Endress+Hauser CM82 Specifications

Page 1
TI01397C/07/EN/01.18 71406282 2018-01-31
Products
Solutions Services
Technical Information
Liquiline Compact CM82
Configurable multiparameter transmitter for Memosens sensors
Space-saving transmitter for monitoring and controlling processes in industry and the environmental sector
Application
The CM82 can be used in all sectors and by plant manufacturers in these sectors and supports all sensors with the blue Memosens plug-in head:
• pH, ORP and pH-ORP combined electrodes
• Conductive conductivity
• 4 to 20 mA
• HART
• Bluetooth® LE interface for commissioning and maintenance
Your benefits
• Space-saving installation and storage: – The two-wire device fits into an assembly and does not require a separate power
supply.
– Minimum inventory
• Maximum safety: Tried and trusted Memosens technology
• Easy operation – Use your existing tablet and smartphone for operation and commissioning. – Standardized operating concept across all devices from the Liquiline platform
• Fast and reliable A reliable Bluetooth® LE connection allows you to check measuring points that are dangerous or difficult to access from a safe distance.
• Suitable for all locations Regardless of whether your measuring point is exposed to dust, steam, rain, snow, heat or cold, the CM82 is always exactly the transmitter you need!
Page 2

Function and system design

1
2
3
4
5
Liquiline Compact CM82

Measuring system

The overview shows examples of measuring systems. Other sensors and assemblies can be ordered for conditions specific to your application (www.endress.com/products).
A complete measuring system comprises:
• Liquiline compact transmitter
• Sensors with Memosens technology
• Assemblies to suit the sensors used
A0036772
 1 Example of a measuring system
1 Measuring point with assembly and Memosens sensor 2 Liquiline Compact CM82 3 Junction box, optionally available 4 PLC (programmable logic control) 5 Bluetooth LE optional for mobile terminals, e.g. tablets
2 Endress+Hauser
Page 3
Liquiline Compact CM82
MEMO SENS

Application example

A0035115
 2 Application example in a fermenter, using a tablet for operation

Reliability

Communication and data processing

Types of communication:
• 4 to 20 mA
• HART fieldbus
• Bluetooth® LE wireless technology (optional) The device drivers available make it possible to perform a basic setup and display measured
values and diagnostics information via the fieldbus. Full device configuration is possible via the fieldbus and Bluetooth.

Dependability

Memosens
Memosens makes your measuring point safer and more reliable:
• Non-contact, digital signal transmission enables optimum galvanic isolation
• No contact corrosion
• Completely watertight
• Sensor can be calibrated in a lab, thus increasing the availability of the measuring point in the
process
• Predictive maintenance thanks to recording of sensor data, e.g.:
– Total hours of operation – Hours of operation with very high or very low measured values – Hours of operation at high temperatures – Number of steam sterilizations – Sensor condition
Endress+Hauser 3
Page 4
Liquiline Compact CM82
 3 Plug & Play with Memosens technology
The status of the transmitter and the connected sensor is indicated by a red/green LED.
A0035116
 4 LED display
USP and EP
• "Water for Injection" (WFI) as per USP <645> and EP
• "Highly Purified Water" (HPW) as per EP
• "Purified Water" (PW) as per EP The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared against the tables defined in the standards. An alarm is triggered if the limit value is exceeded. Furthermore, it is also possible to configure an early warning alarm that signals undesired operating states before they occur.

Safety Safe signal transmission via Bluetooth® LE

Signal transmission via Bluetooth® wireless technology uses a cryptographic technique tested by the Fraunhofer Institute.
Security levels for Endress and Hauser Bluetooth infrastructure:
• Protocol: High
• Algorithms: High Measured against:
• the security objectives, e.g. confidentiality, integrity, availability, etc.
• the risk analysis, e.g. key distribution, authentication, password recovery, etc.
• the attack model, e.g. motivation for attack, time required, expertise in electronics, etc.
• the weak-point analysis For comparison: The general Bluetooth standard is classified as "Low".
Protection against unauthorized access:
• Password-protected
• Without the SmartBlue app, the device is not visible via Bluetooth® wireless technology.
• Only one point-to-point connection is established between a sensor and a smartphone or tablet.
A0036843
1)
:
1) Multi-level scale for security assessments in accordance with Fraunhofer AISEC cryptographic technique: "Very low", "Low", "High", "Very high"
4 Endress+Hauser
Page 5
Liquiline Compact CM82
• The Bluetooth® wireless technology interface can be disabled via SmartBlue.
• Bluetooth® is optional. It can be ordered with this functionality enabled.
If ordered with Bluetooth® disabled, Bluetooth® can be enabled at a later stage by means of an activation code (accessory kit) linked to the serial number.
• A disabled Bluetooth® interface can be re-enabled only via HART.
Measured value compensation
pH:
Temperature
Oxygen:
• Temperature
• Air pressure
Conductivity:
Temperature Various methods are available to compensate for the temperature dependency: – Linear – NaCl (IEC 746-3) – Water ISO7888 (20°C) – Water ISO7888 (25°C)
IT security
Our warranty is valid only if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer, must be implemented by the operators themselves in line with their security standards.

Measured values

Measuring ranges

Types of input

Cable specification

Output signal

Linearization/transmission behavior

Input

The transmitter is designed for digital Memosens sensors with inductive plug-in head:
• pH, ORP. pH/ORP combined sensors
• Conductive Conductivity
• Dissolved oxygen
→ Documentation of the connected sensor
Digital sensor inputs for Memosens-sensors
Cable length:
• Max. 3 m (10 ft)
• Max. 7 m (23 ft)
• Max. 15 m (49 ft)

Output

4 ... 20 mA/HART, galvanically isolated from the sensor circuits
Linear
Endress+Hauser 5
Page 6

Power supply

V
12
13
14
15 16
17
18
19
20
0
100
300
500
800
700
600
12,6
200
400
21
22
23
24 25
26
27
28
29 30
Liquiline Compact CM82

Supply voltage

12.6 to 30 VDC (with setting error current > 20 mA)
14 to 30 VDC (with setting error current < 4 mA)
A0036752
 5 Supply voltage and load
The lower voltage value in each case applies only to a load resistance of 0 Ohm.

Overvoltage protection

Sensor connection

Response time of current output

Tolerance of current output

Resolution of current output

NOTICE
The device does not have a power switch
At the supply point, the power supply must be isolated from dangerous live cables by double or
reinforced insulation in the case of devices with a 24 V power supply.
IEC 61 000-4-4 and IEC 61 000-4-5 with +/- 1 kV
Sensors with Memosens protocol
Sensor types Sensors
Digital Sensors with inductive memosens plug head • pH sensors
• ORP sensors
• pH/ORP combination sensors
• Oxygen sensors
• Conductivity sensors

Performance characteristics

t90 = max. 500 ms for an increase from 0 to 20 mA
Typical measuring tolerances:
< ±20 µA (if current value = 4 mA) < ±50 µA (for current values 4 to 20 mA) at 25 ˚C (77˚ F) each
additional tolerance depending on the temperature:
< 1.5 µA/K
< 5 µA
6 Endress+Hauser
Page 7
Liquiline Compact CM82
Ambient temperature
Process temperature

Repeatability

Ambient temperature range

→ Documentation of the connected sensor

Environment

-20 to +85 °C (0 to 185 °F)
The maximum ambient temperature depends on the process temperature and the installation situation of the transmitter.
Make sure that the ambient temperature at the transmitter does not exceed +85 °C (185 °F).
Example of environmental conditions in Endress+Hauser fittings:
• with open installation (without protective cover, i.e. free convection at the transmitter), e.g.
CPA442, CPA842
• for enclosed installation (with protective cover), e.g. CPA871, CPA875, CPA842
T
= max. 60 °C (140 °F)
ambient
T
= max. 100 °C (212 °F), in continuous operation
prozess
T
= max. 140 °C (284 °F), < 2h (for sterilization)
prozess

Storage temperature

Humidity

Degree of protection

Electromagnetic compatibility

Electrical safety

Max. altitude above MSL

Degree of contamination

A0036933-EN
 6 Mounting situation of the transmitter with or without protective cover
-40 to +85 °C (-40 to 185 °F)
5 to 95 %
IP 67
IP 68
NEMA Type 6
• EN 61326-1
• EN 61326-2-3
• EN 301489-1
• EN 301489-17
• NAMUR NE 21
EN 61010-1
< 2000 m (< 6562 ft) above MSL
Complete device: Pollution level 4
Internal: Pollution level 2
Endress+Hauser 7
Page 8
Liquiline Compact CM82
110.8 (4.33)
99.8 (3.92)
32.75 (1.29)
114.7 (4.52)
124.6 (4.90)
Ø
18.95 (0.75)
Ø
5.5 (0.22)

Radio standards

Dimensions

• EN 300 328 (Europa)
• 47 CFR 15.247 (USA)
• RSS-247 Issue 1 (Kanada)
• RSS-GEN Issue 4 (Kanada)

Mechanical construction

Materials

Weight

A0033272
 7 Dimensions in mm (inch)
Components Material
Housing, cover Peek 151
Strain relief EPDM (peroxide crosslinked)
Axial ring Peek 450 G
Optical waveguide PC transparent
Impact loads
The product is designed for mechanical impact loads of 1 J (IK06) as per the requirements of EN61010-1.
without cable Approx. 42 g (1.5 oz)
3 m (9 f) cable Approx. 190 g (7 oz)
7 m (23 f) cable Approx. 380 g (13 oz)
15 m (49 f) cable Approx. 760 g (27 oz)
For every 1 m (3 f) of cable Approx. 48 g (2 oz)

Operating concept

Operability

• Bluetooth® LE wireless technology
• HART
8 Endress+Hauser
Page 9
Liquiline Compact CM82
EH_FMRx0_0123456
CM82_M50053105G11
Operation via SmartBlue (app)
SmartBlue is available as a download for Android terminals from the Google Playstore and for iOS devices from the Apple iTunes Store.
If you scan the QR code, you will be brought directly to the app:
A0031189-EN
 8 Download links
 9 SmartBlue App
 10 Livelist
The Livelist displays all of the devices that are within range.
System requirements
• iOS devices: iPhone 4S or higher from iOS9.0; iPad2 or higher from iOS9.0; iPod Touch 5.
Generation or higher from iOS9.0
• Devices with Android: from Android 4.4 KitKat and Bluetooth® 4.0
A0029747
A0035117
Endress+Hauser 9
Page 10

Remote operation HART

ESC
E
+
1
3
4
2
5
6
 11 Wiring options for remote operation via HART protocol
1 PLC (programmable logic control) 2 RIA15 loop-powered process indicator, optional 3 Junction box 4 HART operating device (e.g. Fieldcare), optional 5 Transmitter with optional Bluetooth® LE wireless technology 6 Optional: Smartphone / tablet with SmartBlue (app)
Liquiline Compact CM82
A0036740

Certificates and approvals

mark

FCC/IC Radio approval for USA/Canada

Endress+Hauser Conducta GmbH+CO. KG hereby declares that the radio system type CM82 complies with directives 2014/53/EU and 2011/65/EU.
The complete text of the EU declaration of conformity can be found at the following web address: "http://www.endress.com/CM82"
Frequency band: 2400-2483.5 MHz, power output: < 10dBm EIRP
This device complies with Part 15 of the FCC Rules [and with Industry Canada license- exempt RSS standard(s)]. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Any changes or modifications made to this equipment not expressly approved by Endress+Hauser may void the FCC authorization to operate this equipment. This device complies with Part 15 of the FCC Rules and with Industry Canada license- exempt RSS standard(s).
Operation is subject to the following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may cause undesired operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes:
• l'appareil ne doit pas produire de brouillage, et
• l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses
10 Endress+Hauser
Page 11
Liquiline Compact CM82
202-LSF040
R
R-CRM-E1H-CM82A
and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Increase the separation between the equipment and receiver.
• Consult the dealer or an experienced radio technician for help.
Radio approval for Japan
A0036603
Japanese Radio Law and Japanese Telecommunications Business Law Compliance. This device is granted pursuant to the Japanese Radio Law (電波法). This device should not be modified (otherwise the granted designation number will become invalid).
Radio approval for China
Certification number: CMIIT ID: 2017DJ6495
Radio approval for South Korea
Brazil radio approval
Mexico radio approval
In development
A0036602
A0036857
Endress+Hauser 11
Page 12

Ordering information

Liquiline Compact CM82

Product page

Product Configurator

Scope of delivery

www.endress.com/CM82
On the product page there is a Configure button to the right of the product image.
1. Click this button.  The Configurator opens in a separate window.
2. Select all the options to configure the device in line with your requirements.  In this way, you receive a valid and complete order code for the device.
3. Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right above the selection window.
For many products you also have the option of downloading CAD or 2D drawings of the selected product version. Click the CAD tab for this and select the desired file type using picklists.
The scope of delivery includes:
• CM82
• Brief Operating Instructions

Accessories

Sensors Glass electrodes

Orbisint CPS11D
• pH electrode for process technology
• Optional SIL version for connecting to SIL transmitter
• With dirt-repellent PTFE diaphragm
• Product Configurator on the product page: www.endress.com/cps11d Technical Information TI00028C
Memosens CPS31D
• pH electrode with gel-filled reference system with ceramic diaphragm
• Product Configurator on the product page: www.endress.com/cps31d Technical Information TI00030C
Ceraliquid CPS41D
• pH electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps41d Technical Information TI00079C
Ceragel CPS71D
• pH electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps71d Technical Information TI00245C
Memosens CPS171D
• pH electrode for bio-fermenters with digital Memosens technology
• Product Configurator on the product page: www.endress.com/cps171d Technical Information TI01254C
Orbipore CPS91D
• pH electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps91d Technical Information TI00375C
12 Endress+Hauser
Page 13
Liquiline Compact CM82
Orbipac CPF81D
• Compact pH sensor for installation or immersion operation
• In industrial water and wastewater
• Product Configurator on the product page: www.endress.com/cpf81d Technical Information TI00191C
Enamel pH electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Product Configurator on the product page: www.endress.com/cps341d Technical Information TI00468C
ORP sensors
Orbisint CPS12D
• ORP sensor for process technology
• Product Configurator on the product page: www.endress.com/cps12d Technical Information TI00367C
Ceraliquid CPS42D
• ORP electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps42d Technical Information TI00373C
Ceragel CPS72D
• ORP electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps72d Technical Information TI00374C
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Product Configurator on the product page: www.endress.com/cpf82d Technical Information TI00191C
Orbipore CPS92D
• ORP electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps92d Technical Information TI00435C
pH ISFET sensors
Tophit CPS441D
• Sterilizable ISFET sensor for low-conductivity media
• Liquid KCl electrolyte
• Product Configurator on the product page: www.endress.com/cps441d Technical Information TI00352C
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
• Water treatment and biotechnology
• Product Configurator on the product page: www.endress.com/cps471d Technical Information TI00283C
Tophit CPS491D
• ISFET sensor with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps491d Technical Information TI00377C
Endress+Hauser 13
Page 14
pH and ORP combined sensors
Memosens CPS16D
• Combined pH/ORP sensor for process technology
• With dirt-repellent PTFE diaphragm
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps16D Technical Information TI00503C
Memosens CPS76D
• Combined pH/ORP sensor for process technology
• Hygienic and sterile applications
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps76d Technical Information TI00506C
Memosens CPS96D
• Combined pH/ORP sensor for chemical processes
• With poison-resistant reference with ion trap
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps96d Technical Information TI00507C
Liquiline Compact CM82
Conductivity sensors with conductive measurement of conductivity
Condumax CLS15D
• Conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• Product Configurator on the product page: www.endress.com/CLS15d Technical Information TI00109C
Condumax CLS16D
• Hygienic, conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• With EHEDG and 3A approval
• Product Configurator on the product page: www.endress.com/CLS16d Technical Information TI00227C
Condumax CLS21D
• Two-electrode sensor in plug-in head version version
• Product Configurator on the product page: www.endress.com/CLS21d Technical Information TI00085C
Memosens CLS82D
• Four-electrode sensor
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cls82d Technical Information TI01188C
Oxygen sensors
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• With Memosens technology or as an analog sensor
• Product Configurator on the product page: www.endress.com/cos22d Technical Information TI00446C
14 Endress+Hauser
Page 15
Liquiline Compact CM82
Oxymax COS51D
• Amperometric sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos51d Technical Information TI00413C
Memosens COS81D
• Sterilizable, optical sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos81d Technical Information TI01201C

Software

Other accessories Activation codes

Memobase Plus CYZ71D
• PC software to support laboratory calibration
• Visualization and documentation of sensor management
• Sensor calibrations stored in database
• Product Configurator on the product page: www.endress.com/cyz71d Technical Information TI00502C
You must quote the serial number of the device when ordering the activation code.
Activation code: Bluetooth
Order No. 71401176
Cable junction with Velcro strip
Cable junction with Velcro strip
• 4 pieces, for sensor cable
• Order No. 71092051
Communication-specific accessories
Commubox FXA195
Intrinsically safe HART communication with FieldCare via the USB port
Technical Information TI00404F
Wireless HART adapter SWA70
• Wireless device connection
• Easily integrated, offers data protection and transmission safety, can be operated in parallel with
other wireless networks, minimum cabling complexity
Technical Information TI00061S
System components
RIA15
• Process display unit, Digital display unit for integration into 4-20 mA circuits
• Panel mounting
• With optional HART communication Technical Information TI01043K
Endress+Hauser 15
Page 16
www.addresses.endress.com
Loading...