Emerson Process Management CFA-3000 User Manual

CFA-3000
Ion Selective Electrode (ISE) Analyzer
Instruction Manual
PN 51CFA3000ISE/rev.F December 2013
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in oper ating and maintaining the instrument. To ensure their continued operation to the design specifications, per sonnel should read this manual thoroughly before proceeding with installation, commissioning, opera tion, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 18008548257 or 9497578500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pressure sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
• All equipment doors must be closed and protective covers must be in place unless qualified personnel are performing maintenance.
• If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
WARNINGS
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
• Installation of cable connections and servicing of this product require access to shock hazard voltage levels.
• Main power and relay contacts wired to separate power source must be disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least 240 V.
• Nonmetallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bush ings and jumper wires.
• Unused cable conduit entries must be securely sealed by nonflammable closures to provide enclosure integrity in compliance with personal safety and environmental protec tion requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA70) and/or any other applicable national or local codes.
• Operate only with front and rear panels fastened and in place over terminal area.
• Safety and performance require that this instrument be connected and properly grounded through a threewire power source.
Proper relay use and configuration is the responsibility of
the user.
CAUTION
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.
WARNING
This product is not intended for use in the light industrial, residential or commercial environments per the instrument’s certification to EN500812.
Emerson Process Management
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 7578500 Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2012
CFA3000 ISE Analyzer TABLE OF CONTENTS Instruction Manual
1.0 INTRODUCTION...........................................................................................................................1
1.1 Thoroughly Read This Manual......................................................................................................2
1.2 Analyzer Benefits...........................................................................................................................2
1.3 WARNING  Electrical Shock Hazard..........................................................................................3
2.0 INSTALLING THE ANALYZER....................................................................................................4
2.1 Unpack Parts .................................................................................................................................4
2.2 Mount Main Cabinet ......................................................................................................................6
2.3 Mount Reagent Cabinet ................................................................................................................6
2.4 Install Waste Tube .........................................................................................................................8
2.5 Mount Overflow Sampling Assembly............................................................................................8
2.6 Prepare Electrical Connections...................................................................................................10
2.7 Install Valve Pump Assembly ......................................................................................................11
2.8 Prepare Reagent Containers ......................................................................................................13
3.0 OPERATING THE ANALYZER ..................................................................................................14
3.1 Operational Overview..................................................................................................................14
3.2 Check Sample Stream ................................................................................................................15
3.2.1 Check Fluidics .............................................................................................................................15
3.2.2 Check Reagents..........................................................................................................................15
3.2.3 Turn On Power ............................................................................................................................15
3.2.4 Test Valve Pump Operation ........................................................................................................16
3.2.5 Set Up Stream Parameters ........................................................................................................20
3.3 Set Reagent Supply Time ..........................................................................................................22
3.4 Set Scaling Outputs.....................................................................................................................22
3.4.1 Record mV ................................................................................................................................23
3.4.2 Initiate Auto Calibration ..............................................................................................................23
3.5 Process a Grab Sample .............................................................................................................24
3.6 Manually Sample a Particular Stream .......................................................................................25
3.7 Shutdown ................................................................................................................................25
3.8 Quick Start ................................................................................................................................26
4.0 MAINTAINING THE ANALYZER ...............................................................................................27
4.1 Analyzer Equipment ....................................................................................................................27
5.0 TROUBLESHOOTING PROCEDURES....................................................................................28
5.1 Troubleshooting Chart.................................................................................................................28
5.2 Test Functions .............................................................................................................................29
6.0 REPAIR PROCEDURES ............................................................................................................31
6.1 Replace Valve Pump Assembly..................................................................................................31
6.2 Replace Circuit Boards ...............................................................................................................32
6.3 Replace Fuses.............................................................................................................................33
6.4 Replace ISE Components...........................................................................................................33
7.0 INSTRUMENT DESCRIPTION ..................................................................................................34
7.1 Analyzer Front View ....................................................................................................................34
7.1.1 Main Cabinet ...............................................................................................................................34
7.1.2 Reagent Cabinet .........................................................................................................................34
7.2 Keypad Panel ..............................................................................................................................36
7.3 Card Cage 41 ............................................................................................................................37
7.3.1 Card Cage and Door ...................................................................................................................37
7.3.2 CPU Board 41 ............................................................................................................................37
7.3.3 Analog Board...............................................................................................................................37
7.3.4 Valve Board 41 ............................................................................................................................37
7.3.5 Power Supply Board ...................................................................................................................37
i
CFA3000 ISE Analyzer
TABLE OF CONTENTS
ii
CFA3000 ISE Analyzer TABLE OF CONTENTS Instruction Manual
7.3.6 Analog Board DIP Switches........................................................................................................39
7.4 Rear of Main Cabinet ..................................................................................................................40
7.5 Valve Pump Assembly.................................................................................................................41
7.6 Interface Ports .............................................................................................................................42
7.7 Relay Output Option....................................................................................................................42
7.8 MultiOutput Board ......................................................................................................................42
7.9 Overflow Sampling Assemblies ..................................................................................................43
8.0 ADDITIONAL FEATURES .........................................................................................................45
8.1 Chart Recorders and Data Loggers............................................................................................45
8.2 Computer Interface......................................................................................................................48
8.2.1 Data Output Format.....................................................................................................................48
8.2.2 Data Configuration.......................................................................................................................48
8.2.3 RS232C Function Commands...................................................................................................48
8.2.4 Initiating RS232C Communications to Computer .....................................................................49
8.2.5 Printer Interface ...........................................................................................................................51
8.2.6 Set Printer Internal Clock ............................................................................................................51
Appendix A. CHEMISTRIES..................................................................................................................52
Appendix B. SPECIFICATIONS ........................................................................................................5354
Appendix C. POWER INTERCONNECT DESCRIPTION .....................................................................55
Appendix D. BYPASS FOR CALIBRATION..........................................................................................56
Appendix E. KEYPAD HELP SHEETS .................................................................................................57
ORDERING INFORMATION ....................................................................................................................63
WARRANTY ................................................................................................................................64
9.0 RETURN OF MATERIAL............................................................................................................65
TABLE OF CONTENTS (CON’T)
iii
TABLE OF CONTENTS (CON’T)
CFA3000 ISE Analyzer TABLE OF CONTENTS Instruction Manual
LIST OF FIGURES
Figure 21 Opened Assembled Cabinet....................................................................................5
Figure 22 Mounting Main and Reagent Cabinets ....................................................................7
Figure 23 Attaching Grounding Strap.......................................................................................7
Figure 24a Singlestream Overflow Sampling Assembly...........................................................9
Figure 24b Multistream Overflow Sampling Assembly.............................................................9
Figure 25 Input/Output Panel..................................................................................................11
Figure 26 Valve Pump Assembly............................................................................................12
Figure 27 Pump Connectors ..................................................................................................12
Figure 28 Sample Bottle Placement in Cabinet (for all methods; colors refer to
reagent straw colors)..............................................................................................13
Figure 31 Simple Block Diagram............................................................................................14
Figure 32 Numbered Positions on a SingleStream Keypad.................................................17
Figure 33 Valve Pump LED Designations..............................................................................19
Figure 34 Logic Flow for Singlestream Parameter Definition...............................................20
Figure 35 Logic Flow for Multistream Parameter Definition .................................................21
Figure 61 Remove Valve Pump Assembly.............................................................................31
Figure 62 Card Cage ..............................................................................................................32
Figure 63 ISE Reaction Cell Assembly, Electrodes ...............................................................33
Figure 71 Analyzer Front View ...............................................................................................35
Figure 72 Card Cage and Boards ..........................................................................................38
Figure 73 Analog Board DIP Switches (with #3 set in OFF position for multi..........................
stream analyzer)....................................................................................................39
Figure 74 Electrical Wiring and Components in Rear Cabinet..............................................40
Figure 75 Valve Pump Assembly............................................................................................41
Figure 76 Interface Ports ........................................................................................................42
Figure 77 "D" Connector for Relay Output.............................................................................42
Figure 78 SingleStream Overflow Sampling Assembly ........................................................43
Figure 79 Multistream Overflow Sampling Assembly (threestream option)........................44
Figure 81 Input/Output Panel .................................................................................................45
Figure 82 Typical Trace Recording with Calibration of Single Stream System.....................46
Figure 83 Typical Stream Trace Recording with Calibration of Multistream
System....................................................................................................................47
Figure 84 Numbered Positions on a SingleStream Keypad................................................49
Figure D1 Calibration Bypass Wiring......................................................................................56
Figure E1 Keypad Panel for Single Stream Option................................................................57
Figure E2 Number Positions on a SingleStream Keypad....................................................57
Figure E3 Keypad Panel for Multistream Option...................................................................60
iv
CFA3000 ISE Analyzer TABLE OF CONTENTS Instruction Manual
TABLE OF CONTENTS (CON’T)
LIST OF TABLES
Table 51 Troubleshooting Chart.....................................................................................28
Table 52 Test Functions .................................................................................................29
Table 71 Main Cabinet Parts..........................................................................................34
Table 72 Reagent Cabinet Parts ....................................................................................34
Table 73 Buttons/LED Functions ...................................................................................36
Table 74 Card Cage Parts .............................................................................................37
Table 75 CPU Board Components.................................................................................37
Table 76 Analog Board Components .............................................................................37
Table 77 Valve Board Components ...............................................................................37
Table 78 Power Supply Board Components ..................................................................37
Table 79 Analog Board DIP Switch Positions ................................................................39
Table 710 Back Cabinet Parts .........................................................................................40
Table 711 Valve Pump Parts............................................................................................41
Table 712 Interface Ports.................................................................................................42
Table 713 Relay Output Option Pin Assignments ............................................................42
Table 714 Multioutput Pin Assignments on “D” Connector .............................................42
Table 715 Overflow Sampling Assembly Parts ................................................................43
Table 81 Stream Identification Blips (SIBs)....................................................................45
Table 82 Normal Data Output Format............................................................................48
Table 83 Data Configuration ..........................................................................................48
Table 84 RS232C Pin Configuration .............................................................................48
Table 85 RS232C Commands ......................................................................................50
Table 86 Printer Port Pin Configuration .........................................................................51
Table E1 Singlestream Keypad Help Sheet.............................................................5859
Table E2 Multistream Keypad Help Sheet ...............................................................6062
Table E3 Replacement Parts .........………..……………………………………….…….63
1
CFA3000 MANUAL SECTION 1
INTRODUCTION
SECTION 1
INTRODUCTION
The Rosemount Analytical CFA3000 Series analyzers are designed for minimum maintenance and maximum reliability. The touch of a button starts the microprocessorcontrolled program, calibrates the analyzer, keeps it calibrated, and reports results directly  unattended  for months at a time. Maintenance is simplified by a modular design with special fittings, labeled tubing, and colorcoded components.
The CFA3000 Series includes analyzers for parameters commonly measured using Ion Selective Electrode (ISE) technology including sodium, fluoride and chloride. Sample concentration is clearly indicated on the digital display panel, and output signals are provided to drive recorders, alarms, or other external devices such as printers or computers. The specific chemistry of your individual analyzer is discussed in Appendix A, Chemistries, or in an accompanying method sheet..
Your purchase from Rosemount Analytical provides you with one of the finest instruments available for your particular application. These instruments have been designed and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowl edge of the user in operating and maintaining the instrument.
2
CFA3000 ISE Analyzer Instruction Manual
1.1 Thoroughly Read This Manual
This instruction manual describes installation, operation, and maintenance of the CFA3000 ISE Analyzer. To ensure continued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, operation, or maintenance of the analyzer.
Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call: 18008548257 or 9497578500 to request the correct manual.
For any inquiries about your analyzer, contact your Rosemount Analytical representative.
Failure to follow the documented instructions may cause any one of the following situations to occur: damage
to this instrument and warranty invalidation.
Follow all warnings, cautions, and instructions supplied with the product.
Only qualified personnel should install, operate, update, program, and maintain the product.
All personnel must learn proper installation, operation, and maintenance of the product.
Install, operate, and maintain equipment exactly as specified in this manual.
Follow appropriate local and national codes.
Only connect the analyzer to electrical sources specified in this manual.
Use only components supplied by Rosemount Analytical or recommended by them for replacement.
Tampering or unauthorized substitution of parts and procedures may affect the performance of the analyzer as well as void the warranty.
All protective covers must be in place and cabinet panels must be closed at all times after installation or
maintenance.
1.2 Analyzer Benefits
Rosemount Analytical listened to you and decided to move away from a peristaltic pump assembly requiring monthly replacement of the pump tubing. The new pump assembly has a vacuum/pressure valve system which eliminates pump tubing and used significantly less reagent. Since the new valve pump swiftly moves fluids through the analyzer, the potential for clogging within any internal component is reduced. Significantly. As a result, mainte nance is required less frequently and procedures such as cleaning flow cells and the mixing system are no longer required.
The upgraded software now identifies a fixed standard that allows one to change the range of the 05 V and the 420 mA output signals to your requirements.
3
CFA3000 ISE Analyzer Instruction Manual
1.3 WARNING – Electrical Shock Hazard
Installation of cable connections and servicing of this product require access to shock hazard voltage levels.
Main power and relay contacts wired to separate power sources must be disconnected before servicing.
Nonmetallic cable strain relief does not provide grounding between conduit connections! Use grounding
type bushings and jumper wires.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA 70) and/or any
other applicable national or local codes.
Operate only with:
1. Main console rear cover fastened.
2. Left side power terminal cover closed.
3. Keypad assembly latched closed.
Safety and performance require that this instrument be connected and properly grounded through a three
wire power source.
Proper configuration and use is the responsibility of the user.
Mark
ID Description
EN 610101 Safety of electrical equipment
EN 55011 Limits of RF equipment
EN 500821:92
Generic immunity from electrostatic discharge, RF & Fast
Transient/Burst Immunity
EtL
9700516
ETL LISTED CONFORMS TO UL STD. 31011
EtL
ETL LISTED CERTIFIED TOCAN/CSA C22.2 No. 1010.192
4
CFA3000 ISE Analyzer Instruction Manual
2.0 INSTALLING THE ANALYZER
2.1 UNPACK PARTS
When you receive your new CFA3000 analyzer, carefully unpack the items in the following list.
Main cabinet
Reagent cabinet
Mounting brackets
Overflow sampler assembly
Waste tube assembly
Accessories pack
Startup reagents
Valve pump assembly (integrated housing includes pump, labeled tubes with straws)
If any of the items are missing or damaged, contact Rosemount Analytical at: our:
Tollfree number: (800) 8548257
Direct number: (949) 7578500
Fax: (949) 8639159
Inspect each piece to confirm that there is no damage or loose parts. If anything is missing or appears broken or damaged, contact Rosemount Analytical at the tollfree phone number above.
SECTION 2
INSTALLING THE ANALYZER
5
CFA3000 ISE Analyzer Instruction Manual
After analyzer has been unpacked and all parts accounted for, install as described in following sections. Refer to Figure 21 to view a completely installed analyzer.
Figure 21. Opened Assembled Cabinet
To mount main cabinet, proceed as follows (see Figure 22):
1. Securely attach supplied "L" brackets to wall using suitable hardware. Brackets must be anchored properly
to support 250 pounds (113.6 kg).
2. Attach main cabinet to "L" brackets using supplied four bolts, split ring washers, and hex nuts.
2.3 Mount Reagent Cabinet
To mount reagent cabinet, proceed as follows:
1. Loosen the four bolts in bottom of main cabinet.
2. Lift reagent cabinet and attach to main cabinet by placing heads of bolts into keyhole slots in top of
reagent cabinet.
3. Slide reagent cabinet back.
4. Feed grounding strap on bottom of main cabinet through notched hole in left rear corner and attach it to
stud on back of reagent cabinet (see Figure 23).
5. Tighten screws to attach reagent cabinet to main cabinet.
6. For multistream units, plug valve bank cable into positive fit receptacle at left side of cabinet (refer to
Figure 71, “Multistream Valve Panel Connector”).
6
CFA3000 ISE Analyzer Instruction Manual
2.2 Mount Main Cabinet
General Recommendations
This cabinet is not for outdoor installation.Environmental conditions of enclosed facility:
• Temperature: min. 50°F to max. 113°F (min 10°C – max. 45°C) ±10° within regulated range.
• Humidity: noncondensing
Mount cabinet on a wall or other suitable vertical surface, at least 42” (107 cm) wide and
431/2” (111 cm) off the floor.
Mounting surface must be able to support 250 pounds (113.6 kg).
Allow 11/2 feet (.5 m) of clearance at left side of analyzer for access to input/output panel,
and allow for swing out for rear access.
Allow 12” (.3 m) at right side for access to overflow tube assembly (if single stream).
Mount cabinet level, or with a slight backward tilt for proper leak detector operation.
7
CFA3000 ISE Analyzer Instruction Manual
Figure 22. Mounting Main and
Reagent Cabinets
Figure 23. Attaching
Grounding Strap
8
CFA3000 ISE Analyzer Instruction Manual
2.4 Install Waste Tube
To install waste tube, proceed as follows (refer to Figure 21):
1. Guide waste tube through main cabinet into reagent cabinet and out through bottom of reagent cabinet.
2. Secure plastic waste tube in clamp in lower right side of main cabinet.
3. Insert Colorimeter drain tube into waste tube.
2.5 Mount Overflow Sampling Assembly
To mount overflow sampling assembly proceed as follows (for sample configurations, see Figures 24a and 24b):
1. Mount overflow sampler assembly on outer, right side of reagent cabinet using the provided four machine screws. For multistream units, mount overflow assembly on left side of cabinet. (If using IF100 InLine Filter, refer to instruction sheet accompanying filter for complete installation instructions.)
2. Place ends of air vent tubing and overflow waste tubing into waste pipe.
3. Connect waste pipe to drain.
4. Supply sample stream to regulating or intake valve with 1/4” tubing (plastic or steel) as follows: Insert tubing into tube fitting. Make sure tubing rests firmly on shoulder of fitting and nut is fingertight. Mark nut at 6 o’clock position. While holding fitting body steady with a wrench, tighten nut 11/4 turns (go to 9 o’clock position).
5. For multistream assembly, connect sample intake line to bulkhead fitting.
6. For singlestream assembly, attach sample intake line to valve pump.
7. Make sure that sample liquid overflows into overflow chamber.
The Sample intake tube should extend into
The overflow tube approximately 1½” – 2”
3.81 –5.08 cm).
9
CFA3000 ISE Analyzer Instruction Manual
Figure 24b. Multi
stream Overflow Sampling Assembly, threestream option)
Figure 24a. Singlestream Overflow Sampling Assembly
10
CFA3000 ISE Analyzer Instruction Manual
Electrical connections are made to and from the cabinet at the input/output panel, located on the left side of the analyzer. See Figure 25.
• Terminals are provided for ground, power supply, high/low alarms, output signals, calibration and status relays
• Output signals  a voltage output of 05 Vdc and a floating, ungrounded current output of 420 mA are available for remote readout or control of a process.
• Also available are contact closures for the following:
High alarm Calibration Low Alarm Leak Alarm
To make electrical connections, proceed as follows:
1. Remove solid caps from the holes above the input/output panel by puncturing with an awl and prying them out. (Note: remove caps only from those holes that will be used.)
2. Replace with permanent connectors.
3. Connect alarms and external computer interfaces (if any).
4. Feed 3wire power cable into analyzer through hole above the input/output panel.
5. Connect wires to appropriate contacts on Terminal Strip TB3.(See next page)
2.6 Prepare Electrical Connections
Use cable glands to hold cables securely in holes above input/output panel.
WARNING:
Connect output signals, if any before connecting the power supply.
CAUTION:
For proper operation and for safety, electrical connections to this equipment must be
made in accordance with local or national electrical code as applicable.
This equipment must be grounded. A qualified electrician should wire this equipment to an
electrical circuit.
11
CFA3000 ISE Analyzer Instruction Manual
Figure 25.
Input/Output Panel
2.7 Install Valve Pump Assembly
The pump assembly has a vacuum pressure system that:
l Eliminates internal pump tubing l Uses significantly less reagent l Accurately moves fluids through the analyzer.
Perform the following procedure to install valve pump assembly:
1. Tilt pump assembly towards you.
2. Align the hinge pins and sockets, and slide assembly tray as far as possible to right (see Figure 27). Carefully lay tray down into cabinet
3. Plug in the two connectors to the back of the cabinet (6position connector for heating bath; 9position connector for pump) (see Figure 28).
4. Connect 16pin connector from pump assembly to small connector on CPU board (yellow ejector) above connector for keypad panel.
5. Attach colorimeter inlet tubing on left side of pump to flowcell inlet (see Figure 21).
6. Continue with procedure depending on whether system is singlestream or multistream.
For singlestream, continue:
6. Thread tube labeled Sample" through hole 0n bottom of shelf to the right of pump assembly, and out through hole on left side of cabinet.
7. Connect Sample tube to top of overflow sample on left side of cabinet.
8. Guide all other tubes that are attached to valve pump assembly through reagent cabinet access hole.
For multistream, continue:
6. Thread Sample tube from pump assembly through reagent access hole on the right side of the valve pump assembly, and connect to Common (C) on #1 multistream solenoid valve on left side of reagent
cabinet.
Guide all other tubes that are attached to valve pump assembly through reagent cabinet access hole.
12
CFA3000 ISE Analyzer Instruction Manual
Figure 26. Valve Pump Assembly
Figure 27. Pump Connector
13
CFA3000 ISE Analyzer Instruction Manual
2.8 Prepare Reagent Containers
Prepare reagent containers as follows:
1. Remove screw caps from reagent bottles.
2. Using an awl, make one hole in foil for the reagent straw. Do not allow awl to touch reagent or standards.
3. Place Standard and Reagent bottles into reagent cabinet, as shown in Figure 28.
4. Insert colorcoded reagent straws and labeled Standard straw leading from valve pump assembly into holes punched in corresponding bottles.
Figure 28. Sample Bottle Placement in Cabinet (for all methods; colors refer to reagent straw colors)
14
CFA3000 ISE Analyzer Instruction Manual
3.0 Operating the Analyzer
3.1 Operational Overview
The analyzer uses a batch flow method for automatically analyzing a sample against a fixed standard (refer to block diagram in Figure 31). That is, the touch of a button starts the microprocessorcontrolled program, calibrates the analyzer, takes measurements of samples, and reports results directly  unattendedfor months at a time.
The automated monitoring program determines the sequence for activating each of the valve pumps in the valve pump assembly to inject air and fluids into the reaction chamber. Each valve pump is associated with a specific input via the quick disconnect connector on the side of the assembly (the number of reagents is determined by your particular chemistry and procedure).For example, the inputs for sodium chemistry would be:
• Buffer or Ionic Strength Adjustor, reagent #1
• Sample
• Grab sample
• Baseline
• Standard
• Air
• Evacuation of chamber
• Recirculate
For a multistream system, solenoid valves in the analyzer reagent cabinet open and close streams for sampling. When sampling, the keypad panel LEDs identify which sample stream (valve number) is associated with the dis played value, and which valve is now open for sampling. Although the system runs samples on its own, you may also perform a manual grab sampling or a manual calibration at any time.
Figure 31. Simple Block Diagram
Valve
Pump
Electronics
Data output
ISE
(ion selective
electrode)
Reaction
Cell
SECTION 3
OPERATING THE ANALYZER
15
CFA3000 ISE Analyzer Instruction Manual
3.2 Check Sample Stream
• Check that the sample intake tube is 1½ to 2 inch (3.81 to 5.08 cm) into the liquid.
• Sample should just overflow the overflow tube. There should be no particulate or bubbles in the tube.
Using a MultiStream Overflow Sampler
Multistream analysis requires an overflow sampler assembly that will accommodate the extra sample streams. One multistream chamber allows for up to three sample streams; for 4  6 streams, use two chambers.
The sample tubes connect to solenoid valves in the reagent cabinet. The analyzer uses the valves to cycle through all sample lines, in sequence. First, stream 1 valve is opened, and Valve 1 LED on the keypad panel is lit. When the measurement is complete, Valve 1 will be closed, and Valve 2 is opened to begin sampling stream
2. The keypad panel will show Valve 1 LED lit as the Displayed Stream, and Valve 2 LED is lit for the Selected
Stream.
LED's on the valve pump assembly light up when a particular fluid is taken up by the valve pump assembly. Figure 32 identifies the valve pump LED designations.
3.2.1 Check Fluidics
Check the following:
• Liquid should overflow in overflow chamber.
• Waste tubing is in waste pipe.
• Sample tube is connected to Overflow Sampler Assembly for single stream; and, for multistream, it is connected to Common (C) port of sample solenoid valve #1, on left side of reagent cabinet.
• System has no leaks.
3.2.2 Check Reagents
Check the following:
• One hole has been punched in all reagent cap liners.
• Reagent straws are placed in appropriate reagent containers.
• High and low standard tubing are in appropriate containers.
3.2.3 Turn On Power
1. Open the keypad panel door to expose the circuit boards.
2. Press each board in to ensure proper seating.
3. Simultaneously press Power button on the back wall of upper cabinet and Reset button on the CPU board (yellow ejector), then release.
4. First release the Power button, then the Reset button. The Keypad should display ”H E L O", then four dashes "   ".
5. Pull pump forward and check leak detector by dipping finger in tap water and touching detector behind pump. Leak detector LED on keypad panel will light up.
6. Turn main power switch to OFF.
7. Dry leak detector
8. Wait 20 seconds, then switch main power switch back to ON. Leak detector LED should be off.
16
CFA3000 ISE Analyzer Instruction Manual
3.2.4 Test Valve Pump Operation
Reaction cell takes approx. 30 minutes to heat up.
If leak detector LED does not light up:
Adjust potentiometer on analog board counterclockwise for more sensitivity.
Turn off power to reset system.
Wait 20 seconds, and simultaneously press Power and Reset buttons.
Dip finger in tap water and touch leak detector contacts on bottom of cabinet behind the valve
pump. LED should light up.
If problem persists, contact Rosemount Analytical.
The test procedure allows you to confirm that all physical connections are installed properly, and to ensure that the pump is taking up the proper fluids at the correct time.
Enter Test Mode:
1. Put all straws and Sampling tube into a beaker of DI water.
2. Turn power on by simultaneously pressing Power button on the back of upper cabinet and Reset button on CPUboard (yellow ejector).
3. First release Power button, then Reset button. Keypad should display "H E L O",
then four dashes "   ".
To enter test mode, continue with the step appropriate for your analyzer setup:
For singlestream, continue:
For multistream, continue:
Referring to Figure 32, on keypad panel, press posi tion 15 and then 2
On keypad panel, press: then
.
DECIMAL
MANUAL
17
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
FIGURE 32. KEYPAD PANEL
The keypad buttons may now be used to test pump operation. Do not rush through
the test steps – when a specific step is initiated, it takes approx. 2 seconds to pump air
through the system. Actual pumping of a liquid takes approximately 4 seconds. After this operation
is finished, another 4 seconds elapses before the pump is primed for the next test step.
To exit the test at any time, press the ENTER button on the keypad panel.
If a test step does not work properly, try the following:
Take a syringe and shoot air through straw to make sure it is not clogged.
Check that corresponding LED on valve pumps lights up (refer to Figure 41).
Check that fluid is picked up in correct tube.
If the problem still persists, contact Rosemount Analytical.
Begin Test:
Confirm that the corresponding Pump LED lights up for each step (refer to Figure 33 for LED designations during test):
18
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
2. Press to simulate one shot of Reagent #4. Look at blue straw to confirm that water is pumped up through tube. Repeat a few times until you see water is in clear section of tube
3. Press to simulate one shot of Reagent #3. Check green straw and its clear tube
4. Press to simulate one shot of Reagent #2. Check orange straw and its clear tube.
5. Press to simulate one shot of Reagent #1. Check red straw and its clear tube
6. Press to simulate one shot of Sample coming from overflow sampler. Check tube labeled "Sample."
7. Confirm that fluid is going to flowcell.
8. Press to simulate one shot of Standard. Check corresponding clear straw and its clear tube.
9. Press to pump one shot of air into analyzer.
10. Press to simulate one shot of Baseline solution. Check clear straw labeled, BASELINE, and tube.
11. Repeat Step 9 until you see fluid shoot into reaction chamber.
12. Press to simulate evacuation from the reaction chamber to the flowcell. Check the tube going up to flowcell.
13. Press to simulate one shot of Grab solution. Check clear straw and tube.
14. Press to wash flowcell.
1. Press ZERO to test air. Press ZERO to turn off.
19
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
15. Press to empty reaction chamber. If there is any fluid left, it will be shot through to Waste.
16. Press once to exit test.
NOTE: The LCD on the display board will display the keypad position which is assigned to a specific
valvepump. When the LCD number disappears, the next position may be pressed. The sequence is
given on page 24.
Figure 33. Valve Pump LED Designations
20
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
3.2.5 Set Up Stream Parameters
The analyzer is programmed to measure a singlestream system at 4 sample cycles per hour. The multistream default is 1 cycle per stream, but you may define parameters for individual streams, such as modifying the num ber of cycles or the alarm threshold values.
The alarm values are used to activate alarms or controllers when the sample is outside of the specified concen tration range. A high alarm will activate when the sample concentration exceeds a preset limit. A low alarm will activate when the sample concentration is below the preset limit.
The contacts for the alarm outputs are located at the input/output panel on the cabinet side. The system must be in an alarm condition for at least one minute to activate the alarm contact.
Use the logical flows below to define desired stream parameters.
Figure 34. Logic Flow for Singlestream Parameter Definition
21
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
Figure 35. Logic Flow for Multistream Parameter Definition
(Note: Use decimal on the keypad using arrows to place decimal in the correct position)
22
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
3.3 Set Reagent Supply Time
Use the keypad panel to remind you when you need to replace the reagent. The analyzer will count pump cycles and count down the number of days left. To specify the duration, proceed as follows:
1. Press once to set counter for 90 days (3 months).
Sample Display: 0 0 9 0
2. Press to exit and clear display.
To determine how many days are left until you have to replace reagent,
Press: once.
Look at the LCD and record the number of days;
then press: again to get out of mode.
3.4 Set Scaling Outputs
Set output 420 mA 05 Vdc from the keypad panel, (refer to Figures 72 and 73), then adjust 420 mA and 05 Vdc at your external device.
Set output scale to: 4 mA/0 V
1. Press: .once then .to display 4 mA and for 0 Vdc.
The sample Display for 4 mA and for 0 Vdc: 0 0 0 4
2. To exit, press
Set output scale to: 20 mA/5 V
1. Press once then to display 20 mA and 5 Vdc.
23
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
2. The sample display for 20 mA and for 5 Vdc: 0020
2. To exit, press .
To display the fullscale concentration embedded in the EProm:
Press two times. To exit, press .
3.4.1 Record
mV
Use the mV value to determine that the method’s span and slope are correct. The change in millivolts (mV) is the spread between a baseline and a fullscale standard at the time of calibration. The values are relative to the chemical parameter being measured
Record the millivolt values for both baseline and fullscale standard.
1. Press
The sample Display for value of 150 mV is: 0 1 5 0
2. To exit function, press once.
3.4.2 Initiate Auto Calibration
The analyzer is calibrated automatically usually every two days, depending on the chemistry. Auto calibration performs both baseline and fullscale calibration. Baseline calibration measures a solution with 10% concentra tion of the species of interest. Fullscale calibration measures a standard solution in the upper range of the parameter of interest.
Before starting calibration, make sure that the analyzer is running reaqgents and baseline
solution for at least 20 minutes.
24
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
To initiate auto calibration, press on the keypad panel – the CALIBRATION LED will light. When multi stream systems are being calibrated, only Stream 1 LED is lit.
Baseline calibration will run first for approximately 15 minutes. Fullscale calibration follows, lasting approximately 15 minutes (the actual time for calibration depends on the chemistry). When fullscale calibration is finished, CALIBRATION LED goes out and the value is displayed on the LCD. The “   ” display signifies that a delay is required before the next sample is displayed.
Sample analysis automatically begins after calibration. The time delay from the start of sample analysis to dis play of results is approximately 15 minutes. This allows sufficient time for the previous solution to wash through system, for sample and reagents to react, for analyzer to measure sample and display concentration value.
When the CALIBRATION LED and the ”000” display in a singlestream system go out, the displayed value on the keypad panel is the actual stream value.
For a multistream system, after auto calibration, Sample Stream status LED’s will light to show sample stream valve currently selected and sample stream concentration currently displayed. For example, for the first stream sample, VALVE SELECTED LED is lit for that stream; dashes display on the keypad panel. When analyzer finish es sampling one stream and moves on to the next, the display shows the previous sample stream concentration value, and its VALVE DISPLAYED LED lights up. VALVE SELECTED LED will be lit for next valve in sequence.
Note: If for any reason a calibration is not wanted and needs to be aborted, press:
then, , then to abort the calibration. Press twice in rapid succession on initial
startup for priming of sample & reagents
3.5 Process a Grab Sample
The analyzer allows you to rapidly analyze a grab sample.
Press on the keypad panel, to introduce a grab sample into the analyzer from the holder on the overflow sample panel.
If visible, suspended solids are present in the sample, filter it to 8 microns or less.
25
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
Place the clear grab sample inlet tube into a beaker containing the grab sample.
2. Press one time. The display shows flashing dashes, “    “ .
The system analyzes the grab sample and after 15 minutes, displays the results when complete (The results
are “flashed” on the display).
Grab Sample Display: 10.0 (flashing for three minutes)
After the display of the grab sample’s value, a multistream analyzer begins processing the next stream in sequence.
3. To display measurement for previous grab sample, press twice.
4. Press button to return to normal operation.
3.6 Manually Sample a Particular Stream.
Use this procedure to manually switch between sample streams in a multistream system. This operation can be performed at any time, except during calibration.
Press one time. Sample Display: 0 0 0 1 .
Press and to change stream number. Sample Display: 0 0 0 3
Press to start sampling desired stream.
Press to return to multistream automatic analysis.
3.7 Shutdown
To shut down the CFA3000 for only a few days, press main Power button in back of upper cabinet to OFF position. Leave reagent straws in reagent containers.
26
CFA3000 MANUAL SECTION 3
INSTRUMENT DESCRIPTION
To shut down the analyzer for more than a day, perform the following procedure
1. Remove straws from reagent containers and place them into a beaker of DI water.
2. Press twice within a twosecond interval to activate priming. This allows wash solution or DI water to be injected rapidly through system.
3. After priming finishes (after 15 minutes or when “auto cal” starts), press the main Power button in back of upper cabinet to OFF position.
3.8 Quick Start
If the analyzer has been shut down during normal operations, turn power on and
press once.
If the CFA3000 has been turned off for an extended shutdown:
Press twice to prime; the system will automatically go into :”Auto Cal” after priming
Some chemistries may require specific wash solutions. Check Appendix A,
"Chemistries," for details.
27
CFA3000 MANUAL SECTION 4
MAINTAINING THE ANALYZER
Your CFA3000 Series analyzer is a precision instrument which, if maintained properly, should provide years of accurate and reliable service.
4.1 Analyzer Equipment.
Even though the analyzer normally runs automatically, perform the following activities, quarterly:
1. Start Auto Calibration after reagent replenishment (see Section, "Perform Auto Calibration").
2. Check mV value
SECTION 4
MAINTAINING THE ANALYZER
28
CFA3000 MANUAL SECTION 5
TROUBLESHOOTING PROCEDURES
This chapter is divided into two sections:
• Troubleshooting Chart: a list of symptoms, probable causes, and remedies.
• Test Functions:– test functions for diagnostic information.
5.1 Troubleshooting Chart
Symptoms observed during operation are generally enough to identify a faulty component.Such symptoms, together with probable causesand remedies are listed in the following troubleshooting chart.
Problem Probable Cause Remedy
No power
Power switch
OFF
Press power
switch ON
Power switch
failed
Call Rosemount
Analytical
Pump is
inoperative
Valve pump
switch OFF
Press switch ON
Blown fuse Replace fuse
Leak detector
activated
Find and fix
leaking; check
pump’s electrical
connection
Bad electrical
connection
Check pump’s
plugs to CPU &
connector
Still inoperative
Call Rosemount
Analytical
Keypad lights or
display fails to
light up
or buttons fail to
operate
Circuit boards
are not seated
properly
Remove and
reinstall circuit
boards
Display or
associated
power circuit
faulty
Call Rosemount
Analytical
Problem Probable Cause Remedy
Keypad lights or
display fails to
light up
or buttons fail to
operate
Display or associated
power circuit
faulty
Call Rosemount
Analytical
Connector faulty
or not properly
inserted
Check
connection
SECTION 5
TROUBLESHOOTING PROCEDURES
Table 51: Troubleshooting Chart
29
CFA3000 MANUAL SECTION 5
TROUBLESHOOTING PROCEDURES
Problem Probable Cause Remedy
Alarms not
activated
Incorrect
settings
Readjust
settings, making
sure that low
alarm is set
lower than high alarm
Insufficient time
for alarm to
respond
Alarm condition
must exist for
minimum of 60
seconds before
alarm activates
Signal not
repetitive
(same sample
not repetitive)
Small leak or
clogging in
fluidics
Check all tubing
lines and
connections
Clogged tubing Replace tubing
No signal, or signal “offscale” high
Connector or
wires loose
Faulty circuit
board
Incorrect mV
value
Check circuit
boards for
proper seating
Replace power
supply circuit
board
Loss of mV
Sensitivity
LED ISE on
Reagents,
standards old,
contaminated or
improperly
prepared
Change
reagents
Reagent straws
in wrong
reagent or
standard
Check
placement of
straws
Problem Probable Cause Remedy
Loss of sample
flow
Replace sample
filter on overflow
panel
Decreased pumping volume
Tubes pinched
shut or clogged
from deposits
Remove
condition
causing “pinched” lines; replace sample
filter at overflow
panel
Leak detector LED lights up
Leak in fluidics Repair leak and
dry leak detector
contacts; turn
main power
OFF for
20 seconds
High humidity Adjust screw on
leak detector
potentiometer
on analog
board; turn
clockwise to
decrease
sensitivity
Connector faulty
or not properly
insertede
Check
connection to
analog board;
replace analog
board.
5.2 Test Functions
Table 52: Test Functions
Problem Probable Cause Remedy
Front panel
display
Test LED’s and
LCD on keypad:
a. Press Power
button and reset
button on the
CPU board (yel
low ejector) b. Release power, then
Reset button.
Press power
Press:
30
CFA3000 MANUAL SECTION 5
TROUBLESHOOTING PROCEDURES
Check
button.
Test Procedure
Exit
Front panel
display
c. Press the following keys in sequence: (key #’s 15, 5, 6)
Press:
Sensitivity
(mV)
Press:
then
Press:
then
31
SECTION 6
REPAIR PROCEDURES
CFA3000 MANUAL SECTION 6
REPAIR PROCEDURES
The Ionselective electrode analyzers are designed with modular components for quick and easy replacement. The following sections contain instructions to remove parts and replace them. For more complex repairs, or diag nostic assistance, contact Rosemount Analytical at:
• Telephone: (800) 8548257, (949) 7578500
• Fax: (949) 4747250
• Email: www.raihome.com
6.1 Replace Valve Pump Assembly
Use the following procedure to replace pump assembly:
1. Turn off power to analyzer.
2. Pull back of pump assembly towards you to a 45degree angle (see Figure 61).
3. Unplug the two connectors from the back of the cabinet. (9position connector for pump).
4. Disconnect pump assembly’s “D” connector from right side of pump.
5. Remove all liquid tubes from pump.
6. Disconnect waste tube from bottom of reaction cell (right side of valve pump).
7. Slide assembly to the left to disengage from pins, and remove pump assembly.
8. To install a new pump assembly, refer to Section, " Install Valve Pump Assembly".
Figure 61. Remove Valve Pump Assembly
32
6.2 Replace Circuit Boards
Use the following procedure to replace circuit boards:
1. Turn off power.
2. Open latch from right side of keypad panel; and open door.
3. Disconnect all cables and plugs from board to be replaced.
4. Unlatch colored clip from board to be replaced (refer to Figure 62).
5. Pull circuit board straight out.
6. Reverse steps to replace board and reconnect cables. Press firmly to ensure that boards are properly seated.
7. Turn power on.
8. Perform Automatic Calibration (refer to Section, “Initiate Auto Calibration”).
Please NOTE: Valvepump connector is on CPU board
Figure 62. Card Cage
CFA3000 MANUAL SECTION 6
REPAIR PROCEDURES
IMPORTANT: Do Not remove any PC board or component without first shutting off Power to the analyzer.
33
6.3 Replace Fuses
Use the following procedure to replace fuses (all fuses are 1A):
1. Turn off power.
2. From inside upper left corner near pump switch, press down on fuse cap and turn counterclock wise to remove.
3. Pull fuse and cap straight out.
4. Replace fuse in cap (make sure the new fuse is the same type and rating as the one you are replacing).
5. Press fuse cap in place and turn clockwise.
6. Turn power on.
6.4 ISE Reaction Cell Assembly, Electrodes
Figure 63. ISE Reaction Cell Assembly, Electrodes
CFA3000 MANUAL SECTION 6
REPAIR PROCEDURES
6.4.1 The sodium electrode used in the CFA3011 sodium analyzer is a proprietary “solid state” electrode
which is joined with a reference electrode in a sealed cartridge. The buffered sample stream flows into the reaction cell—through the ground stud to the sodium electrode, then into the reference electrode, out through the ground stud and then to waste. The sample concentration is measured in the sodium electrode, compared to the reference value and the sodium concentration is determined.
34
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
SECTION 7
INSTRUMENT DESCRIPTION
Part Function
Main Cabinet Splashproof enclosure
protects analyzer
Reagent Cabinet Splashproof enclosure
protects reagents and solenoid valves
Door Hinges Pin hinges mount doors
on main and reagent cabinets
Door Latch Secures cabinet door
Keypad Panel and Driver Board
External side contains touch buttons for controlling analyzer, digital display, and status panels. Internal side contains driver board
HB Lamp Indicates when propor
tional controller is heating the heat bath
HB T 1 A 1 A fuse for heating bath
LAMP PS T 1.0 A 1 A fuse for power supply
to ise lamp
PUMP T 1.0 A 1 A fuse for pump
XFMR T 1.0 A 1 A fuse for transformer
Pump On/Off Lighted switch controls
power to pump
On/Off Main power switch
Colorimeter Detector
Waste Drain Pan Drain pan directs waste
solutions to waste tube
Pump Connector 9position connector
connects pump to main cabinet electronics
Heating Bath Connector 6position connector
connects to heating bath assembly
Valve Pump Assembly Includes pump, PC
board, labeled tubes, and reaction vessel assembly (chemistry module with integrated
Part Function
Chemistry Module Tubing Carries sample and
reagent to valve pump assembly
Quickdisconnects Connects reagents,
samples, and standards to valve pump assembly
Waste Tube Carries waste solutions
to drain
Leak Detector Senses leakage of
solutions in main cabinet, shuts pump off, and activates leak LED
Part Function
MultiStream Valve Panel (MultiStream option only)
Allows sequential analysis of up to six sample streams
MultiStream Connector (MultiStream option only)
Positive fit connector connects Multistream valve panel to main cabi net electronics
Bulkhead Fitting Fitting that directs
individual sample stream to sample valve
Overflow Sample Assembly
Supplies continuous sample streams for analyzer
7.1 Analyzer Front View (Refer to Fig. 71).
Table 71.1: Main Cabinet Parts
7.1.1 Main Cabinet
7.1.2 Reagent Cabinet
Table 72. Reagent Cabinet Parts
(heating bath)
35
Note: For multistream monitor, Overflow Sampling assembly mounts on left side, facing the analyzer
Figure 71. Analyzer Front View
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
36
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
Button/LED Function
For Both Single and Multistream Options
LEAK LED When lit, indicates leak in
main cabinet
CALIBRATION LED When lit, indicates
calibration in progress
ISE/LAMP LED Will light when millivolt
level is below default levels.e
REAGENTS LED When lit, indicates
reagents need to be replaced
UP or DOWN Arrow
Increases or decreases displayed value
Sets value or exits function
Sets high alarm limit
Sets low alarm limit
Two functions:  Sets 90 day clock for reagent usage  Displays number of days remaining until next replacement of reagents
Sets scale to adjust out put to 4 mA and 0 Vdc on recording device
Sets scale to adjust out put to 20 mA and 5 Vdc on recording device
Button/LED Function
Function: Displays Systems  mV
Initiates processing from grab sample
Initiates automatic calibration of analyzer
Used for setting alarm values.
For Multistream option only
Sample Streams LED’s
VALVE SELECTED shows which valve (stream) is being analyzed, and VALVE DISPLAYED shows which valve is associated with completed and displayed value
Initiates automatic multistream analysis
Initiates single stream analysis
Enters function to identify parameters for each stream (such as number of cycles and high and low alarms)
(Not used)
7.2 Keypad Panel
The keypad panel has a fourdigit liquid crystal display with an adjustable decimal point. The panel also has indi cator LED’s and pressuresensitive buttons.
Table 73 Buttons/LED Functions
37
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
Part Function
Keypad Panel Door Opens to access card
cage
Display Board Circuit board contains
electronics for keypad and LCD display
CPU Board 180B02801
Central processing circuit board processes the signal fed to the digital display and any external devices. (Yellow ejector)
Valve Board 180B01601 (only supplied for multistream option or for RS232 output on single stream)
Circuit board controls multistream accessory. Jumper should be installed on back of valve board. (Purple ejector)
Analog Board 180B02502 For multistream configurations: 180B02501*
Circuit board converts signals from analog to digital and processes signals from "fault" detec tors (Orange ejector)
Power Supply Board 180B01203
Supplies voltage to electrical circuitry (Blue ejector)
7.3 Card Cage
The card cage is located directly behind the keypad panel. To access the card cage, lift latches from right side of keypad panel and open hinged door. (Refer to Figure 72.)
7.3.1 Card Cage and Door
Table 74: Card Cage Parts
Component Function
0Vdc  5Vdc Connector To recording device.
4mA –20mA Connector To recording device.
Computer Memory Reset Clears microprocessor
memory
Computer Memory Reset Clears microprocessor
memory
Valve Pump Connector Connector for cable from
valve pump
Display Board Cable Connector
Connector for display board cable
Component Function
Potentiometer Adjusts sensitivity of leak
detector – clockwise for less sensitivity; counterclockwise for greater sensitivity
Analog Board DIP Switch Contains 8 rocker switches
for setting various functional processes (see Figure 73)
Signal Input Connector For process signals
On 180B02501: 20 multistream output potentiometers
For each stream 2 – 6, potentiometers for 4 mA, 20 mA, 0 V, 5 V (multistream only)
Component Function
Multistream Panel Connector
Connector for cable connector from display board
7.3.2 CPU Board
Table 75: CPU Board Components
7.3.3 Analog Board
Table 76: Analog Board Components
7.3.4 Valve Board
Table 77: Valve Board Components
Part Function
18pin Female Connector For relays
6pin Male Connector Operating power
6pin Female Connector Valve Pump Power
7.3.5 Power Supply Board
Table 78 Power Supply Board
38
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
PLEASE NOTE: VALVEPUMP CONNECTOR IS ON CPU BOARD
FIGURE 72. CARD CAGE AND BOARDS
IMPORTANT: DO NOT remove any PC board or component without first shutting off Power to the analyzer.
39
Figure 73. Analog Board DIP Switches
(with #3 set in OFF position for multistream analyzer) Colorimeter Assembly
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
7.3.6. Analog Board DIP Switches
The following list describes the conditions for each switch to be in the ON or OFF position:
Table 79. Analog Board DIP Switch Positions
Switch No. Position Condition Remarks
1 ON
When using Analog Board, Part # 180B02602
Analog Board has no adjustment ports for individual output device and/or 05 Vdc, 420 mA outputs.
1 OFF
When using Analog Board, Part # 180B02601
Analog Board has individual ports for adjusting each output device (up to 6) and/or 05 Vdc, 420 mA outputs.
2 ON Quick calibration  on
2 OFF Quick calibration  off
3 ON
4 ON Nonfunctioning switch
5 ON Nonfunctioning switch
6 ON Nonfunctioning switch
7 ON
For stream ID and “brackets” before and after calibration
Allows markers to be recorded on strip charts and other recording devices.
7 OFF
No markers will be recorded on strip charts and other recording devices.
8 ON For 80 column printers Formats printer output
8 OFF For 40 column printers Formats printer output
40
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
Component Function
Backplane Board Circuit boards plug into
front side
Power Transformer Supplies ac voltage to
electronics
Heating Bath Connector Interconnect to heating
bath
Pump Connector Interconnect to valve
pump
220Vdc Transformer Optional part for specific
electrical requirements (208 V – 240 V)
Table 710: Back Cabinet Parts
FIGURE 74. REAR OF MAIN CABINET
7.4 Rear of Main Cabinet
In the rear of the main cabinet are a number of electrical components and connectors. The following diagram is for reference, only.
41
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
Part Function
Pump Housing Contains and protects
valve pump electronics
Pump Valve pump routes
sample and reagents through the analyzer
Reagent Barb Fittings Male half of tubing
connector. Tubes are connected to reagents and sample
Reaction Vessel
Assembly
Contains reaction cell, heating block
6position Connector Electrical power to
heating bath
9position Connector Electrical power to pump
LED Green is OK; RED is
Low pressure
Part Function
Sample and Reagent Tubing
Colorcoded straws and labeled tubes for transport of solutions
“D” Connector Connects to small, 16pin
connector on CPU boards
Vent Line Vents reaction cell to
waste
Pump Drip Tray Catches leaks and
directs them to the leak detector
LEDS Identify which injector is
active.
7.5 Valve Pump Assembly
The vacuum pump assembly includes the pump, PC board, and reaction cell.
Table 7 11: Valve Pump Parts
* Vacuum/Pressure LED is located on the upper right corner of the ValvePump as it is installed
Figure 75. Valve Pump Assembly
42
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
Port Function
RS232C For connection of
computer interface or other accessories
Relay Output Optional relay contact
(multistream analyzer only)
Printer For connection of printer
cable
Recorder For connection of
external recorder; for multioutput, 420 mA and 05 Vdc outputs for streams 2 through 6.
Pin No. Description
1 NO 9 C Stream #1 Alarm or ID
2 NO 10 C Stream #2 Alarm or ID
3 NO 11 C Stream #3 Alarm or ID
4 NO 12 C Stream #4 Alarm or (#1*) ID
5 NO 13 C Stream #5 Alarm or (#2*) ID
6 NO 14 C Stream #6 Alarm or (#3*) ID
7.6 Interface Ports
The interface ports (labeled for recorder, printer, relay output, and RS232) are located on the lower, left outside wall.
Table 712: Interface Ports
If there are 3 streams or less, both stream ID and alarms may be optionally configured.
7.8 MultiOutput Board
The output for Stream #1 is obtained from the Input/Output Panel (see Figure 25). The additional five outputs are obtained from the lower lefthand side of the analyzer. Table 714 gives the pin assignments on the “D” connector for a multistream system:
Figure 76. Interface Ports
7.7 Relay Output Option
This port (see Figure 77) is a collection of contact closures that can be used to identify either the stream being displayed or the alarm that is activated (multistream systems only). these are not “powered” contacts.
Relay Contact Ratings:
100 Vdc  max voltage 10 Vdc max DC current
1.0 A  max carry current
0.5 A  max switched current
Table 713: Relay Output Option
Pin Assignments
\ 1 2 3 4 5 6 7 8 9 10 11 12 13 14 /
\
15 16 17 18 19 20 21 22 23 24 25 /
Figure 77. "D" Connector for Relay Output
Table 714: Multioutput Pin Assignments on
“D” Connector
2 15 2 14 1
3 17 4 16 3
4 19 6 18 5
5 21 8 20 7
6 23 10 22 9
43
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
Part Function
Intake/Regulating Valve Regulates flow of
sample into overflow sampler assembly
Overflow Tube Assures adequate supply
of sample and reproducible results by eliminating positive pressure from sample stream.
Sample Intake Tubing Tubing which carries
sample from overflow tube to analyzer (one to six stream configurations) NOTE: Sample intake tube should be no more than 1 to 11/2" into the sample.
Air Vent Allows air to escape
overflow tube as it fills with sample.
Part Function
Overflow Waste Vent Allows excess sample
to escape.
Waste Pipe Carries waste to drain.
Overflow Tubing All overflow tubes feed
into common waste pipe.
Filters
(180134108)
Replaceable filters are mounted down stream from each intake regulating valve.
Sample Inlet Tubing 1/8" x 1/4" tubing
(provided by customer).
Common Panel Mounts multistream
Overflow Sampler Assembly to left side of analyzer (facing front panel). Panel with dimensions 221/4” x 9" will support up to six streams.
7.9: Overflow Sampling Assemblies
Table 715: Overflow Sampling Assembly Parts
Figure 78. SingleStream Overflow Sampling Assembly
44
Figure 79. Multistream Overflow Sampling Assembly
(threestream option)
CFA3000 MANUAL SECTION 7
INSTRUMENT DESCRIPTION
45
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
SECTION 8
ADDITIONAL FEATURES
8.1 Chart Recorders and Data Loggers
A chart recorder or other data acquisition device may be connected to the analyzer for a permanent record of sam ple concentration. Setting the output 420 mA 05 Vdc from the keypad panel allows you to adjust 420 mA and 0 5 Vdc to a recording device.
These input devices must be isolated or have a floating ground (not reference to ground). The contacts for recorder output are located at the input/output panel on the side of the analyzer (see Figure 81).
The analyzer will operate normally with data output to the chart recorder or other device. During normal operation, calibration is performed biweekly. This may vary by Chemistry. To indicate the beginning and end of a calibration sequence, the recorder makes a quick trace down to zero, and up to full scale. This is called a calibration marker. After calibration, the recorder resumes recording the sample values (refer to Figure 82).
When a recorder is connected to a multistream system, a ”stream identification blip” (SIB) is produced prior to the output signal for that particular stream. This allows quick identification of individual stream readings on the recorder. SIB values are shown in Table 81, and tracings are illustrated in Figure 83.
Figure 81. Input/Output Panel
Table 81: Stream Identification Blips
(SIB’s)
Stream #
Voltage
Output
% Scale of
Recording
Marking
1 0.5 10
2 1.0 20
3 1.5 30
4 2.0 40
5 2.5 50
6 3.0 60
Grab 3.5 70
46
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
Figure 82. Typical Trace Recording with
Calibration of Single Stream System
47
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
Figure 83. Typical Stream Trace Recording with
Calibration of Multistream System
8.2.2 Data Configuration
The computer must be set to match the fixed data configuration of the analyzer, which is:
8.2.3 RS232C Function Commands
The RS232C interface can be used as follows:
• By direct hookup to a nonMacIntosh® based computer A program on disc can be obtained from
Rosemount Analytical for remote control. Cable length must less than 150 feet.
• By direct hookup to a nonintelligent terminal. No
program disc is required. Cable length must be less than 150 feet.
• By communication through a modem. One modem
located at the analyzer can communicate over a
telephone network to another modem hooked up to an IBM PC or compatible computer or a nonintelli gent terminal.
Pin configuration is described in Table 84.
Table 84: RS232C Pin Configuration
8.2 Computer Interface
The computer interface is an RS232C serial data interchange port operating at 1200 baud. The serial port receives the same information as the printer, and has the capability to transmit information, enabling remote control of the analyzer.
8.2.1 Data Output Format
The data output formats contain no control characters, with the exception of <CR> carriage return, <LF> line feed, and the preamble/post amble described below:
Table 82: Normal Data Output Format
48
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
Data Format
Preamble NUL, STX, NUL =
(00, 02, 00 Hex notation)
Date/Time ASCII string of 25
characters, followed by 7 spaces
Concentration ASCII string of 5
characters, followed by 9 spaces
Stream Number One ASCII character
Mode An ASCII string,
consisting of 7 spaces, followed by three possible messages: NOT calibrated, Manual, or Automatic.
Calibration Status There are two ASCII
messages transmitted: CAL when calibration is in progress, and blank when good data is available. Postamble: NUL, EOT, NUL = (00, 04, 00 Hex notation)
Postamble NUL, EOT, NUL =
(00,04,00 Hex notation)
Pin No.
Circuit EIA
Circuit CCITT
Direction of Signal
Description
1 AA 101    
Protective Ground
*2 BA 103
From Terminal
Transmitted Data
*3 BB 104 To Terminal
Received Data
*4 CA 105
From Terminal
Request to Send
*5 CB 106 To Terminal Clear to Send
Parameter. Setting
Baud Rate 1200
Data
8 bits, high order bit set to zero
Parity None
Stop Bits 1
49
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
*6 CC 107 To Terminal
Data Set Ready
*7 AB 102    
Signal to Ground
*8 CF 109 To Terminal
Received Line Signal Detector
9  
Reserved for Data Set Testing
10  
Reserved for Data Set Testing
Rate Selector
24
DA
Transmit Signal Element Timing
25 Unassigned
Pin No.
Circuit EIA
Circuit CCITT
Direction of Signal
Description
11 Unassigned
*12
SCF (High
Speed Indicator)
Secondary Received Line Signal Detector
13 SCB
Secondary Clear to Send
14 SBA
Secondary Transmitted Data
15 DB
Transmission Signal Timing Source
16 SBB
Secondary Received Data
17 DD
Receiver Element Timing (DCE Source)
18 Unassigned
19 SCA
Secondary Request to Send
*20 CD 108.2
From Terminal
Data Terminal Ready
21 CG
Signal Quality Detector
*22 CE
Ring Indicator
23 CH/CI Data Signal
8.2.4 Initiating RS232Communications to the Computer
1. With the computer and CFA3000 power turned off, connect RS232C cable to the computer and
to the RS232C port on the CFA3000 analyzer. (If
a printer is to be used, connect it to the printer port
on the analyzer).
2. Turn on power to the analyzer.
3. From the analyzer keypad panel, turn on the printer
by pressing the following buttons:
• From a multistream keypad:
and , then .
• From a singlestream keypad, referring to Figure 84, press position 5, then 1, and then 10.
4. Turn on computer
Figure 84. Numbered Positions on a
SingleStream Keypad
50
5. If you purchased program from Rosemount Analytical, run it from your computer, by typing in
CFATERM, then press Return.
6. When a blinking cursor appears on the screen,
press Return. The screen will display M>.
7. At this point, all of the RS232C functions listed in
Table 84 are accessible.
8. To set up a file to collect data in the computer, press <altL> (formatted) or <altC> (capture).
9. Type in file name (up to 8 characters).
10. To start collecting data, press Return. The display will show a blinking cursor.
11. To access the command functions again, press
Return. The screen will display M>.
Use the commands in Table 85 to control the analyzer.
Table 85: RS232C Commands
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
Command Function
AUTOCAL Runs fullscale
automatic calibration
AUTOVALV Resumes automatic valve
selection
DATA Forces printing of
analyzer data
DISVALV
Command
Displays stream parameters, including sample cycles, low and Function
MANVALV To operate valves
manually (for multi stream, one stream continuously). For example, to monitor stream 3, enter MANVALV 3.
RESET Resets the computer for
a full wipeout
SETCLO Sets clock calendar, Date
is entered as follows: day of week, year, month, date, hour, minute, second. Day of week is coded as Sunday = 0, Monday = 1, Tuesday = 2, etc. Year is entered as last two digits of year. Month and date must be two digits, use leading zeros as necessary. A 24 hour format is used for hour entry. For example, to set the clock at Thursday, April 12, 2003 at 4:35 p.m., enter SETCLO
4030412163500.
SETHAL Changes stream high
alarm setting. For
example, to change stream 1 to 25, enter SETHAL 1,25.
SETLAL Changes stream low
alarm setting. For example, to change stream 2 to 50, enter SETLAL 2,50.
SETVTIM Changes stream valve
cycles. For example, to change stream 2 to 2 cycles, enter SETVTIM 2,20.
Press…
Analyzer Displays
Press…
5 ENTER
03 ENTER
08 ENTER
16 ENTER
16 ENTER
35 ENTER
51
8.2.5 Printer Interface
The printer interface is a female parallel port designed to be compatible with IBM PCtype cables and printers. This builtin capability permits text printing only, no graphics.
Table 86: Printer Port Pin Configuration
8.2.6 Set Printer Internal Clock
Set printer's internal clock before printing; perform the following procedure:
1. From analyzer keypad panel, turn on printer:
• From a multistream keypad, press:
and , then .
• From a singlestream keypad, referring to Figure 84, press position 15, then 1, and then 10
2. Enter data as follows: a. Day of the week first, one digit:
Sunday = 0, Monday =1
Press enter.
b. Year, two digits: 2003 = 03, 2007= 07
Press enter.
c. Month, two digits: Jan = 01,
Feb = 02
Press enter.
d. Day of the month, two digits: 01, 02
Press enter.
e. Hour, using 24hour format:
8:00 AM = 08, 1:00 PM = 13
Press enter.
f. Minute, two digits: 01, 02
Press enter.
Example: To set the clock for Friday, August 16, 2003
at 4:35 PM:
CFA3000 MANUAL SECTION 8
ADDIDIONAL FEATURES
Pin No. Signal Name Description
1 STB/ Negative going pulse
2 D0 Data Bit 0
3 D1 Data Bit 1
4 D2 Data Bit 2
5 D3 Data Bit 3
6 D4 Data Bit 4
7 D5 Data Bit 5
8 D6 Data Bit 6
9 D7 Data Bit 7
10 ACK/ Acknowledge from printer
11 BUSY Busy, from printer
12 PE Paper empty, from printer
13 SEL/
Printer selected, from printer
14 PFD/ Paper feed, to printer
15 ERR/ Error, from printer
16 INIT/ Initialize, to printer
17 SEL Select, to printer
18 GND Ground
19 GND Ground
20 GND Ground
21 GND Ground
22 GND Ground
23 GND Ground
24 GND Ground
25 GND Ground
52
CFA3000 MANUAL APPENDIX
Method: Sodium01A (0.110, 1100, 101000 ppb to 1000 ppm (contact Rosemount Analytical for other ranges)
General Description:
For this procedure for the determination of sodium, the sodium ion is measured potentiometrically using ion selective electrodes . The addition of an ionic strength adjustor to the sample maintains constant background ionic strength and pH.
Sodium in Water
Range: 0.110, 1100, 101000 ppb up to 1000 ppm or higher.
Reagents Kit: (reagent & standards) 185301130
REAGENT #1, Buffering Reagent: 185301134
Working Standards 10 ppb: 185301133 100 ppb: 182301132
Operating Notes
1. The chemicals used for reagents and distilled water
should be free of sodium.
2. The sample stream must be filtered to 10 microns
or less before being introduced to the system with
an inline filter such as Rosemount Analytical' IF
100
(Part #180140001).
APPENDIX A CHEMISTRIES
53
CFA3000 MANUAL APPENDIX
APPENDIX B SPECIFICATIONS
Accuracy: All twodecade ranges up to 1000 ppb
maximum: ±10% of reading or ±2 ppb, whichever is greater
0.110 ppm range = ±10% of reading or ±0.5 ppm, whichever is greater
Automatic Calibration: 40 minutes
Total Calibration: 40 minutes, twice/week
Reagent Consumption: 5 L of each reagent per 3
months
Standard Consumption: 1 L of high standard,1 L of low standard per 3 months
Sample Requirements:
Flow: 4050 mL/min Current standard bottles are:
“0”(baseline): 1.0 L. Fullscale: 5.0 L
Pressure: <10 psig (<170 kPa) (overflow cup must drain to atmosphere)
Temperature: 050° C (32122° F)
Filter: 8 micron filter in sample line is
Recommended; (#180134108).
Multiple Stream Option: General: up to 6 sample streams; measurement is sequential
Response Time: 15 minutes (methoddependent)
Outputs: 420 mA or 05 Vdc (nonisolated).
(Isolated output is optional.)
Load requirements for 420 mA output: 450 ohms minimum, 1200 ohms? maximum
Load requirements for 05 Vdc output: 10 k Ohms, minimum
The range of outputs is the same as the measure ment range. Parallel output for printer is optional for
single stream, standard for multistream monitors. RS232C output is optional for single stream monitors, standard for multistream monitors.
Alarms:
General: contact closures for
leak detection,
calibration in progress,
Contact closures are rated:2 A at 250 Vac or 3 A
at 30 Vdc; delay on/off is 1 minute
Streams 2 through 6 (multistream option):
• One alarm relay for each stream; contacts have dc ratings
• 100 Vdc (maximum  carry only)
• 0.5 A (maximum  switched)
• delay on/off time is 1 minute
Display: Single line LCD, 4 1/2 digits, readout in concentration units. For multistream option display remains at last value until new stream reading is stable.
Status LEDs:
• Leak,
• Lamp failure,
• Quarterly maintenance due
• Calibration in progress.
For multistream analyzers.
StatusLEDs show:
• the sample stream currently selected
• the sample stream for which data is being displayed.
Environmental: The analyzer is not intended for out door use. Install in a clean, indoor environment.
Ambient Temperature:
40120° F (1045° C) must be regulated to ± 10° within range, avoid temperature swings greater than ± 10° F (± 5± C) over a 12hour period
Humidity: 595% (noncondensing) Pollution degree: Standard Altitude: Not applicable
54
Dimensions: Height of Analyzer Cabinet:
25.75 in. (654 mm) Height of Reagent Cabinet: 18.25 in. (464 mm) Width: 18 in. (457 mm) Depth: 16 in. (406 mm)
Space Requirements: Wall mounting: Allow 18 in. (457 mm) on one side for swing out access to rear panel. Allow 12 in. (305 mm) on the other side for access to input/output panel and overflow sampler. Panel Mounting: Allow 12 in. (305 mm) on both sides for access to input/output panel and overflow panel.
Weight: 117 lb (53 kg).
Add up to 60 lb (27.3 kg) for full weight of reagents and standards. Shipping Weight of Analyzer: 138 lb (63 kg) Shipping Weight of Reagents: up to 64 lb (29 kg) per kit (3 months supply)
APPENDIX B SPECIFICATIONS Cont.
CFA3000 MANUAL APPENDIX
55
The block diagram in Figure C1 graphically shows the distribution of power and signals throughout the analyzer. The raised numbers (e.g., (6) ) in the description below relate to the circled numbers on the diagram. For clarity, minor components such as fuses are not shown.
1. For 110 Vac 50/60 Hz option, line power enters the analyzer at the Power Entry barrier strip TB3 (1), with the following connections (refer to Figure 25 and Figure C1):
• Line (hot) connection is TB31
• Neutral is TB32
• Safety (chassis) ground connection is TB34
For 220 Vac 50/60 Hz option, the following con nections are used:
• L1 connection is TB37
• L2 connection is TB38
• Safety (chassis) ground connection is TB34
2. For 110 VAC 50/60 Hz option, line power flows through the Power Switch and other minor compo nents to barrier strip TB1. TB1 serves as the line power distribution point.
For 220 VAC 50/60 Hz option, line power flows through the Power Switch and other minor compo nents to the primary of the stepdown Transformer (2). The Stepdown transformer converts the input 220 VAC to 110 VAC for compatibility with the standard CFA3000 electronics. From the Stepdown Transformer secondary power flows to barrier strip TB1.
3. From TB1, for both options, line power flows to the Colorimeter Lamp Power Supply, (4) , the Power Transformer (3) , the Pump Power Supply (5) and the Power Supply Board (8e) located in the Card Cage (8).
4. The Colorimeter Lamp Power Supply produces highly regulated DC current which energizes the Colorimeter Lamp (4a).
5. The Power Transformer converts standard line voltage to lowvoltage AC for use by the CFA3000 electronics. The Power Transformer secondaries feed the Power Supply Board, which converts the lowvolt age AC to regulated DC current, and to the heater element in the Heating Bath (7) . Relay contacts on the Power Supply Board are externally available at the I/O Panel (10) for activation of alarm and status indicators.
Multistream Analyzer Option
In units equipped with the Multistream option, a Valve Board (8c) will be installed. This board contains relay contacts that are externally available at the Interface Ports (11) as well as LED drivers to activate the sample stream indicators located on the Keypad Panel (9) . The Valve Board also provides interfaces for a printer and serial data communications, exter nally available at the Interface Ports.
If the monitor includes a multioutput analog board, analog recorder drivers for each stream are installed on the Analog Board (8d) . These outputs are exter nally available at the Interface Ports.
APPENDIX C POWER INTERCONNECT DESCRIPTION
CFA3000 MANUAL APPENDIX
56
CFA3000 MANUAL APPENDIX
APPENDIX D BYPASS FOR CALIBRATION
To avoid extraneous alarms from external devices, rewire output as shown below
Figure D1. Calibration Bypass Wiring
57
CFA3000 MANUAL APPENDIX
APPENDIX E KEYPAD HELP SHEETS
E.1. SingleStream Keypad "Help Sheet"
Figure E1. Keypad Panel for Single Stream Option
Figure E2. Number Positions on a SingleStream Keypad
Press … Function To Exit, Press …
Initiates automatic calibration No action required
Initiates priming and auto calibration No action required
Sets scale to adjust output to 4mA
and 0 Vdc on external device
Referring to Figure E2, on keypad panel, press position 15 and then 2.
Initiate valve pump test operation
Display high alarm setting
or Set high alarm value
Display low alarm setting
or
Change low alarm value
(MUST BE SET LOWER
THAN HIGH ALARM)
CFA3000 MANUAL APPENDIX
Table E1: Singlestream Keypad Help Sheet
58
59
CFA3000 MANUAL APPENDIX
Press … Function To Exit, Press …
Set reagent supply time
Display number of days remaining
until replacement of reagents
Set output scale for 20mA and
5 Vdc
. Display mV
Display Gain value
Initiate grab sample
Display measurement of previous
grab sample
Referring to Figure 84, press
position 15, then 1, and then 10
Turn on printer
No action required
Referring to Figure 84, press posi
tion 15, then 1, and then 10
Turn on printer No action required
Press … Function To Exit, Press …
Initiates automatic calibration No action required
Initiates priming and auto calibration No action required
Sets scale to adjust output to 4mA
and 0 Vdc on external device
Initiate valve pump test operation
60
CFA3000 MANUAL APPENDIX
Table E2. Multistream Keypad Help Sheet
Figure E3. Keypad Panel for Multistream Option
61
CFA3000 MANUAL APPENDIX
Press … Function To Exit, Press …
or
Display a particular stream in multi
stream system, to set parameters
Then, or
Set number of cycles for a particular
stream in multistream system
or
Set high alarm value for stream
(IF ONLY A LOW ALARM VALUE IS
DESIRED, GO TO SET LOW
ALARM
Press … Function To Exit, Press …
or
Change low alarm value
(MUST BE SET LOWER
THAN HIGH ALARM SETTING)
Display low alarm setting
Display high alarm setting
Initiates multistream sampling No action required
Set reagent supply time (90 days)
62
Press … Function To Exit, Press …
Display number of days remaining
until replacement of reagents
Set output scale for 20mA and 5Vdc
Display mV
Display Gain value
Press … Function To Exit, Press …
Initiate grab sample
Display measurement of
previous grab sample
Turn on printer No action required
Turn on printer No action required
CFA3000 MANUAL APPENDIX
63
CFA3000 MANUAL ORDERING
ORDERING INFORMATION
Reagents
Analyzer reagents are available from Rosemount Analytical on a contract basis. Refer to APPENDIX A, "Chemistries" for specific ordering information or contact Rosemount Analytical
Replacement Parts and Accessories
Table E3 Replacement Parts,
Part Part No.
Valve Pump Assembly 185A00101
Solenoid Valve, multistream
180B30101
Waste Tube Assembly 180134601
Mounting Brackets 180200401
Overflow Sampler
Assembly:
1 Stream Analyzer 180130001
2 Stream Analyzer 180130002
3 Stream Analyzer 180130003
4 Stream Analyzer 180130004
5 Stream Analyzer 180130005
6 Stream Analyzer 180130006
IF100 InLine Filter 180140001
Fuse, T 1 A/250 Vdc 625000302
Clamp, singlestream overflow
180132701
Clamp, multistream overflow
180133701
Clamp, drain 180202901
CPU Board 180B02801
Power Supply Board 180B01203
Analog Board 180B02602
Multi Output Board 180B02601
Valve Board 180B01601
Keypad & Display Board
Assembly, single stream
185B00301
Keypad & Display Board
Assembly, multistream
185B00401
Part Part No.
E Prom
(NOTE: specify chem istry range with decimal places, part #'s of analog, and line frequency, 50/60 Hz, number of streams)
Sample Line Tubing (.0625" ID)
180285201
Filter, disposable (for overflow panel assembly)
180134108
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, orings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or nonwarranty service is requested.
5. Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges for inspection and testing to determine the problem with the device.
If the material is returned for outofwarranty repairs, a purchase order for repairs should be enclosed.
65
SECTION 9.0
RETURN OF MATERIAL
9.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 19497578500 for a Return Materials Authorization (RMA) number.
9.2 WARRANTY REPAIR.
The following is the procedure for returning instruments still under warranty:
1. Call Rosemount Analytical for authorization.
2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or subassemblies, the serial number on the unit must be supplied.
3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same manner as they were received.
4. Send the package prepaid to:
Rosemount Analytical Liquid Division 2400 Barranca Parkway Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
9.3 NONWARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individ ual to be contacted should additional information be needed.
3. Do Steps 3 and 4 of Section 17.2.
NOTE
Consult the factory for additional information regarding service or repair.
IMPORTANT
Please see second section of “Return of Materials Request” form. Compliance with the OSHA requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required.
CFA3000 MANUAL SECTION 9
RETURN OF MATERIALS
66
A Worldwide Network of Sales and Service
Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical data and application information.
For more information, please contact your nearest Emerson Process Management sales office.
Immediate, Reliable Analytical Support
Now there’s a way to quickly get the right answers for your liquid analytical instrumentation ques tions: the Analytical Customer Support Center.
Our staff of trained professionals is ready to provide the information you need. If you are placing an order, verifying delivery, requesting application information, or just want to contact a Rosemount Analytical representative, a call to the Customer Support Center will provide you with the right peo ple, the right answers, right now.
The right people, the right answers, right now.
THE AMERICAS  HEADQUARTERS
Emerson Process Management Rosemount Analytical Inc. Liquid Center of Excellence 2400 Barranca Parkway Irvine, CA 92606 Phone: +1.949.757.8500 Toll Free: +1.800.854.8257 Fax: +1.949.474.7250
ASIAPACIFIC
Emerson Process Management Asia Pacific Private Ltd. 1 Pandan Crescent Singapore 0512 Republic of Singapore Phone: 65.777.8211 Fax: 65.777.0947
EUROPE
Emerson Process Management AG Blegistrasse 21 CH6341 BaarWalterswil Switzerland T 41.41.768.6111 T 41.41.761.8740
VISIT OUR WEBSITE AT
www.rosemountanalytical.com
GERMANY
Emerson Process Management Process Gas Analyzer Center of Excellence
GmbH & Co. OHG
Industriestrasse 1
63594 Hasselroth
Germany
T 49.6055.884.0 F 49.6055.884.20
LATIN AMERICA
Emerson Process Management Rosemount Analytical 10241 West Little York, Suite #200 Houston, TX 77040 USA T 713.467.6000 F 713.827.3328
MIDDLE EAST AND AFRICA
Emerson Process Management EPM Building P. O. Box 17033 Jebe Ali Free Zone Dubai, United Arab Emirates T 971.4.8835235 F 971.4.8835312
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ONLINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.rosemountanalytical.com
Emerson Process Management
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 7578500 Fax: (949) 4747250
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2013
8
Loading...