Emerson 2081T, 2081C User Manual

Model 2081C/T
Conductivity Microprocessor Transmitter
Instruction Manual
PN 51-2081C/T/rev.B April 2003
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
• Read all instructions prior to installing, operating, and servic­ing the product. If this Instruction Manual is not the correct manual, telephone 1-800-654-7768 and the requested man­ual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical haz­ards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being per­formed by qualified persons, to prevent electrical shock and personal injury.
NOTE
The Model 2081C/T Two-wire transmitter is certified for use in areas requiring intrinsic safety area by FM, CSA, and CEN­ELEC. However, the CENELEC certification requires that a Model 2081C/T used with contacting sensors be certified sep­arately from those using inductive (toroidal) sensors. The type of sensor must be specified with the order for a CENELEC certified transmitter.
Since this decision is required prior to shipment, the instruction manual covers both configurations. However, if the trans­mitter is reconfigured in the field, it may not meet CENELEC certification conditions.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2003
About This Document
This manual contains instructions for installation and operation of the Model 2081C/T Conductivity Transmitter. The following list provides notes concerning all revisions of this document.
Rev. Level Date Notes
A 7/01 This is the initial release of the product manual. The manual has been refor-
matted to reflect the Emerson documentation style and updated to reflect any changes in the product offering.
B 4/03 Updated CE certification
MODEL 2081C/T CONDUCTIVITY
MICROPROCESSOR TRANSMITTER
TABLE OF CONTENTS
Section Title Page
1.0 INSTALLATION ...................................................................................................... 1
1.1 Unpacking and Inspection....................................................................................... 1
1.2 Mechanical Installation............................................................................................ 1
1.3 Wiring ...................................................................................................................... 1
1.4 Fault Mode and Security Jumpers .......................................................................... 2
1.5 Calibration and Setup.............................................................................................. 3
1.6 Hazardous Locations — Explosion Proof Installation ............................................. 3
1.7 Hazardous Locations — Intrinsically Safe Installation ............................................ 3
2.0 DESCRIPTION OF CONTROLS ............................................................................ 17
2.1 General.................................................................................................................... 17
2.2 Menu Selection ....................................................................................................... 17
2.3 Value Adjustment .................................................................................................... 17
3.0 RANGE CONFIGURATION .................................................................................... 18
3.1 General.................................................................................................................... 18
4.0 START UP AND CALIBRATION ............................................................................ 19
4.1 General.................................................................................................................... 19
4.2 Temperature Calibration.......................................................................................... 19
4.3 Entering the Cell Constant ...................................................................................... 19
4.4 Zeroing the System ................................................................................................. 19
4.5 Initial Loop Calibration............................................................................................. 19
4.6 Routine Standardization.......................................................................................... 21
4.7 Hold Mold for Maintenance ..................................................................................... 21
4.8 Sensor Maintenance ............................................................................................... 21
5.0 KEYBOARD SECURITY ........................................................................................ 22
5.1 Push Button Security............................................................................................... 22
5.2 Eeprom Security...................................................................................................... 22
6.0 DIAGNOSTICS AND TROUBLESHOOTING......................................................... 23
6.1 General.................................................................................................................... 23
6.2 Contact Conductivity Mode Bench Check............................................................... 24
6.3 Toroidal Conductivity Mode Bench Check .............................................................. 25
6.4 Systematic Troubleshooting.................................................................................... 25
7.0 THEORY OF OPERATION ..................................................................................... 29
7.1 General.................................................................................................................... 29
7.2 Overview ................................................................................................................. 29
8.0 DESCRIPTION AND SPECIFICATIONS................................................................ 31
8.1 Features .................................................................................................................. 31
8.2 Functional Specifications ........................................................................................ 32
8.3 Performance Specifications .................................................................................... 32
8.4 Physical Specifications............................................................................................ 33
8.5 Recommended Sensors.......................................................................................... 33
8.6 Ordering Information ............................................................................................... 34
8.7 Accessories ............................................................................................................. 34
9.0 RETURN OF MATERIAL........................................................................................ 36
i
MODEL 2081C/T TABLE OF CONTENTS
ii
TABLE OF CONTENTS CONT’D.
LIST OF FIGURES
Figure No. Title Page
1-1 Model 2081C/T Dimensional/Mounting Instructions
with Power Supply/Loop Signal Wiring ................................................................... 4
1-2 Model 2081C/T Contacting Electrode Wiring.......................................................... 5
1-3 Model 2081C/T Inductive Sensor Wiring ................................................................ 6
1-4 Fault and Security Jumper Locations...................................................................... 7
1-5 Sensor To Transmitter Switch Settings ................................................................... 7
1-6 Explosion Proof System Installation Schematic ..................................................... 8
1-7 Intrinsically Safe (BASEEFA) Wiring for Model 2081C ........................................... 9
1-8 Intrinsically Safe (BASEEFA) Wiring for Model 2081T .......................................... 11
1-9 Intrinsically Safe (CSA) Wiring for Model 2081C/T ................................................ 13
1-10 Intrinsically Safe (FM) Wiring for Model 2081C/T .................................................. 15
3-1 Menu Description .................................................................................................... 17
5-1 Jumper Locations.................................................................................................... 22
6-1 Bench Check (Contact Sensor) .............................................................................. 26
6-2 Bench Check (Toroidal Sensor) .............................................................................. 26
6-3 Troubleshooting Flow Chart .................................................................................... 28
7-1 Block Diagram......................................................................................................... 29
8-1 Load/Power Supply Requirements.......................................................................... 32
8-2 Model 275 HART
®
Communication Menu Tree-Model 2081C/T............................ 35
LIST OF TABLES
Table No. Title Page
3-1 Sensors used with Model 2081C/T conductivity..................................................... 18
3-2 Measurement Ranges............................................................................................. 18
6-1 Diagnostic Messages.............................................................................................. 23
6-2 Resistance Values (Temperature) ........................................................................... 24
6-3 Resistance Values (Conductivity)............................................................................ 24
6-4 Quick Troubleshooting Guide.................................................................................. 27
8-1 Local Interference Functions................................................................................... 32
8-2 Measurement Ranges............................................................................................. 32
MODEL 2081C/T TABLE OF CONTENTS
SECTION 1.0 INSTALLATION
1.1 UNPACKING AND INSPECTION. Inspect the
transmitter for shipping damage. If damaged, notify the carrier immediately.
Confirm that all items shown on the packing list are present. Notify Rosemount Analytical if any items are missing.
If the transmitter appears to be in satisfactory condition proceed to Section 1.2.
NOTE
Save the original packing cartons and materi­als as most carriers require proof of damage due to mishandling.
1.2 MECHANICAL INSTALLATION.
1.2.1 General. The transmitter may be installed in
harsh environments. However, it should be installed in an area where sources of extreme temperature fluctu­ation, vibration and shock are at a minimum or absent. Installation site should:
1. permit the use of the standard cable lengths (unless a junction box is used),
2. be easily accessed by operating and maintenance personnel, and
3. be at least 12 inches (.3 m) from sources of high voltage.
NOTE
Before installing the transmitter, it should be determined whether the Model 275 will be used with the 2081C/T.
1.2.2 Mounting. The transmitter may be mounted on a
flat surface using the two threaded mounting holes located on the bottom of the transmitter or through the use of an optional 2-inch pipe/wall mounting bracket, Code 07 (Figure 1-1).
NOTE
The meter may be installed in 90-degree increments for easy viewing. Remove the four screws holding the meter in place and change the meter to the desired angle. Plug in the display and tighten the four screws.
1.3 WIRING. The transmitter is equipped with two 1/2-
inch NPT conduit openings, one on each side of the housing. One is for the power supply/signal wiring and the other is for the sensor wiring.
The use of weathertight cable glands or conduit is recommended to prevent moisture from entering the housing. Conduit should be positioned to prevent con­densation from draining into the housing. Conduit con­nections on the transmitter housing should be plugged and sealed to avoid moisture accumulation inside the terminal section of the housing.
CAUTION
If the connections are not sealed, mount trans­mitter with the electrical housing positioned downward for drainage. Wiring should be installed with a drip loop. The bottom of the drip loop should be lower than the conduit con­nections or the transmitter housing.
1.3.1 Power Supply/Signal Wiring. It is recommend­ed that the signal wiring be shielded, twisted pairs that are grounded. The best place to ground the loop is at the negative terminal of the power supply. Do not ground the signal loop at more than one point. The transmitter case shall be grounded. The power and signal wiring terminal is TB2 terminals 1 through 4 as shown in either Figures 1-1, 1-2, and 1-3.
Signal or sensor wiring should never be run in the same conduit or open tray as AC power or relay actu­ated signal cables. Keep signal or sensor wiring at least 12 in. (.3 m) from heavy electrical equipment.
NOTE
For best EMI/RFI protection the power sup­ply/signal cable should be shielded and enclosed in an earth grounded, rigid metal conduit. Connect the cable’s outer shield to the ground terminal near TB2, Figure 1-1. The sensor cable should also shielded. The cable’s outer shield shall be connected to the earth ground terminal provided near TB2, Figure 1-
1. If the outer shield is braided an appropriate metal cable gland fitting maybe used to con­nect the braid to earth ground via the instru­ment case.
A new addition to the suite of tests done to ensure CE compliance is IEC 1000-4-5. This is a surge immunity test that simulates overvolt­ages due to switching and lightning transients.
In order to meet the requirements of this test, additional protection must be added to the instrument in the form of a Transient Protector such as the Rosemount Model 470D. This is a 3½-inch tube with ½-inch MNPT threads on both ends. Inside the tube are gas discharge and zener diode devices to limit surges to the transmitter from the current loop. No addition­al protection is needed on the sensor connec­tions.
1
MODEL 2081C/T SECTION 1.0
INSTALLATION
2
1.3.2 Sensor Wiring. The sensor wiring terminals are located on the side opposite to the LCD meter. Remove the housing cover to gain access. Pass the sensor cable through the transmitter’s conduit opening. Connect the sensor wiring to TB1 terminals 1 through 12 as shown on Figures 1-2 and 1-3.
NOTE
If the standard cable length is not sufficient for the planned installation, the use of a junction box with extension cable is strong­ly recommended. Do not exceed 250 feet (76m) total cable length from the sensor to the transmitter.
1.3.3 Matching Transmitter to sensor. Switches S4 and S5 are used to normalize the transmitter for Toroidal or contact sensors. See Figure 1-5.
Switches S4 and S5 should both be set to T when Toroidal sensors are installed, C when Contact sensors are installed.
Switches S1 and S2 are used to select the sensitivity range of the LCD. To obtain desired switch settings, multi­ply the cell constant of the probe being used by the maxi­mum low range value. If the working range of the process is below this calculated value then set both switch to Lo. If the working range of the process is above this calculat­ed value then set both switch to Hi (see table below).
Sample: Model 142 sensor: .2 (cell constant) x 4,000µS (maximum low range value)= 800µS (calcu­lated value).
1.4 FAULT MODE AND SECURITY JUMPER
1.4.1 General This section describes how to set the
jumpers on the CPU board for the following (See Figure 1-4).
1. Output value in fault (alarm) mode
2. Security - Write Disable
3. Security - Push Buttons
CAUTION
The circuit board is electrostatically sensitive. Be sure to observe handling precautions for static-sensitive components.
1.4.2 Fault Mode Output. The default output of the transmitter during a fault condition is determined by the position of the default current output jumper JP3 (refer to Figure 1-4). The output can be set to default either:
– below 4mA - JP3 with jumper.
– above 20mA - JP3 without jumper (factory setting).
Store jumper on one post to prevent misplacing it. See Section 8.0 for corrective actions for fault messages.
1.4.3 Security. Also explained in Section 6.0. These jumpers are located on the CPU board. See Figure 1-4 for location.
– Disable 2081 push buttons - JP4 without jumper.
(Store jumper on one post to prevent misplacing it).
– Enable 2081 push buttons - JP4 with jumper (fac-
tory setting).
– Disable all changes to configuration - JP1. Jumper
in place disables any (EEPROM or push button) changes.
– Allow changes to configuration - JP1 Jumper on
one post only will enable configuration changes.
Mnemonic Fault
SLP – FAIL - 2 point calibration error AdC – FAIL - Transmitter electronics failure
(A/D converter)
rngE LOOP - Conductivity value outside 4-20 mA
range points
tEnP – Lo - Temperature too low or RTD shorted tEnP – Hi - Temperature too high or RTD open
Type Maximum Switch
Sensor Range Setting
Toroidal Below 400,000µS x Cell Constant Both “Lo”
Above 400,000µS x Cell Constant Both “Hi”
Contacting Below 4,000µS x Cell Constant Both “Lo”
Above 4,000µS x Cell Constant Both “Hi”
MODEL 2081C/T SECTION 1.0
INSTALLATION
3
1.5 CALIBRATION AND SET-UP
Instructions for output current ranging, security set ups, and calibration of sensor temperature, cell constant etc, are contained in Sections 2.0 through 5.0.
1.6 HAZARDOUS LOCATIONS — EXPLOSION PROOF INSTALLATIONS
In order to maintain the explosion proof rating for installed transmitter, the following conditions must be met.
1. Code 67 must be specified when ordering F.M. (Factory Mutual) units.
2. Explosion proof installation must be in accordance with Drawing Number 1400160 (see Figure 1-6).
3. The transmitter enclosure covers must be on hand tight and the threads must not be damaged.
NOTE
These covers seat on rings which serve to provide a dust proof enclosure for Class II and Class III installations.
4. Explosion proof “Y” fittings must be properly installed and plugged with a sealing compound to prevent explosive gases from entering the trans­mitter. CSA has determined that the transmitter housing is “Factory Sealed”. Installation of “Y” fit­tings and the use of sealing compound is not required for CSA approved Explosion Proof instal­lations.
NOTE
Do not install sealing compound until all field wiring is completed.
CAUTION
Sealing compound must be installed prior to applying power to the transmitter.
5. If one of the conduit connections on the housing is not used, it must be closed with a threaded metal plug with at least five threads engaged.
1.7 HAZARDOUS LOCATIONS —
INTRINSICALLY SAFE INSTALLATIONS
See Figure 1-7 for Model 2081C intrinsically safe instal­lation (BASEEFA). See Figure 1-8 for Model 2081T intrinsically safe installation (BASEEFA). See Figure 1­9 for Model 2081C and 2081T intrinsically safe installa­tion (CSA). See Figure 1-10 for Model 2081C and 2081T intrinsically safe installation (FM).
MODEL 2081C/T SECTION 1.0
INSTALLATION
4
FIGURE 1-1. Model 2081C/T Dimensional/Mounting Instructions with Power Supply/Loop Signal Wiring
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40208120 A
MODEL 2081C/T SECTION 1.0
INSTALLATION
5
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
FIGURE 1-2. Model 2081C Contacting Electrode Wiring
PN 9200275
CABLE PN 9200275 - UNPREPPED CABLE PN 23747-00 - PREPPED
CE approved cable for Sensor Models 140, 141, 142, 150, 400, 401, 402, 403, & 404 using cable PN 9200275 or 23747-00
WIRE FUNCTIONS
DWG. NO. REV.
40208119 A
MODEL 2081C/T SECTION 1.0
INSTALLATION
6
FIGURE 1-3. Model 2081T Inductive Sensor Wiring
DWG. NO. REV.
40208106 C
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
MODEL 2081C/T SECTION 1.0
INSTALLATION
7
FIGURE 1-5. Sensor to Transmitter Switch Settings
FIGURE 1-4. Fault and Security Jumper Locations
DWG. NO. REV.
40208110 B
DWG. NO. REV.
40208109 A
MODEL 2081C/T SECTION 1.0
INSTALLATION
8
FIGURE 1-6. Explosion Proof System Installation Schematic (FM)
DWG. NO. REV.
1400160 A
MODEL 2081C/T SECTION 1.0
INSTALLATION
9
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-7. Intrinsically Safe (BASEEFA) Wiring for Model 2081C
Part 1 of 2
DWG. NO. REV.
1700294 C
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-7. Intrinsically Safe (BASEEFA) Wiring for Model 2081C
Part 2 of 2
DWG. NO. REV.
1700294 C
10
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-8. Intrinsically Safe (BASEEFA) Wiring for Model 2081T
Part 1 of 2
DWG. NO. REV.
1700291 D
11
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-8. Intrinsically Safe (BASEEFA) Wiring for Model 2081T
Part 2 of 2
DWG. NO. REV.
1700291 D
12
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-9. Intrinsically Safe (CSA) Wiring for Model 2081C/T
Part 1 of 2
DWG. NO. REV.
1400169 B
13
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-9. Intrinsically Safe (CSA) Wiring for Model 2081C/T
Part 2 of 2
DWG. NO. REV.
1400169 B
14
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-10. Intrinsically Safe (FM) Wiring for Model 2081C/T
Part 1 of 2
DWG. NO. REV.
1400166 C
15
MODEL 2081C/T SECTION 1.0
INSTALLATION
FIGURE 1-10. Intrinsically Safe (FM) Wiring for Model 2081C/T
Part 2 of 2
DWG. NO. REV.
1400166 C
16
17
SECTION 2.0 DESCRIPTION OF CONTROLS
Figure 3-1. Menu Description
2.2.2 Display Flags and Hold Mode. The transmitter
displays a flag on the LCD corresponding to the appro­priate units.
When the transmitter is in hold output mode a flashing flag will appear on the middle right hand side of the dis­play and HoLd will flash periodically. The output will remain at the last process value. To return the trans­mitter to normal operation, scroll to hold again and release both buttons.
2.3 VALUE ADJUSTMENT. Selection of a menu item
that has a user-adjustable value will cause a numeric display with the right-most digit flashing.
Depressing and holding push button #1 will cause
the number flashing to increment upwards, looping back to zero.
Depressing push button #2 will shift the flashing
cursor to the next decade.
Decimal position is shifted by pressing button #2 when
the left most digit is flashing under the Std menu. Decimal will start flashing. Change location by pressing button #1.
Select millisiemens/microsiemens flag by going
through decimal position sequence and then pressing button #2. Flag will start flashing. Press button #1 to position the flag on the proper choice. Press button #2 once more and sensor will return to the starting point to change values. Press both buttons in momentarily to enter dis­played value into memory.
NOTE
The autoranging feature on the 2081C/T will make decimal point setting and engr. unit settings unnecessary.
MAIN MENU
dISP – – – – – – – – – –
Ê
Std
Standardize Conductivity
hold Initiate and remove hold output mode
-0- Zero sensor loop LoC 4 mA range point
HiC 20 mA range point CELL Cell constant tAdJ Standardize temperature, °C
Atc Automatic temperature compensation
SLP Electrode slope (mV/pH)
tP1 Calibration point one
tP2 Calibration point two
DISPLAY SUBMENU
Cond Conductivity value
fct Cell factor
curr Output in mA
DIo
Output in % of full scale
DC Temperature, °C DF Temperature, °F
Cin Absolute conductivity
2.1 GENERAL. Nearly all functions of the transmitter are accessed through the dual push buttons. The transmitter uses no potentiometer.
2.2 MENU SELECTION. The dual push buttons are located on the side of the transmitter. Press and hold both buttons to display the transmitter menu items. See Figure 3-1. The display will show each item for about one second then scroll to the next item. It will continue to loop through the items until one is select­ed. To select an item, release both push buttons when the desired item is displayed.
NOTE
When no push button is pressed for a peri­od of 60 seconds, the transmitter defaults to reading conductivity. If the push buttons are accidentally released and this would upset the process (e.g. tP2), escape by waiting 60 seconds for the transmitter to default to the conductivity reading.
2.2.1 Display Submenu. The display submenu (dISP) is used to access secondary process values that cannot be changed.
To enter the submenu, scroll thru the main menu. When dISP is on the LCD release both buttons. The display will read Cond. To see the conductivity value in this submenu, press both push buttons momentarily.
To scroll thru the submenus, press both buttons and HOLD. Submenu mnemonics will start scrolling. Release both but­tons when mnemonic desired is displayed. Screen will dis­play to value of submenu item selected. To view another value in the submenu press both buttons momentarily. Screen will revert to Cond, the entry screen for the sub­menus. – Repeat previous submenu actions.
To exit the submenu, press both push buttons and hold. Display will start scrolling thru the main menu items or take no action for 60 seconds and the display will default to the conductivity reading.
Press and hold both buttons auto
scrolls through main menu and submenu.
Release buttons to select item.
Push button #1: Press and hold
auto scrolls digits.
Push button #2: Press and hold
auto shifts decades.
Enter a value: Depress both but-
tons briefly.
MODEL 2081C/T SECTION 2.0
DESCRIPTION OF CONTROLS
SECTION 3.0 RANGE CONFIGURATION
3.1 GENERAL. The conductivity range of the transmit-
ter depends upon the sensor used. Table 3-1 can be used to determine ranges available with the sensor used. As shipped from the factory, 4 mA represents 0 siemens and 20 mA represents 2 siemens. The value displayed for these output currents are user selec­table. To change the displayed readings in siemens for 4mA and 20mA proceed:
A. Output Zero (4 mA) LoC
1. Depress and hold both push buttons. The display will begin to auto scroll.
2. Release both keys when LoC is displayed. The present 4 mA conductivity value in memory will be displayed with the last digit flashing.
3. Depress and hold push button #1 (scroll) and #2 (shift) as needed to display the desired conductiv­ity value.
4. Enter the value into memory by depressing both push buttons briefly. The display will return to dis­playing the present conductivity value.
B. Full Scale (20 mA) HiC
1. Depress and hold both push buttons. The display will begin to auto scroll.
2. Release both keys when HiC is displayed. The present 20 mA conductivity value in memory will be displayed with the last digit flashing.
3. Depress and hold push button #1 (scroll) and #2 (shift) as needed to display the desired conductiv­ity value.
4. Enter the value into memory by depressing both push buttons briefly. The display will return to dis­playing the present conductivity value.
NOTE
For a reverse output, enter the higher value for LoC, and the lower value for HiC.
TABLE 3-1
SENSORS USED WITH MODEL 2081 CONDUCTIVITY
Conductivity Sensor 142 142 140 140, 141
Model Number 400 150, 400 141 150 150, 400 401 222 225 226 228
Cell Constant** 0.01 0.1 0.2 0.5 1.0 10.0 * 3.0 1.0 3.0
Min. Range 1 2 4 10 20 200 500 250 50 250
Max. Range 200 2,000 4,000 10,000 20,000 200,000
2,000,000 2,000,000 1,000,000 2,000,000
FULL SCALE MICROSIEMENS/cm
* 1 in. diameter = 6.0, 2 in. diameter = 4.0 (Typical) ** Typical
MODEL 2081C/T SECTION 3.0
RANGE CONFIGURATION
18
TABLE 3-2. Measurement Ranges
Type Maximum Switch
Sensor Range Setting
Toroidal Below 400,000µS x Cell Constant Both “Lo”
Above 400,000µS x Cell Constant Both “Hi”
Contacting Below 4,000µS x Cell Constant Both “Lo”
Above 4,000µS x Cell Constant Both “Hi”
19
SECTION 4.0 START-UP AND CALIBRATION
4.1 GENERAL. A sensor must be wired to the trans-
mitter for calibration. See the appropriate sensor man­ual for additional instructions relating specifically to the sensor.
4.2 TEMPERATURE CALIBRATION. For the most accurate temperature compensation, the temperature reading may need adjusting. The following steps should be performed with the sensor in a grab sample or process of known temperature.
NOTE
Calibrate at or near the process tempera­ture for greatest accuracy.
1. Scroll through the menu items. Press and hold both push buttons simultaneously.
2. When tAdJ, is displayed, release both buttons. The display will show the present temperature value in degrees Celsius with the last digit flashing. The tenths digit will alternate between a value and °C.
3. Compare the displayed reading with a calibrated temperature reading device. If the reading requires adjustment, proceed to step 4. Other­wise, depress both push buttons briefly to accept the displayed value.
4. Depress and hold push buttons #1 (scroll) and #2 (shift) as needed to display the desired tempera­ture value.
5. Enter the correct value into memory by depressing both push buttons briefly. The display will return to displaying the present conductivity value.
4.3 ENTERING THE CELL CONSTANT. The first time
the analyzer is calibrated and any time there is a sen­sor change, the sensor cell constant must be entered into memory. Entering a cell constant into memory will reset the cell factor Fct to 1.0 and will initiate the ana­lyzer. The cell factor gives an indication of sensor foul­ing or coating. Refer to Section 6.0.
NOTE
The cell constant (K) will be found on the sensor label (i.e. K = 3.04, K = 1.00) locat­ed on the cable. For Models 140, 141, and
142. The cell constant is shown on the junction box label. (For the Model 222 the label is on spindle).
1. Scroll through the menu items. Press and hold both push buttons simultaneously.
2. When CELL, is displayed release both buttons. The display will show the cell constant in memory with the right most digit flashing.
3. Press and hold buttons #1 (scroll) and #2 (shift) as needed to display the correct cell constant.
4. Enter the correct value into memory by briefly pressing both push buttons simultaneously. The display will return to displaying the present con­ductivity value.
4.4 ZEROING THE SYSTEM. The transmitter must
calibrate the zero point before the sensor is placed into the process solution.
CAUTION
DO NOT PLACE THE SENSOR IN THE PROCESS. The sensor must be placed into solution only after performing a system zero.
1. Assure that the sensor is properly wired and out of process (in air).
2. Scroll through the menu items. Press and hold both push buttons simultaneously.
3. When -0-, is displayed, release both buttons. The transmitter will calculate the loop zero point. This takes about 10 to 20 seconds. -0- will flash on the display while the transmitter is calibrating. The dis­play will return to displaying conductivity when the zero calibration is complete.
4. Place the sensor in solution and proceed with the system calibration.
4.5 INITIAL LOOP CALIBRATION. Please read the
entire calibration section before proceeding to deter­mine the best calibration procedure. Also, please check the appropriate sensor manual’s calibration sec­tion for any specific instructions.
MODEL 2081C/T SECTION 4.0
START-UP AND CALIBRATION
20
4.5.1 Two Point Calibration - Process temperature slope not known. Recommended procedure for the ini­tial calibration if the process temperature slope is not known. If any of the steps below are impossible or impractical to perform, refer to alternate Section 4.5.2.
1. Obtain a grab sample of the process to be meas­ured.
2. Determine the sample’s conductivity using a cali­brated instrument or portable analyzer. The instru­ment must be able to reference the conductivity to 25°C, or the solution must be measured at 25°C. Record the reading.
NOTE
The transmitter must be in hold or off line.
3. Immerse the measuring portion of the sensor in process solution. (Model 222 users should refer to the sensor’s manual for special instructions). The sensor body must be held away from the bottom and sides of the sample’s container at a distance at lease the diameter of the sensor. Shake the sensor to ensure that no air bubbles are present.
4. Adjust the sample’s temperature to either the nor­mal high or normal low process temperature. (To raise the sample’s temperature, a hot plate with stirrer may be used. To lower the sample temper­ature, place the grab sample’s container in an ice bath or let it slowly cool down).
5. Allow the sensor to acclimate to the sample temper­ature. (The temperature reading should be stable).
6. Scroll through the menu items. Press and hold both push buttons simultaneously.
7. When tP1, is displayed, release both buttons. The display will show the conductivity value in memo­ry with the right most digit flashing.
8. Depress and hold buttons #1 (scroll) and #2 (shift) as needed to display the grab sample’s conductiv­ity value at 25°C as noted in step 2.
9. Enter the correct value into memory by briefly depressing both push buttons simultaneously. The display will return to displaying the present con­ductivity value.
10. Adjust the sample’s temperature to the process’s other normal temperature extreme.
11. Allow the sensor to acclimate to the sample tempera­ture. (The temperature reading should be stable).
12. Scroll through the menu items. Press and hold both push buttons simultaneously.
13. When tP2 is displayed, release both buttons. The display will show the conductivity value in memo­ry with the right most digit flashing.
14. Depress and hold buttons #1 (scroll) and #2 (shift) as needed to display the grab sample’s conductiv­ity value at 25°C as noted in step 2.
15. Enter the correct value into memory by briefly depressing both push buttons simultaneously. The display will return to displaying the present con­ductivity value.
The analyzer then calculates the true cell constant and the processes temperature slope, then returns to read­ing conductivity.
The temperature slope of the process can now be read by selecting SLP from the menu.
The sensor may now be installed in the process.
NOTE
If the 2081 was placed in hold, be sure to remove it from hold after the sensor is on line.
4.5.2. Single Point Calibration - Process Temp-
erature Slope Known. This is the recommended proce­dure for the initial calibration if the temperature slope of the process is known. If you do not know the exact temperature slope value, but wish to approximate it, refer to the following guide.
Acids: 1.0 to 1.6%/°C Bases: 1.8 to 2.2%/°C Salts: 2.2 to 3.0%/°C Water: 2.0%/°C
1. Scroll through the menu items. Press and hold both push buttons simultaneously.
2. Release both buttons simultaneously when SLP is displayed. The slope value in memory will display with the right most digit flashing.
3. Depress and hold buttons #1 (scroll) and #2 (shift) as needed to display the desired temperature slope value.
4. Enter the correct value into memory by briefly depressing both push buttons simultaneously. The display will return to displaying the present con­ductivity value.
MODEL 2081C/T SECTION 4.0
START-UP AND CALIBRATION
6. Depress and hold buttons #1 (scroll) and #2 (shift) as needed to display the conductivity value noted in step 3.
Depress both push buttons simultaneously to accept this value.
The transmitter will then calculate the proper cell factor.
7. Depress both push buttons simultaneously and release when diSP is displayed.
Depress both push buttons again and release when Fct is displayed.
Note this value to determine a sensor mainte­nance schedule.
4.7 HOLD MODE FOR MAINTENANCE. Before per-
forming maintenance of the sensor, or buffer checks, the transmitter should be placed in the HoLd mode. This mode of operation maintains the output current at the last process value. To initiate HoLd mode:
1. Depress and hold both push buttons simultaneously until HoLd is displayed on the screen.
2. Release both buttons.
a. The HoLd flag will flash, and
b. The display will show HoLd every two seconds
to confirm the hold status.
To place the transmitter back into normal operation, (remove from HoLd mode):
1. Depress and hold both push buttons simultan­eously until HoLd is displayed on the screen.
2. Release both buttons.
a. The HOLD, flag will stop flashing.
b. The display will stop showing HoLd every two
seconds.
c. The transmitter output current will return to
normal operation.
4.8 SENSOR MAINTENANCE. Always calibrate after
cleaning or repair of the conductivity sensor.
Always place transmitter into the HoLd mode of opera­tion while performing any maintenance to the sensor to avoid loss of process control.
Always return transmitter to normal operation after installing the sensor back into the process.
5. Obtain a grab sample of the process to be meas­ured.
NOTE
Be sure the transmitter is in Hold or Off-Line.
6. Determine the sample’s conductivity using a cali­brated instrument or a portable analyzer. The instrument must be able to reference the conduc­tivity to 25°C, or the solution must be at 25°C. Write down the reading. Ensure that the analyzer is in hold.
7. With the sensor in sample, depress and hold both push buttons simultaneously to enter the transmit­ter menu. The transmitter will auto scroll through the menu items. Release both push buttons when Std is displayed. The conductivity value will dis­play with the right most digit flashing.
8. Depress and hold buttons #1 (scroll) and #2 (shift) as needed to display the conductivity value noted in step 5. Depress both push buttons simultaneously to accept this value. The transmitter will then calcu­late the proper cell constant.
NOTE
Return transmitter to normal operation from Hold.
4.6 ROUTINE STANDARDIZATION. The sensor should
be standardized in-line routinely if it is suspected that the process might degrade or coat the sensor. After the ini­tial calibration, each time a standardization is performed the cell factor Fct is changed. Refer to Section 4.3 for a description of the cell factor.
To perform a standardization do the following:
1. Take a grab sample which is as close to the sensor as possible. Record transmitter reading.
2. Determine the sample’s conductivity using a cali­brated instrument or a portable analyzer. The instrument must be able to reference the conduc­tivity to 25°C, or the solution must be at 25°C. Record the reading.
3. Note any change in the 2081’s present reading from first reading recorded in Step 1. Add this change (either + or –) to the calibrated instruments value. Record this figure.
4. Scroll through the menu items. Press and hold both push buttons simultaneously.
5. Release both push buttons when Std is displayed. The conductivity value will display with the right most digit flashing.
21
MODEL 2081C/T SECTION 4.0
START-UP AND CALIBRATION
SECTION 5.0 KEYBOARD SECURITY
5.1 PUSH BUTTON SECURITY. This feature disables
the dual push buttons on the transmitter to prevent accidental or unauthorized changes to the calibration and configuration. HART communication is not affected by this change. Perform the following steps to disable the push buttons. See Figure 5-1 for jumper locations on CPU board.
CAUTION
The circuit board is electrostatically sensitive. Be sure to observe handling precautions for static-sensitive components.
1. Remove the display side cover.
2. For easy access, the electronics assembly may be removed by pulling the assembly straight out.
3. Remove jumper JP4 from the CPU board. To pre­vent misplacing the jumper, slide it back onto one post only.
4. Replace the display side cover.
CAUTION
A minimum of seven cover threads must be engaged in order for the transmitter to meet explosion-proof requirements.
Replace the jumper to enable push button operation.
5.2 FAULT MODE SELECTION This feature allows the user to select the desired mA output during a system disabling fault condition.
CAUTION
The circuit board is electrostatically sensitive. Be sure to observe handling precautions for static-sensitive components.
1. Remove the display side cover.
2. For easy access, the electronics assembly may be removed by pulling the assembly straight out.
3. JP3's Fault Mode Selection
Jumper on both posts: below 4.0 mA
Jumper on one post: above 20.0 mA
4. Replace the display side cover
CAUTION
A minimum of seven cover threads must be engaged in order for the transmitter to meet explosion-proof requirements.
FIGURE 5-1. Jumper Locations
NOTE:
JP1 and JP3 shown with jumpers in open (store) position.
DWG. NO. REV.
40208109 A
MODEL 2081C/T SECTION 5.0
KEYBOARD SECURITY
22
23
SECTION 6.0 DIAGNOSTICS AND TROUBLESHOOTING
6.1 GENERAL: The Model 2081C/T Transmitter auto-
matically searches for fault conditions that would cause an error in the measurement. If such a condition exists, the 2081C/T transmitter will flash a diagnostic message. If more that one fault exists, the display will sequence through the diagnostic messages.This will continue until the cause of the fault has been correct­ed.
Troubleshooting is easy as 1,2,3. . .
Step 1 Look for a diagnostic fault message on the display
to help pinpoint the problem. Refer to Table 6-1 for an explanation of the message and a list of the pos­sible problems that might have triggered it.
Step 2 Refer to the Quick Troubleshooting Guide,
Table 6-4 for common loop problems and the recommended actions to resolve them.
Step 3 Follow the step-by-step troubleshooting
approach offered in Section 6.2 to diagnose less common or more complex problems.
In addition, a theory of operation explanation is offered in section 7.0 for those that may want a more detailed understanding of the instrument.
NOTE
During fault conditions the output of the transmitter is determined by the placement of Jumper JP3 on the CPU board.
1. JP3 in place (shorted across both pins) Output driven below 4 mA.
2. JP3 on one pin only - Output driven above 20 mA.
6.1.1 Diagnostic Messages. Table 6-1 lists the fault
displays, describes the meaning of each and lists some appropriate corrective actions for each.
TABLE 6-1. Diagnostic Messages
Display Description Corrective Action
Adc - FaiL Transmitter electronics failure Turn power off, wait one minute, and power on.
Replace stack.
SnSr - rngE Conductivity value outside 4-20 Check Sensor cell constant and measurement range.
mA range points. Return conductivity to within range, or replace sensor.
SnSr - FaiL Sensor failure or A-D failure Replace sensor. Replace stack. tEnP - Lo Temperature too low or RTD shorted
Check wiring, calibrate temperature, bring sensor within temperature specifications, replace stack, replace sensor. Note: if in manual temp mode, RTD or appropriate resistor must be in place. See Table 6-2.
SLP - FaiL Invalid temp. slope calculated. Turn power off and on.
Did not accept entry. Repeat temp. slope calibration procedure. (Must be performed at 2 different process temperatures.)
-0- FaiL Too much resistance is sensed to Check wiring. zero (over 1Meg). Did not accept entry.
Verify sensor is in air during sensor 0: Not partially submerged.
Std - FaiL Entered conductivity value is unacceptable
Conductivity calibration failed: verify conductivity standard or
(+/- 150 C). grab sample.
Temp. calibration failed: verify temp at the sensor: allow 20 minutes for stabilization.
LooP - rngE Conductivity value is outside of the present Bring conductivity back into the programmed range.
Lo and Hi range setpoints. Re-adjust Lo and Hi range setpoints.
Flashes You have attempted to enter a value that Verify the attempted procedure’s instructions.
bAd exceeds the allowable software. Check standard or instrumentation used.
FAIL The transmitter recognizes that its Check wiring
electronics have malfunctioned Turn power off and on.
Replace electronic stack.
NOTE: Some of the fault codes can be cleared by cycling the power off and on.
MODEL 2081C/T SECTION 6.0
DIAGNOSTICS AND TROUBLESHOOTING
24
6.1.2 Temperature Compensation. Table 6-2 (below) is a reference of RTD resistance values at various tem­peratures. These are used for test and evaluation of the sensor.
NOTE
Ohmic values are read across the RTD ele­ment and are based on the manufacturer’s stated values (+/- 1%). Allow enough time for the RTD element to stabilize to the sur­rounding temperature.
Table 6-2. Resistance Values (Temperature)
6.2 CONTACT CONDUCTIVITY MODE BENCH CHECK: Sensor simulation may be used to check the
operation of the model 2081C/T. See Figure 6-1 for wiring diagram to perform analyzer check.
CAUTION
Do not use over 55 volts to check the loop. Damage to the transmitter may result.
1. Set up transmitter as shown in Figure 6-1.
2. Simulate a conductivity value by choosing a typi­cal resistance value from Table 6-3 and entering this resistance value into the decade box.
TABLE 6-3. Resistance Values (Conductivity)
NOTE
Chart assumes a cell constant of 1.0. Therefore, Resistance = 1/Conductivity.
3. Apply power to the Loop.
4. Set the CELL constant to 1.00.
5. Set the atc to USEr. Enter the user value of 25° C. (This eliminates temperature condition.)
6. Remove one test lead from the resistance decade box to simulate 0.00 Microsiemens.
7. Select the zero function -0- from the Level 1 menu.
8. The display will flash -0- while the unit zeros.
9. Reconnect the test lead to the decade box to sim­ulate full-scale conductivity.
10. If calibrated, the display should read close to the full-scale conductivity value.
11. Scroll to Std and enter the desired full-scale con­ductivity value
12. Check half-scale reading by doubling the resist­ance value entered into the decade box.
13. The transmitter conductivity circuitry is operating correctly if it will zero, read the full-scale and half­scale conductivity values.
Simulate Full Scale
Conductivity Of:
20 microsiemens 200 microsiemens 2000 microsiemens 20,000 microsiemens
Resistance Value
To Use:
50,000 ohms 5,000 ohms 500 ohms 50 ohms
Temperature Resistance
0
0
C 100.00 Ohms 100C 103.90 Ohms 200C 107.90 Ohms 300C 111.67 Ohms 400C 115.54 Ohms 500C 119.40 Ohms 600C 123.24 Ohms 700C 127.07 Ohms 800C 130.89 Ohms 900C 134.70 Ohms
1000C 138.50 Ohms
MODEL 2081C/T SECTION 6.0
DIAGNOSTICS AND TROUBLESHOOTING
25
6.3 TOROIDAL CONDUCTIVITY MODE BENCH CHECK: Sensor simulation may be used to check the
operation of the Model 2081C/T. See Figure 6-2 for wiring diagram to perform analyzer check.
CAUTION
Do not use over 55 volts to check the loop. Damage to the transmitter may result.
1. Set up transmitter as shown in Figure 6-2.
2. Simulate a full scale conductivity value by choos­ing a typical resistance value from Table 6-3 and entering this resistance value into the decade box.
TABLE 6-3. Resistance Values (Conductivity)
NOTE
Chart assumes a cell constant of 1.0. Therefore, Resistance = 1/Conductivity.
3. Apply power to the Loop.
4. Set the CELL constant to 1.00.
5. Set the atc to USEr. Enter the user value of 25° C.
(This eliminates temperature correction.)
6. Remove one test lead from the resistance decade box to simulate 0.00 Microsiemens.
7. Select the zero function -0- from the Level 1 menu.
8. The display will Flash -0- while the unit zeros.
9. Reconnect the test lead to the decade box to sim­ulate full-scale conductivity.
10. If calibrated, the display should read close to the full-scale conductivity value.
11. Scroll to Std and enter the desired full-scale con­ductivity value.
12. Check half-scale reading by doubling the resist­ance value entered into the decade box.
13. The transmitter conductivity circuitry and the sen­sor are operating correctly if it will zero, read the full-scale and half-scale conductivity value. If not, refer to the appropriate sensor instruction manual to perform tests on the toroidal sensor. If the sen­sor tests O.K, then replace the transmitters elec­tronic stack.
6.4 SYSTEMATIC TROUBLESHOOTING.
Not all problems that you encounter will be typical. If you are unable to resolve your problem using the Quick Troubleshooting Guide, Table 6-4, then try the step-by-step approach offered in Figure 6-3.
Simulate Full Scale
Conductivity Of:
20 microsiemens 200 microsiemens 2000 microsiemens 20,000 microsiemens
Resistance Value
To Use:
50,000 ohms 5,000 ohms 500 ohms 50 ohms
MODEL 2081C/T SECTION 6.0
DIAGNOSTICS AND TROUBLESHOOTING
26
FIGURE 6-1. Bench Check (Contact Sensor)
FIGURE 6-2. Bench Check (Toroidal Sensor)
DWG. NO. REV.
40208113 B
DWG. NO. REV.
40208113 B
MODEL 2081C/T SECTION 6.0
DIAGNOSTICS AND TROUBLESHOOTING
27
SYMPTOM PROBLEM ACTION
fct below 0.5 or above 2.0. 1. Old or coated sensor. 1. Clean or replace glass electrode. Actual range determined by user. 2.
A preceding standardization was incorrect.
Value locks up (no change of 1. Incorrect wiring. 1. Verify wiring. reading in different standards). 2. Open sensor. 2. Perform sensor checks.
3. Replace sensor.
2081C/T value not the same as 1. Grab sample incorrect. 1. Re-evaluate sample technique and
grab sample of process. equipment.
2. Unclear what is correct. 2. Test with standard solution.
3. 2081C/T out of calibration. 3.
Recalibrate per start-up and calibration sections.
Fault Code TEnP-Lo, or 1. Incorrect wiring. 1. Check wiring between the sensor and the
TEnP-Hi transmitter.
2. Open or shorted RTD. 2. Replace RTD or sensor.
3. Manual temperature 3. Appropriate resistor must be in place ­mode. TB1-6 to 8. See Table 6-2.
Zero Conductivity reading. 1. Sensor wired incorrectly. 1. Repair wire connection.
2. Solids coating sensor. 2. Clean sensor.
3. Open wire in sensor. 3. Replace sensor.
Fault code rn9E-LOOP. 1. Process value is outside 1. Return process to normal. Check
4-20 mA range points. grab sample.
Fault Code FAIL. 1.Range selection switches 1. Calculate the maximum low range value
S1 and S2 are set and set the switches accordingly. See incorrectly. Section 1.3.3.
2. Defective PCB stack. 2. Reconnect power or if necessary,
replace PCB stack.
No output current. 1. Defective PCB stack. 1. Replace PCB stack.
2. Incorrect wiring. 2. Check for short.
Low output current. 1. Circuit loading with excess- 1. consult output loading limits 2081C/T
sive resistance on output. specifications (1840 ohms max load).
Drifting or Unstable readings. 1. Air bubbles in process 1. Mount sensor to decrease exposure to
affecting reading. air bubbles.
2. Electrical or grounding 2. Verify sensor works in a standard solu­interference. tion outside of the process.
3.
Verify near by vendor pumps are properly
grounded.
4. Verify that sensor shield never touches an Earth ground.
5. Run sensor cable though grounded metal conduit to reduce RFI effects.
Fault code "LOOP" flashing inter- 1. Model 268 or other host 1. Model 268 or other host can return
mittently with Conductivity and has set the output to a transmitter to normal operation. output value. specific value.
Can't calibrate new sensor. 1. Wrong sensor for meas- 1. Verify that cell content is correct for the
urement range. measurement range.
2. Sensor incompatible with 2. Verify sensor has a PT-100 temperature analyzer. compensator.
Transmitter won't 1. Incorrect wiring. 1.
Check wiring and model 268 connections.
communicate. 2. Insufficient load resistance. 2. Verify minimum of 500 Ohm loop
resistance.
3. Electronics failure. 3. Replace electronic stack.
TABLE 6-4. Quick Troubleshooting Guide
MODEL 2081C/T SECTION 6.0
DIAGNOSTICS AND TROUBLESHOOTING
FIGURE 6-3. TROUBLE SHOOTING FLOW CHART
28
Conductivity Measurement
Problem (in the process)
Remove sensor from process and place sensor in air. Zero instru­ment. Refer to Section 4.4.
OK?
OK?
Place sensor in process and
standardize. Refer to Section 4.6
Check wiring again
for short/open
NOTE:
Before starting this procedure make sure that all wiring is correct.
YES
YES
NO
NO
Does problem
still exist?
YES
NO
Check diagnostic
messages
Refer to Table 6-1
Restart
Analyzer
Check for ground
loops and/or improp-
er installation
Remove sensor from process and test in known conductivity solution
OK?
Refer to Sensor man-
ual for sensor checks
Verify earth ground is not connect-
ed to sensor shields
Isolate sensor shield from any ground metal (conduit)
Consult service center
Check T.S. Guide
Table 6-4
Verify transmitter
operation with bench
check procedures
Section 6.2 & 6.3
YES
NO
MODEL 2081C/T SECTION 6.0
DIAGNOSTICS AND TROUBLESHOOTING
29
7.1 GENERAL: Conductivity is the conductance per unit of length. The 2081C/T measures and calculates conductivity using the relation ship I=GV. Where:
G = conductance,
V = the reference voltage, and,
I = the resultant current directly proportional to
conductance.
7.2. OVERVIEW: Refer to Figure 7-1 2081C/T Block Diagram.
Power Supply: Power for the 2081C/T is derived from the +12 to 55 volt DC loop power. This power is dropped to a regulated +9 volts DC and maintained by com­parison with the output of a 2.5 volt op-amp comparator. The 9 volts DC is switched by a 28.8 KHz square wave signal from the Oscillator across the center tapped pri­mary of the drive transformer. This provides the raw 4.5 VAC for the 4.5 Volt DC power supply.
The 28.8 KHz square wave signal from the oscillator also develops an AC at the secondary that is used to drive a toroidal (or contacting) conductivity sensor.
When a contacting conductivity sensor is used this iso­lates the 2081C/T electronics from the process, and the AC signal also prevents polarizing of the fluid sur­rounding a contacting conductivity sensor.
Oscillator: The 1.8432 MHz oscillator output is divided down to provide:
1. a 921.6 KHz CPU clock signal. 2 a 460.8 KHz MODAC clock signal. 3 a 57.6 KHz demodulator sample signal.
4. the 28.8 KHz transformer drive signal.
Input Signals: RTD and sensed cable line voltages appear at an input to the Mux. These voltages are converted to digital values by the A/D circuits, and stored by the CPU. The CPU subtracts the unwanted cable volt­age from the RTD voltage and uses these readings with a reference voltage to calculate the sensor temperature. This value is used to compensate the absolute conductiv­ity using the temperature slope of the liquid, and can be sent to the LCD to be displayed.
1. Sensors: This transmitter can be used with either type of sensor. Contacting or Toroidal inductive).
Power &
28.8 KHz
Drive
Osc.
and
Divider
Preamp
Demodulate & Auto-Zero
Mux
A/D
CPU
ROM
Display
Drivers
DAC
MODAC
Modem
4 to 20
mA
Current
Loop
Summer
HART
Input Filter
+12 to 55 Volts Loop Power & HART Comm
Conductivity Probe Out (Toroidal or contacting
Conductivity Probe In (Toroidal or contacting)
PT 100 RTD
Receive Transformer
RTD source
RTD return
)( )( )(
)( )( )(
Drive Transformer
+ 9 Volts + 4.5 Volts
921.6 KHz (CPU Clock)
460.8 KHz (MODAC Clock)
57.6 KHz (Sample)
28.8 KHz (Drive)
Sense Line
SECTION 7.0 THEORY OF OPERATION
Figure 7-1. Block Diagram
MODEL 2081C/T SECTION 7.0
THEORY OF OPERATION
30
Sensors: This transmitter can be used with either type of sensor. Contacting or toroidal (inductive). Currents developed by the conductivity probe through the Receive Transformer are converted to a voltage and amplified by the preamplifier block. This voltage is demodulated (in the demodulate and auto-zero block) in phase with the transmitter toroid resulting in a DC voltage representing the absolute conductivity.
a. Contact Conductivity Probe. In the contacting sen-
sor, the area and distance between the two electrodes determine a cell constant, the volume of fluid between the electrodes acts as a resistor whose value depends on the conductance of the fluid.
b. Toroidal Conductivity Sensor. Consists of two
independent transformers that have a common turn, the path of the liquid. The transmitter toroid, or driver, induces a voltage in the path of the liquid which is the voltage across the toroid winding divided by the number of turns in the winding. The voltage times the conductance results in a current that is proportional to the conductance. This cur­rent is sensed by the receiver toroid. The current out of this toroid is the current in the liquid loop divided by the number of turns in its winding. The result is I = G*V/(N1*N2), where:
N1 is the number of turns of the transmitter toroid.
N2 is the number of turns of the receiver toroid.
V is the voltage across the receive transformer.
G is the conductance of the liquid loop.
I is the resultant current developed in the receiver
transformer by the receive toroid.
HART communication signals. Super-imposed on the 4 to 20 mA current loop, are passed through the HART Input filter (to reduce noise). This FSK signal is then demodulate by the MODAC ASIC (Application Specific Integrated Circuit) and sent on to the CPU via the MODAC. A FSK signal is an AC signal, whose fre­quency is shifted higher or lower depending on the condition of the digital signal (High or Low), This Frequency Shift Keying is mixed with the DC value of the 4 to 20 mA signal, using Bell 202 protocol. This communication conforms to the Rosemount HART specification and may be used to configure and inter­rogate the transmitter.
CPU: A 68HC11 micro controller computer chip contains the central processing unit (CPU). This chip also contains volatile Random Access Memory (RAM) and nonvolatile Electrically Erasable Programmable Read Only Memory (EEPROM). CPU instructions, however, are stored in an external ROM. The CPU communicates with this external ROM, Mux, Analog to Digital (A/D) converter, the two push buttons on the side of the 2081C/T, and the Model 275 Smart Interface through the MODAC. It also controls the display drivers,
When requested by the CPU the input signals from either the Contacting or Toroidal sensor are zeroed and scaled between user supplied setpoints to deter­mine what the proportional current value of the 4 to 20 mA loop should be. This value is sent to control the Digital to Analog Converter (DAC) in the MODAC ASIC. The MODAC also produces a weighted Pulse Width Modulated output representing digital informa­tion generated by the CPU to communicate with other digital devices on the current loop. Both of these sig­nals are sent to the current loop summer. The summer then mixes this PWM signal with the current loop value to produce the correct 4 to 20 mA output value for the transmitter.
MODEL 2081C/T SECTION 7.0
THEORY OF OPERATION
31
SECTION 8.0 DESCRIPTION AND SPECIFICATIONS
UTILIZES EITHER CONTACTING OR INDUCTIVE CONDUCTIVITY SENSORS to meet a
wide range of application requirements.
FIELD MOUNTED TRANSMITTER ideal for central data processing and control.
MEMBER OF THE ROSEMOUNT SMART FAMILY
®
.
LOCAL, PUSH-BUTTON INTERFACE for convenient calibration and range adjustment.
WEATHERPROOF, CORROSION-RESISTANT ENCLOSURE.
CONTINUOUS DIAGNOSTICS drive the output to a jumper-selectable high or low value in
the event of a failure.
HOLD OUTPUT MODE for manual control during sensor maintenance.
8.1 FEATURES
The Model 2081 C Smart Transmitters, with the appro­priate sensor, are designed to continuously measure contacting, toroidal, or conductivity in industrial and municipal processes. These two-wire transmitters are members of the Rosemount SMART FAMILY of instru­ments, which are designed to communicate with the hand-held Model 275 SMART FAMILY Interface and any other hosts that support the HART®communica­tions protocol. See Figure 8-1. The Model 2081 design permits remote configuration, interrogation, testing, and diagnostics.
The Model 2081 C features an easy-to-use, dual push­button interface on the transmitter housing for local cal­ibration and range adjustment. The LCD display indi­cates the conductivity value as well as temperature, current output value, range values, temperature slope, and fault messages. Unauthorized adjustments may be prevented by a jumper-selectable setting on a circuit board. All functions are microprocessor based.
The isolated 4-20 mA output is continuously expand­able over the measurement range, and may be dis­played in either milliamps or percent of full scale. A hold output function is available for manual control during routine sensor maintenance. While in hold mode the current output signal (refer to Table 1) will remain con­stant at the last current value.
In the event of a failure, the transmitter diagnostic rou­tine will drive the output below 4 mA or above 20 mA (user selectable) in addition to displaying a fault code on the transmitter. This routine alerts the user to errors due to temperature slope, faulty temperature compen­sation element, open wiring and transmitter electronics failure. The transmitter will also go into a fault mode if the conductivity value is outside the output range set­tings.
With a two-point calibration, the transmitter automati­cally calculates the temperature slope and the cell con­stant. On-line standardization is easily accomplished by simply entering the conductivity value of a grab sam­ple into the transmitter. The transmitter uses a Pt100 RTD signal from the sensor to reference the conductiv­ity reading to 25°C.
The Model 2081 C transmitter is designed to be weath­erproof and corrosion resistant to meet NEMA 4X (IP65) standards. An optional pipe or wall mounting bracket is available.
The transmitter communicates via the HART®protocol, which uses an industry standard BELL 202 frequency shift keying (FSK) technique. Communication is accomplished by superimposing a digital signal on top of the 4-20 mA output signal. The Rosemount imple­mentation of this technique allows simultaneous com­munication and output without compromising loop integrity.
The Model 2081 requires an external power supply such as the Model 515 Isolated Power Supply or DC loop power from a distributed control system. The Model 515 Power Supply provides power for up to 10 transmitters. Two transmitters may be wired directly to the power supply. For more than two transmitters, junc­tion boxes are available, each accommodating wiring for a maximum of ten transmitters.
Remote alarms with independently adjustable setpoints and hysteresis are available in the Model 230A Alarm Module. Contacts of the Model 230A may be specified for high/low, high/high, or low/low operation.
MODEL 2081C/T SECTION 8.0
DESCRIPTION AND SPECIFICATIONS
32
8.2 FUNCTIONAL SPECIFICATIONS
Output: Two-wire 4-20 mA with superimposed HART
digital signal.
Power Supply and Load Requirements: See Figure
8-1 below. A minimum loop resistance (load) of 250 ohms and minimum power supply voltage of 18 volts DC is required for digital communication.
Local Indication: Four digit LCD Ambient Temperature: –30 to 70°C (–22 to 158°F) Failure Mode Alarm: The analog signal will be driven
either below 4 mA or above 20 mA (user-selec­table)
Transmitter Security: Security activation (by internal
jumper) prevents changes to the transmitter con­figuration from the local interface.
Relative Humidity : 0-95% Automatic Temperature Compensation: Pt100 RTD
0 to 200°C (32 to 392°F)
Enclosure: Weatherproof and corrosion-resistant,
explosion proof
EMI/RFI: EN-61326
8.3 PERFORMANCE SPECIFICATIONS
Range*: See Table 8-2 Output Scale: Zero Suppression: Up to 90% of full
scale Span: From 10 to 100% of full scale
Accuracy: ±0.5% full scale @ 25°C Resolution: 0.1% full scale, 0.1°C Repeatability: ±0.25% full scale Stability: 1%/ year at 25°C Temperature Effect: 0.02% F.S./°C Vibration Effect: ±1.0% of F.S. per SAMA PMC
31.1
* Not applicable for ultra pure water applications. See the
products data sheets for Model 3081C/81C or Model 1054B LC for these applications.
• Display output in mA/% of full scale
• Hold output mode
• Set 4-20 mA output range
• Display temperature in °C/°F
• Input or display cell constant
• Display cell factor
• Electronic zero
• Auto/manual temperature compensation
• Two point calibration
• Standardize temperature
• Standardize conductivity
• Input or display temperature slope
• Display absolute conductivity
TABLE 8-1. Local Interface Functions
TABLE 8-2. Measurement Ranges
FIGURE 8-1. Load/Power Supply Requirements
MODEL 2081C/T SECTION 8.0
DESCRIPTION AND SPECIFICATIONS
Type Maximum Switch
Sensor Range Setting
Toroidal Below 400,000µS x Cell Constant Both “Lo”
Above 400,000µS x Cell Constant Both “Hi”
Contacting Below 4,000µS x Cell Constant Both “Lo”
Above 4,000µS x Cell Constant Both “Hi”
Resistivity
(-cm)
Conductivity
(µS/cm)
.05k
20,000µS
2,000µS
500µS
200µS
100µS
20µS
10µS
2µS
1µS
.2µS
.1µS
.05µS
.02µS
.5k
2k
5k
10k
50k
100k
500k
1M
5M
10M
20M
50M
Operating Ranges for Various Contacting Cells Constants (not to scale)
1.0 cm
-1
10.0 cm
-1
0.01 cm
-1
0.1 cm
-1
33
8.4 PHYSICAL SPECIFICATIONS
Electrical Connections: 1/2 in. NPT.
Model 275 SMART FAMILY Interface connections permanently fixed to terminal block. The terminal block also has removable connectors for bare wire sensor connections.
Housing: Epoxy-polyester painted over low
coppercast aluminum. BUNA-N O-rings on
cover.
Recommended Cable: Transmitter to power supply
two-wire, 18 AWG, shielded, Belden 8760 or equal (Rosemount Analytical PN 9200001).
Weight/Shipping Weight: 2.18 kg/2.68 kg (4.8 lb/5.9 lb)
Hazardous Area Classification:
FM Certification:
Type Class Div. Groups I.S. I, II, III I ABCDEFG Dust/Ignition II, III I EFG N.I. I 2 ABCD
CENELEC Certification*:
Contacting Conductivity-
I.S. EEXia IIC T5 (Tamb = 40°C) or
EEXia IIC T4 (Tamb = 70°C)
Toroidal Conductivity- I.S. same specifications
* user must specify sensor type in order to receive an
Intrinsically safe loop
8.5 RECOMMENDED SENSORS:
(CONTACTING)
Model 140 Retractable Conductivity
Model 141/142 Insertion Conductivity
Model 150 Insertion/Submersion Conductivity
Model 400 Screw-In Low Conductivity
Model 401 Screw-In High Conductivity
INDUCTIVE FULL SCALE
Minimum Maximum
Model 222 Sensor 0-500 µS/cm 0-2000 mS/cm
Model 226 Sensor 0-50 µS/cm 0-1000 mS/cm
Model 225 Sensor 0-250 µS/cm 0-2000 mS/cm
Model 228 Sensor 0-250 µS/cm 0-2000 mS/cm
MODEL 2081C/T SECTION 8.0
DESCRIPTION AND SPECIFICATIONS
NOTE: Not applicable for conductivity less than 10 mS/cm. For these applications, please
see product data sheets for Models 1054BLC, 1055C, or 3081/81C.
CONTACTING SENSORS
Conductivity Sensor 142 142 140, 141 401-14
Model Number 400 150 150 402/403
402/403/404 400/402/403/404 400/402/403
Cell Constant 0.01 0.1 1.0 10.0
FULL SCALE MICROSIEMENS/cm
INDUCTIVE SENSORS
Conductivity Sensor
Model Number 226 228 225 222 (1in.) 222 (2 in.)
Cell Constant* 1.0 3.0 3.0 6.0 4.0
Minimum Range 50 250 250 500 500
Maximum Range 1,000,000 2,000,000 2,000,000 2,000,000 2,000,000
* Typical
34
8.6 ORDERING INFORMATION
2081C/T 67 EXAMPLE
Code Input (Required selection)
C Contacting Conductivity
T Toroidal Conductivity
Code Options
11 Stainless steel tag (specify marking)
Code Agency Approvals
67 FM approved, Intrinsically Safe (when used with approved sensor and safety barrier) and
explosion-proof, Contacting and Toroidal
69 CSA approved, Intrinsically Safe (when used with approved sensor and safety barrier) and
explosion-proof, Contacting and Toroidal
73 CENELEC approved, Intrinsically Safe (safety barrier required), Contacting and Toroidal
MODEL
2081 HART SMART TWO-WIRE CONDUCTIVITY TRANSMITTER
Model 2081C/T conductivity Two-Wire Transmitter is compatible with both contacting and inductive conductiv-
ity sensors (indicate sensor type if known). It is housed in a NEMA 4X (IP65) weatherproof, corrosion-resistant enclosure suitable for pipe mounting. Standard features include HART digital communications capability, LCD digital display, isolated 4-20 mA output, and automatic temperature compensation.
PN DESCRIPTION RECOMMENDED SPARES
23419-00 PCB, 2081 LCD Display 1
23533-00 PCB, 2081 Interface
23421-02 PCB Stack, 2081C (CPU/Sensor loop and display) 1
23421-03 PCB Stack, 2081T (Sensor loop and display)
23519-00 Push Buttons, 2081, Kit 1
33197-00 Enclosure, Middle
2002518 Enclosure Cover, LCD Display Side
2002577 Pipe Mounting Bracket
2002603 O-ring Kit (Qty 2), Window
2002604 O-ring Kit (Qty 2), Enclosure 2
3002468 Enclosure Cover, Tall
9200001 Cable, 2 Conductor, 18 AWG, Shielded
9240008-00 Overlay, 2081pH LCD Display 1
9240864 Stainless steel tag (formerly Code -07)
8.7 ACCESSORIES
MODEL 2081C/T SECTION 8.0
DESCRIPTION AND SPECIFICATIONS
35
FIGURE 8-2. MODEL 275 HART COMMUNICATION MENU TREE- MODEL 2081C
MODEL 2081C/T SECTION 8.0
DESCRIPTION AND SPECIFICATIONS
SECTION 9.0 RETURN OF MATERIAL
9.1 GENERAL.
To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Call 1-949-757-8500 for a Return Materials Authorization (RMA) number.
9.2 WARRANTY REPAIR.
The following is the procedure for returning instru­ments still under warranty:
1. Call Rosemount Analytical for authorization.
2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the unit must be supplied.
3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the materials in the same manner as they were received.
4. Send the package prepaid to:
Rosemount Analytical Inc., Uniloc Division Uniloc Division 2400 Barranca Parkway Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
9.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair instruments that are no longer under warranty:
1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed.
3. Do Steps 3 and 4 of Section 9.2.
NOTE
Consult the factory for additional informa­tion regarding service or repair.
MODEL 2081C/T SECTION 9.0
RETURN OF MATERIAL
36
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira­tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90 days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man­ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por­tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit­able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli­gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per­sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri­od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT­NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material. Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Emerson Process Management Rosemount Analytical Inc.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2003
Loading...