Configurable multiparameter transmitter for
Memosens sensors
Space-saving transmitter for monitoring and
controlling processes in industry and the
environmental sector
Application
The CM82 can be used in all sectors and by plant manufacturers in these sectors and
supports all sensors with the blue Memosens plug-in head:
• pH, ORP and pH-ORP combined electrodes
• Conductive conductivity
• Oxygen
Direct connection to PLC via:
• 4 to 20 mA
• HART
• Bluetooth® LE interface for commissioning and maintenance
Your benefits
• Space-saving installation and storage:
– The two-wire device fits into an assembly and does not require a separate power
supply.
– Minimum inventory
• Maximum safety:
Tried and trusted Memosens technology
• Easy operation
– Use your existing tablet and smartphone for operation and commissioning.
– Standardized operating concept across all devices from the Liquiline platform
• Fast and reliable
A reliable Bluetooth® LE connection allows you to check measuring points that are
dangerous or difficult to access from a safe distance.
• Suitable for all locations
Regardless of whether your measuring point is exposed to dust, steam, rain, snow,
heat or cold, the CM82 is always exactly the transmitter you need!
Page 2
Function and system design
1
2
3
4
5
Liquiline Compact CM82
Measuring system
The overview shows examples of measuring systems. Other sensors and assemblies can be ordered
for conditions specific to your application (www.endress.com/products).
A complete measuring system comprises:
• Liquiline compact transmitter
• Sensors with Memosens technology
• Assemblies to suit the sensors used
A0036772
1Example of a measuring system
1Measuring point with assembly and Memosens sensor
2Liquiline Compact CM82
3Junction box, optionally available
4PLC (programmable logic control)
5Bluetooth LE optional for mobile terminals, e.g. tablets
2Endress+Hauser
Page 3
Liquiline Compact CM82
MEMO SENS
Application example
A0035115
2Application example in a fermenter, using a tablet for operation
Reliability
Communication and data processing
Types of communication:
• 4 to 20 mA
• HART fieldbus
• Bluetooth® LE wireless technology (optional)
The device drivers available make it possible to perform a basic setup and display measured
values and diagnostics information via the fieldbus. Full device configuration is possible via the
fieldbus and Bluetooth.
Dependability
Memosens
Memosens makes your measuring point safer and more reliable:
• Non-contact, digital signal transmission enables optimum galvanic isolation
• No contact corrosion
• Completely watertight
• Sensor can be calibrated in a lab, thus increasing the availability of the measuring point in the
process
• Predictive maintenance thanks to recording of sensor data, e.g.:
– Total hours of operation
– Hours of operation with very high or very low measured values
– Hours of operation at high temperatures
– Number of steam sterilizations
– Sensor condition
Endress+Hauser3
Page 4
Liquiline Compact CM82
3Plug & Play with Memosens technology
The status of the transmitter and the connected sensor is indicated by a red/green LED.
A0035116
4LED display
USP and EP
• "Water for Injection" (WFI) as per USP <645> and EP
• "Highly Purified Water" (HPW) as per EP
• "Purified Water" (PW) as per EP
The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared against the tables defined in the standards. An alarm
is triggered if the limit value is exceeded. Furthermore, it is also possible to configure an early
warning alarm that signals undesired operating states before they occur.
SafetySafe signal transmission via Bluetooth® LE
Signal transmission via Bluetooth® wireless technology uses a cryptographic technique tested
by the Fraunhofer Institute.
Security levels for Endress and Hauser Bluetooth infrastructure:
• Protocol: High
• Algorithms: High
Measured against:
• the security objectives, e.g. confidentiality, integrity, availability, etc.
• the risk analysis, e.g. key distribution, authentication, password recovery, etc.
• the attack model, e.g. motivation for attack, time required, expertise in electronics, etc.
• the weak-point analysis
For comparison: The general Bluetooth standard is classified as "Low".
Protection against unauthorized access:
• Password-protected
• Without the SmartBlue app, the device is not visible via Bluetooth® wireless technology.
• Only one point-to-point connection is established between a sensor and a smartphone or tablet.
A0036843
1)
:
1)Multi-level scale for security assessments in accordance with Fraunhofer AISEC cryptographic technique: "Very low", "Low", "High", "Very high"
4Endress+Hauser
Page 5
Liquiline Compact CM82
• The Bluetooth® wireless technology interface can be disabled via SmartBlue.
• Bluetooth® is optional. It can be ordered with this functionality enabled.
If ordered with Bluetooth® disabled, Bluetooth® can be enabled at a later stage by means of an
activation code (accessory kit) linked to the serial number.
• A disabled Bluetooth® interface can be re-enabled only via HART.
Measured value compensation
pH:
Temperature
Oxygen:
• Temperature
• Air pressure
Conductivity:
Temperature
Various methods are available to compensate for the temperature dependency:
– Linear
– NaCl (IEC 746-3)
– Water ISO7888 (20°C)
– Water ISO7888 (25°C)
IT security
Our warranty is valid only if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer,
must be implemented by the operators themselves in line with their security standards.
Measured values
Measuring ranges
Types of input
Cable specification
Output signal
Linearization/transmission
behavior
Input
The transmitter is designed for digital Memosens sensors with inductive plug-in head:
• pH, ORP. pH/ORP combined sensors
• Conductive Conductivity
• Dissolved oxygen
→ Documentation of the connected sensor
Digital sensor inputs for Memosens-sensors
Cable length:
• Max. 3 m (10 ft)
• Max. 7 m (23 ft)
• Max. 15 m (49 ft)
Output
4 ... 20 mA/HART, galvanically isolated from the sensor circuits
Linear
Endress+Hauser5
Page 6
Power supply
V
12
13
14
15 16
17
18
19
20
0
100
300
500
800
700
600
12,6
Ω
200
400
21
22
23
24 25
26
27
28
29 30
Liquiline Compact CM82
Supply voltage
12.6 to 30 VDC (with setting error current > 20 mA)
14 to 30 VDC (with setting error current < 4 mA)
A0036752
5Supply voltage and load
The lower voltage value in each case applies only to a load resistance of 0 Ohm.
Overvoltage protection
Sensor connection
Response time of current
output
Tolerance of current output
Resolution of current output
NOTICE
The device does not have a power switch
At the supply point, the power supply must be isolated from dangerous live cables by double or
‣
reinforced insulation in the case of devices with a 24 V power supply.
IEC 61 000-4-4 and IEC 61 000-4-5 with +/- 1 kV
Sensors with Memosens protocol
Sensor typesSensors
Digital Sensors with inductive memosens plug head• pH sensors
• ORP sensors
• pH/ORP combination sensors
• Oxygen sensors
• Conductivity sensors
Performance characteristics
t90 = max. 500 ms for an increase from 0 to 20 mA
Typical measuring tolerances:
< ±20 µA (if current value = 4 mA)
< ±50 µA (for current values 4 to 20 mA)
at 25 ˚C (77˚ F) each
additional tolerance depending on the temperature:
< 1.5 µA/K
< 5 µA
6Endress+Hauser
Page 7
Liquiline Compact CM82
Ambient temperature
Process temperature
Repeatability
Ambient temperature range
→ Documentation of the connected sensor
Environment
-20 to +85 °C (0 to 185 °F)
The maximum ambient temperature depends on the process temperature and the installation
situation of the transmitter.
Make sure that the ambient temperature at the transmitter does not exceed +85 °C (185 °F).
Example of environmental conditions in Endress+Hauser fittings:
• with open installation (without protective cover, i.e. free convection at the transmitter), e.g.
CPA442, CPA842
• for enclosed installation (with protective cover), e.g. CPA871, CPA875, CPA842
T
= max. 60 °C (140 °F)
ambient
T
= max. 100 °C (212 °F), in continuous operation
prozess
T
= max. 140 °C (284 °F), < 2h (for sterilization)
prozess
Storage temperature
Humidity
Degree of protection
Electromagnetic
compatibility
Electrical safety
Max. altitude above MSL
Degree of contamination
A0036933-EN
6Mounting situation of the transmitter with or without protective cover
-40 to +85 °C (-40 to 185 °F)
5 to 95 %
IP 67
IP 68
NEMA Type 6
• EN 61326-1
• EN 61326-2-3
• EN 301489-1
• EN 301489-17
• NAMUR NE 21
EN 61010-1
< 2000 m (< 6562 ft) above MSL
Complete device:Pollution level 4
Internal:Pollution level 2
Endress+Hauser7
Page 8
Liquiline Compact CM82
110.8 (4.33)
99.8 (3.92)
32.75 (1.29)
114.7 (4.52)
124.6 (4.90)
Ø
18.95 (0.75)
Ø
5.5 (0.22)
Radio standards
Dimensions
• EN 300 328 (Europa)
• 47 CFR 15.247 (USA)
• RSS-247 Issue 1 (Kanada)
• RSS-GEN Issue 4 (Kanada)
Mechanical construction
Materials
Weight
A0033272
7Dimensions in mm (inch)
ComponentsMaterial
Housing, coverPeek 151
Strain reliefEPDM (peroxide crosslinked)
Axial ringPeek 450 G
Optical waveguidePC transparent
Impact loads
The product is designed for mechanical impact loads of 1 J (IK06) as per the requirements of
EN61010-1.
without cableApprox. 42 g (1.5 oz)
3 m (9 f) cableApprox. 190 g (7 oz)
7 m (23 f) cableApprox. 380 g (13 oz)
15 m (49 f) cableApprox. 760 g (27 oz)
For every 1 m (3 f) of cableApprox. 48 g (2 oz)
Operating concept
Operability
• Bluetooth® LE wireless technology
• HART
8Endress+Hauser
Page 9
Liquiline Compact CM82
EH_FMRx0_0123456
CM82_M50053105G11
Operation via SmartBlue (app)
SmartBlue is available as a download for Android terminals from the Google Playstore and for iOS
devices from the Apple iTunes Store.
If you scan the QR code, you will be brought directly to the app:
A0031189-EN
8Download links
9SmartBlue App
10Livelist
The Livelist displays all of the devices that are within range.
System requirements
• iOS devices: iPhone 4S or higher from iOS9.0; iPad2 or higher from iOS9.0; iPod Touch 5.
Generation or higher from iOS9.0
• Devices with Android: from Android 4.4 KitKat and Bluetooth® 4.0
A0029747
A0035117
Endress+Hauser9
Page 10
Remote operationHART
ESC
E
+
1
3
4
2
5
6
11Wiring options for remote operation via HART protocol
1PLC (programmable logic control)
2RIA15 loop-powered process indicator, optional
3Junction box
4HART operating device (e.g. Fieldcare), optional
5Transmitter with optional Bluetooth® LE wireless technology
6Optional: Smartphone / tablet with SmartBlue (app)
Liquiline Compact CM82
A0036740
Certificates and approvals
mark
FCC/ICRadio approval for USA/Canada
Endress+Hauser Conducta GmbH+CO. KG hereby declares that the radio system type CM82 complies
with directives 2014/53/EU and 2011/65/EU.
The complete text of the EU declaration of conformity can be found at the following web address:
"http://www.endress.com/CM82"
Frequency band: 2400-2483.5 MHz, power output: < 10dBm EIRP
This device complies with Part 15 of the FCC Rules [and with Industry Canada license- exempt RSS
standard(s)]. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Any changes or modifications made to this equipment not expressly approved by Endress+Hauser
may void the FCC authorization to operate this equipment. This device complies with Part 15 of the
FCC Rules and with Industry Canada license- exempt RSS standard(s).
Operation is subject to the following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may cause undesired
operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts
de licence. L'exploitation est autorisée aux deux conditions suivantes:
• l'appareil ne doit pas produire de brouillage, et
• l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d'en compromettre le fonctionnement.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates, uses
10Endress+Hauser
Page 11
Liquiline Compact CM82
202-LSF040
R
R-CRM-E1H-CM82A
and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
• Increase the separation between the equipment and receiver.
• Consult the dealer or an experienced radio technician for help.
Radio approval for Japan
A0036603
Japanese Radio Law and Japanese Telecommunications Business Law Compliance. This device is
granted pursuant to the Japanese Radio Law (電波法). This device should not be modified (otherwise
the granted designation number will become invalid).
Radio approval for China
Certification number: CMIIT ID: 2017DJ6495
Radio approval for South Korea
Brazil radio approval
Mexico radio approval
In development
A0036602
A0036857
Endress+Hauser11
Page 12
Ordering information
Liquiline Compact CM82
Product page
Product Configurator
Scope of delivery
www.endress.com/CM82
On the product page there is a Configure button to the right of the product image.
1.Click this button.
The Configurator opens in a separate window.
2.Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3.Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right
above the selection window.
For many products you also have the option of downloading CAD or 2D drawings of the
selected product version. Click the CAD tab for this and select the desired file type using
picklists.
The scope of delivery includes:
• CM82
• Brief Operating Instructions
Accessories
SensorsGlass electrodes
Orbisint CPS11D
• pH electrode for process technology
• Optional SIL version for connecting to SIL transmitter
• With dirt-repellent PTFE diaphragm
• Product Configurator on the product page: www.endress.com/cps11d
Technical Information TI00028C
Memosens CPS31D
• pH electrode with gel-filled reference system with ceramic diaphragm
• Product Configurator on the product page: www.endress.com/cps31d
Technical Information TI00030C
Ceraliquid CPS41D
• pH electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps41d
Technical Information TI00079C
Ceragel CPS71D
• pH electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps71d
Technical Information TI00245C
Memosens CPS171D
• pH electrode for bio-fermenters with digital Memosens technology
• Product Configurator on the product page: www.endress.com/cps171d
Technical Information TI01254C
Orbipore CPS91D
• pH electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps91d
Technical Information TI00375C
12Endress+Hauser
Page 13
Liquiline Compact CM82
Orbipac CPF81D
• Compact pH sensor for installation or immersion operation
• In industrial water and wastewater
• Product Configurator on the product page: www.endress.com/cpf81d
Technical Information TI00191C
Enamel pH electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Product Configurator on the product page: www.endress.com/cps341d
Technical Information TI00468C
ORP sensors
Orbisint CPS12D
• ORP sensor for process technology
• Product Configurator on the product page: www.endress.com/cps12d
Technical Information TI00367C
Ceraliquid CPS42D
• ORP electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps42d
Technical Information TI00373C
Ceragel CPS72D
• ORP electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps72d
Technical Information TI00374C
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Product Configurator on the product page: www.endress.com/cpf82d
Technical Information TI00191C
Orbipore CPS92D
• ORP electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps92d
Technical Information TI00435C
pH ISFET sensors
Tophit CPS441D
• Sterilizable ISFET sensor for low-conductivity media
• Liquid KCl electrolyte
• Product Configurator on the product page: www.endress.com/cps441d
Technical Information TI00352C
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
• Water treatment and biotechnology
• Product Configurator on the product page: www.endress.com/cps471d
Technical Information TI00283C
Tophit CPS491D
• ISFET sensor with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps491d
Technical Information TI00377C
Endress+Hauser13
Page 14
pH and ORP combined sensors
Memosens CPS16D
• Combined pH/ORP sensor for process technology
• With dirt-repellent PTFE diaphragm
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps16D
Technical Information TI00503C
Memosens CPS76D
• Combined pH/ORP sensor for process technology
• Hygienic and sterile applications
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps76d
Technical Information TI00506C
Memosens CPS96D
• Combined pH/ORP sensor for chemical processes
• With poison-resistant reference with ion trap
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps96d
Technical Information TI00507C
Liquiline Compact CM82
Conductivity sensors with conductive measurement of conductivity
Condumax CLS15D
• Conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• Product Configurator on the product page: www.endress.com/CLS15d
Technical Information TI00109C
Condumax CLS16D
• Hygienic, conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• With EHEDG and 3A approval
• Product Configurator on the product page: www.endress.com/CLS16d
Technical Information TI00227C
Condumax CLS21D
• Two-electrode sensor in plug-in head version version
• Product Configurator on the product page: www.endress.com/CLS21d
Technical Information TI00085C
Memosens CLS82D
• Four-electrode sensor
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cls82d
Technical Information TI01188C
Oxygen sensors
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• With Memosens technology or as an analog sensor
• Product Configurator on the product page: www.endress.com/cos22d
Technical Information TI00446C
14Endress+Hauser
Page 15
Liquiline Compact CM82
Oxymax COS51D
• Amperometric sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos51d
Technical Information TI00413C
Memosens COS81D
• Sterilizable, optical sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos81d
Technical Information TI01201C
Software
Other accessoriesActivation codes
Memobase Plus CYZ71D
• PC software to support laboratory calibration
• Visualization and documentation of sensor management
• Sensor calibrations stored in database
• Product Configurator on the product page: www.endress.com/cyz71d
Technical Information TI00502C
You must quote the serial number of the device when ordering the activation code.
Activation code: Bluetooth
Order No. 71401176
Cable junction with Velcro strip
Cable junction with Velcro strip
• 4 pieces, for sensor cable
• Order No. 71092051
Communication-specific accessories
Commubox FXA195
Intrinsically safe HART communication with FieldCare via the USB port
Technical Information TI00404F
Wireless HART adapter SWA70
• Wireless device connection
• Easily integrated, offers data protection and transmission safety, can be operated in parallel with
other wireless networks, minimum cabling complexity
Technical Information TI00061S
System components
RIA15
• Process display unit, Digital display unit for integration into 4-20 mA circuits
• Panel mounting
• With optional HART communication
Technical Information TI01043K
Endress+Hauser15