Endress+Hauser CPA140 Specifications

TI00178C/07/EN/13.18 71436283 2018-08-31
Products
Solutions Services
Technical Information
Dipfit CPA140
Immersion assembly with flange connection for very demanding processes for the installation of 12 mm sensors
Sensor holder with bayonet lock
Application
• Chemical industry, e.g. in
• production of synthetic materials and dyes
• production or pesticides and fertilizers
• oil or wastewater separation
• condensate treatment
• Power stations and incinerator plants, e.g. in
• flue gas cleaning
• Metal extraction and metal processing
Your benefits
• Simple installation and removal of the sensor holder thanks to bayonet lock technology
• Reliable sealing even in media containing solid particles
• Suitable for use at high pressures and temperatures
• Immersion depth from 500 to 2500 mm
• Integrated potential matching pin made of Alloy C4 or tantalum
• Reduced condensation thanks to breathable GORE-TEX® filter
• 3 slots for 12 mm sensors and cleaning head
• Chemical and ultrasonic cleaning can be easily retrofitted
• Variety of flange connections (DIN, ANSI, JIS) for flexible connection to the process

Function and system design

1 2
3
5
4
Dipfit CPA140

Measuring system

A complete measuring system comprises:
• Immersion assembly Dipfit CPA140
• 1-3 pH, ORP, pH/ORP combined sensors or temperature sensors (12 mm), e.g. CPS11D, CPS12D
• 1-3 measuring cables, e.g. CYK10 or CPK9
• Transmitter, e.g. Liquiline CM442 Optional:
Extension cable, e.g. CYK11
 1 Example of a measuring system (process and process connections are not illustrated)
1 Immersion assembly Dipfit CPA140, here as PVDF version 2 Transmitter CM442 3 Sensor cable CYK10 4 pH sensor CPS11D 5 ORP sensor CPS12D
A0037542
2 Endress+Hauser
Dipfit CPA140
200
2
4
6
8
10
40 60 80 100 120 140 160
a
b
p, abs
[bar][psi]
[°C]
[°F]
T
6832 104 140 176 212 248 284 320
29
58
87
116
145
-15
5

Environment

Ambient temperature range

Storage temperature

Degree of protection

Process temperature

Process pressure

Pressure-temperature ratings

-10 to +70 °C (+10 to +160 °F)
-10 to +70 °C (+10 to +160 °F)
IP65

Process

PVDF version Stainless steel version -15 to 150 °C (5 to 300 °F), for all seals except EPDM
PVDF version Stainless steel version Max. 10 bar (145 psi), absolute
0 to 120 °C (32 to 250 °F)
-15 to 140 °C (5 to 280 °F), for EPDM seal
Max. 6 bar (87 psi), absolute
 2 Pressure-temperature ratings
a Stainless steel version a PVDF version
Endress+Hauser 3
A0037394-EN

Mechanical construction

2
3
4
5
1
6
7
8

Design PVDF version

Dipfit CPA140
 3 PVDF version
1 Assembly head 2 Cable gland Pg 13.5 and 2x dummy plug Pg 16 3 GORE-TEX® filter 4 3 sensor slots for 120 mm sensors 5 Potential matching pin 6 Shock-protection stud 7 Sensor holder with bayonet lock 8 Lap joint flange, depending on version
A0037531
4 Endress+Hauser
Loading...
+ 8 hidden pages