80 Volt EXT400/200/30H Compound Turbomolecular PumpB756-60-000
Page 2
Declaration of Conformity
We, Edwards,
Manor Royal,
Crawley,
West Sussex, RH10 9LW, UK
declare under our sole responsibility, as manufacturer and person within the EU authorised
to assemble the technical file, that the product(s)
EXT Compound Turbomolecular pump:
EXT400/200/30H B756-60-000
to which this declaration relates is in conformity with the following standard(s) or other
normative document(s)
EN ISO 12100-2: 2003 Safety of Machinery. Basic Concepts, General Principles for
+ A1: 2009 Design. Technical Principals
EN1012-2:1996, A1: 2009 Compressors and Vacuum Pumps. Safety Requirements.
Vacuum Pumps
EN61010-1: 2001 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use. General Requirements*
EN 61326-1: 2006 Electrical equipment for measurement, control and laboratory
Use. EMC requirements. General requirements.
†
C22.2 61010-1-04: 2004
Control and laboratory use – Part 1: General requirements
UL61010-1: 2002
Control and laboratory use – Part 1: General requirements
* The pumps comply with EN61010-1: 2001 when installed in accordance with the instruction manual supplied.
†
Pumps only comply with the Canadian Standards Authority and Underwriters
Laboratory when used with EXDC 80V family of controllers, EXC100E and EXC100L controllers.
and fulfils all the relevant provisions of
2006/42/EC Machinery Directive
2006/95/EC Low Voltage Directive
2004/108/EC Electromagnetic Compatibility (EMC) Directive
Note: This declaration covers all product serial numbers from the date this Declaration was
signed onwards.
†
Safety requirements for electrical equipment for measurement,
Safety requirements for electrical equipment for measurement,
9 December 2009
B. D. Brewster, Technical Manager,
Burgess Hill Products
This product has been manufactured under a quality system registered to ISO9001
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 5
B756-60-880 Issue D
CAUTION
WARNING
1Introduction
1.1Scope and definitions
This manual provides installation, operation, maintenance and storage instruction s for the Edwards EXT400/200/30H
Split Flow Compound Turbomolecular Vacuum Pump. Please read and follow all instructions in this manual.
The EXT400/200/30H Split Flow Compound Turbomolecular Vacuum Pump is designed for use with an Edwards EXDC
Controller. Read this manual and the instruction manual supplied with your controller before you attempt to install
or operate the equipment. The controller manual contains details of electrical installation.
Important safety information in this manual is highlighted as WARNING and CAUTION instructions; please obey these
instructions. The use of WARNINGS and CAUTIONS is defined below.
Warnings are given where failure to observe the instruction could result in injury or death to
people.
Introduction
Cautions are given where failure to observe the instruction could result in damage to the equipment, associated
equipment and process.
In accordance with the recommendations of EN61010, the following warning symbols may appear on the p ump or its
accessories:
Warning - refer to accompanying documentation.
Warning - risk of electric shock.
Warning - hot surfaces.
The units used throughout this manual conform to the SI international system of units of measurement. Also
throughout this manual, wherever flow rates are specified, the abbreviation 'sccm' is used to mean standard
3
cm
min-1: this is a flow of 1 cm3min-1 at an ambient temperature of 0°C and a pressure of 1 013 mbar (1.013 x 105 Pa).
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 7
B756-60-880 Issue D
CAUTION
WARNING
WARNING
WARNING
WARNING
2Technical data
2.1General
Table 1 - General data
General itemsReference data
PerformanceRefer to Table 3
DimensionsRefer to Figure 1
Maximum inlet-flange temperature70°C
Maximum magnetic field3.5 mT
Installation categoryEN61010 part 1, Category 1
Pollution degreeCAN/CSA, UL, EN61010 part 1, Category 2
Altitude2000 m
Humidity10 to 90 % non condensing
Electrical supply80 V
Equipment typeFixed equipment, for indoor use only
Technical data
2.2Pumping media
Vent dangerous gases and gas mixtures safely. Do not expose people to these gases.
Do not use the EXT400/200/30H Split Flow pump to pump explosive gas mixtures as the pump is
not suitable for this purpose.
If pyrophoric gases are pumped then the customer must supply adequate dilution with an inert gas.
If pumping hazardous gases or vapours, the customer must observe the safety recommendations
of the supplier of the gas or vapour.
Do not use the EXT400/200/30H Split Flow pump to pump gases containing more than 20 % oxygen. If you do,
the lubricant will polymerise and the pump will fail prematurely.
Note:Concentrations of gases may be modified by the compression of the pump.
The pump is designed to pump the following residual gases normally used in high-vacuum systems:
z Airz Nitrogenz Carbon dioxide
z Carbon monoxidez Kryptonz Helium
z Neonz Argonz Hydrogen
z Ethanez Propanez Butane
z Methane
You can use the pump to pump oxygen and water vapour, subject to the following conditions:
z Oxygen – The oxygen concentration must be less than 20 % by volume.
z Water vapour - You must ensure that vapour does not condense inside the pump.
If you wish to pump a gas not in the list above, co ntact your supplier for advice. If you do not contact your supplier,
you may invalidate the warranty on the pump. The pumps are not suitable for pumping aggressive or corrosive gases.
2.3Vent gas specification and vent control data
Although the pump may be vented to atmosphere, high relative humidity of the air may greatly increase the
subsequent pump-down time. To reduce pump-down times you should vent the pump with dry, clean gases.
Table 2 - Vent gas and vent control data
Vent itemsReference data
Vent gasDry air, nitrogen, argon or other inert gases
Maximum dew point at atmospheric pressure-22°C
Maximum size of particulate1 µm
Maximum concentration of oil0.1 parts per million
Time for rotational speed to reduce to 50 %≥15 seconds
Purge gas pressure
*
If you wish to use purge pressures in excess of this value, consult Edwards.
*
≤ 0.5 bar gauge
Table 3 - Technical data
ParameterEXT400/200/30HNotes
Mass14.67 kg
Main Inlet PortRectangular96 x 210 mm
Side Inlet PortRectangular96 x 60 mm
Interstage Inlet PortRectangular96 x 20 mm
Outlet-flangeDN25NW
Vent-port1/8 inch BSP
Main Inlet Pumping Speed
Pumping speeds are without inlet-screen.
Inlet-screens reduce speed by approximately 10% for main and side inlets and by approximately 4 % at
interstage inlet.
‡
With an E2M28 backing pump.
**
Point at which pump folds back.
††
Above this pressure, rotational speed drops below nominal.
‡‡
The customer must ensure that the exhaust system and relief valves are capable of passing the maximum
throughput without causing the maximum pump outlet pressure to be exceeded.
2.4Materials exposed to gases pumped
The following materials and component types are exposed to the gases pumped: aluminium alloys, stainless steels,
fluoroelastomer and nitrile O-rings, hydrocarbon lubricant, felt, rare earth magnets, silicon nitride, phenolic resin
and carbon-fibre reinforced epoxy resin.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 13
B756-60-880 Issue D
WARNING
3Installation
3.1Unpack and inspect
The pump weighs approximately 15 kg (33 lbs). Appropriate care should be taken when lifting and
moving the pump to avoid injury.
Take care when you unpack the pump to avoid excessive shocks w hich could damage the bearings and reduce the life
of the pump. The pump is supplied with the inlet and outlet sealed to p re vent entry of dust and vapour. Do not
remove these seals until you are ready to install the pump on your vacuum system.
Cut the plastic fibre banding strip/s with suitable sidecutters/knife, being aware that the strip is under tension and
could therefore cause a whipping hazard. Open the cardboard box from the top by firstly removing the tape and then
by removing the staples. Once the box lid is open, remove the upper piece of foam. The cardboard can then be lifted
away from the pallet by pulling in a vertical direction. The pump can now be lifted vertically from the packaging
using appropriate mechanical lifting equipment and the two M8 tapped holes provided in the top of the pump. If the
pump is damaged, notify your supplier and the carrier in writing within three days; state the Item Number of the
pump together with your order number and your supplier’s invoice number. Retain all packing materials for
inspection. Do not use the pump if it is damaged.
Installation
Check that your package contains the items listed in Table 4. If any of these items is missing, notify your supplier in
writing within three days.
If the pump is not to be used immediately, store the pump in suitable conditions as described in Section 6.1.
It is advised to retain all packing materials for use, should you return it for service.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 15
3.3Connection to the vacuum system
WARNING
WARNING
WARNING
WARNING
Install the pump in the vacuum system before you connect the controller to the power supply. This
will ensure that the pump cannot operate and injure people during installation.
The impeller blades on an EXT400/200/30H are very sharp. Care should be taken when handling
the pump not to dislodge or damage the protective inlet screens.
Under no circumstances must a part of the human body be exposed to a vacuum.
Carefully remove the pump from packaging media and connect to vacuum system.
Table 4 - List of items supplied
B756-60-880 Issue D
Installation
QtyDescriptionCheck (9)
1EXT400/200/30H Split Flow Turbomolecular Pump
1Large inlet-screen (fitted)
1Medium inlet-screen (fitted)
1Small inlet-screen (fitted)
1Large inlet seal (O-ring)
1Medium inlet seal (O-ring)
1Small inlet seal (O-ring)
1EXDC160 controller (fitted)
*
EXDC160 controller only supplied with system variants.
3.3.1Mechanical fixing
Ensure the EXT400/200/30H Split Flow pump is securely fixed to the vacuum system via its
inlet-flange using all available mounting points. If the pump were to seize when not securely
mounted, the stored energy of the rotor could cause rapid movement of the pump, which may
cause further damage and injury to personnel.
Removal of inlet-screens will expose the risk of injury from sharp edges.
Do not remove the inlet-screens unless you can be sure that there is no danger that debris can fall into the pump.
To remove an inlet-screen carefully extract it from the inlet-flange using a bent wire hook. Once removed, a screen
is not reusable. Spares are available (refer to Section 7).
Figure 3 - Correct installation of inlet-screen
3.3.3Inlet connection and orientation
Do not store or orientate the pump such that the finned body is above the pump inlets.
The pump can be mounted in any atti tude apart from one where the finned body (with electrical connector) is above
the inlets of the pump. Mounting in such a condition may allow oil from the lower lubricated bearing to drain into
the pumping mechanism, reducing performance and pump life.
Make sure that the pump's inlets and all components fitted to the pump-inlets are clean and dust-free. If the
pump-inlets are not kept clean, the pump-down time may be increased.
Use the O-rings supplied with the pump and use six M8 bolts tightened to a torque of 20 Nm (15 lbf ft) to connect the
inlet-flange of the pump to the vacuum system.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 17
Figure 4 - Mounting attitude of the EXT400/200/30H split flow pump
CAUTION
WARNING
WARNING
1. Inlets
2. Backing port
B756-60-880 Issue D
Installation
3.3.4Backing connection
The customer must ensure safe ducting of the backing line if oil mist or hazardous substances are
present.
The EXT400/200/30H must not be operated, or vented from a positive pressure gas supply, with a
restricted/blocked backing line.
Do not use the EXT400/200/30H Split Flow pump with a backing pressure below 5 x 10-4 mbar (5 x 10-2 Pa).
Lower backing pressures will increase the evaporation rate of the lubricating oil and so will reduce the life of
the bearings.
Use suitable vacuum tubing and connectors to connect the NW25 flange of the backing-port to your backing-pump.
If necessary, use flexible pipe or bellows to reduce the transmission of vibration from the backing-pump to the
EXT400/200/30H Split Flow pump.
We recommend that you use an Edwards E2M28 backing-pump. The minimum size of the backing-pump required is
given in Table 3. You may have to use a larger backing-pump if you run the pump at a high inlet pressure. The
EXT400/200/30H is also suitable for use with diaphragm backing-pumps although the effect of higher backing
pressure on the pump's performance and cooling requirements should be noted (refer to Table 3 and Section 3.6).
3.4Vent options, vent valve connection and control
To maintain the cleanliness of your vacuum system, we recommend that, whenever you switch the pump off, you
vent the pump (or vacuum system) when the speed of the EXT400/2 00/30H Split Flow pump is between full rotational
speed and 50 % of full rotational speed.
At and above 50 % of full rotational speed, the rotor spins fast enough to suppress any backstreaming of hydrocarbon
oil from your backing pump. Venting may be accomplished by one of the following methods:
z Use a TAV5 or TAV6 solenoid vent-valve accessory (refer to Section 7) in place of the manual vent-valve.
z Use a TAV5 or TAV6 solenoid vent-valve connected to a convenient flange on your vacuum system.
z Use an alternative valve, with an appropriate restriction, connected to your vacuum system. For further
details contact Edwards.
However, if you manually vent the pump when it is at full rotational speed and the rate of pressure rise is too high,
the pump life may be reduced. When using the manual vent valve supplied, we recomme nd that you either limit the
vent or only open the vent-valve after the EXT400/2 00/30H Split Flow pump speed has falle n to 50 % of full rotational
speed.
Do not vent from the backing line, this may lead to contamination. If you vent into your vacuum system, select a
point upstream of the pump, to prevent oil backstreaming from the backing line.
If you use the TAV5 or TAV6 vent-valve you can control it using an EXC controller, or by other methods. Table 5 gives
an indication of the approp riate orifice size to be fitted to the vent valve for given vacuum system volumes in order
that the vent rate is kept within the limits given in Section 4.
Table 5 - Vent-restrictor orifice diameter (with atmospheric pressure at the inlet of the vent-valve)
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 19
3.5Electrical installation
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
The customer must ensure that any electrical circuits are protected from dripping water caused
by condensation on cold surfaces.
It is the responsibility of the customer to ensure that the power supply used is correctly
rated/protected.
This product requires a separate power supply (not included). The power supply should be
adequately protected against a hazardous live condition (for example in case of a short circuit).
Do not remove the EXDC/EXC controller cable from the pump until the pump is completely at rest.
This could expose the user to hazardous voltage and potentially damage the EXDC/EXC controller.
B756-60-880 Issue D
Installation
The customer must ensure appropriate routing of cables and pipework to avoid slip/trip hazards.
The EXT400/200/30H Split Flow pump should be electrically bonded to earth using the connection provided
(M5 thread located next to the pump electrical connector). Refer to the instruction manual supplied with the
controller to complete the electrical installation.
3.5.1EXDC controllers
The customer must provide an emergency stop circuit to turn off power to the EXDC controller.
An EXDC controller requires connection to a suitably rated power supply. The EXDC Controller is designed to allow a
pumping system to operate in a fully automatic system.
3.5.2EXC controllers
The EXC Controller provides the electrical supply to the EXT400/2 00/30H Split Flow pump through the
pump-to-controller cable. Connect and lock the bayonet-connectors at the ends of the cable to the mating
connectors on the pump and the EXC Controller (if applicable). If the cable is disconnected at the pump, the
EXC Controller output is switched off, making the cable safe. However, if the cable is disconnected from the
EXC Controller it should be noted that the other end of the cable is still connected to the pump. Refer to the
appropriate warning in Section 3.5 regarding exposure to hazardous voltages whilst the pump is still rotating.
The EXC Controller is designed to allow a pumping system to be configured in a variety of ways, from a basic
manually-operated system to a fully automatic system with remote control.
You must cool the pump by forced-air to prevent damage to the bearing.
You must use forced-air cooling to cool the pump. You must ensure that there is an adequate supply of cooling-air to
the pump.
During operation, if the temperature of any surface of the pump is higher than 60°C, the pump is too hot and you
must increase the cooling.
3.6.2Forced-air cooling
An air-cooler accessory is available for the EXT400/2 00/30H pump (refer to Section 7). Fit the air-cooler as described
in the instruction manual supplied with it. I f you w ish to us e an alternative fan for air-cooling, ensure that the flow
rate is above 70 m3h-1 (40 cfm).
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 21
4Operation
WARNING
WARNING
WARNING
WARNING
Do not operate the EXT400/200/30H Split Flow pump unless it is connected to your vacuum
system. If you do, the pump rotor can cause injury. The pump rotor rotates at very high speeds
and the rotating blades might not be visible.
Do not move the pump whilst it is running. The gyroscopic forces generated by this movement can
cause excessive use of the back-up bearing and may result in catastrophic failure of the pump.
When power is restored following a power cut, the pump will re-start automatically. The pump
must remain connected to the vacuum system to prevent risk of injury.
B756-60-880 Issue D
Operation
After power to the pump has been switched off, either through emergency or as a requirement,
the rotor will continue to spin at very high speeds. Until the rotor has stopped it possesses
considerable mechanical energy.
4.1Start-up
Use the procedure below to start up a basic, manually-controlled pumping system with a manual vent-valve and an
EXDC version controller. Refer to the EXDC instruction manual.
1. Ensure any vent-valves are closed (including the manual vent-valve on the pump body).
2. Switch on the power to the air cooler.
3. Start the backing-pump.
4. When the vacuum system pressure is approximately 10 mbar or less, switch on the power supply to start the
EXT400/200/30H Split Flow pump.
5. The pump will then accelerate to full operating speed. Once nominal operating speed is reached, the normal
speed LED on the EXDC Controller will illuminate.
Note:Refer to the controller manual if using the EXC Controller.
4.2Shut-down
Note:In an emergency only, open the vent-valve quickly to decelerate the pump rotor in the shortest possible
time.
Use the procedure below to shut down a basic, manually-controlled pumping system with a manual vent-valve and
an EXDC Controller. Refer to the EXC Controller instruction manual when using an EXC Controller.
1. Close the valve in the backing-line connecting the EXT400/200/30H Split Flow pump to the backing-pump.
2. Switch off the backing-pump.
3. Switch off the power supply to the EXT400/200/30H Split Flow pump.
4. When the EXT400/200/30H Split Flow pump rotational speed has fallen below 50 % of full rotational speed, turn
the manual vent-valve anti-clockwise to open it. Refer to Section 3.4 for venting procedure.
4.3Safety interlocks and control system
The pump protection and safety interlock features are listed below. Refer to the instruction manual supplied with
the controller for a full description of these features.
4.3.1EXDC controllers
z The EXDC Controller monitors the temperature of the EXT400/200/30H Split Flow pump and the electrical
power consumption of the pump. If the EXDC Controller detects excessive power consumption or
temperature, the rotational speed of the pump motor is reduced until the power and temperature return to
normal.
4.3.2EXC controllers
z The EXC Controller monitors the temperature of the EXT400/200/30H Split Flow pump and the electrical
power consumption of the pump. If the EXC Controller detects excessive power consumption or temperature,
the rotational speed of the pump motor is reduced until the power and temperature return to normal.
z If the rotational speed is reduced to 50 % of nominal speed, then power is removed (or after a user defined
time delay) and the FAIL LED on the EXC Controller lights.
z If pump rotational overspeed is detected by the EXC Controller, the pump is stopped immediately and the
FAIL LED on the EXC Controller lights.
If the FAIL LED lights, switch off the backing-pump i mmediately and vent the EXT400/200/30H Sp lit Flow pump. Once
the EXT400/200/30H Split Flow pump has stopped, rectify the cause of the failure (refer to Section 5.5), press the
EXC Controller Start/Stop button to reset the Fail condition, and restart the EXT400/200/30H Split Flow pump. If the
pump is hot, allow sufficient time for it to cool before you restart it.
Note:Any references to LEDs and buttons do not apply to OEM versions of the EXC controller.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 23
B756-60-880 Issue D
WARNING
WARNING
5Maintenance
The EXT400/200/30H Split Flow pump is not to be serviced by the customer. Pumps requiring
servicing should be returned to Edwards or serviced by a qualified Edwards engineer.
Allow the pump-rotor to stop, then disconnect the controller before you remove the pump from
your vacuum system for maintenance or fault-finding procedures.
5.1Introduction
The maintenance operations for the EXT400/200/30H Turbomolecular vacuum pump are described in the following
sections. The inlet-screens and inlet-flange seals are available as spares (refer to Section 7). Fit the inlet-screens as
shown in Figure 3.
Maintenance
5.2Bearing maintenace
When supplied, the pump contains sufficient lubricant to supply the bearings for life. No routine maintenance is
therefore required between bearing replacements. The bearings are not user-serviceable. The bearings will need to
be replaced when they reach the end of their service life. This is typically more than 20,000 hours, but may be less;
this depends on the type of pumping duty on which the pump is used.
When the bearings need replacement, we recommend that you exchange your pump for a factory reconditioned
replacement. Alternatively, you can send your pump to an Edwards Service Centre to have the bearings replaced.
When you return the EXT400/200/30H Split Flow pump to Edwards Service Centres please use the procedure included
at the end of this manual. However, the instruction to drain all fluids does not apply to the lubricant in the
EXT400/200/30H Split Flow pump oil-r eservoir.
5.3Rotor life
The fatigue life of the EXT400/200/30H rotor is typically 40,000 to 50,000 cycles of acceleration from rest to full
speed and then back to rest. As a precauti onary measure, Edwards recommends tha t pumps are returned for a major
service (rotor replacement) after 20,000 cycles or ten years, whichever occurs first.
Clean the external surfaces of the EXT pump in a well-ventilated location. When you use cleaning
solutions and solvents to clean the pump, observe all precautions specified by the manufacturer.
Avoid inhalation of any particulates which may be present in the pump.
Do not attempt to clean any parts of the EXT pump other than external surfaces. Organic solvents may damage
internal pump components. Do not use abrasive materials to clean any part of the pump.
If the inside of the EXT pump is contaminated, it may not be possible to achieve the specified ultimate vacuum, or
pump-down time may increase. In these circumstances the pump must be returned to an Edwards Service Centre,
where the pump will be dismantled and cleaned. Use the procedure given in the forms at the end of this manual to
return the pump.
You can use any organic solvent to clean the external surfaces of the EXT pump. Edwards recommend the use of
non-CFC solvents such as isopropanol or ethanol. Use a cleaning solution which is suitable for the contaminants on
the pump surfaces.
For environmental reasons, keep wastage of cleaning solutions and solvents to a minimum.
5.5Fault finding
Refer to Table 6 for the possible causes of faults and for the recommended actions to rectify faults. Table 6 is
applicable to a basic, manually controlled pumping system with an EXC Controller configured for local (manual)
operation.
Note that if you use an EXDC Pump Drive Module to control the EXT400/200/30H Split Flow pump, or if you use an
EXC Controller configured for remote operation to control the EXT400/200/30H Split Flow pump, some of the checks
and actions in Table 6 may not apply to your system. Refer to the fault finding section of the instruction manual
supplied with your EXDC Pump Drive Module or EXC Controller for further fa ult finding information.
Table 6 - Fault finding
SymptomCheckAction
Pumps driven by EXDC controllers:
The impeller does not rotate after
giving start signal
Are the three red phase LEDs on the
EXDC Controller lit?
Is the EXDC correctly connected to
the EXT pump?
Is the start signal being given across
the correct pins of the EXDC logic
interface plug?
If not, check that power is supplied
to the EXDC.
If power is supplied but one or more
of the LEDs are not lit, then the EXDC
Controller may be
faulty - change the EXDC and try to
start the pump.
If the problem persists then the EXT
pump may be faulty . Consult Edwards
or your supplier.
If the red LEDs are all lit, check that
the EXDC is properly connected to
the pump.
Check that the start signal is being
given across the correct pins of the
EXDC.
SymptomCheckAction
Pumps driven by EXC controllers
(additional checks):
The impeller does not rotate after
giving start signal
The EXC Controller trips into fail (at
any speed)
Is the EXC Controller power LED lit?If not, check that the electrical
Is the EXC Controller start/stop LED
flashing?
Is the EXC Controller first speed
indication lit?
Are the system interlocks correctly
connected?
Is the timer set incorrectly?Increase the timer setting (refer to
Is the inlet pressure too high?
Is the backing pressure too high?
Is the EXT pump running too hot?Increase the cooling or reduce the
Does the pump rotor rotate freely?If not, the EXT pump bearings may
None of the above.Consult Edwards or your supplier.
supply is on, check that the switch
on the rear of the EXC Controller is
on, check the fuse at the rear of the
EXC Controller.
If all of the above are ok then the
EXC Controller is faulty. Consult
Edwards or your supplier.
If so, check that the correct links are
made on the EXC Controller logic
interface (refer to the instruction
manual supplied with the EXC
Controller).
Check that any system interlocks are
correctly made (refer to the
instruction manual supplied with the
EXC Controller).
Check that the pump to Contro ller
lead is connected.
If not, the EXC Controller is faulty. If
lit, the EXT pump is faulty. Consult
Edwards or your supplier.
Ensure that the system interlocks do
not open after the EXT pump has
started.
the EXC Controller instruction
manual).
If so, check for a gross leak in the
system; if the system is leak tight,
then reduce the pumping load.
pumping load; check that external
heat sources are not excessive.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 27
B756-60-880 Issue D
WARNING
6Storage and disposal
6.1Storage
Use the following procedure to store the pump.
1. Place protective covers over the main inlet, side inlet, interstage inlet, outlet and vent ports.
2. Place the pump in its packing materials. For fastest pump-down when the pump is put back into service, seal the
pump inside a plastic bag together with a suitable desiccant.
3. Store the pump in cool, dry conditions, preferably not exposed to atmospheric air until required for use. When
required, prepare and install the pump as described in Section 3.
4. Keep the pump upright at all times to prevent the drainage of oil from the bearing reservoir.
5. Avoid long-term storage if possible. When long-term storage is necessary, the pump should be set up and run for
at least eight hours every six months.
6.2Disposal
Storage and disposal
In the unlikely event of a failure of the pump rotor, dust can be generated from the carbon fibre
reinforced components.
In this event, use appropriate Personal Protective Equipment when handling and disposing of the
pump, and ensure that all pump inlets and outlets are capped off before disposal.
Dispose of the EXT400/200/30H Split Flow Turbomolecular Pump and any components and accessories safely in
accordance with all local and national safety and environmental requirements.
Particular care must be taken with any components which have been contaminated with dangerous process
substances.
Take appropriate action to avoid inhalation of any particulates which may be present in the pump. Do not incinerate
the pump. The pump contains phenolic and fluorosilicone materials which can decompose to very dangerous
substances when heated to high temperatures.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 29
B756-60-880 Issue D
7Service, spares and accessories
7.1Introduction
Edwards products, spares and accessories are available from Edwards compani es in Belgium, Brazil, Canada, France,
Germany, Hong Kong, Italy, Japan, Korea, United Kingdom, USA and a worldwide network of distributors. The
majority of these employ service engineers who have undergone comprehensive Edwards training courses.
Order spare parts and accessories from your nearest Edwards company o r di st ri buto r. When you order, pleas e state
for each part required:
z Model and item number of your equipment.
z Serial number (if any).
z Item number and description of the part.
7.2Service
Edwards products are supported by a worldwide network of Edwards Service Centres. Each Service Centre offer s a
wide range of options including: equipment decontamination; service exchange; repair; rebuild and testing to factory
specifications. Equipment which has been serviced, repaired or rebuilt is returned with a full warranty.
Service, spares and accessories
Your local Service Centre can also provide Edwards engineers to support on-site maintenance, service or repair of
your equipment.
For more information about service options, contact your nearest Service Centre or other Edwards company.
7.3Spares
7.3.1Inlet-screens
Inlet-screens are fitted to your pump as supplied to prevent damage from the entry of debris into the pump. The Item
Number of replacement inlet-screens is given in Table 7.
7.3.2Inlet-flange seals
The EXT400/200/30H Split Flow pump is suppli ed with inlet seals. The Item Number of replacement seals is given in
The accessories available for use with the EXT400/200/30H Split Flow turbomolecular vacuum pump are described
in the following sections. Figure 5 shows how the accessories are fitted to an EXT400/200/30H Split Flow pump.
The Item Number of the EXDC Controller is shown in Table 8.
Table 8 - EXDC drive modules
Drive ModuleItem Number
EXDC160 80VD396-41-000
7.4.3EXC controllers
Fit an EXC Controller and pump-to-controller cable as an alternative to an EXDC Controller. Refer to Table 9.
Table 9 - EXC controllers
ControllerVoltageItem Number
EXC100E90 - 264 V a.c.D396-20-000
EXC100L90 - 264 V a.c.D396-22-000
EXC12090 - 264 V a.c.D396-16-000
EXC30090 - 132/180 - 264 V a.c.D396-14-000
7.4.4Pump-to-controller cables
You must fit a pump-to-controller cable between an EXC Controller and the EXT400/200/30H Split Flow pump. A
cable is not supplied with the EXT400/200/30H Pump or the EXC Controller (except EXC100L). Refer to Table 10.
Two solenoid-operated vent-valves are available for system venting. The valves are 24 V d.c., normally-open, and
can be driven automatically from the EXC Controller. The solenoid-valve is fitted in place of the manual-valve, or
alternatively can be fitted with an adaptor (supplied with th e valve) and be used with any suitable NW10 flanged p ort
on your vacuum system. The vent-port adaptor allows the vent-port or the purge-port to be used with any suitable
NW10 fitting. Refer to Table 11.
Table 11 - TAV vent-valve and vent-port adaptor
ProductItem Number
TAV5 vent-valveB580-66-010
TAV6 vent-valveB580-66-020
NW10-1/8 inch BSP male adaptorB580-66-011
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 31
B756-60-880 Issue D
7.4.6ACX air-cooler
An ACX air-cooler can be fitted to the EXT400/200/30H Split Flow pump. However, please refer to Section 3.6 to
check the suitability of air-cooling in a particular application. Refer to Table 12.
Table 12 - AXC air-cooler
Air-coolerItem Number
ACX250HB580-53-160
7.4.7VRX vent-restrictor
Use a VRX fixed orifice vent-restrictor to restrict the flow of vent gas into the EXT400/200/30H Split Flow pump.
Refer to Section 3.4 for information on the selection of the correct VRX vent-restrictor. Refer to Table 13.