Conductivity sensor with inductive measurement of
conductivity and hygienic design for applications in
the food, beverage and pharmaceutical industries
and in biotechnology
Indumax CLS54DTable of contents
Table of contents
1Document information ......... 4
1.1Warnings ........................... 4
1.2Symbols used ........................ 4
1.3Symbols on the device ................ 5
2Basic safety instructions ....... 6
2.1Requirements for the personnel ........ 6
2.2Designated use ...................... 6
2.3Occupational safety .................. 6
2.4Operational safety ................... 7
2.5Product safety ....................... 7
3Incoming acceptance and
product identification .......... 8
3.1Incoming acceptance ................. 8
3.2Scope of delivery ..................... 8
3.3Product identification ................. 9
3.4Certificates and approvals ............ 10
4Installation .................... 11
4.1Installation conditions ............... 11
4.2Installation ........................ 13
4.3Post-installation check ............... 13
5Electrical connection .......... 14
5.1Connecting to the transmitter ......... 14
5.2Ensuring the degree of protection ..... 15
5.3Post-connection check ............... 15
6Maintenance .................. 16
7Repairs ......................... 16
7.1Return ............................ 16
7.2Disposal ........................... 16
8Accessories .................... 17
8.1Cable extension ..................... 17
8.2Calibration solutions ................ 17
9Technical data ................. 18
Index ................................. 21
Endress+Hauser3
Document informationIndumax CLS54D
1 Document information
1.1 Warnings
Structure of informationMeaning
DANGER
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
WARNING
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
CAUTION
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
NOTICE
Cause/situation
If necessary, Consequences of noncompliance (if applicable)
Action/note
‣
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols used
SymbolMeaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
4Endress+Hauser
Indumax CLS54DDocument information
1.3 Symbols on the device
SymbolMeaning
Reference to device documentation
Endress+Hauser5
Basic safety instructionsIndumax CLS54D
2 Basic safety instructions
2.1 Requirements for the personnel
• Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified
activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
• Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried out only
directly at the manufacturer's site or by the service organization.
2.2 Designated use
Indumax CLS54D is designed for the inductive measurement of the conductivity of liquids in
the food and beverages industry.
The six-decade measuring range and the excellent chemical resistance properties of the
materials in contact with the medium make it possible to use this sensor in a wide range of
applications, such as:
• Concentration measurement of acids and bases
• Phase separation of products
The sensor is used with Liquiline CM44x/R/P, Liquiline CM42 or Liquiline CM14.
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
NOTICE
Non-designated use
Incorrect measurements, malfunctions and even measuring point failure could result
Only use the product in accordance with the product specifications.
‣
Pay particular attention to the technical data on the nameplate.
‣
2.3 Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
6Endress+Hauser
Indumax CLS54DBasic safety instructions
2.4 Operational safety
1.Before commissioning the entire measuring point, verify that all connections are
correct. Ensure that electrical cables and hose connections are undamaged.
2.Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Label the damaged product as defective.
3.If faults cannot be rectified:
Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5 Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
Endress+Hauser7
Incoming acceptance and product identificationIndumax CLS54D
3 Incoming acceptance and product identification
3.1 Incoming acceptance
1.Verify that the packaging is undamaged.
Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2.Verify that the contents are undamaged.
Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3.Check the delivery for completeness.
Check it against the delivery papers and your order.
4.Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
3.2 Scope of delivery
The scope of delivery includes:
• Sensor in the version ordered
• Operating Instructions
8Endress+Hauser
Indumax CLS54DIncoming acceptance and product identification
3.3 Product identification
3.3.1 Nameplate
The nameplate can be found on the sensor.
The following information is provided on the nameplate:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Cell constant (nominal value)
• Protection class
• Pressure specification at 20 °C
• Continuous service temperature
Compare the data on the nameplate with your order.
3.3.2 Product identification
Product page
www.endress.com/cls54D
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1.Go to the product page for your product on the Internet.
2.At the bottom of the page, click the link Online Tools and then select Access devicespecific information.
An additional window opens.
3.Enter the order code from the nameplate into the search field and then select Showdetails.
You will receive information on each feature (selected option) of the order code.
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
Endress+Hauser9
Incoming acceptance and product identificationIndumax CLS54D
3.4 Certificates and approvals
3.4.1 Hygiene
FDA
All materials in contact with the product are listed by the FDA.
EHEDG
Certified cleanability according to EHEDG TYPE EL-class I.
When using the sensor in hygienic applications, please note that the cleanability of the
sensor also depends on the way the sensor is installed. To install the sensor in a pipe, use
the appropriate and EHEDG-certified flow vessels for the particular process connection.
3-A
Certified according to 3-A Standard 74- ("3-A Sanitary Standards for Sensor and Sensor
Fittings and Connections Used on Milk and Milk Products Equipment").
Biological reactivity (USP class VI) (option)
Biological reactivity test certificate (Certificate of Compliance) according to USP (United States
Pharmacopoeia) part<87> and part <88> class VI with lot number traceability of materials in
contact with the medium.
3.4.2 mark
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EU directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
10Endress+Hauser
Indumax CLS54DInstallation
> 1 m (3.3 ft)
4 Installation
4.1 Installation conditions
4.1.1 Orientation
The sensor must be completely immersed in the medium. Avoid air bubbles in the area of the
sensor.
A0017691
1Installation positions of the conductivity sensor
If the flow direction changes (after pipe bends), turbulence in the medium can result.
Install the sensor at a distance of at least 1 m (3.3 ft) downstream from a pipe bend.
The product should flow along the hole of the sensor (see the arrows on the housing). The
symmetrical measuring channel allows flow in both directions.
For a 3-A compliant installation, the following must be noted:
After the instrument is installed its hygienic integrity shall be maintained. All process
connections must be 3-A compliant.
Endress+Hauser11
InstallationIndumax CLS54D
a
1
2
a[inch]
0
5
1015
20
2525
a[mm]
0.80
1.00
1.20
1.40
f
0.20
0.39
0.59
0.79
0.98
4.1.2 Installation factor
The ionic current in the liquid is affected by the walls in
confined installation conditions. This effect is compensated
by what is referred to as the installation factor. The
installation factor can be entered in the transmitter for the
measurement or the cell constant is corrected by multiplying
by the installation factor.
The value of the installation factor depends on the diameter
and the conductivity of the pipe nozzle as well as the
distance a between the sensor and the wall.
The installation factor f (f = 1.00) can be disregarded if the
distance to the wall is sufficient (a > 15 mm, from DN 65).
If the distance to the wall is smaller, the installation factor
increases for electrically insulating pipes (f > 1), and
decreases for electrically conductive pipes (f < 1).
It can be measured using calibration solutions, or a close
approximation can be determined from the diagram below.
2Installation of CLS54D
aWall distance
A0005453
3Relationship between installation factor f and wall distance a
The digital sensor has already been adjusted at the factory. Onsite compensation is not
required.
12Endress+Hauser
A0005441
Indumax CLS54DInstallation
a
1
4.2 Installation
A0032586
4Installed length of the sensor
1Direction of medium flow
aDistance from pipe wall
When installing, align the sensor in such a way that the medium flows through the flow
‣
opening of the sensor in the direction of medium flow. The sensor head must be completely
immersed in the medium.
4.3 Post-installation check
Put the sensor into operation only if you can answer "yes" to the following questions:
• Are the sensor and cable undamaged?
• Is the orientation correct?
• Has the sensor been installed in the process connection, and does not suspend freely from
the cable?
Endress+Hauser13
Electrical connectionIndumax CLS54D
2
A
B
20 (0.79)
20 (0.79)
PK
85
GY
86
BN
87
WH
88
GN
97
YE
98
63 (2.48)
75 (2.95)
80 (3.15)
1
5 Electrical connection
WARNING
L
Device is live
Incorrect connection may result in injury or death
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions and
‣
must follow the instructions contained therein.
Prior to commencing connection work, ensure that no voltage is present on any cable.
‣
5.1 Connecting to the transmitter
The sensor is supplied with a fixed cable. The wiring diagram is provided in the Operating
Instructions of the transmitter used.
Connection via a junction box is necessary for a cable connection. The extension to the
transmitter is via the CYK11 cable.
5Connection with CYK11 cable extension via junction box, dimensions in mm (inch)
1Cable glands - shield fixed in gland
2Shielding
ACYK11 from transmitter
BSensor cable
Sensors with a fixed cable and M12 plug can be extended with the CYK11 measuring cable
and an M12 socket.
14Endress+Hauser
A0032587
Indumax CLS54DElectrical connection
A0017842
6CYK11 for extension with M12 connection
1Transmitter
2CYK11 measuring cable with M12 connection
ACLS54D connecting cable with M12 plug
BSensor CLS54D
5.2 Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions and
which are necessary for the required, designated use, may be carried out on the device
delivered.
Exercise care when carrying out the work.
‣
Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC
interference immunity) agreed for this product can no longer be guaranteed due, for example,
to covers being left off or cable (ends) which are loose or insufficiently secured.
5.3 Post-connection check
Device condition and specificationsNotes
Are the outside of the sensor, assembly, cable undamaged?Visual inspection
Electrical connectionNotes
Are the installed cables strain-relieved and not twisted?
Is a sufficient length of the cable cores stripped, and is it positioned
in the terminal correctly?
Are all the screws terminals properly tightened?Tighten
Are all cable entries mounted, tightened and leak-tight?For lateral cable entries, make sure the cables
Are all cable entries installed downwards or mounted laterally?
Endress+Hauser15
Check the fit (by pulling gently)
loop downwards to allow water to drip off
MaintenanceIndumax CLS54D
6 Maintenance
CAUTION
L
Corrosive chemicals
Danger of chemical burns to the eyes and skin. Danger of damage to clothing and equipment
It is absolutely essential to protect the eyes and hands properly when working with acids,
‣
bases and organic solvents!
Wear protective goggles and safety gloves.
‣
Clean away splashes on clothes and other objects to prevent any damage.
‣
Pay particular attention to the information provided in the safety data sheets for the
‣
chemicals used.
As there is no galvanic contact with the medium, inductive sensors are considerably less
sensitive to dirt and fouling than conventional conductive sensors.
However, dirt can clog the measuring channel which, in turn, can alter the cell constant. In
such cases, an inductive sensor also needs to be cleaned.
Clean away fouling on the sensor as follows depending on the type of fouling:
• Oily and greasy films:
Clean with grease remover, e.g. alcohol, acetone, possibly hot water and dishwashing
detergent.
• Lime and metal hydroxide buildup:
Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with plenty
of clear water.
• Sulfidic buildup (from flue gas desulfurization or sewage treatment plants):
Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available) and
then rinse thoroughly with plenty of clear water.
• Buildup containing proteins (e.g. food industry):
Use a mixture of hydrochloric acid (0.5 %) and pepsin (commercially available) and then
rinse thoroughly with plenty of clear water.
7 Repairs
7.1 Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at www.endress.com/support/return-material.
7.2 Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
16Endress+Hauser
Indumax CLS54DAccessories
Observe the local regulations.
8 Accessories
8.1 Cable extension
8.1.1 Measuring cable
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C
8.1.2 Junction box
Junction box, M12 socket/cable
• Material: aluminum, painted
• Cable extension: Memosens sensors, Liquiline
• Order No.: 71145498
Junction box, cable/cable
• Material: aluminum, painted
• Cable extension: Memosens sensors, Liquiline
• Order No.: 71145499
8.2 Calibration solutions
Conductivity calibration solutions CLY11
Precision solutions referenced to SRM (Standard Reference Material) by NIST for qualified
calibration of conductivity measuring systems in accordance with ISO 9000
• CLY11-B, 149.6 μS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081903
• CLY11-C, 1.406 mS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081904
• CLY11-D, 12.64 mS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081905
• CLY11-E, 107.00 mS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081906
Technical Information TI00162C
Endress+Hauser17
Technical dataIndumax CLS54D
9 Technical data
9.1 Input
9.1.1 Measured values
• Conductivity
• Temperature
9.1.2 Measuring range
ConductivityRecommended range: 100 μS/cm to 2000 mS/cm
(uncompensated)
Temperature–10 to +150 °C (+14 to +302 °F)
9.1.3 Cell constant
k = 6.3 cm
9.1.4 Temperature measurement
Pt1000 (Class A according to DIN EN 60751)
9.2 Performance characteristics
9.2.1 Temperature response time
t90 ≤ 26 s
–1
9.2.2 Conductivity response time
t95 ≤ 2 s
9.2.3 Maximum measured error
< 100 °C (212 °F):
> 100 °C (212 °F):±(25 μS/cm + 0.5 % of reading), after calibration
9.2.4 Repeatability
0.2 % of reading + 3 µS/cm
±(10 μS/cm + 0.5 % of reading), after calibration
9.3 Environment
9.3.1 Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
9.3.2 Storage temperature
-25 to +80 °C (-13 to +176 °F)
18Endress+Hauser
Indumax CLS54DTechnical data
A
5070
1 011 05
1030
125
90
0
–10
[° C]
9
13
[bar]
32
50
86
122
158
230
302194
257
[°F]
130.5
188.5
[psi]
6
B
1
87
14.5
p (abs.)
9.3.3 Relative humidity
5 to 95 %
9.3.4 Degree of protection
IP 68 / NEMA type 6P (1 m water column, 25 °C, 168 h)
9.4 Process
9.4.1 Process temperature
-10 to +125 °C (+14 to +257 °F)
9.4.2 Sterilization
150 °C (302 °F) / 6 bar (87 psi) absolute, (max. 60 min.)
9.4.3 Process pressure (absolute)
13 bar (188.5 psi) up to 90 °C (194 °F)
9 bar (130.5 psi) at 125 °C (257 °F)
Underpressure down to 0.1 bar (1.45 psi)
9.4.4 Temperature/pressure ratings
7Pressure/temperature ratings
ATemporarily for sterilization (max. 60 min.)
BMAWP (maximum allowable working pressure) according to ASME-BPVC Sec. VIII, Div 1 UG101 for
CRN registration
Endress+Hauser19
A0008379
Technical dataIndumax CLS54D
9.5 Mechanical construction
9.5.1 Weight
0.3 to 0.5 kg (0.66 to 1.1 lb.) depending on version plus cable
9.5.2 Materials
In contact with mediumVirgin PEEK
Not in contact with mediumPPS-GF40