“Damping electronics are available as an option. Transmitters with standard electronics
can be retrofitted with the adjustable damping feature by changing out both the amplifier
board and the calibration board. Please reference Table 6-2 (Parts List Rosemount
Steam Pressure/Temperature
Accuracy as specified below, during and after sequential exposure to steam at the
following temperatures and pressures, concurrent with chemical spray for the first 24
hours.
420 °F (215.6 °C), 85 psig for 3 minutes
350 °F (176.6 °C), 85 psig for 7 minutes
320 °F (160 °C), 75 psig for 8 hours
265 °F (129.4 °C), 24 psig for 56 hours
Range codes 4-8: Accur ac y within ±(1.0% of upper range limit + 1.0% of span)
(1)
Range code 9: Accur acy within ±(2.0% of upper range limit + 0.5% of span)
Note
(1) For Range codes 4-8, if pressure applied at the 4mA point is greater than 50% of the upper range limit, the
follows:
Amplifier Circuit Board, Output Code R:
In all locations, replaced “01154-0001-000 5” with “01154-0153-0001”
Amplifier Circuit Board with Damping, Output Code R:
In all locations, replaced “01154-0021-000 4” with “01154-0156-0001”
Amplifier Circuit Board, Output Code R, N0026:
following part numbers are updated:
Amplifier Circuit Board, Output Code R:
“01154-0153-0001” is replaced by “01154-0263-0001”
Amplifier Circuit Board with Damping, Output Code R:
“01154-0156-0001” is replaced by “01154-0266-0001”
Amplifier Circuit Board, Output Code R, N0026:
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Rosemount 1154 Series H
Alphaline
®
Pressure
Transmitters
NOTICE
Read this manual before working with the product. For personal and system safety, and for
optimum product performance, make sure you thoroughly understand the contents before
installing, using, or maintaining this product.
For equipment service needs outside the United States, contact the nearest Rosemount
representative.
Within the United States, the North American Response Center is at your service 24 hours a
day, and is a single-point contact for all Rosemount equipment service needs. If at any time
you are not sure what to do, you have a question about using the product, or you have a
service or support request, call the center toll free at 1-800-654-RSMT (7768). This contact
is your fastest link to quick and complete answers about any Rosemount group, product, or
service.
Alphaline, Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc.
-Cell is a trademark of Rosemount Inc.
D.C. 55 is a registered trademarks of Dow Corning.
Loctite is a registered trademark of Henkel KGaA Corporation.
Grafoil is a trademark of Union Carbide Corp.
Swagelok is a registered trademark of Crawford Fitting Company.
Lubri-Bond is a registered trademark of E/M Corporation.
Cover Photo: 1153-001AB
Rosemount Nuclear Instruments, Inc. satisfies all obligations coming
from legislation to harmonize product requirements in the European
Union.
www.emersonprocess.com/rosemount/nuclear
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007
Rosemount Nuclear Instruments, Inc. Warranty and
Limitations of Remedy
The warranty and limitations of remedy applicable to this Rosemount equipment are as stated on the reverse of the
current Rosemount quotation and customer acknowledgment forms.
RETURN OF MATERIAL
Authorization for return is required from Rosemount Nuclear Instruments, Inc. prior to shipment. Contact Rosemount Nuclear Instruments,
Inc. (1-952-949-5210) for details on obtaining Return Material Authorization (RMA). Rosemount Nuclear Instruments will not accept any returned material without a Returned Material Authorization. Material returned without authorization is subject to return to
customer.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Nuclear Instruments, Inc.
8200 Market Blvd.
Chanhassen, MN 55317
USA
IMPORTANT
The Rosemount 1154 Series H Pressure Transmitter is designed for Nuclear Class IE usage, has been tested per IEEE Std 323-1974,
IEEE 344-1975 as defined in the Rosemount Report D8700096 and is manufactured to the requirements of NQA-1; 10CFR50, Appendix
B quality assurance programs; and 10CFR Part 21. During qualification testing, interfaces were defined between the transmitter and its
environment that are essential to meeting IEEE Std 323-1974 requirements. To ensure compliance with 10CFR Part 21, the transmitter
must comply with the requirements herein and in Report D8700096 throughout its installation, operation, and maintenance. It is incumbent
upon the user to ensure that the Rosemount Nuclear Instruments, Inc. component traceability program where applicable is continued
throughout the qualified life of the transmitter.
In order to maintain the qualified status of the transmitter, the essential environmental interfaces must not be compromised. Performance
of any operations on the transmitter other than those specifically authorized in this manual has the potential for compromising an essential
environmental interface.
Where the manual uses the terms requirements, mandatory, must, or required, the instructions so referenced must be carefully followed.
Rosemount Nuclear Instruments, Inc. expressly disclaims all responsibility and liability for transmitters for which the foregoing has not
been complied with by the user.
www.emersonprocess.com/rosemount/nuclear
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Revisions
Changes From June 1999 to April 2007
Page (Old) Page (New)Changes
CoverCoverDocument revision date change from June 1999 to April 2007, rev. from AA to BA.
Inside cover
i, ii, 5-7 and
back cover
3-6, and 6-8 3-6, and 6-8 Include errata sheet information on circuit board number changes:
ThroughoutThroughout References to Fisher-Rosemount were changed to Emerson Process Management.
i, back
cover
--Cover, i,
2-1, 3-1,
4-1, 5-1, 6-1
2-2, 6-72-2, 6-7Removed word “process” from sentence indicating user assumes responsibility for qualifying the connection
2-22-3Updated reference to Swagelok catalog and added web address, removed street address.
2-42-4Removed reference to Model 353C.
2-42-5Rearranged wording on shielded cable.
2-6 and 2-72-7 and 2-8 Added word ‘nominal’ to Notes in drawings. Changed significant digits to conform to standard.
2-72-8Changed height from 9.0 max (229) to 9.3 max (236). Changed shroud width from 3.40 (86.4) to 3.48 (88.3).
6-16-1Changed ISO 9001 to ISO 9001:2000.
--Back cover Added trademark and registration information.
ii, 5-7 and
back cover
Cover, i, ii
and back
cover
and back
page
2-1, 3-1,
4-1, 5-1, 6-1
Include errata sheet information on address and phone number.
Replaced amplifier circuit board, output code R P/N 01154-0001-0001 with 01154-0001-0005.
Replaced amplifier circuit board with damping, output code R P/N 01154-0021-0002 with 01154-0021-0004.
Replaced amplifier circuit board for N0026, output code R P/N 01154-0001-0002 with 01154-0001-0006.
Web address changed from www.rosemount.com to www.emersonprocess.com/rosemount/nuclear.
Added reference to European Union product requirement (CE).
Added table of contents to each section.
interface.
NOTE
The above Revision Status list summarizes the changes made. Please refer to both manuals for complete
comparison details.
OVERVIEWThis manual is designed to assist in installing, operating, and maintaining the
Rosemount 1154 Series H Alphaline® Nuclear Pressure Transmitter. The
manual is organized into the following sections:
Section 2: Installation
Provides general, mechanical, and electrical installation considerations to
guide you through a safe and effective transmitter installation.
Section 3: Calibration
Provides transmitter calibration procedures.
Section 4: Operation
Provides descriptions of how the transmitter operates.
Section 5: Maintenance and Troubleshooting
Provides basic hardware troubleshooting considerations including sensing
module checkout, disassembly and reassembly procedures, and
post-assembly tests.
ABOUT THE
TRANSMITTER
Section 6: Specifications and Reference Data
Provides nuclear, performance, functional, and physical transmitter
specifications; also includes ordering information, and a list of spare parts.
Rosemount 1154 Series H Alphaline Pressure Transmitters are designed for
precise pressure measurements in nuclear applications requiring reliable
performance and safety over a specified qualified life. These transmitters
were generically tested to IEEE Std 323-1974 and IEEE Std 344-1975 per the
Qualification Test Report D8700096. The Rosemount 1154 Series H has been
qualification tested to environments typical of pressurized water reactors
(PWR) under accident conditions. Stringent quality control during the
manufacturing process includes traceability of pressure retaining parts,
special nuclear cleaning, and hydrostatic testing.
Rosemount 1154 Series H Transmitters are of a design unique to Class 1E
nuclear service while retaining the working concept and design parameters of
the Rosemount 1151 that has become a standard of reliable service. Units are
available in sealed reference (S), differential (D), and high-line differential (H)
configurations, with a variety of pressure range options (see Table 6-1 on
page 6-7). Figure 2-5 on page 2-8 shows dimensional drawings of the
transmitters.
OVERVIEWThis section contains information and instructions regarding the following
installation-related information:
•General Considerations
•Mechanical Considerations
Process Connections
Conduit
•Electrical Considerations
Mechanical
Electrical
GENERAL
CONSIDERATIONS
The quality and accuracy of flow, level, or pressure measurement depends
largely on the proper installation of the transmitter and its associated impulse
piping and valves. For flow measurement, proper installation of the primary
measuring element is also critical to the accuracy of the measurement.
Transmitter installation should minimize the effects of temperature gradients
and temperature fluctuations, and avoid vibration and shock during normal
operation. Take care when designing the measurement to minimize the error
caused by incorrect installation. The temperature of the transmitter ambient
environment affects the qualified life of the transmitter. (See Figure 2-1.)
www.emersonprocess.com/rosemount/nuclear
Rosemount 1154 Series H
Figure 2-1. Qualified Life vs.
Ambient Temperature.
Reference Manual
00809-0100-4631, Rev BA
April 2007
Module
Qualified Life
Electronics
Qualified Life
Time (Years)
MECHANICAL
CONSIDERATIONS
Ambient Temperature (°F)
This section contains information you should consider when preparing to
mount the transmitter. Read this section carefully before proceeding to the
1154-0293A
mechanical installation procedure.
Mount the Rosemount 1154 Series H transmitter to a rigid support (i.e., one
with a fundamental mechanical resonant frequency of 40 Hz or greater). A
mounting bracket included with the transmitter facilitates panel mounting.
Figure 2-4 on page 2-7 shows the qualified mounting configurations. The
transmitter was seismic tested and qualified with the bracket mounted with
3
/8 in. diameter bolts. Orientation with respect to gravity is not critical to
four
qualification. However, if the transmitter is mounted with the flanges in a
horizontal position, zero the transmitter to cancel the liquid head effect caused
by the difference in height of the process connections.
If the transmitter is mounted to a non rigid panel, ensure that seismic input to
the mounting bracket does not exceed qualification levels given in Rosemount
Report D8700096.
Process ConnectionsProcess tubing installation must prevent any added mechanical stress on the
transmitter under seismic disturbances. This may be done by using
stress-relief loops in the process tubing or by separately supporting the
process tubing near the transmitter.
3
The process connections to the transmitter flanges were qualified with
tubing using Swagelok
®
compression fittings. For options using 1/4–18 NPT
connections, the user assumes responsibility for qualifying the interface.
/8 in.
2-2
Transmitters with Flange Options A or H are shipped with Swagelok fittings for
process connections. Included are front ferrule, rear ferrule, and nut. Ensure
the fittings are placed on the tubing with the orientation and relative position
shown in Detail A in Figure 2-5 on page 2-8.
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Process tubing used is 3/8 in. outside diameter, and of suitable thickness for
the pressure involved.
The Swagelok tube fittings are shipped completely assembled for immediate
use. Do not disassemble them before use; doing so may contaminate the
fittings and result in leaks. Insert the tubing into the Swagelok tube fitting,
making sure that the tubing rests firmly on the shoulder of the fitting and that
the nut is finger tight. Tighten the nut one-and-one-quarter turns past finger
tight to prepare the transmitter for use. Do not overtighten.
The connections can be loosened and re-tightened 20–30 times without
compromising the leak-proof seal. To reconnect, insert the tubing with
pre-swaged ferrules into the fitting until the front ferrule sits in the fitting.
Tighten the nut by hand, then rotate one-quarter turn more or to the original
one-and-one-quarter tight position. Then snug it slightly with a wrench. For
more information regarding the use of Swagelok tube fittings, refer to:
Fittings Catalog MS-01-140
“Gaugeable Tube Fittings and Adapter Fittings”
www.swagelok.com
If the drain/vent valves must be opened to bleed process lines, torque them to
1
/2 ft-lb (10 N-m) when closing.
7
Proper location of the transmitter with respect to the process tubing depends
on various process parameters. When determining the best location, consider
the following:
•Keep hot or corrosive fluids from contacting the transmitter.
•Prevent sediment from depositing in the impulse tubing.
•Ambient temperature gradients and fluctuations can result in erroneous
transmitter readings.
•Keep impulse tubing as short as possible.
•For differential transmitters, balance the liquid head on both legs of the
impulse tubing.
•For liquid flow or pressure measurements, make taps on the side of the
line to avoid sediment deposits and mount the transmitter beside or
below the taps so gases vent into the process line (see Figure 2-6 on
page 2-9).
•For gas flow or pressure measurements, make taps on the top or side
of the line and mount the transmitter beside or above the taps so liquid
drains into the process line (see Figure 2-6 on page 2-9).
•For steam flow or pressure measurements, make taps on the side of
the line and mount the transmitter below the taps so the impulse tubing
stays filled with condensate (see Figure 2-6 on page 2-9).
•For steam service, fill the lines with water to prevent steam from
contacting the transmitter. Condensate chambers are not necessary
since the volumetric displacement of the transmitter is negligible.
2-3
Rosemount 1154 Series H
The piping between the process and the transmitter must transfer the
pressure measured at the process taps to the transmitter. Possible sources of
error in this pressure transfer are:
•Leaks
•Friction loss (particularly if purging is used)
•Trapped gas in a liquid line or trapped liquid in a gas line (head error)
•Temperature-induced density variation between legs (head error) for
differential transmitters
To minimize the possibility of errors, take the following precautions:
•Make impulse tubing as short as possible.
•Slope tubing at least one inch per foot up toward the process
connections for liquid and steam.
•Slope tubing at least one inch per foot down toward the process
connections for gas.
•Avoid high points in liquid lines and low points in gas lines.
•Use impulse tubing of sufficient diameter to avoid friction effects.
•Ensure that all gas is vented from liquid tubing legs.
•Ensure that impulse tubing is of adequate strength to be compatible
with anticipated pressures.
Reference Manual
00809-0100-4631, Rev BA
April 2007
For differential transmitters, also consider the following:
•Keep both impulse legs at the same temperature.
•When using sealing fluid, fill both piping legs to the same level.
•When purging, make the purge connection close to the process taps
and purge through equal lengths of the same size tubing. Avoid purging
through the transmitter.
ConduitThe conduit connection to the transmitter is
conduit seal at the conduit entry to prevent moisture from accumulating in the
terminal side of the housing during accident conditions. To prevent the conduit
from adding mechanical stress to the transmitter during seismic disturbances,
use flexible conduit or support the conduit near the transmitter. Install the
conduit seal in accordance with the manufacturer’s instructions or use the
procedure on page 2-9.
ELECTRICAL
CONSIDERATIONS
This section contains information that you should consider when preparing to
make electrical connections to the transmitter. Read this section carefully
before proceeding to the electrical installation procedures.
The Rosemount 1154 Series H Pressure Transmitter provides a 4–20 mA
signal when connected to a suitable dc power source. Figure 2-2 on page 2-5
shows a typical signal loop consisting of transmitter, power supply, and
various receivers (controller, indicator, computer, etc.). The power supply
must supply 12 volts minimum to the transmitter terminals at 30 mA
(overscale) signal, or the maximum output current required for proper system
operation. Any power supply ripple will be seen in the output load. The supply
voltage versus load limitation relationship is shown in Figure 2-3 on page 2-5.
See qualification report D8700096 for details. The load is the sum of the
resistance of the signal leads and the load resistance of the receivers.
1
/2–14 NPT. Use a qualified
2-4
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Signal wiring need not be shielded, but twisted pairs should be used for best
results. In electrically noisy environments, shielded cable should be used for
best results. Do not run signal wiring in conduit or open trays with power
wiring, nor near heavy electrical equipment. Signal wiring may be ungrounded
(floating) or grounded at any place in the signal loop. The transmitter case
may be grounded or ungrounded.
The capacitance sensing element uses alternating current to generate a
capacitance signal. This alternating current is developed in an oscillator circuit
with a frequency of 32,000 ±10,000 Hz. This 32,000 Hz signal is capacitor
coupled to transmitter case ground through the sensing element. Because of
this coupling, a voltage may be imposed across the load, depending on
choice of grounding.
This impressed voltage, which is seen as high frequency noise, will have no
effect on most instruments. Computers with short sampling times in a circuit
where the negative transmitter terminal is grounded, will detect a significant
noise signal. Filter this noise with a large capacitor (1 µf) or by using a 32,000
Hz LC filter across the load. Signal loops grounded at any other point are
negligibly affected by this noise and do not need filtering.
Figure 2-2. Transmitter Wiring
Connections.
Figure 2-3. Transmitter Load
Limits.
Terminal Side
(cover removed)
1825
1575
1500
1000
Power
Supply
1154-G05A
4–20 mA dc
Qualified
Region
45
Design
Region
50
2-5
500
0
Load Resistance ()
0
12
13.5
Power Supply (V dc)
35
40
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
INSTALLATION
PROCEDURES
Installation consists of mounting the transmitter and conduit and making
electrical connections. Following are procedures for each operation.
MechanicalTransmitter
Be careful not to break the neck seal between the sensor module and the electronics
housing.
The threaded interface between the sensor module and the electronics
housing is hermetically sealed before shipment. The integrity of this seal is
necessary for the safe operation of the transmitter during accident conditions.
If the seal is broken, reseal it according to “Connecting Electrical Housing to
Module Shroud” on page 5-4.
1.Mount bracket to a panel or other flat surface as shown in Figure 2-4
on page 2-7. Use four
SAE grade 2, AISI 316 SST, or ASTM A193 grade 8M class 1 bolts
were used during qualification testing. Torque each bolt to 33.5 ft-lb
(45.4 N-m).
2.Attach the transmitter to the mounting bracket, as shown in Figure 2-4
on page 2-7. Use four
with unit). Torque each bolt to 21 ft-lb (29 N-m).
NOTE
All dimensions are nominal in inches (millimeters).
1154-A99B, B99B
2-7
Rosemount 1154 Series H
d
Figure 2-5. Transmitter Dimensional Drawings.
ROSEMOUNT 1154DH AND 1154HH
9.3 Max.
(236)
4.7 Max.
(119.4)
Dim.
A
4.3
(109)
7
/16 - 20
UNF
(Typical)
1.63
(41.3)
1
/2 - 14 NPT
Conduit
connection
(1 Place)
4.2
(107)
Compression fittings (2) Swagelok for 3/8 in.
4.72 Max.
(119.9)
Welded
drain/vent
3.48 (88.3)
0.8(20) to end
of mating
tubing
Tubing (optional 1/4 - 18 NPT available)
valve (2)
(optional
1
available)
Reference Manual
00809-0100-4631, Rev BA
April 2007
0.75 (19)
clearance for
cover
removal
Nameplate
(remove for
zero and
span adjust)
Transmitter
Circuitry
(this side)
Terminal
connections
(this side)
/4 - 18
NPT
3
/8 in.
Mating
Tubing
Detail A
0.8
(20)
4.7 Max.
(119.4)
7
9.3 Max.
(236)
/16 - 20
UNF
(typical)
1.63
(41.3)
Dim. A
Welde
4.3
Plug
(109)
NOTE
All dimensions are nominal in inches (millimeters).
1
ROSEMOUNT 1154SH
/2 - 14 NPT
Conduit
connection
(1 Place)
4.2
(107)
Compression fittings (1) Swagelok for 3/8 in.
Tubing (optional 1/4 - 18 NPT available)
4.72 Max.
(119.9)
3.48 (88.3)
0.8(20) to end
of mating
tubing
Welded
drain/vent
valve (1)
(optional
1
/4 - 18
NPT
available)
Pressure
Range Code
4, 52.13 (54.0)
6, 72.19 (55.6)
82.25 (57.2)
92.28 (57.9)
Nameplate
(remove for
zero and
span adjust)
Transmitter
Circuitry (this side)
Ter min al
connections
(this side)
Dimension A
0.75 (19)
clearance for
cover
removal
Welded
Plug
2-8
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Conduit 1.Seal the conduit threads with thread sealant. (The transmitter conduit
™
seal interface was qualified using Grafoil
mate with a standard
2.Starting at zero thread engagement, install the conduit into the
transmitter between 4 and 7 turns, or a minimum of 12.5 ft-lb (16.9
N-m). Hold the electronics housing securely to avoid damaging the
threaded neck seal between the sensor module and the electronics
housing during conduit installation.
3.Provide separate support for the conduit if necessary.
Electrical1.Remove the cover from the terminal side of the transmitter
(see Figure 2-5 on page 2-8).
2.Connect the power leads to the “SIGNAL” terminals on the transmitter
terminal block (see Figure 2-7 on page 2-10). Torque the terminal
screws to 5 in-lb (0.6 N-m) or hand tight.
Do not connect signal leads to the “TEST” terminals.
3.Recheck connections for proper polarity.
4.Check the cover O-ring grooves for cleanliness. If chips or dirt are
present, clean the seat and mating portion of the cover with alcohol.
Lubricate replacement O-ring with O-ring grease (Rosemount P/N
01153-0248-0001 or P/N 01153-0053-0001). The transmitter was
qualified using Dow Corning
5.Spray the inside threads of the electronics covers with cover lubricant
(Rosemount P/N 01153-0333-0001 or equivalent) if necessary; if
covers are already sufficiently lubricated, do not spray.
6.Carefully replace cover and tighten to 16.5 ft-lb (22.4 N-m).
OVERVIEWEach transmitter is factory calibrated to the range specified by the customer.
This section contains the following transmitter calibration information:
•Calibration
Span Adjustment
Zero Adjustment
•Calibration Procedures
Zero and Span Adjustment
Linearity Adjustment
Damping Adjustment
Correction for High Line Pressure
CALIBRATIONThe Rosemount 1154DH, HH, and SH transmitters are factory calibrated to
the range shown on the nameplate. This range may be changed within the
limits of the transmitter. Zero may also be adjusted to elevate or suppress.
The span and zero adjustments are external and located under the
nameplate.
Span AdjustmentThe span on any Rosemount 1154 Series H Transmitter is continuously
1
adjustable to allow calibration anywhere between maximum span and
maximum span. For example, the span on a Range Code 4 transmitter can be
continuously adjusted between 0–150 inH
0–25 inH
O.
2
O and
2
/6 of
Zero AdjustmentThe zero can be adjusted for up to 500 percent of span suppression or 600
percent of span elevation (see Figure 3-1).
The zero may be elevated or suppressed to these extremes with the limitation
that no applied pressure within the calibrated range exceeds the full-range
pressure limit. For example, a Range Code 4 transmitter cannot be calibrated
for 150 inH
200 inH
Code 4.
The transmitter may be calibrated to cross zero (e.g., –75 inH
75 inH
O to 200 inH2O (only 300 percent zero suppression) because the
2
O exceeds the 150 inH2O upper range pressure limit of a Range
2
O to
O), but this may result in a slight loss of linearity.
2
2
www.emersonprocess.com/rosemount/nuclear
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
Figure 3-1. Zero Adjustment
Range.
20
600% Zero Elevation
–150 –125 –100
Output
(mA)
0
Pressure (inH2O)
No Zero Elevation or Suppression
20
Output
(mA)
Rosemount 1154 Series H with a calibrated span of 25 inH
4
➀
Graphs based on a Range 4 (0–25 to 0–150 inH2O)
500% Zero Suppression
25 50
0
–75 –50
Pressure (inH2O)
600% Zero Elevation
20
4
25
75 100 125 150
Pressure (inH2O)
500% Zero Suppression
–250
➀
➀
Output
(mA)
4
25
➀
2
O
1152-0282A
CALIBRATION
PROCEDURE
Zero and Span
Adjustment
NOTE
The Rosemount 1154 Series H Pressure Transmitter contains electronic
circuit boards which may be static sensitive.
NOTE
Covers need not be removed for zero and span adjustment.
The zero and span adjustment screws are accessible externally. They are
located behind the nameplate on the side of the electronics housing (see
Figure 3-2). The output of the transmitter increases with clockwise rotation of
the adjustment screws.
The zero adjustment screw has very little effect on the span. The span
adjustment, however, does affect zero. The effect of interaction is more
apparent with suppression or elevation. The span adjustment changes the
zero output and the full-scale output by approximately the same percentage.
Therefore, it is best to calibrate the transmitter from 0 to the desired span and
finish the calibration by adjusting the zero screw to achieve the desired
elevation or suppression.
3-2
Reference Manual
00809-0100-4631, Rev BA
April 2007
Figure 3-2. Zero and Span
Adjustment Screws.
Rosemount 1154 Series H
Zero
Screw
Span
Screw
1154-0314A
Example (for Range Code 4)
Initial transmitter calibration: 25 to 125 inH
suppressed 25 inH
2
O).
Desired transmitter calibration: –75 to –25 inH
elevated 75 inH
O).
2
O (100 inH2O span with zero
2
O (50 inH2O span with zero
2
1.Adjust the zero to eliminate any existing zero elevation or
suppression. With 0 inH
O pressure applied to the transmitter, turn
2
zero adjustment until output reads 4 mA. The unit is now calibrated
for 0 to 100 inH
O.
2
2.Adjust the span to the desired new span. To reduce the span, turn the
span screw until the output, with 0 in H
Existing Span
4mA
⎛⎞
---------------------------------------
⎝⎠
Desired Span
3.Adjust the zero screw to bring the output, with 0 inH
O pressure input, equals:
2
100 inH
4mA
⎛⎞
------------------------------
⎝⎠
50 inH
2
O
2
O
2
8mA==
O input, back to
4 mA. The transmitter calibration should now be very close to 0 to 50
inH
O.
2
4.Check full-scale output and fine tune the span and zero adjustment if
required. Remember zero adjustments do not affect span, but span
adjustments do affect zero predictably. Adjusting the span screw will
affect the zero 1/5 as much as it affects the span. To compensate for
this effect, simply overadjust by 25 percent. For example, if after
completing Step 3, the transmitter output read 19.900 mA at 50
O, turn the span potentiometer until the output (at 50 inH2O)
Since the span adjustment affects zero one-fifth as much as the span,
the 0.125 mA increase in span causes a 0.025 mA increase in zero.
Therefore, turn the zero adjustment (at 50 inH
reads 20.000 mA. The unit should now be calibrated for 0 to 50 in
H
5.Zero Elevation/Suppression. Elevate zero. Turn screw until the output
reads 4 mA with –75 inH
(applying 75 inH
output may stop changing before the desired 4 mA reading is
obtained. If this occurs, turn off power to the unit and unplug the
amplifier board (see Figure 5-2 on page 5-6 to locate the amplifier
board). To elevate or suppress zero a large amount, use the following
procedure:
•Solder: 60% tin, 40% lead (60/40)—Fed Spec QQ-S-571
•Flux: MIL F 14256 Type A; Fed Spec QQ-S-571 Type RA
Method
a. Locate the three turret terminals on the component side of the
b. To elevate zero, connect a jumper wire between the middle
c. Wrap the jumper wire once around each terminal and cut off any
d. Solder the jumper wire to the terminals using proper electronics
e. Plug the amplifier board back in and complete the zero adjustment.
Reference Manual
00809-0100-4631, Rev BA
April 2007
O) until the output
2
O.
2
O applied to the high side of the transmitter
O to the low side will give the same result). The
2
amplifier board. Remove any jumper wires between them (see
Figure 3-3).
terminal and the terminal marked “EZ” (see Figure 3-3, Detail B).
excess.
soldering techniques. Clean solder joints thoroughly with isopropyl
alcohol.
2
3-4
To suppress zero, follow the same procedure, except connect the
jumper wire between the middle terminal and the terminal
marked “SZ” (see Figure 3-3, Detail C).
6.Recheck full scale and zero and fine tune if necessary.
NOTE
There is some mechanical backlash in the zero and span adjustments, so
there is a dead band when you change the direction of adjustment. Because
of the backlash, the simplest procedure, if the desired setting is overshot, is to
intentionally overshoot a larger amount before reversing the direction of the
adjustment.
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Figure 3-3. Jumper Wire
Placement.
DETAIL A
Moderate
Elevation/
Suppression
(No Jumper Wire)
DETAIL B
(To Elevate Zero)
Jumper Wire
Jumper
DETAIL C
(To Suppress Zero)
Wire
Linearity AdjustmentIn addition to the span and zero adjustments, there is a linearity adjustment
located inside the transmitter on the amplifier board (see Figure 3-4 on
page 3-6). Linearity is factory calibrated for optimum performance over the
calibrated range of the instrument and is not normally adjusted in the field. If
you want to maximize linearity over some particular range, use the following
procedure:
1.Apply mid-range pressure and note the error between theoretical and
actual output signal.
2.Apply full-scale pressure. Multiply the error noted in Step 1 by six and
by the following rangedown factor:
3.Add the result to the full-scale output for negative errors, or subtract
the result from the full-scale output for positive errors, by adjusting the
linearity trimmer (see Figure 3-4). Example: At
4-to-1 rangedown the midscale point is low by 0.05 mA. Therefore,
adjust the “Linearity” trimmer until full-scale output increases by (0.05
mA 6 4) = 1.2 mA.
4.Readjust zero and span.
NOTE
If you remove either cover during the above procedures, replace the O-ring
and torque the cover per the instructions given in Section 5: Maintenance and Troubleshooting. Spare cover O-rings are supplied with each
transmitter.
3-5
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007
Figure 3-4. Linearity and
Damping Adjustment.
Damping
Adjustment
(optional)
Linearity
Adjustment
Electronics Side of Transmitter
Housing (cover removed)
Damping AdjustmentDamping electronics are available as an option. Transmitters with standard
electronics can be retrofitted with the adjustable damping feature by changing
out both the amplifier board (RMT P/N 01154-0021-0004) and the calibration
board (RMT P/N 01154-0023-0002).
The damping adjustment permits damping of rapid pressure variations by
adjusting the single-turn trim potentiometer located on the upper right-hand
side of the amplifier board (see Figure 3-4). The available settings, when
adjusted to the maximum position, provide time-constant values of at least 1.2
seconds for Range Code 4 and 0.8 seconds for Range Codes 5–9.
Transmitters with the electronic damping option are calibrated and shipped
with the adjustment set at the counterclockwise stop, which gives the
minimum time-constant.
1154-0299B
Correction For High Line
Pressure (Rosemount
1154DH and 1154HH
only)
To adjust the damping, turn the damping adjustment potentiometer until the
desired time-constant is obtained. It is best to set the damping to the shortest
possible time-constant. Since transmitter calibration is not affected by the
damping setting, you may adjust the damping with the transmitter installed on
the process.
The damping adjustment potentiometer has positive stops at both ends. Forcing the
potentiometer beyond the stops may cause permanent damage.
NOTE
If you remove either cover during the above procedures, replace the
O-ring and torque the cover per the instructions given in Section 5:
Maintenance and Troubleshooting of this manual. Spare cover
O-rings are supplied with each transmitter.
If a differential transmitter is calibrated with the low side at ambient pressure
but will be used at high line pressure, correct the span adjustment to
compensate for the effect of static pressure on the unit. If zero is elevated or
suppressed, also correct the zero adjustment. Correction factors, expressed
in percent of differential pressure input at end points per 1,000 psi static
pressure, are:
3-6
Reference Manual
00809-0100-4631, Rev BA
April 2007
SpanRange Codes 4, 5, and 8:
+0.75% of input/1,000 psi
Range Codes 6 and 7:
+1.25% of input/1,000 psi
The correction procedure below uses the following example:
Rosemount 1154 Series H
Range Code 5, calibrated at –100 to 300 inH
O to be operated at 1,200 psi
2
line pressure. Note that steps 3–6 are omitted for ranges based at zero
differential pressure.
1.Calibrate the unit per the “Calibration Procedure” on page 3-2 to
output = 4 mA at –100 inH
O and 20 mA at 300 inH2O.
2
2.Calculate correction factor:
0.75 %
-----------------------1,000 psi
1 200 psi0.9% differential input=,×
3.Calculate zero adjustment correction in terms of pressure:
0.9 %100 inH
–×O 0.9 inH2O–=
2
4.Convert pressure correction to percent of input span:
10. Add the mA correction to the ideal full-scale output (20 mA). This is
the corrected ideal full-scale output.
20.00 mA 0.108 mA20.108 mA=+
3-7
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007
11. Readjust zero and span adjustments for corrected outputs:
3.964 mA at –100 inH
20.108 mA at 300 inH
O
2
O
2
There is an uncertainty of ±0.5 percent of input reading per 1,000 psi
associated with the span correction.
ZeroZero shift with static pressure is not systematic. However, if the calibrated
range includes zero differential pressure, the effect can be trimmed out after
installation and with the unit at operating pressure.
Equalize pressure to both process connections, and turn the zero adjustment
until the ideal output at zero differential input is observed. Do not readjust the
span potentiometer.
If the transmitter does not include zero differential pressure within its
calibrated span, the zero effect or zero correction can be determined before
the unit is suppressed or elevated to eliminate the zero effect after correcting
for the span effect.
The following procedure illustrates how to eliminate the zero effect for a
non-zero differential pressure calibration. The example uses a Range Code 5
calibrated from 100 to 500 inH
1.Using standard calibration procedures, calibrate the unit to the
required span, with the 4 mA or zero point corresponding to zero
differential pressure:
O with 1,200 psi static line pressure.
2
4 mA at 0 inH
O and 20 mA at 400 inH2O
2
2.Apply static pressure to both high and low process connections with
zero differential pressure across the transmitter, and note the zero
correction (zero shift). For example, if the output reads 4.006 mA, the
zero correction is calculated as:
4.00 mA – 4.006 mA = –0.006 mA
Note the sign associated with this correction, as this result is added
when determining the final, ideal transmitter output.
3.Remove static pressure and correct for the span effect as outlined in
the span correction procedure. Calibrate the unit to the calculated
output values. If, for example, the span correction procedure yielded
4.029 mA and 20.144 mA, calibrate the unit for:
4.029 mA at 100 inH
20.144 mA at 500 inH
O
2
O
2
4.Add the zero correction (–0.006 mA), found in Step 2, to the ideal
zero point value calculated in Step 3.
4.029 mA + (–0.006 mA) = 4.023 mA
5.To eliminate the zero effect, readjust the zero potentiometer so the
output reads the ideal zero point calculated in Step 4 (do not readjust
the span potentiometer). Note that all the calibration points will shift
the same amount toward the correct reading. The example output is
now 4.023 mA at 100 inH
O.
2
The transmitter output is now 4–20 mA over its calibrated span when the unit
is operated at 1,200 psi static line pressure.
The block diagram in Figure 4-2 on page 4-3 illustrates the operation of the
transmitter.
The Rosemount 1154 Series H Alphaline Pressure Transmitters have a
variable capacitance sensing element, the -Cell (Figure 4-1 on page 4-2).
Differential capacitance between the sensing diaphragm and the capacitor
plates is converted electronically to a 2-wire 4–20 mA dc signal.
C2C1–
⎛⎞
--------------------
=
PK
⎜⎟
1
+
C
⎝⎠
1C2
Where:
Pis the process pressure.
Rosemount 1154 Series H
K1is a constant.
C
is the capacitance between the high pressure side and the sensing
1
diaphragm.
C
is the capacitance between the low pressure side and the sensing
2
diaphragm.
Where:
I
is the current source.
ref
V
is the peak-to-peak oscillation voltage.
p-p
fis the oscillation frequency.
fV
pp–
--------------------=
C1C2+
Reference Manual
00809-0100-4631, Rev BA
April 2007
I
ref
Figure 4-1. The -Cell.
Leadwires
Isolating
Diaphragm
I
difffVpp–C2C1
–()=
Where:
is the difference in current between C1 and C2.
I
diff
Capacitor
Plates
Sensing Diaphragm
Rigid
Insulation
Silicone Oil
Welded Seals
1151-0142A
4-2
Therefore:
C2C1–
⎛⎞
PConstant I
==
×I
diff
--------------------
⎜⎟
ref
+
C
⎝⎠
2C1
Reference Manual
00809-0100-4631, Rev BA
April 2007
Figure 4-2. Electrical Block
Diagram.
Rosemount 1154 Series H
SENSOR
OSCILLATOR
DEMODULATOR
OSC.
CONTROL
AMP.
CURRENT
DETECTOR
VOLTAGE
REGULATOR
CURR.
CONTROL
AMP.
CURRENT
LIMITER
CURRENT
CONTROL
REVERSE
POLARITY
PROTECTION
-
TEST
+
SIGNAL
THE -CELL™ SENSORProcess pressure is transmitted through an isolating diaphragm and silicone
oil fill fluid to a sensing diaphragm in the center of the
pressure is transmitted in like manner to the other side of the sensing
diaphragm.
The position of the sensing diaphragm is detected by the capacitance plates
on both sides of the sensing diaphragm. The capacitance between the
sensing diaphragm and either capacitor plate is approximately 150 pf. The
sensor is driven through transformer windings by an oscillator at roughly 32
kHz and 30 V
p-p
.
-Cell. The reference
1154-0295A
DEMODULATORThe demodulator consists of a diode bridge that rectifies the ac signal from
the sensor cell to a dc signal.
The oscillator driving current, I
(the sum of the dc currents through two
ref
transformer windings) is controlled to be a constant by an integrated circuit
amplifier.
The dc current through a third transformer winding is a current directly
proportional to pressure, i.e.,
I
diff
fV
–()=
pp–C2C1
The diode bridge and span temperature compensating thermistor are located
inside the sensor module. The effect of the thermistor is controlled by
resistors located in the electronics housing.
LINEARITY
ADJUSTMENT
Linearity is adjusted by a variable-resistance network, capacitor, and diodes.
The currents generated through this part of the circuit are summed into the
inputs of the oscillator control circuit. This provides a programmed correction
that raises the oscillator peak-to-peak voltage to compensate for first-order
nonlinearity of capacitance as a function of pressure.
OSCILLATORThe oscillator has a frequency determined by the capacitance of the sensing
element and the inductance of the transformer windings.
4-3
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
The sensing element capacitance is variable. Therefore, the frequency is
variable about a nominal value of 32 kHz.
An integrated circuit amplifier is used as a feedback control circuit and
controls the oscillator drive voltage such that:
fV
pp–
--------------------=
C1C2+
I
ref
April 2007
VOLTAGE REGULATORThe transmitter uses a zener diode, transistor, and resistors to provide a
constant voltage of 6.4 V dc for the reference and 7 V dc for the oscillator.
ZERO AND SPAN
ADJUSTMENTS
Zero adjustment components consist of a potentiometer and resistor that
develop a separate adjustable current that sums with the sensor current. The
coarse zero switch switches resistors into the circuit as needed.
Span adjustment is performed with a potentiometer which determines the
amount of loop current which is sensed and fed back to the current control
amplifier.
CURRENT CONTROLThe current control amplifier consists of an integrated circuit, two transistors,
and associated components. The IC reference voltage is established at the
junction of a resistor network. The current control amplifier drives the current
control to a level such that the current detector feeds back a signal equal to
the sum of the zero current and the variable sensor current.
CURRENT LIMITA current limiter prevents the output current from exceeding 30 mA in an
overpressure condition.
REVERSE POLARITY
A zener diode provides reverse polarity protection.
Use only the procedures and new parts specifically referenced in this manual to ensure
specification performance and certification compliance. Unauthorized procedures or parts
can render the instrument dangerous to life, limb, or property.
OVERVIEWThis section contains a technique for checking out the components, a method
for disassembly and reassembly, and a troubleshooting guide.
NOTE
Maintenance of traceability of any replacement parts is the responsibility of
the user (see Important Notice on page 6-9 and Important Notice at the
beginning of this manual, preceding the Table of Contents).
The Rosemount 1154 Series H has no moving parts and requires a minimum
of scheduled maintenance. Calibration procedures for range adjustment are
outlined in Section 3: Calibration. A calibration check should be conducted
after inadvertent exposure to overpressure, unless your plant considers this
factor separately in the plant error analysis.
Test terminals are available for in-process checks. For further checks, the
transmitter can be divided into three active physical components: the sensing
module, the amplifier board, and the calibration board.
An exploded view of the transmitter is provided in Figure 5-2 on page 5-6. In
the following procedures, numbers in parenthesis refer to item numbers in the
exploded view.
www.emersonprocess.com/rosemount/nuclear
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007
SAFETY MESSAGESInstructions and procedures in this section may require special precautions to
ensure the safety of the people performing the operations. Information that
raises potential safety issues is indicated by a warning message. The
following warning messages appear in this section.
Use only the procedures and new parts specifically referenced in this manual to ensure
specification performance and certification compliance. Unauthorized procedures or parts
can render the instrument dangerous to life, limb, or property.
TEST TERM IN ALSThe test terminals are connected across a diode through which the loop
signal current passes. The indicating meter or test equipment shunts the
diode when connected to the test terminals. As long as the voltage across the
terminals is kept below the diode threshold voltage, no current passes
through the diode. To ensure that there is no current leaking through the diode
when making a test reading or when connecting an indicating meter, the
resistance of the test connection or meter should not exceed 10 Ω.
BOARD CHECKOUTNOTE
Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.
Figure 5-1. Header Board
Connections.
SENSING MODULE
CHECKOUT
NOTE
The Rosemount 1154 Series H Pressure Transmitter contains electronic
circuit boards which may be static sensitive.
You can easily check the printed circuit boards (5 and 6) for a malfunction by
substituting spare boards into the circuit. If this procedure turns up a
malfunctioning board, return the defective board to Rosemount Nuclear
Instruments, Inc. for replacement. Because of parts traceability, qualification
becomes the responsibility of the customer in the event of unauthorized board
repairs.
(Component Side Up)
1151-0180A
NOTE
Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.
5-2
The sensing module is contained within the module shroud (8) and is not field
repairable. If no defect such as loss of fill fluid is observed, check the sensing
module in the following manner.
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
1.Disengage the header assembly board (4) as described in Step 4 of
the electrical housing disassembly procedure on page 5-3. You need
not remove the module shroud from the electrical housing for
checkout.
2.Jump connections 1 and 2 on the header assembly board (see Figure
5-1).
3.Using a low-voltage ohmmeter, check the resistance between the
jumper wire and the module shroud. This resistance should be
greater than 10 MΩ. Remove the jumper wire.
4.Jump connections 3 and 4 on the header assembly board and repeat
Step 3 (see Figure 5-1).
NOTE
The above procedure does not completely test the sensing module. If circuit
board replacement does not correct the abnormal condition, and no other
problems are obvious, the transmitter should be replaced.
DISASSEMBLY
PROCEDURE
Electrical Housing
Disassembly
NOTE
Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.
NOTE
The Rosemount 1154 Series H Pressure Transmitter contains electronic
circuit boards which may be static sensitive.
1.The signal terminals and the test terminals are accessible by
unscrewing the cover (1) on the terminal side. This compartment is
identified as “Terminal Side” on the nameplate. The terminals are
permanently attached to the housing and must not be removed.
2.Circuit boards are located in a separate compartment identified as
“Circuit Side” on the nameplate. Remove power from the transmitter
before removing the circuit cover. Unscrew the cover (1) on the circuit
side to access the circuit boards. A special cover wrench (Rosemount
Part Number 01153-0382-0001) is available from Rosemount to
remove and replace the housing covers.
3.Unplug the amplifier board (6) after removing 3 holding screws (7).
4.The header assembly board (4) is permanently attached to the sensor
module within the module shroud (8) and contains the
temperature-compensating resistors. Carefully pull this board off the
bayonet pins and rotate the board 180° about the axis formed by the
connecting leads. This allows access to the calibration board (5).
5.Disconnect the calibration board (5) by aligning the zero and span
adjust screws so that their slots are perpendicular to the board.
Remove the board by inserting a 6-32 screw in the rivnut on the
board, and carefully pull the board off the bayonet pins.
6.If replacement of the zero and span adjustment screws (10) is
necessary, remove the nameplate (13) and detach the snap rings (12)
inside the housing.
5-3
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
Removing Module
Shroud From Electrical
Housing
REASSEMBLY
PROCEDURE
1.Remove amplifier board and calibration board as described in the
Electrical Housing Disassembly Section above.
2.Loosen the lock nut (9).
3.Unscrew the module shroud (8) from the electronics housing,
simultaneously turning the header board and leads to prevent them
from being twisted or damaged. The threaded connection has a
sealing compound on it and must be broken loose. Then carefully pull
the header assembly board (4) through the hole.
4.The module shroud (8) containing the sensor module is a complete
assembly and cannot be further disassembled.
NOTE
Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.
NOTE
The Rosemount 1154 Series H Pressure Transmitter contains electronic
circuit boards which may be static sensitive.
Preliminary1.Replace the cover O-rings (2) whenever you remove a cover. Clean
the sealing areas with alcohol, if necessary, and lightly grease the
O-ring with Dow Corning
Part Number 01153-0248-0001 or P/N 01153-0053-0001). Spray the
inside threads of the electronics covers with cover lubricant
(Rosemount P/N 01153-0333-0001 or equivalent) if necessary. If
covers are already sufficiently lubricated, do not spray.
2.Verify that the circuit boards are clean.
3.Verify that the bayonet pins on the connection board are clean.
4.If you remove the module shroud, clean the thread sealant from
module shroud threads, lock nut, and electronics housing threads
with a wire brush.
®
55 Silicone O-ring Grease (Rosemount
Connecting Electrical
Housing to Module
Shroud
5-4
1.Run the lock nut down to the base of the sensor threads.
3
2.Apply a heavy, continuous bead (about
580-PST sealant (Rosemount Part Number 01153-0329-0001)
around the top module shroud threads.
3.Insert header assembly board (4) through the hole in the bottom of
the electronics housing.
4.Screw module shroud (8) into electrical housing (3) making sure that
five full threads are engaged. Be careful not to damage or twist the
sensor leads. Turn the header board to avoid twisting wires.
5.Align the module shroud with the high and low pressure sides
oriented per Figure 2-5 on page 2-8. Alternately, tighten the module
shroud one-half turn further to reverse the orientation of the module
shroud about the electronics housing.
6.Tighten the lock nut (9) to 35 ft-lb (48 N-m) torque.
7.Wipe off excess sealant.
8.Place the assembled unit in an oven at 200 ± 5 °F (93 ± 3 °C) for
12 hours to cure the sealant.
/8-in. wide) of Loctite®
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Electrical Housing
Reassembly
1.Replace the zero and span adjustment screw O-rings (11) whenever
you remove the zero or span adjustment screws (10). Lightly grease
the O-rings with Dow Corning 55 Silicone O-ring Grease, (Rosemount
P/N 01153-0248-0001). Reinstall adjustment screws and secure with
snap rings (12).
2.Align the zero and span adjustment screws with the potentiometer
stems on the calibration board (5) and push the calibration board onto
the bayonet pins.
3.Slide the header assembly board (4) onto the bayonet pins with the
component side toward the pins. Slide any excess wire behind the
calibration board, taking care to avoid kinks.
4.Push the amplifier board (6) onto the bayonet pins and secure with
the holding screws (7). Use nominal torque of 10 in-lb
(1.1 N-m).
5.Carefully replace the cover and tighten to 16.5 ft-lb (22.4 N-m)
(see “Preliminary” on page 5-4).
6.Replace the nameplate (13) and attach with two nameplate screws
(14).
POST-ASSEMBLY TESTS1.Calibrate transmitter per the calibration section of this manual.
2.Conduct nuclear cleaning to 1 ppm chloride content of transmitter
“wetted parts.”
5-5
Rosemount 1154 Series H
Figure 5-2. Exploded View Parts Drawing, Rosemount 1154 Series H.
1
Reference Manual
00809-0100-4631, Rev BA
April 2007
2
3
5
13
4
12
10
11
9
14
6
7
2
1
8
Table 5-1. Rosemount 1154 Series H Parts List.
PartDescriptionPartDescription
1Electronics Cover 8Module Shroud
2O-Ring for Electronics Cover9Module Shroud Lock Nut
3Electronics Housing10Zero and Span Adjustment Screws
4Header Assembly Board11O-Ring for Adjustment Screw
5Calibration Board12Snap Rings
6Amplifier Board13Nameplate
7Holding Screws14Nameplate Screws
Table 5-2. Torque Reference.
Items(s) to be TorquedTorque ValueTol er ance
Bracket to Mounting Panel Bolts
Transmitter to Bracket Bolts
Swagelok Process Fittings
Drain/Vent Valves
Covers
Module Neck Lock Nut
Conduit Fitting
Terminal Block Screws
Amplifier Board Screws
High OutputPrimary ElementCheck for restrictions at primary element, improper installation or poor condition. Note any changes
Impulse PipingCheck for leaks or blockage.
Transmitter ElectronicsMake sure that post connectors and the sensor connections are clean.
Transmitter Electronics
Failure
Sensing ModuleSee sensing module checkout section. The sensing module is not field repairable and must be
Power SupplyCheck the power supply output voltage at the transmitter.
Low Output or
No Output
Primary ElementCheck for restrictions at primary element, improper installation or poor condition. Note any changes
Loop Wiring
in process fluid properties.
Ensure that blocking valves are fully open.
Check for entrapped gas in liquid lines and for liquid in dry lines.
Ensure that density of fluid in impulse lines is unchanged.
Check for sediment in the transmitter process flanges.
If the electronics are still suspect, substitute new electronics.
Determine faulty circuit board by trying spare boards. Replace faulty circuit board.
replaced if found to be defective. See “Disassembly Procedure” for instructions on disassembly. Check
for obvious defects, such as punctured isolating diaphragm or fill fluid loss, and contact Rosemount
Nuclear Instruments, Inc. at 952-949-5210.
in process fluid properties.
Do not use over 100 volts to check the loop, or damage to the transmitter electronics
may result.
Check for adequate voltage to the transmitter.
Check the milliamp rating of the power supply against the total current being drawn for all transmitters
being powered.
Check for shorts and multiple grounds.
Check for proper polarity at the signal terminal.
Check loop impedance.
Check wire insulation to detect possible shorts to ground.
Impulse PipingEnsure that the pressure connection is correct.
Transmitter Electronics
Connections
Test Diode FailureReplace electronics housing.
Transmitter Electronics
Failure
Sensing ModuleSee sensing module checkout section. The sensing module is not field repairable and must be
Power SupplyCheck the power supply output voltage at transmitter.
Check for leaks or blockage.
Check for entrapped gas in liquid lines.
Check for sediment in the transmitter process flange.
Ensure that blocking valves are fully open and that bypass valves are tightly closed.
Ensure that density of the fluid or other fluid properties in the impulse piping are unchanged.
Ensure that calibration adjustments are in allowable range.
Check for shorts in sensor leads.
Make sure post connectors are clean, and check the sensor connections.
If the electronics are still suspect, substitute new electronics.
replaced if found to be defective. See “Disassembly procedure” for instructions on disassembly. Check
for obvious defects, such as punctured isolating diaphragm or fill fluid loss, and contact Rosemount
Nuclear Instruments, Inc. at 952-949-5210.
Continued On Next Page
5-7
Rosemount 1154 Series H
SymptomPotential SourceCorrective Action
Erratic OutputLoop Wiring
Do not use over 100 volts to check the loop, or damage to the transmitter electronics
may result.
Check for adequate voltage to the transmitter.
Check for intermittent shorts, open circuits, and multiple grounds.
Impulse Piping and
Process Connections
Transmitter ElectronicsCheck for intermittent shorts or open circuits.
Transmitter Electronics
Failure
Power SupplyCheck power supply output voltage.
Check for entrapped gas in liquid lines and for liquid in dry lines.
Make sure that post connectors and the sensor connectors are clean and properly connected.
Determine faulty board by trying spare boards. Replace faulty circuit board.
Qualified to IEEE Std 323-1974 and IEEE Std 344-1975 per Rosemount
Report D8700096.
Radiation
Accuracy within ±(0.25% of upper range limit + 0.75% of span) during first 30
minutes; ±(0.5% upper range limit + 1.0% span) thereafter up to 55 megarads
total integrated dosage (TID); ±(0.75% upper range limit + 1.0% span) after
110 megarads TID gamma radiation exposure
Seismic
Accuracy within ±0.5% of upper range limit during and after a disturbance
defined by a required response spectrum with a horizontal ZPA of 8.5 g and a
vertical ZPA of 5.2 g
Steam Pressure/Temperatures
Accuracy within ±(1.0% of upper range limit + 1.0% of span) for range codes
4–8; ±(2.0% upper range limit + 0.5% span) for range code 9 during and after
sequential exposure to steam at the following temperatures and pressures,
concurrent with chemical spray for the first 24 hours:
420 °F (215.6 °C), 85 psig for 3 minutes
350 °F (176.6 °C), 85 psig for 7 minutes
320 °F (160 °C), 75 psig for 8 hours
265 °F (129.4 °C), 24 psig for 56 hours
www.emersonprocess.com/rosemount/nuclear
Chemical Spray
Chemical Spray composition is 0.28 molar boric acid, 0.064 molar sodium
thiosulfate, and sodium hydroxide to make an initial pH of 11.0 and a
subsequent pH ranging from 8.5 to 11.0. Chemical spray is sprayed at a rate
of 0.25 gal/min/ft
Post DBE Operation
Accuracy at reference conditions shall be within ±2.5% of upper range limit for
one year following DBE.
Quality Assurance Program
In accordance with NQA-1, 10CFR50 Appendix B, and ISO 9001:2000
2
.
Rosemount 1154 Series H
Nuclear Cleaning
To 1 ppm maximum chloride content
Hydrostatic Testing
To 150% of maximum working pressure or 2,000 psi (13.8 MPa), whichever is
greater
Traceability
In accordance with NQA-1 and 10CFR50 Appendix B; chemical and physical
material certification of pressure-retaining parts.
Qualified Life
Dependent on continuous ambient temperature at the installation site,
illustrated in Figure 6-1 on page 6-2. Replacement of amplifier and calibration
circuit boards at the end of their qualified life permits extension of the
transmitter’s qualified life to the module’s qualified life. See Rosemount
Report D8700096.
Reference Manual
00809-0100-4631, Rev BA
April 2007
Figure 6-1. Qualified Life vs.
Ambient Temperature.
PERFORMANCE
SPECIFICATIONS
Module
Qualified Life
Electronics
Qualified Life
Time (Years)
Temperature (°F)
Based on zero-based ranges under reference conditions
Accuracy
1154-0293A
6-2
±0.25% of calibrated span; includes combined effects of linearity, hysteresis,
and repeatability
Deadband
None
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Drift
±0.2% of upper range limit for 30 months
Temperature Effect
Range Codes 4–8:
±(0.15% upper range limit + 0.35% span) per 50 °F (27.8 °C) ambient
temperature change between 40 °F (4.4 °C) and 130 °F (54.4 °C)
Range Code 9:
±(0.25% of upper range limit + 0.5% span) per 50 °F (27.8 °C) ambient
temperature change between 40 °F (4.4 °C) and 130 °F (54.4 °C)
All Range Codes:
±(0.75% of upper range limit + 0.5% span) per 100 °F (55.6 °C) ambient
temperature change between 40 °F (4.4 °C) and 200 °F
(93.3 °C)
Overpressure Effect
Rosemount 1154DH:
Maximum zero shift after 2,000 psi (13.8 MPa) overpressure:
Range CodeOverpressure Effect
4±0.25% of upper range limit
5±1.0% of upper range limit
6–7±3.0% of upper range limit
8±6.0% of upper range limit
Rosemount 1154HH:
Maximum zero shift after 3,000 psi (20.68 MPa) overpressure:
Range CodeOverpressure Effect
4±1.0% of upper range limit
5±2.0% of upper range limit
6–7±5.0% of upper range limit
Rosemount 1154SH:
Maximum zero shift after 4,500 psi (31.0 MPa) overpressure:
Range CodeOverpressure Effect
9±0.5% of upper range limit
Static Pressure Zero Effect
Rosemount 1154DH:
Per 1,000 psi (6.89 MPa):
Range CodeStatic Pressure Zero Effect
4–5±0.2% of upper range limit
6–8±0.5% of upper range limit
Rosemount 1154HH:
Per 1,000 psi (6.89 MPa):
Range CodeStatic Pressure Zero Effect
4–7±0.66% of upper range limit
6-3
Rosemount 1154 Series H
Static Pressure Span Effect
Effect is systematic and can be calibrated out for a particular pressure before
installation; correction uncertainty: ±0.5% of input
reading/1,000 psi (6.89 MPa)
Power Supply Effect
Less than 0.005% of output span/volt
Load Effect
No load effect other than the change in voltage supplied to the transmitter
Mounting Position Effect
Reference Manual
00809-0100-4631, Rev BA
April 2007
FUNCTIONAL
SPECIFICATIONS
No span effect; zero shift of up to 1.5 inH
O (372 Pa) which can be calibrated
2
out
Response Time
Fixed time constant (63%) at 100 °F (37.8 °C) as follows:
Range CodeResponse Time
40.5 seconds or less
all others0.2 seconds or less
Adjustable damping is available through a special
N option.
Service
Liquid, gas, or vapor
Output
4–20 mA dc
Power Supply
Design limits as shown in Figure 6-2. See Rosemount Report D8700096 for
additional detail.
Figure 6-2. Load Limitations.
6-4
1575
Load Resistance ()
1825
1500
1000
500
4–20 mA dc
Qualified
Region
Design
Region
0
0
12
13.5
Power Supply (V dc)
35
40
45
50
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
Span and Zero
Continuously adjustable externally
Zero Elevation and Suppression
Maximum zero elevation: 600% of calibrated span
Maximum zero suppression: 500% of calibrated span
Zero elevation and suppression must be such that neither the calibrated span
nor the upper or lower range value exceeds 100% of the upper range limit.
Temperature Limits
Normal operating design limits: 40 to 200 °F (4.4 to 93.3 °C)
Qualified storage limits: –40 to 120 °F (–40.0 to 48.9 °C)
Humidity Limits
0 to 100% relative humidity (NEMA 4X)
Volumetric Displacement
3
Less than 0.01 in
(0.16 cm3)
Turn-on Ti m e
Two seconds maximum. No warm-up required.
Pressure Ranges
Rosemount 1154DH and 1154HH:
Range CodePressure Range
40–25 to 0–150 inH2O (0–6.22 to 0–37.3 kPa)
50–125 to 0–750 inH2O
60–17 to 0–100 psi (0–0.12 to 0–0.69 MPa)
70–50 to 0–300 psi (0–0.34 to 0–2.07 MPa)
80–170 to 0–1,000 psi (Rosemount 1154 D only)
(0–31.08 to 0–186.4 kPa)
(0–1.17 to 0–6.89 MPa)
Rosemount 1154SH:
Range CodePressure Range
90–500 to 0–3,000 psi (0–3.45 to 0–20.68 MPa)
Maximum Working Pressure
Rosemount 1154DH and 1154HH:
Static Pressure Limit
Rosemount 1154SH:
Upper range limit
Static Pressure and Overpressure Limits
Rosemount 1154DH:
0.5 psia to 2,000 psig (3.4 kPa abs to 13.8 MPa) maximum rated static
pressure for operation within specifications; overpressure limit is 2,000
psig (13.8 MPa) on either side without damage to the transmitter
6-5
Rosemount 1154 Series H
Rosemount 1154HH:
0.5 psia to 3,000 psig (3.4 kPa abs to 20.7 MPa) maximum rated static
pressure for operation within specifications; overpressure limit is 3,000
psig (20.7 MPa) on either side without damage to the transmitter
Overpressure Limits
Rosemount 1154SH:
Operates within specification from 0.5 psia (3.4 kPa abs) to upper range
limit; overpressure limit is 4,500 psig (31.0 MPa) for Range Code 9,
without damage to the transmitter
Reference Manual
00809-0100-4631, Rev BA
April 2007
PHYSICAL
SPECIFICATIONS
Materials of Construction
Isolating Diaphragms:
316L SST
Drain/Vent Valves:
316 SST
Process Flanges:
CF-8M (cast version of 316 SST)
O-rings:
316L SST
Electronics Housing O-rings:
Ethylene propylene
Fill Fluid:
Silicone oil
Flange Bolts:
Plated alloy steel, per ASTM A-540
Electronics Housing:
316 SST
Module Shroud:
304L SST
6-6
Module Shroud Potting:
Silicone RTV
Mounting Bracket:
316L SST
Mounting Bolts:
SAE J429 Carbon steel, Grade 2 or Grade 5
Weight
26.6 lb (12.1 kg) including mounting bracket
Electrical Connections
1
/2–14 NPT conduit with screw terminals
Process Connections
3
/8 in. Swagelok compression fitting, 316 SST (¼–18 NPT optional)
Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
ORDERING INFORMATION
Table 6-1. Transmitter Design Specifications.
ModelProduct Description
115 4Alphaline Pressure Transmitters for Nuclear Applications (qualified per IEEE Std 323-1974 and IEEE Std 344-1975)
HTransmitter Enclosed in a Stainless Steel Module Shroud
Pressure Ranges at 68 °F
Code
40–25 to 0–150 inH2O
50–125 to 0–750 inH2O
60–17 to 0–100 psi
70–50 to 0–300 psi
80–170 to 0–1,000 psi
9N/AN/A0–500 to 3,000 psi
CodeOutput
(1)
R
Code
AWelded 3/8 in. Swagelok compression fitting process connection and welded drain/vent valve
(2)
B
(2)
C
HWelded 3/8 in. Swagelok compression fittings on both process connection and drain/vent connection
Typical Model Number: 1154DH 4 R A
(1) The Rosemount 1154 Series H with the R Output Code Electronics is also available with adjustable damping. Specify adjustable damping by adding “N0037”
to the end of the complete model number, for example: 1154DH4RAN0037.
(2) Customer assumes responsibility for qualifying connection interfaces on these options. Contact Rosemount Nuclear Instruments, Inc. for details.
Rosemount 1154DH (Differential)
(0–6.22 to 0–37.3 kPa)
(0–31.08 to 0–186.4 kPa)
(0–0.12 to 0–0.69 MPa)
(0–0.34 to 0–2.07 MPa)
(0–1.17 to 0–6.89 MPa)
Standard 4–20 mA
Flange Option
¼–18 NPT process connection and welded drain/vent valve
¼–18 NPT process connection and drain hole (drain/vent valve not supplied)
Panel Mounting Bracket with Bolts
Universal Mounting Bracket with Bolts
Bolts and Washers for Bracket (Pkg of 4)
Conduit Elbow (M22)
Conduit Elbow (½–14 NPT)
(1) Rosemount recommends one spare part or kit for every 25 transmitters in Category “A,” one spare part or kit for every 50 transmitters in Category
“B,” and one spare part or kit for every 5 transmitters in Category “C.”
(2) For use with existing N0026 transmitter only.
NOTE: Rosemount 1154 Series H Spare Parts not hydrostatic tested nor nuclear cleaned.
Store all spare transmitters and spare component parts in accordance with
ANSI N45.2.2 level B.
Qualified transmitters, spare circuit boards, spare O-rings: The qualified life
(as defined in Qualification Test Report D8700096) plus the shelf life is equal
to the typical design life of the plant (40 years) when the ambient storage
temperature is below 90 °F.
Lubricants and sealant: The date of the end of shelf life (use by date) is
provided with the lubricants and/or sealant, at the time of shipment. The
product has a minimum of six months shelf life at the time of shipment.
All other parts: Shelf life is not applicable.
IMPORTANT NOTICEBecause of the nuclear use intended for these parts, there are factors to
consider concerning maintenance of qualification and traceability during
on-site instrument repair. Rosemount Nuclear Instruments, Inc. rigidly
controlled the original assembly of the instrument to ensure that the
specifications were met. Since we are not installing the replacement parts in
the instruments, Rosemount Nuclear Instruments, Inc. is unable to ensure
that the specifications are being satisfied. This responsibility is shifted to the
user. The integrity of the instrument as originally assembled is broken.
Replacing parts has ramifications under 10CFR21, for which the user is
responsible. These same regulations additionally mandate a component
traceability program, which the user must undertake for the replacement
parts. In view of this, and in order to maintain the qualification of the product,
the user must ensure that all replacement parts are installed in accordance
with the Rosemount Nuclear Instruments, Inc. approved installation and
calibration procedures herein.
1.Rosemount 1154 Series H spare parts are not hydrostatic tested nor
nuclear cleaned.
2.Part numbers shown may differ from those currently supplied. The
part numbers shown are current at the time of printing of this manual,
but may be revised in the future. Parts provided are compatible and
interchangeable with those listed on your order as to the form, fit, and
function of the part required. Please adjust your needs accordingly.
The Emerson logo is a trade mark and service mark of Emerson Electric Co.
Rosemount, the Rosemount logotype, and Alphaline are registered trademarks of Rosemount, Inc.
δ−Cell is a trademarks of Rosemount, Inc.
Swagelok is a registered trademark of Swagelok Co.
Emerson Process Management
Rosemount Nuclear Instruments, Inc.
8200 Market Boulevard
Chanhassen, MN 55317 USA
T (952) 949-5210
F (952) 949-5201
Rosemount Nuclear Instruments, Inc. satisfies all obligations coming from
legislation to harmonize product requirements in the European Union.
¢00809-XXXX-XXXX ¤
www.emersonprocess.com/rosemount/nuclear
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.