Emerson NGA2000 User Manual

NGA 2000
Software Manual
CLD - Analyzer Module
Chemiluminescence - Measurement
(combined with NGA 2000 - Platform / MLT - Analyzer)
CLD Software V e rsion 2.3
(Platform / MLT Software Version 3.1.X)
Catalog No.: 90 003 749
90003749(1) [NGA-e (CLD-Software 2.3 / MLT-Software 3.1.X)] 04/98
Managing The Process Better
This Operation Manual includes information about the operation of the instrument. Information about the additional indications and notes regarding maintenance, troubleshooting and repair are found in the accompanying Maintenance & Operation Manual.
T roubleshooting, component replacement and internal adjustments must be made by qualified service personnel only.
Fisher-Rosemount GmbH & Co does not take responsibility for an y omissions or errors in this manual. Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of this manual, is expressly e xcluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations. T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
Misprints and alterations reserved
©
1998 by FISHER-ROSEMOUNT GmbH & Co. (PAD)
1. Edition: 04/98
Read this operation manual carefully before attempting to operate the analyzer ! For expedient handling of reports of defects , please include the model and serial number which can be read on the instrument identity plate.
Fisher - Rosemount GmbH & Co.
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
Telefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
90003749(1) [NGA-e (CLD-Software 2.3 / MLT-Software 3.1.X)] 04/98

Contents

1 Introduction 1 - 1
2 Structure of Menus 2 - 1
3 Display and Keyboard 3 - 1
3.1 Starting and Initializing...................................................................................3 - 1
3.2 Display and Function...................................................................................... 3 - 1
3.3 Operating Keys ..............................................................................................3 - 2
3.4 Lines...............................................................................................................3 - 2
3.5 Important Functions of the Softkeys...............................................................3 - 3
3.6 Entering/Changing of Variables......................................................................3 - 4
3.7 The STATUS Key...........................................................................................3 - 5
4 Basic Controls 4 - 1
4.1 Measurement 4 - 3
4.1.1 Multi Component Display: Change of Channel ..............................................4 - 3
4.1.2 Multi Component Display: Sequence Setup...................................................4 - 5
4.1.3 Range Setting ................................................................................................4 - 7
4.1.4 Range and Functional Control .......................................................................4 -11
4.1.5 Enable/Disable Ozonator Power....................................................................4 -15
4.1.6 Measurement Mode: NO ↔ NOx....................................................................4 -19
4.2 Calibration 4 -21
4.2.1 Zeroing...........................................................................................................4 -21
4.2.2 Spanning........................................................................................................ 4 -25
4.2.3 If It Won't Calibrate.........................................................................................4 -29
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
I
5 Expert Controls and Set up 5 - 1
5.1 Analyzer Module Set up 5 - 3
5.1.1 Calibration/Calibration Gas List......................................................................5 - 4
5.1.2 Calibration Parameters...................................................................................5 - 9
5.1.3 Gas Measurement Parameters ......................................................................5 -11
5.1.4 Linearization Parameters................................................................................5 -12
5.1.5 Response Time/Delay Parameters.................................................................5 -14
5.1.6 Range Settings...............................................................................................5 -16
5.1.7 Units ...............................................................................................................5 -17
5.1.8 Linearization Functions...................................................................................5 -18
5.1.9 Analyzer Parameter List .................................................................................5 -23
5.1.10 Physical Measurement Parameters................................................................5 -26
5.1.11 Displayed Parameters ....................................................................................5 -28
5.2 Auxiliary Module Controls 5 -29
5.2.1 Local SIO Configuration Sections...................................................................5 -30
5.2.2 Set-up DIO Module(s).....................................................................................5 -39
5.3 Auxiliary Module Set up 5 -45
5.4 System Set up 5 -47
5.4.1 Front Panel Control ........................................................................................5 -48
5.4.2 Display Resolution..........................................................................................5 -50
5.4.3 Auxiliary Lines.................................................................................................5 -51
5.5 Expert Analyzer Controls 5 -53
5.5.1 Range Settings...............................................................................................5 -55
5.5.2 Zero/Span Calibration.....................................................................................5 -56
5.5.3 Physical Measurements..................................................................................5 -60
II
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
6 Technical Level Configuration 6 - 1
6.1 System Set up 6 - 3
6.1.1 Main Display Configuration ............................................................................6 - 4
6.1.2 Front Panel Control........................................................................................6 - 9
6.1.3 Date and Time Set up....................................................................................6 -11
6.1.4 Module Binding ..............................................................................................6 -12
6.1.5 System Reset.................................................................................................6 -14
6.1.6 Security Codes...............................................................................................6 -15
6.2 Service Menus 6 -17
6.2.1 Control Module Data......................................................................................6 -18
6.2.2 Analyzer Module Data.................................................................................... 6 -19
6.2.3 Control Module History...................................................................................6 -21
6.2.4 Analyzer Module History ................................................................................6 -23
6.3 Diagnostic Menus 6 -25
6.3.1 Control Module Diagnostics ...........................................................................6 -26
6.3.2 Analyzer Module Diagnostics.........................................................................6 -27
6.3.2.1 Power Supply Voltages ...................................................................6 -28
6.3.2.2 Primary Variable Parameters ..........................................................6 -29
6.3.2.3 Physical Measurements ..................................................................6 -33
6.3.2.4 Temperature Control.......................................................................6 -35
6.3.2.5 Flow Control Parameters.................................................................6 -36
6.3.2.6 Miscellaneous Control Parameters..................................................6 -38
6.3.2.7 Trend Display Control......................................................................6 -39
6.3.2.8 Software Diagnostics.......................................................................6 -41
6.3.2.9 Analyzer Start up............................................................................. 6 -42
6.3.2.10 NOx Converter Efficiency.................................................................6 -43
6.4 Other Module Diagnostic Menus 6 -44
6.5 Listing of all Modules 6 -45
Supplement: System Calibration
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
III
IV
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
1 Introduction
This software manual describes step by step how to operate successfully with the ½ 19" and 19" MLT analyzer module and analyzer of the NGA 2000 Series from . Chapter two shows the structure of the software menus. Chapter three describes the display and the keyboard of the analyzer. Chapter four describes the basic controls with detailed illustrations. So you can easily compare the actual display of the analyzer module with the illustrations of the manual. Chapter five describes the expert controls and chapter six describes the technical level configuration. The layout of both chapters is not as detailed as in chapter four. Normally, the way to a certain software menu is described with the software catchwords, you have to press to reach this menu. You will find the illustration of the corresponding LCD screen at the end of the catchword listing. After that you can read the meaning of the functions and variables of each expert or technical level menu. Some contents of the expert controls are not important for every customer. It depends on the configuration of your NGA 2000 system, relative to the following components:
Control Module CMAnalyzer Module AMInput/Output Modules I/O's (SIO, DIO)
You can distinguish the following system units and SIO/DIO configurations:
System Unit SIO/DIO-Configuration Corresponding Chapter
1
CLD analyzer module (AM):
without front panel i.e.
Local I/O's are not existing
⁄⁄
without control unit
can be combined with a
platform or an MLT analyzer
Platform (CM software):
Control unit with front panel
without measurement
channels
1 SIO and up to 4 DIO's
can be installed in the platform
SIO and DIO can be
chapter 5.2
p. 29 to 44
configured for all channels combined with the platform
MLT analyzer (CM and MLT AM software):
Analyzer with front panel
CM and AM software in the
same analyzer, i.e. all functions of the control unit and of the AM are united
1 SIO and 1 DIO can be
installed in the MLT analyzer
SIO and DIO can be con-
figured for all channels combined with the MLT analyzer
chapter 5.2
p. 29 to 44
in the same implement
1
The configuration of other analyzer modules like MLT modules is described in their own manuals !
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
1 - 1
The following illustrations shall make plain the connection between the hardware configu­ration and the software setup of the modules:
NGA 2000 System via Platform
(see 5.1.17) (see 5.1.18)
(other manuals)
(other manuals)
(other manuals)
Local
I/O's
SIO DIO
Analyzer
Modules (AM's)
Control
Module (CM)
MLT
CLD
FID
Platform
PMD
NGA 2000 System via MLT Analyzer
Auxiliary Modules
SIO
DIO
Other I/O´s
(see 5.2.1)
(see 5.2.2)
(other manuals)
Local
I/O's
(see 5.1/5.5)
SIO DIO
Analyzer
Modules (AM's)
CLD
MLT
FID(other manuals)
PMD(other manuals)
Control
Modul (CM)
MLT Analyzer
Auxiliary Modules
SIO
DIO
(see 5.2.1)
(see 5.2.2)
1 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Notes:
*
In the menu structure shown above, you will find the most important ramifications of the MLT software 3.1.X for the
MLT platform and analyzer versions. The menu points related to the CLD analyzer module are marked with
They belong to the CLD software revision number 2.3.
*
At each menu point of the "Expert controls" and "Technical level" you can find which functionality will be set up:
odule), CM (
odule), I/O (
utput Module).
*
Menu lines ending with three points (...) are followed by submenus with further functions and set-ups.
-
-
-
-
-
Technical level
configuration...
2 Structure of Menus
"AM".
Main Menu
(CM)
- System set up...
and set up...
Expert controls
controls...
- Expert analyzer
(AM, CM, I/O's:
SIO, DIO and other I/O's)
- Service menus...
(AM)
controls...
- Auxiliary module
(AM, CM, I/O's:
SIO, DIO and other I/O's)
- Diagnostic menus...
(I/O's: SIO, DIO)
(CM)
- System set up...
diagnostic menus...
(Other I/O's)
- Other module
(AM)
- Analyzer module set up...
(AM, CM, I/O's:
SIO, DIO and other I/O's)
- Listing of all modules...
(Other I/O's)
- Auxiliary module set up...
(CM)
- System calibration...
Input-/O
Control M
Basic controls... (AM)
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Range set-up
Range and
functional control
Ozonator power
enable/disable
NGA 2000
Measurement mode
selection: NO/NOx-Zeroing...
Spanning...
AM (Analyzer M
2 - 1
2 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X] 04/98
3 Display and Keyboard

3.1 Starting and Initializing

After switching on the CLD analyzer module (in a platform or part of a NGA network), the initialization procedure will be performed. A self control of the analyzer modules or the analyzer(s) is running. You can see a sequence of several displays. They show the status of initialization, revision notes of the MLT software and the tag. At the end of this procedure you can see the single component display of channel one. It is the origin to all the other channels, menus and submenus. The instructions of the basic controls (chapter four) are all beginning with the single com­ponent display. The actual display might differ from the shown one because the customer can configure it according to his requirements.
TAG
here e.g.: CLD Analyzer Module
CLD
95.00 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Typical starting menu after switching on
Range: 4
340 hPa
0.5 C
51.5 C 360 C

3.2 Display and Function

The LCD-screen shows all measurement values of the analyzer and all customer instruc­tions. You can operate with five function keys, four arrow keys (cursors) and the enter key. The function of each key depends on:
the type of analyzer/analyzer module usedthe optional auxiliary modules (e.g. I/O boards) usedthe individual menu displayed
In case of power failure all customer specific module parameters are saved by a battery­powered buffer.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 1

3.3 Operating Keys

CLD
95.00 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 4
340 hPa
0.5 C
51.5 C 360 C
Function Keys:
Keys without defined functionsThe current function depends on
the menu selected
The softkey legend is shown on
the display above the key

3.4 Lines

Cursor keys:
↑↑ -key / ↓↓ -key:
Line up / line down
within the same menu
Alteration of numbers,
variables or digits
-key/ -key:
Moving back/forwards
between the pages of a menu
Selection of digits
Enter Key:
To confirm a previously entered value
(variable)
To start a selected function
(Alternative: -key)
To go into a menu (via menu line)
Lines can be selected by the ↓↓ -key or the ↑↑ -key. The selected line is displayed white on black. You have four different types of lines in the menu:
Menu line...
Line ending with three dots.You will go to a submenu by pressing the ENTER key or the → -key in such a line.
Function line !
Line ending with an exclamation-mark.You will start a function (e.g. NO/NO
-toggle) by pressing the ENTER key or
x
the → -key in such a line.
Line of variables:
Line ending with a colon.Display of module parameters (variables).Some parameters can be changed (e.g. range number), some parameters display only
a status (e.g. calibration status) and cannot be changed.
Text line
Line without any punctuation marks.Only display of informations (e.g. analyzer module tag).
3 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
3 Display and Keyboard
Text Line
Line of Variables
Function Line
Menu Line
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Zero/span calibration
Measurement range number: 3 Zero gas concentration: 0.0 ppm Span gas concentration: 100.0 ppm Sample flow: 1100 ml/min Raw measurement signal: 521590 Measurement gas: NOx NO/NOx toggle! Status: READY Result... Calibration adjustment limits: Disabled
HOME FACTORS ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Function Key Legend
3.5 Important Functions of the Softkeys DISPLAY
Change from the single component display to the multi component display.
MEASURE
Change from the main menu to the single component display.
STATUS (see 3.7!)
Change to the status report: Display of the most important parameters and
status informations of the CLD module or of the current channel.
If available, this command is always assigned to the F2 key.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 3
MENU
Change from the single component display to the main menu or from several submenus
to higher menus.
ESCAPE or <<<
Changing back to the last menu page selected.Reset of a changed but not confirmed parameter to the former value resp. to zero.
BACK
Reset of a changed but not confirmed parameter to the former value.
LOCK
Lock of any operation level, if activated. Each of the three operation levels has to be
activated separately.
F4 key of the main menu.
NEXT
Change of the channel in the same menu. This will only be possible, if several channels
of an analyzer or of combined analyzers are existing. Then, all available channels can be selected one after another.
In the main menu you can move among all channels of the connected analyzers and
analyzer modules. In the submenus you can only move among the channels of the current analyzer or analyzer module.
INFO
Change to menus containing status or help informations of the actual menu.If available, this function is always assigned to the F5 key.
3.6 Entering/Changing of Variables ENTER Key
If you will press this key in a line of variables, the parameter will be displayed white on
black and can be changed. After you will have set up a new value, you can confirm it by pressing the ENTER key again.
↑↑ -key / ↓↓ -key
Function will depend on the variable selected: - Changing of the parameter values
- Scrolling among variables selected
- Changing of digits or characters
Numbers consisting of several digits will increase/decrease totally, not digit by digit.
-key / -key
Selection of digits within a parameter.For some variables you can change the quantity of digits or characters.
3 - 4
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
3 Display and Keyboard

3.7 The STATUS Key

From the single component display or from the main menu you can change with the F2 key (STATUS) to the menu "Current measurement parameters":
CLD
Current measurement parameters
Analyzer gas measured: NOx Measurement range number: 3 Range change control: LOCAL
Linearization mode: Disabled Analyzer operational state: STANDBY
Analyzer alarm state: NORMAL Alarms reported: FAILURE
HOME ESCAPE MORE INFO
F1 F2 F3 F4 F5
CLD
Current measurement parameters
Response time: 3.0 s Bypass flow: 1100 ml/min Capillary pressure: 340 hPa Detector temperature: 0.5 C
95.0 ppm
95.0 ppm
HOME ESCAPE MORE INFO
F1 F2 F3 F4 F5
In the menu "Current measurement parameters" you can control the status of the CLD analyzer module. With the F3 key (MORE) you can change to a further menu page of this menu. With the F2 key (ESCAPE) you can always go back to the main menu. From the second menu page you can return to the first with the F4 key (MORE) or the ← -key. In the menu "Current measurement parameters" you will only find status informations. You cannot change the setups there. If you want to modify some parameters, you have to change to the menu "Basic Controls" (see chapter 4) or to the menu "Analyzer module set up" (see chapter 5.1) resp. "Expert analyzer controls" (see chapter 5.5).
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 5
3 - 6
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4 Basic Controls
In the chapter "Basic Controls" you will find the most important functions to set up your CLD analyzer module via an NGA front panel:
1) Measurement p. 4 - 3pp (Chapter 4.1.1 and 4.1.2, p. 4-3 to 4-6, are only important for systems with several channels.)
2) Calibration p. 4 - 21pp
All steps are figured with detailed illustrations and operation instructions. In the left column you can see display and keyboard of the NGA front panel. In the right column you can read the instructions and notes. All instructions will begin with any single component display and will end with the corresponding single component display after the setups are done. The keys you have to press are illustrated in black. The instructions are completed with notes and further informations.
Example: You want to change from the single component display of NOx to the main menu
Picture one shows the starting situation: single component display of NO
.
x
If you press the F3 key you will change to menu page shown in picture 2: Main Menu.
Left column: Right column: LCD and keyboard Instructions and notes
CLD
95.00 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
CLD
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
F1 F2 F3 F4 F5
Range: 4
340 hPa
0.5 C
51.5 C 360 C
95.0 ppm
Main Menu
Change to the
Main Menu
F3
Press
or
Next instruction
or step, here e.g.: Change to the Basic Controls:
Press
or
90003479(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 1
4 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

4.1.1 Measurement - Multi Component Display:

CLD
95.00 ppm NOx
0 250
Range: 4
Change of Channel
Change to the multi
component display
F1
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
340 hPa
0.5 C
51.5 C 360 C
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
Note: You can change to the multi component display from each single component display.
Enable the selecting:
">"
F1
Press
or
Note: If no channels of other analyzer modules are combined to the CLD, only the bargraph of the CLD module will be displayed. Then, a change of channel will not be possible !
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
NGA 2000
Select any channel
Press or as often as necessary to
put the > - mark into the selected line.
Example: Change from NOX (CLD) to CO (MLT channel 1)
4 - 3
CLD
95.00
45.00
333.0
150.0
20.00
CLD
ppm NOx
MLT25/CH1/R2
ppm CO
MLT25/CH2/R2
ppm SO2
MLT25/CH3/R2
ppm NO
MLT25/CH4/R2
%O2
Change to the single
component display of the channel selected
F1
Press
SELECT STATUS TAGS OFF
INFO
F1 F2 F3 F4 F5
MLT25/CH1/R2
45.00 ppm CO
0 250
Temperature: Function Control: Maintenance Request: Failure:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 2
25.0 C No No No
Single component
display of the channel selected will appear
4 - 4
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4.1.2 Measurement - Multi Component Display:
CLD
95.00 ppm NOx
0 250
Range: 4
Sequence setup
Change to the multi
component display
F1
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
340 hPa
0.5 C
51.5 C 360 C
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
Enable the selecting:
">"
F1
Press
or
Note: If no channels of other analyzer modules are combined to the CLD, only the bargraph of the CLD module will be displayed. Then, the sequence setup will not be possible !
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
NGA 2000
Select the channel
for position one
Press or as often as necessary to
put the > - mark into the selected line.
Example: MLT channel 2 (SO2) shall appear at position one of the multi component display.
4 - 5
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
MLT25/CH2/R2
333.0
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
MLT25/CH2/R2
95.00
45.00
150.0
20.00
SELECT STATUS TAGS OFF
MLT25/CH1/R2 MLT25/CH3/R2 MLT25/CH4/R2
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
ppm SO2
CLD
ppm NOx
ppm CO ppm NO
%O2
INFO
INFO
Assignation of the
channel selected to position one
Press
Note: The channel selected will appear at position one. All the other channels will automatically appear one line below their last position.
Multi component
display will appear with the changed sequence
Note: Now, you can change to any available single component display. The method is described in the last chapter (4.1.1).
F1 F2 F3 F4 F5
4 - 6
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.00 ppm NOx
0 250
Range: 4

4.1.3 Measurement - Range Setting

Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Choose the
measurement range number
Press
or
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 7
CLD
Basic Controls
Measurement range number: Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
4
ESCAPE INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
4
ESCAPE INFO
95.0 ppm
F1 F2 F3 F4 F5
Select the
range number
4
Press or as often as necessary to select another range number.
Example: Change from range 4 (250 ppm) to range 3 (100 ppm)
Confirm the
range selected
3
Press
Note: If you don't want to set the new range number, you can reactivate the former selection: Press F2 (ESCAPE) and confirm it with ENTER.
CLD
Basic Controls
Measurement range number: 3
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
F1 F2 F3 F4 F5
4 - 8
95.0 ppm
NGA 2000
Change to the
main menu
Press or
F1
2 times
Note: The range upper limit (line 2) of the range selected will be set up automatically after the range setting is confirmed.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
F1 F2 F3 F4 F5
4.1.3 Measurement - Range Setting
Change to the single
component display
F1
Press
CLD
95.00 ppm NOx
0 100
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 3
340 hPa
0.5 C
51.5 C 360 C
Single component
display will appear with the new range
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 9
4 - 10
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

4.1.4 Measurement - Range and Functional Control

CLD
95.00 ppm NOx
0 250
Range: 4
Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Cahnge to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the
line "Range and functional control"
Press or as often as necessary until the line "Range and functional control" will appear white on black.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 11
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm
Range and functional control: LOCAL
Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
LOCAL
ESCAPE INFO
95.0 ppm
LOCAL
F1 F2 F3 F4 F5
Choose the
parameter
Press
or
Set up the range
and functional control you want
++ Pay attention to
the note at p. 4-14 !
Press or as often as necessary to select the parameter you want.
Options: LOCAL, AUTO, REMOTE
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
LOCAL
ESCAPE INFO
F1 F2 F3 F4 F5
4 - 12
95.0 ppm
AUTO
NGA 2000
Confirm the
parameter selected
Press
Note: If you don't want to set the new parameter, you can reactivate the former selection: Press F2 (ESCAPE) and confirm it with ENTER.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4.1.4 Measurement - Range and Functional Control
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm
Range and functional control: AUTO
Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
F1 F2 F3 F4 F5
95.0 ppm
Change to the
main menu
Press or
F1
2 times
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
95.00 ppm NOx
0 250
Range: 4
Change to the single
component display
F1
Press
Single component
display will appear after the changing of the range and functional control
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 13
Range and functional control – Options:
LOCAL:
The platform will be controlled via the display.
AUTO:
The range will be selected automatically ("auto range"). This function will only be available with an I/O board with 3 alarms (PIN 70 656 193), because the auto ranging parameters are located on the I/O board (see its own manual)!
REMOTE:
The range selection will be controlled by digital inputs, e.g. via a DIO board or I/O board with 3 alarms.
4 - 14
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

4.1.5 Measurement - Enable/Disable Ozonator Power

CLD
95.00 ppm NOx
0 250
Range: 4
Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the line
"Ozonator power"
Press or as often as necessary until the line "Ozonator power" will appear white on black.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 15
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On
Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
Enabled
ESCAPE INFO
95.0 ppm
Enabled
Choose the
parameter
Press
or
Select the ozonator
power status you want
Press or to select "Enabled" or "Disabled".
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
Enabled
ESCAPE INFO
95.0 ppm
Disabled
F1 F2 F3 F4 F5
Confirm the setup
Press
Note: If you don't want to confirm the new selection, you can reactivate the former one: Press F2 (ESCAPE) and confirm it with ENTER.
4 - 16
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4.1.5 Measurement - Enable/Disable Ozonator Power
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: OFF - DISABLED
Ozonator power: Disabled
HOME NO/NOx ZERO SPAN INFO
F1 F2 F3 F4 F5
95.0 ppm
Change to the
main menu
Press or
F1
2 times
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
95.00 ppm NOx
0 250
Range: 4
Change to the single
component display
F1
Press
Single component
display will appear after the setup of the ozonator power status
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 17
4 - 18
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4.1.6 Measurement - Measurement Mode: NO NO
CLD
95.00 ppm NOx
0 250
Range: 4
x
Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Select NO NO
x
F2
Press to select the measurement mode you want.
The current selection will be displayed in the line "Measurement mode".
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 19
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NO Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
13.8 ppm
Change to the
main menu
Press or
F1
2 times
Change to the single
component display
F1
Press
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
F1 F2 F3 F4 F5
CLD
13.75 ppm NO
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
Range: 4
340 hPa
0.5 C
51.5 C 360 C
Single component
display will appear with the measure­ment mode selected
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
4 - 20
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.00 ppm NOx
0 250
Range: 4

4.2.1 Calibration - Zeroing

Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the
submenu "Analyzer zero"
F3
Press
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 21
CLD
Analyzer zero
Are you sure? You must have zero gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press ZERO again now. Press the left arrow key when you are done. Calibration status: READY Error message for last zero: CAL OK
HOME ESCAPE ZERO INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Analyzer zero
Are you sure? You must have zero gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press ZERO again now. Press the left arrow key when you are done. Calibration status: ZEROING - WAIT Error message for last zero: CAL OK
ABORT
13.7 ppm
Start Zeroing
++ Pay attention to
the note at p. 4-23 !
F3
Press
Caution:
You have to
flow enough zero gas
through the analyzer
look for a stabilized signal
before starting the zero gas calibration. Note: You can go back to the menu "Basic Controls" with the F2 key.
Zero gas calibration
is running
Note: You can abort the procedure at any time with the F2 key.
F1 F2 F3 F4 F5
CLD
Analyzer zero
Are you sure? You must have zero gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press ZERO again now. Press the left arrow key when you are done. Calibration status: READY Error message for last zero: CAL OK
HOME ESCAPE ZERO INFO
0.057 ppm
F1 F2 F3 F4 F5
Change to the
main menu
F1
Press
Alternative: If you want to continue with the span gas calibration after the zeroing is done, you can go back to the basic controls with the F2 key or the ←← -key. (The spanning will be described in the next chapter: 4.2.2).
4 - 22
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
Main Menu
0.017 ppm
4.2.1 Calibration - Zeroing
Change to the
single component display
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
F1 F2 F3 F4 F5
CLD
0.017 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
Range: 4
340 hPa
0.5 C
51.5 C 360 C
F1
Press
Single component
display will appear after the zero gas calibration
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Notes for zero gas calibration:
You can calibrate all ranges together or each range separately
with zero gas.
You have to set up the parameter you want in the menu
"Calibration Parameters" (see chapter 5.1.2 S. 5-9).
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 23
4 - 24
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.00 ppm NOx
0 250
Range: 4

4.2.2 Calibration - Spanning

Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the
submenu "Analyzer span"
F4
Press
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 25
CLD
Analyzer span
Are you sure? You must have span gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press SPAN again now. Press the left arrow key when you are done. Calibration status: READY Error message for last span: CAL OK
HOME ESCAPE SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Analyzer span
Are you sure? You must have span gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press SPAN again now. Press the left arrow key when you are done. Calibration status: SPANNING - WAIT Error message for last span: CAL OK
ABORT
95.0 ppm
Start Spanning
++ Pay attention to
the note at p. 4-27 !
F4
Press
Caution:
You have to
flow enough span gas
through the analyzer
look for a stable signal
before starting the span gas calibration. Note: You can go back to the menu "Basic Controls" with the F2 key.
Span gas calibration
is running
Note: You can abort the procedure at any time with the F2 key.
F1 F2 F3 F4 F5
CLD
Analyzer span
Are you sure? You must have span gas flowing through the analyzer. This control does NOT control any auto-calibration
module bound to this analyzer! If you are sure, press SPAN again now. Press the left arrow key when you are done. Calibration status: READY Error message for last span: CAL OK
HOME ESCAPE SPAN INFO
225.0 ppm
F1 F2 F3 F4 F5
Change to the
main menu
F1
Press
Alternative: If you want to continue with further setups of the basic controls, you can go back to the corresponding menu with the F2 key or the ←← -key.
4 - 26
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
F1 F2 F3 F4 F5
225.0 ppm
4.2.2 Calibration - Spanning
Change to the single
component display
F1
Press
Single component
CLD
225.00 ppm NOx
display will appear after the span gas calibration
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 4
340 hPa
0.5 C
51.5 C 360 C
Notes for the span gas calibration:
You can calibrate all ranges together or each range separately with span gas. You have to set up
the parameter you want in the menu "Calibration Parameters" (see chapter 5.1.2 p. 5-9).
If it will not be possible to calibrate all ranges with the same span gas, you have to calibrate each
range separately!
The desired value of span gas has to be (!) a value between 10% and 110% of each end of range.
Otherwise, no calibration will be running.
Example:
End of range 1: 50 ppm → Potential span gas setpoints: 5 – 55 ppm
End of range 2: 250 ppm → Potential span gas setpoints: 25 – 275 ppm
End of range 3: 1000 ppm → Potential span gas setpoints: 100 – 1100 ppm
End of range 4: 2500 ppm → Potential span gas setpoints: 250 – 2750 ppm
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 27
4 - 28
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.00 ppm NOx
0 250
Range: 4

4.2.3 Calibration - If It Won´t Calibrate

Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the line
"If it won´t calibrate..."
Press or as often as necessary until the line "If it won´t calibrate..." will appear white on black.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 29
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY
If it won´t calibrate...
Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Check that you are flowing the correct gas, and the gas concentration is what it is supposed to be. Make sure that the reading is stable before starting. If you have changed the range full scale value, or any linearizer coefficients, or enabled or disabled it, or done anything else that would affect how it measures the gas, you may have made it hard for the algorithm to get to a calibration. In this case, manually adjust the coefficients until the readings are close to correct, and try again.
HOME ESCAPE INFO
If it won´t calibrate...
95.0 ppm
Change to the
menu page "If it won´t calibrate..."
Press
Notes:
Follow the instructions
on the screen to calibrate succesfully!
If you have further
questions, contact your customer service, please!
Change to the
main menu
F1
Press
Alternative: If you want to continue with further setups of the basic controls, you can go back to the corresponding menu with the F2 key or the ←← -key.
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the single
component display
F1
Press
4 - 30
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.00 ppm NOx
4.2.3 Calibration - If It Won´t Calibrate
Single component
display will appear
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
Range: 4
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 31
4 - 32
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5 Expert Controls and Set up
If you press the ENTER key or the → -key in the line "Expert controls and set up..." of the "Main Menu" you will change to the following display:
CLD

Expert controls and set up

Expert analyzer controls...
Auxiliary module controls...
System set up... Analyzer module set up... Auxiliary module set up...
System calibration...
MAIN
F1 F2 F3 F4 F5
<<<
NEXT INFO
95.0 ppm
From the menu "Expert controls and set up" you can change to several submenus to set up parameters for the measurement and calibration of your analyzer and analyzer module. Besides you can set up the configuration for auxiliary modules. Which part of these menus are important for you depends on the configuration of your NGA 2000 system. In the following table you will find a short overview about the contents of the menus and where you will find the description in this manual:
Menu Important Contents Chapter in this manual
Expert analyzer controls...
Auxiliary module controls...
Zero/Span calibrationRange settingsOzonator powerConfiguration of the SIO and of
the DIO's installed in a platform
5.5 p. 53 to 60
5.2 p. 29 to 44
or in an MLT analyzer
System set up...
Front panel controlSet up of the
5.4 p. 47 to 52
single component display
Display resolution
Analyzer module set up...
Measurement/Calibration
5.1 p. 3 to 28
parameters set up
Linearizer parameters set upSensor/Converter/Block
parameters set up and optimization
Auxiliary module set up...
Set up for micro processor
5.3 p. 45
controlled auxiliary I/O modules
System calibration...
Zero gas and span gas
See Supplement!
calibration of all channel automatically
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 1
Structure of chapter five:
At the begin of any chapter you will find the way to a certain submenu of the MLT software starting from the line "Expert controls and set up" in the "Main Menu". The way will be described with the software catchwords that you have to enter one after another to reach the corresponding submenu. At the end of the catchword listing you will find the illustration of the LCD screen. After that you can read the set up instructions and explanations. Sometimes the illustration will be completed by sketches or further menu pictures.
Example: You want to set up the zero/span gas concentrations for all ranges of the
analyzer module.
Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Calibration gas list
↓↓
CLD
Calibration Gas List
Zero gas - range 1: 0.00 ppm
NO Span gas - range 1: 2.00 ppm NOx Span gas - range 1: 2.00 ppm
Zero gas - range 2: 0.00 ppm NO Span gas - range 2: 25.00 ppm NOx Span gas - range 2: 25.00 ppm
Calibration...
HOME ESCAPE MORE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Calibration Gas List" you can set up the desired zero and span gas values for all ranges. On the first menu page you will find the parameters of range 1 and 2, on the second page you will find the parameters of range 3 and 4. You can reach the second menu page with the F3 key (MORE)...
Further explanations and instructions to set up the parameters will follow!
5 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.1 Analyzer Module Set up
Main Menu — Expert controls and set up
↓↓

Analyzer module set up

↓↓
CLD
Analyzer module set up
Calibration gas list...
Calibration parameters... Gas measurement parameters... Analyzer parameter list... Physical measurement parameters...
Displayed parameters... Analyzer tag: CLD
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
From the menu "Analyzer module set up" you can change to several submenus, where you can set up the measurement and calibration parameters of the CLD analyzer module. Besides, you can go to those submenus where you can set up the parameters of the single component display.
Caution:
Normally, you will find "CLD" as analyzer module tag in the line "analyzer tag".
If the CLD module is part of an analyzer network: You have to change the configu­ration of the programmable digital inputs and analog outputs, if you change the identification tag ! Because, the parameters of the module source will change ! So, the relations will disappear !
Set up parameters:
Select the menu line you want or the line of variables "Analyzer tag" with
the ↓↓ -key or the ↑↑ -key.
Select the variable or change to the submenu with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key; Maximum of digits: 30 Options: alphabet (capital and small letters), umlauts, letters from different languages, 0, 1, 2, ..., 9, +, -, *, /, =, ?, !, %, $ and many others.
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 3
5.1.1 Analyzer Module Set up

Calibration/Calibration Gas List

Main Menu — Expert controls and set up
Analyzer module set up
Calibration gas list
↓↓ ↓↓ ↓↓
CLD
Calibration Gas List
Zero gas - range 1: 0.00 ppm
NO Span gas - range 1: 2.00 ppm NOx Span gas - range 1: 2.00 ppm
Zero gas - range 2: 0.00 ppm NO Span gas - range 2: 25.00 ppm NOx Span gas - range 2: 25.00 ppm
Calibration...
HOME ESCAPE MORE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Calibration Gas List" you can set up the desired zero and span gas values for all ranges. On the first menu page you will find the parameters of range 1 and 2, on the second page you will find the parameters of range 3 and 4. You can reach the second menu page with the F3 key (MORE). Via the menu line "Calibration..." you can change to the menu "Zero/span calibration", where you can start zeroing and spanning (see next page).
Set up parameters:
Select the line of variables you want or the menu line "Calibration..." with
the ↑↑ -key or the ↓↓ -key.
Select the parameter or change to the menu "Zero/span calibration" with
the ENTER key or the -key.
Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key. Options zero gas: 0.00 ppm up to the end of range value Options span gas: 0.00 ppm up to 120% of the end of range value
You can set up the end of range values in the menu "Range Settings" (see 5.1.6 p. 5-16).
Confirm the new value with the ENTER key or cancel and go back to the last value
with the F2 key.
5 - 4
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Span gas calibration notes:
The span gas setpoint of NO and NO
have to be the same value, because both
x
components will be calibrated simultaneously!
You can calibrate all ranges together or each range separately with span gas. You have
to set up the parameter you want in the menu "Calibration Parameters" (see chapter 5.1.2 p. 5-9).
If it will not be possible to calibrate all ranges with the same span gas, you have to
calibrate each range separately!
The desired value of span gas has to be (!) a value between 10% and 110% of each end
of range. Otherwise, no calibration will be running:
Example:
End of range 1: 50 ppm Potential span gas setpoints: 5 – 55 ppm End of range 2: 250 ppm Potential span gas setpoints: 25 – 275 ppm End of range 3: 1000 ppm Potential span gas setpoints: 100 – 1100 ppm End of range 4: 2500 ppm Potential span gas setpoints: 250 – 2750 ppm
How to calibrate:
Change to the menu "Zero/span calibration" via the menu line "Calibration...":
CLD
Zero/span calibration
Measurement range number: 3
Zero gas concentration: 0.0 ppm Span gas concentration: 100.0 ppm Sample flow: 1100 ml/min Raw measurement signal: 521590 Measurement gas: NOx NO/NOx toggle! Status: READY Result... Calibration adjustment limits: Disabled
HOME FACTORS ZERO SPAN INFO
F1 F2 F3 F4 F5
95.0 ppm
1) Select the measurement range:
Press the ENTER key or the -key in the line "Measurement range number" to enter
the parameter. Select the range you want to calibrate with the ↑↑ -key or the ↓↓ -key and confirm it with ENTER. The zero gas concentration and the span gas concentration of the range selected will appear in the corresponding lines, automatically.
2) Select the gas component:
Go to the line "NO/NOx toggle!" with the ↓↓ -key and select the gas component you
want with the ENTER key or the → -key. The component selected will appear in the line "Measurement gas", automatically.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 5
3) Zeroing:
Change to the submenu "Analyzer zero" with the F3 key (ZERO).
Pay attention to the notes on the screen !
Flow enough zero gas through the CLD.
Wait for a stable signal (look at the headline of the display). Then, start the zero gas
calibration with the F3 key (ZERO). The running procedure will be the same as it is described detailed in chapter 4.2.1 of the basic controls (p. 4-22pp).
Go back to the menu "Zero/span calibration" with the F2 key or the -key after the
zero gas calibration will be done (Messages on the display: READY/CAL OK).
4) Spanning:
Change to the submenu "Analyzer span" with the F4 key (SPAN).
Pay attention to the notes on the screen !
Flow enough span gas through the CLD.
Wait for a stable signal (look at the headline of the display). Then, start the span gas
calibration with the F4 key (SPAN). The running procedure will be the same as it is described detailed in chapter 4.2.2 of the basic controls (p. 4-26pp).
Go back to the menu "Zero/span calibration" with the F2 key or the -key after the
span gas calibration will be done (Messages on the display: READY/CAL OK). Alternative: Go back to the "Main Menu" with the F1 key (HOME).
5) Check the calibration results:
You can change to the submenu "Zero/span diagnostic data" via the line "Result..." of
the menu "Zero/span calibration". There you can control the results of the last zero and span calibration. You can go back to the menu "Zero/span calibration" with the F2 key or the ← -key.
Manual input of calibration factors:
The calibration may be wrong because of a trouble in the gas flow. If a calibration was not successful, you have to repeat it. It is possible that a new start of calibration will not solve the problem, if the measured zero or span gas value will differ from normal signals largely. In that case you have to put in the calibration factors manually:
From the menu "Zero/span calibration" you can change to the submenu "Calibration
Factors" with the F2 key (FACTORS):
CLD
Calibration Factors
Only those factors appropriate for the current range will affect the reading on the current range. Make sure you are using the right ones!
Measurement range number: 3
Range 1 factors... Range 2 factors... Range 3 factors... Range 4 factors...
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
5 - 6
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Select the range you want in the line "Measurement range number" and change via the
line "Range 1/2/3/4 factors..." to the corresponding submenu:
CLD
Range 3 Factors
Zero offset: 521580.6
Span factor: 0.002608716 Full scale range at calibration: 100.0 ppm Measurement range number: 3
Raw measurement signal: 521660
HOME STORE NEXT HISTORY INFO
F1 F2 F3 F4 F5
95.0 ppm
Set up parameters:
Press the ENTER key or the -key in the line "Zero offset" to select the parameter and
set up "32700" resp. "524000" with the ↑↑ -key or the ↓↓ -key. You may select any digit with the → -key or the ←← -key. Note: The zero offset is about 32000 counts for the CLD software revision 2.2.1 (16 bit D/A transformer) and about 520000 counts for the revision 2.3 (20 bit D/A transformer).
Confirm your setup with the ENTER key.Go to the line "Span factor" with the ↓↓ -key.Press the ENTER key or the → -key to select the parameter and set up "0.00015" with
the arrow keys.
Confirm your setup with the ENTER key.
Notes:
You have to set up the factors for all ranges.With the parameter "Zero offset" you will fix the zero value. The changing of the zero
offset will influence the signal immediately (look at the headline).
With the parameter "Span factor" you will fix the signal at the span gas setpoint. The
changing of the span factor will influence the signal immediately (look at the headline).
Conditions for the manual input of factors:
"Disabled" has to be set up in the line "Calibration adjustment limits" of the menu
"Zero/span calibration".
"SEPARATELY" has to be set up in the lines "Zero ranges" and "Span ranges" of the
menu "Calibration Parameters" (see 5.1.2, p. 5-9).
Attention:
After the calibration with factors you have to calibrate with zero gas and span gas in the usual way!
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 7
Saving the factors:
You can store the actual zero offset and span factor with the F2 key (STORE) of the
menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu page (see next illustration).
You can change to the second menu page of the menu "Range 1/2/3/4 Factors" with
the F4 key (HISTORY):
CLD
Range 3 Factors
Zero offset: 0.000000 Span factor: 0.000000
Zero offset: 0.000000 Span factor: 0.000000
HOME NEXT RSTR MN RSTR ST INFO
F1 F2 F3 F4 F5
Manufacturer´s settings.
Stored settings
95.0 ppm
You can change the stored settings against the manufacturer's settings with the F3 key
(RSTR MN: Restore the manufacturer's values). The "Zero offset" and the "Span factor" of the first menu page will change according to the setups of the manufacturer's settings.
You can change the manufacturer's settings against the stored settings with the F4 key
(RSTR ST: Restore the "stored" values). The "Zero offset" and the "Span factor" of the first menu page will change according to the setups of the stored settings.
5 - 8
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.1.2 Analyzer Module Set up
Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓

Calibration parameters

↓↓
Calibration Parameters
CLD
Calibration Parameters
Calibration adjustment limits: Disabled
Calibration averaging time: 5 s Calibration failure alarm: No Cal. failure error allowed: 50 %
Calibration time out: 60 s Zero ranges: TOGETHER Span ranges: SEPARATELY
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Calibration Parameters" you can set up several parameters of the zero and span calibration.
Set up parameters:
Select the line of variables you want with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Line of variables "Calibration adjustment limits:
Enabled: The calibration result will only be accepted within the limit settings. If the
results will be out of these limits, you will find a corresponding error message in the menu "Analyzer zero" respectively "Analyzer span". Illustrations of these menus you will find in chapter 4.2.1 p. 4-22 resp. 4.2.2 p. 4-26.
Disabled: The calibration will be possible, even if the calibration results will leave the
allowed limits. Compare with the explanations in chapter 5.1.1 p. 5-6/7.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 9
Line of variables "Calibration averaging time": The value in this line will determine the time range the analyzer will use to get the
calibration average. The precision will be better with longer times. Options: 0 to 60 s; Standard: 5 s
Line of variables "Calibration failure alarm": The accuracy of the signals may go down after a long use of the analyzer. If you will set
up "Yes" in the line "Calibration failure alarm", a warning message will display a wrong or impossible calibration.
Line of variables "Cal. failure error allowed": If the setpoint will differ more than allowed by the value in this line, a warning message
may be displayed. To enable this warning, "Yes" has to be set up in the line "Calibration failure alarm". Options: 5 to 80%; Standard: 50%
Line of variables "Calibration time out": If the signal will not be stable after the time out set up in this line, a warning message may
be displayed. To enable this warning, "Yes" has to be set up in the line "Calibration failure alarm". Options: 0 to 300s; Standard: 60s
Lines of variables "Zero ranges/Span ranges":
TOGETHER: All four ranges will be calibrated automatically one after another by zero
and span gas.
SEPARATELY: Only the range selected will be calibrated by zero and span gas. You
have to set up this parameter, if you want to calibrate with manual factors (see chapter.
5.1.1 p. 5-6/7).
Note: Only, if it is possible to get a span gas setpoint between 10% and 110% of each
end of range, you may calibrate all ranges together. Otherwise, only the ranges with this condition will be calibrated automatically. The other ranges will be ignored!
5 - 10
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.1.3 Analyzer Module Set up

Gas Measurement Parameters

Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Gas measurement parameters
↓↓
CLD
Gas Measurement Parameters
Linearization parameters...
Response time / delay parameters... Range setting...
Units...
Linearization functions...
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
From the menu "Gas Measurement Parameters" you can change to several submenus, where you can set up parameters of ranges, of measurement units and of linearization:
Select the line you want with the ↓↓ -key or the ↑↑ -key.Press the ENTER key or the → -key to change to the corresponding menu.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 11
5.1.4 Analyzer Module Set up

Linearization Parameters

Main Menu — Expert controls and set up
Analyzer module set up
Gas measurement parameters
Linearization parameters
↓↓ ↓↓ ↓↓ ↓↓
CLD
Linearization parameters
Range 1 linearizer: Disabled
If enabled, uses curve no.: 1 Range 2 linearizer: Disabled If enabled, uses curve no.: 2 Range 3 linearizer: Disabled If enabled, uses curve no.: 3 Range 4 linearizer: Disabled If enabled, uses curve no.: 4
Set coefficients...
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Linearization parameters" you can enable the linearization for each range and you can select the linearization curve.
Note:
Normally, the CLD is working linearly in the whole range. Therefore you do not have to linearize generally ! If you will find deviations from the linearization, you should control
the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for deviations will be a defect in the ozone supply.
Set up parameters:
Select any line of variables or the menu line "Set coefficients..." with the ↓↓ -key
or the ↑↑ -key.
Select the variable or change to the submenu "Linearity coefficients" with
the ENTER key or the -key.
Select the parameter you want with the ↑↑ -key or the ↓↓ -key.Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
5 - 12
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Lines of variables "Range 1/2/3/4 linearizer": In these lines you can enable the linearization for the range selected.
You have to set up "Disabled" in these lines as long as you will determine the data for linearization. This will be necessary to avoid interferences with existing values during the calculation of the linearization curve.
Lines of variables "If enabled, uses curve no.": You can select for each range one of the four linearization curves from the menu
"Linearity coefficients" (see next illustration). Menu "Linearity coefficients": You can change to the menu "Linearity coefficients" via the line "Set coefficients...":
CLD
Linearity coefficients
A0 coefficient: 0.000000
A1 coefficient: 1.000000 A2 coefficient: 0.000000 A3 coefficient: 0.000000 A4 coefficient: 0.000000 Curve upper limit: 100.0 ppm Curve over-range: 5.0 % Curve under-range: 5.0 %
Status: Disabled
HOME ESCAPE NEXT LAST INFO
F1 F2 F3 F4 F5
Curve 3
95.0 ppm
In the menu "Linearity coefficients" you can put in the desired coefficients for the linearization curve selected. With the F3 key (NEXT) you can scroll between the menu pages of all four linearization curves. You will find the number of the curve selected in the head line of each "Linearity coefficients" menu.
Line of variables "Curve upper limit": In this line you should set up the end of range value for the range you want to determine
the linearization curve. Lines of variables "Curve over-range/Curve under-range": In these two lines you can determine how many per cent the linearization curve for the
range selected shall be over resp. under the range limits. Line of variables "Status": In this line you can enable or disable the linearization curve selected. This is the same
setup as in the lines "Range 1/2/3/4 linearizer" of the menu "Linearization parameters".
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 13
5.1.5 Analyzer Module Set up

Response Time/Delay Parameters

Main Menu — Expert controls and set up
Analyzer module set up
Gas measurement parameters
Response time/delay parameters
↓↓ ↓↓ ↓↓ ↓↓
CLD
Response time/delay parameters
Range 1 t90 time: 3.0 s
Range 2 t90 time: 3.0 s Range 3 t90 time: 3.0 s Range 4 t90 time: 3.0 s
LON update rate: 10 per sec Output delay time: 0.0 s
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Response time/delay parameters" you can set up the response time (t time) of the analyzer for each range and the output delay of the signal.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.
90
Select any digit with the -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Lines of variables "Range 1/2/3/4 t90 time": The t90 time is defined as the time that has to run until the analyzer displays 90% of the
current concentration after a jump in the concentration level. Options: 0 to 30 seconds for every range of a channel.
5 - 14
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Line of variables "LON update rate": In this line you can set up the update rate of the network variable.
Options: 1 per sec, 10 per sec, "ASAP" (As fast as possible)
Line of variable "Output delay time":
Example:
You can shift the absolute time of the analog output signal for one channel. If you use the right output delay time you can watch the signals of several channels at the same starting time. Options: 0 to 30 seconds.
Conc.
CO
CO
NO
2
Time
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 15
5.1.6 Analyzer Module Set up

Range Settings

Main Menu — Expert controls and set up
Analyzer module set up
Gas measurement parameters
Range setting
↓↓ ↓↓ ↓↓ ↓↓
CLD
Range Settings
Minimum range: 10.0 ppm Maximum range: 10000 ppm
Range 1 lower limit: 0.0 ppm
Range 1 upper limit: 10.0 ppm Range 2 lower limit: 0.0 ppm Range 2 upper limit: 25.0 ppm Range 3 lower limit: 0.0 ppm Range 3 upper limit: 100.0 ppm Range 4 lower limit: 0.0 ppm Range 4 upper limit: 250.0 ppm
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Range Settings" you can set up the begin and the end of each range. These values will only be limited by the setups in the line "Minimum range" resp. "Maximum range". The concentrations of these lines are manufacturer's settings and cannot be changed by the customer.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
5 - 16
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.1.7 Analyzer Module Set up
Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Gas measurement parameters
↓↓

Units

↓↓
Units
CLD
Units
Gas measurement units: ppm
Pressure measurement units: hPa Temperature measurement units: C ppm to mg/Nm3 conversion factor: 1.000
Variables are still sent as the basic SI unit.
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Units" you can set up the units of the analyzer for pressure, temperature and gas concentration.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the 88 -key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key. Options: "Gas measurement units" ppb, ppm, mg/Nm3, %
"Pressure measurement units" hPa, psig "Temperature measurement units" C, F "Conversion factor ppm – mg/Nm3" 0 to 100
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 17
5.1.8 Analyzer Module Set up

Linearization Functions

Main Menu — Expert controls and set up
Analyzer module set up
Gas measurement parameters
Linearization functions
↓↓ ↓↓ ↓↓ ↓↓
CLD
Linearization functions
Polynomial set up...
Midpoint correction set up...
Use the polynomial set up to generate a linearizing polynomial from up to 20 gases. With more than 6 gases it will produce a fourth order polynomial linearizer. Use the midpoint correction for a piecewise-linear final correction, to bring up to three points precisely onto the curve.
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
From the menu "Linearization functions" you can change to several submenus, where you can linearize the ranges and optimize the linearization functions:
Change to the line you want with the ↓↓ -key or the ↑↑ -key.
Note:
Normally, the CLD is working linearly in the whole range. Therefore you do not have to linearize generally ! If you will find deviations from the linearization, you should control
the ozone and sample gas supply. Optimize it, if necessary! Usually, the reason for deviations will be a defect in the ozone supply.
Polynomial set up:
Press the ENTER key or the → -key in the line "Polynomial set up..." to change to the corresponding menu:
5 - 18
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
Polynomial set up
Range to be linearized: 3
Current span gas: 100.0 ppm Calculated polynomial order: 4
Gas values shown as: ppm Gas concentrations...
Analyzer function: READY
HOME ESCAPE CALC INFO
F1 F2 F3 F4 F5
95.0 ppm
1) Select the range to be linearized:
Press the ENTER key or the -key in the line "Range to be linearized" to select the
parameter.
Select the range you want with the ↑↑ -key or the ↓↓ -key and confirm it with ENTER.
2) Zeroing and Spanning of the range selected:
e.g.: Change to the "Main Menu" with the F1 key. Change after that via the line "Basic
controls..." to the corresponding menu. Press the F3 key(ZERO) resp. the F4 key (SPAN) to start the calibration (Look at the descriptions in the chapters 4.2.1 p. 4-21 and 4.2.2 p. 4-26).
Alternative: Change to the menu "Zero/span calibration" of the "Expert controls" to
start the calibration (see 5.1.1 p. 5-5).
3) Determine the display of gas concentration:
Press the ENTER key or the -key in the line "Gas values shown as" to select the
parameter.
Select the concentration setup you want with the ↑↑ -key or the ↓↓ -key and confirm it
with the ENTER key:
"ppm" for absolute concentrations
"Percent of span gas" for relative concentrations, i.e. the gas value will be displayed
as percentage of the span gas. This will be necessary, if you will use diluted span gas.
4) How to get the x,y-values to determine the polynomial curve:
Press the ENTER key or the -key in the line "Gas concentrations...“ to change to
the corresponding submenu:
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 19
CLD
Gas concentrations
Gas value: -1.00
Raw reading: 0.00 ppm Linearized value: 0.00 ppm
Gas value: -1.00 Raw reading: 0.00 ppm Linearized value: 0.00 ppm
Point to be measured: Point 1 Analyzer function: READY
HOME ESCAPE DATA NEXT INFO
F1 F2 F3 F4 F5
Point 1
Point 2
95.0 ppm
Input of the desired values:
You will need minimum 6 values to calculate a polynomial curve. In the menu "Gas
concentration" are maximum 20 values available.
Press the ENTER key or the -key in the line "Gas value" of "Point 1" to select the
parameter.
Set up the concentration you want with the ↑↑ -key or the ↓↓ -key. If necessary, change
the sign with the F4 key ("+/-"). Confirm your input with the ENTER key. It is also possible to select the digits with the ← -key or the →→ -key and to change the number with the ↑↑ -key or the ↓↓ -key.
Measuring of the raw values:
Flow the analyzer with a defined gas concentration.
After the signal will be stable press the F3 key (DATA): The analyzer will store the
desired value and the corresponding raw value as "Point 1" to calculate the linearization polynom.
Repeat the procedure described to get all values you need to calculate the polynom:
Go to the line "Gas value" of "Point 2" with the ↓↓ -key and set up the desired value.
Go to the line "Point to be measured" with the ↓↓ -key and select "Point 2".
Measure the corresponding raw value, store the values with the F3 key and so on...
To set up the values of point 3, 4 and all the other points you have to change to further
menu pages with the F4 key (NEXT). You can always go back to the last menu page with the F2 key (ESCAPE).
5) Calculation of the polynomial function:
You can calculate the linearization polynom in the following way, after you will be
ready with the registration of the points:
Go back to the menu "Polynomial set up" with the F2 key (ESCAPE) or the -key.
Then, press the F3 key (CALC): The analyzer will calculate the polynom with the available points and store the coefficients of this function for the range selected.
5 - 20
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
6) Correction of the linearization polynom, if necessary:
You can optimize the calculated linearization with maximum three additional points.
Press the ENTER key or the → -key in the line "Midpoint correction set up..." of the
menu "Linearization functions" to change to the corresponding submenu:
CLD
Midpoint correction set up
Correction: DISABLED
Point being measured: Point 3 Point 1 gas concentration: -1.00 ppm Point 2 gas concentration: -1.00 ppm Point 3 gas concentration: -1.00 ppm Point 1 reading: -1.00 ppm Point 2 reading: -1.00 ppm Point 3 reading: -1.00 ppm Span gas value: 0.00 ppm Analyzer functioning: READY
HOME ESCAPE SET RANGE 4 INFO
F1 F2 F3 F4 F5
Range 3
95.0 ppm
a) Range selecting:
Range 1 will be selected automatically after entering the menu. If you want to select
another range to correct, press the F4 key (RANGE 2/3/4). The range selected will be displayed in the headline of the menu.
b) Disable the correction:
You have to disable the correction mode before calculating the correction data.
Otherwise, you will have interferences during the calculation with existing values.
Press the ENTER key or the -key in the line "Correction" to select the parameter.
Set up "DISABLED" with the ↑↑ -key or the ↓↓ -key and confirm it with the ENTER key.
c) Measured point selecting:
Go to the line "Point being measured" with the ↓↓ -key.
Press the ENTER key or the → -key to select the parameter. Select "Point 1" with the
the ↑↑ -key or the ↓↓ -key and confirm it with the ENTER key.
d) Set up the gas concentration for point 1:
Go to the line "Point 1 gas concentration" with the ↓↓ -key.
Press the ENTER key or the → -key to select the parameter.
Select any digit with the ← -key or -key and set up the value you want with the the
↑↑ -key or the ↓↓ -key. If necessary, change the sign with the F4 key ("+/-").
Confirm your set up with the ENTER key.
e) Measuring of the gas concentrations:
Flow the analyzer with the gas concentration set up of the line "Point 1 gas
concentration" and measure the raw value.
After the signal will be stable press the F3 key (SET): The current signal will be
displayed in the line "Point 1 reading". But, the analyzer will equate the raw value with the desired value of this point automatically.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 21
f) Set up further correction points:
You can repeat the steps c) to e) to determine maximum two further correction points.
Select "Point 2" resp. "Point 3" in the line "Point being measured". Set up the value
you want in the line "Point 2 gas concentration" resp. "Point 3 gas concentration". Measure the gas concentrations and set up the correction with the F3 key.
g) Enable the correction:
Select "ENABLED" in the line "Correction": The midpoint correction will be done auto-
matically for the range selected with the determined values.
The midpoint correction can be done for each range of the CLD analyzer module.
5 - 22
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.1.9 Analyzer Module Set up

Analyzer Parameter List

Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Analyzer Parameter List
↓↓
CLD
Analyzer Parameter List
Analyzer tag: CLD
Ozonator status: OFF - PRESSURE Ozonator power: Enabled First line´s parameter: Sample press.: Second line´s parameter: Detector temperature: Third line´s parameter: Block temperature: Fourth line´s parameter: Converter temp.:
Linearizer parameters...
HOME ESCAPE NEXT LAST INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Analyzer Parameter List" you can set up parameters of the single component display, of the ozonator and the analyzer tag. You can also change to several submenus, where you can set up linearizer parameters, concentration and calibration parameters.
Set up parameters:
Select any line of variables or the menu line "Linearizer parameters..." with
the ↓↓ -key or the ↑↑ -key.
Select the variable or change to the submenu "Linearity coefficients" with
the ENTER key or the -key.
Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Line of variables "Analyzer tag": You can input a tag of maximum thirty digits. You will find the same line in the menu
"Analyzer module set up" (see chapter 5.1 p. 5-3).
Caution:
Normally, you will find "CLD" as analyzer module tag in the line "analyzer tag".
If the CLD module is part of an analyzer network: You have to change the configu­ration of the programmable digital inputs and analog outputs, if you change the identification tag ! Because, the parameters of the module source will change ! So, the relations will disappear !
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 23
Line of variables "Ozonator power": Options: Enabled/Disabled
The setup of this line will change the displayed message of the line "Ozonator status". That line is not available. The ozonator power setup is also possible in the menu "Basic Controls" (see chapter 4.1.5 p. 4-15).
Lines of variables "First/Second/Third/Fourth line's parameter: In these lines you can set up which of the following parameters shall be displayed in each
of the last four lines of the single component display:
Sample pressure (bargraph with measured value)Sample flow (bargraph with measured value)Raw signal (bargraph with measured value)Validity (report: valid/invalid)Interference (report: On/Off)Health (report: Normal/...)Operational state (report: STANDBY/...)Measurement mode (report: 1/...)Linearizer (report: Enabled/Disabled)Calibration status (report: READY/...)Output delay time (bargraph with value selected)t
time (bargraph with value selected)
90
Noise level (of concentration)NO/NODetector temperature (bargraph with measured value)Block temperature (bargraph with measured value)Converter temperature (bargraph with measured value)Ozonator (report: OFF - ENABLED/OFF - DISABLED)
(Display: NO/NOx)
x
Menu line "Linearizer parameters...": Via this line you can change to the submenu "Linearity coefficients", where you can set up
the coefficients of the linearization curves manual. You can also reach the menu "Linearity coefficients" via the menu "Linearization parameters". These menus are described detailed in chapter 5.1.4 p. 5-12/13.
Ramifications via function keys: With the F3 key (NEXT) of the menu "Analyzer Parameter List" you can change to further
menu pages, where you can set up the following concentration and calibration parameters:
Control modeOutput delay timeBegin and end of all four rangest
time (response time) for all ranges
90
Linearizer enabling/disabling for each rangeCalibration averaging time
5 - 24
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Calibration failure alarmCalibration failure error allowedCalibration time outZero and span gas calibration of the ranges together or separatelyEnabling/Disabling of the calibration adjustment limitsSet up of the zero gas concentration of each rangeSet up of the NO and NO
span gas concentration of each range
x
You can also set up all these parameters in the menus "Calibration Gas List", "Calibration Parameters", "Response time/delay parameters" and "Range Settings". You can find the descriptions of these menus in the chapters 5.1.1 p. 5-4, 5.1.2 p. 5-9/10 and 5.1.5/6 p. 5-14 to 16.
With the F4 key (LAST) of the menu "Analyzer Parameter List" you can change to the menu page "Calibration Gases", where you can set up the NO and NOx span gas concentrations of each range. This menu page is the same page you will reach by pressing the F3 key (NEXT) several times.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 25
5.1.10 Analyzer Module Set up

Physical Measurement Parameters

Main Menu — Expert controls and set up
Analyzer module set up
Physical measurement parameters
↓↓ ↓↓ ↓↓
CLD
Sample capillary pressure: 340 hPa Ozone supply pressure: 1030 hPa Ozonator status: ON Ozonator power: Enabled Bypass flow: 1100 ml/min Converter temperature: 360 C Sensor temperature: 0.5 C Block temperature: 51.5 C Capillary flow rate: 200 ml/min
Pressure limits...
Temperature limits...
HOME ESCAPE HISTORY INFO
F1 F2 F3 F4 F5
Physical Measurements
95.0 ppm
In the menu "Physical Measurements" you can find several physical measurement parameters of the CLD analyzer module like ozonator power, converter temperature or capillary flow rate. Via the menu lines "Pressure limits..." and "Temperature limits..." you can change to the corresponding submenus, where you can set up several pressure parameters and temperature parameters of the analyzer. The violation of the limits in the submenus may start alarm reports, if the alarm report is enabled in the diagnostic menus of the analyzer module (see 6.3.2 p. 6-27). Besides, the limits of the submenus will be begin and end of the bargraphs in the single component display, if activated (see 5.1.9 p. 5-24).
Changing to the submenus:
Select the line you want with the ↑↑ -key or the ↓↓ -key.Press the ENTER key or the → -key to change to the corresponding submenu.
Note: You should control the sample capillary pressure, the ozone supply pressure and the
bypass flow in this menu after starting up your CLD. The values displayed in these lines should be equivalent to the original manufacturer's data sheet to avoid measurement errors.
5 - 26
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Menu "Pressure Limits":
CLD
Sample capillary upper limit: 490.0 hPa
Sample capillary lower limit: 50.0 hPa Ozone supply upper limit: 1050 hPa Ozone supply lower limit: 700.0 hPa
Barometric pressure: 1013 hPa
HOME ESCAPE INFO
F1 F2 F3 F4 F5
Menu "Temperature Limits":
CLD
Converter upper limit: 500.0 C
Converter lower limit: 150.0 C
95.0 ppm
Pressure Limits
95.0 ppm
Temperature limits
Sensor upper limit: 50.0 C Sensor lower limit: 0.0 C Block upper limit: 55.0 C Block lower limit: 45.0 C
HOME ESCAPE INFO
F1 F2 F3 F4 F5
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
Options:
Pressures: 0 to 3000 hPaTemperatures: 0 to 500 °C for converter; 0 to 75 °C for sensor and block
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 27
5.1.11 Analyzer Module Set up

Displayed Parameters

Main Menu — Expert controls and set up
Analyzer module set up
Displayed parameters
↓↓ ↓↓ ↓↓
CLD
Displayed parameters
First line´s parameter: Sample press.:
Second line´s parameter: Detektor temperature: Third line´s parameter: Block temperature: Fourth line´s parameter: Converter temp.:
May be displayed on the appropriate line of the single analyzer display screen.
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Displayed Parameters" you can determine which parameters shall be displayed in the last four lines of the single component display. These are the same parameters you can set up in the menu "Analyzer Parameter List" (see 5.1.9 p. 5-23).
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.Select the parameter you want with the ↑↑ -key or the ↓↓ -key.Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
5 - 28
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.2 Auxiliary Module Controls
Main Menu — Expert controls and set up
↓↓

Auxiliary module controls

↓↓
CLD
Auxiliary module controls
Set-up SIO module...
Set-up DIO module(s)...
MAIN BACK
F1 F2 F3 F4 F5
>>> <<<
95.0 ppm
INFO
From the menu "Auxiliary module controls" you can change to some submenus where you can set up the parameters for the SIO module and the DIO modules of the MLT analyzer or platform:
Press the ENTER key or the -key in the line "Set-up SIO module..." or in the line
"Set-up DIO module(s)..." to change to the corresponding submenus.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 29
5.2.1 Auxiliary Module Controls

Local SIO Configuration Sections

Main Menu — Expert controls and set up
↓↓
Auxiliary module controls
↓↓
Set-up SIO module
↓↓
CLD
- LOCAL SIO CONFIGURATION SECTIONS -
Analog output setup...
Serial interface setup... Configuration of Relay Outputs... Module installed: Yes
EXPERT INFO
F1 F2 F3 F4 F5
95.0 ppm
From the menu "LOCAL SIO CONFIGURATION SECTIONS" you can change to submenus to control and set up several output configurations of the auxiliary SIO module. General Configuration of a SIO board (the whole specification is in another manual):
Analog outputs: minimum 2, maximum 8Serial interface (RS 232 or RS 485) to connect the analyzer with a external computerThree relay outputs
If there is a SIO module installed in the platform or in the MLT analyzer, you have to put in "Yes" in the line "Module installed", otherwise "No".
Set up parameters:
Select the line of variables or any menu line with the ↓↓ -key or the ↑↑ -key.Select the variable or change to the submenu with the ENTER key or the → -key.Select the parameter with the ↑↑ -key or the ↓↓ -key.Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
5 - 30
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Analog Output Setup:
Press the ENTER key or the → -key in the line "Analog output setup..." to change to the corresponding submenu:
CLD
- ANALOG OUTPUT SETUP -
Output number: 1
Choose signal source module... Choose signal... Signal value for 0% output: 0.00 ppm Signal value for 100% output: 2500.00 ppm Output current range: 0...20 mA Hold output during calibration? No Signal name: NO/NOx Current signal value: 95.00 ppm Source module: CLD
EXPERT ADJUST INFO
F1 F2 F3 F4 F5
95.0 ppm
Configuration of the analog output:
1) Select the analog output number: You can set up all parameters in the menu "ANALOG OUTPUT SETUP" separately for each of the available analog outputs. The number of outputs depends on the hardware configuration of the SIO module (minimum: 2, maximum: 8).
Press the ENTER key or the -key to select the output number and adjust the value
you want with the ↑↑ -key or the ↓↓ -key.
Confirm the number selected with the ENTER key.
2) Choose the signal source module:
Change with the ↓↓ -key to the line "Choose signal source module..."Change to the submenu "ANALYZER MODULES" with the ENTER key or the → -key.
The tags of all analyzer modules combined with the platform or the MLT analyzer will appear:
CLD
- ANALYZER MODULES -
MENU BACK
F1 F2 F3 F4 F5
>>> <<<
95.0 ppm
CLD
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 31
Go to the line you want with the ↑↑ -key or the ↓↓ -key.Select the tag of the channel with the ENTER key or the → -key:
The display will go back to the menu "ANALOG OUTPUT SETUP" automatically. The tag of the channel selected will appear in the line "Source module".
3) Select any signal:
Change to the line "Choose signal" with the ↓↓ -key.Press the ENTER key or the → -key to change to the submenu "SIGNALS":
CLD
- SIGNALS -
Converter temp.:
Block temperature:
Detector temperature:
MENU BACK
F1 F2 F4 F5
>>> <<<
F3
95.0 ppm
Sample flow:
Sample press.:
Ozonator:
NO/NOx:
Noise level:
INFO
Change to the line you want with the ↑↑ -key or the ↓↓ -key.
You can change with the F3 key (>>>) to further menu pages containing other available signals:
t
time
90
Output delay time
Calibration status
Linearizer
Measurement mode
Operational state
Health
Interference
Validity
Raw signal
Primary Variable
Press the ENTER key or the → -key to select the signal:
The display will go back to the menu "ANALOG OUTPUT SETUP" automatically. The signal selected will appear in the line "Signal name".
4) Determination of the output signal values: You can determine the signal value for the 0% output and the 100% output in the lines "Signal value for 0% output" or "Signal value for 100% output". So you have the possibility to zoom out a certain part of the whole range.
5 - 32
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Example:
Range from 0 to 1000 ppm
0% value shall be 400 ppm, 100% value shall be 700 ppm
Analog output is normally: 0V = 0 ppm 10V = 1000 ppm
After changing the output signal: 0V = 400 ppm 10V = 700 ppm
Change to the line "Signal value for 0% output" or "Signal value for 100% output" with
the ↓↓ -key.
Select the value with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with the ↑↑ -key or
the ↓↓ -key. Confirm it with the ENTER key.
Note: If you change the measurement range the adjustment done in this menu will disappear
and go back to the standard values of the range! To change the output signal values permanently, you have to change the adjustments in the menu "Range Settings"! (see 5.1.6 p. 5-16)
Attention: The signal range at the analog output should not be less than the littlest measurement
range. Otherwise, it is possible that the noise level will be very high at the analog output!
5) Determination of the output current range:
Select to the line "Output current range" with the ↓↓ -key.Press the ENTER key or the → -key to select the range and select the value you want
with the ↑↑ -key or the ↓↓ -key.
Confirm the new adjustment with the ENTER key.
Options: 0...20 mA (and automatically 0...10V) or 4...20 mA (2...10V).
6) Analog output during calibration: In the line "Hold output during calibration?" you can determine the signal drift of the analog output and of the limits during calibration:
Yes: During the whole calibration the analog output signal and the status of the limits
are fixed to the last value before calibration.
No: The analog output signal and the status of the limits are following to the
measurement signal during the whole calibration procedure.
Sample Gas Zeroing Spanning Sample Gas
Hold output during calibration?
Hold output during calibration?
No
Yes
0V
xV
Note:
The adjustments "Yes/No" are valid for all kinds of calibrations of a platform SIO or MLT
analyzer SIO: manual, time controlled, AK controlled and system calibration.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 33
7) Fine adjustment of the analog output: If you press the F4 key in the menu "ANALOG OUTPUT SETUP" you can change to the submenu "OUTPUT FINE ADJUSTMENT", where you can do the fine adjustments of the analog output:
CLD
- OUTPUT FINE ADJUSTMENT -
Output number: 1
Operation mode: Normal Fine adjustment for 0% output: 4097 Fine adjustment for 100% output: 799
EXPERT BACK
F1 F2 F3 F4 F5
95.0 ppm
INFO
In the line "Output number" the output number selected in the menu "ANALOG OUTPUT SETUP" will appear. If you want you can change the number in the submenu "OUTPUT FINE ADJUSTMENT":
Press the ENTER key or the -key to select the output number and adjust the value
you want with the ↑↑ -key or the ↓↓ -key. Options: 1, ..., 8.
Confirm the number selected with the ENTER key.
Set up fine adjustment of the analog output:
Press the ENTER key or the -key in the line "Operation mode" to select the
parameter and adjust one of the following parameters with the ↑↑ -key or the ↓↓ -key:
Normal: The absolute measurement signal will be sent to the analog output.
Adjust 0V: Coordination between the display and the analog output for 0V with fine
adjustment 0%.
Adjust 10V: Coordination between the display and the analog output for 10V with fine
adjustment 100%.
Confirm the new parameter with the ENTER key.Press the ENTER key or the → -key in the line "Fine adjustments for 0% output" or
"Fine adjustment for 100% output" to select the corresponding value.
Adjust the value you want with the ↑↑ -key or the ↓↓ -key and confirm it with the ENTER
key. Options: 3500 - 4800 for 0% and 750 - 900 for 100%.
The last three lines of variables in the menu "Analog output setup" only display some values of the analog output! They are not available.
5 - 34
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Serial interface setup:
Press the ENTER key or the → -key in the line "Serial interface setup..." of the menu "LOCAL SIO CONFIGURATION SECTIONS" to change to the corresponding submenu:
CLD
- SERIAL INTERFACE SETUP -
Baudrate: 19200
Data bits: 8 Stop bits: 1 Parity: None Echo mode: Disabled Handshake: Xon/Xoff
Type of installed serial interface: RS232
EXPERT
F1 F2 F3 F4 F5
95.0 ppm
INFO
In the submenu "SERIAL INTERFACE SETUP" you can determine the parameters of the data transfer between the MLT analyzer or platform and external implements. The setups in this menu are depending on the configuration of the analyzer resp. platform and the corresponding implement. The specification of the serial interface is described in another manual.
Set up parameters:
Select any line of variables with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
Options: Baudrate: 300 1200 2400 4800 9600 19200
Data bits: 7 8 Stop bits 1 2 Parity: None Even Odd Echo mode: Enabled Disabled Handshake: None Xon/Xoff Type of installed serial interface:
RS 232 RS 485_2X RS 485_4X None
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 35
Configuration of Relay Outputs:
Press the ENTER key or the → -key in the line "Configuration of Relay Outputs..." of the menu "LOCAL SIO CONFIGURATION SECTIONS" to change to the corresponding menu:
CLD
- CONFIGURATION of RELAY OUTPUTS -
Outputnumber: 1
Invert signal: Disabled Choose source module... Choose signal...
Signal comes from: NGA_Control_module Signal name: Function control Actual state: Off
EXPERT INFO
F1 F2 F3 F4 F5
95.0 ppm
There are three relays on any SIO board. With a jumper you can determine the relay contact status: NO (normally opened) or NC (normally closed). Informations about the whole specification of the SIO board you can find in another manual. In the menu "CONFIGURATION of RELAY OUTPUTS" you have to determine the con­figuration of the three relay outputs!
Configuration set up:
1) Select the output number:
Press the ENTER key or the -key in the line "Outputnumber" and adjust the output
number one, two or three with the ↑↑ -key or the ↓↓ -key.
Confirm the number selected with the ENTER key.
2) Choose the source module:
Change with the ↓↓ -key to the line "Choose source module...".Change to the submenu "Analyzer Modules" with the ENTER key or the → -key.
(illustration see next page) All analyzer modules joint to the platform or to the MLT analyzer and the control module are available.
Go to the line you want with the ↑↑ -key or the ↓↓ -key. If there are more than eight source
modules available you have to press the F3 key to go to another menu page.
Select the tag of the module with the ENTER key or the -key:
The display will go back to the menu "CONFIGURATION of RELAY OUTPUTS" automatically. The tag of the module selected will appear in the line "Signal comes from".
5 - 36
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.0 ppm
Analyzer Modules
EXPERT BACK
NGA Control module
>>> <<<
CLD
INFO
F1 F2 F3 F4 F5
3) Select any signal:
Change to the line "Choose signal..." with the ↓↓ -key.Press the ENTER key or the → -key to change to the submenu "Signals":
CLD
Signals
MAINTENANCE REQUEST
CAL. IN PROGRESS ZERO IN PROGRESS SPAN IN PROGRESS
HOME ESCAPE MORE BACK INFO
95.0 ppm
NORMAL
FAILURE
ZERO FAILED SPAN FAILED
F1 F2 F3 F4 F5
Go to the line you want with the ↓↓ -key or the ↑↑ -key.
With the F3 key you can go to other menu pages containing further available signals:
RANGE OVERFLOW
RANGE UNDERFLOW
FLOW TO LOW
FLOW TO HIGH
RANGE 1, 2, 3, 4
OZONATOR OFF
NOx MODE
NO MODE
or further signals, if need be:
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 37
All signals of the control module and of the analyzer modules joint to the platform or MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "SIGNALS". The signals shown above will appear by using a CLD module as source module.
Select the signal with the ENTER key or the -key:
The display will go back to the menu "CONFIGURATION of RELAY OUTPUTS" automatically. The signal selected will appear in the line "Signal name".
4) Invert signal: You can decide if the measurement signal should be inverted at the relay output. This can be necessary to combine a failure with an alarm control.
Press the ENTER key or the -key in the line "Invert signal" to select the parameter
and adjust "Enabled" or "Disabled" with the ↑↑ -key or the ↓↓ -key.
Confirm the parameter selected with the ENTER key.
The last three lines of variables in the menu "Configuration of relay outputs" only display some adjustments of the relay output selected! They are not available:
"Signal comes from:" displays the source module selected for the current relay output."Signal name:" displays the signal selected for the current relay output."Actual state:" displays if the relay status of the current relay is "On" or "Off".
5 - 38
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.2.2 Auxiliary Module Controls
Main Menu — Expert controls and set up
↓↓
Auxiliary module controls
↓↓

Set-up DIO module(s)

↓↓
Set-up DIO module(s)
CLD
- DIO-MODULE OUTPUTS -
Inputs...
Outputnumber: 1 Choose module... Choose signal... Invert output: Disabled Module status: Normal Slot ID: 1 Signal name: ???? Signal level: Off Signal comes from: ????
EXPERT
F1 F2 F3 F4 F5
>>>
ACKNOWL
95.0 ppm
INFO
In the menu "DIO-MODULE OUTPUTS" you can set up the output configuration of the DIO modules combined to the control module (platform or MLT analyzer). Besides you can change to the submenu "DIO-MODULE INPUTS" to configure the inputs of the auxiliary DIO modules. Every DIO module is composed of 8 digital inputs and 24 digital outputs. If there exist several DIO modules in the platform you can change to the setup menu of each with the F3 key. The tag number of the DIO board selected will appear in the line "Slot ID". The changing with the F3 key is not available for MLT analyzers because only one DIO can be installed in it. The 24 digital outputs are composed of 3 units with 8 outputs. If there is a short circuit or an overload in one unit it will be switched off and saved against destruction. After the defect is removed the board is available immediately. You only have to press the F4 key. Informations about the whole specification of the DIO board you will find in another manual.
Set up parameters:
Select any line of variables or any menu line with the ↓↓ -key or the ↑↑ -key.Select the variable or change to the submenu with the ENTER key or the → -key.Adjust a new value or select the whole parameter with the ↑↑ -key or the ↓↓ -key.Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 39
Line of variables "Invert output": Select "Enabled" in the line "Invert output" if you want to invert the measurement signal at
the actual digital output, otherwise select "Disabled". The output inversion can be neces­sary for instance, if you want to combine a failure with an alarm control.
The last five lines of variables in the menu "DIO-module outputs" only display some adjustments of the DIO output selected! They are not available:
"Module status:" displays the status of the current DIO board."Slot ID:" displays the tag number of the DIO board selected."Signal name:" displays the signal selected for the current DIO output."Signal level:" displays the status of the signal selected."Signal comes from:" displays the source module selected for the current DIO output.
Configuration of the DIO module inputs:
Press the ENTER key or the → -key in the line "Inputs..." to change to the corresponding submenu:
CLD
- DIO-MODULE INPUTS -
Inputnumber: 1
Choose module... Choose function...
Slot ID: 1 Signal name: ???? Signal level: Off Signal comes from: ????
EXPERT
F1 F2 F3 F4 F5
>>>
95.0 ppm
INFO
In the menu "DIO-MODULE INPUTS" you have to determine the configuration of all of the eight DIO module inputs!
DIO input configuration setup:
1) Select the input number:
Press the ENTER key or the -key in the line "Inputnumber" and adjust an input
number (1, ..., 8) with the ↑↑ -key or the ↓↓ -key.
Confirm the number selected with the ENTER key.
2) Choose the module:
Change with the ↓↓ -key to the line "Choose module...".Change to the submenu "ANALYZER MODULES" with the ENTER key or the → -key.
All modules joint to the platform or to the analyzer are available.
5 - 40
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.0 ppm
Analyzer Modules
EXPERT BACK
NGA Control module
>>> <<<
CLD
INFO
F1 F2 F3 F4 F5
Go to the line you want with the ↑↑ -key or the ↓↓ -key. If there are more than eight source
modules available you have to press the F3 key to go to another menu page.
Select the tag of the module with the ENTER key or the -key:
The display will go back to the menu "DIO-MODULE INPUTS" automatically. The tag of the module selected will appear in the line "Signal comes from".
3) Select any function:
Change to the line "Choose function..." with the ↓↓ -key.Press the ENTER key or the → -key to change to the submenu "FUNCTIONS":
CLD
EXPERT BACK
- FUNCTIONS -
SYS:Zero/Span-Cal
>>> <<<
95.0 ppm
AM:Zero-Cal
AM:Span-Cal.
AM:Range-1 AM:Range-2 AM:Range-3 AM:Range-4
SYS:Zero-Cal
INFO
F1 F2 F3 F4 F5
Go to the line you want with the ↓↓ -key or the ↑↑ -key. With the F3 key you can go to
other menu pages containing further available signals:
SYS:Program-Cal
SYS:Cancel-Cal
SYS:CAL-Test-Mode
SYS:AM-Zero-Gas
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 41
SYS:AM-Span-Gas1, 2, 3, 4
SYS: Function control
AM: Hold Outputs
CLD-AM: NOx
Select the function with the ENTER key or the → -key:
The display will go back to the menu "DIO-MODULE INPUTS" automatically. The function selected will appear in the line "Signal name".
The last four lines of variables in the menu "DIO-module inputs" only display some adjust­ments of the DIO input selected! They are not available:
"Slot ID:" displays the tag number of the DIO board selected. If there exist several DIO
modules in the platform you can change to the setup menu of each with the F3 key. The tag number of the DIO board selected will appear in the line "Slot ID".
"Signal name:" displays the signal selected for the current DIO input."Signal level:" displays the status of the signal selected."Signal comes from:" displays the source module selected for the current DIO input.
Configuration of the DIO module outputs:
1) Select the output number:
Press the ENTER key or the -key in the line "Outputnumber" of the menu
"DIO-MODULE OUTPUTS" and adjust an output number (1, ..., 24) with the
↑↑ -key or the ↓↓ -key.
Confirm the number selected with the ENTER key.
2) Choose the module:
Change with the ↓↓ -key to the line "Choose module...".Change to the submenu "ANALYZER MODULES" with the ENTER key or the → -key.
All modules joint to the platform or to the analyzer are available:
CLD
Analyzer Modules
NGA Control module
EXPERT BACK
F1 F2 F3 F4 F5
>>> <<<
95.0 ppm
CLD
INFO
5 - 42
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Go to the line you want with the ↑↑ -key or the ↓↓ -key. If there are more than eight source
modules available you have to press the F3 key to go to another menu page.
Select the tag of the module with the ENTER key or the -key:
The display will go back to the menu "DIO-MODULE OUTPUTS" automatically. The tag of the module selected will appear in the line "Signal comes from".
3) Select any signal:
Change to the line "Choose signal..." with the ↓↓ -key.Press the ENTER key or the → -key to change to the submenu "Signals":
CLD
Signals
MAINTENANCE REQUEST
CAL. IN PROGRESS ZERO IN PROGRESS SPAN IN PROGRESS
HOME ESCAPE MORE BACK INFO
F1 F2 F3 F4 F5
95.0 ppm
NORMAL
FAILURE
ZERO FAILED SPAN FAILED
Go to the line you want with the ↓↓ -key or the ↑↑ -key. With the F3 key you can go to
other menu pages containing further available signals:
RANGE OVERFLOW
RANGE UNDERFLOW
FLOW TO LOW
FLOW TO HIGH
RANGE 1, 2, 3, 4
OZONATOR OFF
NOx MODE
NO MODE
and further signals, if need be:
All signals of the control module and of the analyzer modules joint to the platform or MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "SIGNALS". The signals shown above will appear by using a CLD module as source module.
Select the signal with the ENTER key or the -key:
The display will go back to the menu "DIO-MODULE OUTPUTS" automatically. The signal selected will appear in the line "Signal name".
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 43
5 - 44
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.3 Auxiliary Module Set up
Main Menu — Expert controls and set up
↓↓

Auxiliary module set up

↓↓
CLD
Select any auxiliary module for set up.
Parameter: 3
MAIN BACK
F1 F2 F3 F4 F5
Auxiliary module set up
>>> <<<
95.0 ppm
INFO
In the menu "Auxiliary module set up" you can select any of the following types of micro processor controlled auxiliary modules:
1) I/O Module with three alarms
2) Autocal I/O
3) Syscal I/O If one of these auxiliary modules is available you will find its corresponding tag in a line of
this menu. If there exist more than eight of such modules you can change to another menu page with the F3 key. Press the ENTER key or the → -key in the line where the tag of the module is displayed to change to the setup submenus of the corresponding I/O board.
For further Informations please contact your customer service or look at the manual of the I/O board!
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 45
5 - 46
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.4 System Set Up
Main Menu — Expert controls and set up
↓↓

System set up

↓↓
CLD
System Set Up
Front panel control...
Display resolution... Auxiliary lines...
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
From the menu "System Set Up" you can change to several submenus to set up different parameters of the LCD and the single component display of the platform.
Scroll with the ↓↓ -key or the ↑↑ -key to select the menu line you want.Press the ENTER key or the → -key in the line selected to change to the corresponding
submenu.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 47
5.4.1 System Set Up

Front panel control

Main Menu — Expert controls and set up
↓↓
System set up
↓↓
Front panel control
↓↓
CLD
Front Panel Control
LCD brightness: 70 %
LCD viewing angle: 25 % Beeper enable: Off Beeper frequency: 1.00 kHz
Reading refresh rate: 1 s Drop out of menus after: 10 min
Date and time...
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Front panel control" you can set up several display parameters. Set up parameters:
Select any line of variables or the menu line with the ↓↓ -key or the ↑↑ -key.Select the variable with the ENTER key or the → -key.Select any digit with the ← -key or the -key and adjust a new value with
the ↑↑ -key or the ↓↓ -key resp. select the whole parameter with the ↑↑ -key or the ↓↓ -key.
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
Lines of variables "LCD brightness" and "LCD viewing angle": In these two lines you can determine the quality of the LCD reproduction. The best values
depend on your position to the screen and on the light of the environment.
Options: 20 to 100% for LCD brightness and 10 to 80% for LCD viewing angle.Usually: 70 to 90% (brightness) and 20 to 30% (angle).Caution: Be careful with the changing of these values! Perhaps you can not read the
display any longer if you use very extreme values! It is not possible to load standard
values via keyboard! Lines of variables "Beeper enable" and "Beeper frequency": The functionality of these two lines is not available at the moment!
5 - 48
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Line of variables "Reading refresh rate": In this line you can determine how often the display shall be built up in a certain time
interval. Options: 1s to 60s. Standard: 1s. Line of variables "Drop out of menus after": In this line you can decide if and after which time without front panel operations the display
shall change from the current menu to the single component display.
Options: 10s, 30s, 1min, 5min, 10min, 30min, Never.
Date and time set up:
Press the ENTER key or the → -key in the line "Date and time..." to change to the corresponding submenu:
CLD
Date and Time
Time display format (am/pm): 24 Hrs
Network updating: Enabled Minutes: 0 Hours: 12 (Entered as 24Hrs format)
Day: 21 Month: 2 Year: 1998
Current time: 13:48:20 21 Feb 1998
HOME ESCAPE SET INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Date and Time" you can set up the date and the time of the control module (platform or MLT analyzer).
Line of variables "Time display format (am/pm)": In this line you can decide if the time of the control module should be displayed in the 12
hours format or in the 24 hours format. Line of variables "Network updating": Select "Enabled" if you want that all modules combined with the platform or the MLT
analyzer can use the date and time setup of the control module. Else, select "Disabled". Set up a new date or time:
1) Press the ENTER key or the → -key in the line "Minutes", "Hours", "Day", "Month" or "Year" to select the corresponding number.
2) Select any digit with the ← -key or the →→ -key and adjust a new value with the ↑↑ -key or the ↓↓ -key.
Options: Minutes: 0 to 59, Hours: 0 to 23, Day: 1 to 28/30/31 depending on the month
selected, Month: 1 to 12, Year: 1996 to 2035.
3) Press the F4 key to set the new date or time. The new adjustment will appear in the line "Current time". The display in this line will be refreshed every five second.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 49
5.4.2 System Set Up

Display resolution

Main Menu — Expert controls and set up
↓↓
System set up
↓↓
Display resolution
↓↓
CLD
Display resolution
Maximum total number of digits: 4
Digits after the decimal point: 2
HOME ESCAPE INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Display resolution" you can determine the number of digits for the display of the measured values. These numbers do not alter the accuracy of measurement. The adjustments are not valid for the analyzer module!
Set up parameters:
Select one of the two lines of variables with the ↓↓ -key or the ↑↑ -key.Select the number with the ENTER key or the → -key.Select the digit with the ↑↑ -key or the ↓↓ -key.
Options: Maximum total number of digits: 3, 4, 5, 6
Digits after the decimal point: 0, 1, 2, 3.
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
Example:
Measured number: 1.2345Select 3 for total number of digits and 2 for digits after decimal point.The screen will display: 1.23
5 - 50
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
5.4.3 System Set Up
Main Menu — Expert controls and set up
↓↓
System set up
↓↓

Auxiliary lines

↓↓
Auxiliary Lines
CLD
Auxiliary lines
First auxiliary line: Enabled
Second auxiliary line: Enabled Third auxiliary line: Enabled Fourth auxiliary line: Enabled
First line source... Second line source... Third line source... Last line source...
HOME ESCAPE NEXT INFO
F1 F2 F3 F4 F5
95.0 ppm
In the menu "Auxiliary lines" you can determine if the last four lines in the single component display of the current channel shall appear. If you want to set up the lines of the other channels of the joint analyzer modules, press the F3 key.
Set up parameters:
Select any line of variables or any menu line with the ↓↓ -key or the ↑↑ -key.Select the parameter with the ENTER key or the → -key.Select "Enabled" or "Disabled" with the ↑↑ -key or the ↓↓ -key to enable or disable the
display of the corresponding auxiliary line in the single component display.
Confirm the parameter selected with the ENTER key or cancel and go back to the last
adjustment with the F2 key.
Choose the source module for the auxiliary lines:
Press the ENTER key or the → -key in the line "First line source...", "Second line source...", "Third line source..." or "Last line source..." to change to the corresponding submenu. The tags of all modules combined with the platform or the MLT analyzer are available:
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 51
CLD
First auxiliary line source
Select the module to use as source for this line in the single analyzer display screen.
HOME ESCAPE MORE BACK INFO
95.0 ppm
CLD
F1 F2 F3 F4 F5
Go to the line you want with the ↑↑ -key or the ↓↓ -key. If there are more than eight source
modules available you have to press the F3 key to go to another menu page.
Select the tag of the module with the ENTER key or the -key:
The display will go back to the menu "Auxiliary lines" automatically. The selection will be done.
5 - 52
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Loading...