Dwyer DPGW User Manual

Bulletin A-34C
Series DPGA & DPGW Digital Pressure Gages
Specifications - Installation and Operating Instructions
STANDARD RANGES
CERAMIC RANGES
3-11/16
[93.66]
4-9/64
[105.17]
The Series DPGA & DPGW Digital Pressure Gages have a precise +/- 1.0% full scale accuracy. The 4 digit digital display will reduce the potential for errors in readings by eliminating parallax error commonly produced with analog gages.
The DPGA & DPGW are packaged in an ABS plastic case. The unit is battery powered and has an auto-shut off to conserve battery life. A two button key pad allows easy access to features without the need to work through complex menus. These features include, auto zero and conversion of the pressure units. Readings may be converted to various engineering units. See reverse page for available units.
OPERATING INSTRUCTIONS
For complete instructions on the two button operation see the flow chart provided on back. On/Off: Press ON/OFF to toggle on and off. Device will automatically shut off after 20 minutes if not used. Zero: Zeros the display - Push & hold for 2 seconds to zero the display. If after pressing zero the display does not re-zero, the unit is no longer meeting specified accuracy and should be sent back to the factory for calibration. Do not use the zero button when pressure is applied. Units: To change units, press both On/Off and Zero buttons simultaneously. Overpressure Indicator: The LCD will flash “OFL”, if the pressure applied is over 105% F.S.
ø2-43/64
[67.87]
1-1/2 [38.10]
1-1/4 NPT
SPECIFICATIONS Service: DPGA: Air and compatible gases; DPGW: Liquids
and compatible gases.
Wetted Materials:
DPGA: 316L SS, Silicone sensor; DPGW: Type 316L SS; Ceramic Ranges: Type 316L SS, ceramic, fluoroelastomer.
Housing Materials: ABS plastic. Accuracy: ±1.0% F.S. (Includes linearity, hysteresis,
repeatability). Pressure Limits: 2X pressure range. Vacuum range max. pressure is 30 psig.
Temperature Limits: 30 to 120°F (-1 to 49°C). Thermal Effect: 0.05% FS/°F. Size: 2.62˝ O.D. x 1.52˝ deep. Process Connections: 1/4˝ male NPT. Display: 4-digit LCD (.425˝ H x .234˝ W digits). Power Requirements: 9 volt alkaline battery. Battery included
but not connected.
Auto Shut-off: 20 minute auto shut-off. Weight: 5.6 oz (160 g).
INSTALLATION
MAINTENANCE
Battery Removal: Remove the screw on backplate. Lift backplate off by hand. To reassemble replace backplate and screw.
Replace batteries per polarity indicators.
A “LOW BAT” descriptor indication will appear on the display when batteries need to be replaced.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. Box 373 • Michigan City, IN 46361-0373, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
DPGA & DPGW BUTTON OPERATION
To Turn On & Off:
Press to turn unit on. Press again to turn unit off.
To Zero Display:
To zero display, disconnect gage from any pressure source. Press and hold zero button until the LCD displays
“- - -”, then release.
To Change Units:
To change units, disconnect gage from any pressure source. Press both & buttons simultaneously until desired units are displayed.
DPGA Models
Model DPGA-00 DPGA-01 DPGA-02 DPGA-03 DPGA-04 DPGA-05 DPGA-06 DPGA-07 DPGA-08 DPGA-09 DPGA-10 DPGA-11
DPGW Models
Model DPGW-00 DPGW-00C DPGW-04 DPGW-05 DPGW-06 DPGW-07 DPGW-08 DPGW-09 DPGW-10 DPGW-11
Note: Gages with a “C” in the range have a ceramic sensor.
Range
30˝ Hg to 0 (vac) 0 to 20˝ w.c. 0 to 1 psi 0 to 2 psi 0 to 5 psi 0 to 15 psi 0 to 30 psi 0 to 50 psi 0 to 100 psi 0 to 200 psi 0 to 300 psi 0 to 500 psi
Range
30˝ Hg to 0 (vac) 30˝ Hg to 0 (vac) 0 to 5 psi 0 to 15 psi 0 to 30 psi 0 to 50 psi 0 to 100 psi 0 to 200 psi 0 to 300 psi 0 to 500 psi
psi
-14.70
0.722
1.000
2.000
5.000
15.00
30.00
50.00
100.0
200.0
300.0
500.0
psi
-14.70
-14.70
5.000
15.00
30.00
50.00
100.0
200.0
300.0
500.0
kg/cm
-1.033 .0508 .0703 .1406 .3515
1.055
2.109
3.515
7.03
14.06
21.09
35.15
kg/cm
-1.033
-1.033 .3515
1.055
2.109
3.515
7.03
14.06
21.09
35.15
2
2
bar
-1.013 .0498 .0689 .1379 .3447
1.034
2.068
3.447
6.89
13.79
20.68
34.47
bar
-1.013
-1.013 .3447
1.034
2.068
3.447
6.89
13.79
20.68
34.47
in Hg
-29.93
1.471
2.036
4.072
10.18
30.54
61.1
101.8
203.6
407.2 611 1018
in Hg
-29.93
-29.93
10.18
30.54
61.1
101.8
203.6
407.2 611 1018
ft wc
-33.94
1.667
2.307
4.614
11.53
34.60
69.2
115.3
230.7
461.3 692 1153
ft wc
-33.94
-33.94
11.53
34.60
69.2
115.3
230.7
461.3 692 1153
Pressure Ranges
kPa
oz/in
-235.2
-101.4
11.55
4.980
16.00
6.89
32.00
13.79
80.0
34.47
240.0
103.4
480.0
206.8 800
344.7 1600
689
3200
1379
4800
2068 3447
Pressure Ranges
kPa
oz/in
-235.2
-101.4
-235.2
-101.4
80.0
34.47
240.0
103.4
480.0
206.8 800
344.7 1600
689
3200
1379
4800
2068 3447
2
2
in wc
-407.3
20.00
27.68
55.4
138.4
415.2 830 1384 2768
in wc
-407.3
-407.3
138.4
415.2 830 1384 2768
mbar
-1013
49.80
68.9
137.9
344.7 1034 2068 3447
mbar
-1013
-1013
344.7 1034 2068 3447
cm wc
-1034
50.8
70.3
140.6
351.5 1055 2109 3515
cm wc
-1034
-1034
351.5 1055 2109 3515
mm Hg
-761
37.37
51.7
103.4
258.6 776 1551 2586
mm Hg
-761
-761
258.6 776 1551 2586
Resolution psi
0.01
0.001
0.001
0.001
0.002
0.01
0.01
0.02
0.1
0.1
0.1
0.2
Resolution psi
0.01
0.01
0.002
0.01
0.01
0.02
0.1
0.1
0.1
0.2