Combination of pH and ORP electrode with
Memosens technology
With open aperture diaphragm for heavily
contaminated media and suspended solids
Poison-resistant reference with ion trap
Application
• Chemical processes
• Pulp and paper industry
• Contaminated media and suspended solids
– Dispersions
– Precipitation reactions
– Emulsions
With ATEX, IEC, FM and CSA approval for use in hazardous areas.
Your benefits
• Simultaneous measurement of pH, ORP and rH value (in rH mode)
• Additional electrode made of platinum, for measuring reference impedance
• Open aperture diaphragm enables use in processes containing solids
• Low-maintenance owing to firm, stabilized gel with good chemical resistance
properties
• Not affected by fluctuations in pressure and temperature
• Fast response time
• Integrated temperature sensor for effective temperature compensation
• Long service life thanks to poison-resistant reference with ion trap
• Glass breakage and blockage detection by measuring:
– Resistance of glass membrane
– Reference impedance
Other advantages of Memosens technology
• Maximum process safety owing to non-contact, inductive signal transmission
• Data security thanks to digital data transmission
• Very easy to use as sensor data saved in the sensor
• Predictive maintenance possible as sensor load data logged in the sensor
Page 2
Function and system design
Memosens CPS96D
Measuring principle
pH measurement
The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The
membrane glass of the electrode supplies an electrochemical potential which is dependent upon the
pH value of the medium. This potential is generated by the selective penetration of H+ ions through
the outer layer of the membrane. An electrochemical boundary layer with an electric potential forms
at this point. An integrated Ag/AgCl reference system serves as the required reference electrode.
The transmitter converts the measured voltage into the corresponding pH value using the Nernst
equation.
ORP measurement
The ORP potential is a unit of measurement for the state of equilibria between oxidizing and
reducing components of a medium. The ORP is measured using a platinum or gold electrode instead
of the pH-sensitive glass membrane. Analog to the pH measurement, an integrated Ag/AgCl
reference system is used as a reference electrode.
rH measurement
The rH value is defined as the negative logarithm of partial pressure of hydrogen in a solution. The
pH value and ORP value of a solution must be measured simultaneously to calculate the rH value.
The value is calculated using the following equation:
rH = 2 . (mV/S) + 2 pH
pH = measured pH value
mV = measured ORP value in mV + 207 mV (Ag/AgCl system)
S = slope of pH electrode
The rH value is an indicator of the oxidation or reducing ability of a process solution. The measuring
range is from 0 to 42. A process medium with rH values between 0 and 9 has a high reducing ability;
a low reducing ability with rH 9-17; an indeterminate ability with rH 17-25; a low oxidation ability
with rH 25-34, and a high oxidation ability with rH 34-42.
2Endress+Hauser
Page 3
Memosens CPS96D
1
2
3
4
Measuring system
A complete measuring system comprises:
• pH/ORP combined electrode CPS96D
• Transmitter, e.g. Liquiline CM44x
• Measuring cable CYK10
• Immersion, fixed installation, flow or retractable assembly, e.g. CPA442
Always connect digital Memosens sensors with a combination of pH and ORP electrode to a
transmitter with the relevant functionality and Memosens technology.
Such transmitters are Liquiline CM42 (software package 9 or higher) and Liquiline CM44x (software
version 01.04.00, from 4th quarter 2012). To upgrade a Liquiline CM42 to software package 9 or
higher, the digital sensor input module FSDG1 with firmware version 1.01.01 or higher is required.
In the case of a FSDG1 with firmware version lower than 1.01.01, the sensor input module must be
replaced. Order numbers: 51518007 (non-Ex) and 51517469 (Ex). The installation of software
package 9 on the CM42 is performed using DAT CY42-S1.
The transfer of data to older transmitters, including those with Memosens capability, is possible only
to a limited extent (pH measurement only). In rare cases, it is possible that no data transfer is
performed.
Endress+Hauser3
Page 4
Memosens CPS96D
The digital sensors are able to store the following system data in the sensor.
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated slope at 25 °C (77 °F)
– Calibrated zero point at 25 °C (77 °F)
– Calibrated offset (ORP mV measuring mode)
– Slope as % (ORP % measuring mode)
– Temperature offset
– Number of calibrations
– Serial number of the transmitter used for the last calibration
– Calibration database (stores the last 8 calibrations in the Memosens head)
• Application data
– Temperature application range
– pH application range
– ORP application range
– Date of first commissioning
– Maximum temperature value
– Operating hours at temperatures above 80 °C (176 °F) and 100 °C (212 °F)
– Operating hours at very low and very high pH values (Nernst voltage below -300 mV, above
+300 mV)
– Number of sterilizations
Reliability
Maintainability
Integrity
Maximum process safety
With inductive transmission of the measured value using a non-contact connection, Memosens
guarantees maximum process safety and offers the following benefits:
• All problems caused by moisture are eliminated.
– Plug-in connection free from corrosion
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.
• The transmitter is galvanically decoupled from the medium. Issues concerning "symmetrical highimpedance" or "asymmetry" or an impedance converter are a thing of the past.
• EMC safety is guaranteed by screening measures for the digital transmission of measured values.
Low-maintenance
The electrode has a sterilizable, dirt-repellent PTFE ring junction that prevents clogging, making the
electrode reliable and stable over the long term.
Data security thanks to digital data transmission
Memosens technology digitizes the measured values in the sensor and transmits the data to the
transmitter using a non-contact connection that is free from potential interference. The result:
• Automatic error message if sensor fails or connection between sensor and transmitter is
interrupted
• Immediate error detection increases measuring point availability
Easy handling
Sensors with Memosens technology have integrated electronics that allow for saving calibration data
and further information such as total hours of operation and operating hours under extreme
measuring conditions. Once the sensor has been connected, the sensor data are transferred
automatically to the transmitter and used to calculate the current measured value. As the calibration
data are stored in the sensor, the sensor can be calibrated and adjusted independently of the
measuring point. The result:
• Easy calibration in the measuring lab under optimum external conditions increases the quality of
the calibration.
• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the
availability of the measuring point .
• Maintenance intervals can be defined based on all stored sensor load and calibration data and
predictive maintenance is possible.
• The sensor history can be documented on external data carriers and evaluation programs at any
time. Thus, the current application of the sensors can be made to depend on their previous history.
4Endress+Hauser
Page 5
Memosens CPS96D
GN/YE
YE
GN
BN
WH
GND
+
–
Com A
Com B
GY
Input
Measured variables
Measuring range
Connecting to the
transmitter
• pH value
• ORP
• rH value
• Temperature
pHORPrHTemperature
Version BT (process)0 ... 14-1500 to 1500 mV0 to 420 to 110 °C (32 to 230 °F)
Electrical connection
The electrical connection of the sensor to the transmitter takes place via special measuring cable
CYK10 or CYK20.
A0024019
2Measuring cable CYK10/CYK20
Endress+Hauser5
Page 6
Installation
15°
15°
A
B
Memosens CPS96D
Installation instructions
Installation angle
• Before screwing in the electrode, make sure the assembly thread is clean and runs smoothly.
• Screw in the electrode finger-tight (3 Nm)! (Data apply only if installing in Endress+Hauser
assemblies.)
• Also pay attention to the installation instructions provided in the Operating Instructions of the
assembly used.
Do not install the electrodes upside down. The angle of inclination must be at least 15° from the
horizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in the
glass sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.
A0024316
3Installation angle
APermitted orientation
BForbidden orientation
Environment
Ambient temperatureNOTICE
Risk of damage due to frost
The sensor must not be used if the temperature drops below –15 °C (5 °F).
‣
Storage temperature
Degree of protection
0 to 50 °C (32 to 120 °F)
IP 68 (10 m (33 ft) head of water at 25 ˚C (77 ˚F) over 45 days, 1 mol/l KCl)
6Endress+Hauser
Page 7
Memosens CPS96D
11
50100150200250300
15
203
p [psi] [bar]
T [°F]
400206080
1
5
7
9
3
T [°C]
23
110
14
230
140
absolute
Process
Process temperature
Process pressure
Pressure-temperature
ratings
BT version:0 to 110 °C (32 to 230 °F)
CAUTION
L
Pressurization of sensor due to prolonged use under increased process pressure
Risk of injury due to glass breakage
Avoid excessive heating of these sensors if using them under reduced process pressure or under
‣
atmospheric pressure.
When handling these sensors, wear protective goggles and suitable gloves.
‣
BT version:1 to 14 bar (15 to 203 psi)
Minimum conductivity
A0024728
4Pressure-temperature ratings CPS96D
500 μS/cm
Endress+Hauser7
Page 8
Design, dimensions
3
4
2
5
7
1
6
12
(0.47)
120, 225, 360
(4.72, 8.86, 14.17)
50 (1.97)
Memosens CPS96D
Mechanical construction
Weight
Materials
Process connection
Temperature sensor
Plug-in head
A0017212
5Design
1Memosens plug-in head, Pg 13.5
2Viton-O-ring with thrust collar
3Glass shaft
4Platinum round plate (ORP element)
5pH glass membrane
6Ag/AgCl reference lead - pH
7Reference with ion trap
6Measurements in mm (inch)
0.1 kg (0.2 lbs)
Electrode shaft
Glass to suit process
pH membrane glassesType B
ORP measuring elementPlatinum round plate
JunctionOpen aperture junction
Metal leadAg/AgCl
GelStabilized and firm
Pg 13.5
NTC 30KΩ
Memosens plug-in head for digital, non-contact data transmission
A0017211
Reference system
Ag/AgCl reference lead with stabilized gel and ion trap in reference
8Endress+Hauser
Page 9
Memosens CPS96D
Certificates and approvals
Ex approval
Electromagnetic
compatibility
Product page
Product Configurator
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• IECEx Ex ia IIC T3/T4/T6 Ga
• FM/CSA IS/NI Cl. I. Div 1+2, Group A-D
Hazardous area versions of digital sensors with Memosens technology are indicated by a red-
orange ring in the plug-in head.
Interference emission and interference immunity as per EN 61326: 2006
Ordering information
www.endress.com/cps96d
The navigation area is located on the right of the product page.
1.Under "Device support" click "Configure your selected product".
The Configurator opens in a separate window.
2.Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3.Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of
the screen.
Endress+Hauser9
Page 10
Memosens CPS96D
Accessories
The following are the most important accessories available at the time this documentation was
issued. For accessories not listed here, please contact your service or sales office.
Assemblies
Dipfit CPA111
• Immersion and installation assembly made of plastic for open and closed vessels
• Product Configurator on the product page: www.endress.com/cpa111
Technical Information TI00112C
Dipfit CPA140
• pH/ORP immersion assembly with flange connection for very demanding processes
• Product Configurator on the product page: www.endress.com/cpa140
Technical Information TI00178C
Unifit CPA442
• Installation assembly for food, biotechnology and pharmaceutics
• With EHEDG and 3A certificate
• Product Configurator on the product page: www.endress.com/cpa442
Technical Information TI00306C
Cleanfit CPA871
• Flexible process retractable assembly for water, wastewater and the chemical industry
• For applications with standard 12mm sensors
• Product Configurator on the product page: www.endress.com/cpa871
Technical Information TI01191C
Not suitable for combined sensors with length of 120 mm.
Cleanfit CPA472
• Compact plastic retractable assembly for installation in tanks and pipes
• For manual or pneumatic, remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa472
Technical Information TI00223C
Cleanfit CPA472D
• Robust retractable assembly for pH, ORP and other industrial sensors
• Heavy-duty version made of durable materials
• For manual or pneumatic, remote-controlled operation
• Product Configurator on the product page: www.endress.com/cpa472d
Technical Information TI00403C
Cleanfit CPA473
• Stainless steel process retractable assembly with ball valve shutoff for particularly reliable
separation of the medium from the environment
• Product Configurator on the product page: www.endress.com/cpa473
Technical Information TI00344C
Cleanfit CPA474
• Plastic process retractable assembly with ball valve shutoff for particularly reliable separation of
the medium from the environment
• Product Configurator on the product page: www.endress.com/cpa474
Technical Information TI00345C
Cleanfit CPA875
• Retractable process assembly for sterile and hygienic applications
• For in-line measurement with standard 12 mm sensors for parameters such as pH, ORP and
oxygen
• Product Configurator on the product page: www.endress.com/cpa875
10Endress+Hauser
Page 11
Memosens CPS96D
Technical Information TI01168C
Cleanfit CPA450
• Manual retractable assembly for installing 120 mm sensors in tanks and pipes
• Product Configurator on the product page: www.endress.com/cpa450
Technical Information TI00183C
Flowfit CPA250
• Flow assembly for pH/ORP measurement
• Product Configurator on the product page: www.endress.com/cpa250
Technical Information TI00041C
Flowfit CPA240
• pH/redox flow assembly for processes with stringent requirements
• Product Configurator on the product page: www.endress.com/cpa240
Technical Information TI00179C
Ecofit CPA640
• Set comprising adapter for 120 mm pH/ORP electrodes and sensor cable with TOP68 coupling
• Product Configurator on the product page: www.endress.com/cpa640
Technical Information TI00246C
Buffer solutions
Measuring cable
Only versions CPA640-A***, CPA640-B***, CPA640-G*** and CPA640-I*** are suitable for
installing combined sensors.
High-quality buffer solutions from Endress+Hauser - CPY20
The secondary buffer solutions have been referenced to primary reference material of the PTB
(German Federal Physico-technical Institute) and to standard reference material of NIST (National
Institute of Standards and Technology) according to DIN 19266 by a DKD (German Calibration
Service) accredited laboratory.
Product Configurator on the product page: www.endress.com/cpy20
Memosens data cable CYK10
• For digital sensors with Memosens technology
• Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C