Emerson Rosemount 1154DH, Rosemount 1154HH, Rosemount 1154SH Reference Manual

Reference Manual
00809-0100-4631, Rev BA April 2007

Rosemount 1154 Series H

Alphaline® Nuclear Pressure Transmitter

www.emersonprocess.com/rosemount/nuclear
IMPORTANT NOTICE – ERRATA
Affected
Pages
Effect.
Date
1
6-6
Process Flagel – CF3M (Cast version of 316L SST)
optional)
10/21/09
2
3-6
Change the first paragraph on the page to read as follows:
1154DH, 1154HH, and 1154SH) for the applicable part numbers.”
4/13/12
3
6-1
Steam Pressure/Temperature specification is updated as follows:
specification is ±2% of upper range limit.
4/13/12
4
6-8
Table 6-2, “Parts List Rosemount 1154DH, 1154HH, and 1154SH”, is updated as
In all locations, replaced “01154-0001-000 6” wit h “01154-0153-0002”
4/13/12
5
6-11
Table 6-2, “Parts List Rosemount 1154DH, 1154HH, and 1154SH”, in all locations the
“01154-0153-0002” is replaced by “01154-0263-0002”
6/9/17
Model 1153 Series H Product Manual 00809-0100-4631 Rev BA (April 2007)
No.
Description of Change
Drain/Vent Valves – 316L SST Process Connections – 3/8-inch Swagelok compression fitting, 316L SST (1/4-18 NPT
“Damping electronics are available as an option. Transmitters with standard electronics can be retrofitted with the adjustable damping feature by changing out both the amplifier board and the calibration board. Please reference Table 6-2 (Parts List Rosemount
Steam Pressure/Temperature Accuracy as specified below, during and after sequential exposure to steam at the
following temperatures and pressures, concurrent with chemical spray for the first 24 hours.
420 °F (215.6 °C), 85 psig for 3 minutes 350 °F (176.6 °C), 85 psig for 7 minutes 320 °F (160 °C), 75 psig for 8 hours 265 °F (129.4 °C), 24 psig for 56 hours
Range codes 4-8: Accur ac y within ±(1.0% of upper range limit + 1.0% of span)
(1)
Range code 9: Accur acy within ±(2.0% of upper range limit + 0.5% of span)
Note (1) For Range codes 4-8, if pressure applied at the 4mA point is greater than 50% of the upper range limit, the
follows: Amplifier Circuit Board, Output Code R:
In all locations, replaced “01154-0001-000 5” with “01154-0153-0001” Amplifier Circuit Board with Damping, Output Code R:
In all locations, replaced “01154-0021-000 4” with “01154-0156-0001” Amplifier Circuit Board, Output Code R, N0026:
following part numbers are updated: Amplifier Circuit Board, Output Code R:
“01154-0153-0001” is replaced by “01154-0263-0001” Amplifier Circuit Board with Damping, Output Code R:
“01154-0156-0001” is replaced by “01154-0266-0001” Amplifier Circuit Board, Output Code R, N0026:
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Rosemount 1154 Series H Alphaline
®
Pressure
Transmitters
NOTICE
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.
For equipment service needs outside the United States, contact the nearest Rosemount representative.
Within the United States, the North American Response Center is at your service 24 hours a day, and is a single-point contact for all Rosemount equipment service needs. If at any time you are not sure what to do, you have a question about using the product, or you have a service or support request, call the center toll free at 1-800-654-RSMT (7768). This contact is your fastest link to quick and complete answers about any Rosemount group, product, or service.
Alphaline, Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc. -Cell is a trademark of Rosemount Inc. D.C. 55 is a registered trademarks of Dow Corning. Loctite is a registered trademark of Henkel KGaA Corporation. Grafoil is a trademark of Union Carbide Corp. Swagelok is a registered trademark of Crawford Fitting Company. Lubri-Bond is a registered trademark of E/M Corporation.
Cover Photo: 1153-001AB
Rosemount Nuclear Instruments, Inc. satisfies all obligations coming from legislation to harmonize product requirements in the European Union.
www.emersonprocess.com/rosemount/nuclear
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007
Rosemount Nuclear Instruments, Inc. Warranty and Limitations of Remedy
The warranty and limitations of remedy applicable to this Rosemount equipment are as stated on the reverse of the current Rosemount quotation and customer acknowledgment forms.
RETURN OF MATERIAL
Authorization for return is required from Rosemount Nuclear Instruments, Inc. prior to shipment. Contact Rosemount Nuclear Instruments, Inc. (1-952-949-5210) for details on obtaining Return Material Authorization (RMA). Rosemount Nuclear Instruments will not accept any returned material without a Returned Material Authorization. Material returned without authorization is subject to return to customer.
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Nuclear Instruments, Inc. 8200 Market Blvd. Chanhassen, MN 55317 USA
IMPORTANT
The Rosemount 1154 Series H Pressure Transmitter is designed for Nuclear Class IE usage, has been tested per IEEE Std 323-1974, IEEE 344-1975 as defined in the Rosemount Report D8700096 and is manufactured to the requirements of NQA-1; 10CFR50, Appendix B quality assurance programs; and 10CFR Part 21. During qualification testing, interfaces were defined between the transmitter and its environment that are essential to meeting IEEE Std 323-1974 requirements. To ensure compliance with 10CFR Part 21, the transmitter must comply with the requirements herein and in Report D8700096 throughout its installation, operation, and maintenance. It is incumbent upon the user to ensure that the Rosemount Nuclear Instruments, Inc. component traceability program where applicable is continued throughout the qualified life of the transmitter.
In order to maintain the qualified status of the transmitter, the essential environmental interfaces must not be compromised. Performance of any operations on the transmitter other than those specifically authorized in this manual has the potential for compromising an essential environmental interface.
Where the manual uses the terms requirements, mandatory, must, or required, the instructions so referenced must be carefully followed. Rosemount Nuclear Instruments, Inc. expressly disclaims all responsibility and liability for transmitters for which the foregoing has not been complied with by the user.
www.emersonprocess.com/rosemount/nuclear
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Revisions

Changes From June 1999 to April 2007
Page (Old) Page (New) Changes
Cover Cover Document revision date change from June 1999 to April 2007, rev. from AA to BA.
Inside cover
i, ii, 5-7 and back cover
3-6, and 6-8 3-6, and 6-8 Include errata sheet information on circuit board number changes:
Throughout Throughout References to Fisher-Rosemount were changed to Emerson Process Management.
i, back
cover
-- Cover, i,
2-1, 3-1,
4-1, 5-1, 6-1
2-2, 6-7 2-2, 6-7 Removed word “process” from sentence indicating user assumes responsibility for qualifying the connection
2-2 2-3 Updated reference to Swagelok catalog and added web address, removed street address. 2-4 2-4 Removed reference to Model 353C. 2-4 2-5 Rearranged wording on shielded cable.
2-6 and 2-7 2-7 and 2-8 Added word ‘nominal’ to Notes in drawings. Changed significant digits to conform to standard.
2-7 2-8 Changed height from 9.0 max (229) to 9.3 max (236). Changed shroud width from 3.40 (86.4) to 3.48 (88.3). 6-1 6-1 Changed ISO 9001 to ISO 9001:2000.
-- Back cover Added trademark and registration information.
ii, 5-7 and
back cover
Cover, i, ii
and back
cover
and back
page
2-1, 3-1,
4-1, 5-1, 6-1
Include errata sheet information on address and phone number.
Replaced amplifier circuit board, output code R P/N 01154-0001-0001 with 01154-0001-0005. Replaced amplifier circuit board with damping, output code R P/N 01154-0021-0002 with 01154-0021-0004. Replaced amplifier circuit board for N0026, output code R P/N 01154-0001-0002 with 01154-0001-0006.
Web address changed from www.rosemount.com to www.emersonprocess.com/rosemount/nuclear.
Added reference to European Union product requirement (CE).
Added table of contents to each section.
interface.
NOTE
The above Revision Status list summarizes the changes made. Please refer to both manuals for complete comparison details.
www.emersonprocess.com/rosemount/nuclear
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
1-2
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Table of Contents

SECTION 1 Introduction

SECTION 2 Installation

SECTION 3 Calibration

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
About the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
General Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mechanical Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Process Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Conduit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Electrical Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Conduit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Span Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Zero Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Zero and Span Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Material. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Linearity Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Damping Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Correction For High Line Pressure
(Rosemount 1154DH and 1154HH only). . . . . . . . . . . . . . . . . . . . 3-6
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8

SECTION 4 Operations

SECTION 5 Maintenance and Troubleshooting

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Transmitter Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
The -Cell
Demodulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Linearity Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Oscillator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Voltage Regulator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Zero and Span Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Current Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Current Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Reverse Polarity Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Test Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Board Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
TOC-1
Rosemount 1154 Series H
Sensing Module Checkout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Disassembly Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Electrical Housing Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Removing Module Shroud From Electrical Housing. . . . . . . . . . . . 5-4
Reassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Preliminary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Connecting Electrical Housing to Module Shroud . . . . . . . . . . . . . 5-4
Electrical Housing Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Post-Assembly Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Reference Manual
00809-0100-4631, Rev BA
April 2007

SECTION 6 Specifications and Reference Data

Nuclear Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Seismic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Steam Pressure/Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Chemical Spray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Post DBE Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Quality Assurance Program . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Nuclear Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Hydrostatic Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Traceability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Qualified Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Temperature Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Overpressure Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Static Pressure Zero Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Static Pressure Span Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Power Supply Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Load Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Mounting Position Effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Response Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Span and Zero. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Zero Elevation and Suppression . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Temperature Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Humidity Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Volumetric Displacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Turn-on Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Pressure Ranges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Maximum Working Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Static Pressure and Overpressure Limits . . . . . . . . . . . . . . . . . 6-5
Overpressure Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
TOC-2
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Materials of Construction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Electrical Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Process Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Spare Parts Shelf Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Important Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
TOC-3
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
TOC-4
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Section 1 Introduction

OVERVIEW This manual is designed to assist in installing, operating, and maintaining the

Rosemount 1154 Series H Alphaline® Nuclear Pressure Transmitter. The manual is organized into the following sections:
Section 2: Installation
Provides general, mechanical, and electrical installation considerations to guide you through a safe and effective transmitter installation.
Section 3: Calibration
Provides transmitter calibration procedures.
Section 4: Operation
Provides descriptions of how the transmitter operates.
Section 5: Maintenance and Troubleshooting
Provides basic hardware troubleshooting considerations including sensing module checkout, disassembly and reassembly procedures, and post-assembly tests.

ABOUT THE TRANSMITTER

Section 6: Specifications and Reference Data
Provides nuclear, performance, functional, and physical transmitter specifications; also includes ordering information, and a list of spare parts.
Rosemount 1154 Series H Alphaline Pressure Transmitters are designed for precise pressure measurements in nuclear applications requiring reliable performance and safety over a specified qualified life. These transmitters were generically tested to IEEE Std 323-1974 and IEEE Std 344-1975 per the Qualification Test Report D8700096. The Rosemount 1154 Series H has been qualification tested to environments typical of pressurized water reactors (PWR) under accident conditions. Stringent quality control during the manufacturing process includes traceability of pressure retaining parts, special nuclear cleaning, and hydrostatic testing.
Rosemount 1154 Series H Transmitters are of a design unique to Class 1E nuclear service while retaining the working concept and design parameters of the Rosemount 1151 that has become a standard of reliable service. Units are available in sealed reference (S), differential (D), and high-line differential (H) configurations, with a variety of pressure range options (see Table 6-1 on page 6-7). Figure 2-5 on page 2-8 shows dimensional drawings of the transmitters.
www.emersonprocess.com/rosemount/nuclear
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
1-2
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Section 2 Installation

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
General Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1
Mechanical Considerations . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-4
Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-6

OVERVIEW This section contains information and instructions regarding the following

installation-related information:
General Considerations
Mechanical Considerations Process Connections Conduit
Electrical Considerations Mechanical Electrical

GENERAL CONSIDERATIONS

The quality and accuracy of flow, level, or pressure measurement depends largely on the proper installation of the transmitter and its associated impulse piping and valves. For flow measurement, proper installation of the primary measuring element is also critical to the accuracy of the measurement. Transmitter installation should minimize the effects of temperature gradients and temperature fluctuations, and avoid vibration and shock during normal operation. Take care when designing the measurement to minimize the error caused by incorrect installation. The temperature of the transmitter ambient environment affects the qualified life of the transmitter. (See Figure 2-1.)
www.emersonprocess.com/rosemount/nuclear
Rosemount 1154 Series H

Figure 2-1. Qualified Life vs. Ambient Temperature.

Reference Manual
00809-0100-4631, Rev BA
April 2007
Module Qualified Life
Electronics
Qualified Life
Time (Years)

MECHANICAL CONSIDERATIONS

Ambient Temperature (°F)
This section contains information you should consider when preparing to mount the transmitter. Read this section carefully before proceeding to the
1154-0293A
mechanical installation procedure.
Mount the Rosemount 1154 Series H transmitter to a rigid support (i.e., one with a fundamental mechanical resonant frequency of 40 Hz or greater). A mounting bracket included with the transmitter facilitates panel mounting. Figure 2-4 on page 2-7 shows the qualified mounting configurations. The transmitter was seismic tested and qualified with the bracket mounted with
3
/8 in. diameter bolts. Orientation with respect to gravity is not critical to
four qualification. However, if the transmitter is mounted with the flanges in a horizontal position, zero the transmitter to cancel the liquid head effect caused by the difference in height of the process connections.
If the transmitter is mounted to a non rigid panel, ensure that seismic input to the mounting bracket does not exceed qualification levels given in Rosemount Report D8700096.

Process Connections Process tubing installation must prevent any added mechanical stress on the

transmitter under seismic disturbances. This may be done by using stress-relief loops in the process tubing or by separately supporting the process tubing near the transmitter.
3
The process connections to the transmitter flanges were qualified with tubing using Swagelok
®
compression fittings. For options using 1/4–18 NPT
connections, the user assumes responsibility for qualifying the interface.
/8 in.
2-2
Transmitters with Flange Options A or H are shipped with Swagelok fittings for process connections. Included are front ferrule, rear ferrule, and nut. Ensure the fittings are placed on the tubing with the orientation and relative position shown in Detail A in Figure 2-5 on page 2-8.
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Process tubing used is 3/8 in. outside diameter, and of suitable thickness for the pressure involved.
The Swagelok tube fittings are shipped completely assembled for immediate use. Do not disassemble them before use; doing so may contaminate the fittings and result in leaks. Insert the tubing into the Swagelok tube fitting, making sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger tight. Tighten the nut one-and-one-quarter turns past finger tight to prepare the transmitter for use. Do not overtighten.
The connections can be loosened and re-tightened 20–30 times without compromising the leak-proof seal. To reconnect, insert the tubing with pre-swaged ferrules into the fitting until the front ferrule sits in the fitting. Tighten the nut by hand, then rotate one-quarter turn more or to the original one-and-one-quarter tight position. Then snug it slightly with a wrench. For more information regarding the use of Swagelok tube fittings, refer to:
Fittings Catalog MS-01-140 “Gaugeable Tube Fittings and Adapter Fittings” www.swagelok.com
If the drain/vent valves must be opened to bleed process lines, torque them to
1
/2 ft-lb (10 N-m) when closing.
7
Proper location of the transmitter with respect to the process tubing depends on various process parameters. When determining the best location, consider the following:
Keep hot or corrosive fluids from contacting the transmitter.
Prevent sediment from depositing in the impulse tubing.
Ambient temperature gradients and fluctuations can result in erroneous transmitter readings.
Keep impulse tubing as short as possible.
For differential transmitters, balance the liquid head on both legs of the impulse tubing.
For liquid flow or pressure measurements, make taps on the side of the line to avoid sediment deposits and mount the transmitter beside or below the taps so gases vent into the process line (see Figure 2-6 on page 2-9).
For gas flow or pressure measurements, make taps on the top or side of the line and mount the transmitter beside or above the taps so liquid drains into the process line (see Figure 2-6 on page 2-9).
For steam flow or pressure measurements, make taps on the side of the line and mount the transmitter below the taps so the impulse tubing stays filled with condensate (see Figure 2-6 on page 2-9).
For steam service, fill the lines with water to prevent steam from contacting the transmitter. Condensate chambers are not necessary since the volumetric displacement of the transmitter is negligible.
2-3
Rosemount 1154 Series H
The piping between the process and the transmitter must transfer the pressure measured at the process taps to the transmitter. Possible sources of error in this pressure transfer are:
•Leaks
Friction loss (particularly if purging is used)
Trapped gas in a liquid line or trapped liquid in a gas line (head error)
Temperature-induced density variation between legs (head error) for differential transmitters
To minimize the possibility of errors, take the following precautions:
Make impulse tubing as short as possible.
Slope tubing at least one inch per foot up toward the process connections for liquid and steam.
Slope tubing at least one inch per foot down toward the process connections for gas.
Avoid high points in liquid lines and low points in gas lines.
Use impulse tubing of sufficient diameter to avoid friction effects.
Ensure that all gas is vented from liquid tubing legs.
Ensure that impulse tubing is of adequate strength to be compatible with anticipated pressures.
Reference Manual
00809-0100-4631, Rev BA
April 2007
For differential transmitters, also consider the following:
Keep both impulse legs at the same temperature.
When using sealing fluid, fill both piping legs to the same level.
When purging, make the purge connection close to the process taps and purge through equal lengths of the same size tubing. Avoid purging through the transmitter.

Conduit The conduit connection to the transmitter is

conduit seal at the conduit entry to prevent moisture from accumulating in the terminal side of the housing during accident conditions. To prevent the conduit from adding mechanical stress to the transmitter during seismic disturbances, use flexible conduit or support the conduit near the transmitter. Install the conduit seal in accordance with the manufacturer’s instructions or use the procedure on page 2-9.

ELECTRICAL CONSIDERATIONS

This section contains information that you should consider when preparing to make electrical connections to the transmitter. Read this section carefully before proceeding to the electrical installation procedures.
The Rosemount 1154 Series H Pressure Transmitter provides a 4–20 mA signal when connected to a suitable dc power source. Figure 2-2 on page 2-5 shows a typical signal loop consisting of transmitter, power supply, and various receivers (controller, indicator, computer, etc.). The power supply must supply 12 volts minimum to the transmitter terminals at 30 mA (overscale) signal, or the maximum output current required for proper system operation. Any power supply ripple will be seen in the output load. The supply voltage versus load limitation relationship is shown in Figure 2-3 on page 2-5. See qualification report D8700096 for details. The load is the sum of the resistance of the signal leads and the load resistance of the receivers.
1
/2–14 NPT. Use a qualified
2-4
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Signal wiring need not be shielded, but twisted pairs should be used for best results. In electrically noisy environments, shielded cable should be used for best results. Do not run signal wiring in conduit or open trays with power wiring, nor near heavy electrical equipment. Signal wiring may be ungrounded (floating) or grounded at any place in the signal loop. The transmitter case may be grounded or ungrounded.
The capacitance sensing element uses alternating current to generate a capacitance signal. This alternating current is developed in an oscillator circuit with a frequency of 32,000 ±10,000 Hz. This 32,000 Hz signal is capacitor coupled to transmitter case ground through the sensing element. Because of this coupling, a voltage may be imposed across the load, depending on choice of grounding.
This impressed voltage, which is seen as high frequency noise, will have no effect on most instruments. Computers with short sampling times in a circuit where the negative transmitter terminal is grounded, will detect a significant noise signal. Filter this noise with a large capacitor (1 µf) or by using a 32,000 Hz LC filter across the load. Signal loops grounded at any other point are negligibly affected by this noise and do not need filtering.

Figure 2-2. Transmitter Wiring Connections.

Figure 2-3. Transmitter Load Limits.

Terminal Side (cover removed)
1825
1575
1500
1000
Power
Supply
1154-G05A
4–20 mA dc
Qualified Region
45
Design Region
50
2-5
500
0
Load Resistance ()
0
12
13.5
Power Supply (V dc)
35
40
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007

INSTALLATION PROCEDURES

Installation consists of mounting the transmitter and conduit and making electrical connections. Following are procedures for each operation.

Mechanical Transmitter

Be careful not to break the neck seal between the sensor module and the electronics housing.
The threaded interface between the sensor module and the electronics housing is hermetically sealed before shipment. The integrity of this seal is necessary for the safe operation of the transmitter during accident conditions. If the seal is broken, reseal it according to “Connecting Electrical Housing to Module Shroud” on page 5-4.
1. Mount bracket to a panel or other flat surface as shown in Figure 2-4
on page 2-7. Use four SAE grade 2, AISI 316 SST, or ASTM A193 grade 8M class 1 bolts were used during qualification testing. Torque each bolt to 33.5 ft-lb (45.4 N-m).
2. Attach the transmitter to the mounting bracket, as shown in Figure 2-4
on page 2-7. Use four with unit). Torque each bolt to 21 ft-lb (29 N-m).
3
/8-in. diameter bolts (not supplied with unit).
7
/16–20 3/4 in. bolts with washers (supplied
2-6
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Figure 2-4. Typical Transmitter Mounting Bracket Configuration.

Center of Gravity (bracket included)
0.06 (1.5)
1.31
(30.5)
7.55
(191.8)
4.93
(125)
2.81
(71.4)
5
(127)
PANEL MOUNTING HOLE PATTERN
(Back Side)
1.69 (42.9)
3
/8 in. Bolts (4)
(customer
supplied)
2.81
(71.4)
2.75
(69.9)
10 (254) Minimum Clearance
MOUNTING BRACKET FOR PANEL MOUNT SHOWN IN
TYPICAL MOUNTING CONFIGURATION
2.75 (69.9)
0.06 (1.5)
Center of Gravity (bracket included)
ACCEPTABLE ALTERNATE MOUNTING
NOTE All dimensions are nominal in inches (millimeters).
1154-A99B, B99B
2-7
Rosemount 1154 Series H
d

Figure 2-5. Transmitter Dimensional Drawings.

ROSEMOUNT 1154DH AND 1154HH
9.3 Max. (236)

4.7 Max. (119.4)

Dim.
A
4.3
(109)
7
/16 - 20
UNF
(Typical)
1.63
(41.3)
1
/2 - 14 NPT
Conduit
connection
(1 Place)
4.2
(107)
Compression fittings (2) Swagelok for 3/8 in.

4.72 Max. (119.9)

Welded
drain/vent
3.48 (88.3)
0.8(20) to end of mating tubing
Tubing (optional 1/4 - 18 NPT available)
valve (2)
(optional
1
available)
Reference Manual
00809-0100-4631, Rev BA
April 2007
0.75 (19) clearance for cover removal
Nameplate
(remove for
zero and
span adjust)
Transmitter
Circuitry
(this side)
Terminal
connections
(this side)
/4 - 18
NPT
3
/8 in. Mating Tubing

Detail A

0.8
(20)

4.7 Max. (119.4)

7
9.3 Max. (236)
/16 - 20
UNF
(typical)
1.63
(41.3)
Dim. A
Welde
4.3
Plug
(109)
NOTE All dimensions are nominal in inches (millimeters).
1
ROSEMOUNT 1154SH
/2 - 14 NPT
Conduit
connection
(1 Place)
4.2
(107)
Compression fittings (1) Swagelok for 3/8 in.
Tubing (optional 1/4 - 18 NPT available)
4.72 Max. (119.9)
3.48 (88.3)
0.8(20) to end of mating tubing
Welded
drain/vent
valve (1)
(optional
1
/4 - 18
NPT
available)
Pressure
Range Code
4, 5 2.13 (54.0) 6, 7 2.19 (55.6)
8 2.25 (57.2) 9 2.28 (57.9)
Nameplate
(remove for
zero and
span adjust)
Transmitter
Circuitry (this side)
Ter min al
connections
(this side)
Dimension A
0.75 (19) clearance for cover removal
Welded
Plug
2-8
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Conduit 1. Seal the conduit threads with thread sealant. (The transmitter conduit

seal interface was qualified using Grafoil mate with a standard
2. Starting at zero thread engagement, install the conduit into the transmitter between 4 and 7 turns, or a minimum of 12.5 ft-lb (16.9 N-m). Hold the electronics housing securely to avoid damaging the threaded neck seal between the sensor module and the electronics housing during conduit installation.
3. Provide separate support for the conduit if necessary.

Figure 2-6. Transmitter Installation Configuration.

Liquid Service
Sufficient
Length for
Cooling
Plugged Tees for Steam Service or Sealing Fluid
Blocking Valves
3-Valve
L
Manifold
H
Drain/Vent Valves
Flow
Plugged Tees for Steam Service or Sealing Fluid
Sufficient
Length for
Cooling
Blocking Valve
Drain/Vent Valve
1
/2–14 NPT male fitting.
Gas Service
Drain/Vent
H
Valves
tape.) Conduit threads
L
3-Valve Manifold
Flow
Flow
Drain/Vent Valve
H
Rosemount 1154DH, HH Rosemount 1154SH
Rosemount 1154DH, HH
Rosemount 1154SH

Electrical 1. Remove the cover from the terminal side of the transmitter

(see Figure 2-5 on page 2-8).
2. Connect the power leads to the “SIGNAL” terminals on the transmitter terminal block (see Figure 2-7 on page 2-10). Torque the terminal screws to 5 in-lb (0.6 N-m) or hand tight.
Do not connect signal leads to the “TEST” terminals.
3. Recheck connections for proper polarity.
4. Check the cover O-ring grooves for cleanliness. If chips or dirt are present, clean the seat and mating portion of the cover with alcohol. Lubricate replacement O-ring with O-ring grease (Rosemount P/N 01153-0248-0001 or P/N 01153-0053-0001). The transmitter was qualified using Dow Corning
5. Spray the inside threads of the electronics covers with cover lubricant (Rosemount P/N 01153-0333-0001 or equivalent) if necessary; if covers are already sufficiently lubricated, do not spray.
6. Carefully replace cover and tighten to 16.5 ft-lb (22.4 N-m).
®
55 Silicone O-ring Grease.
1154-0312AV2
2-9
Rosemount 1154 Series H
Figure 2-7. Transmitter Terminal Block.
Reference Manual
00809-0100-4631, Rev BA
April 2007
Signal Terminals
Test Terminals
2-10
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Section 3 Calibration

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-2

OVERVIEW Each transmitter is factory calibrated to the range specified by the customer.

This section contains the following transmitter calibration information:

Calibration

Span Adjustment Zero Adjustment
Calibration Procedures
Zero and Span Adjustment Linearity Adjustment Damping Adjustment Correction for High Line Pressure
CALIBRATION The Rosemount 1154DH, HH, and SH transmitters are factory calibrated to
the range shown on the nameplate. This range may be changed within the limits of the transmitter. Zero may also be adjusted to elevate or suppress. The span and zero adjustments are external and located under the nameplate.

Span Adjustment The span on any Rosemount 1154 Series H Transmitter is continuously

1
adjustable to allow calibration anywhere between maximum span and maximum span. For example, the span on a Range Code 4 transmitter can be continuously adjusted between 0–150 inH 0–25 inH
O.
2
O and
2
/6 of

Zero Adjustment The zero can be adjusted for up to 500 percent of span suppression or 600

percent of span elevation (see Figure 3-1).
The zero may be elevated or suppressed to these extremes with the limitation that no applied pressure within the calibrated range exceeds the full-range pressure limit. For example, a Range Code 4 transmitter cannot be calibrated for 150 inH 200 inH Code 4.
The transmitter may be calibrated to cross zero (e.g., –75 inH 75 inH
O to 200 inH2O (only 300 percent zero suppression) because the
2
O exceeds the 150 inH2O upper range pressure limit of a Range
2
O to
O), but this may result in a slight loss of linearity.
2
2
www.emersonprocess.com/rosemount/nuclear
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007

Figure 3-1. Zero Adjustment Range.

20
600% Zero Elevation
–150 –125 –100
Output (mA)
0
Pressure (inH2O) No Zero Elevation or Suppression
20
Output (mA)
Rosemount 1154 Series H with a calibrated span of 25 inH
4
Graphs based on a Range 4 (0–25 to 0–150 inH2O)
500% Zero Suppression
25 50
0
–75 –50
Pressure (inH2O) 600% Zero Elevation
20
4
25
75 100 125 150
Pressure (inH2O)
500% Zero Suppression
–25 0
Output (mA)
4
25
2
O
1152-0282A

CALIBRATION PROCEDURE

Zero and Span Adjustment

NOTE

The Rosemount 1154 Series H Pressure Transmitter contains electronic circuit boards which may be static sensitive.

NOTE

Covers need not be removed for zero and span adjustment.
The zero and span adjustment screws are accessible externally. They are located behind the nameplate on the side of the electronics housing (see Figure 3-2). The output of the transmitter increases with clockwise rotation of the adjustment screws.
The zero adjustment screw has very little effect on the span. The span adjustment, however, does affect zero. The effect of interaction is more apparent with suppression or elevation. The span adjustment changes the zero output and the full-scale output by approximately the same percentage. Therefore, it is best to calibrate the transmitter from 0 to the desired span and finish the calibration by adjusting the zero screw to achieve the desired elevation or suppression.
3-2
Reference Manual
00809-0100-4631, Rev BA April 2007

Figure 3-2. Zero and Span Adjustment Screws.

Rosemount 1154 Series H
Zero
Screw
Span
Screw
1154-0314A
Example (for Range Code 4)
Initial transmitter calibration: 25 to 125 inH suppressed 25 inH
2
O).
Desired transmitter calibration: –75 to –25 inH elevated 75 inH
O).
2
O (100 inH2O span with zero
2
O (50 inH2O span with zero
2
1. Adjust the zero to eliminate any existing zero elevation or suppression. With 0 inH
O pressure applied to the transmitter, turn
2
zero adjustment until output reads 4 mA. The unit is now calibrated for 0 to 100 inH
O.
2
2. Adjust the span to the desired new span. To reduce the span, turn the span screw until the output, with 0 in H
Existing Span
4mA
⎛⎞
---------------------------------------
⎝⎠
Desired Span
3. Adjust the zero screw to bring the output, with 0 inH
O pressure input, equals:
2
100 inH
4mA
⎛⎞
------------------------------
⎝⎠
50 inH
2
O
2
O
2
8mA==
O input, back to 4 mA. The transmitter calibration should now be very close to 0 to 50 inH
O.
2
4. Check full-scale output and fine tune the span and zero adjustment if required. Remember zero adjustments do not affect span, but span adjustments do affect zero predictably. Adjusting the span screw will affect the zero 1/5 as much as it affects the span. To compensate for this effect, simply overadjust by 25 percent. For example, if after completing Step 3, the transmitter output read 19.900 mA at 50
O, turn the span potentiometer until the output (at 50 inH2O)
inH
2
reads 20.025 mA.
19.900 + (20.000 – 19.900) 1.25 = 19.900 + 0.125 = 20.025
3-3
Rosemount 1154 Series H
Since the span adjustment affects zero one-fifth as much as the span, the 0.125 mA increase in span causes a 0.025 mA increase in zero. Therefore, turn the zero adjustment (at 50 inH reads 20.000 mA. The unit should now be calibrated for 0 to 50 in H
5. Zero Elevation/Suppression. Elevate zero. Turn screw until the output reads 4 mA with –75 inH (applying 75 inH output may stop changing before the desired 4 mA reading is obtained. If this occurs, turn off power to the unit and unplug the amplifier board (see Figure 5-2 on page 5-6 to locate the amplifier board). To elevate or suppress zero a large amount, use the following procedure:
Material
Wire: 22-gauge tinned solid copper—Fed Spec QQW343, ASTM B33
Solder: 60% tin, 40% lead (60/40)—Fed Spec QQ-S-571
Flux: MIL F 14256 Type A; Fed Spec QQ-S-571 Type RA
Method
a. Locate the three turret terminals on the component side of the
b. To elevate zero, connect a jumper wire between the middle
c. Wrap the jumper wire once around each terminal and cut off any
d. Solder the jumper wire to the terminals using proper electronics
e. Plug the amplifier board back in and complete the zero adjustment.
Reference Manual
00809-0100-4631, Rev BA
April 2007
O) until the output
2
O.
2
O applied to the high side of the transmitter
O to the low side will give the same result). The
2
amplifier board. Remove any jumper wires between them (see Figure 3-3).
terminal and the terminal marked “EZ” (see Figure 3-3, Detail B).
excess.
soldering techniques. Clean solder joints thoroughly with isopropyl alcohol.
2
3-4
To suppress zero, follow the same procedure, except connect the jumper wire between the middle terminal and the terminal marked “SZ” (see Figure 3-3, Detail C).
6. Recheck full scale and zero and fine tune if necessary.
NOTE
There is some mechanical backlash in the zero and span adjustments, so there is a dead band when you change the direction of adjustment. Because of the backlash, the simplest procedure, if the desired setting is overshot, is to intentionally overshoot a larger amount before reversing the direction of the adjustment.
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Figure 3-3. Jumper Wire Placement.

DETAIL A
Moderate
Elevation/
Suppression
(No Jumper Wire)
DETAIL B
(To Elevate Zero)
Jumper Wire
Jumper
DETAIL C
(To Suppress Zero)
Wire

Linearity Adjustment In addition to the span and zero adjustments, there is a linearity adjustment

located inside the transmitter on the amplifier board (see Figure 3-4 on page 3-6). Linearity is factory calibrated for optimum performance over the calibrated range of the instrument and is not normally adjusted in the field. If you want to maximize linearity over some particular range, use the following procedure:

1. Apply mid-range pressure and note the error between theoretical and actual output signal.

2. Apply full-scale pressure. Multiply the error noted in Step 1 by six and by the following rangedown factor:

1152-0289A, 0289B, 0289C
Maximum Allowable Span
Rangedown Factor
-------------------------------------------------------------------- -=
Calibrated Span
3. Add the result to the full-scale output for negative errors, or subtract the result from the full-scale output for positive errors, by adjusting the linearity trimmer (see Figure 3-4). Example: At 4-to-1 rangedown the midscale point is low by 0.05 mA. Therefore, adjust the “Linearity” trimmer until full-scale output increases by (0.05 mA 6 4) = 1.2 mA.

4. Readjust zero and span.

NOTE

If you remove either cover during the above procedures, replace the O-ring and torque the cover per the instructions given in Section 5: Maintenance and Troubleshooting. Spare cover O-rings are supplied with each transmitter.
3-5
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007

Figure 3-4. Linearity and Damping Adjustment.

Damping Adjustment (optional)
Linearity Adjustment
Electronics Side of Transmitter
Housing (cover removed)

Damping Adjustment Damping electronics are available as an option. Transmitters with standard

electronics can be retrofitted with the adjustable damping feature by changing out both the amplifier board (RMT P/N 01154-0021-0004) and the calibration board (RMT P/N 01154-0023-0002).
The damping adjustment permits damping of rapid pressure variations by adjusting the single-turn trim potentiometer located on the upper right-hand side of the amplifier board (see Figure 3-4). The available settings, when adjusted to the maximum position, provide time-constant values of at least 1.2 seconds for Range Code 4 and 0.8 seconds for Range Codes 5–9. Transmitters with the electronic damping option are calibrated and shipped with the adjustment set at the counterclockwise stop, which gives the minimum time-constant.
1154-0299B

Correction For High Line Pressure (Rosemount 1154DH and 1154HH only)

To adjust the damping, turn the damping adjustment potentiometer until the desired time-constant is obtained. It is best to set the damping to the shortest possible time-constant. Since transmitter calibration is not affected by the damping setting, you may adjust the damping with the transmitter installed on the process.
The damping adjustment potentiometer has positive stops at both ends. Forcing the potentiometer beyond the stops may cause permanent damage.

NOTE

If you remove either cover during the above procedures, replace the O-ring and torque the cover per the instructions given in Section 5: Maintenance and Troubleshooting of this manual. Spare cover O-rings are supplied with each transmitter.
If a differential transmitter is calibrated with the low side at ambient pressure but will be used at high line pressure, correct the span adjustment to compensate for the effect of static pressure on the unit. If zero is elevated or suppressed, also correct the zero adjustment. Correction factors, expressed in percent of differential pressure input at end points per 1,000 psi static pressure, are:
3-6
Reference Manual
00809-0100-4631, Rev BA April 2007

Span Range Codes 4, 5, and 8:

+0.75% of input/1,000 psi

Range Codes 6 and 7:

+1.25% of input/1,000 psi
The correction procedure below uses the following example:
Rosemount 1154 Series H
Range Code 5, calibrated at –100 to 300 inH
O to be operated at 1,200 psi
2
line pressure. Note that steps 3–6 are omitted for ranges based at zero differential pressure.
1. Calibrate the unit per the “Calibration Procedure” on page 3-2 to output = 4 mA at –100 inH
O and 20 mA at 300 inH2O.
2
2. Calculate correction factor:
0.75 %
-----------------------­1,000 psi
1 200 psi 0.9% differential input=,×
3. Calculate zero adjustment correction in terms of pressure:
0.9 % 100 inH
× O 0.9 inH2O=
2
4. Convert pressure correction to percent of input span:
0.9 i nH2O
---------------------------------------------------------------- 0.225 % span= 400 inH2O input span
5. Calculate correction in terms of output span (mA):
0.225 % 16 mA span× 0.036 mA=
6. Add the milliamp correction to the ideal zero output (4 mA). This is the corrected ideal zero output:
4.00 mA 0.036 3.964 mA=
7. Calculate full-scale adjustment correction in terms of pressure:
0.9 % 300 in H
× O 2.7 i nH2O=
2
8. Repeat Step 4 with the results of Step 7:
2.7 inH2O
---------------------------------------------------------------- 0.675 % span= 400 inH2O input span
9. Repeat Step 5 with the result of Step 8:
0.675 % 16 mA span× 0.108 mA=
10. Add the mA correction to the ideal full-scale output (20 mA). This is the corrected ideal full-scale output.
20.00 mA 0.108 mA 20.108 mA=+
3-7
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007
11. Readjust zero and span adjustments for corrected outputs:
3.964 mA at –100 inH
20.108 mA at 300 inH
O
2
O
2
There is an uncertainty of ±0.5 percent of input reading per 1,000 psi associated with the span correction.

Zero Zero shift with static pressure is not systematic. However, if the calibrated

range includes zero differential pressure, the effect can be trimmed out after installation and with the unit at operating pressure.
Equalize pressure to both process connections, and turn the zero adjustment until the ideal output at zero differential input is observed. Do not readjust the span potentiometer.
If the transmitter does not include zero differential pressure within its calibrated span, the zero effect or zero correction can be determined before the unit is suppressed or elevated to eliminate the zero effect after correcting for the span effect.
The following procedure illustrates how to eliminate the zero effect for a non-zero differential pressure calibration. The example uses a Range Code 5 calibrated from 100 to 500 inH
1. Using standard calibration procedures, calibrate the unit to the required span, with the 4 mA or zero point corresponding to zero differential pressure:
O with 1,200 psi static line pressure.
2
4 mA at 0 inH
O and 20 mA at 400 inH2O
2
2. Apply static pressure to both high and low process connections with zero differential pressure across the transmitter, and note the zero correction (zero shift). For example, if the output reads 4.006 mA, the zero correction is calculated as:
4.00 mA – 4.006 mA = –0.006 mA
Note the sign associated with this correction, as this result is added when determining the final, ideal transmitter output.
3. Remove static pressure and correct for the span effect as outlined in the span correction procedure. Calibrate the unit to the calculated output values. If, for example, the span correction procedure yielded
4.029 mA and 20.144 mA, calibrate the unit for:
4.029 mA at 100 inH
20.144 mA at 500 inH
O
2
O
2
4. Add the zero correction (–0.006 mA), found in Step 2, to the ideal zero point value calculated in Step 3.
4.029 mA + (–0.006 mA) = 4.023 mA
5. To eliminate the zero effect, readjust the zero potentiometer so the output reads the ideal zero point calculated in Step 4 (do not readjust the span potentiometer). Note that all the calibration points will shift the same amount toward the correct reading. The example output is now 4.023 mA at 100 inH
O.
2
The transmitter output is now 4–20 mA over its calibrated span when the unit is operated at 1,200 psi static line pressure.
3-8
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Section 4 Operation

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
Transmitter Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
The -Cell
Demodulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-3
Linearity Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-3
Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-3
Voltage Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-4
Zero and Span Adjustments . . . . . . . . . . . . . . . . . . . . . . . page 4-4
Current Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-4
Current Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-4
Reverse Polarity Protection . . . . . . . . . . . . . . . . . . . . . . . . page 4-4

OVERVIEW This section provides brief descriptions of basic transmitter operations in the

following order:
Transmitter Operation
The -Cell
Demodulator
Linearity Adjustment
Oscillator
Voltage Regulator
Zero and Span Adjustments
Current Control
Current Limit
Reverse Polarity Protection
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-3
Sensor

TRANSMITTER OPERATION

www.emersonprocess.com/rosemount/nuclear
The block diagram in Figure 4-2 on page 4-3 illustrates the operation of the transmitter.
The Rosemount 1154 Series H Alphaline Pressure Transmitters have a variable capacitance sensing element, the -Cell (Figure 4-1 on page 4-2). Differential capacitance between the sensing diaphragm and the capacitor plates is converted electronically to a 2-wire 4–20 mA dc signal.
C2C1–
⎛⎞
--------------------
=
PK
⎜⎟
1
+
C
⎝⎠
1C2
Where:
P is the process pressure.
Rosemount 1154 Series H
K1is a constant.
C
is the capacitance between the high pressure side and the sensing
1
diaphragm.
C
is the capacitance between the low pressure side and the sensing
2
diaphragm.
Where:
I
is the current source.
ref
V
is the peak-to-peak oscillation voltage.
p-p
f is the oscillation frequency.
fV
pp
--------------------=
C1C2+
Reference Manual
00809-0100-4631, Rev BA
April 2007
I
ref
Figure 4-1. The -Cell.
Leadwires
Isolating Diaphragm
I
difffVpp–C2C1
()=
Where:
is the difference in current between C1 and C2.
I
diff
Capacitor Plates
Sensing Diaphragm
Rigid Insulation
Silicone Oil
Welded Seals
1151-0142A
4-2
Therefore:
C2C1–
⎛⎞
P Constant I
==
× I
diff
--------------------
⎜⎟
ref
+
C
⎝⎠
2C1
Reference Manual
00809-0100-4631, Rev BA April 2007

Figure 4-2. Electrical Block Diagram.

Rosemount 1154 Series H
SENSOR

OSCILLATOR

DEMODULATOR

OSC. CONTROL AMP.
CURRENT DETECTOR
VOLTAGE REGULATOR
CURR. CONTROL AMP.
CURRENT LIMITER
CURRENT CONTROL
REVERSE POLARITY PROTECTION
-
TEST
+
SIGNAL
THE -CELL™ SENSOR Process pressure is transmitted through an isolating diaphragm and silicone
oil fill fluid to a sensing diaphragm in the center of the pressure is transmitted in like manner to the other side of the sensing diaphragm.
The position of the sensing diaphragm is detected by the capacitance plates on both sides of the sensing diaphragm. The capacitance between the sensing diaphragm and either capacitor plate is approximately 150 pf. The sensor is driven through transformer windings by an oscillator at roughly 32 kHz and 30 V
p-p
.
-Cell. The reference
1154-0295A
DEMODULATOR The demodulator consists of a diode bridge that rectifies the ac signal from
the sensor cell to a dc signal.
The oscillator driving current, I
(the sum of the dc currents through two
ref
transformer windings) is controlled to be a constant by an integrated circuit amplifier.
The dc current through a third transformer winding is a current directly proportional to pressure, i.e.,
I
diff
fV
()=
ppC2C1
The diode bridge and span temperature compensating thermistor are located inside the sensor module. The effect of the thermistor is controlled by resistors located in the electronics housing.

LINEARITY ADJUSTMENT

Linearity is adjusted by a variable-resistance network, capacitor, and diodes. The currents generated through this part of the circuit are summed into the inputs of the oscillator control circuit. This provides a programmed correction that raises the oscillator peak-to-peak voltage to compensate for first-order nonlinearity of capacitance as a function of pressure.
OSCILLATOR The oscillator has a frequency determined by the capacitance of the sensing
element and the inductance of the transformer windings.
4-3
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
The sensing element capacitance is variable. Therefore, the frequency is variable about a nominal value of 32 kHz.
An integrated circuit amplifier is used as a feedback control circuit and controls the oscillator drive voltage such that:
fV
pp
--------------------=
C1C2+
I
ref
April 2007

VOLTAGE REGULATOR The transmitter uses a zener diode, transistor, and resistors to provide a

constant voltage of 6.4 V dc for the reference and 7 V dc for the oscillator.

ZERO AND SPAN ADJUSTMENTS

Zero adjustment components consist of a potentiometer and resistor that develop a separate adjustable current that sums with the sensor current. The coarse zero switch switches resistors into the circuit as needed.
Span adjustment is performed with a potentiometer which determines the amount of loop current which is sensed and fed back to the current control amplifier.

CURRENT CONTROL The current control amplifier consists of an integrated circuit, two transistors,

and associated components. The IC reference voltage is established at the junction of a resistor network. The current control amplifier drives the current control to a level such that the current detector feeds back a signal equal to the sum of the zero current and the variable sensor current.

CURRENT LIMIT A current limiter prevents the output current from exceeding 30 mA in an

overpressure condition.
REVERSE POLARITY
A zener diode provides reverse polarity protection.
PROTECTION
4-4
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Section 5 Maintenance and
Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2
Test Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2
Board Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2
Sensing Module Checkout . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2
Disassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-3
Reassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-4
Post-Assembly Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-5
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-7
Use only the procedures and new parts specifically referenced in this manual to ensure specification performance and certification compliance. Unauthorized procedures or parts can render the instrument dangerous to life, limb, or property.

OVERVIEW This section contains a technique for checking out the components, a method

for disassembly and reassembly, and a troubleshooting guide.

NOTE

Maintenance of traceability of any replacement parts is the responsibility of the user (see Important Notice on page 6-9 and Important Notice at the beginning of this manual, preceding the Table of Contents).
The Rosemount 1154 Series H has no moving parts and requires a minimum of scheduled maintenance. Calibration procedures for range adjustment are outlined in Section 3: Calibration. A calibration check should be conducted after inadvertent exposure to overpressure, unless your plant considers this factor separately in the plant error analysis.
Test terminals are available for in-process checks. For further checks, the transmitter can be divided into three active physical components: the sensing module, the amplifier board, and the calibration board.
An exploded view of the transmitter is provided in Figure 5-2 on page 5-6. In the following procedures, numbers in parenthesis refer to item numbers in the exploded view.
www.emersonprocess.com/rosemount/nuclear
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H
April 2007

SAFETY MESSAGES Instructions and procedures in this section may require special precautions to

ensure the safety of the people performing the operations. Information that raises potential safety issues is indicated by a warning message. The following warning messages appear in this section.
Use only the procedures and new parts specifically referenced in this manual to ensure specification performance and certification compliance. Unauthorized procedures or parts can render the instrument dangerous to life, limb, or property.

TEST TERM IN ALS The test terminals are connected across a diode through which the loop

signal current passes. The indicating meter or test equipment shunts the diode when connected to the test terminals. As long as the voltage across the terminals is kept below the diode threshold voltage, no current passes through the diode. To ensure that there is no current leaking through the diode when making a test reading or when connecting an indicating meter, the resistance of the test connection or meter should not exceed 10 Ω.

BOARD CHECKOUT NOTE

Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.

Figure 5-1. Header Board Connections.

SENSING MODULE CHECKOUT

NOTE

The Rosemount 1154 Series H Pressure Transmitter contains electronic circuit boards which may be static sensitive.
You can easily check the printed circuit boards (5 and 6) for a malfunction by substituting spare boards into the circuit. If this procedure turns up a malfunctioning board, return the defective board to Rosemount Nuclear Instruments, Inc. for replacement. Because of parts traceability, qualification becomes the responsibility of the customer in the event of unauthorized board repairs.
(Component Side Up)
1151-0180A

NOTE

Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.
5-2
The sensing module is contained within the module shroud (8) and is not field repairable. If no defect such as loss of fill fluid is observed, check the sensing module in the following manner.
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
1. Disengage the header assembly board (4) as described in Step 4 of the electrical housing disassembly procedure on page 5-3. You need not remove the module shroud from the electrical housing for checkout.
2. Jump connections 1 and 2 on the header assembly board (see Figure 5-1).
3. Using a low-voltage ohmmeter, check the resistance between the jumper wire and the module shroud. This resistance should be greater than 10 MΩ. Remove the jumper wire.
4. Jump connections 3 and 4 on the header assembly board and repeat Step 3 (see Figure 5-1).

NOTE

The above procedure does not completely test the sensing module. If circuit board replacement does not correct the abnormal condition, and no other problems are obvious, the transmitter should be replaced.

DISASSEMBLY PROCEDURE

Electrical Housing Disassembly

NOTE

Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.

NOTE

The Rosemount 1154 Series H Pressure Transmitter contains electronic circuit boards which may be static sensitive.
1. The signal terminals and the test terminals are accessible by unscrewing the cover (1) on the terminal side. This compartment is identified as “Terminal Side” on the nameplate. The terminals are permanently attached to the housing and must not be removed.
2. Circuit boards are located in a separate compartment identified as “Circuit Side” on the nameplate. Remove power from the transmitter before removing the circuit cover. Unscrew the cover (1) on the circuit side to access the circuit boards. A special cover wrench (Rosemount Part Number 01153-0382-0001) is available from Rosemount to remove and replace the housing covers.

3. Unplug the amplifier board (6) after removing 3 holding screws (7).

4. The header assembly board (4) is permanently attached to the sensor module within the module shroud (8) and contains the temperature-compensating resistors. Carefully pull this board off the bayonet pins and rotate the board 180° about the axis formed by the connecting leads. This allows access to the calibration board (5).
5. Disconnect the calibration board (5) by aligning the zero and span adjust screws so that their slots are perpendicular to the board. Remove the board by inserting a 6-32 screw in the rivnut on the board, and carefully pull the board off the bayonet pins.

6. If replacement of the zero and span adjustment screws (10) is necessary, remove the nameplate (13) and detach the snap rings (12) inside the housing.

5-3
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007

Removing Module Shroud From Electrical Housing

REASSEMBLY PROCEDURE

1. Remove amplifier board and calibration board as described in the Electrical Housing Disassembly Section above.

2. Loosen the lock nut (9).

3. Unscrew the module shroud (8) from the electronics housing, simultaneously turning the header board and leads to prevent them from being twisted or damaged. The threaded connection has a sealing compound on it and must be broken loose. Then carefully pull the header assembly board (4) through the hole.

4. The module shroud (8) containing the sensor module is a complete assembly and cannot be further disassembled.

NOTE

Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-6.

NOTE

The Rosemount 1154 Series H Pressure Transmitter contains electronic circuit boards which may be static sensitive.

Preliminary 1. Replace the cover O-rings (2) whenever you remove a cover. Clean

the sealing areas with alcohol, if necessary, and lightly grease the O-ring with Dow Corning
Part Number 01153-0248-0001 or P/N 01153-0053-0001). Spray the inside threads of the electronics covers with cover lubricant (Rosemount P/N 01153-0333-0001 or equivalent) if necessary. If covers are already sufficiently lubricated, do not spray.

2. Verify that the circuit boards are clean.

3. Verify that the bayonet pins on the connection board are clean.

4. If you remove the module shroud, clean the thread sealant from module shroud threads, lock nut, and electronics housing threads with a wire brush.

®
55 Silicone O-ring Grease (Rosemount

Connecting Electrical Housing to Module Shroud

5-4

1. Run the lock nut down to the base of the sensor threads.

3
2. Apply a heavy, continuous bead (about 580-PST sealant (Rosemount Part Number 01153-0329-0001) around the top module shroud threads.

3. Insert header assembly board (4) through the hole in the bottom of the electronics housing.

4. Screw module shroud (8) into electrical housing (3) making sure that five full threads are engaged. Be careful not to damage or twist the sensor leads. Turn the header board to avoid twisting wires.

5. Align the module shroud with the high and low pressure sides oriented per Figure 2-5 on page 2-8. Alternately, tighten the module shroud one-half turn further to reverse the orientation of the module shroud about the electronics housing.

6. Tighten the lock nut (9) to 35 ft-lb (48 N-m) torque.

7. Wipe off excess sealant.

8. Place the assembled unit in an oven at 200 ± 5 °F (93 ± 3 °C) for 12 hours to cure the sealant.
/8-in. wide) of Loctite®
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Electrical Housing Reassembly

1. Replace the zero and span adjustment screw O-rings (11) whenever you remove the zero or span adjustment screws (10). Lightly grease the O-rings with Dow Corning 55 Silicone O-ring Grease, (Rosemount P/N 01153-0248-0001). Reinstall adjustment screws and secure with snap rings (12).

2. Align the zero and span adjustment screws with the potentiometer stems on the calibration board (5) and push the calibration board onto the bayonet pins.

3. Slide the header assembly board (4) onto the bayonet pins with the component side toward the pins. Slide any excess wire behind the calibration board, taking care to avoid kinks.

4. Push the amplifier board (6) onto the bayonet pins and secure with the holding screws (7). Use nominal torque of 10 in-lb (1.1 N-m).

5. Carefully replace the cover and tighten to 16.5 ft-lb (22.4 N-m) (see “Preliminary” on page 5-4).

6. Replace the nameplate (13) and attach with two nameplate screws (14).

POST-ASSEMBLY TESTS 1. Calibrate transmitter per the calibration section of this manual.

2. Conduct nuclear cleaning to 1 ppm chloride content of transmitter “wetted parts.”

5-5
Rosemount 1154 Series H

Figure 5-2. Exploded View Parts Drawing, Rosemount 1154 Series H.

1
Reference Manual
00809-0100-4631, Rev BA
April 2007
2
3
5
13
4
12
10
11
9
14
6
7
2
1
8

Table 5-1. Rosemount 1154 Series H Parts List.

Part Description Part Description
1 Electronics Cover 8 Module Shroud
2 O-Ring for Electronics Cover 9 Module Shroud Lock Nut
3 Electronics Housing 10 Zero and Span Adjustment Screws
4 Header Assembly Board 11 O-Ring for Adjustment Screw
5 Calibration Board 12 Snap Rings
6 Amplifier Board 13 Nameplate
7 Holding Screws 14 Nameplate Screws

Table 5-2. Torque Reference.

Items(s) to be Torqued Torque Value Tol er ance
Bracket to Mounting Panel Bolts Transmitter to Bracket Bolts Swagelok Process Fittings Drain/Vent Valves Covers Module Neck Lock Nut Conduit Fitting Terminal Block Screws Amplifier Board Screws
5-6
4 to 7 turns or a min. of 12.5 ft-lb (16.9 N-m)
33.5 ft-lb (45.4 N-m) 21 ft-lb (29 N-m)
See Installation Instructions
7.5 ft-lb (10 N-m)
16.5 ft-lb (22.4 N-m) 35 ft-lb (48 N-m)
5 in-lb (0.6 N-m)
Nominal 10 in-lb (1.1 N-m)
±1 ft-lb (±1 N-m) ±1 ft-lb (±1 N-m)
±0.5 ft-lb (±0.7 N-m)
±1 ft-lb (±1 N-m) ±1 ft-lb (±1 N-m) ±1 ft-lb (±1 N-m)
±1 in-lb (±0.1 N-m)
1154-A02B
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

TROUBLESHOOTING

Table 5-3. Troubleshooting.
Symptom Potential Source Corrective Action
High Output Primary Element Check for restrictions at primary element, improper installation or poor condition. Note any changes
Impulse Piping Check for leaks or blockage.
Transmitter Electronics Make sure that post connectors and the sensor connections are clean.
Transmitter Electronics Failure
Sensing Module See sensing module checkout section. The sensing module is not field repairable and must be
Power Supply Check the power supply output voltage at the transmitter.
Low Output or No Output
Primary Element Check for restrictions at primary element, improper installation or poor condition. Note any changes
Loop Wiring
in process fluid properties.
Ensure that blocking valves are fully open. Check for entrapped gas in liquid lines and for liquid in dry lines. Ensure that density of fluid in impulse lines is unchanged. Check for sediment in the transmitter process flanges.
If the electronics are still suspect, substitute new electronics. Determine faulty circuit board by trying spare boards. Replace faulty circuit board.
replaced if found to be defective. See “Disassembly Procedure” for instructions on disassembly. Check for obvious defects, such as punctured isolating diaphragm or fill fluid loss, and contact Rosemount Nuclear Instruments, Inc. at 952-949-5210.
in process fluid properties.
Do not use over 100 volts to check the loop, or damage to the transmitter electronics may result.
Check for adequate voltage to the transmitter. Check the milliamp rating of the power supply against the total current being drawn for all transmitters being powered. Check for shorts and multiple grounds. Check for proper polarity at the signal terminal. Check loop impedance. Check wire insulation to detect possible shorts to ground.
Impulse Piping Ensure that the pressure connection is correct.
Transmitter Electronics Connections
Test Diode Failure Replace electronics housing. Transmitter Electronics
Failure Sensing Module See sensing module checkout section. The sensing module is not field repairable and must be
Power Supply Check the power supply output voltage at transmitter.
Check for leaks or blockage. Check for entrapped gas in liquid lines. Check for sediment in the transmitter process flange. Ensure that blocking valves are fully open and that bypass valves are tightly closed. Ensure that density of the fluid or other fluid properties in the impulse piping are unchanged.
Ensure that calibration adjustments are in allowable range. Check for shorts in sensor leads. Make sure post connectors are clean, and check the sensor connections. If the electronics are still suspect, substitute new electronics.
Determine faulty circuit board by trying spare boards. Replace faulty circuit board.
replaced if found to be defective. See “Disassembly procedure” for instructions on disassembly. Check for obvious defects, such as punctured isolating diaphragm or fill fluid loss, and contact Rosemount Nuclear Instruments, Inc. at 952-949-5210.
Continued On Next Page
5-7
Rosemount 1154 Series H
Symptom Potential Source Corrective Action
Erratic Output Loop Wiring
Do not use over 100 volts to check the loop, or damage to the transmitter electronics may result.
Check for adequate voltage to the transmitter. Check for intermittent shorts, open circuits, and multiple grounds.
Impulse Piping and Process Connections
Transmitter Electronics Check for intermittent shorts or open circuits.
Transmitter Electronics Failure
Power Supply Check power supply output voltage.
Check for entrapped gas in liquid lines and for liquid in dry lines.
Make sure that post connectors and the sensor connectors are clean and properly connected. Determine faulty board by trying spare boards. Replace faulty circuit board.
Reference Manual
00809-0100-4631, Rev BA
April 2007
5-8
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Section 6 Specifications and Reference
Data
Nuclear Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page 6-2
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . .page 6-4
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-6
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-7

NUCLEAR SPECIFICATIONS

Qualified to IEEE Std 323-1974 and IEEE Std 344-1975 per Rosemount Report D8700096.

Radiation

Accuracy within ±(0.25% of upper range limit + 0.75% of span) during first 30 minutes; ±(0.5% upper range limit + 1.0% span) thereafter up to 55 megarads total integrated dosage (TID); ±(0.75% upper range limit + 1.0% span) after 110 megarads TID gamma radiation exposure

Seismic

Accuracy within ±0.5% of upper range limit during and after a disturbance defined by a required response spectrum with a horizontal ZPA of 8.5 g and a vertical ZPA of 5.2 g

Steam Pressure/Temperatures

Accuracy within ±(1.0% of upper range limit + 1.0% of span) for range codes 4–8; ±(2.0% upper range limit + 0.5% span) for range code 9 during and after sequential exposure to steam at the following temperatures and pressures, concurrent with chemical spray for the first 24 hours:
420 °F (215.6 °C), 85 psig for 3 minutes
350 °F (176.6 °C), 85 psig for 7 minutes
320 °F (160 °C), 75 psig for 8 hours
265 °F (129.4 °C), 24 psig for 56 hours
www.emersonprocess.com/rosemount/nuclear

Chemical Spray

Chemical Spray composition is 0.28 molar boric acid, 0.064 molar sodium thiosulfate, and sodium hydroxide to make an initial pH of 11.0 and a subsequent pH ranging from 8.5 to 11.0. Chemical spray is sprayed at a rate of 0.25 gal/min/ft

Post DBE Operation

Accuracy at reference conditions shall be within ±2.5% of upper range limit for one year following DBE.

Quality Assurance Program

In accordance with NQA-1, 10CFR50 Appendix B, and ISO 9001:2000
2
.
Rosemount 1154 Series H

Nuclear Cleaning

To 1 ppm maximum chloride content

Hydrostatic Testing

To 150% of maximum working pressure or 2,000 psi (13.8 MPa), whichever is greater

Traceability

In accordance with NQA-1 and 10CFR50 Appendix B; chemical and physical material certification of pressure-retaining parts.

Qualified Life

Dependent on continuous ambient temperature at the installation site, illustrated in Figure 6-1 on page 6-2. Replacement of amplifier and calibration circuit boards at the end of their qualified life permits extension of the transmitter’s qualified life to the module’s qualified life. See Rosemount Report D8700096.
Reference Manual
00809-0100-4631, Rev BA
April 2007

Figure 6-1. Qualified Life vs. Ambient Temperature.

PERFORMANCE SPECIFICATIONS

Module Qualified Life
Electronics
Qualified Life
Time (Years)
Temperature (°F)
Based on zero-based ranges under reference conditions

Accuracy

1154-0293A
6-2
±0.25% of calibrated span; includes combined effects of linearity, hysteresis, and repeatability

Deadband

None
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

Drift

±0.2% of upper range limit for 30 months

Temperature Effect

Range Codes 4–8:
±(0.15% upper range limit + 0.35% span) per 50 °F (27.8 °C) ambient
temperature change between 40 °F (4.4 °C) and 130 °F (54.4 °C)
Range Code 9:
±(0.25% of upper range limit + 0.5% span) per 50 °F (27.8 °C) ambient
temperature change between 40 °F (4.4 °C) and 130 °F (54.4 °C)
All Range Codes:
±(0.75% of upper range limit + 0.5% span) per 100 °F (55.6 °C) ambient
temperature change between 40 °F (4.4 °C) and 200 °F (93.3 °C)

Overpressure Effect

Rosemount 1154DH:
Maximum zero shift after 2,000 psi (13.8 MPa) overpressure:
Range Code Overpressure Effect
4 ±0.25% of upper range limit
5 ±1.0% of upper range limit
6–7 ±3.0% of upper range limit
8 ±6.0% of upper range limit
Rosemount 1154HH:
Maximum zero shift after 3,000 psi (20.68 MPa) overpressure:
Range Code Overpressure Effect
4 ±1.0% of upper range limit
5 ±2.0% of upper range limit
6–7 ±5.0% of upper range limit
Rosemount 1154SH:
Maximum zero shift after 4,500 psi (31.0 MPa) overpressure:
Range Code Overpressure Effect
9 ±0.5% of upper range limit

Static Pressure Zero Effect

Rosemount 1154DH:
Per 1,000 psi (6.89 MPa):
Range Code Static Pressure Zero Effect
4–5 ±0.2% of upper range limit
6–8 ±0.5% of upper range limit
Rosemount 1154HH:
Per 1,000 psi (6.89 MPa):
Range Code Static Pressure Zero Effect
4–7 ±0.66% of upper range limit
6-3
Rosemount 1154 Series H

Static Pressure Span Effect

Effect is systematic and can be calibrated out for a particular pressure before installation; correction uncertainty: ±0.5% of input reading/1,000 psi (6.89 MPa)

Power Supply Effect

Less than 0.005% of output span/volt

Load Effect

No load effect other than the change in voltage supplied to the transmitter

Mounting Position Effect

Reference Manual
00809-0100-4631, Rev BA
April 2007

FUNCTIONAL SPECIFICATIONS

No span effect; zero shift of up to 1.5 inH
O (372 Pa) which can be calibrated
2
out

Response Time

Fixed time constant (63%) at 100 °F (37.8 °C) as follows:
Range Code Response Time
4 0.5 seconds or less
all others 0.2 seconds or less
Adjustable damping is available through a special N option.

Service

Liquid, gas, or vapor

Output

4–20 mA dc

Power Supply

Design limits as shown in Figure 6-2. See Rosemount Report D8700096 for additional detail.

Figure 6-2. Load Limitations.

6-4
1575
Load Resistance ()
1825
1500
1000
500
4–20 mA dc
Qualified Region
Design Region
0
0
12
13.5
Power Supply (V dc)
35
40
45
50
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H
Span and Zero
Continuously adjustable externally
Zero Elevation and Suppression
Maximum zero elevation: 600% of calibrated span Maximum zero suppression: 500% of calibrated span Zero elevation and suppression must be such that neither the calibrated span
nor the upper or lower range value exceeds 100% of the upper range limit.
Temperature Limits
Normal operating design limits: 40 to 200 °F (4.4 to 93.3 °C) Qualified storage limits: –40 to 120 °F (–40.0 to 48.9 °C)
Humidity Limits
0 to 100% relative humidity (NEMA 4X)
Volumetric Displacement
3
Less than 0.01 in
(0.16 cm3)
Turn-on Ti m e
Two seconds maximum. No warm-up required.
Pressure Ranges
Rosemount 1154DH and 1154HH:
Range Code Pressure Range
4 0–25 to 0–150 inH2O (0–6.22 to 0–37.3 kPa)
5 0–125 to 0–750 inH2O
6 0–17 to 0–100 psi (0–0.12 to 0–0.69 MPa)
7 0–50 to 0–300 psi (0–0.34 to 0–2.07 MPa)
8 0–170 to 0–1,000 psi (Rosemount 1154 D only)
(0–31.08 to 0–186.4 kPa)
(0–1.17 to 0–6.89 MPa)
Rosemount 1154SH:
Range Code Pressure Range
9 0–500 to 0–3,000 psi (0–3.45 to 0–20.68 MPa)
Maximum Working Pressure
Rosemount 1154DH and 1154HH:
Static Pressure Limit
Rosemount 1154SH:
Upper range limit
Static Pressure and Overpressure Limits
Rosemount 1154DH:
0.5 psia to 2,000 psig (3.4 kPa abs to 13.8 MPa) maximum rated static pressure for operation within specifications; overpressure limit is 2,000 psig (13.8 MPa) on either side without damage to the transmitter
6-5
Rosemount 1154 Series H
Rosemount 1154HH:
0.5 psia to 3,000 psig (3.4 kPa abs to 20.7 MPa) maximum rated static pressure for operation within specifications; overpressure limit is 3,000 psig (20.7 MPa) on either side without damage to the transmitter
Overpressure Limits
Rosemount 1154SH:
Operates within specification from 0.5 psia (3.4 kPa abs) to upper range limit; overpressure limit is 4,500 psig (31.0 MPa) for Range Code 9, without damage to the transmitter
Reference Manual
00809-0100-4631, Rev BA
April 2007

PHYSICAL SPECIFICATIONS

Materials of Construction

Isolating Diaphragms:
316L SST
Drain/Vent Valves:
316 SST
Process Flanges:
CF-8M (cast version of 316 SST)
O-rings:
316L SST
Electronics Housing O-rings:
Ethylene propylene
Fill Fluid:
Silicone oil
Flange Bolts:
Plated alloy steel, per ASTM A-540
Electronics Housing:
316 SST
Module Shroud:
304L SST
6-6
Module Shroud Potting:
Silicone RTV
Mounting Bracket:
316L SST
Mounting Bolts:
SAE J429 Carbon steel, Grade 2 or Grade 5

Weight

26.6 lb (12.1 kg) including mounting bracket

Electrical Connections

1
/2–14 NPT conduit with screw terminals

Process Connections

3
/8 in. Swagelok compression fitting, 316 SST (¼–18 NPT optional)
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

ORDERING INFORMATION

Table 6-1. Transmitter Design Specifications.

Model Product Description
115 4 Alphaline Pressure Transmitters for Nuclear Applications (qualified per IEEE Std 323-1974 and IEEE Std 344-1975)
Code Pressure Measurement
D Differential Pressure, 2,000 psig (13.8 MPa) Static Pressure Rating H Differential Pressure, 3,000 psig (20.68 MPa) Static Pressure Rating S Sealed Reference Pressure
Code Series
H Transmitter Enclosed in a Stainless Steel Module Shroud
Pressure Ranges at 68 °F
Code
4 0–25 to 0–150 inH2O
5 0–125 to 0–750 inH2O
6 0–17 to 0–100 psi
7 0–50 to 0–300 psi
8 0–170 to 0–1,000 psi
9 N/A N/A 0–500 to 3,000 psi
Code Output
(1)
R
Code
A Welded 3/8 in. Swagelok compression fitting process connection and welded drain/vent valve
(2)
B
(2)
C
H Welded 3/8 in. Swagelok compression fittings on both process connection and drain/vent connection
Typical Model Number: 1154DH 4 R A
(1) The Rosemount 1154 Series H with the R Output Code Electronics is also available with adjustable damping. Specify adjustable damping by adding “N0037”
to the end of the complete model number, for example: 1154DH4RAN0037.
(2) Customer assumes responsibility for qualifying connection interfaces on these options. Contact Rosemount Nuclear Instruments, Inc. for details.
Rosemount 1154DH (Differential)
(0–6.22 to 0–37.3 kPa)
(0–31.08 to 0–186.4 kPa)
(0–0.12 to 0–0.69 MPa)
(0–0.34 to 0–2.07 MPa)
(0–1.17 to 0–6.89 MPa)
Standard 4–20 mA
Flange Option
¼–18 NPT process connection and welded drain/vent valve ¼–18 NPT process connection and drain hole (drain/vent valve not supplied)
Rosemount 1154HH (Differential) Rosemount 1154SH (Sealed Ref.)
0–25 to 0–150 inH2O
(0–6.22 to 0–37.3 kPa)
0–125 to 0–750 inH2O N/A
(0–31.08 to 0–186.4 kPa)
0–17 to 0–100 psi N/A
(0–0.12 to 0–0.69 MPa)
0–50 to 0–300 psi N/A
(0–0.34 to 0–2.07 MPa)
N/A N/A
N/A
N/A
N/A
N/A
(0-3.45 to 0-20.68 MPa)
6-7
Reference Manual
00809-0100-4631, Rev BA
Rosemount 1154 Series H

Parts List

Rosemount 1154DH, 1154HH, and 1154SH
Spare Parts Category
Traceable Part
Quantity Required
Item Number (see Figure 5-2 on page 5-6)
Part Description Order No. Order No. Order No.
Amplifier Cir. Board, Output Code R Calib. Cir. Board, Output Code R Amplifier Cir. Board with Damping,
Output Code R
Calib. Cir. Board with Damping,
Output Code R
Amplifier Cir Board
Output Code R, N0026
Electronics Housing, Austenitic SST Electronics Cover, Austenitic SST Cover Wrench
Hollow Terminal Block Screw Kit Solid Terminal Block Screw Kit
(Kit contains 20 each)
Valve Stem, 316 SST Valve Stem and Seat Kit, 316 SST
Axial Drain/Vent Valve Kit Quick Disconnect Axial Drain/Vent Valve
Kit
Adjustment Screw Kit
Adjustment Screw (2) Retaining Ring (2) O-ring for Adjustment Screw (2)
O-ring for Electronics Cover
(Kit contains 20 each)
O-ring for Electronics Cover
(Kit contains 1 each)
D.C . 5 5 O-ring Lubricant (0.25 oz) D.C . 5 5 O-ring Lubricant (5.3 oz) Loctite 580 PST Thread Sealant (50 ml) Lubri-Bond® A Cover Lubricant (12 oz.)
Electronics Assembly Hardware
Electronics Screw (3) Nameplate Screw (2) Locknut
Jumper Wire Kit (36 in.) 01153-0055-0001 01153-0055-0001 01153-0055-0001
Panel Mounting Bracket with Bolts Universal Mounting Bracket with Bolts Bolts and Washers for Bracket (Pkg of 4)
Conduit Elbow (M22) Conduit Elbow (½–14 NPT)
(1) Rosemount recommends one spare part or kit for every 25 transmitters in Category “A,” one spare part or kit for every 50 transmitters in Category
“B,” and one spare part or kit for every 5 transmitters in Category “C.” (2) For use with existing N0026 transmitter only. NOTE: Rosemount 1154 Series H Spare Parts not hydrostatic tested nor nuclear cleaned.
(1)
Rosemount
1154D H
6
1
5
1
6
1
5
1
(2)
6 1 A 01154-0001-0006 01154-0001-0006
3
1
1
2
10 2 X A 01153-0277-0001
1 01153-0294-0001 01153-0294-0001 01153-0294-0001 10 12 11
2
2
2
2
1 A 01153-0040-0001 01153-0040-0001 01153-0040-0001
7
14
9
1
1
1
A
01154-0001-0005
A
01154-0002-0001
A
01154-0021-0004
A
01154-0023-0002
01153-0211-0001 01153-0204-0001 01153-0382-0001
01153-0041-0001 01153-0330-0001
01153-0038-0001
01153-0350-0002 01153-0373-0001
CC01153-0039-0001
01153-0039-0003
01153-0053-0001 01153-0248-0001 01153-0329-0001 01153-0333-0001
01153-0013-0001 01153-0013-0003 01153-0321-0001
01154-0035-0001 01154-0040-0001
Rosemount
1154H H
01154-0001-0005 01154-0002-0001
01154-0021-0004
01154-0023-0002
01153-0211-0001 01153-0204-0001 01153-0382-0001
01153-0041-0001 01153-0330-0001
01153-0277-0001 01153-0038-0001
01153-0350-0002 01153-0373-0001
01153-0039-0001
01153-0039-0003
01153-0053-0001 01153-0248-0001 01153-0329-0001 01153-0333-0001
01153-0013-0001 01153-0013-0003 01153-0321-0001
01154-0035-0001 01154-0040-0001
Rosemount
1154S H
01154-0001-0005 01154-0002-0001
01154-0021-0004
01154-0023-0002
01153-0211-0001 01153-0204-0001 01153-0382-0001
01153-0041-0001 01153-0330-0001
01153-0277-0001 01153-0038-0001
01153-0350-0002 01153-0373-0001
01153-0039-0001
01153-0039-0003
01153-0053-0001 01153-0248-0001 01153-0329-0001 01153-0333-0001
01153-0013-0001 01153-0013-0003 01153-0321-0001
01154-0035-0001 01154-0040-0001
April 2007
6-8
Reference Manual
00809-0100-4631, Rev BA April 2007
Rosemount 1154 Series H

SPARE PARTS SHELF LIFE

Store all spare transmitters and spare component parts in accordance with ANSI N45.2.2 level B.
Qualified transmitters, spare circuit boards, spare O-rings: The qualified life (as defined in Qualification Test Report D8700096) plus the shelf life is equal to the typical design life of the plant (40 years) when the ambient storage temperature is below 90 °F.
Lubricants and sealant: The date of the end of shelf life (use by date) is provided with the lubricants and/or sealant, at the time of shipment. The product has a minimum of six months shelf life at the time of shipment.
All other parts: Shelf life is not applicable.

IMPORTANT NOTICE Because of the nuclear use intended for these parts, there are factors to

consider concerning maintenance of qualification and traceability during on-site instrument repair. Rosemount Nuclear Instruments, Inc. rigidly controlled the original assembly of the instrument to ensure that the specifications were met. Since we are not installing the replacement parts in the instruments, Rosemount Nuclear Instruments, Inc. is unable to ensure that the specifications are being satisfied. This responsibility is shifted to the user. The integrity of the instrument as originally assembled is broken.
Replacing parts has ramifications under 10CFR21, for which the user is responsible. These same regulations additionally mandate a component
traceability program, which the user must undertake for the replacement parts. In view of this, and in order to maintain the qualification of the product, the user must ensure that all replacement parts are installed in accordance with the Rosemount Nuclear Instruments, Inc. approved installation and calibration procedures herein.

1. Rosemount 1154 Series H spare parts are not hydrostatic tested nor nuclear cleaned.

2. Part numbers shown may differ from those currently supplied. The part numbers shown are current at the time of printing of this manual, but may be revised in the future. Parts provided are compatible and interchangeable with those listed on your order as to the form, fit, and function of the part required. Please adjust your needs accordingly.
6-9
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
6-10
Reference Manual
00809-0100-4631, Rev BA April 2007

Index

Rosemount 1154 Series H
A
Accuracy . . . . . . . . . . . . . . . . 6-2
Adjustment . . . . . . . . . . . . . . 3-7
Damping . . . . . . . . . . . . . 3-6
Span . . . . . . . . . . . . . . . 3-1
Zero . . . . . . . . . . . . . . . . 3-1
B
BOARD CHECKOUT . . . . . . . 5-2
C
Calibration
Correction for High
Line Pressures
Damping Adjustment . . . . 3-6
Procedure . . . . . . . . . . . . 3-1
Span Adjustment . . . . . . . 3-1
Zero Adjustment . . . . . . . 3-1
Chemical Spray . . . . . . . . . . . 6-1
Cleaning
Nuclear . . . . . . . . . . . . . 6-2
Conduit Installation Procedure . 2-9 Connecting Electrical Housing
to Module Shroud . . . . . . . . . 5-4
Connections
Electrical
Process . . . . . . . . . . . . . 6-6
Considerations
Electrical
Construction
Materials . . . . . . . . . . . . 6-6
Correction
zero Correction For High Line Pressures
Current Control . . . . . . . . . . . 4-4
Current Limit . . . . . . . . . . . . . 4-4
. . . . . . . . . . . . 6-6
. . . . . . . . . . . . 2-4
. . . . . . . . . . . . . . . . 3-8
. . . . . . . . . . . . . . . 3-6
. . 3-6
D
Damping Adjustment . . . . . . . 3-6
Deadband . . . . . . . . . . . . . . . 6-2
Demodulator . . . . . . . . . . . . . 4-3
Differential
transmitter
Disassembly Procedure . . . . . 5-3
Electrical Housing
Removing Module Shroud
Drift . . . . . . . . . . . . . . . . . . . 6-2
. . . . . . . 2-3, 2-4
Disassembly
From Electrical Housing
. . . 5-3
. . . . . . . 5-4
E
Electrical Connection Procedure
Electrical Connections . . . . . . 6-6
Electrical Housing
Electrical Housing Disassembly 5-3 Electrical Housing Reassembly 5-5 Electrical Installation
Considerations . . . . . . . . . . . 2-4
Errors
. . . . . . . . . . . . . . 2-9
Connecting to
Module Shroud
minimizing in
measurements
. 5-4
. . 2-4
F
Flow . . . . . . . . . . . . . . . . . . . 2-3
Functional Specifications . . . . 6-4
Humidity Limits . . . . . . . . 6-5
Output . . . . . . . . . . . . . . 6-4
Power Supply . . . . . . . . . 6-4
Service . . . . . . . . . . . . . 6-4
Span and Zero . . . . . . . . 6-4
Temperature Limits . . . . . 6-5
Turn-On Time . . . . . . . . . 6-5
Volumetric Displacement . 6-5
G
Gas . . . . . . . . . . . . . . . . . . . 2-4
H
Header Board Connections . . 5-2
Humidity Limits . . . . . . . . . . . 6-5
Hydrostatic Testing . . . . . . . . 6-2
I
Impulse . . . . . . . . . . . . .2-3, 2-4
Installation
Conduit Installation
Procedure
. . . . . 2-9
J
Jumper Wire Placement . . . . . 3-5
L
Level . . . . . . . . . . . . . . . . . . 2-4
Limits
Humidity . . . . . . . . . . . . 6-5
Temperature . . . . . . . . . 6-5
Linearity Adjustment . . . .3-5, 4-3
Load Effect . . . . . . . . . . . . . . 6-4
M
Maintenance
Disassembly Procedure
Post-Assembly Tests . . . . 5-5
Reassembly Procedure . . . 5-4 Sensing Module Checkout .5-2
Test Terminals . . . . . . . . . 5-2
Troubleshooting . . . . . . . .5-7
Materials of Construction . . . . . 6-6
Mechanical Considerations
Mounting Procedure . . . . .2-2
Process Connections . . . . 2-2
Module Shroud
Removal . . . . . . . . . . . . .5-4
Mounting Position Effect . . . . .6-4
Mounting Procedure . . . . . . . . 2-2
. .5-3
N
Nuclear Cleaning . . . . . . . . . . 6-2
Nuclear Specifications
Chemical Spray
Hydrostatic Testing . . . . . .6-2
Nuclear Cleaning . . . . . . .6-2
Post DBE Operation . . . . .6-1
Qualified Life . . . . . . . . . .6-2
Quality Assurance
Radiation . . . . . . . . . . . . .6-1
Seismic . . . . . . . . . . . . . . 6-1
Steam
Pressure/Temp
Traceability . . . . . . . . . . . 6-2
Program
. . . . . . . .6-1
. . . . . . . 6-1
. . . . . 6-1
O
Operation
Current Control
Current Limit . . . . . . . . . .4-4
Demodulator . . . . . . . . . .4-3
Linearity Adjustment . . . . . 4-3
Oscillator . . . . . . . . . . . . .4-4
Reverse Polarity
The -Cell Sensor . . . . . .4-3
Voltage Regulator . . . . . .4-4
Zero and Span
Oscillator . . . . . . . . . . . . . . . . 4-4
Output . . . . . . . . . . . . . . . . . .6-4
Overpressure Effect . . . . . . . . 6-3
. . . . . . . .4-4
Protection
Adjustments . . . .4-4
. . . . . . 4-4
Index-1
Rosemount 1154 Series H
Reference Manual
00809-0100-4631, Rev BA
April 2007
P
Performance Specifications . . . 6-2
Accuracy . . . . . . . . . . . . 6-2
Deadband . . . . . . . . . . . . 6-2
Drift . . . . . . . . . . . . . . . . 6-2
Load Effect . . . . . . . . . . . 6-4
Mounting Position Effect . . 6-4
Overpressure Effect . . . . . 6-3
Power Supply Effect . . . . 6-4
Static Pressure Span Effect 6-4 Static Pressure Zero Effect 6-3
Temperature Effect . . . . . 6-3
Physical Specifications . . . . . . 6-6
Electrical Connections . . . 6-6 Materials of Construction . 6-6
Process Connections . . . . 6-6
Weight . . . . . . . . . . . . . . 6-6
Post DBE Operation . . . . . . . . 6-1
Post-Assembly Tests . . . . . . . 5-5
Potentiometer . . . . . . . . . . . . 3-8
Power Supply . . . . . . . . . . . . 6-4
Power Supply Effect . . . . . . . . 6-4
Preliminary . . . . . . . . . . . . . . 5-4
Process
connections . . . . . . 2-2, 3-8
loosening and
retightening
line . . . . . . . . . . . . . . . . 2-3
taps . . . . . . . . . . . . . . . . 2-4
Process Connections . . . 2-2, 6-6
Purging . . . . . . . . . . . . . . . . . 2-4
. 2-3
Q
Qualified Life . . . . . . . . . . . . . 6-2
Quality Assurance Program . . 6-1
R
Radiation . . . . . . . . . . . . . . . 6-1
Reassembly Procedure . . . . . 5-4
Connecting Electrical
Housing to Module Shroud
Electrical Housing
Reassembly
Preliminary . . . . . . . . . . . 5-4
Removing Module Shroud From Electrical Housing Reverse Polarity Protection . . . 4-4
. . . . . . . . . . 5-5
. . . . . . . . . 5-4
5-4
Shelf Life, Spare Parts . . . . . . 6-9
Signal
terminals
Span Adjustment . . . . . . . . . . 3-1
Span and Zero . . . . . . . . . . . 6-4
Specifications and Reference Data
Functional Specifications Performance Specifications 6-2
Physical Specifications . . 6-6 Static Pressure Span Effect . . 6-4 Static Pressure Zero Effect . . . 6-3 Steam Pressure/Temperatures 6-1
Swagelok Tube Fittings . . . . . 2-3
installing . . . . . . . . . . . . 2-3
more information about . . 2-3
. . . . . . . . . . . . 2-9
. 6-4
T
Temperature . . . . . . . . . .2-3, 2-4
storage . . . . . . . . . . . . . 6-9
Temperature Effect . . . . . . . . 6-3
Temperature Limits . . . . . . . . 6-5
Terminals
Test . . . . . . . . . . . . . . . . 5-2
Test Terminals . . . . . . . . . . . 5-2
The -Cell Sensor . . . . . . . . . 4-3
Torque Specifications
table . . . . . . . . . . . . . . . . . . 5-6
terminal screws
Traceability . . . . . . . . . . .6-2, 6-9
Transmitter
terminal block
Troubleshooting . . . . . . . . . . 5-7
Turn-On Time . . . . . . . . . . . . 6-5
. . . . . . . 2-9
. . . . . . . . 2-10
V
Voltage Regulator . . . . . . . . . 4-4
Volumetric Displacement . . . . 6-5
W
Weight . . . . . . . . . . . . . . . . . 6-6
Z
Zero . . . . . . . . . . . . . . . . . . . 3-8
correction . . . . . . . . . . . . 3-8
shift . . . . . . . . . . . . . . . . 3-8
Zero Adjustment . . . . . . . . . . 3-1
Zero and Span Adjustments . . 4-4
S
Seal . . . . . . . . . . . . . . . . . . . 2-3
Sediment
in impulse tubing
preventing deposits . . . . . 2-3
Seismic . . . . . . . . . . . . . . . . . 6-1
Sensing Module Checkout . . . 5-2
Service . . . . . . . . . . . . . . . . . 6-4
Index-2
. . . . . . . 2-3
Reference Manual
00809-0100-4631, Rev BA
April 2007
The Emerson logo is a trade mark and service mark of Emerson Electric Co. Rosemount, the Rosemount logotype, and Alphaline are registered trademarks of Rosemount, Inc. δ−Cell is a trademarks of Rosemount, Inc.
Swagelok is a registered trademark of Swagelok Co.
Emerson Process Management
Rosemount Nuclear Instruments, Inc.
8200 Market Boulevard Chanhassen, MN 55317 USA T (952) 949-5210 F (952) 949-5201
Rosemount Nuclear Instruments, Inc. satisfies all obligations coming from legislation to harmonize product requirements in the European Union.
¢00809-XXXX-XXXX ¤
www.emersonprocess.com/rosemount/nuclear
Loading...