Hygienic conductivity sensors, analog or digital
with Memosens technology
Cell constant k = 0.1 cm
Application
Measurement in pure and ultrapure water range:
• Monitoring of ion exchangers
• Reverse osmosis
• Distillation
• Electrodeionization
• WFI (water for injection) in the pharmaceutical industry
Sensors with temperature probes are used in conjunction with conductivity
measuring devices that support automatic temperature compensation:
• Liquiline CM442/CM444/CM448 (only CLS16D)
• Liquiline CM42
• Liquiline CM14 (only CLS16D)
• Liquisys CLM223/253 (only CLS16)
• Mycom CLM153 (only CLS16)
The resistivity in MΩ · cm can also be measured using these transmitters.
Your benefits
• High measuring accuracy as cell constant is individually measured
• Hygienic process connections for installation in pipes or flow vessel
• With plug-in head (IP68) or fixed cable (IP67)
• Easy to clean thanks to electropolished surfaces
• Sterilizable up to 150 °C (302 °F)
• Quality certificate stating the individual cell constant
• Certified according to EHEDG, Document 8
• Certificate according to United States Pharmacopeia Class VI Part 87 and 88
(optional)
• Inspection certificate EN 10204 3.1 (optional)
Other advantages of Memosens technology
• Maximum process safety owing to non-contact, inductive signal transmission
• Data security thanks to digital data transmission
• Very easy to use as sensor data saved in the sensor
• Predictive maintenance possible as sensor load data logged in the sensor
–1
Function and system design
AC
I
U
1
2
3
Condumax CLS16D/CLS16
Measuring principle
Measuring system
Conductivity of liquids is determined with a measuring
arrangement where two electrodes are located in the
medium. An alternating voltage that causes a current
to flow through the medium is applied at these
electrodes. The electrical resistance, or its reciprocal
value - conductance G - is calculated based on Ohm's
law. The specific conductance κ is determined from the
conductance value using the cell constant k, which
depends on the sensor geometry.
A0024260
1Conductive measurement of conductivity
AC Alternating voltage source
ICurrent intensity measurement
UVoltage measurement
A complete measuring system comprises the following components at least:
• The conductively measuring conductivity sensor CLS16D or CLS16
• A transmitter, e.g. Liquiline M CM42
• A measuring cable, e.g. Memosens data cable CYK10 or CPK9 for analog sensors
2Example of a measuring system (with Memosens sensor)
1Liquiline M CM42 transmitter
2Memosens data cable
3Condumax CLS16D
2Endress+Hauser
A0024286
Condumax CLS16D/CLS16
Communication and data processing (only CLS16D)
Communication with the transmitter
Always connect digital sensors with Memosens technology to a transmitter with Memosens
technology. Data transmission to a transmitter for analog sensors is not possible.
Digital sensors are able to store the following system data in the sensor
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Cell constant
– Delta cell constant
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– Conductivity application range
– Date of first commissioning
– Maximum temperature value
– Hours of operation at high temperatures
Reliability
Maintainability
Integrity
Dependability (only CLS16D)
Memosens technology digitizes the measured values in the sensor and transmits the data to the
transmitter using a non-contact connection that is free from potential interference. The result:
• Automatic error message if sensor fails or connection between sensor and transmitter is
interrupted
• Immediate error detection increases measuring point availability
Easy handling
Sensors with Memosens technology have integrated electronics that allow for saving calibration data
and further information such as total hours of operation and operating hours under extreme
measuring conditions. Once the sensor has been connected, the sensor data are transferred
automatically to the transmitter and used to calculate the current measured value. As the calibration
data are stored in the sensor, the sensor can be calibrated and adjusted independently of the
measuring point. The result:
• Easy calibration in the measuring lab under optimum external conditions increases the quality of
the calibration.
• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the
availability of the measuring point .
• Maintenance intervals can be defined based on all stored sensor load and calibration data and
predictive maintenance is possible.
• The sensor history can be documented on external data carriers and evaluation programs at any
time. Thus, the current application of the sensors can be made to depend on their previous history.
With inductive transmission of the measured value using a non-contact connection, Memosens
guarantees maximum process safety and offers the following benefits:
• All problems caused by moisture are eliminated.
– Plug-in connection free from corrosion
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.
• The transmitter is galvanically decoupled from the medium.
• EMC safety is guaranteed by screening measures for the digital transmission of measured values.
Endress+Hauser3
Input
GN/YE
YE
GN
BN
WH
GND
+
–
Com A
Com B
GY
Outer electrode
Inner electrode
BN
n.c.
BK
WH
GN
WH
YE
Temp.
Temp.
Temp.
GY
Outer shield
(see wiring diagram
of transmitter)
Condumax CLS16D/CLS16
Measured variables
• Conductivity
• Temperature
Measuring ranges
Conductivity(in relation to water at 25 °C (77 °F))
CLS16D/CLS160.04 to 500 μS/cm
Temperature
CLS16D
CLS16
Cell constant
Temperature compensation
CLS16D/CLS16
NTC 30K (CLS16D)
Pt 100 (class A as per IEC 60751) (CLS16)
Pt 1000 (class A as per IEC 60751)(CLS16, optional)
Power supply
Electrical connectionCLS16D
The sensor is electrically connected to the transmitter via the CYK10 measuring cable.
-5 to 100 °C (23 to 212 °F)
-5 to 150 °C (23 to 300 °F)
k = 0.1 cm
–1
A0024019
3Measuring cable CYK10
CLS16
The sensor is electrically connected via the CPK9 measuring cable (plug-in head versions) or the
sensor's fixed cable. The wiring diagram is provided in the Operating Instructions of the transmitter
used.
A0024206-EN
4Measuring cable CPK9
A VMB junction box and a CYK71 cable are required for the cable extension.
4Endress+Hauser
Condumax CLS16D/CLS16
Performance characteristics, general
Uncertainty of measurement
Response time
Maximum measured error
Repeatability
Each individual sensor is factory-measured in a solution of approx. 5 μS/cm using a reference
measuring system traceable to NIST or PTB. The exact cell constant is entered into the quality
certificate supplied. The uncertainty of measurement in determining the cell constant is 1.0 %.
Performance characteristics (only CLS16D)
Conductivity
Temperaturet90 ≤ 13 s
2 % of reading up to 200 μS/cm
3 % of reading from 200 to 500 μS/cm
0.2% of reading
t95 ≤ 3 s
Endress+Hauser5
Installation
Condumax CLS16D/CLS16
Installation instructions
Ambient temperature range
Storage temperature
The sensors are installed directly via the process connection. Take the flow direction into
consideration when installing in pipes.
A0019016
5Permitted flow direction
6Inadmissible flow direction
Environment
-20 to +60 °C (0 to 140 °F)
-25 to +80 °C (-10 to +180 °F)
A0019017
Humidity
Degree of protection
5 to 95 %
CLS16D
CLS16
Fixed cable versionIP 67 / NEMA 6
TOP68 plug-in systemIP 68 / NEMA 6
IP 68 / NEMA type 6P (1 m water column, 25 °C, 24 h)
6Endress+Hauser
Condumax CLS16D/CLS16
6
–5
2060100 120 150
23
60
68
120
180
240
248
300
2060100 120 150
302
A
[psi]
[°F]
[bar]
[°C]
p (abs.)
T
14
43
72
101
130
159
188
1
3
5
7
9
11
13
87
1.50.1
}
1
2
3
4
Process
Process temperature
Process pressure (absolute)
Temperature-pressure
ratings
Normal operation-5 to 120 °C (23 to 248 °F)
Sterilization (max. 45 min.)max. 150 °C (302 °F) at 5 bar (73 psi)
The maximum temperature for communication with the transmitter is 130 °C (266°F) for the
Memosens versions.
CLS16D/CLS1613 bar (188 psi) at 20 °C (68 °F)
9 bar (130 psi) at 120 °C (248 °F)
0.1 bar (1.5 psi (negative pressure)) at 20 °C (68 °F)
A0024880
7Mechanical pressure-temperature resistance
ACan be sterilized for a short time (45 min.)
Design CLS16D
Endress+Hauser7
Mechanical construction
A0024294
8CLS16D
1Memosens plug-in head
2Process connection made of electropolished stainless steel 1.4435 (AISI 316 L) (Clamp, Varivent, BioControl)
3Coaxial measuring electrode made of electropolished stainless steel 1.4435 (AISI 316 L)
4Minimum immersion depth
Clamp 1", 1½", 2" as per ISO 2852 (also suitable for TRI-CLAMP, DIN 32676)
Tuchenhagen VARIVENT N DN 50 to 125
NEUMO BioControl D50
Ra ≤ 0.8 μm, electropolished
Ra ≤ 0.38 μm, electropolished, optional
Certificates and approvals
Declaration of Conformity
The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EC directives. The manufacturer confirms successful testing of the
product by affixing to it the mark.
CLS16D-**G
ATEX / NEPSI II 1G Ex ia IIC T3/T4/T6 Ga, IECEx Ex ia IIC T3/T4/T6 Ga
CLS16D-**O
FM/CSA IS/NI Cl. I Div.1&2 Gr. A-D in conjunction with Liquiline M CM42 transmitter
CLS16D-**V
ATEX/NEPSI II 3G Ex ic IIC T3/T4/T6 Gc for use in Zone 2 with Liquiline M CM42-KV*** transmitter
CLS16
ATEX II 1G EEx ia IIC T3/T4/T6
FM/CSA IS/NI Cl. I Div.1&2 Gr. A-D in conjunction with Liquiline M CM42 transmitter
ATEX and FM/CSA versions of digital sensors with Memosens technology are indicated by a
red-orange ring in the plug-in head.
EHEDG
FDA
Quality certificate
Biological reactivity test
(USP class VI)
Inspection certificate in
accordance with EN 10204
3.1
ASME
Validated as follows:
• Cleanability in accordance with EHEDG, Document 2
• Sterilizability in accordance with EHEDG, Document 5
• Bacteria-tightness in accordance with EHEDG, Document 7
All materials in contact with the product are listed by the FDA.
Stating the individual cell constant
Biological reactivity test certificate according to USP (United States Pharmacopeia) class VI part
<87> and part <88> with batch traceability of materials in contact with the medium
A test certificate 3.1 in accordance with EN10204 is supplied depending on the version (→ Product
Configurator on the product page).
Manufactured in accordance with ASME criteria (American Society of Mechanical Engineers)
Endress+Hauser11
Ordering information
Condumax CLS16D/CLS16
Product page
Product Configurator
Scope of delivery
www.endress.com/cls16d
www.endress.com/cls16
The navigation area is located on the right of the product page.
1.Under "Device support" click "Configure your selected product".
The Configurator opens in a separate window.
2.Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3.Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of
the screen.
The scope of delivery includes:
• Sensor in the version ordered
• Operating Instructions
12Endress+Hauser
Condumax CLS16D/CLS16
Accessories
The following are the most important accessories available at the time this documentation was
issued. For accessories not listed here, please contact your service or sales office.
Measuring cable
Junction boxes (only CLS16)
CYK10 Memosens data cable
• For digital sensors with Memosens technology
• Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C
CPK9
• Terminated measuring cable for connecting analog sensors with TOP68 plug-in head
• Selection in accordance with product structure
For more information and to order, please contact your sales office.
• Junction box for cable extension in hazardous area
• 10 terminal strips (blue)
• Cable entries: 2 x Pg 13.5
• Material: aluminum
•
Degree of protection: IP 65 ( NEMA 4X)
• Order number: 50003991
Sensor regeneration
Calibration solutions
Calibration set
Replacement of seals and recalibration in the factory
Order No. 51505585
Conductivity calibration solutions CLY11
Precision solutions referenced to SRM (Standard Reference Material) by NIST for qualified
calibration of conductivity measuring systems in accordance with ISO 9000
• CLY11-A, 74 μS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081902
• CLY11-B, 149.6 μS/cm (reference temperature 25 °C (77 °F)), 500 ml (16.9 fl.oz)
Order No. 50081903
Technical Information TI00162C
Conducal CLY421
• Conductivity calibration set (case) for ultrapure water applications
• Complete, factory-calibrated measuring system with certificate, traceable to SRM by NIST and PTB,
for comparison measurement in ultrapure water up to max. 20 μS/cm
• Product Configurator on the product page: www.endress.com/cly421
Technical Information TI00496C/07/EN
Endress+Hauser13
www.addresses.endress.com