Dynalco TMP-100 User Manual

Product Data Sheet
March 2009
TMP-100
Multi-Point
Digital Pyrometer
Instrumentation & Controls
SPECIFICATIONS
2-Year
Warranty
FEATURES
z Digital readout with 1º resolution. z Intrinsically safe. Third–party approved. No
expensive hazardous enclosure required.
z Heavily gold-plated selector switch for long life,
trouble-free operation.
z Integral 2-pole selector switch permits using
grounded or ungrounded thermocouples.
z Requires no external dc or ac power. z Fits existing cutouts and mounting holes of other
popular pyrometers.
z Compensation for lead lengths or lead resistances
not required; use inexpensive small gauge thermocouple extension wire and save cost and space.
z Fully gasketed; rugged construction. z Readout unaffected by power line noise and
spikes (no line connections).
Display: 3-1/2 digit (–999 to +1999), 0.5" high, liquid crystal display, 1º increment. Brightness
increases with ambient light (no outdoors fadeout).
z Open thermocouple or RTD indicated by a
number 1 in the thousands column; all other digits blanked.
z Momentarily pressing the front display push
button activates the display for 50 seconds.
Environment Temperature: 0ºF to +165ºF (–18ºC to +74ºC); 0.25% maximum effect on readout.
Selector Switch: Two-pole, 19-position, double gold plating of contacts.
Sensor:
z Thermocouples:Grounded or ungrounded;
integral cold junction compensation. Thermocouple extension wire resistance of up to
100Ω introduces less than 1º error.
z RTDs: Accepts 2-wire RTDs only. Use with heavy
copper extension leads to minimize error, i.e., 16 AWG for up to 50 feet, 14 AWG to 100 feet.
Power: Two common 9-volt batteries such as Mallory M1604HD2, Eveready 1222, Ray-O-Vac D1604. Battery life is approximately 4000 viewings, or a couple of years, under typical operation. Appearance of decimal points after every digit indicates discharged batteries.
INTRINSIC SAFETY
*THIRD PARTY APPROVALS
Canadian Standards Assoc:
Intrinsically safe for Class I, Division 1, Groups C and D hazardous locations, when used only with previously mentioned 9-volt batteries.
.
Per ISA-RP-12.2 criteria. All circuits and electrical components operate at low energy levels, incapable of releasing sufficient electrical or thermal energy under normal or abnormal conditions to cause ignition of hazardous atmospheric mixtures of pentane, ethylene, or methane in their most easily ignited concentrations.
OUTLINE DRAWING
A division of Crane Co.
(13.9)
(13.6)
(4.4)
(4.4)
(11.3)
(6.3)
(12.9)
Dimensions in inches (cm)
Toll Free 800.368.6666 (US & Canada) • Tech Support 800.950.2382
Main 954.739.4300 • Fax 954.484.3376 • www.dynalco.com
3690 NW 53rd Street, Ft. Lauderdale, FL 33309 USA