Please take your time to read this Instructions book in order to understand the
CONTENTS
Section 1
Page 3
Section 2
Page 4
Section 3
Page 5
Section 4
Page 7
Section 5
Page 8
Section 6
Page
9
Section 7
Page 1
0
Section 8
Pag
e 11
Section 9
Page 1
2
Section 10
Page 1
4
Section 11
Page 1
5
Section 12
Page 1
7
Section 13
Page 20
Appendix A
Page 16
Bibby Scientific Limited.
safe and correct use of your new Bibby Scientific product.
It is recommended the responsible Body for the use of this equipment reads this
instruction book and ensures the user(s) are suitably trained in its operation.
In the interest of continued development Bibby Scientific Limited reserve the right to alter or
modify the design and /or assembly process of their products without prior notification.
This product is manufactured in Great Britian by Electrothermal, part of the Bibby Scientific Group
of companies.
Beacon Road,
Stone,
Staffordshire ST15 0SA,
Great Britain.
Tel: +44(0)1785 812121
Fax: +44(0)1785 810405
Page 2 of 20 M6895 Issue 6.4
1. INTRODUCTION.
The Macro-Kjeldahl Equipment.
The Bibby Scientific MQ Series of products has been designed for Kjeldahl
extraction on a macro scale.
These multibank units are available with either 2 or 6 recesses. The units have a
stainless steel housing. There are two capacity sizes available, one for 100300ml vessels and the second for 500-800ml vessels.
The lower part of the unit houses the control equipment. This is separated from
the heating element by a stainless steel tray and a well ventilated air space.
Each heating mantle has its own energy regulator incorporating an ON/OFF
switch and an amber, mains to heater neon indicator. There is also a mains On
indicator (clear) on the front panel. The units are also fused on the supply line
and neutral except MQ3868B/Ex1 which has no fuse ( see Section 5).
Note: MQ products incorporate earth screen and are CE marked.
The six recessed versions when specified for use with 500-800ml vessels is only
available in a 230 V~AC rating. All other MQ’s are available in 115Volt~AC or
230V~AC. Note the MQ3868B/Ex1 is only for sale outside of Europe and is not
CE approved.
Page 3 of 20 M6895 Issue 6.4
2. SYMBOLS AND USING THIS INSTRUCTION BOOK.
2.1. Throughout this Instruction book the following symbols are shown
to identify conditions which pose a hazard to the user, or to identify actions that
should be observed. These symbols are also shown on the product, or its packaging.
When a symbol is shown next to a paragraph or statement it is recommended the
user takes particular note of that instruction in order to prevent damage to the
equipment or to prevent injury to one’s self or other people.
The Responsible Body and the Operator should read and be familiar with this
Instruction book in order preserve the protection afforded by the equipment.
To prevent injury or equipment damage it is the manufacturers
recommendation that all persons using this equipment are suitably trained
before use.
2.2. Symbols defined.
Caution, risk of danger. See note or adjacent symbol.
Protective conductor terminal to be earthed.
(Do not loosen or disconnect).
Caution / risk of electric shock
Recyclable Packing Material.
Do not dispose of product in normal domestic waste.
Caution. Hot surface.
Material irritant to skin. When handling wear face mask to BS/EN 149 and protective gloves
Bio Chemical Hazard. Caution required. Will require decontamination.
Refer to Instruction book
Hot Zone on
MQ product.
Page 4 of 20 M6895 Issue 6.4
3. SAFETY INFORMATION.
DO NOT
Never
Do not
Do
not
Do not
Do not
Do no
t
Do not
Do not
Do not
This product has been designed for safe operation when used as detailed in accordance
with the Manufacturers instructions.
NOTE: Failure to use this equipment in accordance with the manufactures instruction
book may compromise your basic safety protection afforded by the equipment and may
invalidate the warranty / guarantee. The warranty / guarantee does not cover damaged
caused by faulty installation or misuse of the equipment
3.1. Prevention of Fire and Electric Shock.
To prevent a risk of fire or electric shock,
case without authorisation. Only qualified Service personnel should
attempt to repair this product.
Replace fuses only with the type as listed in section, ‘Technical
Specifications and Parts and Accessories’ (See fuse type and rating).
Ensure the Mains Power Supply conforms to rating found on the data
plate located on the back of this product.
Operate this equipment without connection to earth / ground.
Ensure the mains supply voltage is correctly earthed / grounded in
accordance with current area legislation.
3.2. General Safe Operating Practice.
Always follow good laboratory practice when using this
equipment. Give due recognition to your company’s safety and
legislative health & safety procedures and all associated legislation
applicable to your areas of operation. Check laboratory procedures for
substances being heated and ensure all hazards (e.g. explosion,
implosion or the release of toxic or flammable gases) that might arise
have been suitably addressed before proceeding. When heating
certain substances the liberation of hazardous gases may require the
use of a fume cupboard or other means of extraction.
Ensure equipment is used on a clean, dry, non-combustible, solid work
surface with at least 300mm suitable clearance all around from other
equipment.
position the product so that it is difficult to disconnect from the
mains supply.
touch the heating element or any glass vessel whilst in use.
open your product
lean or stretch over equipment, glassware and fixings when in
use.
immerse unit in water or fluids.
spill substances onto the MQ. If spillage does occur,
disconnect unit from mains supply and follow instructions as detailed
in Maintenance. (Section 9).
cover the MQ whilst in use.
block or obstruct ventilation
slots / airways.
Only use an Earth screened mantle with electrically conductive
vessels.
leave equipment switched on without a charged flask(s).
Page 5 of 20 M6895 Issue 6.4
Do not
Do not
ATTENTION:
-
as the insulation used may obstruct the convection
cooling airways around the rim of the cartridge enclosure and cause
the mantle to overheat.
It is not recommended to leave any heating apparatus unattended
during operation.
Only use Original Equipment manufactures spares and accessories.
Ref Section 10.
The equipment is not spark, flame or explosion proof and has not been
designed for use in hazardous areas in terms of BSEN 6007914:1997. Keep flammable, low flash point substances away from the
apparatus.
Keep the mains cord away from the heating surface.
thermally insulate the exposed upper section of the vessel(s),
operate or handle any part of the product with wet hands.
With high energy input and certain configurations of glassware in MQ
products, where the heating contact to glassware is relatively small,
localised heating and subsequent ‘bumping’ of the fluid being heated
may occur. Application advice should be sought from the
manufacturer.