This Operation Manual includes information about the operation of the instrument.
Information about the additional indications and notes regarding maintenance, troubleshooting and repair
are found in the accompanying Maintenance & Operation Manual.
T roubleshooting, component replacement and internal adjustments must be made by qualified
service personnel only.
Fisher-Rosemount GmbH & Co does not take responsibility for an y omissions or errors in this manual.
Any liability for direct or indirect damages, which might occur in connection with the deliv ery or the use of
this manual, is expressly e xcluded to the extend permitted by applicable law.
This instrument has left the works in good order according to safety regulations.
T o maintain this operating condition, the user must strictly follow the instructions and consider the warnings
in this manual or provided on the instrument.
Read this operation manual carefully before attempting to operate the analyzer !
For expedient handling of reports of defects , please include the model and serial number which
can be read on the instrument identity plate.
This software manual describes step by step how to operate successfully with the ½ 19"
and 19" MLT analyzer module and analyzer of the NGA 2000 Series from .
Chapter two shows the structure of the software menus. Chapter three describes the
display and the keyboard of the analyzer. Chapter four describes the basic controls with
detailed illustrations. So you can easily compare the actual display of the analyzer module
with the illustrations of the manual.
Chapter five describes the expert controls and chapter six describes the technical level
configuration. The layout of both chapters is not as detailed as in chapter four. Normally,
the way to a certain software menu is described with the software catchwords, you have to
press to reach this menu. You will find the illustration of the corresponding LCD screen at
the end of the catchword listing. After that you can read the meaning of the functions and
variables of each expert or technical level menu.
Some contents of the expert controls are not important for every customer. It depends on
the configuration of your NGA 2000 system, relative to the following components:
After switching on the CLD analyzer module (in a platform or part of a NGA network), the
initialization procedure will be performed. A self control of the analyzer modules or the
analyzer(s) is running. You can see a sequence of several displays. They show the status
of initialization, revision notes of the MLT software and the tag. At the end
of this procedure you can see the single component display of channel one. It is the origin
to all the other channels, menus and submenus.
The instructions of the basic controls (chapter four) are all beginning with the single component display. The actual display might differ from the shown one because the customer
can configure it according to his requirements.
The LCD-screen shows all measurement values of the analyzer and all customer instructions. You can operate with five function keys, four arrow keys (cursors) and the enter key.
The function of each key depends on:
♦ the type of analyzer/analyzer module used
♦ the optional auxiliary modules (e.g. I/O boards) used
♦ the individual menu displayed
In case of power failure all customer specific module parameters are saved by a batterypowered buffer.
♦ Keys without defined functions
♦ The current function depends on
the menu selected
♦ The softkey legend is shown on
the display above the key
3.4Lines
Cursor keys:
↑↑ -key / ↓↓ -key:
♦ Line up / line down
within the same menu
♦ Alteration of numbers,
variables or digits
←← -key/ →→ -key:
♦ Moving back/forwards
between the pages of
a menu
♦ Selection of digits
Enter Key:
♦ To confirm a previously entered value
(variable)
♦ To start a selected function
(Alternative: →→-key)
♦ To go into a menu (via menu line)
Lines can be selected by the ↓↓ -key or the ↑↑ -key. The selected line is displayed white on
black. You have four different types of lines in the menu:
Menu line...
♦ Line ending with three dots.
♦ You will go to a submenu by pressing the ENTER key or the →→ -key in such a line.
Function line !
♦ Line ending with an exclamation-mark.
♦ You will start a function (e.g. NO/NO
-toggle) by pressing the ENTER key or
x
the →→ -key in such a line.
Line of variables:
♦ Line ending with a colon.
♦ Display of module parameters (variables).
♦ Some parameters can be changed (e.g. range number), some parameters display only
a status (e.g. calibration status) and cannot be changed.
Text line
♦ Line without any punctuation marks.
♦ Only display of informations (e.g. analyzer module tag).
Measurement range number:4
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
HOMENO/NOxZEROSPANINFO
95.0 ppm
F1F2F3F4F5
CLD
Zero/span calibration
Measurement range number:3
Zero gas concentration:0.0 ppm
Span gas concentration:100.0 ppm
Sample flow:1100 ml/min
Raw measurement signal:521590
Measurement gas:NOx
NO/NOx toggle!
Status:READY
Result...
Calibration adjustment limits:Disabled
HOMEFACTORSZEROSPANINFO
95.0 ppm
F1F2F3F4F5
Function Key Legend
3.5Important Functions of the Softkeys
DISPLAY
♦ Change from the single component display to the multi component display.
MEASURE
♦ Change from the main menu to the single component display.
STATUS (see 3.7!)
♦ Change to the status report: Display of the most important parameters and
status informations of the CLD module or of the current channel.
♦ If available, this command is always assigned to the F2 key.
Response time:3.0 s
Bypass flow:1100 ml/min
Capillary pressure:340 hPa
Detector temperature:0.5 C
95.0 ppm
95.0 ppm
HOMEESCAPEMOREINFO
F1F2F3F4F5
In the menu "Current measurement parameters" you can control the status of the CLD
analyzer module. With the F3 key (MORE) you can change to a further menu page of this
menu. With the F2 key (ESCAPE) you can always go back to the main menu. From the
second menu page you can return to the first with the F4 key (MORE) or the ←← -key.
In the menu "Current measurement parameters" you will only find status informations. You
cannot change the setups there. If you want to modify some parameters, you have to
change to the menu "Basic Controls" (see chapter 4) or to the menu "Analyzer module set
up" (see chapter 5.1) resp. "Expert analyzer controls" (see chapter 5.5).
In the chapter "Basic Controls" you will find the most important functions to set up your
CLD analyzer module via an NGA front panel:
1) Measurement p. 4 - 3pp (Chapter 4.1.1 and 4.1.2, p. 4-3 to 4-6, are only important
for systems with several channels.)
2) Calibration p. 4 - 21pp
All steps are figured with detailed illustrations and operation instructions. In the left column
you can see display and keyboard of the NGA front panel. In the right column you can
read the instructions and notes. All instructions will begin with any single component
display and will end with the corresponding single component display after the setups are
done. The keys you have to press are illustrated in black. The instructions are completed
with notes and further informations.
Example: You want to change from the single component display of NOx to the main menu
• Picture one shows the starting situation: single component display of NO
.
x
• If you press the F3 key you will change to menu page shown in picture 2: Main Menu.
Left column:Right column:
LCD and keyboardInstructions and notes
Note:
You can change to the
multi component display
from each single
component display.
⇒ ⇒ Enable the selecting:
">"
F1
Press
or
Note:
If no channels of other
analyzer modules are
combined to the CLD,
only the bargraph of the
CLD module will be displayed.
Then, a change of channel
will not be possible !
Note:
If no channels of other
analyzer modules are
combined to the CLD,
only the bargraph of the
CLD module will be displayed.
Then, the sequence setup
will not be possible !
Expert controls and set up...
(Operational configuration)
Technical level configuration...
(Diagnostic and manufacturing/service)
MEASURESTATUSNEXTLOCKINFO
95.0 ppm
or
⇒ ⇒ Change to the
basic controls
Press
or
F1F2F3F4F5
CLD
Basic Controls
Measurement range number:4
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
Measurement range number:
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
4
ESCAPEINFO
95.0 ppm
F1F2F3F4F5
CLD
Basic Controls
Measurement range number:
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
4
ESCAPEINFO
95.0 ppm
F1F2F3F4F5
⇒ ⇒ Select the
range number
4
Press or
as often as necessary
to select another range
number.
Example:
Change from
range 4 (250 ppm) to
range 3 (100 ppm)
⇒ ⇒ Confirm the
range selected
3
Press
Note:
If you don't want to set the
new range number, you
can reactivate the former
selection:
Press F2 (ESCAPE) and
confirm it with ENTER.
CLD
Basic Controls
Measurement range number:3
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
HOMENO/NOxZEROSPANINFO
F1F2F3F4F5
4 - 8
95.0 ppm
NGA 2000
⇒ ⇒ Change to the
main menu
Press or
F1
2 times
Note:
The range upper limit (line 2)
of the range selected will be
set up automatically after the
range setting is confirmed.
Expert controls and set up...
(Operational configuration)
Technical level configuration...
(Diagnostic and manufacturing/service)
MEASURESTATUSNEXTLOCKINFO
95.0 ppm
or
⇒ ⇒ Cahnge to the
basic controls
Press
or
F1F2F3F4F5
CLD
Basic Controls
Measurement range number:4
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
HOMENO/NOxZEROSPANINFO
95.0 ppm
F1F2F3F4F5
⇒ ⇒ Change to the
line "Range and
functional control"
Press or
as often as necessary
until the line "Range and
functional control" will
appear white on black.
Measurement range number:4
Range upper limit:250.0 ppm
Range and functional control:LOCAL
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
HOMENO/NOxZEROSPANINFO
95.0 ppm
F1F2F3F4F5
CLD
Basic Controls
Measurement range number:4
Range upper limit:250.0 ppm
Range and functional control:
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
LOCAL
ESCAPEINFO
95.0 ppm
LOCAL
F1F2F3F4F5
⇒ ⇒ Choose the
parameter
Press
or
⇒ ⇒ Set up the range
and functional
control you want
++ Pay attention to
the note at p. 4-14 !
Press or
as often as necessary
to select the parameter
you want.
Options:
LOCAL, AUTO, REMOTE
CLD
Basic Controls
Measurement range number:4
Range upper limit:250.0 ppm
Range and functional control:
Ranges with valid calibration:4
Calibration status:READY
If it won´t calibrate...
Status:STANDBY
Measurement mode:NOx
Ozonator status:On
Ozonator power:Enabled
LOCAL
ESCAPEINFO
F1F2F3F4F5
4 - 12
95.0 ppm
AUTO
NGA 2000
⇒ ⇒ Confirm the
parameter selected
Press
Note:
If you don't want to set the
new parameter, you can
reactivate the former
selection:
Press F2 (ESCAPE) and
confirm it with ENTER.
The range will be selected automatically ("auto range"). This
function will only be available with an I/O board with 3 alarms(PIN 70 656 193), because the auto ranging parameters are
located on the I/O board (see its own manual)!
• REMOTE:
The range selection will be controlled by digital inputs, e.g. via
a DIO board or I/O board with 3 alarms.