Emerson NGA2000 User Manual

NGA 2000
Software Manual
CLD - Analyzer Module
Chemiluminescence - Measurement
(combined with NGA 2000 - Platform / MLT - Analyzer)
CLD Software V e rsion 2.3
(Platform / MLT Software Version 3.1.X)
Catalog No.: 90 003 749
90003749(1) [NGA-e (CLD-Software 2.3 / MLT-Software 3.1.X)] 04/98
Managing The Process Better
This Operation Manual includes information about the operation of the instrument. Information about the additional indications and notes regarding maintenance, troubleshooting and repair are found in the accompanying Maintenance & Operation Manual.
T roubleshooting, component replacement and internal adjustments must be made by qualified service personnel only.
Fisher-Rosemount GmbH & Co does not take responsibility for an y omissions or errors in this manual. Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of this manual, is expressly e xcluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations. T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings in this manual or provided on the instrument.
Misprints and alterations reserved
©
1998 by FISHER-ROSEMOUNT GmbH & Co. (PAD)
1. Edition: 04/98
Read this operation manual carefully before attempting to operate the analyzer ! For expedient handling of reports of defects , please include the model and serial number which can be read on the instrument identity plate.
Fisher - Rosemount GmbH & Co.
Industriestrasse 1
D - 63594 Hasselroth • Germany
Phone + 49 (6055) 884-0
Telefax + 49 (6055) 884-209
Internet: http://www.processanalytic.com
90003749(1) [NGA-e (CLD-Software 2.3 / MLT-Software 3.1.X)] 04/98

Contents

1 Introduction 1 - 1
2 Structure of Menus 2 - 1
3 Display and Keyboard 3 - 1
3.1 Starting and Initializing...................................................................................3 - 1
3.2 Display and Function...................................................................................... 3 - 1
3.3 Operating Keys ..............................................................................................3 - 2
3.4 Lines...............................................................................................................3 - 2
3.5 Important Functions of the Softkeys...............................................................3 - 3
3.6 Entering/Changing of Variables......................................................................3 - 4
3.7 The STATUS Key...........................................................................................3 - 5
4 Basic Controls 4 - 1
4.1 Measurement 4 - 3
4.1.1 Multi Component Display: Change of Channel ..............................................4 - 3
4.1.2 Multi Component Display: Sequence Setup...................................................4 - 5
4.1.3 Range Setting ................................................................................................4 - 7
4.1.4 Range and Functional Control .......................................................................4 -11
4.1.5 Enable/Disable Ozonator Power....................................................................4 -15
4.1.6 Measurement Mode: NO ↔ NOx....................................................................4 -19
4.2 Calibration 4 -21
4.2.1 Zeroing...........................................................................................................4 -21
4.2.2 Spanning........................................................................................................ 4 -25
4.2.3 If It Won't Calibrate.........................................................................................4 -29
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
I
5 Expert Controls and Set up 5 - 1
5.1 Analyzer Module Set up 5 - 3
5.1.1 Calibration/Calibration Gas List......................................................................5 - 4
5.1.2 Calibration Parameters...................................................................................5 - 9
5.1.3 Gas Measurement Parameters ......................................................................5 -11
5.1.4 Linearization Parameters................................................................................5 -12
5.1.5 Response Time/Delay Parameters.................................................................5 -14
5.1.6 Range Settings...............................................................................................5 -16
5.1.7 Units ...............................................................................................................5 -17
5.1.8 Linearization Functions...................................................................................5 -18
5.1.9 Analyzer Parameter List .................................................................................5 -23
5.1.10 Physical Measurement Parameters................................................................5 -26
5.1.11 Displayed Parameters ....................................................................................5 -28
5.2 Auxiliary Module Controls 5 -29
5.2.1 Local SIO Configuration Sections...................................................................5 -30
5.2.2 Set-up DIO Module(s).....................................................................................5 -39
5.3 Auxiliary Module Set up 5 -45
5.4 System Set up 5 -47
5.4.1 Front Panel Control ........................................................................................5 -48
5.4.2 Display Resolution..........................................................................................5 -50
5.4.3 Auxiliary Lines.................................................................................................5 -51
5.5 Expert Analyzer Controls 5 -53
5.5.1 Range Settings...............................................................................................5 -55
5.5.2 Zero/Span Calibration.....................................................................................5 -56
5.5.3 Physical Measurements..................................................................................5 -60
II
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
6 Technical Level Configuration 6 - 1
6.1 System Set up 6 - 3
6.1.1 Main Display Configuration ............................................................................6 - 4
6.1.2 Front Panel Control........................................................................................6 - 9
6.1.3 Date and Time Set up....................................................................................6 -11
6.1.4 Module Binding ..............................................................................................6 -12
6.1.5 System Reset.................................................................................................6 -14
6.1.6 Security Codes...............................................................................................6 -15
6.2 Service Menus 6 -17
6.2.1 Control Module Data......................................................................................6 -18
6.2.2 Analyzer Module Data.................................................................................... 6 -19
6.2.3 Control Module History...................................................................................6 -21
6.2.4 Analyzer Module History ................................................................................6 -23
6.3 Diagnostic Menus 6 -25
6.3.1 Control Module Diagnostics ...........................................................................6 -26
6.3.2 Analyzer Module Diagnostics.........................................................................6 -27
6.3.2.1 Power Supply Voltages ...................................................................6 -28
6.3.2.2 Primary Variable Parameters ..........................................................6 -29
6.3.2.3 Physical Measurements ..................................................................6 -33
6.3.2.4 Temperature Control.......................................................................6 -35
6.3.2.5 Flow Control Parameters.................................................................6 -36
6.3.2.6 Miscellaneous Control Parameters..................................................6 -38
6.3.2.7 Trend Display Control......................................................................6 -39
6.3.2.8 Software Diagnostics.......................................................................6 -41
6.3.2.9 Analyzer Start up............................................................................. 6 -42
6.3.2.10 NOx Converter Efficiency.................................................................6 -43
6.4 Other Module Diagnostic Menus 6 -44
6.5 Listing of all Modules 6 -45
Supplement: System Calibration
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
III
IV
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
1 Introduction
This software manual describes step by step how to operate successfully with the ½ 19" and 19" MLT analyzer module and analyzer of the NGA 2000 Series from . Chapter two shows the structure of the software menus. Chapter three describes the display and the keyboard of the analyzer. Chapter four describes the basic controls with detailed illustrations. So you can easily compare the actual display of the analyzer module with the illustrations of the manual. Chapter five describes the expert controls and chapter six describes the technical level configuration. The layout of both chapters is not as detailed as in chapter four. Normally, the way to a certain software menu is described with the software catchwords, you have to press to reach this menu. You will find the illustration of the corresponding LCD screen at the end of the catchword listing. After that you can read the meaning of the functions and variables of each expert or technical level menu. Some contents of the expert controls are not important for every customer. It depends on the configuration of your NGA 2000 system, relative to the following components:
Control Module CMAnalyzer Module AMInput/Output Modules I/O's (SIO, DIO)
You can distinguish the following system units and SIO/DIO configurations:
System Unit SIO/DIO-Configuration Corresponding Chapter
1
CLD analyzer module (AM):
without front panel i.e.
Local I/O's are not existing
⁄⁄
without control unit
can be combined with a
platform or an MLT analyzer
Platform (CM software):
Control unit with front panel
without measurement
channels
1 SIO and up to 4 DIO's
can be installed in the platform
SIO and DIO can be
chapter 5.2
p. 29 to 44
configured for all channels combined with the platform
MLT analyzer (CM and MLT AM software):
Analyzer with front panel
CM and AM software in the
same analyzer, i.e. all functions of the control unit and of the AM are united
1 SIO and 1 DIO can be
installed in the MLT analyzer
SIO and DIO can be con-
figured for all channels combined with the MLT analyzer
chapter 5.2
p. 29 to 44
in the same implement
1
The configuration of other analyzer modules like MLT modules is described in their own manuals !
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
1 - 1
The following illustrations shall make plain the connection between the hardware configu­ration and the software setup of the modules:
NGA 2000 System via Platform
(see 5.1.17) (see 5.1.18)
(other manuals)
(other manuals)
(other manuals)
Local
I/O's
SIO DIO
Analyzer
Modules (AM's)
Control
Module (CM)
MLT
CLD
FID
Platform
PMD
NGA 2000 System via MLT Analyzer
Auxiliary Modules
SIO
DIO
Other I/O´s
(see 5.2.1)
(see 5.2.2)
(other manuals)
Local
I/O's
(see 5.1/5.5)
SIO DIO
Analyzer
Modules (AM's)
CLD
MLT
FID(other manuals)
PMD(other manuals)
Control
Modul (CM)
MLT Analyzer
Auxiliary Modules
SIO
DIO
(see 5.2.1)
(see 5.2.2)
1 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Notes:
*
In the menu structure shown above, you will find the most important ramifications of the MLT software 3.1.X for the
MLT platform and analyzer versions. The menu points related to the CLD analyzer module are marked with
They belong to the CLD software revision number 2.3.
*
At each menu point of the "Expert controls" and "Technical level" you can find which functionality will be set up:
odule), CM (
odule), I/O (
utput Module).
*
Menu lines ending with three points (...) are followed by submenus with further functions and set-ups.
-
-
-
-
-
Technical level
configuration...
2 Structure of Menus
"AM".
Main Menu
(CM)
- System set up...
and set up...
Expert controls
controls...
- Expert analyzer
(AM, CM, I/O's:
SIO, DIO and other I/O's)
- Service menus...
(AM)
controls...
- Auxiliary module
(AM, CM, I/O's:
SIO, DIO and other I/O's)
- Diagnostic menus...
(I/O's: SIO, DIO)
(CM)
- System set up...
diagnostic menus...
(Other I/O's)
- Other module
(AM)
- Analyzer module set up...
(AM, CM, I/O's:
SIO, DIO and other I/O's)
- Listing of all modules...
(Other I/O's)
- Auxiliary module set up...
(CM)
- System calibration...
Input-/O
Control M
Basic controls... (AM)
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Range set-up
Range and
functional control
Ozonator power
enable/disable
NGA 2000
Measurement mode
selection: NO/NOx-Zeroing...
Spanning...
AM (Analyzer M
2 - 1
2 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X] 04/98
3 Display and Keyboard

3.1 Starting and Initializing

After switching on the CLD analyzer module (in a platform or part of a NGA network), the initialization procedure will be performed. A self control of the analyzer modules or the analyzer(s) is running. You can see a sequence of several displays. They show the status of initialization, revision notes of the MLT software and the tag. At the end of this procedure you can see the single component display of channel one. It is the origin to all the other channels, menus and submenus. The instructions of the basic controls (chapter four) are all beginning with the single com­ponent display. The actual display might differ from the shown one because the customer can configure it according to his requirements.
TAG
here e.g.: CLD Analyzer Module
CLD
95.00 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Typical starting menu after switching on
Range: 4
340 hPa
0.5 C
51.5 C 360 C

3.2 Display and Function

The LCD-screen shows all measurement values of the analyzer and all customer instruc­tions. You can operate with five function keys, four arrow keys (cursors) and the enter key. The function of each key depends on:
the type of analyzer/analyzer module usedthe optional auxiliary modules (e.g. I/O boards) usedthe individual menu displayed
In case of power failure all customer specific module parameters are saved by a battery­powered buffer.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 1

3.3 Operating Keys

CLD
95.00 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 4
340 hPa
0.5 C
51.5 C 360 C
Function Keys:
Keys without defined functionsThe current function depends on
the menu selected
The softkey legend is shown on
the display above the key

3.4 Lines

Cursor keys:
↑↑ -key / ↓↓ -key:
Line up / line down
within the same menu
Alteration of numbers,
variables or digits
-key/ -key:
Moving back/forwards
between the pages of a menu
Selection of digits
Enter Key:
To confirm a previously entered value
(variable)
To start a selected function
(Alternative: -key)
To go into a menu (via menu line)
Lines can be selected by the ↓↓ -key or the ↑↑ -key. The selected line is displayed white on black. You have four different types of lines in the menu:
Menu line...
Line ending with three dots.You will go to a submenu by pressing the ENTER key or the → -key in such a line.
Function line !
Line ending with an exclamation-mark.You will start a function (e.g. NO/NO
-toggle) by pressing the ENTER key or
x
the → -key in such a line.
Line of variables:
Line ending with a colon.Display of module parameters (variables).Some parameters can be changed (e.g. range number), some parameters display only
a status (e.g. calibration status) and cannot be changed.
Text line
Line without any punctuation marks.Only display of informations (e.g. analyzer module tag).
3 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
3 Display and Keyboard
Text Line
Line of Variables
Function Line
Menu Line
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Zero/span calibration
Measurement range number: 3 Zero gas concentration: 0.0 ppm Span gas concentration: 100.0 ppm Sample flow: 1100 ml/min Raw measurement signal: 521590 Measurement gas: NOx NO/NOx toggle! Status: READY Result... Calibration adjustment limits: Disabled
HOME FACTORS ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Function Key Legend
3.5 Important Functions of the Softkeys DISPLAY
Change from the single component display to the multi component display.
MEASURE
Change from the main menu to the single component display.
STATUS (see 3.7!)
Change to the status report: Display of the most important parameters and
status informations of the CLD module or of the current channel.
If available, this command is always assigned to the F2 key.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 3
MENU
Change from the single component display to the main menu or from several submenus
to higher menus.
ESCAPE or <<<
Changing back to the last menu page selected.Reset of a changed but not confirmed parameter to the former value resp. to zero.
BACK
Reset of a changed but not confirmed parameter to the former value.
LOCK
Lock of any operation level, if activated. Each of the three operation levels has to be
activated separately.
F4 key of the main menu.
NEXT
Change of the channel in the same menu. This will only be possible, if several channels
of an analyzer or of combined analyzers are existing. Then, all available channels can be selected one after another.
In the main menu you can move among all channels of the connected analyzers and
analyzer modules. In the submenus you can only move among the channels of the current analyzer or analyzer module.
INFO
Change to menus containing status or help informations of the actual menu.If available, this function is always assigned to the F5 key.
3.6 Entering/Changing of Variables ENTER Key
If you will press this key in a line of variables, the parameter will be displayed white on
black and can be changed. After you will have set up a new value, you can confirm it by pressing the ENTER key again.
↑↑ -key / ↓↓ -key
Function will depend on the variable selected: - Changing of the parameter values
- Scrolling among variables selected
- Changing of digits or characters
Numbers consisting of several digits will increase/decrease totally, not digit by digit.
-key / -key
Selection of digits within a parameter.For some variables you can change the quantity of digits or characters.
3 - 4
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
3 Display and Keyboard

3.7 The STATUS Key

From the single component display or from the main menu you can change with the F2 key (STATUS) to the menu "Current measurement parameters":
CLD
Current measurement parameters
Analyzer gas measured: NOx Measurement range number: 3 Range change control: LOCAL
Linearization mode: Disabled Analyzer operational state: STANDBY
Analyzer alarm state: NORMAL Alarms reported: FAILURE
HOME ESCAPE MORE INFO
F1 F2 F3 F4 F5
CLD
Current measurement parameters
Response time: 3.0 s Bypass flow: 1100 ml/min Capillary pressure: 340 hPa Detector temperature: 0.5 C
95.0 ppm
95.0 ppm
HOME ESCAPE MORE INFO
F1 F2 F3 F4 F5
In the menu "Current measurement parameters" you can control the status of the CLD analyzer module. With the F3 key (MORE) you can change to a further menu page of this menu. With the F2 key (ESCAPE) you can always go back to the main menu. From the second menu page you can return to the first with the F4 key (MORE) or the ← -key. In the menu "Current measurement parameters" you will only find status informations. You cannot change the setups there. If you want to modify some parameters, you have to change to the menu "Basic Controls" (see chapter 4) or to the menu "Analyzer module set up" (see chapter 5.1) resp. "Expert analyzer controls" (see chapter 5.5).
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
3 - 5
3 - 6
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4 Basic Controls
In the chapter "Basic Controls" you will find the most important functions to set up your CLD analyzer module via an NGA front panel:
1) Measurement p. 4 - 3pp (Chapter 4.1.1 and 4.1.2, p. 4-3 to 4-6, are only important for systems with several channels.)
2) Calibration p. 4 - 21pp
All steps are figured with detailed illustrations and operation instructions. In the left column you can see display and keyboard of the NGA front panel. In the right column you can read the instructions and notes. All instructions will begin with any single component display and will end with the corresponding single component display after the setups are done. The keys you have to press are illustrated in black. The instructions are completed with notes and further informations.
Example: You want to change from the single component display of NOx to the main menu
Picture one shows the starting situation: single component display of NO
.
x
If you press the F3 key you will change to menu page shown in picture 2: Main Menu.
Left column: Right column: LCD and keyboard Instructions and notes
CLD
95.00 ppm NOx
0 250
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
CLD
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
F1 F2 F3 F4 F5
Range: 4
340 hPa
0.5 C
51.5 C 360 C
95.0 ppm
Main Menu
Change to the
Main Menu
F3
Press
or
Next instruction
or step, here e.g.: Change to the Basic Controls:
Press
or
90003479(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 1
4 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

4.1.1 Measurement - Multi Component Display:

CLD
95.00 ppm NOx
0 250
Range: 4
Change of Channel
Change to the multi
component display
F1
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
340 hPa
0.5 C
51.5 C 360 C
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
Note: You can change to the multi component display from each single component display.
Enable the selecting:
">"
F1
Press
or
Note: If no channels of other analyzer modules are combined to the CLD, only the bargraph of the CLD module will be displayed. Then, a change of channel will not be possible !
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
NGA 2000
Select any channel
Press or as often as necessary to
put the > - mark into the selected line.
Example: Change from NOX (CLD) to CO (MLT channel 1)
4 - 3
CLD
95.00
45.00
333.0
150.0
20.00
CLD
ppm NOx
MLT25/CH1/R2
ppm CO
MLT25/CH2/R2
ppm SO2
MLT25/CH3/R2
ppm NO
MLT25/CH4/R2
%O2
Change to the single
component display of the channel selected
F1
Press
SELECT STATUS TAGS OFF
INFO
F1 F2 F3 F4 F5
MLT25/CH1/R2
45.00 ppm CO
0 250
Temperature: Function Control: Maintenance Request: Failure:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 2
25.0 C No No No
Single component
display of the channel selected will appear
4 - 4
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4.1.2 Measurement - Multi Component Display:
CLD
95.00 ppm NOx
0 250
Range: 4
Sequence setup
Change to the multi
component display
F1
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
340 hPa
0.5 C
51.5 C 360 C
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
Enable the selecting:
">"
F1
Press
or
Note: If no channels of other analyzer modules are combined to the CLD, only the bargraph of the CLD module will be displayed. Then, the sequence setup will not be possible !
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
INFO
NGA 2000
Select the channel
for position one
Press or as often as necessary to
put the > - mark into the selected line.
Example: MLT channel 2 (SO2) shall appear at position one of the multi component display.
4 - 5
CLD
95.00
45.00
333.0
150.0
20.00
SELECT STATUS TAGS OFF
F1 F2 F3 F4 F5
MLT25/CH2/R2
333.0
MLT25/CH1/R2 MLT25/CH2/R2 MLT25/CH3/R2 MLT25/CH4/R2
MLT25/CH2/R2
95.00
45.00
150.0
20.00
SELECT STATUS TAGS OFF
MLT25/CH1/R2 MLT25/CH3/R2 MLT25/CH4/R2
CLD
ppm NOx
ppm CO
ppm SO2
ppm NO
%O2
ppm SO2
CLD
ppm NOx
ppm CO ppm NO
%O2
INFO
INFO
Assignation of the
channel selected to position one
Press
Note: The channel selected will appear at position one. All the other channels will automatically appear one line below their last position.
Multi component
display will appear with the changed sequence
Note: Now, you can change to any available single component display. The method is described in the last chapter (4.1.1).
F1 F2 F3 F4 F5
4 - 6
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
95.00 ppm NOx
0 250
Range: 4

4.1.3 Measurement - Range Setting

Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Change to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Choose the
measurement range number
Press
or
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 7
CLD
Basic Controls
Measurement range number: Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
4
ESCAPE INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
4
ESCAPE INFO
95.0 ppm
F1 F2 F3 F4 F5
Select the
range number
4
Press or as often as necessary to select another range number.
Example: Change from range 4 (250 ppm) to range 3 (100 ppm)
Confirm the
range selected
3
Press
Note: If you don't want to set the new range number, you can reactivate the former selection: Press F2 (ESCAPE) and confirm it with ENTER.
CLD
Basic Controls
Measurement range number: 3
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
F1 F2 F3 F4 F5
4 - 8
95.0 ppm
NGA 2000
Change to the
main menu
Press or
F1
2 times
Note: The range upper limit (line 2) of the range selected will be set up automatically after the range setting is confirmed.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
F1 F2 F3 F4 F5
4.1.3 Measurement - Range Setting
Change to the single
component display
F1
Press
CLD
95.00 ppm NOx
0 100
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
F1 F2 F3 F4 F5
Range: 3
340 hPa
0.5 C
51.5 C 360 C
Single component
display will appear with the new range
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 9
4 - 10
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

4.1.4 Measurement - Range and Functional Control

CLD
95.00 ppm NOx
0 250
Range: 4
Change to the
main menu
F3
Press
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
or
Cahnge to the
basic controls
Press
or
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4
Range upper limit: 250.0 ppm Range and functional control: LOCAL Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
Change to the
line "Range and functional control"
Press or as often as necessary until the line "Range and functional control" will appear white on black.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 11
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm
Range and functional control: LOCAL
Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
LOCAL
ESCAPE INFO
95.0 ppm
LOCAL
F1 F2 F3 F4 F5
Choose the
parameter
Press
or
Set up the range
and functional control you want
++ Pay attention to
the note at p. 4-14 !
Press or as often as necessary to select the parameter you want.
Options: LOCAL, AUTO, REMOTE
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm Range and functional control: Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
LOCAL
ESCAPE INFO
F1 F2 F3 F4 F5
4 - 12
95.0 ppm
AUTO
NGA 2000
Confirm the
parameter selected
Press
Note: If you don't want to set the new parameter, you can reactivate the former selection: Press F2 (ESCAPE) and confirm it with ENTER.
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
4.1.4 Measurement - Range and Functional Control
CLD
Basic Controls
Measurement range number: 4 Range upper limit: 250.0 ppm
Range and functional control: AUTO
Ranges with valid calibration: 4 Calibration status: READY If it won´t calibrate... Status: STANDBY Measurement mode: NOx Ozonator status: On Ozonator power: Enabled
HOME NO/NOx ZERO SPAN INFO
F1 F2 F3 F4 F5
95.0 ppm
Change to the
main menu
Press or
F1
2 times
CLD
Main Menu
Basic controls...
Expert controls and set up... (Operational configuration)
Technical level configuration... (Diagnostic and manufacturing/service)
MEASURE STATUS NEXT LOCK INFO
95.0 ppm
F1 F2 F3 F4 F5
CLD
95.00 ppm NOx
0 250
Range: 4
Change to the single
component display
F1
Press
Single component
display will appear after the changing of the range and functional control
Sample press.: Detector temperature: Block temperature: Converter temp.:
DISPLAY STATUS MENU NEXT INFO
340 hPa
0.5 C
51.5 C 360 C
F1 F2 F3 F4 F5
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 13
Range and functional control – Options:
LOCAL:
The platform will be controlled via the display.
AUTO:
The range will be selected automatically ("auto range"). This function will only be available with an I/O board with 3 alarms (PIN 70 656 193), because the auto ranging parameters are located on the I/O board (see its own manual)!
REMOTE:
The range selection will be controlled by digital inputs, e.g. via a DIO board or I/O board with 3 alarms.
4 - 14
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Loading...
+ 148 hidden pages