Emerson NGA2000 PMD User Manual

Instruction Manual
760007-A July 2003
Model NGA2000 PMD
Paramagnetic Detector Analyzer Module
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and interna­tional standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions
prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruc- tion Manual and per applicable local and national codes. Connect all products to the proper elec-
trical and pressure sources.
To ensure proper performance, use qualified personnel
to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replace ment parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY
. Look-alike substitutions may result
in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Viton-A® is a registered trademark of E.I. duPont de Nemours & Co. Paliney No. 7™ is a trademark of J.M. Ney Co.
Emerson Process Management Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model NGA2000 PMD

TABLE OF CONTENTS

PREFACE...........................................................................................................................................P-1
Intended Use Statement.....................................................................................................................P-1
Definitions...........................................................................................................................................P-1
Safety Summary.................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
Compliances.......................................................................................................................................P-5
Glossary Of Terms ...........................................................................................................................P-6
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1
Overview................................................................................................................................1-1
1-2 Typical Applications...............................................................................................................1-1
1-3 Theory Of Technology...........................................................................................................1-1
1-4 Features.................................................................................................................................1-2
1-5 Specifications.........................................................................................................................1-3
a. General ...........................................................................................................................1-3
b. Sample............................................................................................................................1-3
c. Physical...........................................................................................................................1-3
Instruction Manual
760007-A July 2003
2-0 INSTALLATION ....................................................................................................................2-1
2-1 Unpacking..............................................................................................................................2-1
2-2 Assembly...............................................................................................................................2-1
2-3 Location.................................................................................................................................2-2
2-4 Gases ....................................................................................................................................2-2
a. Requirements..................................................................................................................2-2
b. Connections....................................................................................................................2-3
c. Leak Test ........................................................................................................................2-3
2-5 Electrical Connections...........................................................................................................2-6
3-0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2 Displays.................................................................................................................................3-1
a. Run Mode Display...........................................................................................................3-1
b. Menu Displays.................................................................................................................3-1
c. Help Displays..................................................................................................................3-1
3-3 Startup Procedure .................................................................................................................3-5
3-4 Binding...................................................................................................................................3-5
3-5 Calibration..............................................................................................................................3-5
3-6 Background Gas Compensation ...........................................................................................3-8
3-7 Barometric Pressure Compensation .....................................................................................3-11
4-0 MAINTENANCE AND SERVICE ..........................................................................................4-1
4-1 Overview................................................................................................................................4-1
4-2 Printed Circuit Board Replacement.......................................................................................4-1
4-3 Module Fan Replacement.....................................................................................................4-2
4-4 Thermal Fuse Replacement..................................................................................................4-3
4-5 Flow Sensor Replacement ....................................................................................................4-4
4-6 Power Fuse Replacement.....................................................................................................4-4
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
760007-A July 2003
5-0 REPLACEMENT PARTS......................................................................................................5-1
5-1 Matrix.....................................................................................................................................5-1
5-2 Replacement Parts................................................................................................................5-2
6-0 RETURN OF MATERIAL......................................................................................................6-1
6-1 Return Of Material.................................................................................................................6-1
6-2 Customer Service..................................................................................................................6-1
6-3 Training..................................................................................................................................6-1
7-0 APPENDIX A. MENU DISPLAYS.........................................................................................7-1
Figure 1-1. Spherical Body in Non-Uniform Magnetic Field..................................................... 1-1
Figure 1-2. Trace Oxygen Detector Coulometric Principle........................................................ 1-2
Figure 2-1. Analyzer Module Installation Into Instrument Platform.......................................... 2-1
Figure 2-2. PMD Front Panel Connections.............................................................................. 2-4
Figure 2-3. PMD Back Panel Connections............................................................................... 2-4
Figure 2-4. Interconnection of Typical Gas Manifold to PMD Analyzer Module ...................... 2-5
Figure 2-5. PMD Wiring Diagram............................................................................................. 2-6
Figure 2-6. PMD Outline and Mounting Dimensions................................................................ 2-7
Figure 3-1. Run Mode Display ................................................................................................. 3-3
Figure 3-2. Main Menu Display ................................................................................................ 3-3
Figure 3-3. Basic Controls Menu.............................................................................................. 3-3
Figure 3-4. Expert Controls and Setup Menu........................................................................... 3-4
Figure 3-5. Technical Level Configuration Menu..................................................................... 3-4
Figure 3-6. Typical Help Screen............................................................................................... 3-4
Figure 4-1. PMD Module – Major Components ....................................................................... 4-1
Figure 4-2. Module Fan Assembly........................................................................................... 4-2
Figure 4-3. Detector Assembly................................................................................................. 4-3
Model NGA2000 PMD

LIST OF ILLUSTRATIONS

LIST OF TABLES

Table 3-1. PMD Analyzer Module Alarms............................................................................... 3-2
Table 3-2. Calibration Range for Various Zero Based Operating Ranges ............................. 3-6
Table 3-3. Calibration Range for Various Suppressed Range Operations.............................3-6
Table 3-4. Oxygen Equivalents of Common Gases.............................................................. 3-10
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 PMD

PREFACE

INTENDED USE STATEMENT

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the Model NGA2000 PMD and the System Accessories of the NGA2000 System.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
760007-A July 2003
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

SAFETY SUMMARY

Some sections may describe equipment not used in your configuration. The user should become thor­oughly familiar with the operation of this module before operating it. Read this instruction manual com­pletely.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
DANGER.
POSSIBLE EXPLOSION HAZARD
This equipment is not designed and should not be used in the analysis of flammable samples. Use of this equipment in this way could result in explosion and death.
NOTE
Apply leak test liquid to cell or detectors only as a last resort.
WARNING
POSSIBLE EXPLOSION HAZARD
Verify that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death. See Section 2-4c on page 2-3 for Leak test procedure.
WARNING.
HIGH PRESSURE GAS CYLINDERS
This analyzer requires use of pressurized gas. See General Precautions for Handling and Storing High Pressure Cylinders, page P-4.
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 PMD
WARNING
OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical po wer supply, adding Rosemo unt Analytical PN 90331 Current Protector in series with the 24 V positive line will prevent over-voltage spiking and resultant fuse flowing when powering up the instrument.
CAUTION
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.
CAUTION
OVERBALANCE HAZARD
760007-A July 2003
This Analyzer Module may tip instrument over if it is pulled out too far and the Platform is not prop­erly supported.
NOTICE
Software compatibility is necessary for all NGA2000 components in your system to work together. The version of your Platform's software must be equal to or greater that the version of any other module(s) for successful compatibility. If it is not, contact Rosemount Analytical at 800-433-6076 to order software upgrade kit PN 657150 for the Platform.
You can locate the version of each NGA2000 component as follows: Platform Controller Board
Turn power ON.
The display will show "Control Module V2. ...". This is the software version.
Analyzer Module Located on the right side of the Analyzer Module case.
I/O Module Located on the backplane connector of the module. If no label is present, the module is Version
2.0.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Com­pressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
°
F (52°C). A flame should
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
9
6
Model NGA2000 PMD

DOCUMENTATION

The following NGA2000 PMD instruction materials are available. Contact Customer Service Center or the local representative to order.
760007 Instruction Manual (this document)

COMPLIANCES

This product may carry approvals from several certifying agencies, including Factory Mutual and the Cana­dian Standards Association (which is also an OSHA accredited, Nationally Recognized Testing Labora­tory), for use in non-hazardous, indoor locations.
760007-A July 2003
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe.
These products comply with the standard level of NAMUR EMC. Recommendation (May 1993).
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
NRTL /C
®
97-C219
NAMUR
N
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

GLOSSARY OF TERMS

Analyzer Module
The module that contains all sensor/detector components for development of a Primary Variable signal; in­cludes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and net­work cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The computer board that serves as the Network Manager and operates the Display and Keypad.
Distribution Assembly
The Backplane and the card cages that hold I/O and Expansion Modules.
Expansion Module
A circuit board that plugs into the Backplane from the front of the Platform and performs special features not related to I/O functions.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function.
Operator Interface
The Display and Keyboard.
Platform
Any workable collection of the following: Controller Board, Power Supply, Distribution Assembly, Enclosure and Operator Interface.
Power Supply
Any of a variety of components that provides conditioned power to other NGA2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, source voltage and other diagnostic information.
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 PMD
Softkeys
The five function softkeys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s).
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD

DESCRIPTION AND SPECIFICATIONS

Instruction Manual
760007-A July 2003
SECTION 1

1-1 OVERVIEW

This manual describes the Paramagnetic De­tector (PMD) Analyzer Module of Rosemount Analytical's NGA2000 Series of gas analysis components.
The PMD Analyzer Module is designed to continuously determine the concentration of oxygen in a flowing gaseous mixture. The concentration is expressed in ppm or percent volume O
2.
The entire Analyzer Module is designed as a slide-in module (if configured in stand-alone instrument fashion), removable from the front of the Platform, with gas connections made from the rear. All electronics relative to sam­ple detection and conditioning are included in this module.

1-2 TYPICAL APPLICATIONS

PMD Analyzer Module applications include:
process control
continuous emissions monitoring sy st ems
(CEMS)
industrial gas production
fermentation process monitoring

1-3 THEORY OF TECHNOLOGY

Oxygen is strongly paramagnetic (i.e., capa­ble of becoming a temporary magnet when placed in a magnetic field) while most other common gases are weakly diamagnetic (i.e., tend to be non-magnetic). See Figure 1-1 below.
The Magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet as­sembly. As shown in Figure 1-2 on page 1-2, a dumbbell shaped, nitrogen-filled, hollow gas test body is suspended on a platinum/nickel alloy ribbon in a non-uniform magnetic field.
Because of a "magnetic buoyancy" effect, the spheres of the test body are subjected to dis­placement forces, resulting in a displacement torque proportional to the magnetic suscepti­bility of the gas surrounding the test body.
Measurement is accomplished by a null­balance system, whereby the displacement torque is opposed by an equal and opposite restorative torque. The restoring current is automatically maintained at the correct level by an electro-optical feedback system. A beam of light from the source LED is reflected off the square mirror attached to the test body onto a bi-cell (dual photodiode).
The current required to keep the test body to the null position is a linear function of the total magnetic susceptibility of the sample gas.
See Figure 4-1 on page 4-1, Figure 4-2 on page 4-2, and Figure 4-3 on page 4-3 for component configuration.
Shaded Pole Piece
Sphere (Magnetic Susceptibility = ko )
F
k
Sample Gas (Magnetic Susceptibility = k )
As percentage of oxygen in sample gas increases, displacement force (F
Note:
) increases.
k
Figure 1-1. Spherical Body in Non-Uniform
Magnetic Field
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
q
A
A
q
SS
760007-A July 2003
Model NGA2000 PMD

1-4 FEATURES

Among the features incorporated into the PMD Analyzer Module is a flow splitter (≈2:1) that allows for greater sample flow, decreased lag time and faster analyzer response.
Balancing Weight
Electromagnetic
xis
Displacement
Restoring
Current
Restoring Current
Current Loop
Mirror
Platinum/Nickel Alloy Suspension Ribbon
TEST BODY DETAIL
Tor
Restoring Torque
The "Time Alignment" feature can be used to delay sending the Primary Variable from the PMD Analyzer Module for up to 30 seconds in
0.1 second intervals. This feature allows Pri­mary Variables form more than one PMD Analyzer Module to be "time aligned" if nec­essary.
ue
Electromagnetic
xis
Balancing Weight
Nitrogen-Filled Hollow Gl ass Test Body
Shaded Pole Pieces (4)
Displacement
Tor
ue
Restoring Torque
Magnet
Test Body
LED Source
Dual Photocell
NULL BALANCE SYSTEM
Restoring
DETECTOR/MAGNET
A
EMBLY
SIGNAL CONDITIONING
Zero
ASSEMBLY
Span
Figure 1-2. Trace Oxygen Detector Coulometric Principle
SIGNAL OUTPUT TO NETWORK
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003

1-5 SPECIFICATIONS

a. General

Measurement Species................... Oxygen
Ranges.......................................... 0 to 100% oxygen; four fullscale selections, including suppressed
Repeatability.................................. ±1% of fullscale (at constant temperature)
Minimum Detectable Level............ 0.01% oxygen
Noise ............................................. <1% of fullscale, peak-to-peak; <±1% for suppressed ranges
Linearity......................................... ±1% of fullscale
Response Time ............................. 0 to 90% of fullscale in 20 seconds (±2 seconds)
Drift (Zero and Span)..................... <±1% of fullscale/24 hours, <±2% of fullscale/week at constant
Effect of Temperature.................... <±1% of fullscale over any 10°C interval for rate of change no
Environment .................................. Location - Class B controlled, indoor, non-hazardous
Ambient Temperature.................... 0 to 45°C (32 to 113°F)
Effect of Flow................................. <±1% of range when sample flow rate is changed by 20 ml/min.
Power Requirements..................... 24 VDC ±5%, 50 W max.; ripple and noise: <100 mV peak-to-peak;

b. Sample

1
zero ranges in 1% increments
temperature;
<±2% of fullscale/24 hours, <±4% of fullscale/week of range for 99 to 100% (at constant temperature)
greater than 10°C per hour
line and load regulations: <±1%
Temperature.................................. Non-flammable;: 10 to 66°C (50 to 150°F)
Flow Rate ...................................... 800 to 1400 ml/min.
Exhaust Pressure.......................... -345 to 690 hPa-gauge (-5 to 10 psig)
Particles......................................... filtered to <2 microns
Dewpoint........................................ below 43°C (110°F), no entrained liquid
Materials in Contact with Sample.. Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy resin, Vi-
ton A, platinum, nickel, rhodium and MgF
Sample Humidity ........................... non-condensing at ambient temperatures

c. Physical

Case Classification........................ General purpose for installation in weather-protected areas
Dimensions.................................... See Outline and Mounting Dimensions, Figure 2-6 on page 2-7
Weight ........................................... 8 kg (17.6 lbs.
Mounting........................................ Inside a Platform or custom-installed in a panel
Maximum Length of LON Cable.... 1600 m (1 mile) between Analyzer Module and Platform
1
See the Platform manual for specifications regarding Platform related components.
2
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A

2-1 UNPACKING

Instruction Manual
760007-A July 2003
SECTION 2

INSTALLATION

WARNING
If the Paramagnetic Analyzer Module is re­ceived as a separate unit, carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping car­rier if the carton or contents is damaged. Re­tain the carton and packing material until all components associated with the Analyzer Module are operational.

2-2 ASSEMBLY

If the Analyzer Module requires assembly with other components (e.g., the Platform and as­sociated I/O Modules), do so at this time. Fol­lowing the guides on the bottom left and bottom center of the Platform, carefully slide the Analyzer Module halfway into place.
HAND INJURY HAZARD
Do not place hands or fingers in the Plat­form front handles when front panel is open. Dropping the front panel of the Plat­form while hand or fingers are inside either handle can cause serious injury.
Lift the spring-loaded pins on the front of the Analyzer Module, and carefully slide it the rest of the distance. Secure the module in position by releasing the pins, which seat in the avail­able holes in the bottom of the case (see Figure 2-1 below). If the module and Platform are difficult to assemble, remove the module, ensure the top cover of the module is firmly seated on the hold-down screws, and repeat the assembly procedure.
Install I/O Module(s) according to guidelines in the I/O manual. After startup and calibration have been performed, secure the front panel with the six screws provided.
Figure 2-1. Analyzer Module Installation Into Instrument Platform
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
NALYZER MODULE GUIDES
PIN SEATS
DISENGAGED FRONT PANEL
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

2-3 LOCATION

Install the Analyzer Module in a clean, non­hazardous, weather protected, vibration free location free from extreme temperature varia­tions. For best results, either install the mod­ule near the sample stream to minimize sample transport time or supply a flow greater than necessary and route only the appropriate amount through the Analyzer Module.
Observing these requirements are critical. Note the following:
Excessive vibration can cause a noisy readout. To minimize vibration effects, the detector/magnet assembly is envel­oped in a shock-mounted compartment.
The user should ensure, when making any internal electrical connections, that no cables are placed in contact with the detector assembly or associated internal sample inlet and outlet tubing.
An oxygen-free gas, typically nitrogen, is required for use as the zero standard gas. Recommendations for span calibration gases, bases on various operating ranges, are tabulated in Table 3-4 on page 3-10. Air (20.93% oxygen) can be used as span gas regardless of the ranges used for sampling, although very low ranges may lose accuracy.
Sample Gas
Sample gas should be non-flammable.
Temperature
Sample temperature at the inlet should be from 50°F to 150°F (10°C to 66°C). A maximum entry temperature of 110°F (43°C) is recommended to prevent cool­ing of the sample and possible internal condensation. Such condensation could damage some components of the Ana­lyzer Module. This recommendation can be ignored if a thoroughly dry sample is examined.
Magnetic susceptibilities and partial pressures of gases vary with tempera­ture. Permissible ambient temperature range is 32°F to 113°F (0°C to 45°C).
The interior of the Detector Assembly is maintained at approximately 144°F (62°C) by an electronically controlled heater. Prior to entering the detector assembly, the sample is heated in a coiled tubing to match the detector's temperature.

2-4 GASES

a. Requirements

Calibration Gases
Analyzer Module calibration requires the establishment of zero and span calibra­tion points. This requires a zero standard gas to set the zero point span gas to es­tablish a calibration point at or near the upper range limit.
Pressure
Sample exhaust pressure limits are -5 to 10 psig (-345 to 690 hPa-gauge). Normal operation is in the positive range, be­tween 0 and 10 psig (0 and 690 hPa­gauge). Negative gauge pressures are not normally recommended, but may be used in certain special applications.
To prevent over-pressurization, insert a pressure relief valve into the sample inlet line. A check valve should also be placed in the outlet line if the Analyzer Module is connected to a manifold associated with a flare or other apparatus that does not op­erate at atmospheric pressure.
The outlet port is commonly vented to the atmosphere. Any change in barometric pressure has a directly proportional effect on the indicated percent of oxygen, and should be neutralized through manual or computer correction of data. Note the fol­lowing example:
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003
Range = 0% to 5% oxygen Barometric pressure change after
calibration = 1% Analyzer Module measurement = 5%
oxygen Measurement error = 0.01 x 5% oxy-
gen Fullscale span = 5% oxygen
0.05% oxygen error = 1% of fullscale
The error is more significant for sup­pressed range 99% to 100%.
An optional barometric pressure compensation board is available to automatically perform this correction.
A general rule regarding calibration gas pressure is that it should be the same as the expected sample gas pressure during routine operation.
The above requirement increases the dif­ficulty of operation at negative gauge pressure. A suction pump can be con­nected to the outlet port for drawing sam­ple through the Analyzer Module. Such operation necessitates special precau­tions to ensure accurate readout, includ­ing the following:
The need for equilibrium between sample and gas calibration pressures.
Any leakage in the sample handling sys­tem will decrease readout accuracy.
Flow Rate
Recommended sample flow rate is 800 to 1400 ml/min., ±40 ml/min. Optimum flow rate is 1100 ml/min.
If flow is held to within tolerance and op­erating pressure remains constant, zero and span drift will meet specified limits.

b. Connections

(See Figure 2-3 on page 2-4) Connect inlet and outlet lines for sample gas to appropriately labeled fittings on the rear panel. Both connections are 1/4 inch fer­rule-type compression fittings.
Zero and span gases use the same inlet and outlet as the sample. Figure 2-4 on page 2-5 shows a typical external sample handling manifold for gas selection. Par­ticulates must be filtered down to two mi­crons, gases generally require pressurization, and flow measurement metering MUST be present.

c. Leak Test

The Analyzer Module is thoroughly tested at the factory for gas leakage. The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. The user is also responsible for in­ternal leak testing periodically and if any internal pneumatic components are ad­justed or replaced (with a test procedure selected by the user).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
(
)
760007-A July 2003
Note: Reference and purge gas connections are applicable only to certain applications.
Model NGA2000 PMD
Figure 2-2. PMD Front Panel Connections
FAN
SAMPLE
10 PSI MAX
69 kPa MAX
Figure 2-3. PMD Back Panel Connections
NETWORK 1
NETWORK 2
POWER
FUSE
OUT
IN
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Z
Instruction Manual
760007-A July 2003
Sample In
ero Standard Gas
Span Standard Gas
PMD Analyzer
Needle Valves
Two Micron
Filter
Flowmeter
Flow Splitter
Figure 2-4. Interconnection of Typical Gas Manifold to PMD Analyzer Module
Module
(2:1)
To Vent
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

2-5 ELECTRICAL CONNECTIONS

NOTE
Electrical connections must be in compli­ance with National Electrical Code (ANSI/NFPA 70) and/or any applicable na­tional or electrical codes.
Two electrical connections are required on the Analyzer Module; POWER and NETWORK.
J5
PRESSURE COMPENSATION BOARD
J6
J10
J11
J13
COMPUTER ANALYSIS
J7
J14
J19
W1
FLOW SENSOR CABLE
PREAMP CABLE ASSEMBLY
CASE TEMP SENSOR
FAN
See Figure 2-2 on page 2-4. On the Analyzer Module, two NETWORK connections are available, either of which is appropriate for : 1) interconnection with Backplane of the Plat­form (see Platform instruction manual) or 2) "daisy chaining" with other NGA2000 compo­nents.
Connect Analyzer Module POWER 24 VDC power source, either the Platform or external power source.
BOARD
J4
FLOW
SENSOR
J1
J2
J3
J16
Figure 2-5. PMD Wiring Diagram
MAIN BOARD
J20
SENSOR
ASSEMBLY
J21 J22
LED SOURCE
THERMAL CUTOFF
MAGNET HEATER AND
OPTICAL BENCH
J17
J9
J12
J15
ASSEMBLY
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
8.4
[213]
6.7
6.2
[157]
[152]
6.0
8.4
[213]
8.2
[208]
17.41 [442]
[109]
[28]
20.0
[508]
.6
[15]
E. FUSE. D. POWER CABLE TO NETWORK. C. NETWORK CABLE CONNECTIONS TO PLATFORM. B. SAMPLE OUT: 1/4" O.D. TUBE FITTING. A. SAMPLE IN: 1/4" O.D. TUBE FITTING.
5. MODULE TO BE INSTALLED WITHIN ±15° OF HORIZONTAL.
4. POWER REQUIREMENTS: 24 VDC 3.5 A.
3. ELECTRICAL INSTALLATION MUST BE IN COMPLIANCE WITH NATIONAL ELECTRICAL CODE (ANSI/NFPA 70) AND/OR ANY APPLICABLE NATIONAL OR LOCAL CODES.
2. MODULE IS NOT WEATHERPROOF.
1. APPROXIMATE WEIGHT: 15 LB (6.8 kg).
[170]
Figure 2-6. PMD Outline and Mounting Dimensions
[15]
10 PSI MAX
2.8
[71]
[13]
1.0
[25]
1.6
[41]
DIMENSIONS
Instruction Manual
760007-A July 2003
1.6
[40]
INCH
[mm]
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003
SECTION 3

OPERATION

3-1 OVERVIEW

Prior to initial startup, the user should perform the leak test procedure outlined in Section 2­4c on page 2-3.
For the remainder of this section Analyzer Module interconnection with a Platform or some interfacing component will be assumed. Display and keypad information shall refer to that which the user can expect to see and do with regard to the front panel of the Platform.
For a complete description of the Platform front panel controls and indicators, see the Platform instruction manual, Displays & Oper­ating Keys.

3-2 DISPLAYS

Three kinds of Display screens are available to the user:
Run Mode
Menu
Help

a. Run Mode Display

The Run Mode is the normal mode of op­eration. In this mode, the Display (see Figure 3-1 on page 3-3) will show current gas measurement, the component of in­terest, the current operations of the soft­keys, and a graphic bar representing the displayed concentration as ppm or as a percent of oxygen.
If more than one Analyzer Module is con­nected to the system, the Run Mode dis­play will show as many as four gas measurements on screen. Alarm mes­sages may also appear on the display (See Table 3-1 on page 3-2).

b. Menu Displays

The menu structure enables the user to access data and functions, and put infor­mation onto the network.
The Main Menu (Figure 3-2 on page 3-3) is subdivided into three levels of control based generally on which personnel is likely to use it: Basic Controls, Expert Controls and Setup, and Technical Level Configuration. See Figure 3-3 on page 3­3, Figure 3-4 on page 3-4, and Figure 3-5 on page 3-4. Many layers of the menu structure are described at appropriate places throughout this manual.
From the Run Mode display, press the MENU softkey to enter the Main Menu (Figure 3-2 on page 3-3).

c. Help Displays

The Help structure is intended to be an on-line "tutorial," context sensitive and topic-interconnected, so that the user can practically operate NGA2000 without need of an instruction manual (Figure 3-6 on page 3-4).
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Instruction Manual
760007-A July 2003
MESSAGE DISPLAY DESCRIPTION TYPE
barometer System Barometer warning case temp Case Temperature warning crude noise Calculated Noise warning currentrnghi Current, High Range warning currentrnglo Current, Low Range warning det tem Detector Temperature warning
Model NGA2000 PMD
fan fet Fan Current warning heater fet Heater Current warning led current LED Current warning lin error Linearizer Error warning loop current PMD Loop Current warning n15 volts Power Supply, -15V warning p15 volts Power Supply, +15V warning p24 volts Power Supply, +24V warning P5 volts Power Supply, +5v warning raw signal Raw Signal warning Samp Pres Sample Pressure warning svflow Sample Bypass Flow warning bicella PMD Photo Sensor failure bicellb PMD Photo Sensor failure sw error Software Error failure
Table 3-1. PMD Analyzer Module Alarms
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
y
y
y
y
Instruction Manual
760007-A July 2003
Analyzer PQ 322-14
23.2 ppm HC
0 ppm
Secondar Secondar Secondar Secondar
Variable: XXXX Variable: XXX Variable: XXXX Variable: XXXX
Display Parms. Menu Dual Info
F1 F2 F3 F4 F5
Figure 3-1. Run Mode Display
23.2 ppm HC Analyzer XXXXXXXX Main Menu
Basic Controls
Expert controls and setup ... (Operational configuration)
Technical level configuration ... (Diagnostic and manufacturing/service)
Delete alarm message!
Display Parms. Info
F1 F2 F3 F4 F5
Figure 3-2. Main Menu Display
23.2 ppm HC Analyzer XXXXXXXX Basic Controls
Measurement range numbers:
Range upper limit: 10 ppm Range and functional control: Local Bypass sample flow: 1000 ml/min Ranges with valid calibration 1&2 Calibration status: Ready If it won’t calibrate… Flame condition: On Light flame…
Home Escape Zero Span Info
F1
F2 F3 F4 F5
Figure 3-3. Basic Controls Menu
50
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
Instruction Manual
g
p
p
p
760007-A July 2003
Model NGA2000 PMD
23.2 ppm HC Analyzer XXXXXXXX ert controls and set u
Ex
Expert analyzer controls ...
Auxiliary module controls ... System set up ... Analyzer module set up ... Auxiliary module set up ... Local I/O set up ...
Home Escape Info
F1 F2 F3 F4 F5
Figure 3-4. Expert Controls and Setup Menu
23.2 ppm HC Analyzer XXXXXXXX Technical confi
System set up ...
Service menus... Diagnostic menus... Other module diagnostic menus... listing of all modules...
Home Escape Cal Info
F1
Figure 3-5. Technical Level Configuration Menu
23.2 ppm HC Analyzer XXXXXXXX
The Main Menu for the analyzer system. Note that this menu refers to the particular analyzer selected from the run screen, when used in a system. The softkey marked “HOME” will always return you to this screen. Help menu system... Help on help... Keyboard controls... Editing controls...
F2 F3 F4 F5
Main Menu Hel
Home Escape Map
F1
F2 F3 F4 F5
Figure 3-6. Typical Help Screen
uration menu
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003

3-3 STARTUP PROCEDURE

Introduce a suitable on-scale gas (NOT actual sample) into sample inlet.
Apply power to the PMD Analyzer Module. If it is associated with a Platform, do this by plugging in the Platform to a power source. The Platform has no ON/OFF power switch. Once power is supplied to the Platform, the PMD Analyzer Module will be energized.
If the user's system contains only one Ana­lyzer Module, all system components, the Controller Board and the network "self-install" (bind together) during initial startup. If the system contains more than one Analyzer Module, the startup sequence will interrogate the network to locate and identify all compo­nents on the network. The user will have to bind appropriate combinations of components after the startup sequence.
After the warm-up period, approximately one hour for PMD Module, all modules are com­pletely functional.
Enter appropriate data in the Calibration Gas List (by making the following display selec­tions: Main Menu, Expert Controls and Setup [enter security code, if necessary], Analyzer Module setup, Calibration Gas List). Also, en­ter appropriate values in the Calibration Pa­rameters menu (by making the following display selections: Main Menu, Expert Con­trols and Setup [enter security code, if neces­sary], Analyzer Module Setup, Calibration Parameters), particularly data related to which ranges are to be zeroed together and how the Analyzer Module is expected to calibrate ranges (separately or otherwise).
The Analyzer Module will not allow the user to increase the upper limit of a range beyond the "maximum range" software setting. To change the "maximum range" value, select the following from the Main Menu: Technical Configuration Menu, Service Menu, Manufac­turing Data, Analyzer Module Data. Select Maximum Range, and use the arrow keys to scroll the indicated value. The same applies for "minimum range" settings.

3-4 BINDING

To achieve full coordination between Analyzer Modules and associated I/O Modules, the user must bind those components together in the System Setup portion of the Technical Configuration Menu in software.

3-5 CALIBRATION

Calibration consists of establishing zero and span calibration points. Generally, zero and span calibration should be performed on the range that will be used during sample analy­sis.
To calibrate the PMD Analyzer Module, intro­duce zero gas into the SAMPLE INLET, and do the following:
1. If the Multi-Analyzer Module, split Run
Mode display is shown, press the DIS­PLAY softkey until the desired Analyzer's Run Mode display is acquired.
2. Press the MENUS softkey to enter the Main Menu.
3. Press the ENTER key to enter the Basic Controls menu.
4. Press the ZERO softkey to enter the Ana- lyzer Zero menu.
5. Press the ZERO softkey and wait.
6. Introduce span gas (Table 3-2 on page 3-
6) into the SAMPLE INLET.
7. Press the SPAN softkey to enter the Ana- lyzer Span menu, press SPAN again and wait.
8. Press the HOME softkey to re-enter the Main Menu.
9. Press the DISPLAY softkey for the Run Mode display.
For users of analyzers with suppressed indi­cating ranges, it may be desireable to cali­brate the analyzer zero and span points within the suppressed range with gases in the sup-
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
pressed range. Table 3-3 below shows rec­ommended zero and span standard gases for
RANGE % OXYGEN
0 to 1 Nitrogen 0.9% O2, balance N2
0 to 2.5 Nitrogen 2.3% O2, balance N2
0 to 5 Nitrogen 4.5% O2, balance N2 0 to 10 Nitrogen 9% O2, balance N2 0 to 25 Nitrogen Air (20.93% O2) 0 to 50 Nitrogen 45% O2, balance N2
0 to 100 Nitrogen 100% O2
Table 3-2. Calibration Range for Various Zero Based Operating Ranges
RANGE % OXYGEN
50 to 100 50 - 50.5% O2 Balance AR 100% O2
suppressed range oxygen indication in an Ar­gon background.
RECOMMENDED ZERO
STANDARD GAS
RECOMMENDED ZERO STANDARD
GAS
RECOMMENDED SPAN
STANDARD GAS
RECOMMENDED SPAN
STANDARD GAS
70 – 100% 70 – 70.5% O2 Balance AR 100% O2 98 – 100% 98 – 98.5% O2 Balance AR 100% O2 99 – 100% 99 – 99.5% O2 Balance AR 100% O2
Table 3-3. Calibration Range for Various Suppressed Range Operations
If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is initiated, nothing happens), a possible solution relates to the use of an incorrect gas for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer Module). Recalibrating with the appropriate gas(es) WILL NOT correct the problem be­cause the ZERO OFFSET or SPAN FACTOR has been set to an extreme value in the proc­ess.
To remedy the problem, do the following:
1. Select the following from the Main Menu: Expert Controls and Setup (enter security
code if necessary), Analyzer Module Setup, and Calibration Parameters.
2. Using the down arrow, select Zero Ranges, press ENTER and, using the up/down arrows, toggle to SEPARATE. Do the same for the Span Ranges selec­tion. Do not press ESCAPE at any time unless retention of prior settings is de­sired.
3. Return to the Main Menu (HOME) and make the following selections: Expert Controls and Setup (enter security code if necessary), Expert Analyzer Controls, CAL softkey, FACTORS softkey, and
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003
Range 1 (2, 3, 4) Factors. (Do steps 4 and 5 for each range.)
4. Select Zero Offset, press ENTER, adjust the value to 525000 with the up/down ar­row keys, and press ENTER. Do not press ESCAPE at any time unless reten­tion of prior settings is desired.
5. Select Span Factor, press ENTER, adjust the value to 0.000015 with the up/down arrow keys, and press ENTER. Do not
press ESCAPE unless retention of prior settings is desired.
6. Attempt to recalibrate the Analyzer Mod­ule according to the procedure outlined at the beginning of Section 3-5 on page 3-5. If re-calibration fails, return to the Range Factors menu, readjust factors, and at­tempt calibration again.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Instruction Manual
A
760007-A July 2003

3-6 BACKGROUND GAS COMPENSATION

Any gas having a composition other than 100% oxygen contains background gas, that is, non-oxygen components. Sometimes, the PMD Module response to background gas is significant, depending largely on the span and range used.
If the operator uses zero and span gases that contain the same background gas as the sampl e, calibration procedures automatically compensate. No adjustments are necessary.
If the background gas in the sample is different from that in the zero and/or span gases, the operator must take into consideration background effects to ensure correct readout. During entry of zero and span ga s values in the Calibration Gas List, the instrument is not set to indicate the true oxygen content of the zero and span standard gases. It is set to indicate a slightly different value, relative to background gas, calcu­lated to provide correct readout during subsequent analysis of sample gas.
Oxygen Equivalent Values of Gases
For computation of background corrections, the analyzer's response to each component of the sample must be known. Table 3-4 on page 3-10 lists the percentage oxygen equivalent values for many common gases. For a more comprehensive list of oxygen equivalent values, refer to a resource text such as the Handbook of Chemistry and Physics for tables of magnetic susceptibility of substances. The percentage oxygen equivalent of a gas can be determined by the following equation, assuming both gases are supplied at the same pressure:
Model NGA2000 PMD
For example, if the analyzer's response to oxygen is +100%, the response to xenon would be -1.34%. The oxygen equivalent of a gas mixture is the sum of the contribution of the individual gas components. Example: Zero Based Range At lower range limit (i.e., 0% O From Table 3-4 on page 3-10, the percent oxygen equivalents are: CO
The percent oxygen equivalent of the mixture = 0.8(-0.623) + 0.2(-0.358) = (-0.4984) + (-0.0716) = 0. 570% O
%O2 Equivalent of Gas = X 100%
2), composition of sample is: 80% CO2, 20% N2.
nalyzer Response to Gas
Analyzer Response to O2
2 -0.623%, N2 -0.358%.
2.
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model NGA2000 PMD
Computing Adjusted Values for Calibration Gas List
Before calibrating the Analyzer Module, values in the Calibration Gas List must be adjusted to correct for magnetic susceptibility of background gas. In the equation that follows, the quantities are defined as fol­lows:
BGGst = oxygen equivalent of background gas in standard gas (Table 3-4 on page 3-10).
BGGs = oxygen equivalent of background gas in sample (Table 3-4 on page 3-10).
OP = operating pressure. Unless special pressure corrections a re to be made, the zero standard, span
standard and sample gases must all be admitted at the same pressure.
Use the following equation to compute the adjusted settings for the Calibration Gas List:
(A)[100 + (B-C)]-100[B-C]
Adjusted percent oxygen for standard gas =
100
760007-A July 2003
Where:
A = true percent oxygen of standard gas B = BGGs
C = BGGst Example: Background gs in sample is CO Zero gas is 100% N
2.
, oxygen equivalent = -0.623%.
2
Span standard gas is air: 21% O2, 79% N Background gas in zero and span standard gases is N With N
2 zero standard gas flowing, zero gas value in the Calibration Gas List would be 0.265% O
termined by the following):
0[100+(-0.623-(-0.358))] - 100{-0.623-(-0.358)]
.
2
100
, oxygen equivalent = 0.358%.
2
= 0.265% O2
(as de-
2
With air flowing, span gas value in the Calibration Gas List would be 21.21% oxygen (as determined by the following):
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
21(100 - 0.265) - 100 (-0.265)
100
= 21.209% O2 ≅ 21.21 O2
Instruction Manual
760007-A July 2003
In two limiting cases, the general equation is reduced to simpler forms.
1. If the span standard gas is 100% oxygen, the adjusted oxygen value is the same as the true value (i.e.,
100% O
2. If the zero standard is an oxygen-free zero gas, the adjusted value for setting the ZERO Control =
BGGst-BGGs. (If the oxygen-free zero gas is more diamagnetic than the background gas in the sam-
ple, this difference is negative. The negative value may be entered in the Calibration Gas List.) Alternately, the user can avoid these compensation calculations by using zero and span gases which have
been specially prepared to contain the expected amounts of background gas. Calibration of the analyzer module will then factor in background gas effects in the same proportions as normal run mode measure­ment.
).
2
Model NGA2000 PMD
GAS EQUIV. % AS O
Acetylene, C2H2 -0.612 Hydrogen Bromide, Hbr -0.968 Allene, C3H4 -0.744 Hydrogen Chloride, HC1 -0.650 Ammonia, NH3 -0.479 Hydrogen Fluoride, HF -0.253 Argon, A -0.569 Hydrogen Iodide, HI -1.403 Bromine, Br2 -0.83 Hydrogen Sulhide, H2S -0.751 1,2-Butadiene, C4H6 -1.047 Krypton, Kr -0.853 1,3-Butadiene, C4H6 -0.944 Methane, CH4 -0.512 n-Butane, C4H10 -1.481 Neon, Ne -0.205 iso-Butane, C4H10 -1.485 Nitric Oxide, NO +44.2 Butene-1, C4H8 -1.205 Nitrogen, N2 -0.358 cis Butene-2, C4H8 -1.201 Nitr o g e n D i o x i d e , N O2 +28.7 iso-Butene, C4H8 -1.274 Nitr o u s O x i d e , N2O -0.560 trans Butene-2, C4H8 -1.274 n-Octane, C8H18 -2.840 Carbon Dioxide, CO2 -0.623 Oxygen, O2 +100.0 Carbon Monoxide, CO -0.354 n-Pentane, C5H12 -1.810 Ethane, C2H6 -0.789 iso-Pentane, C5H12 -1.853 Ethylene, C2H4 -0.553 neo-P entane, C5H12 -1.853 Hel i u m , H e -0.059 Prop a n e , C3H8 -1.135 n-Heptane, C7H16 -2.508 Propylene, C3H6 -0.903 n-Hexane, C6H14 -2.175 Water, H2O -0.381 cyclo-Hexane, C6H12 -1.915 Xenon, Xe -0.340 Hydrogen, H2 -0.117
GAS EQUIV. % AS O2
2
Table 3-4. Oxygen Equivalents of Common Gases
3-10 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003

3-7 BAROMETRIC PRESSURE COMPENSATION

Although normally calibrated for readout in percent oxygen, the PMD Analyzer Module actually responds to oxygen partial pressure. The partial pressure of the oxygen component in a gas mixture is proportional to the total pressure of the mixture. Thus readout is af­fected by pressure variations.
For instance, assume that an instrument is calibrated for correct readout with a standard gas containing 5% oxygen, admitted at the normal sea level atmospheric pressure of 14.7 psia (1013 hPa). If the operating pressure
now drops to one-half of the original value (i.e., to 7.35 psia/506 hPa and the calibration controls are left at the previously established settings, the display reading for the standard gas will drop to 2.5%.
It is therefore necessary to calibrate the in­strument at the same pressure that will be used during subsequent operation, and to maintain this pressure during operation.
Alternatively, an optional Barometric Pressure Compensation Board, typically used for sup­pressed range applications, can perform sig­nal corrections automatically.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-11
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
3-12 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD

MAINTENANCE AND SERVICE

Instruction Manual
760007-A July 2003
SECTION 4
WARNING
QUALIFIED PERSONNEL
This equipment should not be adjusted or repaired by anyone except properly quali­fied service personnel.

4-1 OVERVIEW

PMD Analyzer components that may require replacement include:
All printed circuit board
Thermal fuse inside Detector
Case temperature sensor
Flow sensor
Power fuse
Detector
Module fan
PRESSURE COMPENSATION BOARD
COMPUTER ANALYSIS BOARD
FLOW SENSOR
MODULE BOARD
Figure 4-1. PMD Module – Major Components
The LED bi-cell assembly source required ad­justment (rotation) anytime the Detector is disassembled. Refer to Figure 4-3 on page 4­3 for locations of these components.

4-2 PRINTED CIRCUIT BOARD REPLACEMENT

Refer to Figure 4-1 belowfor locations of the Signal, Microprocessor, Power Supply and (optional) Pressure Compensation Boards.
All boards are secured to a side of the ana­lyzer module that folds out while interconnec­tion wiring is still in place. Remove the securing screws and fold out the entire panel.
To remove individual boards on the fold-out panel, label and unplug all interconnection wiring, and remove securing hardware. Re­verse this procedure for installation.
NETWORK INPUT
DETECTOR HOUSING
CASE TEMPERATURE SENSOR
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-1
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

4-3 MODULE FAN REPLACEMENT

The Analyzer Module fan assembly is disas­sembled as shown in Figure 4-2 below.
Figure 4-2. Module Fan Assembly
4-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Instruction Manual
760007-A July 2003

4-4 THERMAL FUSE REPLACEMENT

See Figure 4-3 below for location of the De­tector Thermal Fuse.

DETECTOR ASSEMBLY

THERMAL FUSE
DETECTOR ASSEMBLY
TOP VIEW
MAGNET/COIL ASSEMBLY
SPLITTER ASSEMBLY
BOTTOM VIEW
Remove the Detector Assembly form the de­tector housing, replace the thermal fuse.
Reassemble in reverse order.
DETECTOR HOUSING
THERMAL FUSE
Figure 4-3. Detector Assembly
DETECTOR ASSEMBLY
LED
OPTICAL BENCH ASSEMBLY
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 4-3
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

4-5 FLOW SENSOR REPLACEMENT

See Figure 4-1 on page 4-1 for location of Flow Sensor. To replace the sensor, re­move connections to sample gas line and disconnect securing hardware. Reassem­ble in reverse order.

4-6 POWER FUSE REPLACEMENT

The power fuse is located in the Network Input Module and is accessible through the front panel of the PMD Analyzer. To re­move the fuse, push and turn the fuse­holder cap 1/4 turn counterclockwise. Verify that the replacement fuse is the same type and rating.
4-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD

REPLACEMENT PARTS

Instruction Manual
760007-A July 2003
SECTION 5
WARNING.
PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.

5-1 MATRIX

Each analyzer is configured per the customer sales order. Below is the PMD sales matrix which lists the various configurations avail­able.
PMD2 NGA2000 PARAMAGNETIC DETECTOR
01 Current Version 02 2.2.1 version 03 3X version – specify version
A1 Calibrated Standard Ranges: 0-5, 0-10, 0-25, 0-100% A2 Calibrated Standard Ranges: 0-1, 0-5, 0-10, 0-25%
H1 H2
99 Special Calibration Ranges
00 None S1 Standard Elevations E1 High Elevations
R1 Rhodium Plated Current Loop T1 Titanium Current Loop (Standard)
PMD2 01 A1 S1 T1 Example
Code Software Version
Code Configuration Identifier
Calibrated Standard Suppressed Ranges/Barometric Pressure Compensation: 90-100, 95-100, 98-100, 99-100% Calibrated Standard Suppressed Ranges/Barometric Pressure Compensation: 50-100, 70-100, 80-100, 90-100%
Code Barometric Pressure Compensation
Code Detector Type
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 5-1
Instruction Manual
760007-A July 2003
Model NGA2000 PMD

5-2 REPLACEMENT PARTS

902931 Sensor, Gas Flow 655856 Source/Holder Assembly 902922 Bi-Cell, Optical 655670 Pressure Compensation Board 903347 Fuse, Time-Delay 6A 250 VAC 657860 Module Board 622917 Sensor RTD 656576 Case Temperature Sensor 655893 Fan 898733 Detector Thermal Fuse 655838 Optical Bench Assembly 658083 Detector, Corrosion Resistant (Option)
5-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD

RETURN OF MATERIAL

6-1 RETURN OF MATERIAL

If factory repair of defective equipment is re­quired, proceed as follows:
1. Secure a return authorization from a
Rosemount Analytical Inc. Sales Office or
Representative before returning the
equipment. Equipment must be returned
with complete identification in accordance
with Rosemount instructions or it will not
be accepted.
Rosemount CSC will provide the shipping
address for your instrument.
In no event will Rosemount be responsi-
ble for equipment returned without proper
authorization and identification.
2. Carefully pack the defective unit in a
sturdy box with sufficient shock absorbing
material to ensure no additional damage
occurs during shipping.
3. In a cover letter, describe completely:
The symptoms that determined the equipment is faulty.
The environment in which the equip­ment was operating (housing, weather, vibration, dust, etc.).
Site from where the equipment was removed.
Whether warranty or non-warranty service is expected.
Complete shipping instructions for the return of the equipment.
4. Enclose a cover letter and purchase order and ship the defective equipment accord­ing to instructions provided in the Rose­mount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
Instruction Manual
760007-A July 2003
SECTION 6
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi­tions listed in the standard Rosemount war­ranty, the defective unit will be repaired or replaced at Rosemount’s option, and an oper­ating unit will be returned to the customer in accordance with the shipping instructions fur­nished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

6-2 CUSTOMER SERVICE

For order administration, replacement Parts, application assistance, on-site or factory re­pair, service or maintenance contract informa­tion, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076

6-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact the Technical Ser- vices Department at:
Rosemount Analytical Inc.
Customer Service Center
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 6-1
Instruction Manual
760007-A July 2003
Model NGA2000 PMD
6-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
A

APPENDIX A. MENU DISPLAYS

Menu: 0 ANALOP
Menu: 1 ANALSET
Menu: 2 FLOCHEK
Measurement range number: Range lower limit: Range upper limit: Linearizer: Range and functional control: Zero/Span calibration… Ranges with val id ca lib r a t i on :
HOME
Calibration gas list… Calibration Parameters… Concentration alarms… Gas measurement parameters…
Analyzer parameter list… Physical measurement parameters…
Displayed parameters…
HOME
Sample flow: Flow lower limit: Flow upper limit: Sample pressure:
Case temperature:
HOME
SECTION 7
Expert controls
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE CAL CAL DAT
nalyzer module set up
ESCAPE INFO
Secondary Measurements
ESCAPE INFO
INFO
000.0
000.0
000.0
000.0
000.0
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-1
Instruction Manual
K
y
760007-A July 2003
Menu: 3 ZEROl1
This allow manual control of the zero and span. Flow zero gas, and make sure the gas value is correct: press the zero key to make the analyzer zero itself. Or select Edit measurement usin g zero offset, th e n scro ll th e reading with the up and down keys. In this way you can make the analyzer read what you want. Then do the same with span gas. If the zero was not a real zero, the span action will change the zero reading; the last zero reading shows you what it would have been on the zero gas with the current span.
HOME
Menu: 4 SPANl1
Use this screen to perform a span Calibration. Either directly adjust the reading with the up and down arrow keys, or press the SPAN softkey to force a span calibr a tion to the span gas concentration.
You should do a zero calibration before
HOME
Menu: 5 FLOCHEKl1
This screen shows the auxiliar measurements made b y the anal yzer mod ul e the limits may be set by the user as warning alarms.
These readings are updated only when they change.
HOME
Zero/Span Calibration help
ESCAPE MORE INFO
Span Calibration help
ESCAPE INFO
Secondary Measurements help
ESCAPE INFO
MORE
Model NGA2000 PMD
BAC
7-2 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 6 ANALOPl1
This screen selects immediately available functions. Lines that are not editable refer to variables set up elsewhere. To zero or span the analyzer, flow the correct gas and press the zero or span button. Remote control does not disable local control. This screen does not control an autocal module. Calibration info…
HOME
ESCAPE INFO
Menu: 7 ACALSET
Calibration adjustment limits: Calibration averaging time: Calibration failure alarm: Cal failure error allowed:
Calibration time out: Zero ranges:
Span ranges:
HOME
ESCAPE INFO
Menu: 8 LINSET
Linearization parameters… Response time/delay parameters…
Range setting… Automatic range change parameters… Units…
Linearization functions…
HOME
ESCAPE INFO
Basic controls help
Calibration Parameters
Gas measurement parameters
Instruction Manual
760007-A July 2003
000.0
000.0
000.0
000.0
000.0
000.0
000.0
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-3
Instruction Manual
A
760007-A July 2003
Menu: 9 APARLST
Analyzer tag: First line’s parameter:
Second line’s parameter: Third line’s parameter: Fourth line’s parameter:
HOME
Menu: 10 ANALSETl1
This lists the operational parameters that can be set up by the user. More detailed information can be seen in the diagnostic menus.
HOME
ESCAPE INFO
Menu: 11 CALLIST
Zero gas Span gas – range 1: Zero gas – range 2: Span gas – range 2: Zero gas – range 3: Span gas – range 3: Zero gas – range 4: Span gas – range 4:
range 1:
HOME
ESCAPE INFO
nalyzer Parameter List
ESCAPE INFO
Measurement Parameters
NEXT
Calibration Gas List
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
LAST
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-4 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
Menu: 12 CALLISTl1
Zero and span gases for each range are shown. Edit these to correspond to the contents of the appropriate calibrat ion gas bottles.
Enter the values of concentration on the bottles of gas used for calibration.
HOME
ESCAPE INFO
Menu: 13 ACALSETl1
Factors that control how the calibration works. Calibration mode allows calibration to be initiated through this screen (local), through an IO module or gateway (remote), or automatically by an auto­calibration module (auto). Calibration averaging time set the time used by the analyzer to average its reading. A longer time will give a
HOME
ESCAPE INFO
Menu: 14 APARLSTl1
This is a listing of all the user editable parameters in the current parameter set.
HOME
ESCAPE INFO
Calibration Gas List
Calibration Parameter help
MORE
nalyzer Parameter List
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-5
Instruction Manual
A
A
760007-A July 2003
Menu: 15 AMMAN
Menu: 16 AMMANl1
Menu: 17 AMSVC
nalyzer manufacturing data
More… Bench configuration code:
Minimum range: Maximum range: Measured gas:
HOME
Shows manufacturing data. Edit at your own risk! The tag is the tag of this analyzer module, and may be modified as desired by user. This appears at the top of this screen, but it won’t be updated until you re-enter the menus from the display screen.
RESET erases ALL EEPROM data! Re-initialize the system after RESET!
HOME
Manufacturing date: In service date: Last zero calibration date: Last span calibration date: Last service date: List notes…
HOME
ESCAPE INFO
nalyzer manufacturing data
ESCAPE INFO
Analyzer module service history
ESCAPE INFO
RESET
STORE
ManData
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-6 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
A
A
Menu: 18 AMSVCl1
Menu: 19 ADIAG
Menu: 20 AMPWR
Add notes as desired. See the control module service screen into for a list of the service note abbrevi a tions.
HOME
Power supply voltages… Primary variable parameters… Physical Measurements… Temperature control parameters… Miscellaneous control parameters… Trend display control… Barometric pressure parameters… Software diagnostics…
HOME
Power supply voltages +15V analog is: +15V analog was:
-15V analog is:
-15V analog was: +5V digital is: +5V digital was: +24V power is: +24V power was:
HOME
nalyzer module service notes
ESCAPE INFO
nalyzer Diagno st ic s
ESCAPE INFO
ESCAPE INFO
REBOOT
nalyzer diagnostics
INIT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-7
Instruction Manual
760007-A July 2003
Menu: 21 AM1V
Menu: 22 AMTEMP
Menu: 23 AMMISC
Primary variable parameters
Raw measurement signal: Bicell #1 signal: Bicell #2 signal: Pk-pk noise: Loop current:
Calibration factors…
HOME
Fan lower set point: Fan upper set point: Minimum fan duty cycle:
Case temperature: Detector set point: Detector P gain: Detector I gain: Detector bias:
HOME
Fan current: Fan duty cycle: Source LED current:
Heated current: Heater duty cycle:
Alarm messages valid for:
HOME
ESCAPE INFO
Temperature control
ESCAPE INFO
Miscellaneous control parameters
ESCAPE INFO
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-8 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
Menu: 24 AMTREND
Menu: 25 ADIAGl1
Menu: 26 RANGESETAM
First displayed variable: Second displayed variable:
Timebase: Drop out to measuring mode:
HOME
Select the area of diagnostics to examine.
Press INIT to re-initialize th e EEPROM data. Press REBOOT to restart the analyzer.
HOME
Minimum range: Maximum range: Range 1 lower limit: Range 1 upper limit: Range 2 lower limit: Range 2 upper limit: Range 3 lower limit: Range 3 upper limit: Range 4 lower limit: Range 4 upper limit:
HOME
Instruction Manual
760007-A July 2003
Trend display control
000.0
000.0
000.0
000.0
ESCAPE INFO
nalyzer Diagno st ic s
ESCAPE INFO
Range Settings
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-9
Instruction Manual
K
K
760007-A July 2003
Menu: 27 RANGESSETl1
Menu: 28 SPANl2
Menu: 29 ACALSETl2
Range Settings
Set the upper and lower limits of the reportable ranges. These values are copied into the output module and used for calculating the analog output. The analyzer uses them to select the closest linearizer polynomial to use if any.
HOME
The calibration is performed on the range selected by the range If the calibration set up selected the ranges to be calibrated together, all the ranges will be calibrated at once by the first calibration. Otherwise they must be calibrated individually. The line showing what the last zero gas would read shows the effect of the most
HOME
Calibration alarms will only work if warning alarms are enabled.
Calibration info…
HOME
ESCAPE INFO
Span Calibration help
ESCAPE INFO
Calibration Parameter help
ESCAPE INFO
BAC
BAC
Model NGA2000 PMD
7-10 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
K
Menu: 30 ZEROl2
Menu: 31 LINRANGE1
Menu: 32 LINRANGE2
Make sure the zero gas is flowing through the analyzer. When the reading has stabilized, either adjust the reading to the zero gas concentration, or, if the zero gas concentration shown is correct, select the automatic zero function. Then if desired, do a span calibration.
HOME
Curve 1 A0 coefficient: A1 coefficient: A2 coefficient: A3 coefficient: A4 coefficient: Curve upper limit: Curve over-range: Curve under-range:
HOME
Curve 2 A0 coefficient: A1 coefficient: A2 coefficient: A3 coefficient: A4 coefficient: Curve upper limit: Curve over-range: Curve under-range:
HOME
Zero Calibration help
ESCAPE INFO
Linearity coefficients
ESCAPE INFOLASTNEXT
Linearity coefficients
ESCAPE INFOLASTNEXT
MORE
BAC
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-11
Instruction Manual
760007-A July 2003
Menu: 33 LINRANGE3
Menu: 34 LINRANGE4
Menu: 35 LINRANGE0
Linearity coefficients
Curve 3 A0 coefficient: A1 coefficient: A2 coefficient: A3 coefficient: A4 coefficient: Curve upper limit: Curve over-range: Curve under-range:
HOME
Curve 4 A0 coefficient: A1 coefficient: A2 coefficient: A3 coefficient: A4 coefficient: Curve upper limit: Curve over-range: Curve under-range:
HOME
Range 1 lineariz er: If enabled, uses curve no.: Range 2 lineariz er: If enabled, uses curve no.: Range 3 lineariz er: If enabled, uses curve no.: Range 4 lineariz er: If enabled, uses curve no.:
HOME
ESCAPE INFOLASTNEXT
Linearity coefficients
ESCAPE INFOLASTFIRST
Linearization parameters
ESCAPE INFO
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-12 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
Menu: 36 AMPWRl1
Menu: 37 FLOCHEK1l1
Menu: 38 FILTER
Power supply voltages The the power supplies as described. The
Unit was manufactured. Changes of more than a few percent should be noted. The 24V supply may differ substantially if the unit is used on anything but a Rosemount power supply.
HOME
These are the measurements made by the analyzer module to make sure that it is working correctly, and the sample and support gases if any are flowing.
The various temperatures are controlled to values set up in the diagnostic menus.
HOME
Range 1 t90 time: Range 2 t90 time: Range 3 t90 time: Range 4 t90 time:
LON update rate: Output delay time:
HOME
Instruction Manual
760007-A July 2003
nalyzer diagnostics
ESCAPE INFO
Physical Measurements
ESCAPE INFO
Response time/delay parameters
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-13
Instruction Manual
760007-A July 2003
Menu: 39 AM1Vl1
Menu: 40 AMTEMPl1
Menu: 41 AM2VA
Model NGA2000 PMD
Primary variable parameters
Shows the parameters used to calculate the primary reading. The bicell readings are the individual outputs of the two photocells. The instrument responds to the difference between them, but each signal should be as large as possible without saturating. The pk-pk noise is a measure of the standard devi a tion of the signal.
HOME
These are the variables used to define the operation of the PID algorithms used for temperature control. Adjust them at your own risk!
This analyzer controls its detec tor temperature, while keeping its case temperature withi n rath er wide lim it s.
HOME
Sample flow: Sample pressure: Case temperature:
Detector temperature: Flow limits… Pressure limits…
HOME
ESCAPE INFO
Temperature control
ESCAPE INFO
Physical Measurements
ESCAPE INFO
000.0
000.0
000.0
000.0
7-14 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 42 PLMITSA
Menu: 43 TLIMITSA
Menu: 44 AMMISCl1
Sample pressure upper limit: Sample pressure lower limit:
HOME
Case upper limit: Case lower limit:
HOME
These are diagnostic varia b les used to determine if the analyzer is operating correctly.
Enable alarms if desired.
HOME
Instruction Manual
760007-A July 2003
Pressure Limits
000.0
000.0
ESCAPE INFO
Temperature limits
000.0
000.0
ESCAPE INFO
Miscellaneous control parameters
ESCAPE INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-15
Instruction Manual
760007-A July 2003
Menu: 45 ANALSIMPLE
Menu: 46 FILTERl1
Menu: 47 LINSET1l1
Basic Controls
Measurement range number: Range upper limit:
Range and functional control: Zero gas concentration: Span gas concentration: Sample flow: Ranges with valid calibration: Calibration status:
HOME
This screen sets the final filtering for the analyzer primary variable output. This is in addition to the inherent filtering in the ana lyzer. The time delay simply delays the output by that time, allowing the fastest responding analyzer systems to be synchronized with the slowest.
HOME
Allows the setting of linearizer coefficients, definition of parameter sets, and filtering and delay. These all apply to the reporting of the analyzer primary variable.
HOME
ESCAPE INFO
Filter and Delay Parameters
ZERO
ESCAPE INFO
Primary Variable Parameters
ESCAPE INFO
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
SPAN
7-16 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 48 LINRANGE0l1
Menu: 49 PLMITSAl1
Menu: 50 CALFACTORS
The linearizer polynomials act over a certain range, not the same as the measurement range. The system use s the linearizer polynomial appropriate for the measurement range chosen. This is the polynomial with the next higher linearizer range than the measurement range. Polynomial coefficients may be edited for custom curves.
HOME
These are settable limits on th e samp le gas pressure and flow. The force alarms to occur and also act as end points on the bar graph display of their variable.
HOME
Range 1 factors… Range 2 factors… Range 3 factors… Range 4 factors…
Zero compensation factor: Span compensation factor:
HOME
Instruction Manual
760007-A July 2003
Set Linearity curve
ESCAPE INFO
Pressure and flow Limits
ESCAPE INFO
Calibrati on F actors
000.0
000.0
ESCAPE INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-17
Instruction Manual
760007-A July 2003
Menu: 51 R1FACTORS
Menu: 52 RN2FACTORS
Menu: 53 RN3FACTORS
Range 1 Factors
Zero offset: Span factor: Full scale range at calibration: Measurement range number:
Hardware zero offset: Raw measurement signal:
HOME
Zero offset: Span factor: Full scale range at calibration: Measurement range number:
Hardware zero offset: Raw measurement signal:
HOME
Zero offset: Span factor: Full scale range at calibration: Measurement range number:
Hardware zero offset: Raw measurement signal:
HOME
STORE INFO
Range 2 Factors
STORE INFO
Range 3 Factors
STORE INFO
NEXT
NEXT
NEXT
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
HISTORY
000.0
000.0
000.0
000.0
000.0
000.0
HISTORY
000.0
000.0
000.0
000.0
000.0
000.0
HISTORY
7-18 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
Menu: 54 RN4FACTORS
Menu: 55 RFACTORSl
Menu: 56 AMHELPINDEX
Zero offset: Span factor: Full scale range at calibration: Measurement range number:
Hardware zero offset: Raw measurement signal:
HOME
Shows the calibration factors for this range. Modify the zero factor for zero calibration, and the span factor for spanning this range. The take effect as soon as you press the enter key.
With zero gas, the raw reading should be the same as the zero offset.
Then do a complete recalibration.
HOME
Paramagnetic Oxygen detector This device uses the paramagnetic nature of oxygen as a measurement technique. Other gases are usually diamagnetic and the detector responds only weakly to them. Nitrogen oxides are the only common exceptions. The analyzer is sensitive to vibratio n and should be installed with care.
HOME
Range 4 Factors
000.0
000.0
000.0
000.0
000.0
000.0
STORE INFO
ESCAPE INFO
ESCAPE INFO
FIRST
Range Factors
nalyzer Module Help
HISTORY
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-19
Instruction Manual
A
760007-A July 2003
Menu: 57 LINRANGE1l1
Menu: 58 CALFACTORSI1
Menu: 59 APARLST2
Model NGA2000 PMD
Linearity coefficients
Edit the polynomial coefficients as desired. Make sure that the curve upper limit is correct, this is the limit of the range that this polynomial will correct. The status line selects wh et her the curve is in use.
HOME
The analyzer uses calibration facto rs fo r each range. You can adjust them while viewing the rea di n g, to achieve an accurate calibr a tion. Make sure the factors are correct for the range you are on. You will not see a change in the reading if you use the wrong ones, but you’ll find out when you change the range! You cannot adjust all ranges at the same time, you must adjust them one by one.
HOME
Primary Variable Parameters Control mod e: Output delay time: Range 1 upper limit: Range 2 upper limit: Range 3 upper limit: Range 4 upper limit: Range 1 lower limit: Range 2 lower limit: Range 3 lower limit:
HOME
STORE INFO
Calibrati on F actors
ESCAPE INFO
nalyzer Parameter List
ESCAPE INFO
NEXT
BACK
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-20 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
A
K
A
Menu: 60 APARLST4
Menu: 61 APARLST5
Menu: 62 APARLST6
Primary Variable Parameters Range 1 t90 time: Range 2 t90 time: Range 3 t90 time: Range 4 t90 time: Linearizer on range 1: Linearizer on range 2: Linearizer on range 3: Linearizer on range 4:
HOME
Calibration Parameters Calibration averaging time:
Calibration failure alarm: Cal failure error allowed: Calibration time out: Ranges zeroed: Calibrate ranges:
HOME
Calibration Gases Zero gas – range 1: Zero gas – range 2: Zero gas – range 3: Zero gas – range 4:
Span gas – range 1: Span gas – range 2: Span gas – range 3: Span gas – range 4:
HOME
nalyzer Parameter List
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
ESCAPE INFO
ESCAPE INFO
NEXT
nalyzer Parameter List
NEXT
nalyzer Parameter List
FIRST
BACK
000.0
000.0
000.0
000.0
000.0
000.0
BAC
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
BACK
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-21
Instruction Manual
760007-A July 2003
Menu: 63 DISPLAY
Menu: 64 MPARMS
Menu: 65 MPARMS2
Displayed parameters
First line’s parameter: Second line’s parameter: Third line’s parameter: Fourth line’s parameter:
Displayed concentration digits: Digits after decimal point:
HOME
Analyzer gas measured: Measurement range number: Range change control:
Linearization mode: Analyzer operational status:
Analyzer alarm state: Alarm reporting level:
HOME
Response time: Sample flow: Sample pressure: Detector temperature:
HOME
ESCAPE INFO
Current measurement parameters
ESCAPE INFO
Current measurement parameters
ESCAPE INFO
MORE
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-22 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 66 MPARMSI1
Menu: 67 BAROM_PARMS
Menu: 68 BAROM_PARMI1
Shows the main measurement parameters. These can be controlled in the various set up menus.
HOME
Pressure transducer: Barometric pressure compensation: Measured pressure:
Transducer offset: Transducer slope:
Transducer PGA gain:
HOME
Shows whether the optional pressure transducer is installed or not. Also this screen allows you to set the transducer. Use the offset at atmospheric pressure, and pull a vacuum of about 25cm H2O on the transducer, and use the slope to make it read correctly.
Choose whether to enable barometric compensation, and whether to allow the analyzer to emit t h e pressure value. Only one analyzer at a time in a
HOME
Instruction Manual
760007-A July 2003
Current measurement parameters help
ESCAPE INFO
Barometric pressure parameters
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
Barometric pressure parameters
ESCAPE INFO
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-23
Instruction Manual
A
760007-A July 2003
Menu: 69 TLIMITSIAI1
Menu: 70 SW_DIAG
Menu: 71 LISTNOTES
Temperature limits
The limits on the temperature beyond which the analyzer will send a warning mes sa ge.
The also act as end points to the bar graph display.
The various temperatures are controlled to values set up in the diagnostic menus.
HOME
Last message: And: And: And: And: And: And: And: Edit to reset:
HOME
You can write up to 22 characters in each line.
HOME
ESCAPE INFO
Software diagnostics
ESCAPE INFO
nalyzer module service notes
ESCAPE INFO
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-24 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
Menu: 72 FLOWLIMITS
Menu: 73 AMPATH
Menu: 74 CALI1
Sample flow upper limit: Sample flow lower limit:
HOME
Select the function you want in the line below, and not the path shown.
Function: Select: Then: Then: Then:
HOME
Use the calibration parameter screen to choos e to calibrate ranges together or not. If together, zeroing or spanning will go through each range one by one. If the change is too great, it will fail, and send an alarm if warning alarms are enabled. In this case, disable calibration limit checking and try again.
If you use non-zero zero gases, or the changes are
HOME
Instruction Manual
760007-A July 2003
Flow limits
000.0
000.0
ESCAPE INFO
nalyzer Module Help
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
Calibration info
ESCAPE INFOMORE
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-25
Instruction Manual
760007-A July 2003
Menu: 75 LINFUNCT
Menu: 76 POLYSETUP
Menu: 77 MIDPOINT1
Model NGA2000 PMD
Linearization functions
Polynomial set up… Midpoint correction set up…
Use the polynomial set up to generate a linearizing polynomial from up to 20 gases. With more than 6 gases it will produce a fourth order polynomial linearizer. Use the midpoint correction for a piecewise-linear final correction, to bring up to three points precisely onto
HOME
Range to be linearized: Current span gas: Calculated polynomial order:
Gas values shown as: Gas concentrations…
HOME
Range 1 Correction: Point being measured: Point 1 gas concentration: Point 2 gas concentration: Point 3 gas concentration: Point 1 reading: Point 2 reading: Point 3 reading: Span gas value:
HOME
ESCAPE INFO
Polynomial set up
ESCAPE INFOCALC
Midpoint correction set up
ESCAPE SET NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
INFO
7-26 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
A
Menu: 78 POLYGAS1
Menu: 79 POLYSETI1
Menu: 80 POLYGAS2
Point 1 Gas value: Raw readi ng : Linearized value: Point 2 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Select the range to linearizer. Make sure that the span gas value is correct. Choose whether to define the gas concentrations as absolute values or as a percent to the span gas. Use percent if you are diluting the sp a n gas fo r thi s. Get into the gas concentrations scree ns, and set the concentration for as many points as you want. At each point, flow the gas of the correct val u e, and when the reading is stable, press The analyzer will store the gas value and the reading
HOME
Point 3 Gas value: Raw readi ng : Linearized value: Point 4 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Gas concentrations
000.0
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
ESCAPE INFOMORE
ESCAPE INFO
DAT
Polynomial set up
Gas concentrations
DAT
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-27
Instruction Manual
A
A
A
760007-A July 2003
Menu: 81 POLYGAS3
Menu: 82 POLYGAS4
Menu: 83 POLYGAS5
Point 5 Gas value: Raw readi ng : Linearized value: Point 6 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Point 7 Gas value: Raw readi ng : Linearized value: Point 8 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Point 9 Gas value: Raw readi ng : Linearized value: Point 10 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
ESCAPE INFO
ESCAPE INFO
ESCAPE INFO
Gas concentrations
DAT
Gas concentrations
DAT
Gas concentrations
DAT
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
7-28 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
A
A
A
Menu: 84 POLYGAS6
Menu: 85 POLYGAS7
Menu: 86 POLYGAS8
Point 11 Gas value: Raw readi ng : Linearized value: Point 12 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Point 13 Gas value: Raw readi ng : Linearized value: Point 14 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Point 15 Gas value: Raw readi ng : Linearized value: Point 16 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Gas concentrations
000.0
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
ESCAPE INFO
ESCAPE INFO
DAT
Gas concentrations
DAT
Gas concentrations
DAT
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-29
Instruction Manual
A
A
760007-A July 2003
Menu: 87 POLYGAS9
Menu: 88 POLYGAS0
Menu: 89 MIDPOINT2
Gas concentrations
Point 17 Gas value: Raw readi ng : Linearized value: Point 18 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Point 19 Gas value: Raw readi ng : Linearized value: Point 20 Gas value: Raw readi ng : Linearized value:
Point to be measured:
HOME
Range 2 Correction: Point being measured: Point 1 gas concentration: Point 2 gas concentration: Point 3 gas concentration: Point 1 reading: Point 2 reading: Point 3 reading: Span gas value:
HOME
ESCAPE INFO
Gas concentrations
ESCAPE INFO
Midpoint correction set up
ESCAPE INFO
DAT
DAT
SET
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
BACK
000.0
000.0
000.0
000.0
000.0
000.0
000.0
0.000
000.0
NEXT
7-30 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 90 MIDPOINT3
Menu: 91 MIDPOINT4
Menu: 92 EXP_CAL
Range 3 Correction: Point being measured: Point 1 gas concentration: Point 2 gas concentration: Point 3 gas concentration: Point 1 reading: Point 2 reading: Point 3 reading: Span gas value:
HOME
Range 4 Correction: Point being measured: Point 1 gas concentration: Point 2 gas concentration: Point 3 gas concentration: Point 1 reading: Point 2 reading: Point 3 reading: Span gas value:
HOME
Measurement range number: Zero gas concentration: Span gas concentration: Sample flow:
Raw measurement signal: Ranges with valid calibration: Status: Result…
HOME
Midpoint correction set up
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
ESCAPE INFO
Midpoint correction set up
ESCAPE INFO
Zero/span calibration
FACTORS INFO
SET
SET
ZERO
NEXT
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
BACK
000.0
000.0
000.0
000.0
000.0
000.0
000.0
SPAN
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-31
Instruction Manual
760007-A July 2003
Menu: 93 EXP_CAL_DAT
Menu: 94 EXP_CAL_DATI
Menu: 95 UNITS
Zero/span diagnostic data
Date of last zero: Error message for last zero: Error percentage for last zero: Raw signal at last zero: Last zero gas would read: Date of last span: Error message for last span: Error percentage for last span: Raw signal at last span:
HOME
Shows what happened at the last calibration. The errors are expressed as a percentage of range. The last zero and span readings are how the analyzer would read on those gases with the current calibrati on factors.
HOME
Gas measurement units: Pressure measurement units: Temperature measurement units: Ppm to mg/Nm3 conversion factor:
Lower explosion limit (LEL): Upper explosion limit (UEL):
HOME
FACTORS INFO
Zero/span diagnostic data
ESCAPE INFO
Units
ESCAPE INFO
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-32 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
K
Menu: 96 UNITSI1
Menu: 97 POLYSETI2
Menu: 98 POLYSETI3
Select the units in which you want the values to be displayed. This does not affect the variabl e contents, it merely affects how the control module display them.
Note that all analyzer ranges wi ll be set as perce n t or ppm, you can’t set some as ppm and others as percent.
HOME
When you have entered the desired number of points, return to the polynomial set up screen, and press Polynomial, and stor e it as the coefficients in the current range’s linearization function. The order of the polynomial is optimized based onthe number of data points provided. You need at least 7 points for a fourth order polynomial correction. You can modify the results with the piecewis e li n e ar correction also provided in thi s sectio n .
HOME
WARNING: The linearization curve must be monoto nic. If it is not, the calibration routine will fail and the analyzer will not calibrate. Test this by copying the values of the linearization coefficients into a spreadsheet program and plotting the result. The analyzer does test for monotonicity when it spans, but this test may not catch all possible errors. Monotonic means that the curve does not roll over and start going back down as the gas concentration
HOME
Units
ESCAPE INFO
Polynomial set up
ESCAPE INFO
MORE
Polynomial set up
BACK
ESCAPE INFOBAC
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-33
Instruction Manual
760007-A July 2003
Menu: 99 RESET
Menu: 100 STORE
Menu: 101 RFHIST1A
Reset
Are you sure? RESET will erase ALL the configuration and
manufacturing data, including serial numbers and everything else.
If you are sure, press RESET again.
HOME
Are you sure? STORE will copy current diagnostic data into the
historical ( currently there.
If you are sure, press STORE again.
HOME
Manufacturer’s settings. Zero offset: Span factor:
Stored settings Zero offset: Span factor:
HOME
ESCAPE INFORESET
Store historical data
ESCAPE INFO
NEXT INFORSTR STRSTR MN
STORE
Range 1 Factors
Model NGA2000 PMD
000.0
000.0
000.0
000.0
7-34 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 102 RFACTORSIA
Menu: 103 RFHIST2A
Menu: 104 RFHIST3A
Shows the calibration factors for this range. Modify the zero factor for zero calibration, and the span factor for spanning this range. The take effect as soon as you press the enter key. With zero gas, the zero factor should be the same as the raw reading. RSTR MN restores the manufacturing values. RSTR ST restore the stored val ues.
HOME
Manufacturer’s settings. Zero offset: Span factor:
Stored settings Zero offset: Span factor:
HOME
Manufacturer’s settings. Zero offset: Span factor:
Stored settings Zero offset: Span factor:
HOME
Range Factors
ESCAPE INFO
Range 2 Factors
000.0
000.0
000.0
000.0
NEXT INFO
NEXT INFORSTR STRSTR MN
RSTR MN
Range 3 Factors
RSTR ST
000.0
000.0
000.0
000.0
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-35
Instruction Manual
K
760007-A July 2003
Menu: 105 RFHIST4A
Menu: 106 TWEAKI1
Menu: 107 ANALSETI3
Model NGA2000 PMD
Range 4 Factors
Manufacturer’s settings. Zero offset: Span factor:
Stored settings Zero offset: Span factor:
HOME
This function allow s you to set up to th r ee midpoints that the analyzer will It does this with a piece-wise linear alg ori t hm. This polynomial linearization. First disable the correction. Set the Then enter the first midpoint gas value, run the gas, and when stable, press SET. the actual reading, but the analyzer will
HOME
Then go to the second set point, and repeat. You can use up to three midpoints. When you are done, set the correction to WARNING: make sure that you do not have excessive corrections. If the correctio n is too odd, the calibration routine will fail, and you will not be able to calibrate the anal yzer. In this case, try it again.
You can perform this correction individually for each range.
HOME
NEXT INFO
Midpoint correction set up
ESCAPE INFO
Midpoint correction set up
ESCAPE INFOBAC
RSTR MN
MORE
RSTR ST
000.0
000.0
000.0
000.0
7-36 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 108 ACALSETI1A
Menu: 109 INIT
Menu: 110 SWDIAGI1
Disable the limits to reco v er from calibration failure Calibration averaging time sets the time used by the analyzer to average its reading. Longer times will give a better calibration. Calibration failure alarm will issue a WARNIN G if the analyzer has to change calibration by more than the Cal failure error, if warning alarms are enabled. Calibration time out sets how long the analyzer will wait for the signal to stabilize before issuing a WARNING. You can zero or span the ranges all at once or not.
HOME
Are you sure? INIT will erase ALL the configuration data, including
manufacturing data, serial numbers etc.
If you are sure, press INIT again.
HOME
Shows the first detected software error since the variable on the bottom line was r eset. Please report any errors to your service representative. They may mean nothing.
The analyzer has a lot of error recovery code. Errors may therefore correct themselves.
HOME
Calibration Parameters
ESCAPE INFO
Re-initialize the analyzer
ESCAPE INFO
Software Diagnostics
ESCAPE INFO
INIT
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-37
Instruction Manual
A
A
760007-A July 2003
Menu: 111 STOREDPVA
Menu: 112 ZERO_NOW
Menu: 113 SPAN_NOW
Model NGA2000 PMD
Trend display control
The analyzer stores 24 hours of 15 minute averages. These values are only accessible via a PC. Use the variables DATA_INDEX and D ATA_POINT to access them.
HOME
Are you sure? You must have zero gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press ZERO again now. Press the left arrow key when you are done. Calibration status: 000.0
HOME
Are you sure? You must have span gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press SPAN again now. Press the left arrow key when you are done. Calibration status: 000.0
HOME
ESCAPE INFO
nalyzer zero
ESCAPE INFO
ESCAPE INFO
ZERO
nalyzer span
SPAN
7-38 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
Menu: 114 CALFAIL
Menu: 115 ABOUT
Menu: 116 ABOUT1
Check that you are flowing the correct gas, and the gas concentration is what it is supposed to be. Make sure that the reading is stable before staring. If you have enabled or disabled the linear iz er, you may have made it hard for the analyzer to calibrate. If so, go to the calibration parameters screen under Expert controls and set up, under Analyzer set up, and disable the limits checki ng. Recali brate, and then enable the limits checking aga in. If all else fails, manually adjust the calibration factors
HOME
(C) Copyright Fisher-Rosemount Analytical Inc., 2000 Manufactured by:
Rosemount Analytical Inc. 4125 East La Palma Avenue Anaheim, CA 92807-1802 /USA Tel: (714) 986-7600 FAX: (714) 577-8739
Measure
-- Analyzer Module Version Information --
Serial number: Manufacturing date:
Hardware revision: Software revision: Revision date: Revision time:
Measure
If it won’t calibrate…
ESCAPE INFO
Back…
Back…
More…
000.0
000.0
000.0
000.0
000.0
000.0
Instruction Manual
760007-A July 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Appendix A. Menu Displays 7-39
Instruction Manual
A
760007-A July 2003
Menu: 117 ALARM1
Menu: 118 MANDATA
Menu: 119 AUTORANGE
Concentration Alarm Setup
Alarm generation is: Level for Low-Low alarm: Level for Low alarm: Level for High alarm: Level for High-High alarm: Alarm delay: Low-Low alarm: Low alarm: High alarm:
HOME
Serial number:
Set manufacturing date!
Actual date:
Measure Back…
Actual switch levels… Switch level hysteresis: Usage of range -1: Usage of range -2: Usage of range -3: Usage of range -4: Automatic range change control:
Absolute, range upper limit:
Measure
ESCAPE
-- Manufacturing data… --
-- Autom a tic Range Control --
CKN
Back…
Model NGA2000 PMD
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
000.0
7-40 Appendix A. Menu Displays Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD, NDIR
VAISALA BAROMETRIC PRESSURE TRANSDUCER
Addendum
748461-B
April 2003
This addendum serves as an amendment to the Model NGA 2000 PMD and NDIR instruc­tion manuals. The following information should be considered part of the manual, if your instrument has a Vaisala Barometric Pressure Transducer installed in it, or if you are retrofitting a PMD or NDIR with a Vaisala transducer kit.
The Vaisala PTB100 series Analog Barometer is a N.I.S.T. traceable precision silicon capaci­tive atmospheric pressure transducer. The sensor delivers an analog voltage output that is inversely proportional to its inlet pressure. It has an operational ambient pressure range of 800 to 1060 hPa.
The original PMD and NDIR instruments were designed for use with a resistance bridge type of pressure transducer requiring an external gain amplifier with an offset trim The Vaisala transducer contains an integral gain amplifier that is scaled for a 0 to +5 Vdc output. The 661380 “Interface Board” contains the cir­cuitry to convert the Vaisala single ended 0 to +5 V dc output into the referenced bridge 0 to +0.625 V dc output configuration that is re­quired by the Analyzer module’s Computer Analysis Board. The Vaisala transducer kit is backward compatible with the original Rose­mount pressure transducer.
Vaisala Interface Board Circuit Function:
The Vaisala Interface Circuit board adapts the single ended analog output of the Vaisala sensor into a Precision Voltage referenced, balanced bridge (at 1 atmosphere) input con­figuration. A “Bridge Balance” pot on the Vaisala Interface Circuit board provides the fine trim “Bridge Balancing” calibration ad­justment that is required to set up each indi-
vidual transducer at a known ambient air pressure. Test points are provided on the cir­cuit board for this adjustment.
The Vaisala transducer output range is 0 to +5 Vdc.
The Vaisala analog output voltage is inverted and fed into a balanced summing node that is referenced by a precision regulator. The summing node voltage is 0.000 Vdc at 1013 hPa (1 atmosphere ambient air pressure). The summing node voltage is then buffered, in­verted, and gain reduced by a factor of 8 to produce the Pressure Compensation Output voltage at J12-1 for the Computer Analysis board.
Low noise, Low input current offset Op-Amps, and common substrate resistors are used to reduce gain tracking errors and null thermal coefficient effects.
1-1 INSTALLATION & RETROFIT
INSTRUCTIONS
1. Remove the original Rosemount trans­ducer circuit board assembly, located on the motherboard at J12 and J13 (if pre­sent).
2. Install the transducer with the pressure inlet facing the rear of the instrument us­ing the mounting studs provided on left­hand side of the detector housing.
3. Install the Interface board into it’s mating sockets (J12 and J13) on the mother­board.
4. Connect the 4-pin cable (P1) from the in­terface board (noting pin polarity) to the Vaisala Pressure Transducer.
http://www.processanalytic.com
Addendum
748461-B April 2003
5. Connect the instrument sample inlet tub­ing to the Vaisala Pressure Inlet port.
Model NGA2000 PMD, NDIR
Barometric Pressure menu reads the correct ambient air pressure.
6. Enable the Barometric Pressure Com­pensation and the Use Local Reading op­tions via the Expert Controls – Analyzer Module Diagnostics – Barometric Pres­sure Compensation menus of the instru­ment. The instrument will automatically detect and report the ‘Presence’ of the Pressure Compensation Option.
1-2 ADJUSTMENT PROCEDURE
Circuit adjustment can be made using Baro­metric Pressure menu display, or an external Digital voltmeter.
Display Method
With no sample inlet lines connected and the instrument exhaust ported to atmosphere, ad­just R1 on the 661380 Interface board until the Barometric Pressure reading on the
DMM Method
Connect the DMM (+) to TP1 on the 661380 Interface board, (-) to TP2. Set the DMM to measure mVDC. With no sample inlet lines connected and the instrument exhaust ported to atmosphere, adjust R1 on the Interface board until the DMM reads 0.000 Vdc ± 0.001 V. The instrument is now calibrated at 1 ATM.
1-3 HIGH ALTITUDE VERSION OPTION
For applications where the instrument is to be used above 5400 Feet (MSL) an alternate transducer option is available from the factory. This version consists of a Vaisala model PTB100B transducer (900-1100 hPa range). Please contact the factory if you require this option.
2 of 2
Addendum – Vaisala Transducer Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy.
DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure.
Seller's direct control.
Seller shall not be liable for failure to perform due to labor strikes or acts beyond
SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY
Instruction Manual
A
760007-A July 2003
Model NGA2000 PMD
EMERSON PROCESS MAN
ROSEMOUNT ANALYTICAL INC. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd. 1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
GEMENT
EUROPEAN TECHNOLOGY CENTER Fisher-Rosemount GmbH & Co. Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd. Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...