4 to 20 mA
A0014197
Liquiport CSP44Electrical connection
+
-
Ui
-
+
Ui
-
+
Ui
1
(A)
(K)
(J)
(H)
(M)
(G)
(F)
(E)
(D)
(C)
(L)
(B)
(A)
(B)
(C)
(D)
(E)
(F)
(G)
(H)
(J)
(K)
(L)
(M)
1
2
3
4
-
+
A
-
+
A
U1-
U1+
B 1+I
B 1-I
B 1+O
B 1-O
B 2+O
B 2-O
I 1+O
I 1-O
I 2+O
I 2-O
WH
BN
BU
RD
GN
YE
GY
PK
BK
VT
GY/PK
RD/BU
A0014198
8Pin assignment and wiring diagram of signal cable (version K5)
1Auxiliary voltage U: 24 V max. 30 mA load capacity
2Binary input BI: > 20 ms, only extra-low voltage Ui £ 30 V DC
3Binary output BO: only extra-low voltage Ui £ 30 V DC, max. current when using ext. auxiliary voltage (max.
200 mA)
4Analog inputs AI: 0 to 20 mA, 4 to 20 mA
6.5 Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions
and which are necessary for the required, designated use, may be carried out on the device
delivered.
Exercise care when carrying out the work.
‣
Individual types of protection permitted for this product (impermeability (IP), electrical
safety, EMC interference immunity, Ex protection) can no longer be guaranteed if, for
example :
• Covers are left off
• Different power units to the ones supplied are used
• Cable glands are not sufficiently tightened (must be tightened with 2 Nm (1.5 lbf ft) for
the permitted level of IP protection)
• Unsuitable cable diameters are used for the cable glands
• Modules are not fully secured
• The display is not fully secured (risk of moisture entering due to inadequate sealing)
• Loose or insufficiently tightened cables/cable ends
• Conductive cable strands are left in the device
Endress+Hauser21
Electrical connectionLiquiport CSP44
6.6 Post-connection check
WARNING
L
Connection errors
The safety of people and of the measuring point is at risk! The manufacturer does not
accept any responsibility for errors that result from failure to comply with the instructions
in this manual.
Put the device into operation only if you can answer yes to all the following questions.
‣
Instrument status and specifications
Are the device and all the cables free from damage on the outside?
‣
Electrical connection
Are the mounted cables strain relieved?
‣
Are the cables routed without loops and cross-overs?
‣
Are the signal cables correctly connected as per the wiring diagram?
‣
Are all plug-in terminals securely engaged?
‣
Are all the connection wires securely positioned in the cable terminals?
‣
22Endress+Hauser
Liquiport CSP44System integration
FieldCare
Service
FXA291
7 System integration
7.1 Service interface
You can connect the device to a computer via the service interface and configure it using
"FieldCare" . Furthermore, configurations can also be saved, transferred and documented.
7.1.1 Connection
1.Connect the service connector to the interface on the Liquiline base module and
connect it to the Commubox.
2.Connect the Commubox via the USB connection to the computer on which FieldCare
is installed.
A0039618
9Connection overview
7.1.2 Establishing the data connection
1.Start FieldCare.
2.Establish a connection to the Commubox. To do so, select the "CDI Communication
FXA291" ComDTM.
3.Then select the "Liquiline CM44x" DTM and start configuration.
You can now start online configuration via the DTM.
Online configuration competes with onsite operation, i. e. each of the two options blocks
the other one. On both sides it is possible to take away access from the other side.
7.1.3 Operation
• In the DTM the menu structure corresponds to the onsite operation. The functions of the
Liquiline soft keys are found in the main window on the left.
• Clicking a menu name or a function corresponds to pressing the navigator.
• You can make your settings conveniently via the computer keyboard.
• You can use FieldCare to save logbooks, make backups of configurations and transfer
configurations to other devices.
• You can also print out configurations or save them as PDFs.
Endress+Hauser23
Operation optionsLiquiport CSP44
EH_CSF
09:11:05 31.03.2015
MODE
4
3
2
1
1
2
3
8 Operation options
8.1 Overview
8.1.1 Display and operating elements
1
LED
2
Display (with red display background in alarm
3
condition)
4
Navigator (jog/shuttle and press/hold function)
Soft keys (function depends on menu)
10Overview of operation
8.1.2 Display
11Display (example)
A0025501
A0029090-EN
1
Menu path and/or device designation
2
Status display
3
Assignment of soft keys, e. g.:
ESC: escape or abortion of a sampling process
MAN: manual sample
?: Help, if available
MODE: switch the device to standby or cancel the
program
24Endress+Hauser
Liquiport CSP44Operation options
MODE
MODE
MODE
Menu/Language
Menu/LanguageMenu/Language
MODE
8.2 Access to the operating menu via the local display
8.2.1 Operating concept
Pressingthe soft key: selecting the menu directly
Pressingthe navigator: launching a function
Turningthe navigator: moving the cursor in the menu
Turningthe navigator: selecting a value (e.g. from a list)
Pressingthe navigator: accepting the new value
New setting is accepted
8.2.2 Locking or unlocking operating keys
Locking operating keys
Press the navigator for longer than 2 s.
‣
A context menu for locking the operating keys is displayed.
You have the choice of locking the keys with or without password protection. "With
Endress+Hauser25
password" means that you can only unlock the keys again by entering the correct
Operation optionsLiquiport CSP44
password. This password is set here: MenuSetupGeneral settingsExtended setupData
managementChange lock password
Choose whether you want to lock without or without a password.
‣
The keys are locked. No more entries can be made. In the soft key bar, you will see
the symbol.
The password is 0000 when the device is delivered from the factory. Make sure tonote down any changes to the password, as otherwise you will not be able to unlock
the keypad yourself.
Unlocking operating keys
1.Press the navigator for longer than 2 s.
A context menu for unlocking the operating keys is displayed.
2.Select Key unlock
The keys are unlocked immediately if you did not choose to lock with a password.
Otherwise you are asked to enter your password.
3.Only if keypad is password-protected: enter the right password.
The keys are unlocked. It is possible to access the entire onsite operation again.
The symbol is no longer visible on the display.
The password is 0000 when the device is delivered from the factory. Make sure tonote down any changes to the password, as otherwise you will not be able to unlock
the keypad yourself.
8.3 Configuration options
8.3.1 Display only
• You can only read the values but cannot change them.
• Typical read-only values are: sensor data and system information
• Example: Menu/Setup/Inputs/../Sensor type
8.3.2 Picklists
• You receive a list of options. In a few cases, these also appear in the form of multiple
choice boxes.
• Usually you just select one option; in rare instances you select one or more options.
• Example: Menu/Setup/General settings/Temperature unit
26Endress+Hauser
Liquiport CSP44Operation options
8.3.3 Numerical values
• You are changing a variable.
• The maximum and minimum values for this variable are shown on the display.
• Configure a value within these limits.
• Example: Menu/Operation/Display/Contrast
8.3.4 Actions
• You trigger an action with the appropriate function.
• You know that the item in question is an action if it is preceded by the following symbol:
• Examples of typical actions include:
• Deleting log entries
• Saving or loading configurations
• Triggering cleaning programs
• Examples of typical actions include:
• Start a sampling program
• Start manual sampling
• Saving or loading configurations
• Example: Menu/Manual sampling/Start sampling
Endress+Hauser27
Operation optionsLiquiport CSP44
E+H CSP4
8.3.5 Free text
• You are assigning an individual designation.
• Enter a text. You can use the characters in the editor for this purpose (upper-case and
lower-case letters, numbers and special characters).
• Using the soft keys, you can:
• Cancel your entries without saving the data ()
• Delete the character in front of the cursor ()
• Move the cursor back one position ()
• Finish your entries and save ()
• Example: Menu/Setup/General settings/Device tag
8.3.6 Tables
• Tables are needed to map mathematical functions or to enter irregular interval samples.
• You edit a table by navigating through rows and columns with the navigator and
changing the values of the cells.
• You only edit the numerical values. The controller automatically takes care of the
engineering units.
• You can add lines to the table ( INSERT) or delete lines from the table ( DEL).
• Afterwards, you save the table ( SAVE).
• You can also cancel your entries any time using the soft key.
• Example: Menu/Setup/Inputs/pH/Medium comp.
28Endress+Hauser
Liquiport CSP44Commissioning
OFFON
9 Commissioning
9.1 Function check
WARNING
L
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions!
Check that all connections have been established correctly in accordance with the
‣
wiring diagram.
Ensure that the supply voltage matches the voltage indicated on the nameplate.
‣
Saving displays as a screenshot
Via the local display, you can take screenshots at any time and save them to an SD
card.
1.Insert an SD card into the SD card slot in the basic module.
2.Press the navigator button for at least 3 seconds.
3.In the context menu select the "Screenshot" item.
The current screen is saved as a bitmap file to the SD card in the "Screenshots"
folder.
9.2 Switching on the measuring device
The device is delivered with a built-in rechargeable battery. The switch in the battery cover
is set to "OFF".
A0035816
12Switch position
1.Prior to initial commissioning, charge the batteries by connecting the power unit.
The battery starts charging as soon as the power unit is connected, regardless of
the switch position. It takes approx. 5 hours to fully charge the battery. Please
refer to the charger's operating manual for detailed information on the charger.
2.Once the charging process is complete, press the switch on the battery cover to the
"ON" position.
The transmitter starts up.
3.Wait for the boot process to complete.
Disconnecting the batteries when the sampler is not in use:
Press the switch on the battery cover to the "OFF" position.
‣
When the switch is in the "OFF" position, it provides reliable and effective
protection against the battery discharging completely and thus becoming
irreparably damaged.
Endress+Hauser29
CommissioningLiquiport CSP44
9.3 Setting the operating language
Configuring the language
If you have not already done so, close the housing cover and screw the device closed.
1.Connect the rechargeable battery (see "Electrical connection" section).
Wait for the initialization to finish.
2.Press the soft key MENU . Set your language in the top menu item.
The device can now be operated in your chosen language.
9.4 Configuring the measuring device
9.4.1 Start screen
You can find the following menu items and soft keys on the initial screen:
• Select sampling program
• Edit program %0V
• Start program %0V
• MENU
• MAN
• MEAS
• MODE
1)
1)
9.4.2 Display behavior
Menu/Operation/Display
FunctionOptionsInfo
Contrast5 to 95 %
Factory setting
50 %
BacklightSelection
• On
• Off
• Automatic
Factory setting
Automatic
Screen rotationSelection
• Manual
• Automatic
Factory setting
Manual
Current program:Read onlyThe name of the sampling program currently
StatusRead onlyActive
StartActionThe selected sampling program is started.
Adjust the screen settings to suit your working
environment.
Backlight = Automatic
The backlighting is switched off automatically
after a short time if a button is not pressed. It
switches back on again as soon as you press the
navigator button.
Backlight = On
The backlighting does not switch off
automatically.
If Automatic is selected, the single‐channel
measured value display switches from one
channel to the next every second.
selected is displayed.
The sampling program has been started and the
device takes a sample as per the set parameters.
Inactive
No sampling program has been started, or a
program that was running has been stopped.
1)"%0V" here stands for text that depends on the context. This text is generated automatically by the software and inserted in place of %0V.
30Endress+Hauser
Loading...
+ 114 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.