Emerson Quick Start Guide: Emerson Location Awareness Manuals & Guides

00825-0100-4542, Rev AB
Emerson Location Awareness
with WirelessHART® Protocol
Quick Start Guide
July 2022
Quick Start Guide July 2022
Emerson Location Awareness safety messages
Emerson has two toll-free assistance numbers and one international number.
NOTICE
Customer Central
1 800 999 9307 (7:00 a.m. to 7:00 p.m. CST)
National Response Center (equipment service needs)
1 800 654 7768 (24 hours a day)
International
1 952 906 8888
Nuclear qualification
The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings. For information on Rosemount™ nuclear-qualified products, contact an Emerson Sales Representative.
Wireless Gateway
The Location Anchor and all other wireless devices should be installed only after the Wireless Gateway has been installed and is functioning properly. Wireless devices should also be powered up in order of proximity from the Wireless Gateway, beginning with the closest. This will result in a simpler and faster network configuration. Use caution when making changes to the TCP/IP network settings. If they are lost or entered incorrectly, the Gateway will require a factory reset. Contact your network administrator for information on the proper TCP/IP network settings.
Hazardous substance
Individuals handling products exposed to hazardous substances can avoid injury if they are informed of and understand the hazard. If the returned product was exposed to a hazardous substance as defined by OSHA, a copy of the Material Safety Data Sheet (MSDS) for each hazardous substance identified must accompany the returned goods.
Shipping considerations
The Location Anchor unit was shipped with the battery installed. Each Location Anchor contains one “D” size primary lithium battery. Primary lithium batteries are regulated in transportation by the U. S. Department of Transportation and are also covered by the International Air Transport Association (IATA), International Civil Aviation Organization (CAO), and European Ground Transportation of Dangerous Goods (ARD). It is the shipper’s responsibility to ensure compliance with these or any other local requirements. Consult current regulations and requirements before shipping. The polymer enclosure has surface resistivity greater than one giga-ohm. Take care during transportation to and from the configuration point to prevent a potential electrostatic charging hazard.
2 Emerson.com
July 2022 Quick Start Guide
NOTICE
Battery considerations
Battery hazards remain when cells are discharged. Electrostatic discharge can damage electronics. Batteries should be stored in a clean dry area. For maximum battery life, storage temperature should not exceed 30 °C (86 °F). The Location Anchor battery shall only be replaced in a non-hazardous area. The Personnel Tag battery is rechargeable and shall only be charged in a non-hazardous area using a Qi certified charger.
Location Awareness tags and anchors use lithium batteries and are subject to local, state, national, and/or international requirements. Lithium batteries may ignite, explode or release toxic and harmful materials or chemicals if mishandled. Do not use a battery that is damaged. Do not incinerate batteries. Do not crush, disassemble, heat batteries above 100 °C (212 °F) or expose contents to water. Please check and comply with all applicable local, state, national, and international regulations before transporting, disposing or recycling lithium batteries. Do not recharge anchor battery. Do not use Anchor battery that has been dropped. Use of damaged equipment can be dangerous and could result in injury. Properly dispose of damaged Anchor batteries or Tags immediately. Do not place the Location Awareness Anchor, anchor battery, or the Location Tag device near fire or heat to reduce the risk of explosion and possible injury.
Installation considerations
Only qualified personnel should perform the installation. Verify the operating atmosphere of the transmitter is consistent with applicable hazardous area certifications. Reference the Product Certifications section of this manual for information regarding hazardous certification. This Location Anchor must be installed to ensure a minimum antenna separation distance of 20 cm (8 in.) from all persons.
WARNING
Failure to follow these installation guidelines could result in death or serious injury.
Ensure only qualified personnel perform installation or service.
Explosions could result in death or serious injury.
Before connecting a handheld communicator in an explosive atmosphere, ensure the instruments are installed in accordance with intrinsically safe or non-incendive field wiring practices.
Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.
Electrical shock could cause death or serious injury.
Physical access
Unauthorized personnel may potentially cause significant damage to an/or misconfiguration of end users' equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end user's assets. This is true for all systems used within the facility.
Contents
Overview......................................................................................................................................5
Installation Plantweb Insight and Plantweb Insight Location application...................................... 7
Quick Start Guide 3
Quick Start Guide July 2022
Location Anchor configuration..................................................................................................... 8
Installation Location Anchor.......................................................................................................11
Location Anchor commissioning................................................................................................ 14
Operation and maintenance.......................................................................................................17
Personnel Tag configuration.......................................................................................................19
Personnel Tag mounting considerations.................................................................................... 20
Personnel Tag operation and maintenance................................................................................ 21
Product certifications................................................................................................................. 23
Declaration of Conformity..........................................................................................................30
4 Emerson.com
July 2022 Quick Start Guide

1 Overview

WARNING
Location Awareness devices and services are meant to augment, not replace, safety and security measures within a plant and must not be used as the sole method of locating or tracking personnel within a customer’s facility. Even when properly installed and maintained, and when working as intended, location position information will only be provided every 32 seconds (or longer depending on system configuration). Location Awareness devices and services cannot detect if a user has fallen.
In addition, structures within a plant, including moveable heavy equipment or other changes made after the initial installation, can prevent or delay detection and impact accuracy. Location Awareness devices and services can also experience a loss or delay of signal as a result of software or hardware malfunction, or as a result of interruption or limitation of services by third party networks. Inform employees or contractors about the features and limitations of the Location Awareness System before allowing use.

1.1 Location Awareness solution components

Location Anchor
Quick Start Guide 5
Quick Start Guide July 2022
Location Personnel Tag
Plantweb™ Insight Location application
1.2
6 Emerson.com

Product recycling/disposal

Consider recycling equipment and packaging. Dispose of the product and packaging in accordance with local and national legislation.
July 2022 Quick Start Guide
2 Installation Plantweb Insight and Plantweb
Insight Location application

2.1 Install Plantweb Insight and Plantweb Insight Location application

Follow instructions in the Plantweb Insight Quick Start Guide.

2.2 Add-ons and tag licensing

The Plantweb Insight application will be provided with as-purchased features and add-ons.

2.3 Location Awareness App configuration

See Location Awareness User Manual.
Quick Start Guide 7
Quick Start Guide July 2022

3 Location Anchor configuration

3.1 Safety messages

WARNING
Failure to follow these installation guidelines could result in death or serious injury.
Ensure only qualified personnel perform the installation.
Explosions could result in death or serious injury.
Before connecting a handheld communicator in an explosive atmosphere, ensure the instruments are installed in accordance with intrinsically safe or non-incendive field wiring practices.
Verify that the operating atmosphere of the transmitter Anchor and Tag is consistent with the appropriate hazardous locations certifications.
Electrical shock could cause death or serious injury.
Use extreme caution when making contact with the leads and terminals. The Anchor enclosure has a high surface resistivity that may present a static
hazard. Take care during transportation and installation.

3.2 Location Anchor configuration

Remove the Location Anchor electronics cover and connect to the HART Communication terminals for configuration.
The Emerson Location Anchor will receive HART Communication from any HART handheld Field Communicator or AMS Device Manager.
AMS Device Manager
AMS Device Manager can connect to devices directly, using a HART modem, or remotely using the Gateway. To configure the Emerson Location Anchor, double click (or right click and select Configure/Setup) on the device icon that will appear below either the HART modem or Gateway connection tree.
3.3
8 Emerson.com

Device network configuration

To communicate with the Gateway (and ultimately the host system), the Location Anchor must be configured to communicate with the wireless network. Using a Field Communicator or AMS Device Manager, enter the network ID and join key so they match the network ID and join key of the Gateway and other devices in the network. If they do not match, the
®
July 2022 Quick Start Guide
Location Anchor will not communicate with the network. The network ID and join key may be obtained from the Gateway on the Systems Settings Network Network Settings page on the web server, shown in Figure 3-1.
Figure 3-1: Gateway Network Settings
AMS Device Manager
Right click on the Location Anchor and select Configure. When the menu opens, select Join Device to Network and follow the method to enter the network ID and join key. See Table 3-1 for menu items.
Table 3-1: Device Configuration
Function Menu items
Join device to network Network ID, Set Join Key
Device information Tag, Long Tag, Descriptor, Message,
Guided setup Basic Setup, Join Device to Network,
Manual setup Wireless, Location Network, HART,
Quick Start Guide 9
Date
Configure Update Rates
Security, Device Information, Power
Quick Start Guide July 2022
Table 3-1: Device Configuration (continued)
Function Menu items
Wireless Network ID, Join Device to Network,
Broadcast Information

3.4 Power down

After the Location Anchor has been configured, turn off power to the electronics by moving the power switch to off and reinstall the cover. Power should only be turned back on when the anchor is ready for commissioning.
10 Emerson.com
July 2022 Quick Start Guide

4 Installation Location Anchor

4.1 Safety messages

WARNING
Failure to follow these installation guidelines could result in death or serious injury.
Ensure only qualified personnel perform the installation.
Explosions could result in death or serious injury.
Before connecting a handheld communicator in an explosive atmosphere, ensure the instruments are installed in accordance with intrinsically safe or non-incendive field wiring practices.
Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.
Electrical shock could cause death or serious injury.
Use extreme caution when making contact with the leads and terminals. The Anchor enclosure has a high surface resistivity that may present a static
hazard. Take care during transportation and installation.

4.2 Wireless considerations

Power-up sequence
The Wireless Gateway should be installed and functioning properly before any wireless devices are powered. Battery is already installed in the Location Anchor. Move power switch to "ON" position to power up. This results in a simpler and faster network configuration. Enabling Active Advertising on the Gateway ensures new devices are able to join the network faster. For more information, see the Wireless Gateway Reference Manual.
Antenna position
The antenna is internal to the Location Anchor. To achieve optimal range and performance, position the Location Anchor with the stem in a horizontal position as shown in Figure 4-1. Good connectivity can also be achieved in other orientations. The antenna should also be approximately 1 m (3 ft.) from any large structure, building, or conductive surface, and a minimum of 2 m (6 ft.) and maximum of 4 m (13 ft.) from the ground to allow for clear communication to other devices. Refer to best practices for additional information on optimal mounting locations of the Location Anchor.
Quick Start Guide 11