Endress+Hauser CM82 Operating Manual

Page 1
BA01845C/07/EN/01.18 71406304 2018-04-30
Products Solutions Services
Operating Instructions
Liquiline Compact CM82
Compact transmitter configurable for Memosens sensors from the Liquiline series
Page 2
Page 3

Table of contents

1 Document information .............. 4
1.1 Warnings ............................ 4
1.2 Symbols .............................. 4
1.3 Symbols at the device .................... 4
1.4 Documentation ........................ 5
2 Basic safety instructions ............ 6
2.1 Requirements for personnel ............... 6
2.2 Designated use ........................ 6
2.3 Occupational safety ..................... 6
2.4 Operational safety ...................... 6
3 Product description ................. 8
3.1 Product design ......................... 8
4 Incoming acceptance and product
identification ....................... 9
4.1 Incoming acceptance .................... 9
4.2 Product identification .................... 9
4.3 Scope of delivery ...................... 10
4.4 Certificates and approvals ............... 10
5 Installation ....................... 12
5.1 Dimensions .......................... 12
6 Electrical connection .............. 13
6.1 Connection .......................... 13
6.2 Post-connection check .................. 14
7 Operation options ................. 15
7.1 Overview of operating options ............ 15
7.2 Access to operating menu via SmartBlue
(app) ............................... 15
7.3 Access to operating menu via RIA15 ........ 17
10 Operation ......................... 35
10.1 Reading measured values ................ 35
10.2 Adapting the measuring device to the process
conditions ........................... 36
11 Diagnostics and troubleshooting ... 38
11.1 Diagnostic information via LED ........... 38
11.2 Adapting the diagnostic information ....... 38
11.3 Adapting sensor diagnostic information ..... 39
11.4 Simulation ........................... 47
11.5 Diagnostic list ........................ 48
11.6 Event logbook ........................ 62
11.7 Resetting the measuring device ........... 62
11.8 Device information .................... 63
11.9 Firmware history ..................... 63
12 Maintenance ...................... 64
12.1 Maintenance tasks ..................... 64
13 Repair ............................ 65
13.1 General notes ........................ 65
13.2 Return .............................. 65
13.3 Disposal ............................ 65
14 Accessories ....................... 66
14.1 Sensors ............................. 66
14.2 Software ............................ 69
14.3 Other accessories ...................... 69
15 Technical data .................... 71
15.1 Input ............................... 71
15.2 Output ............................. 71
15.3 Power supply ......................... 71
15.4 Performance characteristics .............. 72
15.5 Environment ......................... 72
15.6 Mechanical construction ................ 73
8 System integration ................ 18
8.1 Integrating the measuring device into the
system ............................. 18
Index .................................. 75
9 Commissioning .................... 19
9.1 Preparatory steps ...................... 19
9.2 Function check ....................... 19
9.3 Establishing connection via SmartBlue (app) .. 20
9.4 Establishing a connection via RIA15 ........ 21
9.5 Setting the operating language ............ 27
9.6 Date and time ........................ 27
9.7 Configuring the measuring device .......... 27
9.8 Configuration management .............. 34
9.9 Unauthorized access ................... 34
Endress+Hauser 3
Page 4
Document information Liquiline Compact CM82

1 Document information

1.1 Warnings

Structure of information Meaning
DANGER
L
Causes (/consequences)
If necessary, Consequences of non-compliance (if applicable)
Corrective action
WARNING
L
Causes (/consequences)
If necessary, Consequences of non-compliance (if applicable)
Corrective action
CAUTION
L
Causes (/consequences)
If necessary, Consequences of non-compliance (if applicable)
Corrective action
NOTICE
Cause/situation
If necessary, Consequences of non-compliance (if applicable)
Action/note
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.

1.2 Symbols

Symbol Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step

1.3 Symbols at the device

Symbol Meaning
Reference to device documentation
4 Endress+Hauser
Page 5
Liquiline Compact CM82 Document information

1.4 Documentation

The following instructions complement these Operating Instructions and are available on the product pages on the Internet: Operating Instructions Memosens, BA01245C – Software description for Memosens inputs – Calibration of Memosens sensors – Sensor-specific diagnostics and troubleshooting
Endress+Hauser 5
Page 6
Basic safety instructions Liquiline Compact CM82

2 Basic safety instructions

2.1 Requirements for personnel

• Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and must follow the instructions contained therein.
• Faults at the measuring point may only be rectified by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided must be carried out only directly at the manufacturer's site or by the service organization.

2.2 Designated use

The Liquiline CM72 Liquiline CM82 is a transmitter for connecting digital sensors with Memosens technology, configurable, with 4..20 mA/HART communication and optional operation via smartphone or other mobile devices via Bluetooth.
The device is designed for use in the following industries:
• Life science
• Chemical industry
• Water and wastewater
• Food and beverages
• Power stations
• Other industrial applications

2.3 Occupational safety

As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been connected in accordance with these Operating Instructions.

2.4 Operational safety

Before commissioning the entire measuring point:
1. Verify that all connections are correct.
2. Ensure that electrical cables and hose connections are undamaged.
3. Do not operate damaged products, and protect them against unintentional operation.
4. Label damaged products as defective.
6 Endress+Hauser
Page 7
Liquiline Compact CM82 Basic safety instructions
During operation:
If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.
CAUTION
L
Cleaning not switched off during calibration or maintenance activities
Risk of injury due to medium or cleaning agent!
If a cleaning system is connected, switch it off before removing a sensor from the
medium. If you wish to check the cleaning function and have therefore not switched off the
cleaning system, wear protective clothing, goggles and gloves or take other appropriate measures.
Endress+Hauser 7
Page 8
Product description Liquiline Compact CM82
1
2
3
4

3 Product description

3.1 Product design

A0036216
 1 Transmitter design
1 Cable 2 Housing 3 Memosens connection 4 LED, for optical signaling of operating statuses of measuring point

3.1.1 Measuring parameters

The transmitter is designed for digital Memosens sensors with inductive plug-in head:
• pH, ORP. pH/ORP combined sensors
• Conductive Conductivity
• Dissolved oxygen
8 Endress+Hauser
Page 9

Liquiline Compact CM82 Incoming acceptance and product identification

4 Incoming acceptance and product
identification

4.1 Incoming acceptance

1. Verify that the packaging is undamaged.  Notify the supplier of any damage to the packaging.
Keep the damaged packaging until the issue has been resolved.
2. Verify that the contents are undamaged.  Notify the supplier of any damage to the delivery contents.
Keep the damaged goods until the issue has been resolved.
3. Check that the delivery is complete and nothing is missing.  Compare the shipping documents with your order.
4. Pack the product for storage and transportation in such a way that it is protected against impact and moisture.
 The original packaging offers the best protection.
Make sure to comply with the permitted ambient conditions.
If you have any questions, please contact your supplier or your local Sales Center.

4.2 Product identification

4.2.1 Nameplate

The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Firmware version
• Ambient and process conditions
• Input and output values
• Safety information and warnings
• Approvals as per version ordered
Compare the data on the nameplate with your order.

4.2.2 Product identification

Product page
www.endress.com/CM82
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Open the product website.
2. In the page header, select: Product tools.  Online Tools: An additional area opens up.
Endress+Hauser 9
Page 10
Incoming acceptance and product identification Liquiline Compact CM82
3. Select: Access device specific information.  An additional window opens.
4. Enter the order code from the nameplate into the search field. Then select: Show details.
 Details of each feature (selected option) of the order code are displayed.

4.3 Scope of delivery

The scope of delivery includes:
• CM82
• Brief Operating Instructions
If you have any queries:
Please contact your supplier or local sales center.

4.4 Certificates and approvals

4.4.1
Endress+Hauser Conducta GmbH+CO. KG hereby declares that the radio system type CM82 complies with directives 2014/53/EU and 2011/65/EU.
The complete text of the EU declaration of conformity can be found at the following web address: "http://www.endress.com/CM82"
Frequency band: 2400-2483.5 MHz, power output: < 10dBm EIRP
mark

4.4.2 FCC/IC

Radio approval for USA/Canada
This device complies with Part 15 of the FCC Rules [and with Industry Canada license­exempt RSS standard(s)]. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Any changes or modifications made to this equipment not expressly approved by Endress +Hauser may void the FCC authorization to operate this equipment. This device complies with Part 15 of the FCC Rules and with Industry Canada license- exempt RSS standard(s).
Operation is subject to the following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may cause undesired operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes:
• l'appareil ne doit pas produire de brouillage, et
• l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the
10 Endress+Hauser
Page 11
Liquiline Compact CM82 Incoming acceptance and product identification
202-LSF040
R
R-CRM-E1H-CM82A
user is encouraged to try to correct the interference by one or more of the following measures:
• Increase the separation between the equipment and receiver.
• Consult the dealer or an experienced radio technician for help.
Radio approval for Japan
A0036603
Japanese Radio Law and Japanese Telecommunications Business Law Compliance. This device is granted pursuant to the Japanese Radio Law (電波法). This device should not be modified (otherwise the granted designation number will become invalid).
Radio approval for China
Certification number: CMIIT ID: 2017DJ6495
Radio approval for South Korea
A0036602
Endress+Hauser 11
Page 12
Installation Liquiline Compact CM82
110.8 (4.33)
99.8 (3.92)
32.75 (1.29)
114.7 (4.52)
124.6 (4.90)
Ø
18.95 (0.75)
Ø
5.5 (0.22)

5 Installation

5.1 Dimensions

 2 Dimensions in mm (inch)
A0033272
12 Endress+Hauser
Page 13
Liquiline Compact CM82 Electrical connection
GN/YE
BU
WH
+
GN/YE
GND
GY

6 Electrical connection

WARNING
L
Device is live!
Incorrect connection may result in injury or death!
The electrical connection may be performed only by an electrical technician.
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein. Prior to commencing connection work, ensure that no voltage is present on any cable.

6.1 Connection

Supply voltage: 12,6 to 30 VDC (In the case of a residual current > 20 mA)
14 to 30 VDC (When the failure current is set to 3.6 mA.)
Cable length: 3 m (10 ft)
7 m (23 ft) 15 m (46 ft)
Signal output: 4 to 20 mA
Signal on alarm: 3.6 to 23 mA
 3 Electrical connection
Connect ferrules as specified in the table:
Cable Function
Gray (GY) Grounding, GND
BU (blue) 4 to 20 mA +
White (WH) 4 to 20 mA -
The grounding cable must be provided by the customer.

6.1.1 With RIA15

The RIA15 process display unit is loop-powered and does not require any external power supply.
Further information is available in the RIA15 Operating Instructions BA01170K.

6.1.2 With junction box

Max. operating voltage: 30 V
Max. operating current 30 mA
A0033282
Wiring
1. Unscrew cover and remove.  The terminal assignment is indicated in the box.
Endress+Hauser 13
Page 14
Electrical connection Liquiline Compact CM82
1
2
3 4 5 6
7 8
9 10
11 12
0/4...20 mA Source+ 0/4...20 mA Source+ 0/4...20 mA Source-
Display+ Display (Bridge)
Display- or LED Sensor+ (w/Display)
Sensor+ (w/o Display) Sensor-
Shield
HART-
HART+
2. Guide the cable cores through the M16 cable gland.
3. Connect cores in accordance with the assignment provided.
A0034718
 4 Terminal diagram
Further information is available in the Operating Instructions BA01802C.

6.2 Post-connection check

WARNING
L
Connection errors
The safety of people and of the measuring point is under threat. The manufacturer does not accept any responsibility for errors that result from failure to comply with the instructions in this manual.
Put the device into operation only if you can answer yes to all the following questions.
Electrical connection
Is the device or cable undamaged (visual inspection)?
Do the mounted cables have adequate strain relief?
Are the cables routed without loops and cross-overs?
Does the supply voltage match the specifications on the nameplate?
No reverse polarity, is terminal assignment correct?
14 Endress+Hauser
Page 15
Liquiline Compact CM82 Operation options
EH_FMRx0_0123456
CM82_M50053105G11

7 Operation options

7.1 Overview of operating options

Operation and settings via:
• SmartBlue (app)
• RIA15 (with reduced operating function compared to app and HART)
• PLC control station (via HART)

7.2 Access to operating menu via SmartBlue (app)

SmartBlue is available as a download for Android terminals from the Google Playstore and for iOS devices from the Apple iTunes Store.
If you scan the QR code, you will be brought directly to the app:
 5 Download links
 6 SmartBlue App
A0031189-EN
A0029747
 7 Livelist
The Livelist displays all of the devices that are within range.
Endress+Hauser 15
A0035117
Page 16
Operation options Liquiline Compact CM82
System requirements
• iOS devices: iPhone 4S or higher from iOS9.0; iPad2 or higher from iOS9.0; iPod Touch
5. Generation or higher from iOS9.0
• Devices with Android: from Android 4.4 KitKat and Bluetooth® 4.0

7.2.1 Structure and function of the operating menu

In the Home view, the current measured values are displayed along with the device information (tag, serial number, firmware version, order code).
 8 Overview of current measured values
1 System and device information CM82 2 Shortcut to diagnostic list 3 Overview of measured values of connected sensor
The device is operated via four main menus:
• Guidance
• Diagnostics
• Application
• System
 9 Main menu
16 Endress+Hauser
Page 17
Liquiline Compact CM82 Operation options
Menu Function
Guidance Contains functions involving a self-contained sequence of activities, e.g. for calibration (="Wizard",
guided operation).
Diagnostics Contains information regarding operation, diagnostics and troubleshooting, as well as
configuration of the diagnostic behavior.
Application Sensor data for specific optimization and for detailed process adjustment. Adjustment of
measuring point to the application.
System These menus contain parameters for configuring the overall system.

7.3 Access to operating menu via RIA15

The RIA15 process indicator is incorporated into the 4 to 20 mA/HART® loop and displays the measuring signal in digital form. The process indicator does not require an external power supply. It is powered directly from the current loop.
By means of HART® communication, the RIA15 enables configuration and commissioning of selected field devices and readouts of device/sensor status messages.
 10 Process display unit RIA15
A0017816
Endress+Hauser 17
Page 18
System integration Liquiline Compact CM82
1
2

8 System integration

8.1 Integrating the measuring device into the system

Interfaces for measured value transmission:
• 4 to 20 mA
• Bluetooth® LE wireless technology
• HART

8.1.1 Bluetooth® wireless technology

 11 Options for remote operation via Bluetooth® LE wireless technology
1 Smartphone / tablet with SmartBlue (app) 3 Transmitter with Bluetooth® wireless technology

8.1.2 HART

In addition to the analog 4 ... 20 mA signal, as well as the status of the device can be transmitted digitally.
Parameterisation is also possible using an additional control unit and a suitable driver.
HART operation is possible via the following hosts (at least):
• Fieldcare und kompatible DTM-Hosts
• SFX350 Handheld
• Emerson 475 Handheld
• Emerson AMS
• Siemens PDM
A0036075
18 Endress+Hauser
Page 19
Liquiline Compact CM82 Commissioning
ESC
E
+
1
3
4
2
5
6
A0036740
 12 Wiring options for remote operation via HART protocol
1 PLC (programmable logic controller) 2 Loop-powered process display unit RIA15, optional 3 HART control unit (e.g. SFX350) 4 Junction box 5 Transmitter with Bluetooth® LE wireless technology 6 Optional: Smartphone / tablet with SmartBlue (app)

9 Commissioning

9.1 Preparatory steps

Connect the device.
 The device starts up and transmits the measured value as a current value.
To operate via the SmartBlue, the Bluetooth® LE signal on the smartphone or tablet must be switched on.

9.2 Function check

WARNING
L
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions!
Check that all connections have been established correctly in accordance with the
wiring diagram. Ensure that the supply voltage matches the voltage indicated on the nameplate.
Familiarize yourself with the operation of the device before it is first switched on. In particular please read the "Basic safety instructions" sections. After power-up, the device performs a self-test and then goes to the measuring mode.
Endress+Hauser 19
Page 20
Commissioning Liquiline Compact CM82

9.2.1 LED display

LED messages signal the status of the device and sensor.
LED behavior Status
Green Flashes quickly
Green Flashes twice
Green Flashes slowly
Green Flashes quickly three times
Red Flashes quickly
Red, green Three red flashes alternating with three green flashes
Everything OK Device starting up
Everything OK Read out Memosens sensor information from sensor to transmitter (sensor type, calibration data, etc.)
Everything OK Sensor and device OK and functioning correctly.
Everything OK Measured value at PLC in automatic HOLD. If the "Sensor replacement alarm delay" is exceeded, the device transmits a signal on alarm. The automatic hold is set to 30 seconds but can be configured to suit the customer's needs.
Failure of device or sensor Fault state as per NAMUR NE107
Squawk Squawk is signaled briefly while the connection is established. Squawk can also be activated via the app. This makes it possible to locate the device more quickly, e. g. when several devices are installed, you can see which one the connection is established with.

9.3 Establishing connection via SmartBlue (app)

1. Download and install the SmartBlue.
2. Start the SmartBlue.
3. Select device from livelist displayed. All available devices are displayed.
4. Perform login
5. Enter user name -> admin
6. Enter initial password -> device serial number
7. It is advisable to change the user name and password after logging in for the first time.
You can drag additional information (e.g. main menu) onto the screen by swiping across the screen

9.3.1 System settings

Path: Settings
Function Options Info
Information
Version Displays the app version
About Endress+Hauser Manufacturer's information
User interface
Language Picklist of
different languages
Change language
20 Endress+Hauser
Page 21
Liquiline Compact CM82 Commissioning
ESC
E
+
1
4
2
3
A
B
Path: Settings
Function Options Info
Save device login passwords
Device List
Sorting Selection
Show demo devices Selection
Selection
• Off
• 5 minutes
• 15 minutes
• 60 minutes
• Signal strength
• Name
Options for saving password The password is stored temporarily for the selected time period. It does not need to be entered when re-establishing a connection, e.g.  to replace a sensor.
Sorting options
Which devices are displayed in the list

9.4 Establishing a connection via RIA15

The RIA15 can be used as a local indicator of the measured values as well as for basic configuration of the Liquiline CM82 via HART®.
Here, the RIA15 communicates with the CM82 via HART as a secondary master in addition to the PLC or process control system. The RIA15 is not invisible to the PLC in this case. The RIA15 does not alter the current value of the current output of the CM82.
A0036208
 13 Remote operation of CM82 via RIA15
1 PLC 2 RIA15 loop-powered process display unit 3 CM82 transmitter 4 Memosens sensor (e.g. pH sensor) A 4 to 20mA (HART optional) B 4 to 20mA with HART
Endress+Hauser 21
Page 22
Commissioning Liquiline Compact CM82
1
2
3
4
5
6
7
8
A0017719
 14 Display and operating elements of the process display unit
1 Operating menu locked 2 Error 3 Warning 4 HART communication enabled 5 Operating keys "-", "+", "E" 6 14-segment display for unit/TAG 7 Bar graph with indicators for under range and over range 8 5-digit 7-segment display for measured value, digit height 17 mm (0.67 in)
The device is operated using three operating keys on the front of the housing.
The device setup can be disabled with a 4-digit user code. If the setup is disabled, a padlock symbol appears on the display when an operating parameter is selected.
Enter key; calling up the operating menu, confirming the option/setting parameters in the operating menu
A0017716
Selecting and setting/changing values in the operating menu; pressing the '-' and '+' keys simultaneously takes the user back up a menu level. The configured value is not saved.
A0017715
A0017714

9.4.1 RIA15 operating matrix

In HART mode, the RIA15 with "Analysis" option can be used for basic configuration of the Liquiline CM82.
The measuring ranges are dependent on the connected sensor and can be found in the relevant sensor documentation.
Local measured value display and basic configuration of the CM82
The RIA15 can be used as a local indicator of the measured values as well as for basic configuration of the Liquiline CM82 via HART®.
The following values are output here: Digital output (HART®): Measured value and unit depending on the connected sensor
PV: Configured primary value (CMAIN operating parameter)
SV: Temperature (sensor)
TV: Dependent on the connected transmitter parameter + sensor type
QV: Dependent on the connected transmitter parameter + sensor type
Transmitter parameter
pH Glass Raw value in mV Glass impedance in MOhm
pH ISFET Raw value in mV Leak current in nA
pH ORP Relative ORP value as % Raw value in mV
22 Endress+Hauser
Sensor type "TV" value "QV" value
Page 23
Liquiline Compact CM82 Commissioning
Transmitter parameter
pH pH/ORP combined
Conductivity Resistance Conductivity, raw value
Dissolved oxygen Liquid concentration Saturation as %
Sensor type "TV" value "QV" value
pH ORP in mV
sensor
If "UC170" is displayed instead of the unit, see footnote in table.
The following settings for the CM82 can be made using the three operating keys on the front of the RIA15:
• Units of connected sensor
• Current output range
• Retrieval of diagnostic information
Basic configuration of the CM82
The RIA15 must be in the HART mode (MODE = HART) to make the basic settings. The ANALYSIS menu is not visible in analog mode (MODE = 4-20).
1. Press the key.  The Setup menu opens.
2. Press the key.  The CT submenu opens.
3. Set the desired parameters. For parameter descriptions, see the following table.
Setup -> ANALYSIS menu
The CT menu and all of the associated submenus are visible only if the RIA15 was ordered with the "Analysis" option, the HART option has been configured and a CM82 has been detected by the RIA15. Using this menu, the basic settings for the CM82 can be made via the RIA15.
Parameter Values Description
CT This menu contains the parameters for configuring the CM82
compact transmitter.
CSET Access the "CM82 setup" submenu
TUNIT °C
°F °K
OUTS Access the "CM82 - Output Setting" submenu to change the setting
pH glass sensors
CMAIN pH
mV_PH IMPGL TEMP
pH/ISFET sensors
CMAIN pH
mV_PH LEAKC TEMP
pH/ORP sensors
Select the unit for temperature on the CM82.
on the CM82. The primary value (CMAIN) of the CM82 is assigned here and the measuring range (4-20mA) configured.
Depending on the sensor type connected, only certain
measured values can be configured/displayed.
PH: pH measured value in pH mV_PH: pH raw value in mV IMPGL: Glass impedance in MOhm TEMP: Temperature in °C/°F/K (unit as per setting in TUNIT)
PH: pH measured value in pH mV_PH: pH raw value in mV LEAKC: ISFET leak current in "nA" TEMP: Temperature in °C/°F/K (unit as per setting in TUNIT)
1)
1)
Endress+Hauser 23
Page 24
Commissioning Liquiline Compact CM82
Setup -> ANALYSIS menu
The CT menu and all of the associated submenus are visible only if the RIA15 was ordered with the "Analysis" option, the HART option has been configured and a CM82 has been detected by the RIA15. Using this menu, the basic settings for the CM82 can be made via the RIA15.
Parameter Values Description
CMAIN mVORP
%_ORP TEMP
pH/ORP combined sensors
CMAIN pH
mV_PH IMPGL IMPRE mVORP %_ORP RH TEMP
Oxygen sensors
CMAIN PAR_P
%SAT C_LIQ C_GAS CURR RTIME TEMP
UCLIQ
mG_L uG_L PPM PPB
UCGAS
%_VOL PPM_V
Conductivity sensors
CMAIN COND
RESIS RAWC TEMP
URES
KO*CM MO*CM KO*M
UCOND
uS/cm mS/cm S/cm uS/m mS/m S/m
for all sensors
mVORP: ORP measured value in mV %_ORP: Percentage ORP value as % TEMP: Temperature in °C/°F/K (unit as per setting in TUNIT)
PH: pH measured value in pH mV_PH: pH raw value in mV IMPGL: Glass impedance in MOhm
1)
IMPRE: Reference impedance in Ohm mVORP: ORP measured value in mV %_ORP: Percentage ORP value as % RH: rH value in rH TEMP: Temperature in °C/°F/K (unit as per setting in TUNIT)
PAR_P: Partial pressure of oxygen in hPa %SAT: Percentage saturation as % C_LIQ: Liquid concentration (unit as per setting in UCLIQ) C_GAS: Gas concentration (unit as per setting in UCGAS) CURR: Raw value, measuring current of sensor in nA
1)
(visible only in the case of amperometric oxygen sensors) RTIME: Decay time, raw value in μs (visible only in the case of optical oxygen sensors) TEMP: Temperature in °C/°F/K (unit as per setting in TUNIT)
Unit of upper and lower turndown setting if the primary value (CMAIN) is set to C_LIQ mG_L: milligram/liter
1)
uG_L: microgram/liter PPM: parts per million PPB: parts per billion
Unit of upper and lower turndown setting if the primary value (CMAIN) is set to C_GAS %_VOL: percent by volume PPM_V: parts per million
COND: specific conductivity (unit as per setting in UCOND) RESIS: specific resistance (unit as per setting in URES) RAWC: uncompensated conductivity (unit as per setting in UCOND) TEMP: temperature (unit as per setting in TUNIT)
Unit of upper and lower turndown setting if the primary value (CMAIN) is set to RESIS KO*CM: kOhm*cm MO*CM: MOhm*cm KO*M: kOhm*m
Unit of upper and lower turndown setting if the primary value (CMAIN) is set to COND or RESIS uS/cm: microsiemens/cm mS/cm: millisiemens/cm S/cm: siemens/cm uS/m: microsiemens/m mS/m: millisiemens/m S/m: siemens/m
24 Endress+Hauser
Page 25
Liquiline Compact CM82 Commissioning
Setup -> ANALYSIS menu
The CT menu and all of the associated submenus are visible only if the RIA15 was ordered with the "Analysis" option, the HART option has been configured and a CM82 has been detected by the RIA15. Using this menu, the basic settings for the CM82 can be made via the RIA15.
Parameter Values Description
LOW -19,999 to
99,999
Configure turndown of current output. The measured value that corresponds to 4 mA is set here. The limits of adjustment vary depending on the sensor type and measured value. The position of the decimal point is permanently preset depending on the primary value (CMAIN) configured.
Valid ranges of adjustment: pH sensor:
PH: -2.00 to 16.00 pH mV_PH: -2000 to 2000 mV LEAKC: -4000.0 to 4000.0 nA IMPGL: 0 to 99999 MOhm IMPRE: 0 to 99999 Ohm mVORP: -2000 to 2000 mV %_ORP: -3000.0 to 3000.0 % RH: 0.0 to 70.0 rH TEMP: -50.0 to 150.0 °C (depending on the unit configured under TEMP)
-58.0 to 302.0°F
223.1 to 423.1 K
dissolved oxygen sensor:
PAR_P: 0.0 to 2500.0 hPa %SAT: 0.02 to 200.00 % saturation C_LIQ:
-0.02 to 120.00 mg/l
-20.00 to 999.99 ug/l
-0.02 to 120.00 ppm
-20.00 to 999.99 ppb (depending on the unit configured in UCLIQ) C_GAS:
-0.02 to 200 .00 % Vol
-0.02 to 200 .00 % Vol
-200.00 to 999.99 ppm Vol (depending on the unit configured under UCGAS) CURR: 0.0 to 9999.9 nA RTIME: 0.0 to 100.0 µs TEMP:
-10.0 to 140.0 °C
14.0 to 284 °F
263.1 to 413.1 K (depending on the unit configured under TEMP)
Conductivity sensor:
COND:
0.000 to 99.999 uS/cm
0.000 to 99.999 mS/cm
0.000 to 2.000 S/cm
0.000 to 99.999 uS/m
0.000 to 99.999 mS/m
0.000 to 99.999 S/m (depending on the unit configured in UCOND) RESIS:
0.00 to 999.99 kOhm*cm
0.00 to 200.00 MOhm*cm
0.00 to 999.99 kOhm*m (depending on the unit configured in URES) RAWC:
0.000 to 99.999 uS/cm
0.000 to 99.999 mS/cm
0.000 to 2.000 S/cm
0.000 to 99.999 uS/m
0.000 to 99.999 mS/m
0.000 to 99.999 S/m (depending on the unit configured in UCOND)
Endress+Hauser 25
Page 26
Commissioning Liquiline Compact CM82
Setup -> ANALYSIS menu
The CT menu and all of the associated submenus are visible only if the RIA15 was ordered with the "Analysis" option, the HART option has been configured and a CM82 has been detected by the RIA15. Using this menu, the basic settings for the CM82 can be made via the RIA15.
Parameter Values Description
TEMP:
-50.0 to 250.0 °C
-58.0 to 482.0 °F
223.1 to 523.1 K (depending on the unit configured under TEMP)
HIGH -19,999 to
99,999
ERRC 3.6 to 23.0 Configure the error current on the CM82 in mA
CDIAC Access the "CM82 - Device diagnostics" submenu
FCSM Error category as
per NAMUR and error number
DTAG Device tag Display the device tag of the CM82 (use +/- keys to scroll through
DSER Device serial
number
SENOC Sensor order
code
SENSN Sensor serial
number
CTRES Access the "CM82 -Reset" submenu
RBOOT No
YES
FDEF No
YES
CTSIM Access the "CM82 -Simulation" submenu
SIMUL OFF
ON
VALUE 3.6 to 23.0 Configure current output value on CM82 for simulation in mA
Configure turndown of current output. The measured value that corresponds to 20 mA is set here. The limits of adjustment vary depending on the sensor type and measured value. The position of the decimal point is permanently preset depending on the primary value (CMAIN) and units (UCLIQ, UCGAS, URES, UCOND) configured. For valid ranges of adjustment, see LOW (setting for 4 mA)
Display the error message with the highest priority on the CM82
text)
Display the serial number of the CM82 (use +/- keys to scroll through text)
Display the oder code of the sensor (use +/- keys to scroll through text)
Display the serial number of the sensor (use +/- keys to scroll through text)
Trigger a restart of the CM82
Reset the CM82 to factory settings
Switch on simulation for current output value on CM82
1) If "UC170" is displayed instead of unit. To remedy this, the unit must be set individually and manually in the "TEXT1" menu item. (SETUP => HART => HART1-4 => UNIT1-4 => TEXT1-4) →  26
Further information is available in the RIA15 Operating Instructions BA01170K.
"UC170" displayed instead of HART® unit
By default, the unit of the transmitted measured value is automatically read out and displayed using a HART® command. If the transmitted "unit code" cannot be uniquely assigned by the RIA15, the unit code (UC170) is displayed instead of the unit. To remedy this, the unit must be set manually. (SETUP => HART => HART1-4 => UNIT1-4 => TEXT1-4).
26 Endress+Hauser
Page 27
Liquiline Compact CM82 Commissioning
The unit codes 170 to 219 are assigned multiple times as per the HART® specification. As the UC170 is also used with the CM82, the unit must be assigned manually. This applies to the following measured values/units:
PV (TEXT1):
Transmitter parameter Primary value (CMAIN) Unit
pH Leak current (LEAKC) nA
pH Glass impedance (IMPGL) MOhm
Dissolved oxygen Liquid concentration (C_LIQ) mg/l
Dissolved oxygen Raw value of sensor (CURR) nA
QV (TEXT4):
Transmitter parameter Sensor type Unit
pH Glass MOhm
pH ISFET nA

9.5 Setting the operating language

You can change the operating language in the app settings:
Settings/User interface/Language

9.6 Date and time

Configure the date and time under System/Date/Time .
The date and time function runs only while the device is supplied with power. They must be reset if the power supply is interrupted.
(Start time: 01.01.1970 0:00 hours...)

9.7 Configuring the measuring device

Path: Application
Function Options Info
Units
Temperature unit Selection
• °C
• °F
• K
Factory setting
°C
Cond. unit Selection
• Automatic
• µS/cm
• mS/cm
• S/cm
• µS/m
• mS/m
• S/m
Factory setting
mS/cm
To be selected for conductivity
Endress+Hauser 27
Page 28
Commissioning Liquiline Compact CM82
Path: Application
Function Options Info
Conc. (liquid) unit Selection
• mg/l
• µg/l
• ppm
• ppb
Conc. (gaseous) unit
Selection
• %vol
• ppm
To be selected for oxygen.

9.7.1 Changing the device tag

You can change the device tag here:
System/Device management/TAG

9.7.2 Configuring the sensor

Path: Application
Function Options Info
Sensor Sensor-dependent settings
Sensor type Display the sensor type
Order code Order code of sensor
Damping The damping causes a floating average curve of
the measured values over the time specified.
Damping ORP, Damping pH, Damping DO, Damping conductivity
Damping temperature 0 to 60 s
Tag control
Sensor check Selection
Group Range: 0 to 65535
Extended setup
Conductivity:
Current cell constant Read only Value currently saved in the sensor
Compensation Selection
0 ... 60 s
Factory setting
0 s
Factory setting
0 s
• Off
• Tag
• Group
Factory setting
Off
• None
• Linear
• NaCl (IEC 746-3)
• Water ISO7888 (20 °C)
• Water ISO7888 (25°C)
Factory setting
Linear
These functions determine the damping of the primary value of the connected sensor.
These functions determine the damping of the integrated temperature sensor.
Various methods are available to compensate for the temperature dependency. Depending on your process, decide which type of compensation you want to use. Alternatively, you can also select None and thus measure uncompensated conductivity.
28 Endress+Hauser
Page 29
Liquiline Compact CM82 Commissioning
Path: Application
Function Options Info
Meas. ref. temp. -5.0 to 100.0 °C
(23.0 to 212.0 °F)
Factory setting
25.0 °C (77.0 °F)
Factor alpha 0.000 to 20.000 %/K
Factory setting
2.100 %/K
pH:
Temp. compensation Selection
• Off
• Automatic
• Manual
Factory setting
Automatic
Medium comp. Selection
• Off
• 2-point calibration
• Table
Factory setting
Off
Offset -18.00 to 18.00 pH
-100 to 100 mg/l
Factory setting
0.00 pH
0.00 mg/l
Internal buffer pH 0 to 14
Factory setting
pH 7.00
Oxygen:
Medium pressure Selection
• Process pressure
• Air pressure
• Altitude
• Measured value
Factory setting
Air pressure
Air pressure Choose from
Medium pressure
Salinity Selection
• Fixed value
• Measured value
Factory setting
Fixed value
Reference temperature for calculating the temperature-compensated conductivity
Enter the conductivity coefficient of your process medium
Decide how you want to compensate the medium temperature:
• Automatically using the temperature sensor of your sensor (ATC)
• Manually by entering the medium temperature
• Not at all
Take a sample from the medium and determine its pH value at different temperatures in the lab. Decide whether you want to compensate using two points or several points in a table.
The offset compensates for a difference between a laboratory measurement and an online measurement which is caused by interference ions. Enter this value manually. If you are using a compensation electrode, keep the offset at zero.
Only change the value if you are using a sensor with an internal buffer other than pH 7.
Press Measured value you can connect a pressure measured value via a fieldbus input or a current input. This measured value is then used for medium pressure compensation.
For the other types of compensation, specify a compensation value for the measurement in each case.
1. Specify either the altitude (-300 to 4000 m), the process pressure (500 to 9999 hPa) or the air pressure (500 to 1200 hPa) of the measuring point.
 The pressure used during the
calibration is also displayed for information purposes. You can change this pressure in: Calib. settings/ Medium pressure.
2. Confirm.
Specify air pressure (500 to 9999 hPa) of measuring point
Endress+Hauser 29
Page 30
Commissioning Liquiline Compact CM82
Path: Application
Function Options Info
Diagnostic settings For diagnostic settings of sensor, see
section→  39
Format settings Number of decimal places
Calibration settings
Path: Application/Sensor/Extended setup/Calibration settings
Function Options Info
pH:
Stability criteria Once the stability criterion is met, the measured
value is displayed in mV.
Delta mV 0.20 to 2.00 %
Factory setting
0 %
Duration 0 ... 60 s
Factory setting
0 s
Temp. compensation Selection
• Off
• Automatic
• Manual
Factory setting
Automatic
Buffer recognition Selection
• Fixed
• Automatic
• Manual
Factory setting
Fixed
Buffer manufacturer Selection
• Endress+Hauser
• Ingold/Mettler
• DIN 19266
• DIN 19267
• Merck/Riedel
• Hamilton
• Special buffer
Factory setting
Endress+Hauser
Calibration buffer 1 ... 2 The possible options and the factory setting
Oxygen:
1)
Measured value depending on connected sensor
Decide how you want to compensate the buffer temperature:
• Automatically using the temperature sensor of your sensor (ATC)
• Manually by entering the medium temperature
• Not at all
Fixed
You choose values from a list. This list depends on the setting for Buffer manufacturer.
Automatic
The device recognizes the buffer automatically. The recognition depends on the setting for Buffer manufacturer.
As their zero point is offset, enamel pH
sensors cannot be calibrated and adjusted with automatic buffer recognition.
Manual
You enter any two buffer values. These must differ in terms of their pH value.
Temperature tables are stored internally in the unit for the following pH values:
Endress+Hauser
2.00 / 4.00 / 7.00 / (9.00) / 9.22 / 10.00 /
12.00
Ingold/Mettler
2.00 / 4.01 / 7.00 / 9.21
DIN 19266
1.68 / 4.01 / 6.86 / 9.18
DIN 19267
1.09 / 4.65 / 6.79 / 9.23 / 12.75
Merck/Riedel
2.00 / 4.01 / 6.98 / 8.95 / 12.00
Hamilton
1.09 / 1.68 / 2.00 / 3.06 / 4.01 / 5.00 / 6.00
7.00 / 8.00 / 9.21 / 10.01 / 11.00 / 12.00
depend on the Buffer manufacturer
30 Endress+Hauser
Page 31
Liquiline Compact CM82 Commissioning
Path: Application/Sensor/Extended setup/Calibration settings
Function Options Info
Stability criteria
Delta signal 0.1 to 2.0 %
Factory setting
0.2 %
Delta temperature 0.10 to 2.00 K
Factory setting
0.50 K
Duration 5 to 60 s
Factory setting
20 s
Ambient conditions
Medium pressure Selection
• Process pressure
• Air pressure
• Altitude
Factory setting
Air pressure
Process pressure
Medium pressure = Process pressure
Air pressure
Medium pressure = Air pressure
Altitude
Medium pressure = Altitude
Rel. hum. (air variable) 0 to 100 %
Calibration timer
Function Selection
Calibration check The function checks whether the calibration of a
Function Selection
500 to 9999 hPa
Factory setting
1013 hPa
500 to 1200 hPa
Factory setting
1013 hPa
-300 to 4000 m
Factory setting
0 m
Factory setting
100 %
• Off
• On
Factory setting
Off
• Off
• On
Factory setting
Off
Permitted measured value fluctuation during calibration. Referenced to the raw value in nA in the case of amperometric sensors, and referenced to the raw value in µS in the case of optical sensors.
Permitted temperature fluctuation during calibration
Timeframe within which the permitted measured value variation may not be exceeded
Enter the altitude or the average air pressure of the place of calibration (mutually dependent values). If you specify the altitude, the average air pressure is calculated from the barometric altitude formula and vice versa. If you are compensating using the process pressure, enter the pressure in your calibration medium here. The pressure is then independent of the altitude.
sensor is still valid. Example: you install a pre­calibrated sensor. The function checks how long ago the sensor was last calibrated. A diagnostics message is displayed if the time since the last calibration is longer than specified by the predefined warning and alarm limit.
Endress+Hauser 31
Page 32
Commissioning Liquiline Compact CM82
Path: Application/Sensor/Extended setup/Calibration settings
Function Options Info
Warning limit 1 to 50 w
Factory setting
0 w
Alarm limit 1 to 50 w
Factory setting
0 w
1) Only pH sensor or pH/ORP combined sensor
Warning and alarm limits mutually influence each other's possible adjustment range.
Warning and alarm limits mutually influence each other's possible adjustment range.

9.7.3 Current output

Path: Application
Function Options Info
Current output
Output value pH, ISFET, ORP and
combined options
• pH
• Raw value pH
• Impedance glass
• Impedance reference
• rH
• ISFET Leakage current
• ORP mV
• ORP %
• Temperature
Oxygen options
• Partial pressure
• % saturation
• Conc. (liquid) unit
• Conc. (gaseous) unit
• Raw value µs
• Raw value nA
• Temperature
Conductivity options
• Conductivity
• Resistivity
• Raw value (cond. uncomp.)
• Temperature
Range lower value (4mA) The unit depends on the
Range upper value (20mA)
sensor configured.
Depends on connected sensor
Enter the measuring range. The lower and upper range values are assigned to the 3.6 mA value and the 20 mA value respectively. The system uses the engineering unit which you entered beforehand.

9.7.4 HART

Path Application/HART
Function Options Info
Bus address 0 to 63
Factory setting
0
PV value Choose from Current
output/Output value
Partial pressure Partial pressure dependence for oxygen
32 Endress+Hauser
Enter the bus address Address 1 to 63 Multidrop - mode
Primary process value
Page 33
Liquiline Compact CM82 Commissioning
Path Application/HART
Function Options Info
SV value pH, ISFET, ORP and
TV value
QV value
combined units
• pH
• Raw value pH
• Impedance glass
• Impedance reference
• rH
• ISFET Leakage current
• ORP mV
• ORP %
• Temperature
Oxygen units
• Partial pressure
• % saturation
• Conc. (liquid) unit
• Conc. (gaseous) unit
• Raw value µs
• Raw value nA
• Temperature
Conductivity units
• Conductivity
• Resistivity
• Raw value (cond. uncomp.)
• Temperature
Protocol-specific data, dynamic variables of HART communication. SV = Secondary TV = Tertiary QV = Quaternary

9.7.5 Hold

The hold state is a safe condition during configuration and calibration.
Path:System/Hold
Function Options Info
Hold release time 0 to 600 s
Factory setting
0 s
Hold behavior Selection
• None
• Freeze
• Fixed value
Factory setting
Freeze
Hold current 3.6 to 23 mA
Factory setting
0 mA
Manual hold Selection
• Off
• On
Factory setting
Off
Calibration hold Selection
• Off
• On
Factory setting
Off
The hold status is maintained for the duration of the delay time when you switch to the measuring mode.
On You can use this function to set the channel manually to "Hold". Off No channel-specific hold
During calibration, the output signal is set to "HOLD"
Endress+Hauser 33
Page 34
Commissioning Liquiline Compact CM82

9.8 Configuration management

Display the following configurations:
System/System information
• General information
• HART
System/Sensor information
• General information
• Extreme values
• Sensor operation
• Sensor specifications
• Calibration information – Temperature adjustment – Primary value

9.9 Unauthorized access

The compact transmitter is password-protected against unauthorized access.
You can change this password immediately after the password has been entered or under:
System/Bluetooth password

9.9.1 Resetting the password

The recovery password is used to resolve password problems during commissioning of the device. Data security is achieved only if the default setting for the recovery password is changed by the user.
If the user-defined password is lost, access can be restored via a recovery password.
The recovery password is the serial number of the device in reverse.
NOTICE
Forgotten recovery password.
If lost, passwords can be reset via Bluetooth only if HART is used. The device cannot be used if it does not have HART communication.
Ensure that the login and recovery password are stored in a safe place.

9.9.2 Safe signal transmission via Bluetooth® LE

Signal transmission via Bluetooth® wireless technology uses a cryptographic technique tested by the Fraunhofer Institute.
• Without the SmartBlue app, the device is not visible via Bluetooth® wireless technology.
• Only one point-to-point connection is established between a sensor and a smartphone or tablet.
• The Bluetooth® wireless technology interface can be disabled via the SmartBlue.
• Bluetooth® is optional. It can be ordered with this functionality enabled. If ordered with Bluetooth® disabled, Bluetooth® can be enabled at a later stage by means of an activation code (accessory kit) linked to the serial number.
• If the Bluetooth® interface has been disabled, it can be reactivated only via HART.

9.9.3 RIA15 locking

The device setup can be disabled with a 4-digit user code.
Further information is available in the RIA15 Operating Instructions BA01170K.
34 Endress+Hauser
Page 35
Liquiline Compact CM82 Operation

10 Operation

10.1 Reading measured values

The display of the primary values in the app is dependent on the connected sensor.
HOME view
Function
Measurement values
For pH glass, ISFET, ORP or combined sensors:
pH
Raw value pH
Impedance glass
Impedance reference
ORP mV
ORP %
rH
Temperature
For oxygen sensors:
Partial pressure
% saturation
Conc. (liquid) unit
Conc. (gaseous)
Raw value nA
Raw value µs
Temperature
For conductivity sensors:
Conductivity
Resistivity
Raw value (cond. uncomp.)
Temperature
Current output
Data relating to the transmitter:
TAG
Device type
Serial number
Firmware version
Order code

10.1.1 Changing the parameters

The device functions according to the plug & play principle of Memosens technology.
Endress+Hauser 35
Page 36
Operation Liquiline Compact CM82
However, to ensure that the settings of the previous sensor type do not get lost, the new sensor type must be selected in the software.
Path: Guidance/Measurement parameter
Function Options Info
Measurement parameter Selection
• pH
• Dissolved oxygen
• Conductivity
With "Finish" the device will be restarted and the measurement parameter change will be executed. This may take a few minutes.
Select the parameters supported by the device.
10.2 Adapting the measuring device to the process
conditions

10.2.1 Medium compensation (in the process) for oxygen

Path: Application/Sensor/Extended setup
Function Options Info
Medium pressure Selection
• Process pressure
• Air pressure
• Altitude
Salinity 0 to 40 g/kg
Factory setting
0 g/kg
Specify altitude (-300 to 4000 m), process pressure (500 to 9999 hPa) or air pressure (500 to 1200 hPa) of measuring point
The influence of salt content on oxygen measurement is compensated with this function. Example: sea water measurement as per Copenhagen Standard (30 g/kg).
36 Endress+Hauser
Page 37
Liquiline Compact CM82 Operation

10.2.2 LED settings (optical oxygen sensors only)

Path: Application/Sensor/Extended setup
Function Options Info
LED temp. mode Selection
• Off
• On
Factory setting
Off
LED temp. threshold 30 to 130 °C (86 to 266 °F)
Factory setting
80 °C (176 °F)
LED measuring interval Selection
• 1 second
• 3 seconds
• 10 seconds
• 30 seconds
Factory setting
1 second
Measurement filter Selection
Off
Weak
Normal
Strong
Very strong
Switches off the LED when the set temperature threshold is exceeded. This prevents the premature aging of the sensor cap, e.g. during a CIP or SIP cycle.
The LED measuring interval influences the response time on the one hand and the operating life of the sensor cap on the other. Shorter intervals improve the response time but reduce the operating life of the sensor cap. Make your setting depending on the requirements of your process.
Use this function to select how strong or weak the signal filtering in sensor COS81D should be.
Off
No signal filtering takes place  the recorded signals are passed through virtually unfiltered.
Weak
Signal filtering is weak.
Normal
Signal filtering is normal.
Strong
Signal filtering is strong.
Very strong
Signal filtering is very strong. Widely fluctuating raw signals are greatly attenuated by the sensor.
Endress+Hauser 37
Page 38
Diagnostics and troubleshooting Liquiline Compact CM82

11 Diagnostics and troubleshooting

11.1 Diagnostic information via LED

See LED display in Commissioning section. (→  20)

11.2 Adapting the diagnostic information

Path: Diagnostics/Diagnostic settings
Function Options Info
Sensor change alarm delay 0 ... 180 s
Factory setting
30 s
Error current 3.6 to 23.0 mA
Factory setting
22.5 s
LED shows NAMUR status signal Selection
• Off
• On
Factory setting
Off
Diagnostics behavior The list of diagnostic messages
Status signal The messages are divided into
Diagnostics behavior Selection
• Warning
• Alarm
Sensor HOLD
Possible range of error current.
* Additional LED signals for diagnostic messages as per NAMUR NE107 categories.
displayed. There are device­specific messages, and messages that depend on what sensor is connected. Select the message to be adapted. Only then can you make the settings for this message.
different error categories in accordance with NAMUR NE
107.
* LED as per NAMUR NE107 categories:
Three rapid green flashes at the start of the message means: Everything OK - but pay attention!
The more red flashes there are at the end of a message, the more critical the diagnosis as per NE107. Continuous red only flashing means: Error in device or sensor, take action immediately.
LED behavior Status
Three rapid green flashes and a single rapid red flash
Three rapid green flashes and two rapid red flashes Device and sensor are being operated out of specification.
Three rapid green flashes and three rapid red flashes
Red Flashes quickly
Device or sensor requires maintenance. M status signal as per NAMUR NE107
S status as per NAMUR NE107
Device or sensor undergoing function check. C status signal as per NAMUR NE107
Failure of device or sensor F status signal as per NAMUR NE107
38 Endress+Hauser
Page 39
Liquiline Compact CM82 Diagnostics and troubleshooting

11.3 Adapting sensor diagnostic information

This menu branch is used for specifying warning limits, and for defining whether and how diagnostics tools should be used.

11.3.1 Impedance monitoring

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Glass impedance
Upper limit Selection
• Off
• On
Factory setting
Off
Upper limit Selection
• Off
• On
Factory setting
On
Upper alarm limit 0 to 10000 MΩ
Factory setting
3000 MΩ
Upper warning limit 0 to 10000 MΩ
Factory setting
2500 MΩ
Lower limit Selection
• Off
• On
Factory setting
Off
Lower limit Selection
• Off
• On
Factory setting
On
Lower warning limit 0 to 10000 MΩ
Factory setting
0.1 MΩ
Lower alarm limit 0 to 10000 MΩ
Factory setting
0 MΩ
On The Sensor Check System (SCS) operates with the following settings for the upper warning and alarm limits. Off Monitoring of the upper warning and alarm limits is switched off.
On
The Sensor Check System (SCS) operates with the following settings for the upper warning and alarm limits.
Off
Monitoring of the upper warning and alarm limits is switched off.
Diagnostics code and associated message text: 124 Sensor glass
Diagnostics code and associated message text: 125 Sensor glass
On The Sensor Check System (SCS) operates with the following settings for the lower warning and alarm limits. Off Monitoring of the lower warning and alarm limits is switched off.
On
The Sensor Check System (SCS) operates with the following settings for the lower warning and alarm limits.
Off
Monitoring of the lower warning and alarm limits is switched off.
Diagnostics code and associated message text: 123 Sensor glass
Diagnostics code and associated message text: 122 Sensor glass
Endress+Hauser 39
Page 40
Diagnostics and troubleshooting Liquiline Compact CM82

11.3.2 Slope

pH, oxygen
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
pH
Slope The slope characterizes the
sensor condition. The greater the deviation from the ideal value (pH), the worse the condition of the sensor.
Warning limit 5.00 to 99.00 mV/pH
Factory setting
55.00 mV/pH
Oxygen
Upper warning limit 0.0 to 200.0 %
Factory setting
140.0 %
Lower warning limit 0.0 to 200.0 %
Factory setting
60.0 %
Specify your limit values for slope monitoring. Associated diagnostic code and message text: 509 sensor calibration
Associated diagnostics code and message text: 511 Sensor
calibration
Associated diagnostics code and message text: 509 Sensor
calibration

11.3.3 Delta slope

pH, pH/ORP combined sensor, oxygen
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
pH and pH/ORP combined sensors
Delta slope The device determines the
difference in slope between the last calibration and the penultimate calibration, and issues a warning or an alarm depending on the setting configured. The difference is an indicator for the condition of the sensor. The greater the change, the greater the wear experienced by the pH-sensitive glass membrane as a result of chemical corrosion or abrasion.
Function Selection
• Off
• On
Warning limit 0.10 to 10.00 mV/pH
Factory setting
5.00 mV/pH
Oxygen
Switches the function on or off
Specify your limit values for monitoring the slope differential. Associated diagnostics code and message text: 518 Sensor
calibration
40 Endress+Hauser
Page 41
Liquiline Compact CM82 Diagnostics and troubleshooting
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Delta slope The device determines the
difference in slope between the last calibration and the penultimate calibration, and issues a warning or an alarm depending on the setting configured. The difference is an indicator for the condition of the sensor. An increasing change indicates the formation of buildup on the sensor diaphragm or electrolyte contamination. Replace the diaphragm and electrolyte as specified in the instructions in the sensor operating manual.
Function Selection
• Off
• On
Factory setting
Off
Warning limit 0.0 to 50.0 %
Factory setting
5.0 %
Switches the function on or off
Specify your limit values for monitoring the slope differential. Associated diagnostics code and message text: 518 Sensor
calibration

11.3.4 Zero point and operating point

pH, ISFET, oxygen
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
pH, ISFET
Zero point (pH glass) Operating point (ISFET)
Upper warning limit Lower warning limit ...
pH 12.00 Lower warning limit ... 950 mV
1)
2)
Factory setting
pH 8.00 / 300 mV
Lower warning limit pH 2.00 to Upper warning
limit
-950 mV to Upper warning limit
2)
Factory setting
pH 6.00 / 300 mV
Oxygen
The zero point or operating point characterizes the condition of the sensor reference. The bigger the deviation from the ideal value (pH 7.00) the poorer the condition. This can be caused by KCl dissolving away or reference contamination, for example.
Associated diagnostics code and message text: 505 Sensor calibration 515 Sensor calibration
Associated diagnostics code and message text: 507 Sensor calibration 517 Sensor calibration
2)
2)
Endress+Hauser 41
Page 42
Diagnostics and troubleshooting Liquiline Compact CM82
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Zero point The zero point corresponds to
the sensor signal that is measured in a medium in the absence of oxygen. You can calibrate the zero point in water that is free from oxygen or in high-purity nitrogen. This improves accuracy in the trace range.
Warning limit 0.0 to 10.0 nA
Factory setting
3.0 nA
1) pH Glass
2) pH ISFET
Specify the limit values for zero point monitoring in your sensor. Associated diagnostics code and message text: 513 Zero
Warning

11.3.5 Delta zero point/operating point

pH, ISFET, oxygen
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
pH, ISFET
Delta zero point The device determines the
difference between the last calibration and the penultimate calibration, and issues a warning or an alarm depending on the setting configured. The difference is an indicator for the condition of the sensor.
Function Selection
• Off
• On
Warning limit pH 0.00 to 2.00 (pH glass)
Factory setting
pH 0.50 / 25 mV
Oxygen
Function Selection
• Off
• On
Factory setting
Off
Warning limit 0.0 to 10 nA
Factory setting
1.0 nA
Switches the function on or off
Specify your limit values for monitoring the slope differential.
Associated diagnostics code and message text:
• 520 Sensor calibration (pH glass)
• 522 Sensor calibration (ISFET)
Switches the function on or off
Specify your limit values for monitoring the slope differential.
Associated diagnostics code and message text: 520 Sensor
calibration
42 Endress+Hauser
Page 43
Liquiline Compact CM82 Diagnostics and troubleshooting

11.3.6 Operating hours limits

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Limits operating hours The total operating time of the
sensor and its use under extreme conditions is monitored. If the operating time exceeds the defined threshold values, the device issues a corresponding diagnostics message.
Function Selection
• Off
• On
Operating time Factory setting
60000 h
Operating time > 80 °C Factory setting
50000 h
Operating time > 80 °C < 100 nS/cm Factory setting
30000 h
Operating time > 100 °C Factory setting
50000 h
Operating time > 120 °C Factory setting
10000 h
Operating time > 150 °C Factory setting
10000 h
Operating time < -300 mV Factory setting
60000 h
Operating time > 300 mV Factory setting
60000 h
On
The operation of the sensor under extreme conditions is monitored, recorded in the sensor and diagnostics messages are displayed on the controller.
Off
No diagnostics messages. However, the time the sensor operates under extreme conditions is recorded in the sensor and can be read in the sensor information in the diagnostics menu.
Diagnostics code and associated message text: 193 Operating
time
Only conductive sensors
Diagnostics code and associated message text: 194 Operating
time
Diagnostics code and associated message text: 195 Operating
time
Diagnostics code and associated message text: 198 Operating
time
Only pH sensor or pH/ORP combined sensor
Only pH sensor or pH/ORP combined sensor
Endress+Hauser 43
Page 44
Diagnostics and troubleshooting Liquiline Compact CM82

11.3.7 Sterilizations

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Sterilizations The system counts the number
of operating hours in which the sensor is exposed to a temperature that is typical for a sterilization. This temperature depends on the sensor.
Function Selection
• Off
• On
Warning limit 0 to 1000
Factory setting
800
Cap sterilizations (only sterilizable sensors)
Switches the function on or off
Specify the limit value for the number of sensor sterilizations. Diagnostics code and associated message text: 108 SIP, CIP,
autoclaving
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
No. sterilizations cap Not displayed for optical oxygen
sensors. The sterilization counters in the sensor make a distinction between the sensor and the membrane/fluorescence cap currently used. If this cap is replaced, only the (cap) counter is reset.
Function Selection
• Off
• On
Factory setting
Off
Warning limit 0 to 100
Factory setting
25
Specify how many sterilizations may be performed with a membrane cap before the cap has to be replaced. The number depends heavily on the process and must be determined individually.
Associated diagnostics code and message text: 109 Sterilization
cap
44 Endress+Hauser
Page 45
Liquiline Compact CM82 Diagnostics and troubleshooting

11.3.8 Sensor Condition Check (SCC)

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Sensor condition check Sensor condition check (SCC)
monitors the electrode status and the degree of electrode aging. The condition of the electrode is updated after every calibration.
The main reasons for a deteriorating electrode status are:
• Glass membrane blocked or dry
• Diaphragm (reference) blocked
Function Selection
• Off
• On
Switches the function on or off Diagnostics code and associated message text: 127 SCC adequate 126 SCC poor

11.3.9 Process monitoring

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Process check system The process check system (PCS)
checks the measuring signal for stagnation. An alarm is triggered if the measuring signal does not change over a specific period (several measured values).
Function Selection
Duration 0 to 240 min Once this time has elapsed, the
Tolerance width
Not for pH/ORP sensors
• Off
• On
The range depends on the sensor
Factory setting
Depends on the sensor
Switches the function on or off
calibration timer diagnostic message, along with the code 102, appears on the display.
Interval around the measuring signal (raw value) for detecting stagnation. Measured values within the set interval are regarded as stagnating.

11.3.10 Measured value

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
ORP-Meas value
Function Selection
• Off
• On
Upper alarm value 0 to 10000 mV Diagnostics code and associated
Endress+Hauser 45
Switches the function on or off
message text: 124 Sensor glass
Page 46
Diagnostics and troubleshooting Liquiline Compact CM82
Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Upper warning limit 0 to 10000 mV Diagnostics code and associated
message text: 125 Sensor glass
Lower warning limit 0 to 10000 mV Diagnostics code and associated
message text: 123 Sensor glass
Lower alarm value 0 to 10000 mV Diagnostics code and associated
message text: 122 Sensor glass

11.3.11 Pharmaceutical water

Path: Application/Sensor/Extended setup/Diagnostic settings
Pharma water Here you can make settings for
monitoring pharmaceutical water in accordance with the United States Pharmacopoeia (USP) or European Pharmacopoeia (EP). The uncompensated conductivity value and the temperature are measured for the limit functions. The measured values are compared against the tables defined in the standards. An alarm is triggered if the limit value is exceeded. Furthermore, you can also set a preliminary alarm (warning limit) which signals undesired operating states before they occur.
Function Selection
• Off
• EP
• USP
Factory setting
Off
Off 10.0 to 99.9 %
Factory setting
80.0 %
The alarm values are stored in the device in accordance with USP <645> or EP <169> specifications. You define the warning limit as a % of the alarm value.
Diagnostics code and associated message text: 915 USP / EP
warning
If the value exceeds the USP or EP alarm values saved in the software, diagnostics message 914 USP/ EP alarm is displayed.
46 Endress+Hauser
Page 47
Liquiline Compact CM82 Diagnostics and troubleshooting

11.3.12 Cap calibrations

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
No. calibrations cap Not displayed for optical oxygen
sensors The calibration counters in the sensor make a distinction between sensor calibrations and calibrations with the membrane cap currently used. If this cap is replaced, only the (cap) counter is reset.
Function Selection
• Off
• On
Factory setting
Off
Warning limit 0 to 1000
Factory setting
6
Specify how many calibrations may be performed with a membrane cap before the cap has to be replaced. The number depends heavily on the process and must be determined individually.
Associated diagnostics code and message text: 535 Sensor
check

11.3.13 Calibration quality index (optical sensors only)

Path: Application/Sensor/Extended setup/Diagnostic settings
Function Options Info
Cal. quality index Monitoring of calibration
quality index for COS81D. Significant changes in the value may be a sign of spot aging or poor calibration.
Function Selection
• Off
• On
Factory setting
Off
Warning limit 0 ... 100 %
Factory setting
80 %
Switches the function on or off
Associated diagnostics code and message text: 734 Calibration
quality

11.4 Simulation

You can simulate certain parameters for test purposes:
• Current value
• Measured value
• Temperature
Endress+Hauser 47
Page 48
Diagnostics and troubleshooting Liquiline Compact CM82
Main menu/Diagnostics/Simulation
Function Options Info
Current output Simulation of an output current
Simulation Selection
• Off
• On
Factory setting
Off
Simulation value Unit:
Window:
Measurement value Simulation of a measured value
Simulation Selection
• Off
• On
Factory setting
Off
Measured value Select measured variable to be
Simulation value Unit:
Window:
Temperature Simulation of temperature
Simulation Selection
• Off
• On
Factory setting
Off
Simulation value Unit:
Window:
Switch on or off simulation
Set current value
Switch on or off simulation
simulated, depending on connected sensor.
Display the simulated measured value in the selected unit.
Switch on or off simulation
Display the simulated temperature value in the selected unit

11.5 Diagnostic list

11.5.1 Diagnostic messages

In accordance with Namur specification NE 107, the diagnostic messages are characterized by:
• Message number
• Error category (letter in front of the message number) – F = (Failure) a malfunction has been detected – C = (Function check), (no error)
Maintenance work is being performed on the device. Wait until the work has been completed.
S = (Out of specification), the measuring point is being operated outside its
specification Operation is still possible. However, you run the risk of increased wear, shorter operating life or lower measurement accuracy. The cause of the problem is to be found outside the measuring point.
M = (Maintenance required), action should be taken as soon as possible
• Sensor type: – P = ph – C = conductivity – O = oxygen
• Message text
48 Endress+Hauser
Page 49
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
type
1)
S
002 Sensor
unknown
004 Sensor defective F On On All
005 Sensor data
invalid
010 Sensor scanning F Off Off All No
012 Writing data
failed
013 Sensor type
wrong
018 Sensor not
ready
022 Temperature
sensor
061 Sensor
electronic
062 Sensor
connection
100 Sensor
communication
102 Calibration
timer
104 Calibration
validity
F On On All No
F On On All No 1. Check the firmware
F On On All No 1. Repeat write process.
F On On All No
F On On All No Sensor communication
F On On All Yes Temperature sensor defective
F On On All No Sensor electronics defective
F On On All No 1. Check sensor
F On On All No Sensor not communicating
M On Off All Yes Calibration interval elapsed.
M On Off All Yes Last calibration no longer
2)
D
3)
F
ConfigurableTests or remedial action
Replace sensor.
compatibility of the sensor and transmitter or load suitable firmware
2. Set the sensor to the factory settings, disconnect the sensor and reconnect it.
3. Update transmitter data
4. Replace sensor.
Wait for initialization to
be finished.
2. Replace sensor.
Replace sensor, making
sure correct sensor type is used.
blocked
1. Sensor fails tag check. Replace.
2. Internal software error. Contact the Service Department.
Replace sensor.
Replace sensor.
connection.
2. Contact the Service Department.
1. Check sensor connection.
2. Check sensor connector.
3. Contact the Service Department.
Measurement can still take place.
Calibrate sensor.
valid. Measurement can still take place.
Calibrate sensor.
Endress+Hauser 49
Page 50
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
type
1)
S
105 Calibration
validity
106 Sensor TAG F On On All No Sensor has invalid tag or tag
107 Calibration
active
108 SIP, CIP,
autoclaving
109 Sterilization cap M On Off O No Specified number of
111 Operating time
cap
118 Sensor glass
break.
119 Sensor check M On Off P (glass) No
120 Sensor
reference
121 Sensor
reference
M On Off All Yes Last calibration will lose its
C On Off All No
M On Off O Yes Specified number of
M On Off O No Hours of operation
F On Off P (glass) Yes Glass breakage warning,
F On Off P (glass) Yes Reference warning,
M On Off P (glass) No
2)
D
3)
F
ConfigurableTests or remedial action
validity soon. Measurement can still take place.
Calibrate sensor.
group
Wait for calibration to be
finished.
sterilizations will soon be reached. Measurement can still take place.
Replace sensor.
sterilizations for the cap is reached. Measurement can still take place.
Replace membrane cap.
monitoring
The limit set for the total hours of operation for the cap has been reached. Measurement can still take place.
1. Replace the cap.
2. Change monitoring limit.
impedance of pH glass too low
Measuring can continue until the alarm (118) occurs.
1. Inspect sensor for hair-line cracks and breakage.
2. Check medium temperature.
3. Replace sensor.
impedance of reference too low
Measuring can continue until the alarm (120) occurs.
1. Check reference for clogging/ contamination.
2. Clean reference/ junction.
3. Replace sensor.
50 Endress+Hauser
Page 51
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
1)
S
2)
D
3)
F
122 Sensor glass F On Off P (glass) Yes Impedance limit values
123 Sensor glass M On Off P (glass) Yes
124 Sensor glass M On Off P (glass) Yes
125 Sensor glass F On Off P (glass) Yes
exceeded/undershot
Measuring can continue until the alarm (122, 124) occurs.
1. Inspect sensor for hair-line cracks and breakage.
2. Check or change limit values.
3. Replace sensor.
126 Sensor check M On Off P (glass) No Sensor condition check (SCC),
poor sensor condition
Glass membrane fouled or dry, junction blocked
1. Clean sensor, regenerate
2. Replace sensor.
127 Sensor check M On Off P (glass) No Sensor condition check (SCC),
adequate sensor condition
128 Sensor leakage F On Off P (ISFET),
O
Yes Leak current alarm
Defective due to abrasion or damage Damage to the gate (only ISFET)
Replace sensor.
129 Sensor leakage F On Off P (ISFET),
DO
Yes Leak current warning
Measuring can continue until the alarm occurs
130 Sensor supply F On Off P, O Yes Poor sensor power supply
1. Check sensor connection.
2. Replace sensor.
131 Sensor
calibration
132 Sensor
calibration
M On Off O No Limit values for sensor
relaxation time (fluorescence
M On Off O
decay time) exceeded/ undershot
Reasons: high oxygen content, incorrect calibration
1. Repeat calibration.
2. Replace sensor cap.
3. Contact the Service Department.
133 Sensor signal F On Off O No No signal (fluorescence
decay)
1. Replace sensor cap.
2. Contact the Service Department.
Endress+Hauser 51
Page 52
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
type
1)
S
134 Sensor signal M On Off O No Low signal amplitude.
136 Sensor temp.
high
141 Polarization F On Off C No Polarization warning
142 Sensor signal F On Off C No Reasons: sensor in air, sensor
146 Sensor
temperature
154 Sensor data
invalid
160 Sensor data
invalid
164 Sensor data
invalid
168 Polarization S On Off C (cond.) No Polarization warning
S On Off O No Temperature outside
S Off Off All Yes Temperature outside
M Off Off C No Factory calibration is used
F On Off All No No calibration data
M Off Off C No No temperature calibration
2)
D
3)
F
ConfigurableTests or remedial action
Measurement can still take place.
1. Replace sensor cap.
2. Contact the Service Department.
specification
1. Check process.
2. Check installation.
The measured value is corrupted at high conductivity levels.
Use a sensor with a
larger cell constant.
defective
1. Check installation.
2. Replace sensor.
specification
1. Check the temperature.
2. Check electrode system.
3. Replace sensor type.
Calibrate.
Reasons: data deleted
1. Select other data record.
2. Use factory calibration.
3. Contact the Service Department.
data
Factory calibration is used
1. Check process.
2. Check or replace sensor.
The measured value is corrupted at high conductivity levels.
Use a sensor with a
larger cell constant.
52 Endress+Hauser
Page 53
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
1)
S
2)
D
3)
F
178 Operating time M On Off All No Operating hours > 15 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
179 Operating time M On Off P No Operating hours > 300 mV,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
180 Operating time M On Off P No Operating hours < -300 mV,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
183 Operating time M On Off O (amp.) No Operating hours > 10 nA
(COS51D), measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
184 Operating time M On Off O (amp.) No Operating hours > 30 nA
(COS22D), measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
185 Operating time M On Off O (amp.) No Operating hours > 40 nA
(COS51D), measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
186 Operating time M On Off O (amp.) No Operating hours > 160 nA
(COS22D), measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
Endress+Hauser 53
Page 54
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
1)
S
2)
D
3)
F
187 Operating time M On Off C No Operating hours > 80 °C, 100
nS/cm, measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
189 Operating time M On Off O No Operating hours > 5 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
191 Operating time M On Off O No Operating hours > 30 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
192 Operating time M On Off O No Operating hours > 40 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
193 Operating time M On Off P, C, O No Operating hours > 80 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
194 Operating time M On Off P No Operating hours > 100 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
195 Operating time M On Off C No Operating hours > 120 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
54 Endress+Hauser
Page 55
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
1)
S
2)
D
3)
F
197 Operating time M On Off C No Operating hours > 140 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
198 Operating time M On Off C No Operating hours > 150 °C,
measurement can still take place
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
199 Operating time M On Off All No Total operating hours
202 Self-test active F On On All No Self-test active, please wait
215 Simulation
active
C On Off All No Simulation active
End simulation by changing to measuring mode.
216 Hold active C On Off All No HOLD active - output values
and channel status are on hold
Please wait
241 Firmware error F On On All No Software error - internal
1. Carry out software update
2. Replace backplane please
3. Please contact the Service Department and quote the number shown
243 Firmware error F On On All No Software error - internal
1. Carry out software update
2. Replace backplane please
3. Please contact the Service Department and quote the number shown
284 Firmware
update
F On On All No Firmware update active,
please wait.
285 Update error F On On All No Firmware update failed,
possible reasons:
• Read error
• SD card is faulty
• Incorrect firmware on SD card
1. Check device settings
2. Check electronics module type
Endress+Hauser 55
Page 56
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
type
1)
S
373 Temperature of
electronics is high
384 Firmware error F On On All No Sensor measuring sequence
408 Calibration
aborted
411 Up-/Download
active, please wait
460 Output
undershot
461 Output overshot S On Off All No Current output overshot
500 Sensor
calibration
M On Off All No Temperature of backplane
M Off Off P, C, O, No Calibration aborted
C On Off All No Up-/download active
S On Off All No Current output undershot
M On Off All No Calibration aborted, main
2)
D
3)
F
ConfigurableTests or remedial action
electronics is high
Check ambient
temperature and energy consumption
counter. No measurement signal from sensor
1. Update software
2. Contact the Service Department
Please wait
Measured value out of specified current range, possible reasons:
• Sensor / sample line in air
• Air pockets in the assembly
• Incorrect flow to sensor
• Sensor / sample line is contaminated
1. Check sensor.
2. Check application.
3. Check application.
4. Clean sensor / sample line.
Measured value out of specified current range, possible reasons:
• Sensor / sample line in air
• Air pockets in the assembly
• Incorrect flow to sensor
• Sensor / sample line is contaminated
1. Check sensor.
2. Check application.
3. Check application.
4. Clean sensor / sample line.
measured value varies
Reasons: sensor too old, sensor occasionally dry, calibration value not constant
1. Check sensor.
2. Check calibration solution.
56 Endress+Hauser
Page 57
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
type
501 Sensor
1)
S
M On Off All No Calibration aborted,
2)
D
3)
F
calibration
505 Sensor
M On Off P, O No Max. zero point warning,
calibration
507 Sensor
M On Off P, O No Min. zero point warning,
calibration
509 Sensor
M On Off P, O No Min. slope warning,
calibration
ConfigurableTests or remedial action
temperature measured value varies
Reasons: sensor too old, sensor occasionally dry, temperature of calibration solution not constant
1. Check sensor.
2. Regulate calibration solution temperature.
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
Endress+Hauser 57
Page 58
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
511 Sensor
calibration
1)
S
M On Off P, O No Max. slope warning,
2)
D
3)
F
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
513 Zero Warning M On Off O (amp.) No Zero point warning,
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
515 Sensor
calibration
M On Off P (ISFET) No Max. operating point
warning, measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
517 Sensor
calibration
M On Off P (ISFET) No Min. operating point
warning, measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
58 Endress+Hauser
Page 59
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
518 Sensor
calibration
1)
S
M On Off P, O No Delta slope warning,
2)
D
3)
F
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
520 Sensor
calibration
M On Off P, O No Delta zero point warning,
measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
522 Sensor
calibration
M On Off P (ISFET) No Delta operating point
warning, measurement can still take place
Possible reasons: sensor old or defective, reference blocked, calibration solution too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
532 License error M On Off All No License error
534 Electrolyte
warning
M On Off All No Electrolyte depletion warning
The configured electrolyte depletion limit has been reached. Measurement is still possible.
1. Replace electrolyte and, if necessary, the membrane cap.
2. Reset counter for CAL
3. Replace electrolyte, or replace sensor cap and electrolyte.
4. Replace sensor
535 Sensor check M On Off O (amp.) No Specified number of cap
calibrations is reached Measurement can still take place.
Replace sensor cap.
Endress+Hauser 59
Page 60
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
type
550 Process
temperature
551 Process
temperature
552 Conductivity
low
553 Conductivity
high
554 Concentration
low
555 Concentration
high
722 Sensor
reference
723 Sensor
reference
724 Sensor
reference
725 Sensor
reference
1)
S
S On Off C No Process temperature above/
S On Off C
S On Off C No Process concentration above/
S On Off C
S On Off C No Process concentration above/
S On Off C
F On On P Yes Alarm: Reference membrane
M On Off All Yes Warning: Reference
F On On All Yes Alarm: Reference membrane
M On Off All Yes Warning: Reference
2)
D
3)
F
ConfigurableTests or remedial action
below concentration table
• Process value outside specification
• Table incomplete
Extend table.
below concentration table
• Process value outside specification
• Table incomplete
Extend table.
below concentration table
• Process value outside specification
• Table incomplete
Extend table.
impedance too low.
1. Check or replace sensor.
2. Check/correct reference limit value.
membrane impedance too low.
Can continue measuring until the alarm occurs.
1. Check or replace sensor.
2. Check/correct reference limit value.
impedance too high.
1. Check or replace sensor.
2. Check/correct reference limit value.
membrane impedance too high.
Can continue measuring until the alarm occurs.
1. Check or replace sensor.
2. Check/correct reference limit value.
60 Endress+Hauser
Page 61
Liquiline Compact CM82 Diagnostics and troubleshooting
No. Message Factory settings Sensor
ConfigurableTests or remedial action
type
734 Calibration
quality
1)
S
M On Off All No Min. Calibration quality
2)
D
3)
F
warning Measurement is still possible. There is a significant change in calibration quality since the last calibration.
1. Repeat the calibration
2. Check sensor and replace if necessary
740 Sensor defective F On On C No Internal electrode failure
1. Replace sensor.
2. Contact the Service Department.
832 Temp. range
exceeded
S Off Off All Yes Outside temperature
specification
1. Check application.
2. Check temperature sensor.
841 Operating range S Off Off All Yes Process value outside
operational range
1. Check application.
2. Check sensor.
842 Process value S Off Off P Yes Process limit value exceeded/
843 Process value S Off Off P
undershot
Reasons: sensor in air, air pockets in assembly, incorrect flow to sensor, sensor defective
1. Change process value.
2. Check electrode system.
3. Change sensor type.
904 Process check
alarm
F On On All No Stagnating measuring signal
Reasons: sensor in air, sensor fouling, incorrect flow to sensor, sensor defective
1. Check electrode system.
2. Check sensor.
3. Restart the device.
910 Limit switches S On Off No Limit switch activated
914 USP/ EP alarm M On Off C Yes USP limit values exceeded
Check process.
915 USP / EP
M On Off C
warning
942 Process value S Off Off P No Process value high
1. Do not increase process value.
2. Check electrode system.
3. Change sensor type.
Endress+Hauser 61
Page 62
Diagnostics and troubleshooting Liquiline Compact CM82
No. Message Factory settings Sensor
type
1)
S
943 Process value S Off Off P No Process value low
987 Calibration
required
1) Status signal
2) Diagnostic message
3) Error current
M On On I, DI No Electrode replacement
2)
D
3)
F
ConfigurableTests or remedial action
1. Do not decrease process value.
2. Check electrode system.
3. Change sensor type.
Calibrate sensor.

11.6 Event logbook

Main menu/Diagnostics/Diagnostics logbook
Function Info
Time Time when diagnostic message occurred.
Event Indicate whether message is going or coming.
Status signal Error category and fault elimination

11.7 Resetting the measuring device

Path: System/Device management/Reset
Function Options Info
Device restart Press Finish to start the
application Press X to close the wizard without running it and to go back.
Factory default Press Finish to start the
application Press X to close the wizard without running it and to go back.
Restart and keep all the settings
Restart with factory settings Settings that have not been saved are lost.
62 Endress+Hauser
Page 63
Liquiline Compact CM82 Diagnostics and troubleshooting

11.8 Device information

11.8.1 Squawk

Main menu/system
Path: System/Device management/Squawk
Function Options Info
Squawk Selection
• Off
• On
Factory setting
Off
Squawk is signaled briefly while the connection is established. The app can also be used to enable Squawk. This allows the device to be located more quickly in large installations.

11.9 Firmware history

Date Version Changes to firmware Documentation
02/2018 01.01.00 Original firmware BA01845C/07/EN/01.18

11.9.1 Firmware update

Information on available firmware updates can be obtained from your app store or sales office.
The current firmware version and device type can be found under: System/Firmware
update
Endress+Hauser 63
Page 64
Maintenance Liquiline Compact CM82

12 Maintenance

The maintenance of the measuring point comprises:
• Calibration
• Cleaning the controller, assembly and sensor
• Checking the cables and connections.
WARNING
L
Process pressure and temperature, contamination
Risk of serious or fatal injury
If the sensor has to be removed during maintenance work, avoid hazards posed by
pressure, temperature and contamination.
NOTICE
Electrostatic discharge (ESD)
Risk of damaging the electronic components
Take personal protective measures to avoid ESD, such as discharging beforehand at PE
or permanent grounding with a wrist strap.

12.1 Maintenance tasks

12.1.1 Cleaning

NOTICE
Cleaning agents not permitted
Damage to housing surface and optical waveguide
Never use concentrated mineral acids or alkaline solutions for cleaning.
Never use organic cleaners such as acetone, benzyl alcohol, methanol, methylene
chloride, tetrahydrofuran, xylene or concentrated glycerol cleaner.
The device is resistant to:
• Ethanol (for a short time)
• Diluted acids (max. 2% HCl)
• Diluted bases (max. 3% NaOH)
• Soap-based household cleaning agents
• Washing-up liquid

12.1.2 Calibration

To calibrate the sensor, remove it from the medium and calibrate it in the laboratory. Since Memosens sensors save the data, you can always work with "precalibrated" sensors and do not have to stop monitoring the process to perform a calibration.
1. Menu: Guidance/Calibration menu item.
2. Select calibration type.
3. Follow the instructions of the software.
4. Return the sensor to the medium.  This deactivates the hold and the system starts measuring again.
You can cancel the calibration by pressing X. No data are then used to adjust the sensor.
64 Endress+Hauser
Page 65
Liquiline Compact CM82 Repair

13 Repair

13.1 General notes

Only use spare parts from Endress + Hauser to guarantee the safe and stable
functioning of the device.
Detailed information on the spare parts is available at:
www.endress.com/device-viewer

13.2 Return

The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
To ensure the swift, safe and professional return of the device:
Refer to the website www.endress.com/support/return-material for information on the
procedure and conditions for returning devices.

13.3 Disposal

The device contains electronic components. and must therefore be disposed of in accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
Endress+Hauser 65
Page 66
Accessories Liquiline Compact CM82

14 Accessories

14.1 Sensors

14.1.1 Glass electrodes

Orbisint CPS11D
• pH electrode for process technology
• Optional SIL version for connecting to SIL transmitter
• With dirt-repellent PTFE diaphragm
• Product Configurator on the product page: www.endress.com/cps11d
Technical Information TI00028C
Memosens CPS31D
• pH electrode with gel-filled reference system with ceramic diaphragm
• Product Configurator on the product page: www.endress.com/cps31d
Technical Information TI00030C
Ceraliquid CPS41D
• pH electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps41d
Technical Information TI00079C
Ceragel CPS71D
• pH electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps71d
Technical Information TI00245C
Memosens CPS171D
• pH electrode for bio-fermenters with digital Memosens technology
• Product Configurator on the product page: www.endress.com/cps171d
Technical Information TI01254C
Orbipore CPS91D
• pH electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps91d
Technical Information TI00375C
Orbipac CPF81D
• Compact pH sensor for installation or immersion operation
• In industrial water and wastewater
• Product Configurator on the product page: www.endress.com/cpf81d
Technical Information TI00191C

14.1.2 Enamel pH electrodes

Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Product Configurator on the product page: www.endress.com/cps341d
Technical Information TI00468C
66 Endress+Hauser
Page 67
Liquiline Compact CM82 Accessories

14.1.3 ORP sensors

Orbisint CPS12D
• ORP sensor for process technology
• Product Configurator on the product page: www.endress.com/cps12d
Technical Information TI00367C
Ceraliquid CPS42D
• ORP electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps42d
Technical Information TI00373C
Ceragel CPS72D
• ORP electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps72d
Technical Information TI00374C
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Product Configurator on the product page: www.endress.com/cpf82d
Technical Information TI00191C
Orbipore CPS92D
• ORP electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps92d
Technical Information TI00435C

14.1.4 pH ISFET sensors

Tophit CPS441D
• Sterilizable ISFET sensor for low-conductivity media
• Liquid KCl electrolyte
• Product Configurator on the product page: www.endress.com/cps441d
Technical Information TI00352C
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
• Water treatment and biotechnology
• Product Configurator on the product page: www.endress.com/cps471d
Technical Information TI00283C
Tophit CPS491D
• ISFET sensor with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps491d
Technical Information TI00377C
Endress+Hauser 67
Page 68
Accessories Liquiline Compact CM82

14.1.5 pH and ORP combined sensors

Memosens CPS16D
• Combined pH/ORP sensor for process technology
• With dirt-repellent PTFE diaphragm
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps16D
Technical Information TI00503C
Memosens CPS76D
• Combined pH/ORP sensor for process technology
• Hygienic and sterile applications
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps76d
Technical Information TI00506C
Memosens CPS96D
• Combined pH/ORP sensor for chemical processes
• With poison-resistant reference with ion trap
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps96d
Technical Information TI00507C

14.1.6 Conductivity sensors with conductive measurement of conductivity

Condumax CLS15D
• Conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• Product Configurator on the product page: www.endress.com/CLS15d
Technical Information TI00109C
Condumax CLS16D
• Hygienic, conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• With EHEDG and 3A approval
• Product Configurator on the product page: www.endress.com/CLS16d
Technical Information TI00227C
Condumax CLS21D
• Two-electrode sensor in plug-in head version version
• Product Configurator on the product page: www.endress.com/CLS21d
Technical Information TI00085C
Memosens CLS82D
• Four-electrode sensor
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cls82d
Technical Information TI01188C
68 Endress+Hauser
Page 69
Liquiline Compact CM82 Accessories

14.1.7 Oxygen sensors

Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• With Memosens technology or as an analog sensor
• Product Configurator on the product page: www.endress.com/cos22d
Technical Information TI00446C
Oxymax COS51D
• Amperometric sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos51d
Technical Information TI00413C
Memosens COS81D
• Sterilizable, optical sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos81d
Technical Information TI01201C

14.2 Software

Memobase Plus CYZ71D
• PC software to support laboratory calibration
• Visualization and documentation of sensor management
• Sensor calibrations stored in database
• Product Configurator on the product page: www.endress.com/cyz71d
Technical Information TI00502C

14.3 Other accessories

14.3.1 Activation codes

You must quote the serial number of the device when ordering the activation code.
Activation code: Bluetooth
Order No. 71401176

14.3.2 Cable junction with Velcro strip

Cable junction with Velcro strip
• 4 pieces, for sensor cable
• Order No. 71092051

14.3.3 Communication-specific accessories

Commubox FXA195
Intrinsically safe HART communication with FieldCare via the USB port
Technical Information TI00404F
Endress+Hauser 69
Page 70
Accessories Liquiline Compact CM82
Wireless HART adapter SWA70
• Wireless device connection
• Easily integrated, offers data protection and transmission safety, can be operated in
parallel with other wireless networks, minimum cabling complexity
Technical Information TI00061S

14.3.4 System components

RIA15
• Process display unit, Digital display unit for integration into 4-20 mA circuits
• Panel mounting
• With optional HART communication
Technical Information TI01043K
70 Endress+Hauser
Page 71
Liquiline Compact CM82 Technical data
V
12
13
14
15 16
17
18
19
20
0
100
300
500
800
700
600
12,6
200
400
21
22
23
24 25
26
27
28
29 30

15 Technical data

15.1 Input

Measured variables • pH

• ORP
• pH/ORP
• Oxygen
• Conductivity

Measuring ranges → Documentation of the connected sensor

Types of input Digital sensor inputs for Memosens-sensors

Cable specification Cable length:

• Max. 3 m (10 ft)
• Max. 7 m (23 ft)
• Max. 15 m (49 ft)

15.2 Output

Output signal 4 ... 20 mA/HART, galvanically isolated from the sensor circuits

Linearization/transmission
Linear
behavior

15.3 Power supply

Supply voltage 12.6 to 30 VDC (with setting error current > 20 mA)

14 to 30 VDC (with setting error current < 4 mA)
A0036752
 15 Supply voltage and load
Endress+Hauser 71
Page 72
Technical data Liquiline Compact CM82
The lower voltage value in each case applies only to a load resistance of 0 Ohm.
NOTICE
The device does not have a power switch
At the supply point, the power supply must be isolated from dangerous live cables by
double or reinforced insulation in the case of devices with a 24 V power supply.

Overvoltage protection IEC 61 000-4-4 and IEC 61 000-4-5 with +/- 1 kV

Sensor connection Sensors with Memosens protocol

Sensor types Sensors
Digital Sensors with inductive memosens plug head • pH sensors
• ORP sensors
• pH/ORP combination sensors
• Oxygen sensors
• Conductivity sensors

15.4 Performance characteristics

Response time of current
t90 = max. 500 ms for an increase from 0 to 20 mA
output

Tolerance of current output Typical measuring tolerances:

< ±20 µA (if current value = 4 mA) < ±50 µA (for current values 4 to 20 mA) at 25 ˚C (77˚ F) each
additional tolerance depending on the temperature:
< 1.5 µA/K
Resolution of current
< 5 µA
output

Repeatability → Documentation of the connected sensor

15.5 Environment

Ambient temperature -20 to +85 °C (0 to 185 °F)

The maximum ambient temperature depends on the process temperature and the installation situation of the transmitter.
Make sure that the ambient temperature at the transmitter does not exceed +85 °C (185 °F).
Example of environmental conditions in Endress+Hauser fittings:
• with open installation (without protective cover, i.e. free convection at the transmitter),
e.g. CPA442, CPA842
• for enclosed installation (with protective cover), e.g. CPA871, CPA875, CPA842
T
T
= max. 60 °C (140 °F)
ambient
= max. 100 °C (212 °F), in continuous operation
prozess
72 Endress+Hauser
Page 73
Liquiline Compact CM82 Technical data
Ambient temperature
Process temperature
T
= max. 140 °C (284 °F), < 2h (for sterilization)
prozess
A0036933-EN
 16 Mounting situation of the transmitter with or without protective cover

Storage temperature -40 to +85 °C (-40 to 185 °F)

Humidity 5 to 95 %

Degree of protection IP 67

IP 68
NEMA Type 6

Electromagnetic compatibility

• EN 61326-1
• EN 61326-2-3
• EN 301489-1
• EN 301489-17
• NAMUR NE 21

Electrical safety EN 61010-1

Max. altitude above MSL < 2000 m (< 6562 ft) above MSL

Degree of contamination

Complete device: Pollution level 4
Internal: Pollution level 2

Radio standards • EN 300 328 (Europa)

• 47 CFR 15.247 (USA)
• RSS-247 Issue 1 (Kanada)
• RSS-GEN Issue 4 (Kanada)

15.6 Mechanical construction

Materials

Endress+Hauser 73
Components Material
Housing, cover Peek 151
Strain relief EPDM (peroxide crosslinked)
Page 74
Technical data Liquiline Compact CM82
Components Material
Axial ring Peek 450 G
Optical waveguide PC transparent
Impact loads
The product is designed for mechanical impact loads of 1 J (IK06) as per the requirements of EN61010-1.

Weight

without cable Approx. 42 g (1.5 oz)
3 m (9 f) cable Approx. 190 g (7 oz)
7 m (23 f) cable Approx. 380 g (13 oz)
15 m (49 f) cable Approx. 760 g (27 oz)
For every 1 m (3 f) of cable Approx. 48 g (2 oz)
74 Endress+Hauser
Page 75
Liquiline Compact CM82 Index

Index

A
Accessories ................................ 66
Communication-specific .................... 69
Activation codes ............................ 66
Ambient temperature ........................ 72
Approvals ................................. 10
C
Cable specification ...........................71
Calibration ................................ 64
Calibration settings .......................... 30
Calibrations ................................47
Certificates ................................ 10
Check
Installation and function ....................19
Cleaning .................................. 64
Commissioning ............................. 19
Configuration ........................ 27, 28, 34
Connection ................................ 20
Electrical ............................... 13
Sensors ................................ 72
Supply voltage ........................... 71
connection with RIA15 ....................... 13
Current output ............................. 32
D
Date ..................................... 27
Degree of contamination ...................... 73
Degree of protection ......................... 73
Delta slope ................................ 40
Delta zero point .............................42
Designated use .............................. 6
Device information .......................... 63
Diagnostic list .............................. 48
Diagnostic messages ......................... 48
Diagnostics ................................ 38
Dimensions ................................12
Disposal .................................. 65
Documentation .............................. 5
E
Electrical connection ......................... 13
Electrical safety .............................73
Electromagnetic compatibility .................. 73
Event logbook .............................. 62
F
Firmware ................................. 63
Function check ............................. 19
H
HART ................................. 18, 32
Hold ..................................... 33
Humidity ..................................73
I
Impedance ................................ 39
Incoming acceptance ......................... 9
Input
Measured variables ....................... 71
Installation ................................ 12
L
LED display ................................ 20
Limit values ............................... 43
Linearization ...............................71
M
Maintenance ...............................64
Maintenance tasks .......................... 64
Materials ................................. 73
Measured variables .......................... 71
Measuring parameters ........................ 8
Measuring ranges ........................... 71
Medium compensation ....................... 36
N
Nameplate ................................. 9
O
Occupational safety ........................... 6
Operating hours ............................ 43
Operating language ..........................27
Operating menu ............................ 15
Operation .............................. 15, 35
Operation Point .............................41
Operational safety ............................6
Output signal .............................. 71
Overvoltage protection ....................... 72
P
Password ................................. 34
Pharmaceutical water ........................ 46
Post-connection check ........................14
Post-installation check ....................... 19
Power supply ...............................71
Overvoltage protection ..................... 72
Sensor connection ........................ 72
Supply voltage ........................... 71
Process monitoring .......................... 45
Product description ........................... 8
Product design .............................. 8
Product identification ......................... 9
R
Radio standards ............................ 73
Repair ....................................65
Requirements for personnel .....................6
Reset .................................... 62
Return ................................... 65
RIA15 ....................................22
RIA15 operating matrix .......................22
Endress+Hauser 75
Page 76
Index Liquiline Compact CM82
S
Safety
Occupational safety ........................ 6
Operation ................................6
Safety instructions ............................6
SCC ......................................45
Scope of delivery ............................ 10
Sensor
Connection ..............................72
Sensor condition check ....................... 45
Sensor Condition Check ....................... 45
Sensors ...................................66
Simulation ................................ 47
Slope .................................... 40
Software .................................. 66
Squawk ...................................63
Sterilizations ...............................44
Storage temperature ......................... 73
Supply voltage ..............................71
Symbols ................................... 4
System integration .......................... 18
System settings ............................. 20
T
Technical data ..............................71
Environment ............................ 72
Input .................................. 71
Mechanical construction ....................73
Output ................................. 71
Technical personnel ...........................6
Time .....................................27
Transmission behavior ....................... 71
Troubleshooting ............................ 38
Types of input .............................. 71
U
Use
Designated ...............................6
W
Warnings .................................. 4
Z
Zero point ................................. 41
76 Endress+Hauser
Page 77
Page 78
Page 79
Page 80
www.addresses.endress.com
Loading...